AquaUNO TM en 01 12 05
AquaUNO TM en 01 12 05
AquaUNO TM en 01 12 05
AquaUNO
Technical Manual
Software Version: 2.03
Edition: 1/12.05
0123
Manual changes Manual changes will be released as new editions and supplements. In
general - this manual is subject to change without notice.
Purpose This Technical Manual is intended for first studies (to acquire a basic
knowledge) and for reference purposes (for start-up, TSC, maintenance
and repair). This Technical Manual does not replace the training cours-
es offered by the manufacturer.
Specifications For the specifications of the respective system, refer to the latest Oper-
ating Instructions.
Circuit diagrams and The identification on the PCB permits the operator/technician to verify if
component layouts the circuit diagram/component layout matches the PCB actually in-
stalled in the system.
Note
Informs the operator that in case of a failure to follow the steps as de-
scribed, a specific function will be executed incorrectly or will not be ex-
ecuted at all, or will not produce the desired effect.
Caution
Advises the operator against certain procedures or actions that could
cause damage to the equipment or may have adverse effects on oper-
ators and patients.
Precautions When working on the open system, the following precautions must be
respected:
– Protect the components against ingress of fluids.
– Do not touch live parts (e. g. connectors of the power cord or heater).
ESD precautions When repairing and when replacing spare parts, observe the applicable
ESD precautions (e.g. EN 100 015-1).
Local service
1 System Description
1.1 Intended Use .............................................................................................................................. 1-1
1.2 Warnings and Safety Precautions ........................................................................................... 1-2
1.2.1 Operating Conditions................................................................................................................... 1-2
1.2.2 Patient Safety .............................................................................................................................. 1-3
1.2.3 Operator Safety ........................................................................................................................... 1-4
1.2.4 Protection of Buildings................................................................................................................. 1-4
1.2.5 Service Procedures ..................................................................................................................... 1-4
1.2.6 Residual Risks............................................................................................................................. 1-5
1.3 Physical Description ................................................................................................................. 1-6
1.3.1 Front View ................................................................................................................................... 1-6
1.3.2 Rear View .................................................................................................................................... 1-7
1.3.3 Water Inlet Block / Ports .............................................................................................................. 1-7
1.3.4 Interior Left .................................................................................................................................. 1-8
1.4 System Description ................................................................................................................... 1-9
1.4.1 General Notes ............................................................................................................................. 1-9
1.4.2 Physical Background ................................................................................................................. 1-10
1.4.3 Flow Diagram ............................................................................................................................ 1-11
1.4.4 Functional Description ............................................................................................................... 1-12
1.4.5 Overview / Description of Components ..................................................................................... 1-13
1.4.6 Sensors ..................................................................................................................................... 1-14
1.5 Installation and Start-Up ......................................................................................................... 1-15
1.5.1 Installation Requirements .......................................................................................................... 1-15
1.5.2 Start-Up ..................................................................................................................................... 1-18
1.6 Maintenance............................................................................................................................. 1-19
1.6.1 General Notes ........................................................................................................................... 1-19
1.6.2 Overview of Maintenance Procedures....................................................................................... 1-19
1.6.3 Replacing the Optional Filter Element on the Prefilter............................................................... 1-20
1.7 Microbiological Test / Permeate Samples for Determination of CFUs ............................... 1-21
1.8 Optional Accessories.............................................................................................................. 1-23
1.9 Specifications .......................................................................................................................... 1-24
1.9.1 Machine Data ............................................................................................................................ 1-24
1.9.2 Operating Conditions................................................................................................................. 1-25
1.9.3 Electrical Connection................................................................................................................. 1-25
1.9.4 Disposal..................................................................................................................................... 1-26
1.10 Abbreviations and Symbols ................................................................................................... 1-27
1.10.1 AquaUNO Labelling .................................................................................................................. 1-27
1.10.2 AquaUNO Label Types ............................................................................................................. 1-30
1.11 Storage and Transportation ................................................................................................... 1-31
1.12 Certificates ............................................................................................................................... 1-32
1.13 Bibliography............................................................................................................................. 1-36
8 Spare Parts
1 System Description
Caution
Operation, maintenance and repair may only be performed by trained
personnel who have the required technical knowledge and have been
instructed on the risks of a maloperation.
Caution
Disconnect the power plug before opening the unit.
Caution
The following applies to cleaning and disinfection:
Cleaning agents and disinfectants present a hazard for the patient. Prior
to cleaning and disinfection the connected machine must be
disconnected!
Caution
Appropriate tests must be made to check the unit and the ring main for
residual disinfectant.
When using Puristeril® / Puristeril plus® presence of residual
disinfectant must be checked with the peracetic acid test.
When handling the disinfectant upon disinfection of the unit, care must
be taken not to spill any disinfectant concentrate. To prevent acid burns
these fluids must always be handled with care:
– Use rubber gloves to avoid contact of the disinfectant with the skin.
– Do not remove the disinfection connectors while a disinfection is in
progress!
Caution
To prevent any skin contact with the disinfectant it is recommended to
wear rubber gloves (cotton-lined acrylonitrile latex).
When working with acidic substances:
Wear goggles! Carefully read the safety data sheet or the instructions
for use of the disinfectant used!
In the event of contact with acid:
Eye: Immediately flush with flowing water for 15 minutes.
Skin: Use soap under flowing water for neutralization.
Ingestion: Do not induce vomiting, but have victim drink plenty of still
water.
ESD precautions The applicable ESD precautions (EN 1000 1.5/1) must be observed in
the event of all technical interventions in the system.
Spare parts Any installation, modification or repair work requiring the system to be
opened may only be performed by personnel authorized by the
manufacturer and are permitted only when using original spare parts.
Varying (poorer) feed It is imperative that the required feed water quality specified in the
water quality chapter Operating Conditions (see Chapter 1.9.2 , page 1-25) is always
ensured.
Residual disinfectant or The AquaUNO must be thoroughly checked for residual disinfectant
not/ incorrectly performed after each disinfection (see Chapter 3.4.4 , page 3-9).
test for residual
If residual disinfectant is detected, rinse manually until absence of
disinfectant
disinfectant is ensured.
Adding of disinfectant in No disinfectant or cleaning agent may be added to the AquaUNO while
the dialysis mode the unit is in the dialysis mode.
Disinfection: permeate Prior to a disinfection the permeate tubing must be disconnected from
tubing not disconnected the connected machine.
1 2
4 3
Legend:
3 2 1
2
1
1 2 3 4
8 7 6 5
1 Disinfection block
2 Electronics control
3 Display unit
4 Feed water inlet valve V1
5 Water inlet block with feed water and permeate conductivity cell
6 Pump (with motor cooling jacket)
7 Module tube
8 Break tank with level sensor
Diffusion Solutes in a fluid move from the area of higher concentration to the area of lower
concentration.
Osmosis The solvent moves from the area of lower concentration to the area of
higher concentration until equilibrium is reached. The fluid level in the
compartment of the solution with the higher concentration rises. The
difference between these levels corresponds to a static pressure, which
is called osmotic pressure.
Feed water
Permeate
Permeate feed
Concentrate
D4 Inflow restrictor
B1 Break tank
F1 Membrane
The feed water flows across the feed water conductivity cell LFRW, the
feed water inlet valve V1 and the inflow restrictor D4 into the break tank
B1. The water level is monitored by the level sensors. It is distinguished
between 3 levels:
Level 1: Run-dry protection
Level 2: Tank center
Level 3: Tank full
The water is pumped by the rotary slide vane pump P to the reverse
osmosis membrane where the osmosis takes place: The water is
separated into permeate and concentrate.
The permeate produced flows via the permeate conductivity sensor
LFPerm to the dialysis machine. The permeate which is not required,
flows via the check valve VR1 back into the break tank.
After the reverse osmosis module the concentrate flows across the
concentrate drain control valve D2 and is then divided into two streams.
One part is returned via the concentrate return control valve D3 into the
break tank. The other part flows across the motor cooling jacket of the
pump drive B2 to cool the motor and is then diverted to the drain.
During a cleaning or a disinfection procedure, the cleaning agent or
disinfectant is added through the disinfectant connector to the break
tank. A cleaning or disinfection process is then performed automatically.
Component Description
D1 – Water inlet control valve (optional: Water inlet pressures up to a maximum of 6 bar can be regulated
feed water inlet control valve option) down to 2 bar.
Note
If inlet pressures exceed 6 bar, consult the manufacturer.
V1 – Feed water inlet valve The feed water inlet valve opens until the program is started.
D2 – Concentrate drain control valve The concentrate drain control valve determines the concentrate
pressure and thus the permeate volume.
D3 – Concentrate return control valve Due to the concentrate return control valve a certain percentage
of the concentrate is directed to the housing of the pump drive
where it is used to cool the motor.
D4 – Inflow restrictor The inflow restrictor D4 prevents overfilling of the break tank due
to the delayed closing time of V1 and minimizes noise.
P1 – Rotary slide vane pump It is required to achieve the required pressure for the reverse
osmosis unit.
B1 – Break tank The break tank stores the feed water that flows into the tank, the
concentrate returned and the permeate which is not required. It is
also used as a storage tank for the disinfectant and cleaning
agent.
B2 – Pump motor cooling jacket Part of the concentrate is passed through the pump drive motor
cooling jacket. This prevents excessive warming of the motor and
reduces noise.
VR1 – Permeate check valve The permeate check valve VR1 prevents the feed water from
flowing to the dialysis machine. Excess permeate is returned into
the break tank via VR1.
VR2 – Concentrate check valve The concentrate check valve VR2 prevents the concentrate from
flowing back into the break tank.
1.4.6 Sensors
Component Description
LFRW – Feed water conductivity sensor The feed water conductivity sensor LFRW is used to measure the
conductivity of the feed water flowing into the tank. This
information is displayed in the information menu.
The rejection rate is calculated from the feed water conductivity
and the permeate conductivity.
LFPERM – Permeate conductivity sensor The permeate conductivity sensor LFPERM is used to measure the
conductivity of the permeate. This information is displayed in the
information menu.
The rejection rate is calculated from the permeate conductivity
and the feed water conductivity.
NIV1 / NIV2 / NIV3 – Break tank level The level sensors NIV1 to NIV3 are used to monitor the fill level
sensors in the break tank.
NIV1: Switching level to protect from running dry.
NIV2 / NIV3: Control the fill level in the break tank.
NIV2 opens valve V1 and NIV3 closes valve V1.
Caution
The single-station reverse osmosis unit AquaUNO may only be
operated under the conditions specified in the following:
If the quality of the feed water is insufficient, the required water
pretreatment components must be installed after consulting the
Fresenius project planning department. In this event it is necessary to
perform an analysis of the local feed water.
Note
Installation of the water pretreatment system must be completed before
the reverse osmosis unit can be put into use. Planning and
implementation are offered by the Fresenius Wassertechnik.
Note
The unit may only be rinsed with permeate if the jumper 19-22 in the
electronics control has been set.
For further information refer to the Technical Manual.
Part no.: 631 276 1 Prefilter (complete filter unit with filter
element)
5 µm - 5´´
– The feed water inlet pressure at the AquaUNO is 1.5 - 5 bar. If the
feed water inlet pressure is 4 bar or higher, the feed water inlet
control valve option is required.
– A softener is installed if the water hardness is > 25°dH.
– All tubing and piping connections have been checked for leakage.
– A drain connection for concentrate diversion to drain DN 6 is
installed
– A drain pipe is installed.
Caution
Make sure that the drain line is never placed directly into the drain. A
drain pipe in DN 40 with a free fall of 20 – 30 mm to the drain must be
present to prevent back-contamination.
The drain of the unit may not be connected with the drain of the dialysis
machine.
Protective ground:
– When using class I devices, the quality of the protective ground of
the installation is of particular importance. The national
specifications defined by foreign authorities must be considered.
Potential equalization:
– The potential equalization must be connected as prescribed by the
national regulations (e.g. in Germany VDE 0100-710, edition
11.2002).
1.5.2 Start-Up
Caution
Start-up of the system may only be performed by trained personnel.
Condition
Start-up instructions
Note
Trapped air can be indicated by the error message
CD Inlet = 0µS !
In this case the unit should be disconnected from the power supply, be
reconnected and the procedure should be restarted.
1.6 Maintenance
Note
In accordance with the description for cleaning the modules (see
Chapter 3.3 , page 3-4) it is recommended to clean the unit using
Citrosteril® when the cleaning prompt is displayed.
Rinse the unit Run a rinse program. Units for emergency Do not use permeate
use: every day for rinsing!
Note
Replacement of the filter must be entered in the Medical Device
Register. The Medical Device Register is enclosed with the AquaUNO.
Requirements
Caution
Do not use the unit unless it is in a safe and proper condition from the
microbiological point of view!
Caution
Samples may only be collected by persons who have been instructed to
this effect and using aseptic technique.
When collecting a sample it must be ensured the bacteriological
laboratory will receive the sample:
– within 24 hours after taking the sample
– on a workday
Limits
Colony forming units Endotoxin concentration
Note
It is recommended to perform a microbial analysis of the permeate
produced every 3-6 months.
Determination of endotoxins:
Note
It is recomended to determine endotoxins every 3-6 months, however
at the latest after excessively high CFU values have been detected.
Sample collection site The sample is collected at the permeate outlet of the AquaUNO which
can be detached with the collection adapter to take the sample. The
permeate collection adapter is included in the MQS shipping box.
Methods of analysis
Note
The determination of CFUs and endotoxins and the procedure for
shipping the samples is done by the competent laboratory in
compliance with the applicable requirements and directives.
631 278 1 Feed water inlet control With water inlet pressure >4 bar, max. 6 bar
valve
Note
If inlet pressures exceed 6 bar, consult the
manufacturer for an appropriate design.
505 262 1 4008 connection Adapter to connect the AquaUNO to the 4008
dialysis machine
1.9 Specifications
Note
The permeate volume depends on the temperature of the feed water,
the operating differential pressure, the condition of the module used (silt
density).
Efficiency – Up to 50 %
Dimensions
Width 515 mm
Height 455 mm
Depth 200 mm
(without connectors)
Weight Approx. 30 kg
Ambient temperature 5 – 35 °C
Max. inflow pressure 6 bar (optional water inlet control valve D1, pressures up to a maximum
of 6 bar can be regulated down to 2 bar).
Caution
The design of the water treatment must ensure that the following
parameters are fulfilled:
ph 6–8
SDI <3
Water pretreatment Pretreatment of the water is determined after previous water analysis.
Line voltage As specified on the type label (see Chapter 1.10 , page 1-27)
1.9.4 Disposal
Type label
1
2
3
4
11 10 9 8 7
1 Manufacturer's logo
2 Machine type
3 Power connection
4 Power consumption
5 Serial number
6 Manufacturer's address
7 CE mark
8 Disposal information
9 Caution: Consult accompanying documents
10 Type of protection: Drip-water proof equipment
11 Degree of protection / type B
CE mark:
0123
Caution symbol:
Safety class I: The unit is provided with a grounding conductor terminal to prevent high
contact voltages.
Year of manufacture: The year of manufacture is included in the serial number printed on the
type label:
– e.g. serial number 249 S 0001
The first digit indicates the year of production, in the above example:
year of manufacture 2002.
CAUTION!
Make sure that no cleaning agents or disinfectants are injected into the
AquaUNO during dialysis mode.
The abbreviations and symbols on the type label are described on the
example of a AquaUNO as follows:
230 V model
115 V model
Note
Protect the machine from frost!
Transportation
Note
Keep upright, handle with care!
Do not push or pull the AquaUNO across uneven surfaces!
Storage
Note
Store upright!
1.12 Certificates
1.13 Bibliography
Available literature More literature on the AquaUNO is available in the latest edition.
– Operating Instructions and Medical Device Register
– Technical Manual
1 2 3
STANDBY
Select: On
The unit does not have a separate ON/OFF switch. When the power
plug is disconnected, the unit is not supplied with current. When
reconnecting the power plug, the unit will again switch to the STANDBY
mode.
– The key switches between the available program modes in the
lower line of the display.
– Pressing the ↵ key will start the selected program.
Depending on the operating mode of the unit, only one part of the
options which can be selected is available. When the unit is in the
operating mode Cleaning, for example, only the options Test and
Service are available.
2.3.1 Overview
Disinfection Preparing the unit for disinfection and running the DISINF. - Soaking
disinfection
Select: ---------
Service - Set Rinsing Selecting the level (index) for automatic rinsing; 4 | = Set |
e.g. level 4: when the display has been idle for 6
6h/30m | ↵ = Store |
hours, the system will be rinsed for 30 minutes.
STANDBY - NO WATER The water supply to the unit has been interrupted
ON - CDP > 100µS The permeate conductivity is between 100 and 150
µS/cm
CLEANING- NO WATER The water supply to the unit has been interrupted CLEANING
DISINF. - NO WATER The water supply to the unit has been interrupted DISINFECTION
FAULT DRAIN The break tank is not being drained (not enough
water flowing from the tank)
Switching Operating modes are activated by pressing the Enter key for an
appropriate length of time.
Overview – Switching from the Dialysis mode to the Standby mode (for another
dialysis treatment, cleaning or disinfection)
– Switching from the Standby mode to the Dialysis, Cleaning or
Disinfection mode
2.4 Standby
Program structure
STANDBY
Select: On
STANDBY
Select: Cleaning
STANDBY
Select: Disinfection
STANDBY
Select: Man. Rinsing
STANDBY
Select: Test
STANDBY
Select: Service
STANDBY mode When the unit is in the STANDBY mode, it is in a no-program state in
which the following activities can be performed:
– Starting dialysis
– Cleaning
– Disinfection
– Man. Rinsing
– Test
– Service
Standby mode
STANDBY ON
Select: On 2 seconds Select: Off
Setting the unit to the If the unit is in the Dialysis mode, press the key until the following text
Standby mode displays:
ON
Select: Off
– Press and hold the ↵ key for 2 seconds. The following message will
be displayed:
STANDBY
Select: On
Program structure When the unit is in the Dialysis ON mode, it provides permeate for
dialysis. The following activities can be performed:
Dialysis mode
STANDBY ON
Select: On 2 seconds Select: Off
On
Select: Test
ON
Select: Service
Selection OFF: Terminating dialysis, setting the unit to the Standby mode
TEST: Displaying operating data (see Chapter 2.9.2 , page 2-16)
SERVICE: Displaying the Service menu (see Chapter 2.9.3 ,
page 2-18)
Activating the dialysis – If the unit is in the STANDBY mode, press the key until the
mode following text displays:
STANDBY
Select: On
– Press and hold the ↵ key for 2 seconds. The following message will
be displayed:
SELF-TEST
Please wait
Note
There is no need to interrupt a dialysis treatment if the following
message is displayed:
ON -START CLEANING
Function During the T1 test, functions which are relevant to safety will be
checked. The T1 test is considered to be passed if all test criteria have
been fulfilled. After successful completion of the T1 test (duration:
approx. 2 minutes), the ON program will be started.
Phases of the self-test – The pump is turned on and the break tank is drained down to the
lower level.
– The pump is turned off. The break tank is now being filled with the
feed water inlet valve open, up to the upper level.
– The feed water inlet valve is closed. The pump is turned on. The
break tank is now being drained.
– Fill and drain cycles are performed until the permeate conductivity
has dropped below 150µS/cm. The permeate and the feed water
conductivity are also checked. The conductivity must be above 0 µS
(parting of a cable!). Otherwise an error message will be displayed.
ERROR CDP-SENSOR
CD Permeate = 0µS !
or
ERROR CDI-SENSOR
CD Inlet = 0µS !
ERROR CDP-SENSOR
CDP > 150µS !
ON (permeate supply) The self-test (T1 test) is performed automatically after starting ON and
must be successfully completed to be able to continue with the ON
program.
Switching the Dialysis With Select: Off, the dialysis mode can be switched off using the ↵ key.
mode off
ON
Select: Off
If the T1 test was not passed, the unit will be shut down and an
appropriate message will be displayed.
Potential display
messages Display message Test criteria
Note
In the event of an error during the T1 test:
To acknowledge the alarm, the power plug must be disconnected!
Manual rinse
STANDBY
Select: On
STANDBY
Select: Cleaning
STANDBY
Select: Disinfection
Man. Rinsing
Select: Test
Man. Rinsing
Select: Service
Note
The rinse time depends on the programmed level!
Starting the rinse program – Press the Selection key until the following message is displayed:
STANDBY
Select: Man. Rinsing
MANUAL RINSING ON
Select: -----------
Pressing the Enter key ↵ will return the unit to the mode
STANDBY
Select: On
Caution
Unattended flow of water:
To prevent damage caused by water when the rinse process runs
unattended, the manufacturer recommends to install a leakage detector
including a shut-off device in the affected area!
Entry The level (index) is entered in the menu using the Selection key and
the Enter key ↵.
Select: Service ->
Service: Diagnostic ->
Service: Set Rinsing
The levels in the following table show how long the unit will wait
between dialysis treatments before it starts rinsing and how long the
rinsing process lasts:
0 deactivated 5 minutes
(manual rinse)
1 1h 10 minutes
2 2h 15 minutes
3 4h 30 minutes
4 6h
5 12 h
Note
Settings may only be made by trained personnel who have the required
technical knowledge and have been instructed on the risks of a
maloperation.
Operation – Press the Selection key until the following message is displayed:
STANDBY
Select: TEST
Flushing: ##:##:## Time (hours, minutes, seconds) until the start of the automatic rinse or remaining
rinse time while rinsing.
– This information can only be displayed during the automatic rinse.
TEST Sequence
STANDBY
Select: On
STANDBY
Select: Cleaning
STANDBY
Select: Disinfection
STANDBY
Select: Man. Rinsing
STANDBY STANDBY
Select: TEST CD Inlet = xxx µS !
STANDBY
CD permeate = xxx µS !
STANDBY
Operation: xxx h
s s s
STANDBY
Select: Service
STANDBY
Service: Diagnostic
STANDBY
Service: Language
STANDBY
Service: Version No.
STANDBY
Service: Set Rinsing
STANDBY
Service: Set ExtCtrl
STANDBY
Select: Service
s s
STANDBY
Service: Version No.
STANDBY
Service: Set Rinsing
STANDBY
Service: Set ExtCtrl
Operation – Press the Selection key until S>1 displays on the lower left. Then
exit the menu by pressing the Confirm/Enter key ↵.
STANDBY
Select: Service
STANDBY
Service: Diagnostic
s s
STANDBY etc.
Service: Set ExtCtrl
Operation – Press the Selection key until the following message is displayed:
STANDBY
Service: Language
The menu option Version No. is provided for displaying the software
version installed in the system.
Version number - overview
STANDBY
Select: Service
STANDBY
Service: Diagnostic
STANDBY
Service: Language
STANDBY
Service: Set Rinsing
STANDBY
Service: Set ExtCtrl
Operation – Press the Selection key until the following message is displayed:
STANDBY
Service: Version No.
The menu option Set Rinsing is used to program the rinse volumes and
the time intervals to the next automatic rinse. (see Chapter 2.6.2 ,
page 2-13)
Set rinsing - overview
STANDBY
Select: Service
STANDBY
Service: Diagnostic
STANDBY
Service: Language
STANDBY
Service: Version No.
2 | s =Set Standby
2h/20min| =Store | 5 5 sec. Select: On
etc.
Operation – Press the Selection key until the following message is displayed:
STANDBY
Service: Set Rinsing
Note
The software provides the option for the connection of Fresenius
Medical Care dialysis systems of the 4008 series. See Technical
Manual.
When the AquaUNO is supplied, the external control is set to inactive.
The FME interface option is currently not supported.
STANDBY
Select: Service
STANDBY
Service: Diagnostic
STANDBY
Service: Language
STANDBY
Service: Version No.
STANDBY
Service: Set Rinsing
s s
Note
The unit may only be cleaned and disinfected by persons who have
been instructed on the proper handling of the unit.
Preservation of the unit is only permitted after consultation with the
manufacturer of the unit.
Caution
The following applies to cleaning and disinfection:
Cleaning agents and disinfectants present a hazard for the patient. Prior
to cleaning and disinfection the unit must be disconnected from the
dialysis machine!
Caution
When using cleaning agents / disinfectants, observe the manufacturer's
instructions for use.
When using cleaning agents / disinfectants, observe:
– that the cleaning agent / disinfectant matches the respective
cleaning and disinfection program.
– that different cleaning agents and disinfectants are not mixed.
Improper use of cleaning agents and disinfectants (regarding
concentration, temperature range, dwell time) may damage the system.
Note
Leakage caused by damaged seals:
– All couplings must be carefully checked for salt crystals before
connecting them to the matching counter-couplings.
– Remove the salt crystals by rinsing under flowing water and clean
the connectors with commercially available spray disinfectant.
Cleaning agents
Note
– Do not use acetone-containing cleaning agents on a solvent base!
– Never use solvents, diluting agents or chemical cleaning sprays!
– Do not use rough cleaning tools (e.g. scourer, etc.) to clean the unit.
The manufacturer recommends ClearSurfTM to clean the surface of the
AquaUNO.
Note
When using ClearSurfTM:
– Proceed in accordance with the instructions of the manufacturer of
the cleaning agent to clean the surface.
– The manufacturer does not assume any liability for potential damage
to the surface if a different cleaning agent than the one
recommended is used.
Caution
The following must be considered when cleaning and disinfecting the
modules:
Cleaning agents and disinfectants present a hazard for the patient. Prior
to cleaning and disinfection the unit must be disconnected from the
dialysis machine!
Cleaning and disinfection must be performed by personnel who have
been trained in accordance with the latest cleaning and disinfection
instructions.
Note
In the event of a power failure, the remaining time is saved.
Agents for cleaning and The reverse osmosis module is cleaned with Citrosteril®. Approximately
removal of calcium 200 ml Citrosteril®are required to obtain a concentration of 5%.
precipitates / volume
ON
START CLEANING
Note
There is no need to interrupt a dialysis treatment if the following
message is displayed:
ON -START CLEANING
Preparing for cleaning 200 ml of Citrosteril® are required for Cleaning - Injection!
– Disconnect the permeate tube from the dialysis machine
– Provide a syringe
– Provide Citrosteril
Starting the cleaning – In the STANDBY mode, press the Selection key until the following
program text displays:
CLEANING
– Press the Enter key ↵ and keep it pressed for 4 seconds. The display
changes to
CLEANING – Injection!
CLEANING-end
Note
After the AquaUNO has been cleaned, compare the conductivity and
the pH of the permeate displayed in the information menu with the
values in the Settings report.
In the event of deviations from the recorded operating values, contact
the technical service.
Caution
The following must be considered when cleaning and disinfecting the
modules:
Cleaning agents and disinfectants present a hazard for the patient. Prior
to cleaning and disinfection the unit must be disconnected from the
dialysis machine!
Cleaning and disinfection must be performed by personnel who have
been trained in accordance with the cleaning and disinfection
instructions.
Caution
When handling disinfectants, care must be taken not to spill any
disinfectant concentrate. To prevent acid burns these fluids must
always be handled with care!
Cleaning agents / quantity 100 ml of Puristeril 340® / Puristeril plus® are required for
Disinfection - Injection!
Note
Cleaning agent and disinfectant:
Use Puristeril 340® / Puristeril plus® for disinfection.
Caution
If a water supply as specified by the applicable regulations can no
longer be ensured.
– e.g. after repair in the permeate circuit
– or in the event of bacterial growth, the single-station reverse osmosis
unit must be disinfected.
Disinfect the unit With – After the unit has been idle for more than
72 hours!
Puristeril 340®
– Should the microbiological test reveal an
or
elevated microbial count in the permeate
Puristeril plus® (see Chapter 1.7 , page 1-21), disinfect the
unit!
Start condition An automatic disinfection can only be performed with the Disinfection
program.
Preparing for disinfection The module units are disinfected with the AquaUNO.
100 ml Puristeril 340® or Puristeril plus® are required for
Disinfection - Injection!
– Remove the permeate coupling from the dialysis machine
– Provide a syringe,
– provide disinfectant (Puristeril 340® / Puristeril plus®),
Starting the disinfection In the STANDBY mode, press the Selection key until the following
text displays:
Select: Disinfection
– Press the Enter key ↵ and keep it pressed for 10 seconds; the
display changes to
DISINF. - Injection!
DISINFECTION - END
Caution
Appropriate tests must be made to check the AquaUNO for presence
of residual disinfectant.
When using Puristeril 340®/ Puristeril plus® presence of residual
disinfectant must be checked with the peracetic acid test:
Use new (white) indicator strips. Observe the "use by" date! If the test
shows a residual concentration of peracetic acid the rinse program must
be repeated until all residual disinfectant has been completely removed.
Final steps – After absence of residual disinfectant has been ensured, the
permeate tube can be reconnected to the dialysis machine.
3.5 Preservation
Note
Preservation of the unit is only permitted after consultation with the
manufacturer of the unit.
Preserving the unit For detailed information on preservation of the unit, refer to the
Technical Manual.
Preserved 3 months
Note
When storing the unit, it is recommended to visually inspect the unit for
any leaks every week.
Check the pH of the preservation solution once a month. If the pH
drops below < 3, the system must be rinsed clear and then re-
preserved. A permanent pH of < 2 may cause damage to the
membrane.
Maximum storage time in the preserved state: 3 months
If the storage time exceeds 3 months of if the system is to be moved in
preserved state, re-commission the system (rinse and disinfect with
Puristeril) and then re-preserve it.
Procedure
Note
Longer rinse times must be expected when the unit is returned to
service.
For more information on the duration of a rinse procedure and the type
of preservation solution to be used, refer to the Technical Manual.
After rinsing
Note
After rinsing, the unit must be disinfected followed by a check for
residual preservation solution.
Procedure: (see Chapter 3.4 , page 3-7).
4 Error Messages
In the event of malfunctions or if alarm limits are exceeded, the unit
must be restarted after checking the cause of the problem and
correcting the problem.
Procedure In the event of a defect of the control unit, it is imperative that the type
of problem (effect of the malfunction) is recorded before dismantling the
system. A system that has been dismantled can only be repaired
(irrespective of the warranty period) if a detailed description of the
problem is available.
Alert The operator is alerted by a display message and an audible alarm. The
display shows the error message and the prompt to acknowledge with
the Enter key ↵.
Overview The error messages listed in the table are described in the following:
FAULT DRAIN The break tank is not Check It must be ensured that
being drained (not concentrate the concentrate flow to
enough water flowing diversion to the drain is not
from the tank) drain! obstructed.
The solenoid valve
(inlet) has not been
opened during the last
15 minutes.
FAULT CDP MAX The conductivity of the Inform the During the first 15
permeate produced is technical minutes after the units
too high. service! are turned on,
monitoring of the
conductivity is
suppressed, after this
time the AquaUNO is
turned off if the alarm
limit of 150 µS/cm in
the permeate is
exceeded.
Wrong message: Level Self-test failure Inform the Incorrect break tank
technical measurement
Service required
service!
Spare parts Any installation, modification or repair work requiring the system to be
opened may only be performed by personnel authorized by the
manufacturer and are permitted only when using original spare parts.
Requirements The Technical Manual assumes that technical tools and measuring
equipment are available.
Precautions When servicing the system, protect all components from exposure to
fluid.
ESD precautions When repairing and when replacing spare parts, observe the applicable
ESD precautions (e.g. EN 100 015-1).
System type:
Incl. option:
The following inspections must be carried out at no greater than 24 months’ intervals. The
measurements may only be performed by skilled electricians with electrical, system related and
medical/technical knowledge.
1 Visual inspection
3 Sensor check
TSC 3.1 Permeate conductivity sensor Compare the values with the external measuring
instrument
LFPERM
(not temperature-compensated).
Allowable deviation ± 2 μS
Measurement value: ___________________ μS
TSC 3.2 Feed water conductivity sensor Compare the values with the external measuring
instrument
LFRW
(not temperature-compensated).
Deviation ± 50 μS /
Measurement value: ___________________ μS
4 Alarms
TSC 4.1 No water (run-dry protection) Stop the water supply; the AquaUNO switches
off when the lower level has been reached.
5 Function
6 Prefilter option
7 AquaStop option
Comments:
5.2 Description
Note
Disconnect the AquaUNO from the dialysis machine prior to the
measurement. Establish a direct connection between the feed water
inlet and the drain line.
Fig. C.2
Measuring circuit for measurement of the protective earth resistance
on devices disconnected from the power supply
– Measure the protective earth resistance at a measuring current of ≥
200 mA.
– The open-circuit voltage for this measurement has to be 4 … 24 V,
DC (in both polarities) or AC.
– The measured value is the resistance from the grounding contact of
the power line to defined conductive accessible housing parts of the
device which might become live parts in fault conditions.
– When performing the measurement, the power line has to be moved
segment by segment, in order to detect broken cables not
identifiable from the outside.
– During the protective conductor measurement, no other measuring
instrument must be connected to the system.
Measurement points
1 2 3
1. Protective conductor terminal, hood (on the side; not visible on the
above photo)
2. Potential equalization socket
3. Housing screw
Measurement points on
the open system
7. Pump
Fig. C.6
Additional requirements If the value scaled to the nominal voltage is higher than 90% of the
admissible alarm limit (= 450 µA), the last measured value or the first
measured value must additionally be considered for the rating.
If the device leakage current has considerably increased since the last
measurement or has continuously increased since the first
measurement (slow deterioration of the insulation), or if the sum
composed of the current value plus the difference since the last
measurement is > 500 µA, the measurement has not been passed.
1 631 064 1 Secutest S III VDE test device (without printer module)
Caution
The system may only be used in accordance with the accompanying
documents. Only then will the manufacturer consider himself liable for
the safety, reliability and performance of the system.
The start-up must be performed by the Technical Service of Fresenius
Medical Care or a person authorized by them.
In case of a first installation of the system, observe the Specifications!
State upon delivery When the unit is delivered, it is electrically and hydraulically calibrated.
Note
The system is supplied in wet condition.
– The system must be put into operation within the next few weeks
after delivery.
Caution
If the system is preserved, it must be rinsed for at least 30 minutes prior
to its first use!
The system must then be disinfected.
Connection requirements – The supply line must be provided with a stop valve.
– The electrical connection is easily accessible (the system does not
have a separate ON/OFF switch).
– The system must be installed in a place ensuring easy ventilation
and absence of condensate.
– It has been ensured that the relative humidity at a temperature of
20 °C does not exceed 90 %.
– It must be ensured that the ambient temperature is between +10 °C
and +35 °C.
– The system must be set up on a level floor.
– The castors are locked.
Note
Should the feed water fail to fulfill these values, preconnection of an
appropriate water pretreatment is required.
6.2 Start-Up
Hydraulic calibration
Note
All hydraulic settings and modifications may only be performed by an
authorized service technician.
Settings The settings (e.g. language, rinse cycle, etc.) desired by or required for
the customer must be programmed in the Settings menu. These values
must be recorded.
Note
Prior to its start-up, check the unit for damage caused during transport.
If there are signs of damage to the electric or hydraulic components, do
not turn the unit on.
The connectors of the system have been closed with plugs prior to
delivery.
Prior to start-up and connection, check all connectors and the tank
overflow, and remove all plugs.
Step 1 The operating conditions for the water supply of the system and the
required measures to improve the water quality are ensured:
Caution
Appropriate water pretreatment in accordance with the specified inlet
requirements is required.
Do not use the system unless it is in a safe and proper condition from
the microbiological point of view.
– From the Settings menu select and start the ON/ SUPPLY program.
– While the unit is in the ON / SUPPLY mode, collect permeate and
direct it to the drain. This should be done for a period of
20 – 30 minutes.
– In the meantime the permeate conductivity should be verified. The
rinse procedure is complete if a rejection rate of at least 95% has
been reached.
Calculation: 1 – (LFPERM : LFRW) = 0.95
– To improve efficiency of the rinse procedure and to vent the system,
the permeate outlet must be closed for several seconds and then be
opened again during the start-up.
– All tubings and connections must be checked for leakage.
Caution
A standard disinfection has always to be performed after a start-up.
Completion
Customer: Date:
Address: Operator:
City:
Phone: Person of contact:
AquaUNO no.:
AquaUNO version:
PERMEATE CONDUCTIVITY
Miscellaneous:
______________________________________________________________
______________________________________________________________
______________________________________________________________
The complete system has been disinfected/cleaned and checked for absence of disinfectant as
specified in the respective instructions.
Signature [Operator]:______________________
Rev. B / 11.2003
STANDBY
Select: Service
STANDBY
Service: Diagnostic
STANDBY
Service: Language
STANDBY
Service: Version No.
STANDBY
Service: Set Rinsing
STANDBY
Service: Set ExtCtrl
Display Status
3 tmp:** lf:000 mv:000 tmp = motor cooling jacket temperature (not active)
r:00000 s:000:00 S>0 lf = conductivity timer (suppressed at start)
MV = solenoid valve timer Time Out
r = cleaning timer
s = rinse timer
S = selection
Key Status
s = system status
Key Function
12 STANDBY
14 STANDBY - NO WATER
22 ON
28 ON - NO WATER
31 CLEANING- NO WATER
34 CLEANING Initialisation
36 CLEANING – Injection
38 CLEANING-Rinsing in
40 CLEANING- Soaking
42 CLEANING-Rinsing out
62 FAULT SUPPLY
63 FAULT DRAIN
Accessing the menu The language module is used to select another language. The
requested language is selected with the key and confirmed with the
Enter key ↵.
Press ↵ To confirm
Accessing the menu The sub-menu for the version number is used to retrieve the software
version installed: this facilitates responding to questions on site.
Note
There is no need to interrupt a dialysis treatment if the following
message is displayed:
ON -START CLEANING
The following intervals defined by the index list can be programmed for the rinse program:
0 deactivated 5 minutes
(manual rinse)
1 1h 10 minutes
2 2h 15 minutes
3 4h 30 minutes
4 6h 30 minutes
5 12 h 30 minutes
Note
From a microbiological point of view it is recommended to enter index
no. 4.
The values displayed after accessing the editor reflect the current
setting. After pressing the selection key the next allocation (cf. index
table) will be offered.
The preselected times will be saved by pressing and holding the Enter
key for 5 seconds. The display returns to the standard screen. If no key
is pressed for 1 minute, the screen will also be exited. In this case the
original setting will be retained. When setting the value to 0-disabled, no
rinse will be performed.
The external control option (digital input) switches the unit from the
STANDBY mode to ON and vice versa.
Description
As soon as Ext. Control ON has been selected and saved, the control
input is active, otherwise the input will be inactive. The use of the active
external control input generally permits switching from STANDBY<>ON
and vice versa, in addition to key-operated switching.
The respective subprograms STANDBY and ON are not changed, but
can be accessed via the External Control option.
The operator always has the possibility to switch via the external control
from the ON to the STANDBY mode, even if the ON command was
detected by the activated control contact.
It is, however, impossible to switch to the ON mode using the keyboard,
if the activated and external contact signals the command STANDBY.
Note
The disinfection and cleaning program can only be performed if the
activated external contact also signals STANDBY mode.
If the external control is activated, selection of the SUPPLY mode with
keyboard commands is disabled.
Control logic
Activation - deactivation The external control can be activated and deactivated. If the external
control function is deactivated, any change of the potential on the
control input will be without effect. The setting is saved.
Connection
Parameter Connection
P.C.B.
3.0 m 4 x 0.34²
32 Connector
33 gn SV40
32 ws
31 br
2 3
4
1
gn br ws
Control logic
Acknowledgement
Status Output
STANDBY irrelevant
ON clock pulse 1 Hz
Fault high-resistance
6.5 Calibrations
All units are factory-set (see: AquaUNO test report). The conductivity
cell must be re-calibrated if incorrect CD cell values have been detected
(drift) or the CD cell has been replaced.
Displaying conductivities
In the sub-menus the TEST menu for the calibration of the conductivity
which shows the CD values can be activated by pressing the key.
Step 3 – Press the Enter key ↵ to confirm the selection (CD inlet will be
displayed)
Step 5 – Compare the reference value with the value of the internal
conductivity meter.
Note
The temperature compensation of the reference meter must be
switched off.
Potentiometer P1 = CD inlet
Potentiometer P2 = CD permeate
Note
The internal conductivity meter for feed water can only be set in
increments of 5 µS.
Always measure the conductivity under flowing water. Measurements in
a cup or other receptacles will produce incorrect results.
Fig.: P.C.B.
Potentiometer 1 = CD inlet
Potentiometer 2 = CD permeate
Pump pressure
Note
All units are factory-set (see: AquaUNO test report).
A hydraulic calibration is not required.
Data for service and test purposes can be transferred via the serial
interface (see chapter 6.4.1, page -7).
The system will automatically send a data record at regular intervals
(transmission 1x per second; decimal code).
The following data can be – Current operating mode of the AquaUNO (program and alarm
recorded: status, measurement values, pump and valve state).
– Settings
– Counter content (operating hours; remaining cleaning time)
– Software version
6.6 Troubleshooting
The unit is turned on, but does not The inlet pressure is below 1 bar, If the pressure is insufficient, check
produce any permeate. Message the water supply is insufficient. the inflow of water and the prefilter.
displayed:
NO WATER
The unit is turned on, but does not The center contact of the level Open the break tank and manually
produce any permeate. Message sensor is defective. move the float slowly to the top,
displayed: check switching of the center
contact. If the contact fails to
NO WATER
switch, replace the level sensor.
The Diagnostic screen shows that
the lower contact of the level
sensor is switched.
The time until the start of the next The CD measurement on the inlet Compare the value of the CD
cleaning program approaches shows an incorrect value. Or the measurement at the inlet with the
faster than usual. quality of the feed water has reading of the conductivity meter. If
deteriorated. Time until the start of necessary, replace the conductivity
the next cleaning program as a sensor.
function of the CD inlet:
0-300 µS = 100 hours
300-500 µS = 50 hours
500 µS and more = 20 hours
Cooling jacket of the motor Insufficient concentrate flow across Check the concentrate nozzles for
becomes extremely hot. the motor cooling jacket. contamination.
Concentrate flow 70 l/h ± 10 l/h Replace the motor in case of
damage.
The permeate conductivity is above Module defective. If the idle time is relatively short,
150µS/cm check the concentrate flow and the
Insufficient concentrate flow.
cleaning times.
Concentrate flow 2 x 70l/h
Overflow of the break tank. The Level sensor defective Replace the level sensor.
Diagnostic screen shows that the
upper contact of the level sensor is
not switched.
Overflow of the break tank. The The inlet solenoid valve does not Check the connection to the power
Diagnostic screen shows that the close. supply;
level sensor is functional.
It is recommended to replace the
solenoid valve tappet.
Overflow of the break tank. The unit System error Use the Reset key to reset the unit
cannot or cannot be properly to a defined condition.
operated.
Permeate pressure too low Improper function of the permeate Replace the permeate pressure
pressure holding valve. holding valve.
Leakage at connection sites Leakage at plug connections. Check the plug connections,
replace if necessary.
Caution
The acceleration of time-dependent functions may only be used if
sufficient rinsing after abortion of a cleaning, rinse or disinfection
procedure has been ensured.
Description
Service keyboard in
normal mounting position
Selection key
Solenoid valve
Control voltage 24 V DC
Connection DN7
Membrane
Permeate flow and salt rejection are based on the following test
conditions: 2000 ppm NaCL, 15.51 bar, (25°C), pH8
Operational limits
pH (continuous operation) 2 – 11
Check valve
Connections Tubing, d 6 mm
Wattage 230 W
Weight 5,5 kg
Type 133B080F11GAXX
CD cell
Connection G ¼“
Level sensor
Material
Note
In the event of a replacement of the membrane, the new membrane
must be rinsed after installation.
The system must then be disinfected.
Caution
Fuses on the unit may only be changed by a trained technician.
SI 7 SI 6 SI 5 SI 3 SI 4
SI 1 SI 2
230 V / 50 Hz 115 V / 60 Hz
SI 5 1.25 AT 24 V 1,.25 AT 24 V
Note
When updating the software, reset the memory, if necessary, to ensure
proper operation of the reverse osmosis unit (see Chapter Resetting the
Memory for more information).
Observe the manufacturer's instructions regarding the need to reset the
memory when replacing the EPROM!
19 - 22
Functions
19 20 21 22
Automatic cleaning prompt active*
– Connect the power plug, wait until the signal is sounded, the
EPROM has been initialized, the operating hours meter has been
reset.
– Disconnect the power plug
– Remove the jumper (and connect the jumper to 19-20)
– Reconnect the power plug.
– Turn the system on
– The AquaUNO is ready for use.
Note
When the EPROM is initialized, the operating hours meter will be
erased and the programmed rinse cycle will be reset.
6.9 Preservation
Caution
When handling preservation solutions or a preserved system, observe
the respective safety precautions for the preservation solution.
Note
Preservation of the unit is only permitted after consultation with the
manufacturer of the unit.
Preservation of the unit must be performed by a trained technician.
Note
The AquaUNO (up to AC008) includes POM plug connectors. This
material can be damaged by sodium bisulphite.
Requirements
Note
It is recommended to perform a disinfection prior to each preservation
and then to check the unit for absence of disinfectant.
Note
Do not use packing units of sodium bisulphite for preservation which
have already been opened!
Any contact of the preservation solution with air or oxygen will inactivate
the efficiency of the preservation.
Note
Preservation with a 40% sodium bisulphite solution (sodium
hydrogen sulphite) is necessary to prevent clogging/bacterial growth in
the module unit when the AquaUNO is taken out of service for more
than 7 days.
Preparation AquaUNO:
A 2% preservation solution with reference to the system volume is
required.
Volume = approx. 2 liters (2000ml) for control unit + module.
Procedure Initially, preservation of the unit is carried out in the same way as a
standard disinfection.
– The AquaUNO must be disconnected from the distribution system.
The permeate feed end and the permeate return end must be short-
circuited.
– The permeate connection has been removed.
– Fill the injection syringe with 100 ml of sodium bisulphite (40%).
– Start the disinfection program. Remove the plug from the injection
port.
– Inject the preservation solution (sodium bisulphite solution) instead
of the disinfectant.
– Allow approx. 3 minutes for the preservation solution to be
circulated. After it has been circulated, i.e. during the acting time, the
AquaUNO must be turned off by disconnecting the power plug. This
will prevent the preservation solution being rinsed from the unit and
the preservation solution remains in the system.
– Disconnect the tubing and close the ports. Make sure that the
module does not run dry.
– Close and make air tight all hydraulic connections, including the
overflow to the break tank.
– The AquaUNO is now preserved and can be taken out of service.
Note
When storing the unit, it is recommended to visually inspect the unit for
any leaks every week.
Check the pH of the preservation solution once a month. If the pH
drops below < 3, the system must be rinsed clear and then re-
preserved. A permanent pH of < 2 may cause damage to the
membrane.
Maximum storage time in the preserved state: 3 months
If the storage time exceeds 3 months, or if the system is to be moved in
preserved state, re-commission the system (rinse and disinfect with
Puristeril) and then re-preserve it.
Caution
Rinse until no more sodium bisulphite can be detected at the permeate
outlet, in the drain and in the overflow. Rinsing may take several hours.
It is imperative that the test for residual disinfectant be performed both
on the permeate outlet, on the drain and the overflow of the AquaUNO.
Checking for residual – With its reaction zone, immerse the test strip in the water for approx.
sodium bisulphite 1 second. Shake off any excess fluid.
– After approx. 30 seconds, compare the reaction zone with the color
scale of the test tube.
– The reaction zone must stay white and must not show any signs of
red or pink coloring.
Performing a disinfection
Note
Always perform a disinfection after the preservation solution has been
rinsed from the unit and after a successful test for absence of residual
preservation solution.
Caution
When using Puristeril 340®/ Puristeril plus® presence of residual
disinfectant must be checked with the peracetic acid test:
6.10 Accessories
Options overview
631 278 1 Optional feed water inlet With water inlet pressure >4 bar (max. 6 bar)
control valve
505 262 1 Optional connection to 4008 To connect the AquaUNO to the 4008
The feed water inlet control valve is required for a feed water pressure
of > 4 bar. The feed water inlet control valve is installed directly in front
of the feed water inlet nozzle. The maximum allowable pressure for the
feed water inlet control valve is 6 bar.
Note
It is recommended to set the pressure to 3 bar.
AquaUNO 5“ prefilter
The filter holder of the prefilter is affixed to the bottom of the AquaUNO,
as shown. The filter outlet is connected directly to the feed water inlet
nozzle.
6.10.3 AquaStop
Note
The AquaStop option is not intended for use in humid environments.
AquaStop
Function In the event of a leakage (water sensor wet), the unit automatically
generates a warning signal and switches the solenoid valve off. The
alarm will also be activated when disconnecting the water sensor.
To clear the alarm, the control unit must be disconnected. Wait approx.
5 seconds before reconnection.
Instructions for operation Following a short-term power failure, the solenoid valve may close and
fail to open again. Disconnect the control unit and wait approx. 5
seconds before reconnecting it.
Note
Never operate the control unit without connected solenoid valve.
GE/GN
POWER
in
BR
BL
Line filter
SW
M
RT
RT
SW
BL
GN
WS
FME interface
BR
WS
BR
Water inlet valve
Service keyboard
Ext. control
AquaUNO
nc NIV1
1 2345
nc GE
GND GN NIV2 Sensor rod
GR NIV3
WS GND
WS
CD feed water
BR
WS CD permeate
DISPLAY
10 kW
– Power supply unit 24 V 25 VA for load, +5V and –5V for control
The printed circuit board houses the entire control for the AquaUNO
consisting of power supply unit, analog unit, microcontroller and
actuator control.
The core of the control unit, consisting of IC1, type SAB 80C535, the
address latch IC4, type 74HC573, and the program memory IC1, an
EPROM type 27C512, is connected via the multiplexed address/data
bus which is typical of controllers of the 8051 series. The resistor
networks SIL2, SIL3 and SIL4 are provided for interference
suppression.
The serial EPROM IC3, type 24C16 with a memory capacity of 16 384
bit is connected via the 12C bus of the microcontroller (pin 25 and 26)
for filing operating parameters and various counter contents, e.g.
operating hours; included is also a real-time clock chip IC 12, type
PCF8583, with an external quartz ((32,768kHz). A fine adjustment of
the time is possible with the trimmer capacitor C33.
Battery buffering with BAT1 ensures that the system time continues
even when the system is power off.
This diagram shows the connection of the quartz Q1 (12 MHz) to IC1.
The serial interface is provided by IC6, an RS 232 chip. The interface
chip converts the level of the RS 232 interface from TTL to ±10 V and
vice versa.
The actuators are controlled by port P1 of the controller via the buffer
IC7 and the optocouplers. The three upper outputs are designed for
solenoid valves. The channels for triggering the two relays (REL1 and
2) and the buzzer SU are located below these outputs. The output for
the customer-specific remote control is located at port P1.5. (see Page
1/8)
Port P5 of the controller is used for the digital inputs. The optocouplers
ensure electrical isolation between the digital unit and the +24V. The
input pin 63 of the controller is used as input for the customer-specific
remote control. (see Page 1/8)
Fixed voltage regulator V11 for the generation of +12V to control the
outputs. Illustration of the supply voltage connections of the individual
ICs and various decoupling capacitors.
8 Spare Parts
For a list of the currently available spare parts for the AquaUNO, refer
to the FME Water Treatment Spare Parts Catalog.