Nothing Special   »   [go: up one dir, main page]

Photobiological Safety of Lamps and Lamp Systems: Test Report IEC 62471

Download as pdf or txt
Download as pdf or txt
You are on page 1of 22

Test Report issued under the responsibility of:

TEST REPORT
IEC 62471
Photobiological safety of lamps and lamp systems
Report Reference No. .................... : 3150235.50A
Date of issue .................................... : 2014-06-10
Total number of pages ................... : 22

CB Testing Laboratory .................. : DEKRA Testing and Certification (Shanghai) Ltd.


Address ............................................ : 10/F, #250, Jiangchangsan Road building 16 Headquater Economy
Park Shibei Hi-Tech Park, Zhabei District, Shanghai, P.R.C 200436
Applicant’s name ........................... : Philips (China) Investment Co., Ltd.
Address ............................................ : No.1602-1605, No.218 Tianmu W Rd, Shanghai City, 200072,
P. R. China.
Test specification:
Standard .......................................... : IEC 62471:2006 (First Edition)
Test procedure ................................. : Type test
Non-standard test method…………..: N/A
Test Report Form No. .................... : V 1.0
TRF Originator ................................. : DEKRA Testing and Certification (Shanghai) Ltd
Master TRF ...................................... : 2013-10
Test item description ..................... : LED batten
Trade Mark....................................... : PHILIPS
Manufacturer .................................... : Philips (China) Investment Co., Ltd.
No.1602-1605, No.218 Tianmu W Rd, Shanghai City, 200072, P.R.
China.
Model/Type reference : BN 108 series (for details please refer to appendix 3)
Ratings ............................................. : 220 - 240 V~; 50/60 Hz; IP20
Page 2 of 22 Report No.:3150235.50A

Testing procedure and testing location:


CB Testing Laboratory: DEKRA Testing and Certification (Shanghai) Ltd.
Testing location/ address .......................... : 10/F, #250, Jiangchangsan Road building 16 Headquater
Economy Park Shibei Hi-Tech Park, Zhabei District, Shanghai,
P.R.C 200436
Associated CB Laboratory:
Testing location/ address .......................... :

Tested by (name + signature)


........................................................ : Abby Yang
Approved by (+ signature)
........................................................ : Hanson Zhang
Testing procedure: TMP
Tested by (name + signature) ......... :
Approved by (+ signature) ............... :
Testing location/ address .......................... :

Testing procedure: WMT


Tested by (name + signature) ......... :
Witnessed by (+ signature) ............. :
Approved by (+ signature) ............... :
Testing location/ address .......................... :

Testing procedure: SMT


Tested by (name + signature) ......... :
Approved by (+ signature) ............... :
Supervised by (+ signature) ............ :
Testing location/ address .......................... :

Testing procedure: RMT


Tested by (name + signature) ......... :
Approved by (+ signature) ............... :
Supervised by (+ signature) ............ :
Testing location/ address .......................... :
Page 3 of 22 Report No.:3150235.50A

Summary of testing:

Tests performed (name of test and test clause): Testing location:

These tests DEKRA Testing and Certification


Fulfil the requirements of standard ISO/IEC 17025. (Shanghai) Ltd.
When determining the test conclusion, the Measurement 10/F, #250, Jiangchangsan Road
Uncertainty of test has been considered. building 16 Headquater Economy
Park Shibei Hi-Tech Park, Zhabei
District, Shanghai, P.R.C 200436
The tested sample of LED batten light list as below

BN 108 series

have been tested according to the IEC 62471(first edition, 2006-07)


including considering IEC/TR 62778: 2012 and classification as
following:

Model/Type Test condition Risk Group


reference
All models GLS exempt
Non GLS risk 1
Worst Case risk 2
risk 3

Summary of compliance with National Differences:

PASS

Copy of marking plate:

N/A
Page 4 of 22 Report No.:3150235.50A

Test item particulars ..................................................:


Tested lamp ................................................................ : continuous wave lamps pulsed lamps
Tested lamp system .....................................................: N/A
Lamp classification group .............................................: exempt risk 1 risk 2 risk 3
(See table 6.1)
Lamp cap .................................................................... : N/A
Bulb ............................................................................. : LEDs
Rated of the lamp .........................................................: 220 - 240 V~; 50/60 Hz; IP20
Furthermore marking on the lamp................................: N/A
Seasoning of lamps according IEC standard ..............: N/A
Used measurement instrument ....................................: spectroradiometer
Temperature by measurement .....................................: 23-28 °C
Information for safety use .............................................: --
Possible test case verdicts:
– test case does not apply to the test object ............ : N/A
– test object does meet the requirement .................. : P (Pass)
– test object does not meet the requirement ............ : F (Fail)
Testing:
Date of receipt of test item ........................................... : 2014-05
Date (s) of performance of tests................................... : 2014-05 to 2014-06
General remarks:
The test results presented in this report relate only to the object tested.
This report shall not be reproduced, except in full, without the written approval of the Issuing testing laboratory.
"(See Enclosure #)" refers to additional information appended to the report.
"(See appended table)" refers to a table appended to the report.
Throughout this report a comma (point) is used as the decimal separator.
List of test equipment must be kept on file and available for review.

The product complied with the following standards:


GLS Non-GLS Worst case (200mm)
IEC 62471:2006 IEC 62471:2006 IEC 62471:2006
IEC/TR 62471-2:2009 IEC/TR 62471-2:2009 IEC/TR 62471-2:2009
EN 62471:2008 EN 62471:2008 EN 62471:2008
IEC/TR 62778: 2012 IEC/TR 62778: 2012 IEC/TR 62778: 2012

This report should be read in conjunction with the attached pages concerned with the European group
differences and national differences of the standards EN 62471:2008 with the reference number of 3150235.50B.
(2 pages)
Factory:
Philips Luminaire (Chengdu) Co., Ltd.
No.91, Tianyuan Road, Hi-tech West District, Chengdu City, Sichuan Province, 611730, P. R. China.
Page 5 of 22 Report No.:3150235.50A

General product information:

The products are GreenUp Batten. The products have the same electrical and mechanical construction but
different CCTs; length & driver.
For details please refer to appendix 3.

After review, the model BN108C LED28 CW L1200 was selected to be tested.

In order to consider the worst case, the product should be tested at 200mm.

The sample of BN108C LED28 CW L1200 was tested at 200mm from the light source. CCT of the spectral
irradiance was found at 6877 K.

Amendment 1 report:
The original Test report Ref. No. 3141630, dated 2013-10-25 was replaced, include the following additions
and changes, which were considered technical modifications:
1. Add new models. Those models have the same mechanical and electrical construction as the original
types, except different power.
2. Add new LED drivers.
Finally, additional test performed on model BN108C LED36 CW L1200 MB.
The tested samples were considered as GLS product which was tested at illuminance of 500 lx, but not at a
distance less than 200 mm.
In order to consider the worst case, the product should be tested at 200mm.
The sample of BN108C LED36 CW L1200 MB was tested at 200mm from the light source. CCT of the
spectral irradiance was found at 6955 K.
Page 6 of 22 Report No.:3150235.50A

IEC 62471
Clause Requirement + Test Result – Remark Verdict

4 EXPOSURE LIMITS P
4.1 General P
The exposure limits in this standard is not less than P
0,01 ms and not more than any 8-hour period and
should be used as guides in the control of exposure
Detailed spectral data of a light source are generally P
required only if the luminance of the source exceeds see clause 4.3
104 cd.m-2
4.3 Hazard exposure limits P
4.3.1 Actinic UV hazard exposure limit for the skin and eye P
The exposure limit for effective radiant exposure is P
30 J.m-2 within any 8-hour period
To protect against injury of the eye or skin from P
ultraviolet radiation exposure produced by a
broadband source, the effective integrated spectral
irradiance , ES, of the light source shall not exceed
the levels defined by:
P

The permissible time for exposure to ultraviolet P


radiation incident upon the unprotected eye or skin
shall be computed by:
P

4.3.2 Near-UV hazard exposure limit for eye P


For the spectral region 315 nm to 400 nm (UV-A) the P
total radiant exposure to the eye shall not exceed
10000 J.m-2 for exposure times less than 1000 s. For
exposure times greater than 1000 s (approximately
16 minutes) the UV-A irradiance for the unprotected
eye, EUVA, shall not exceed 10 W.m-2.
The permissible time for exposure to ultraviolet P
radiation incident upon the unprotected eye for time
less than 1000 s, shall be computed by:
P

4.3.3 Retinal blue light hazard exposure limit P


To protect against retinal photochemical injury P
from chronic blue-light exposure, the integrated
spectral radiance of the light source weighted
against the blue-light hazard function, B(λ), i.e., the
blue-light weighted radiance , LB, shall not exceed
the levels defined by:
P
for t ≤ 104 s
Page 7 of 22 Report No.:3150235.50A

IEC 62471
Clause Requirement + Test Result – Remark Verdict

P
for t > 104 s

4.3.4 Retinal blue light hazard exposure limit - small source N/A
Thus the spectral irradiance at the eye Eλ, weighted N/A
against the blue-light hazard function B(λ) shall not see table 4.2
exceed the levels defined by:
N/A
for t ≤ 100 s

N/A
for t > 100 s

4.3.5 Retinal thermal hazard exposure limit P


To protect against retinal thermal injury, the P
integrated spectral radiance of the light source, Lλ,
weighted by the burn hazard weighting function R(λ)
(from Figure 4.2 and Table 4.2), i.e., the burn hazard
weighted radiance, shall not exceed the levels
defined by:
P
(10 μs ≤ t ≤ 10 s)

4.3.6 Retinal thermal hazard exposure limit – weak visual stimulus N/A
For an infrared heat lamp or any near-infrared N/A
source where a weak visual stimulus is inadequate
to activate the aversion response, the near infrared
(780 nm to 1400 nm) radiance, LIR, as viewed by the
eye for exposure times greater than 10 s shall be
limited to:
N/A
t > 10 s

4.3.7 Infrared radiation hazard exposure limits for the eye P


The avoid thermal injury of the cornea and possible P
delayed effects upon the lens of the eye
(cataractogenesis), ocular exposure to infrared
radiation, EIR, over the wavelength range 780 nm to
3000 nm, for times less than 1000 s, shall not
exceed:
P
t ≤ 1000 s

For times greater than 1000 s the limit becomes: P


P
t > 1000 s

4.3.8 Thermal hazard exposure limit for the skin P


Visible and infrared radiant exposure (380 nm to P
3000 nm) of the skin shall be limited to:
Page 8 of 22 Report No.:3150235.50A

IEC 62471
Clause Requirement + Test Result – Remark Verdict

5 MEASUREMENT OF LAMPS AND LAMP SYSTEMS P


5.1 Measurement conditions P
Measurement conditions shall be reported as part of P
the evaluation against the exposure limits and the
assignment of risk classification.
5.1.1 Lamp ageing (seasoning) N/A
Seasoning of lamps shall be done as stated in the N/A
appropriate IEC lamp standard.
5.1.2 Test environment P
For specific test conditions, see the appropriate IEC P
lamp standard or in absence of such standards, the
appropriate national standards or manufacturer’s
recommendations.
5.1.3 Extraneous radiation P
Careful checks should be made to ensure that P
extraneous sources of radiation and reflections do
not add significantly to the measurement results.
5.1.4 Lamp operation P
Operation of the test lamp shall be provided in N/A
accordance with:
– the appropriate IEC lamp standard, or N/A
– the manufacturer’ s recommendation P
5.1.5 Lamp system operation N/A
The power source for operation of the test lamp shall N/A
be provided in accordance with:
– the appropriate IEC standard, or N/A
– the manufacturer’ s recommendation N/A
5.2 Measurement procedure P
5.2.1 Irradiance measurements P
Minimum aperture diameter 7mm. P
Maximum aperture diameter 50 mm. P
The measurement shall be made in that position of P
the beam giving the maximum reading.
The measurement instrument is adequate P
calibrated.
5.2.2 Radiance measurements P
5.2.2.1 Standard method P
The measurements made with an optical system. P
Page 9 of 22 Report No.:3150235.50A

IEC 62471
Clause Requirement + Test Result – Remark Verdict

The instrument shall be calibrated to read in P


absolute radiant power per unit receiving area and
per unit solid angle to acceptance averaged over the
field of view of the instrument.
5.2.2.2 Alternative method P
Alternatively to an imaging radiance set-up, an P
irradiance measurement set-up with a circular field
stop placed at the source can be used to perform
radiance measurements.
5.2.3 Measurement of source size P
The determination of α, the angle subtended by a P
source, requires the determination of the 50%
emission points of the source.
5.2.4 Pulse width measurement for pulsed sources N/A
The determination of Δt, the nominal pulse duration N/A
of a source, requires the determination of the time
during which the emission is > 50% of its peak value.
5.3 Analysis methods P
5.3.1 Weighting curve interpolations P
To standardize interpolated values, use linear P
interpolation on the log of given values to obtain see table 4.1
intermediate points at the wavelength intervals
desired.
5.3.2 Calculations P
The calculation of source hazard values shall be P
performed by weighting the spectral scan by the
appropriate function and calculating the total
weighted energy.
5.3.3 Measurement uncertainty P
The quality of all measurement results must be see Annex C in the norm P
quantified by an analysis of the uncertainty.

6 LAMP CLASSIFICATION P
For the purposes of this standard it was decided that see table 6.1 P
the values shall be reported as follows:
– for lamps intended for general lighting service, N/A
the hazard values shall be reported as either
irradiance or radiance values at a distance which
produces an illuminance of 500 lux, but not at a
distance less than 200 mm
– for all other light sources, including pulsed lamp N/A
sources, the hazard values shall be reported at a
distance of 200 mm
6.1 Continuous wave lamps P
6.1.1 Exempt Group P
Page 10 of 22 Report No.:3150235.50A

IEC 62471
Clause Requirement + Test Result – Remark Verdict

In the exempt group are lamps, which does not pose P


any photobiological hazard. The requirement is met
by any lamp that does not pose:
– an actinic ultraviolet hazard (ES) within 8-hours P
exposure (30000 s), nor
– a near-UV hazard (EUVA) within 1000 s, (about P
16 min), nor
– a retinal blue-light hazard (LB) within 10000 s P
(about 2,8 h), nor
– a retinal thermal hazard (LR) within 10 s, nor P
– an infrared radiation hazard for the eye (EIR) P
within 1000 s
6.1.2 Risk Group 1 (Low-Risk) N/A
In this group are lamps, which exceeds the limits for N/A
the except group but that does not pose:
– an actinic ultraviolet hazard (ES) within 10000 s, N/A
nor
– a near ultraviolet hazard (EUVA) within 300 s, nor N/A
– a retinal blue-light hazard (LB) within 100 s, nor N/A
– a retinal thermal hazard (LR) within 10 s, nor N/A
– an infrared radiation hazard for the eye (EIR) N/A
within 100 s
Lamps that emit infrared radiation without a strong N/A
visual stimulus and do not pose a near-infrared
retinal hazard (LIR), within 100 s are in Risk Group 1.
6.1.3 Risk Group 2 (Moderate-Risk) N/A
This requirement is met by any lamp that exceeds N/A
the limits for Risk Group 1, but that does not pose:
– an actinic ultraviolet hazard (ES) within 1000 s N/A
exposure, nor
– a near ultraviolet hazard (EUVA) within 100 s, nor N/A
– a retinal blue-light hazard (LB) within 0,25 s N/A
(aversion response), nor
– a retinal thermal hazard (LR) within 0,25 s N/A
(aversion response), nor
– an infrared radiation hazard for the eye (EIR) N/A
within 10 s
Lamps that emit infrared radiation without a strong N/A
visual stimulus and do not pose a near-infrared
retinal hazard (LIR), within 10 s are in Risk Group 2.
6.1.4 Risk Group 3 (High-Risk) N/A
Lamps which exceed the limits for Risk Group 2 are N/A
in Group 3.
Page 11 of 22 Report No.:3150235.50A

IEC 62471
Clause Requirement + Test Result – Remark Verdict

6.2 Pulsed lamps N/A


Pulse lamp criteria shall apply to a single pulse and N/A
to any group of pulses within 0,25 s.
A pulsed lamp shall be evaluated at the highest N/A
nominal energy loading as specified by the
manufacturer.
The risk group determination of the lamp being N/A
tested shall be made as follows:
– a lamp that exceeds the exposure limit shall be N/A
classified as belonging to Risk Group 3
(High-Risk)
– for single pulsed lamps, a lamp whose weighted N/A
radiant exposure or weighted radiance does is
below the EL shall be classified as belonging to
the Exempt Group
– for repetitively pulsed lamps, a lamp whose N/A
weighted radiant exposure or weighted radiance
dose is below the EL, shall be evaluated using
the continuous wave risk criteria discussed in
clause 6.1, using time averaged values of the
pulsed emission
Page 12 of 22 Report No.:3150235.50A

IEC 62471
Clause Requirement + Test Result – Remark Verdict

Table 4.1 Spectral weighting function for assessing ultraviolet hazards for skin and eye
Wavelength1 UV hazard function Wavelength UV hazard function
λ, nm SUV(λ) λ, nm SUV(λ)
200 0,030 313* 0,006
205 0,051 315 0,003
210 0,075 316 0,0024
215 0,095 317 0,0020
220 0,120 318 0,0016
225 0,150 319 0,0012
230 0,190 320 0,0010
235 0,240 322 0,00067
240 0,300 323 0,00054
245 0,360 325 0,00050
250 0,430 328 0,00044
254* 0,500 330 0,00041
255 0,520 333* 0,00037
260 0,650 335 0,00034
265 0,810 340 0,00028
270 1,000 345 0,00024
275 0,960 350 0,00020
280* 0,880 355 0,00016
285 0,770 360 0,00013
290 0,640 365* 0,00011
295 0,540 370 0,000093
297* 0,460 375 0,000077
300 0,300 380 0,000064
303* 0,120 385 0,000053
305 0,060 390 0,000044
308 0,026 395 0,000036
310 0,015 400 0,000030
1
Wavelengths chosen are representative: other values should be obtained by logarithmic interpolation at
intermediate wavelengths.
* Emission lines of a mercury discharge spectrum.
Page 13 of 22 Report No.:3150235.50A

IEC 62471
Clause Requirement + Test Result – Remark Verdict

Table 4.2 Spectral weighting functions for assessing retinal hazards from broadband optical
sources
Wavelength Blue-light hazard function Burn hazard function
nm B (λ) R (λ)
300 0,01
305 0,01
310 0,01
315 0,01
320 0,01
325 0,01
330 0,01
335 0,01
340 0,01
345 0,01
350 0,01
355 0,01
360 0,01
365 0,01
370 0,01
375 0,01
380 0,01 0,1
385 0,013 0,13
390 0,025 0,25
395 0,05 0,5
400 0,10 1,0
405 0,20 2,0
410 0,40 4,0
415 0,80 8,0
420 0,90 9,0
425 0,95 9,5
430 0,98 9,8
435 1,00 10,0
440 1,00 10,0
445 0,97 9,7
450 0,94 9,4
455 0,90 9,0
460 0,80 8,0
465 0,70 7,0
470 0,62 6,2
475 0,55 5,5
480 0,45 4,5
485 0,40 4,0
490 0,22 2,2
495 0,16 1,6
500-600 10[(450-λ)/50] 1,0
600-700 0,001 1,0
700-1050 10[(700-λ)/500]
1050-1150 0,2
1150-1200 0,2 100,02(1150-λ)
.

1200-1400 0,02
Page 14 of 22 Report No.:3150235.50A

IEC 62471
Clause Requirement + Test Result – Remark Verdict

Table 5.4 Summary of the ELs for the surface of the skin or cornea (irradiance based values)
Wavelength Exposure Limiting EL in terms of
Hazard
Relevant equation range duration aperture constant irradiance
Name
nm sec rad (deg) W•m-2
Actinic UV
ES = ∑Eλ • S(λ) • ∆λ 200 – 400 < 30000 1,4 (80) 30/t
skin & eye
≤1000 10000/t
Eye UV-A EUVA = ∑Eλ • ∆λ 315 – 400 1,4 (80)
>1000 10
Blue-light EB = ∑Eλ • B(λ) • ∆λ ≤100 100/t
small source 300 – 700 < 0,011
>100 1,0
≤1000 18000/t 0,75
Eye IR EIR = ∑Eλ • ∆λ 780 –3000 1,4 (80)
>1000 100
Skin thermal EH = ∑Eλ • ∆λ 380 – 3000 < 10 2π sr 20000/t 0,75

Table 5.5 Summary of the ELs for the retina (radiance based values)
Wavelength Exposure EL in terms of
Field of view
Hazard Name Relevant equation range duration constant radiance
radians
nm sec W•m-2•sr-1)
0,25 – 10 0,011•√(t/10) 106/t
10-100 0,011 106/t
Blue light LB = ∑Lλ • B(λ) • ∆λ 300 – 700
100-10000 0,0011•√t 106/t
≥ 10000 0,1 100
Retinal < 0,25 0,0017 50000/(α•t 0,25)
thermal LR = ∑Lλ • R(λ) • ∆λ 380 – 1400
0,25 – 10 0,011•√(t/10) 50000/(α•t 0,25)
Retinal
thermal
LIR = ∑Lλ • R(λ) • ∆λ 780 – 1400 > 10 0,011 6000/α
(weak visual
stimulus)
Page 15 of 22 Report No.3150235.50A

IEC 62471
Clause Requirement + Test Result – Remark Verdict

Table 6.1 Emission limits for risk groups of continuous wave lamps (BN108C LED28 CW L1200, α= 100 mrad ) P
Test condition: GLS Non GLS Worst Case
Lamp classification group: exempt risk 1 risk 2 risk 3
Emission Measurement
Action
Risk Symbol Units Exempt Low risk Mod risk
spectrum
Limit Result Limit Result Limit Result

Actinic UV SUV(λ) Es W•m-2 0,001 0,0000 0,003 0,03

Near UV EUVA W•m-2 10 0,0000 33 100

Blue light B(λ) LB W•m-2•sr-1 100 26,87 10000 4000000

Blue light,
B(λ) EB W•m-2 1,0* -- 1,0 400
small source

Retinal
R(λ) LR W•m-2•sr-1 28000/α 429,32 28000/α 71000/α
thermal

Retinal
thermal,
R(λ) LIR W•m-2•sr-1 6000/α -- 6000/α 6000/α
weak visual
stimulus**

IR radiation,
EIR W•m-2 100 0,05 570 3200
eye

* Small source defined as one with α < 0,011 radian. Averaging field of view at 10000 s is 0,1 radian.
** Involves evaluation of non-GLS source

TRF No. IEC62471A


Page 16 of 22 Report No.3150235.50A

IEC 62471
Clause Requirement + Test Result – Remark Verdict

Table 6.1 Emission limits for risk groups of continuous wave lamps (BN108C LED36 CW L1200 MB, α= 100 mrad ) P
Test condition: GLS Non GLS Worst Case
Lamp classification group: exempt risk 1 risk 2 risk 3
Emission Measurement
Action
Risk Symbol Units Exempt Low risk Mod risk
spectrum
Limit Result Limit Result Limit Result

Actinic UV SUV(λ) Es W•m-2 0,001 0,0000 0,003 0,03

Near UV EUVA W•m-2 10 0,0000 33 100

Blue light B(λ) LB W•m-2•sr-1 100 24,45 10000 4000000

Blue light,
B(λ) EB W•m-2 1,0* -- 1,0 400
small source

Retinal
R(λ) LR W•m-2•sr-1 28000/α 511,68 28000/α 71000/α
thermal

Retinal
thermal,
R(λ) LIR W•m-2•sr-1 6000/α -- 6000/α 6000/α
weak visual
stimulus**

IR radiation,
EIR W•m-2 100 0,04 570 3200
eye

* Small source defined as one with α < 0,011 radian. Averaging field of view at 10000 s is 0,1 radian.
** Involves evaluation of non-GLS source

TRF No. IEC62471A


Page 17 of 22 Report No.:3150235.50A

Furthermore remarks:

Appenix 1: List of test equipment used:

Clause Measurement/ Registration Testing/measuring


testing Number equipment/material used

5 Irradiance measurements SH 344 MONOCHROMATOR


Radiance measurements
5 Radiance measurements SH 345 S009 TELESCOPE
5 Irradiance measurements SH 346 S400_417 DETECTION
ELECTRONIC
5 Irradiance measurements SH 347 608 CONSTANT
Radiance measurements CURRENT

5 Radiance measurements SH 348 SRS12 RADIANCE

5 Irradiance measurements SH 349 705 DEUTERIUM SUPPLY

5 Irradiance measurements SH 350 CL6 STANDARD

5 Irradiance measurements SH 351 CL7 STANDARD

5 Irradiance measurements SH 352 PHOTOMULTIPLIER


Radiance measurements
5 Irradiance measurements SH 353 INGAAS DETECTOR
Radiance measurements
5 Irradiance measurements SH 354 SILICON DETECTOR
Radiance measurements
5 Irradiance measurements SH 355 PBS-TE DETECTOR

5 Irradiance measurements SH 356 RELAY OPTIC

5 Irradiance measurements SH 357 D8 INTEGRATING SPHER


Radiance measurements
5 Irradiance measurements SH 358 D7 COSINE DIFFUSER

5 Irradiance measurements SH 359 PHOTOMETRIC


DETECTOR

5 Irradiance measurements SH070 WATTMETER


Radiance measurements

TRF No. IEC62471A


Page 18 of 22 Report No.:3150235.50A

Appendix 2: Photo documentation

Overview of BN108C LED28 CW L1200

Overview of BN108C LED36 CW L1200 MB

TRF No. IEC62471A


Page 19 of 22 Report No.:3150235.50A

Appendix 3: Model list

No. Model Power Lamp Ballast/LED driver

1 BN108C LED15 WW L1200

2 BN108C LED15 NW L1200 15W 12×0.3W/LED module UE 16W 200mA

3 BN108C LED15 CW L1200

4 BN108C LED28 WW L1200

5 BN108C LED28 NW L1200 28W 22×0.3W/LED module PLS30-400-FI-LINEAR

6 BN108C LED28 CW L1200

7 BN108C LED15 WW L600

8 BN108C LED15 NW L600 15W 22×0.3W/LED module UE 16W 200mA

9 BN108C LED15 CW L600

10 BN108C LED8 WW L600

11 BN108C LED8 NW L600 8W 12×0.3W/LED module Mad 10W 200mA

12 BN108C LED8 CW L600

BN108C LED20 WW L1200


13
PSD

BN108C LED20 NW L1200 12×0.45W/LED


14 23W
PSD module

BN108C LED20 CW L1200


15
PSD

BN108C LED28 WW L1200 Xitanium 36W


16
PSD 0.12-0.4A 110V TD
230V
BN108C LED28 NW L1200
17
PSD

30W 22×0.3W/LED module

BN108C LED28 CW L1200


18
PSD

TRF No. IEC62471A


Page 20 of 22 Report No.:3150235.50A

BN108C LED15 WW L600


19
PSD

BN108C LED15 NW L600 16W 22×0.3W/LED module


20
PSD

BN108C LED15 CW L600


21
PSD

BN108C LED10 WW L600


22
PSD

BN108C LED10 NW L600 12×0.45W/LED


23 13W
PSD module

BN108C LED10 CW L600


24
PSD

BN108C LED20 WW L1200


25
PSR

BN108C LED20 NW L1200 12×0.45W/LED


26 23W
PSR module

BN108C LED20 CW L1200


27
PSR

BN108C LED28 WW L1200


28
PSR

BN108C LED28 NW L1200


29 30W 22×0.3W/LED module
PSR

BN108C LED28 CW L1200


30
PSR Xitanium 36W
0.12-0.4A 100V 1-10V
BN108C LED15 WW L600 230V
31
PSR

BN108C LED15 NW L600


32 16W 22×0.3W/LED module
PSR

BN108C LED15 CW L600


33
PSR

BN108C LED10 WW L600


34
PSR

BN108C LED10 NW L600 12×0.45W/LED


35 13W
PSR module

BN108C LED10 CW L600


36
PSR

BN108C LED20 WW L1200


37
MB
LED driver 20W
BN108C LED20 NW L1200 12×0.45W/LED
38 22W 280mA 69V 230V 20W
MB module
PCB
BN108C LED20 CW L1200
39
MB

TRF No. IEC62471A


Page 21 of 22 Report No.:3150235.50A

BN108C LED36 WW L1200


40
MB

BN108C LED36 NW L1200 22×0.45W/LED Xitanium 40W 280mA


41 40W
MB module 142V 230V

BN108C LED36 CW L1200


42
MB

BN108C LED18 WW L600


43
MB
LED driver 20W
BN108C LED18 NW L600 22×0.45W/LED
44 21W 280mA 69V 230V 20W
MB module
PCB
BN108C LED18 CW L600
45
MB

BN108C LED10 WW L600


46
MB

BN108C LED10 NW L600 12×0.45W/LED


47 13W PLS12-140-FI-LINEAR
MB module

BN108C LED10 CW L600


48
MB

BN108C LED36 WW L1200


49
PSD MB
Xitanium 75W
BN108C LED36 NW L1200 22×0.45W/LED
50 40W 0.12-0.4A 215V TD
PSD MB module
230V
BN108C LED36 CW L1200
51
PSD MB

BN108C LED18 WW L600


52
PSD MB
Xitanium 36W
BN108C LED18 NW L600 22×0.45W/LED
53 21W 0.12-0.4A 110V TD
PSD MB module
230V
BN108C LED18 CW L600
54
PSD MB

BN108C LED36 WW L1200


55
PSR MB

BN108C LED36 NW L1200 22×0.45W/LED Xitanium 75W 0.2-0.4A


56 40W
PSR MB module 200V 1-10V 230V

BN108C LED36 CW L1200


57
PSR MB

BN108C LED18 WW L600


58
PSR MB
Xitanium 36W
BN108C LED18 NW L600 22×0.45W/LED
59 21W 0.12-0.4A 100V 1-10V
PSR MB module
230V
BN108C LED18 CW L600
60
PSR MB

TRF No. IEC62471A


Page 22 of 22 Report No.:3150235.50A

Appendix 4: Relative spectrum of tested sample

BN108C LED28 CW L1200

BN108C LED36 CW L1200 MB

TRF No. IEC62471A

You might also like