Connect SW
Connect SW
Connect SW
2019
Connect SW
Software version 5.1.x
Operator's Manual
English
Table of contents Dentsply Sirona
Operator's Manual Connect SW
Table of contents
1 Introduction ............................................................................................................. 6
1.1 Dear Customer,............................................................................................ 6
1.1.1 Contact data................................................................................... 6
1.2 Copyright and trademark ............................................................................. 7
66 97 549 D3534
2 D3534.208.18.02.02 08.2019
Dentsply Sirona Table of contents
Operator's Manual Connect SW
4 Configuration .......................................................................................................... 24
4.1 Parameters .................................................................................................. 24
4.2 Devices ........................................................................................................ 24
4.2.1 Scanner.......................................................................................... 25
4.2.1.1 Configuring the scanner ............................................... 25
4.2.1.2 Resetting scanner settings ........................................... 25
4.2.1.3 Calibrating the scanner................................................. 26
4.2.1.4 Color calibration............................................................ 29
4.2.1.5 Scanner heating settings (Omnicam) ........................... 32
4.2.1.6 Updating the firmware .................................................. 32
4.3 Settings ........................................................................................................ 32
4.3.1 ADA/FDI odontogram..................................................................... 32
4.3.2 Notifications ................................................................................... 33
4.3.3 Hub settings ................................................................................... 33
4.3.4 Patient database ............................................................................ 33
4.3.5 Language ....................................................................................... 33
4.3.6 Quality improvement agreement.................................................... 33
4.4 App Center (applications) ............................................................................ 34
66 97 549 D3534
D3534.208.18.02.02 08.2019 3
Table of contents Dentsply Sirona
Operator's Manual Connect SW
7 Page palette............................................................................................................ 43
7.1 View options................................................................................................. 43
7.2 Tools ............................................................................................................ 45
7.2.1 Buccal registration ......................................................................... 45
7.2.2 Cut out model areas....................................................................... 46
7.2.3 Resetting the model ....................................................................... 46
7.2.4 Trimming ........................................................................................ 47
7.2.5 Entering the preparation margin .................................................... 47
7.3 Display objects............................................................................................. 49
7.4 Activating analysis ....................................................................................... 51
9 ACQUISITION phase.............................................................................................. 56
9.1 Image catalogs............................................................................................. 56
9.1.1 Adding image catalogs................................................................... 57
9.2 3D Preview................................................................................................... 57
9.3 Take a scan ................................................................................................. 57
9.3.1 Mode .............................................................................................. 57
9.3.2 CEREC Omnicam, Omnicam......................................................... 58
9.3.2.1 Scanner warm-up time ................................................. 58
9.3.2.2 Scanner guide .............................................................. 59
9.3.2.3 Taking optical impressions with the scanner ................ 65
66 97 549 D3534
4 D3534.208.18.02.02 08.2019
Dentsply Sirona Table of contents
Operator's Manual Connect SW
Index ....................................................................................................................... 97
66 97 549 D3534
D3534.208.18.02.02 08.2019 5
1 Introduction Dentsply Sirona
1.1 Dear Customer, Operator's Manual Connect SW
1 Introduction
66 97 549 D3534
6 D3534.208.18.02.02 08.2019
Dentsply Sirona 1 Introduction
Operator's Manual Connect SW 1.2 Copyright and trademark
66 97 549 D3534
D3534.208.18.02.02 08.2019 7
2 General data Dentsply Sirona
2.1 Certification Operator's Manual Connect SW
2 General data
Please read this document completely and follow the instructions
exactly. You should always keep it within reach.
Original language of the present document: German
2.1 Certification
CE mark
This product bears the CE mark in accordance with the provisions of the
Council Directive 93/42/EEC of June 14, 1993 concerning medical
devices (MDD).
2.3 Accessories
In order to ensure product safety, this device may be operated only with
original Dentsply Sirona accessories or third-party accessories
expressly approved by Dentsply Sirona. The user is responsible for any
damage resulting from the use of non-approved accessories.
66 97 549 D3534
8 D3534.208.18.02.02 08.2019
Dentsply Sirona 2 General data
Operator's Manual Connect SW 2.3 Accessories
66 97 549 D3534
D3534.208.18.02.02 08.2019 9
2 General data Dentsply Sirona
2.4 Structure of the manual Operator's Manual Connect SW
2.3.2 Hub
Hub is a network-based data center for CAD/CAM data in the practice.
Hub stores data from the acquisition unit and enables data exchanges
between devices.
If a Hub is connected to the network, it is detected automatically.
The Hub logo appears in the context bar in the Connect SW.
The Connect SW must be set so that it can save data on the Hub. For
more information refer to the Hub Operator's Manual.
1. In the system menu in the "Configuration" area, under "Settings" the
menu "Patient Database".
Ä The "Database Settings" menu appears.
2. Activate the "Hub Patient Database" option.
Setup and commissioning of the Hub is described for users in the
Operator's Manual. For more information, refer to the Hub Service
Manual.
DANGER
An imminent danger that could result in serious bodily injury or death.
WARNING
A possibly dangerous situation that could result in serious bodily
injury or death.
CAUTION
A possibly dangerous situation that could result in slight bodily injury.
NOTE
A possibly harmful situation which could lead to damage of the
product or an object in its environment.
IMPORTANT
Application instructions and other important information.
66 97 549 D3534
10 D3534.208.18.02.02 08.2019
Dentsply Sirona 2 General data
Operator's Manual Connect SW 2.4 Structure of the manual
66 97 549 D3534
D3534.208.18.02.02 08.2019 11
2 General data Dentsply Sirona
2.4 Structure of the manual Operator's Manual Connect SW
Example Meaning
Tap Single tap on the screen using your finger.
To execute functions in the software you must tap
once on the corresponding button.
Double-tapping Two taps on the screen in rapid succession using
your finger.
Tip: To open programs in Windows you must tap
the corresponding button twice (double-click).
Call up shortcut Tap the corresponding point and hold the finger
menus on the screen for a longer period. A shortcut
menu opens at this point.
Drag & drop .
Tap an element (e.g. pictograph), drag and drop
onto new potential destination.
Item Function
A ➢ Complete a rotary movement using 2 fingers.
Ä The object is rotated in the plane.
B ➢ Drag with 1 finger.
Ä The model is rotated out of its current plane.
C ➢ Pull 2 fingers in together.
Ä The object is minimized.
D ➢ Pull the fingers apart.
Ä The object is maximized.
E ➢ Drag with 2 fingers.
Ä The model is dragged.
66 97 549 D3534
12 D3534.208.18.02.02 08.2019
Dentsply Sirona 2 General data
Operator's Manual Connect SW 2.4 Structure of the manual
66 97 549 D3534
D3534.208.18.02.02 08.2019 13
2 General data Dentsply Sirona
2.5 User interface Operator's Manual Connect SW
2.5.1 New features from the new acquisition unit and Connect
SW software
2.5.1.1 New screen format
With the new CEREC Primescan AC / Primescan AC / CEREC
Omnicam AC / Omnicam AC acquisition unit, you are provided with a
new screen resolution. The wide-screen format in 16:9 offers you more
space for an increased level of precision when working on models and
restorations. Tool windows can be positioned more generously, at the
same time the 3D preview is presented in a rather large format.
66 97 549 D3534
14 D3534.208.18.02.02 08.2019
Dentsply Sirona 2 General data
Operator's Manual Connect SW 2.5 User interface
Touch functionality
The Connect SW software is touch-compatible. This means that you
can interact directly on screen with your finger.
The application can be operated unconditionally by switching between
touch and using a touchpad or trackball.
A number of functions can only be executed by trackball. If this is the
case, this is indicated in the corresponding sections.
Item Function
A ➢ Complete a rotary movement using 2 fingers.
Ä The object is rotated in the plane.
B ➢ Drag with 1 finger.
Ä The model is rotated out of its current plane.
C ➢ Pull 2 fingers in together.
Ä The object is minimized.
D ➢ Pull the fingers apart.
Ä The object is maximized.
E ➢ Drag with 2 fingers.
Ä The model is dragged.
2.5.1.3 Touchpad
A touchpad is installed in the CEREC Primescan AC / Primescan AC /
CEREC Omnicam AC / Omnicam AC acquisition unit as standard. On
the flush-mounted horizontal surface, you can move a mouse cursor
with your finger on the screen. All the usual interactions of a standard
mouse are possible.
2.5.1.4 Trackball
Alternatively, a device model with an integrated trackball can be
ordered. This also enables use of a mouse cursor and represents an
alternative input option for the direct touch operation by finger on the
display.
66 97 549 D3534
D3534.208.18.02.02 08.2019 15
2 General data Dentsply Sirona
2.5 User interface Operator's Manual Connect SW
Phase bar
● ADMINISTRATION
● ACQUISITION
● MODEL
● CONNECT
2.5.2.1 ADMINISTRATION
In this phase, you can perform the following:
ADMINISTRATION ● Manage patients
● Create cases and determine their type,
● Define the tooth numbers and materials.
2.5.2.2 ACQUISITION
In this phase, you can perform the following:
● Create acquisitions
ACQUISITION - lower jaw,
- upper jaw,
- buccal bite registration
● View a 3D preview of the acquisitions
2.5.2.3 MODEL
In this phase, you can perform the following:
MODEL ● The buccal registration of the bite situation
● Adjust the virtual models
● Draw and edit preparation margins
2.5.2.4 CONNECT
In this phase, you can log in to the Connect portal (see “Log in to the
portal from the Connect software [→ 94]”). You must register as a
dentist on the Connect homepage for this.
You can open this phase by clicking the phase in the phase bar
"CONNECT" or the Next arrow in context bar.
66 97 549 D3534
16 D3534.208.18.02.02 08.2019
Dentsply Sirona 2 General data
Operator's Manual Connect SW 2.5 User interface
66 97 549 D3534
D3534.208.18.02.02 08.2019 17
2 General data Dentsply Sirona
2.5 User interface Operator's Manual Connect SW
Status symbols
All steps have a status symbol. Asterisks indicate obligatory steps.
Orange asterisk / The step is mandatory and has not been started yet.
no check mark:
66 97 549 D3534
18 D3534.208.18.02.02 08.2019
Dentsply Sirona 2 General data
Operator's Manual Connect SW 2.5 User interface
66 97 549 D3534
D3534.208.18.02.02 08.2019 19
3 Getting started Dentsply Sirona
3.1 Installing the software Operator's Manual Connect SW
3 Getting started
NOTE
Installing DirectX
If DirectX is not yet installed on your computer, it will be installed now.
Accept the license agreement and decide whether the computer is to
be restarted now or later.
66 97 549 D3534
20 D3534.208.18.02.02 08.2019
Dentsply Sirona 3 Getting started
Operator's Manual Connect SW 3.4 Copy protection
IMPORTANT
Software license for Connect SW is pre-installed.
No separate license needs to be installed for the Connect SW
software. This is pre-installed on every USB license stick.
66 97 549 D3534
D3534.208.18.02.02 08.2019 21
3 Getting started Dentsply Sirona
3.6 Starting the software Operator's Manual Connect SW
66 97 549 D3534
22 D3534.208.18.02.02 08.2019
Dentsply Sirona 3 Getting started
Operator's Manual Connect SW 3.7 License manager
66 97 549 D3534
D3534.208.18.02.02 08.2019 23
4 Configuration Dentsply Sirona
4.1 Parameters Operator's Manual Connect SW
4 Configuration
The "Configuration" menu contains the following submenus:
● "Parameters"
● "Devices"
● "Settings"
● "Apps"
4.1 Parameters
General information
The "Parameters" menu is arranged according to adjustment options.
The parameters in the configuration are used the standard value when
calculating tools.
Preparation Analysis
Under "Preparation Analysis" you can define the ideal distance of the
preparation to the antagonist. A tolerance value can also be specified
here.
4.2 Devices
All connected devices can be displayed and configured under the menu
item "Devices".
If the device is inaccessible, a warning symbol is displayed instead of a
miniature image.
Refresh Devices
With the "Refresh Devices" button you can:
● refresh the status or
● Check the current availability of a device.
66 97 549 D3534
24 D3534.208.18.02.02 08.2019
Dentsply Sirona 4 Configuration
Operator's Manual Connect SW 4.2 Devices
4.2.1 Scanner
4.2.1.1 Configuring the scanner
Audio feedback
Using the "Sound:" selection box, you can switch the audio feedback for
acquisitions on or off. You can set the volume using the slide bar. You
are able to choose from five different sounds.
66 97 549 D3534
D3534.208.18.02.02 08.2019 25
4 Configuration Dentsply Sirona
4.2 Devices Operator's Manual Connect SW
Starting calibration
1. In the software, navigate to the system menu and click on the
"Configuration" button.
2. Press the "Devices" button.
3. Press the "Omnicam" / "Primescan" button.
4. Press the "Calibrate" button.
Ä The camera view is displayed in one window.
5. Enter the 8-digit Sirona ID. You can find this ID on the sticker on the
calibration set (not required for CEREC Primescan / Primescan).
66 97 549 D3534
26 D3534.208.18.02.02 08.2019
Dentsply Sirona 4 Configuration
Operator's Manual Connect SW 4.2 Devices
5. Turn the screw counter-clockwise until you reach the next latching
point.
6. Hold the scanner still.
Ä The software prompts you to proceed to the next latching.
7. CEREC Omnicam / Omnicam: Execute steps 5 and 6 a total of 11
times.
CEREC Primescan / Primescan: Execute steps 5 and 6 a total of 17
times.
Ä The software provides status updates on the calibration and
informs you once the procedure is complete.
Ä You will be prompted to measure the position of the exit
window.
66 97 549 D3534
D3534.208.18.02.02 08.2019 27
4 Configuration Dentsply Sirona
4.2 Devices Operator's Manual Connect SW
End calibration
ü The software indicates that the calibration was completed
successfully.
➢ Press the "OK" button.
Ä The scanner is calibrated.
66 97 549 D3534
28 D3534.208.18.02.02 08.2019
Dentsply Sirona 4 Configuration
Operator's Manual Connect SW 4.2 Devices
General information
NOTE
Faulty color analysis
The color analysis can be negatively impacted due to strong light
incidence and it can lead to varying results.
➢ Set the scanner up so that it is not located directly in the beam
path of an extreme light source (e.g., the treatment light) and not
exposed to direct sunlight.
NOTE
Observe color calibration
A color calibration may only be performed at least 20 minutes after
the system start/cleaning.
The color calibration must be performed regularly.
The scanner must be color calibrated every two weeks in order to carry
out a reliable color analysis. You will achieve the best results if the
scanner is color calibrated immediately before scanning a new case.
Carry out a color calibration also after changing a sleeve/mirror sleeve.
Heavily scratched sleeve window may not be used for a color analysis.
66 97 549 D3534
D3534.208.18.02.02 08.2019 29
4 Configuration Dentsply Sirona
4.2 Devices Operator's Manual Connect SW
NOTE
Only use color calibration set with clean, dry CEREC Primescan /
Primescan / CEREC Omnicam / Omnicam scanner
In order to achieve optimum results, the CEREC Primescan /
Primescan / CEREC Omnicam / Omnicam scanner must be clean,
disinfected and dry before color calibration.
➢ Make sure that the CEREC Primescan / Primescan / CEREC
Omnicam / Omnicam scanner is clean, disinfected and dry.
66 97 549 D3534
30 D3534.208.18.02.02 08.2019
Dentsply Sirona 4 Configuration
Operator's Manual Connect SW 4.2 Devices
NOTE
Regularly replacing the color calibration set
In order to achieve optimum results, the color calibration set must be
replaced regularly.
➢ Observe the following:
66 97 549 D3534
D3534.208.18.02.02 08.2019 31
4 Configuration Dentsply Sirona
4.3 Settings Operator's Manual Connect SW
CAUTION
Hot surface!
The coated sapphire glass of the scanner, is preheated in the scanner
cradle. When removing the scanner from its holder, the surface
temperature of the mirror sleeve can be up to 51°C. This may cause
an unpleasant heat sensation on contact with a person's skin or
mucous membrane. These temperatures will not damage the skin or
mucosal membrane.
After removing the scanner from the scanner cradle, the temperature
of the mirror sleeve drops within a few minutes (< 5 minutes) to less
than 43°C. The scanner is therefore suitable for use in the patient's
mouth for an unlimited period of time.
At an ambient temperature from 30°C, only select the three lower
heater settings.
NOTE
The firmware update is mandatory for operating the CEREC
Omnicam in conjunction with the Connect SW. When starting phase
ACQUISITION, the firmware must be updated.
The firmware update takes around two minutes.
4.3 Settings
The menu item "Settings" has the following subitems:
● ADA/FDI Notation
● Notification Messages
● Hub Settings
● Patient Database
● Language
● Quality Improvement Agreement
66 97 549 D3534
32 D3534.208.18.02.02 08.2019
Dentsply Sirona 4 Configuration
Operator's Manual Connect SW 4.3 Settings
4.3.2 Notifications
Notes may appear in pop-up windows when using the software. Many of
these messages can be deactivated by clicking on the "Don't show this
message again" check box. If this check box is already selected or if a
new user uses the software, all notifications can be reset here. By
pressing the "Reset" button, all notes are displayed again.
4.3.5 Language
Here, you can set the language of the software.
66 97 549 D3534
D3534.208.18.02.02 08.2019 33
4 Configuration Dentsply Sirona
4.4 App Center (applications) Operator's Manual Connect SW
66 97 549 D3534
34 D3534.208.18.02.02 08.2019
Dentsply Sirona 5 System menu
Operator's Manual Connect SW 5.1 Saving the impression
5 System menu
66 97 549 D3534
D3534.208.18.02.02 08.2019 35
5 System menu Dentsply Sirona
5.2 Saving the impression under a different name Operator's Manual Connect SW
NOTE
File types
If the selected file is an optical impression with CEREC Connect SW
or Connect SW this is opened. If not, it will not open, and an error
message will be displayed.
66 97 549 D3534
36 D3534.208.18.02.02 08.2019
Dentsply Sirona 5 System menu
Operator's Manual Connect SW 5.5 Exporting scan data
IMPORTANT
Dentsply Sirona will not be held liable for the further processing of
*.stl data in other/external software.
IMPORTANT
Software license for Connect SW is pre-installed.
No separate license needs to be installed for the Connect SW
software. This is pre-installed on every USB license stick.
5.8 Configuration
The configuration is described in the "Configuration [→ 24]" section.
66 97 549 D3534
D3534.208.18.02.02 08.2019 37
5 System menu Dentsply Sirona
5.9 Window mode Operator's Manual Connect SW
66 97 549 D3534
38 D3534.208.18.02.02 08.2019
Dentsply Sirona 6 Start view
Operator's Manual Connect SW
6 Start view
66 97 549 D3534
D3534.208.18.02.02 08.2019 39
6 Start view Dentsply Sirona
6.1 Create a new patient Operator's Manual Connect SW
Add patients
1. Press the "Add New Patient" button.
Ä An empty patient card is opened.
2. Enter a surname, first name, and date of birth. A real-time search
function is active during the input which should prevent duplicate
entries.
or
➢ Enter the patient ID.
3. Press the "Add New Case" button.
Ä The program switches over to the "ADMINISTRATION" phase.
66 97 549 D3534
40 D3534.208.18.02.02 08.2019
Dentsply Sirona 6 Start view
Operator's Manual Connect SW 6.2 Patient database
6.2.1 Buttons
The order icon opens a case.
The trash can is used for deleting the case or the patient.
66 97 549 D3534
D3534.208.18.02.02 08.2019 41
6 Start view Dentsply Sirona
6.2 Patient database Operator's Manual Connect SW
4. If you do not want to save the changes, click the X icon (Cancel) or
press “Esc” on the keyboard.
66 97 549 D3534
42 D3534.208.18.02.02 08.2019
Dentsply Sirona 7 Page palette
Operator's Manual Connect SW 7.1 View options
7 Page palette
Various functions and options are offered to you in the page palette,
depending on the restoration phase currently active.
You can open several page palettes at the same time. Initially all page
palettes opened on the right side of the main window in a fixed state. In
this state all opened page palettes share the height available there.
Should this display be inadequate for you, you can remove any page
palette of your choosing from the fixed state of the magnet bar. To do
this, drag the palette to the desired position within the main window.
The local views are determined by the element currently selected in the
object bar. Each element in the object bar therefore has its own
coordinate system. Depending on the current step, the following views
are available to you:
66 97 549 D3534
D3534.208.18.02.02 08.2019 43
7 Page palette Dentsply Sirona
7.1 View options Operator's Manual Connect SW
Global views
● "Top"
● "Bottom"
● "Right"
● "Left"
● "Front"
● "Back side"
Local views
● "Mesial"
● "Distal"
● "Buccal" / "Labial"
● "Lingual"
● "Cervical"
● "Occlusal" / "Incisal"
Rotating a 3D preview
1. Press with one finger or click on the 3D preview with the left mouse
button and hold it down.
2. Move your finger or the mouse.
Ä The 3D preview is rotated.
66 97 549 D3534
44 D3534.208.18.02.02 08.2019
Dentsply Sirona 7 Page palette
Operator's Manual Connect SW 7.2 Tools
7.2 Tools
You will find all tools in the page palette under "Tools". The available
tools are displayed for each step.
To change a tool, press the button of another tool in the window of the
of the page palette.
To close a tool, press the button of the active tool once more.
With the "Reset" button in the tools you can undo all changes made on
all restorations since the tool was started.
66 97 549 D3534
D3534.208.18.02.02 08.2019 45
7 Page palette Dentsply Sirona
7.2 Tools Operator's Manual Connect SW
IMPORTANT
For precision reasons, this function can only be operated by trackball
or touchpad.
With the "Cut" function, you can cut out model areas. The cut-out model
areas are then discarded once you exit the "Edit Model" step. You
cannot display discarded areas later on.
66 97 549 D3534
46 D3534.208.18.02.02 08.2019
Dentsply Sirona 7 Page palette
Operator's Manual Connect SW 7.2 Tools
7.2.4 Trimming
("MODEL" phase, "Trim" step)
IMPORTANT
For precision reasons, this function can only be operated by trackball
or touchpad.
With the "Trim" function, you can isolate the preparation. You can thus
e.g. draw in the preparation margin more easily. Trimmed image
regions can be optionally displayed and hidden later on.
66 97 549 D3534
D3534.208.18.02.02 08.2019 47
7 Page palette Dentsply Sirona
7.2 Tools Operator's Manual Connect SW
Manual drawing
With "Manual" you can draw in the preparation margin manually. With
this technique, you must place the individual points close together in
order to clearly define the contour of the preparation margin even in
difficult situations.
ü The "Margin" tool is open.
➢ Press the "Manual" button.
Ä The manual technique is switched on.
Tip: You start the drawing by double-clicking and end it by double-
clicking. The tooth number appears if the preparation margin is drawn. If
amendments are required you can make corrections by clicking on the
line in manual mode. To end the correction, click on the blue line again.
66 97 549 D3534
48 D3534.208.18.02.02 08.2019
Dentsply Sirona 7 Page palette
Operator's Manual Connect SW 7.3 Display objects
66 97 549 D3534
D3534.208.18.02.02 08.2019 49
7 Page palette Dentsply Sirona
7.3 Display objects Operator's Manual Connect SW
66 97 549 D3534
50 D3534.208.18.02.02 08.2019
Dentsply Sirona 7 Page palette
Operator's Manual Connect SW 7.4 Activating analysis
Color model
Using the "Color Model" button, you can change the color of models
that were acquired with the scanner.
66 97 549 D3534
D3534.208.18.02.02 08.2019 51
7 Page palette Dentsply Sirona
7.4 Activating analysis Operator's Manual Connect SW
Undercuts
If the preparation margin has been drawn in then the undercuts can be
displayed in the step "Preparation Analysis". The thickness of the
undercut can be read from the scale.
66 97 549 D3534
52 D3534.208.18.02.02 08.2019
Dentsply Sirona 8 ADMINISTRATION phase
Operator's Manual Connect SW
8 ADMINISTRATION phase
Selecting a restoration type
Tip: Restorations can be selected both in the upper and lower jaw.
➢ Select the restoration type in the page palette: "Single Restoration",
"Bridge Restoration" or "Full arch scan mode".
Bridge restoration
ü You have selected the restoration type "Bridge Restoration".
1. Determine the restoration type and the design mode for the
abutment teeth. With a bridge on implants you can only select
"Implant".
2. Select the positions of the abutment teeth of the bridge.
Ä The selected teeth are marked.
3. Determine the restoration type and the design mode for the
intermediate teeth.
Ä The selected teeth are marked.
4. In the step menu, go to the "Select Material" step to select the
material. The selection of the material class is optional.
5. Confirm your settings with "Ok".
Tip: If you click on "Cancel", the restoration will not be created and
the program will return to the odontogram.
6. If necessary, create further restorations.
66 97 549 D3534
D3534.208.18.02.02 08.2019 53
8 ADMINISTRATION phase Dentsply Sirona
Operator's Manual Connect SW
Select material
You can select the desired material in this optional step.
1. Select the desired material class.
2. Select "See Additional Notes" if you wish to order a certain material
and define it in the text field in the Connect portal.
Color selection
➢ Click "Select Final Shade" or "Select Stump Shade" to proceed to
the step of choosing colors. Stating the color is optional.
Tip: It is also possible to state the color later in the additional
information in the portal.
You can select the stump/restoration color in this step.
1. First select the shade guide.
2. Then select the appropriate color using the coloring group.
3. Press "Ok" to confirm the selection.
CAUTION
The user is responsible for the correct use of scanbodies, which are
offered by other providers. Dentsply Sirona can neither guarantee the
clinical safety and biocompatibility nor the scan quality for all available
scanbodies. Ask your provider whether such use with Omnicam or
Primescan is permitted in your country.
66 97 549 D3534
54 D3534.208.18.02.02 08.2019
Dentsply Sirona 8 ADMINISTRATION phase
Operator's Manual Connect SW
or
➢ Press the Next arrow in the context bar.
Ä The program switches over to the "ACQUISITION" phase.
66 97 549 D3534
D3534.208.18.02.02 08.2019 55
9 ACQUISITION phase Dentsply Sirona
9.1 Image catalogs Operator's Manual Connect SW
9 ACQUISITION phase
Deleting acquisitions
If an acquisition is not suitable, you can delete it. You can then execute
a new acquisition for the corresponding image catalog.
1. Select the image in the image catalog.
2. Move it using drag & drop to the trach can.
3. Press "Delete" in the context bar.
Ä A notification window then appears.
4. Confirm the deletion by selecting "Yes".
Ä The image is deleted.
66 97 549 D3534
56 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.2 3D Preview
9.2 3D Preview
In the default setting, the data are displayed from the occlusal direction
in the 3D preview.
You can freely select the viewing direction of the virtual model in the 3D
preview window by using the mouse.
Rotating a 3D preview
1. Press with one finger or click on the 3D preview with the left mouse
button and hold it down.
2. Move your finger or the mouse.
Ä The 3D preview is rotated.
Taking photos
1. In the step menu, go to "Video / Picture".
2. Select the "Photo" option.
3. Press the button or use the foot switch to trigger the scanner.
66 97 549 D3534
D3534.208.18.02.02 08.2019 57
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
Recording videos
1. In the step menu, go to "Video / Picture".
2. Select the "Video" option.
3. Press the button to start video recording or use the foot control.
4. Repeatedly press the button or actuate the foot control once more
to stop the recording.
66 97 549 D3534
58 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
IMPORTANT
Registration error
Should a registration error occur, you must return to another acquired
point.
To start with, practice this procedure on the model and then on
intraoral areas.
➢ Move the scanner to a position where a successful acquisition
was taken. A point that has already been acquired in the occlusal
area is best.
ð You will be able to hear the sound for registered acquisitions.
➢ Continue the acquisition.
66 97 549 D3534
D3534.208.18.02.02 08.2019 59
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
Important: Ensure that the distance between the coated sapphire glass
of the scanner and the scanned surface is observed.
The distance must be between 0 - 15 mm (ideally: 5 mm). The scanner
does not rest on the teeth or the gums. If the distance is too great, no
data will be obtained.
1. Move the scanner to the starting position. For this purpose, the
scanner is in the occlusal view of the tooth, which is next to the
prepared tooth in the distal direction.
2. Scan in the mesial direction. To do so, move the scanner slowly in
the occlusal direction from the distal-positioned tooth over the
prepared tooth to the mesial-positioned tooth.
With full jaw acquisitions, the scan sequence is different for the
transition to anteriors. Scanning begins with the lingual and labial areas,
before moving on to the incisors.
9.3.2.2.2 Buccal scan
66 97 549 D3534
60 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
66 97 549 D3534
D3534.208.18.02.02 08.2019 61
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
➢ Position the scanner occlusally above the last tooth on the right, to
start the scanning process.
66 97 549 D3534
62 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
66 97 549 D3534
D3534.208.18.02.02 08.2019 63
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
66 97 549 D3534
64 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
NOTE
Image brightness
The image brightness during the acquisition is controlled
automatically, so that there is always optimum image brightness,
largely independent of the distance between the scanner and the
tooth.
The surroundings of the tooth to be scanned should be as weakly
illuminated as possible. Avoid any type of external light. Switch off the
operating light.
IMPORTANT
Do not use cotton rolls in the scan area
Do not use any cotton rolls in the vicinity of the scan area, as they can
reduce the precision of the scan and create image interference.
CAUTION
Prevent cross-contamination
Germs can be transmitted to uncontaminated persons via the hands,
materials or objects.
➢ For hygiene reasons, wear a new set of disposable gloves for
each patient while using the scanner.
CAUTION
In the case of patients with allergies to nickel
Should the scanner mirror sleeve make contact with the skin of
patients allergic to nickel, allergic reactions may occur.
➢ Ensure that in the case of patients with nickel allergies, the parts
of the mirror sleeve which may make contact are protected from
areas of skin.
66 97 549 D3534
D3534.208.18.02.02 08.2019 65
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
WARNING
Risk of injury for those diagnosed with epilepsy
For persons who have been diagnosed with epilepsy, there is a risk of
epileptic shock through the flashing light of the scanner.
➢ Patients who have been diagnosed with epilepsy cannot be
treated with the scanner.
➢ Dentists and dental assistants who have been diagnosed with
epilepsy cannot work with the scanner.
66 97 549 D3534
66 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
NOTE
Image brightness
The image brightness during the acquisition is controlled
automatically, so that there is always optimum image brightness,
largely independent of the distance between the scanner and the
tooth.
The surroundings of the tooth to be scanned should be as weakly
illuminated as possible. Avoid any type of external light. Switch off the
operating light.
IMPORTANT
Do not use cotton rolls in the scan area
Do not use any cotton rolls in the vicinity of the scan area, as they can
reduce the precision of the scan and create image interference.
CAUTION
Prevent cross-contamination
Germs can be transmitted to uncontaminated persons via the hands,
materials or objects.
➢ For hygiene reasons, wear a new set of disposable gloves for
each patient while using the scanner.
WARNING
Risk of injury for those diagnosed with epilepsy
For persons who have been diagnosed with epilepsy, there is a risk of
epileptic shock through the flashing light of the scanner.
➢ Patients who have been diagnosed with epilepsy cannot be
treated with the scanner.
➢ Dentists and dental assistants who have been diagnosed with
epilepsy cannot work with the scanner.
CAUTION
Potentially hazardous optical radiation
The scanner transmits potentially hazardous optical radiation which
may cause harm to the eyes.
➢ During operation, do not look directly at the scanner for long
periods.
IMPORTANT
Potential switch-off procedure
In the case of several repeated scans of the image fields without
model calculation, the scanner can deviate from the calibrated
temperature range. In this case, a warning message appears and you
need to take a scanning break prior to completing the exposures.
Please wait roughly long enough for the remaining exposures to be
performed. The potential switch-off procedure is innocuous for your
scanner and is not a malfunction.
66 97 549 D3534
D3534.208.18.02.02 08.2019 67
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
IMPORTANT
Heating up the scanner
The internal scanner heating ensures that condensation does not
form during scanning. The heating starts immediately after starting up
the exposure unit, and after around five minutes the scanner is free of
condensation. This is usually the case through to navigation into the
exposure phase.
66 97 549 D3534
68 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
IMPORTANT
Registration error
Should a registration error occur, you must return to another acquired
point.
To start with, practice this procedure on the model and then on
intraoral areas.
➢ Move the scanner to a position where a successful acquisition
was taken. A point that has already been acquired in the occlusal
area is best.
ð You will be able to hear the sound for registered acquisitions.
➢ Continue the acquisition.
66 97 549 D3534
D3534.208.18.02.02 08.2019 69
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
Important: Ensure that the distance between the coated sapphire glass
of the scanner and the scanned surface is observed.
The distance must be between 0 - 20 mm (ideally: 2 mm). The scanner
does not rest on the teeth or the gums.
1. Move the scanner to the starting position. For this purpose, the
scanner is in the occlusal view of the tooth, which is next to the
prepared tooth in the distal direction.
2. Scan in the mesial direction. To do so, move the scanner in the
occlusal direction from the distal-positioned tooth over the prepared
tooth to the mesial-positioned tooth.
9.3.3.2.2 Buccal scan
66 97 549 D3534
70 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
66 97 549 D3534
D3534.208.18.02.02 08.2019 71
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
66 97 549 D3534
72 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
Procedure 1
1. Start with the oral surface of the anterior teeth and move the
scanner in the oral direction along the quadrant. Move the scanner
over the distal tooth to the vestibular side and track the first
quadrant to the anterior teeth. Gently tilt the scanner approx. 30° in
the coronal-apical direction.
2. Move the scanner as shown below (1) for the second quadrant.
3. Then scan the anterior teeth from cuspid to cuspid in the coronal-
apical direction. Ensure that both the labial surface and the oral
surfaces are visible.
Extend this third scan to locations where you can view scan holes.
66 97 549 D3534
D3534.208.18.02.02 08.2019 73
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
Procedure 2
1. Start occlusally on the distal tooth, tilt the scanner approx. 60° in an
oral direction and move it orally along the dental arch up to the
opposite distal tooth.
2. Guide the scanner occlusally from the distal tooth across the entire
dental arch back to the other side.
3. To complete the scan, tilt the scanner approx. 60° in a buccal
direction and move it buccally along the entire dental arch.
66 97 549 D3534
74 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
IMPORTANT
For precision reasons, this function can only be operated by trackball
or touchpad.
66 97 549 D3534
D3534.208.18.02.02 08.2019 75
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
66 97 549 D3534
76 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
66 97 549 D3534
D3534.208.18.02.02 08.2019 77
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
User guidance
You are guided through the acquisition steps with tone signals and
graphic displays.
● Starting point for a scanning step: Hold the camera over this marker
for 3 seconds to start scanning the scanning step.
● Resume the scan. If the scan has been interrupted, hold the camera
over this marker to continue the scan.
● Target area: Guide the camera along the arrow into this area. The
partial scan does not end at an exact point.
● End point of a quadrant: Hold the camera over this marker for 3
seconds to mark the end point of the quadrant.
66 97 549 D3534
78 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
66 97 549 D3534
D3534.208.18.02.02 08.2019 79
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
5. Position the camera above the last molar on the right. The starting
point is marked with a start marker (A) and is automatically
recognized.
6. Turn the camera towards the buccal and guide it in the direction of
the arrow over the arch as far as the target marker (B) on the
centerline.
Ä If the centerline has been scanned, a signal tone will be heard.
Ä The software changes to the next "Scan transversal right" step.
7. Guide the camera in the direction of the arrow over the marked area
from the start marker (A) to the target area (B).
Ä If the connection is scanned, a signal tone will be heard.
Ä The software changes to the next left-hand quadrant step.
66 97 549 D3534
80 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
4. Position the camera above the last molar on the left. The starting
point is marked with a start marker (A) and is automatically
recognized.
5. Turn the camera towards the buccal and guide it in the direction of
the arrow over the arch as far as the target marker (B) on the
centerline.
Ä If the centerline has been scanned, a signal tone will be heard.
Ä The software changes to the next "Scan transversal left" step.
6. Guide the camera in the direction of the arrow over the marked area
from the start marker (A) to the target marker (B).
Ä If the connection is scanned, a signal tone will be heard.
Ä The software changes to the "Complete Jaw" step.
66 97 549 D3534
D3534.208.18.02.02 08.2019 81
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
IMPORTANT
Particularly in the area of the preparation as well as neighboring
teeth, the yellow marked areas should be rescanned.
IMPORTANT
For precision reasons, this function can only be operated by trackball,
touchpad, or mouse.
If necessary, edit the blue boundary line of the model. Ensure that only
the areas required for manufacturing the restoration are inside the
boundary line.
1. Align the model in such a way that you can view the area to be
edited orthogonally.
2. Double-click on the point of the boundary line, at which you would
like to start editing.
3. Pull and click to set additional points.
4. To stop editing the boundary line, double-click again.
IMPORTANT
Once the boundary line has been manually edited, the automatic
extension of the boundary line for the current jaw is deactivated
irrevocably.
66 97 549 D3534
82 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
4. Position the camera above the last molar on the right. The starting
point is marked with a start marker (A) and is automatically
recognized.
5. Turn the camera towards the buccal and guide it in the direction of
the arrow over the arch as far as the target marker (B) on the
centerline.
Ä If the centerline has been scanned, a signal tone will be heard.
Ä The software changes to the next "Scan transversal right" step.
66 97 549 D3534
D3534.208.18.02.02 08.2019 83
9 ACQUISITION phase Dentsply Sirona
9.3 Take a scan Operator's Manual Connect SW
IMPORTANT
As the jaw models are displayed with a distance from one another
before the first bite is scanned, the target symbol appears lower in
relation to the lower jaw during scanning than it is in reality.
➢ Before scanning, observe the position of the target marker in the
upper jaw (between the premolars and the molars in the figure)
and scan this region in the mouth.
6. Guide the camera in the direction of the arrow over the marked area
from the start marker (A) as far as the target marker (B).
Ä If the connection is scanned, a signal tone will be heard.
Ä The software changes to the next "Scan palatal left" step.
4. Position the camera above the last molar on the left and hold the
camera in this position to mark the start of the partial scan.
5. Turn the camera towards the buccal and guide it in the direction of
the arrow over the arch as far as the centerline.
Ä If the centerline has been scanned, a signal tone will be heard.
Ä The software changes to the next "Scan transversal left" step.
6. Guide the camera in the direction of the arrow over the marked
area.
Ä If the connection is scanned, a signal tone will be heard.
Ä The software changes to the "Complete Jaw" step.
66 97 549 D3534
84 D3534.208.18.02.02 08.2019
Dentsply Sirona 9 ACQUISITION phase
Operator's Manual Connect SW 9.3 Take a scan
IMPORTANT
Particularly in the area of the preparation as well as neighboring
teeth, the yellow marked areas should be rescanned.
IMPORTANT
For precision reasons, this function can only be operated by trackball,
touchpad, or mouse.
If necessary, edit the blue boundary line of the model. Ensure that only
the areas required for manufacturing the restoration are inside the
boundary line.
1. Align the model in such a way that you can view the area to be
edited orthogonally.
2. Double-click on the point of the boundary line, at which you would
like to start editing.
3. Pull and click to set additional points.
4. To stop editing the boundary line, double-click again.
IMPORTANT
Once the boundary line has been manually edited, the automatic
extension of the boundary line for the current jaw is deactivated
irrevocably.
➢ Click on the Next arrow to switch to the next scan object "Buccal".
66 97 549 D3534
D3534.208.18.02.02 08.2019 85
9 ACQUISITION phase Dentsply Sirona
9.4 Finishing the phase Operator's Manual Connect SW
IMPORTANT
As the jaw models are displayed with a distance from one another
before the first bite is scanned, the target symbol appears lower in
relation to the upper jaw during scanning than it is in reality.
➢ Before scanning, observe the position of the target marker in the
lower jaw (between the premolars and the molars in the figure)
and scan this region in the mouth.
or
➢ Press the Next arrow in the context bar.
Ä The program switches over to the "MODEL" phase.
66 97 549 D3534
86 D3534.208.18.02.02 08.2019
Dentsply Sirona 10 MODEL phase
Operator's Manual Connect SW 10.1 Editing the model
10 MODEL phase
In the "MODEL" phase, the virtual models are reconstructed based on
the acquired image catalogs.
If you would like to edit the model, change to the "Edit Model" step.
The "Edit Model", "Bite Registration" and "Set Model Axis" steps relate
to the entire model (upper and lower jaw).
All other steps in the "MODEL" phase refer to the restoration currently
selected. These steps must be carried out as relevant for all
restorations.
Manual correlation
In this step, the virtual models of the upper and lower jaw should be
aligned with one another with the help of the buccal image in its correct
position.
In the "Buccal Bite Registration" step, you can work with the following
tools in the page palette:
● Drag Buccal
● Turn Buccal Impression
66 97 549 D3534
D3534.208.18.02.02 08.2019 87
10 MODEL phase Dentsply Sirona
10.2 Buccal registration (optional) Operator's Manual Connect SW
Drag Buccal
1. Rotate both models so that you can see the overlap area of the
buccal acquisition and of the upper jaw and lower jaw.
2. Now drag the buccal acquisition to the corresponding area of the
upper jaw with the finger/mouse and let go (drag & drop).
66 97 549 D3534
88 D3534.208.18.02.02 08.2019
Dentsply Sirona 10 MODEL phase
Operator's Manual Connect SW 10.2 Buccal registration (optional)
The buccal image is then displayed upside down. When you begin the
registration, the software detects this and automatically flips the image
right side up.
66 97 549 D3534
D3534.208.18.02.02 08.2019 89
10 MODEL phase Dentsply Sirona
10.3 Manual correlation for image fields Operator's Manual Connect SW
66 97 549 D3534
90 D3534.208.18.02.02 08.2019
Dentsply Sirona 10 MODEL phase
Operator's Manual Connect SW 10.5 Trimming the preparation
IMPORTANT
For precision reasons, this function can only be operated by trackball,
touchpad, or mouse.
66 97 549 D3534
D3534.208.18.02.02 08.2019 91
10 MODEL phase Dentsply Sirona
10.6 Entering the preparation margin Operator's Manual Connect SW
General information
IMPORTANT
For precision reasons, this function can only be operated by trackball,
touchpad, or mouse.
IMPORTANT
You can enlarge or reduce the 3D view during the input or editing of
the preparation margin (see “3D Preview”) in order to ensure the
correct positioning of the preparation margin. Be sure to hold down
the left mouse button for a long time. A short click adds a point to the
preparation margin.
66 97 549 D3534
92 D3534.208.18.02.02 08.2019
Dentsply Sirona 10 MODEL phase
Operator's Manual Connect SW 10.8 Finishing the phase
or
➢ Press the Next arrow in the context bar.
Ä The program switches over to the next phase.
66 97 549 D3534
D3534.208.18.02.02 08.2019 93
11 CONNECT phase Dentsply Sirona
11.1 Registration and functions of the Connect portal Operator's Manual Connect SW
11 CONNECT phase
In this phase, you can log in to the Connect portal (see “Log in to the
portal from the Connect software [→ 94]”). You must register as a
dentist on the Connect homepage for this.
You can open this phase by clicking the phase in the phase bar
"CONNECT" or the Next arrow in context bar.
66 97 549 D3534
94 D3534.208.18.02.02 08.2019
Dentsply Sirona 11 CONNECT phase
Operator's Manual Connect SW 11.3 Check restoration information
IMPORTANT
Maximum 5 labs
You can save a maximum of 5 labs in the list of your favorite labs.
You can add labs to or delete labs from the list via "Add Lab".
Under "Return Date" you can select the delivery date by clicking on the
desired day in the calendar. Under "Time" you can also state a delivery
time.
You can then click "Additional instructions of purchaser" to enter
individual order information, which must be entered in addition to the
return date.
66 97 549 D3534
D3534.208.18.02.02 08.2019 95
11 CONNECT phase Dentsply Sirona
11.7 Shopping cart Operator's Manual Connect SW
66 97 549 D3534
96 D3534.208.18.02.02 08.2019
Dentsply Sirona Index
Operator's Manual Connect SW
Index
A L
Arrow keys, 18 Licenses, 37
B M
Buccal registration Manual
Moving acquisitions, 45 HTML format, 13, 38
Rotating exposures, 45 PDF format, 13
Manufacturer's address, 6
C Model
CE mark, 8 Trimming the preparation, 91
Close Multi-Touch, 12
MODEL, 87 Edit a 3D model, 12, 15
Code library, 37 Enlarge a 3D model, 12, 15
Contact surfaces Navigation, 12
Virtual model, 51 Reduce a 3D model, 12, 15
Customer Service Center, 6 Rotate a 3D model, 12, 15
Cut
Inverting the model area, 46 O
Odontogram
D ADA, 13
Declaration of conformity, 8 FDI, 13
Display objects
Transparency, 49 P
Trimmed model, 50 Product safety, 8
Displaying side surfaces, 49, 51
Displaying the bottom, 49, 51 R
Rotating
E Buccal image, 89
Entering the preparation margin Lower and upper jaw, 87
Automatic, 47
Manual, 48 S
Safety instructions, 10
I Scan objects, 77
Image regions Scanner
Hide, 91 Warm-up time, 58
impression Screen, 12
Exporting, 36 Software
Import, 36 Uninstallation, 20
Save as, 36 Step menu, 18
Saving, 35
66 97 549 D3534
D3534.208.18.02.02 08.2019 97
Index Dentsply Sirona
Operator's Manual Connect SW
T
Tool
Buccal registration, 45
Cut, 46
Entering the preparation margin, 47
Resetting the model, 46
Trimming, 47
Trimming
Inverting the selection, 47
V
View options, 43
Views
Global, 44
Local, 44
66 97 549 D3534
98 D3534.208.18.02.02 08.2019