Time Electronics 5068 Insulation Tester Calibrator User Manual
Time Electronics 5068 Insulation Tester Calibrator User Manual
Time Electronics 5068 Insulation Tester Calibrator User Manual
5068 InsCal
Insulation Tester Calibrator
Version 2.5
11-22
This manual provides operating and safety instructions for the Time Electronics product.
To ensure correct operation and safety, please follow the instructions in this manual.
Time Electronics reserves the right to change the contents, specifications and other
information contained in this manual without notice.
Time Electronics
5068 Insulation Tester Calibrator User Manual
Contents
1 General Description....................................................................................................................... 4
2 Supplied Items ............................................................................................................................... 5
3 Specifications ................................................................................................................................ 6
3.1 Technical Specification .......................................................................................................................... 6
3.2 General Specification ............................................................................................................................ 6
3.3 Ordering Information .............................................................................................................................. 6
4 Controls .......................................................................................................................................... 7
4.1 Descriptions of Controls ........................................................................................................................ 7
5 Operation........................................................................................................................................ 8
5.1 Operational Guidelines .......................................................................................................................... 8
5.2 Typical Operation Procedure ................................................................................................................. 9
5.2.1 Initial Setup ........................................................................................................................................................ 9
5.2.2 Voltage Measurement...................................................................................................................................... 10
5.2.3 Current Measurement ...................................................................................................................................... 11
5.2.4 Insulation Resistance Verification .................................................................................................................... 12
5.2.5 Example Connection........................................................................................................................................ 13
5.2.6 Low Resistance Verification............................................................................................................................. 13
3
Time Electronics
5068 Insulation Tester Calibrator User Manual
1 General Description
This 5068 provides all the equipment necessary to test and calibrate insulation test sets and
megohm meters. It consists of:
4
Time Electronics
5068 Insulation Tester Calibrator User Manual
2 Supplied Items
5068 InsCal Insulation Tester Calibration System
Accessories supplied:
Test Lead Wallet Low resistance test leads 2.5 kV test leads Battery charger
5
Time Electronics
5068 Insulation Tester Calibrator User Manual
3 Specifications
3.1 Technical Specification
Insulation Resistance
Accuracy ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±5%
Switches are a purpose-built design, with break-before-make, silver plated, self-cleaning contacts.
Low Resistance
4 spot values ............... 1 Ω, 10 Ω, 100 Ω and 1 kΩ, 1 %, 0.5 W, TC 50 ppm.
Volt-Meter
Low Range ................. 0 to 1.999 kV, resolution 1 V; 1 % of full-scale reading.
High Range ................. 0 to 2.5 kV, resolution 10 V; 1 % of full-scale reading.
At 2.5 kV the humidity limit should be below 50%.
Terminal impedance ... 1 GΩ.
Milli-Amp Meter
Low Range ................. 0 to 2 mA; 1 % of full-scale reading.
High Range ................. 0 to 20 mA; 1 % of full-scale reading.
Terminal impedance ... 447 Ω (this includes the discharge resistor for short circuit test).
6
Time Electronics
5068 Insulation Tester Calibrator User Manual
4 Controls
V/I meter Function selector and
Voltage/Current V or I range
charging insulation resistance
meter selection button
socket setting knob
V/I meter
On/Off button
7
Time Electronics
5068 Insulation Tester Calibrator User Manual
5 Operation
5.1 Operational Guidelines
Notes:
1. UUT: stands for Unit Under Test (i.e. the insulation tester or the megohm meter).
2. Wherever possible use the manufacturer of the UUT’s supplied test leads.
The 5068 can measure test voltages generated by a UUT up to 2.5 kV.
Be aware that the test voltage for high resistance measurements is dangerous and can
cause electric shock!
Always remove the battery charger leads from both the 5068 and the UUT before
starting the calibration setup.
Use only the measuring leads supplied with the UUT or those supplied with the 5068.
When checking high resistance values the leads may pick up electrical interference,
which can cause a disturbance to the readings obtained. Some manufactures fit a
guard connection for use with dedicated leads, this will stabilise the readings.
It is not recommended to operate the resistance dial switch while a test is made. This
will not automatically do any damage but may cause arcing and create transients which
could give false readings on the UUT display due to the fact that the switch is break-
before-make between each resistor. This will ensure that contact arc is suppressed
with high relative humidity.
Avoid opening the 5068 case lid in the rain or in steamy conditions for the moisture
will impair the readings.
Observe safety at all times and if possible, carry out the tests in a low humidity
environment.
Note: Insulation testers will vary, and the controls may be different than those referenced in this manual.
However, the testing principles should still apply.
8
Time Electronics
5068 Insulation Tester Calibrator User Manual
3) Connect the selected test leads to the 4 mm recessed sockets of the 5068.
NOTE: If the UUT’s own test leads are available, use them as first choice.
9
Time Electronics
5068 Insulation Tester Calibrator User Manual
2) On the 5068, select the O/C VOLTS function using the rotator knob.
3) Using range selection button under the display, choose the required
voltage range of 1.999 kV or 2.50 kV.
4) Using the selected high voltage test leads, connect the 5068 to the UUT, observing their
polarities.
6) When the reading has stabilised, record the O/C voltage reading on the 5068 voltmeter.
10
Time Electronics
5068 Insulation Tester Calibrator User Manual
1) On the 5068, select the S/C CURRENT function using the rotator knob.
2) Using range selection button under the display, choose the required
current range: 2 mA or 20 mA.
3) Press the test button on the UUT, and record the current reading on the 5069 display.
Ensure the output of the UUT is turned off after test.
4) Return the 5068 setting to O/C Voltage, and repeat the Voltage Measurement
procedure above (5.2.2). This is to check the voltage has recovered. Switch off the
output from the UUT.
11
Time Electronics
5068 Insulation Tester Calibrator User Manual
1. Using the rotator knob of 5068, select the required Resistance Spot value.
2. Note the allowed voltage of the spot value resistances shown on the 5068 panel.
Ensure that the voltage set on the UUT does not exceed this voltage rating.
Important Note: Wherever possible use the test leads supplied with the UUT.
5) Press the test button on the UUT. Please note on the higher resistance ranges, due to
the very low currents present, the display on the UUT may fluctuate and take time to
settle, please follow any precautions given by the tester manufacturer.
6) Proceed and record all test results, noting the allowed errors of the 5068 and also for
that of the UUT.
12
Time Electronics
5068 Insulation Tester Calibrator User Manual
Please note that the power dissipation of these resistors is limited to 0.5 W.
13
Time Electronics
5068 Insulation Tester Calibrator User Manual
14
Time Electronics
5068 Insulation Tester Calibrator User Manual
Returning Instruments
Prior to returning your product please contact Time Electronics. We will issue a return
merchandise authorization (RMA) number that is to accompany the goods returning.
Further instructions will also be issued prior to shipment. When returning instruments, please
ensure that they have been adequately packed, preferably in the original packing supplied.
Time Electronics Ltd will not accept responsibility for units returned damaged.
Please ensure that all units have details of the service required and all relevant paperwork.
Email: mail@timeelectronics.co.uk
Web Site: www.timeelectronics.com
1. When this crossed-out wheeled bin symbol is attached to a product it means the product is
covered by the European Directive 2002/96/EC.
2. All electrical and electronic products should be disposed of separately from the municipal
waste stream via designated collection facilities appointed by the government or the local
authorities.
3. The correct disposal of your old appliance will help prevent potential negative consequences
for the environment and human health.
4. For more detailed information about disposal of your old appliance, please contact your city
office, waste disposal service or return to Time Electronics.