Nothing Special   »   [go: up one dir, main page]

Iol 003D

Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

SERVICE

BULLETIN

Instrument: IOLMaster No.: IOL-003D


Subject: IOLMaster Checklist / Validation Form Date: 27 Sep 10

Status: Mandatory Pages: 1

This bulletin replaces Service Bulletin IOL-003C

IOLMaster System Checkout Procedure:


The System Checkout Procedure (for IOL 1-3: Section 4.5 in the IOLMaster Service Instructions Manual
#SM-30-4039; for IOL XP 4-5: pages 11-12 in the IOLMaster XP Service Instructions Manual #SM-30-
4047; other reference is SM-30-4052) is used to verify that the instrument is operating properly within
specifications. The System Checkout Procedure should be performed at the completion of every
installation and service call on the instrument. If an error is detected during the System Checkout
Procedure, troubleshoot and repair the instrument; then start the System Checkout Procedure again
from the beginning. The entire System Checkout Procedure should always be performed from start to
finish without any problems occurring.

IOLMaster System Field Service Checklist:


The Field Service Checklist is a shortened version of the System Checkout Procedure and is used as a
guide for the Service Engineers that are familiar with the procedures. The Field Service Checklist is used
during completion of each IOLMaster System service call (including software upgrades).

An E-copy with writeable fields of the “IOLMaster Instrument Checklist and Maintenance Report”
(Validation Form) is located in Lotus Notes in the Service Bulletin folder under System Operation
Validation Forms.

IOLMaster System Operation Validation Form:


The System Operation Validation Form is a condensed sheet form that captures all of the pertinent
system operation validation information. All pertinent system operation validation information is recorded
on the System Operation Validation Form. The System Operation Validation Form must be filled out
during completion of each IOLMaster System service call (including software upgrades).

Attach the completed System Operation Validation Form in the “Attachments” Assignment Block of the
Service Confirmation.
• All IOLMaster Instruments up to Version 5.4.4 S/W Checklist –
(Checklist / Maintenance Report Version 5.1 can be used for all IOLMaster systems
1,2,3, and XP systems up to and including Version 5.4.4 software.)

• IOLMaster Instrument Version 5.5 S/W Checklist –


(Checklist / Maintenance Report Version 5.5 would be used only for IOLMaster systems
running the 5.5 software. These systems have IOLMaster 500 hardware, running current
version software.)

An E-copy with writeable fields of the “IOLMaster Instrument Checklist and Maintenance Report”
(Validation Form) is located in Lotus Notes in the Service Bulletin folder under System Operation
Validation Forms.

Note – The Validation Form is provided for U. S. Domestic Field Support Engineers only.

ÚÚÚ PROPRIETARY and CONFIDENTIAL ÚÚÚ