Congratulations on your purchase of the Forney CO-PILOT digital force indicator. This
instrument is designed specifically for materials testing and incorporates many features to
streamline that testing process. Please take a few minutes to read through this manual and
get acquainted with the many time-saving features of your CO-PILOT.
1.1 Features
Ø Compression testing
Ø 2 Force channels (in order to monitor 2 test frames)
Ø 7 digit display range
Ø Display live or peak force
Ø Display force rate
Ø Selectable units: in/lbf, mm/KN or mm/Kgf
Ø Store up to 500 tests
Ø Automatic Test Store
Ø Automatic Test ID increments
Ø English or Spanish language
Ø 16-Key keypad for easy setup and operation
Ø Large 16-Character backlit LCD display
Ø RS232 serial port
Ø Battery-backed clock with time and date
2 Installation
2.1 Make certain that you have received these items:
Ø CO-PILOT
Ø AC adapter
Ø AC power cord
Ø 2 Mounting knobs, 5/16”-18 x ¾”lg.
Ø 2 Rubber washers
3 Rear Panel
3.1 ON/OFF Switch
4 Front Panel
4.1 16-Character LCD display.
__0__ Zero the Force channel. Note: only the channel currently being displayed is
ZERO zeroed.
__7__ Toggle the display between Force and Force rate. This key is ignored when
FORCE peak Force is displayed.
5 Operation
5.1 Power Up.
When switched on, the CO-PILOT briefly displays the model number and software version
followed by the date. The four LED mode indicators also flash briefly. It then restores the
test frame and settings from the most recent setup. The active test frame is shown by the
mode indicators and force is displayed. CO-PILOT is now in “live” display mode.
In general, the steps to set up a test are:
Ø choose a test frame and set preferences from the UTILITIES menu
Ø press New key
When repeated tests are run, and Auto-Store is turned on, pressing the New key is all that is
necessary to start another test.
See section 6 for more details on how to set up and run a test.
Press the key to scroll through menu options , then press to make your selection. To
exit the UTILITIES menu you can press 3/UTIL from any of the menu options or scroll to
EXIT and press .
Selects test frame 1 or 2. To select a test frame press , then or . Then press
to return to the utility menu. Press to advance through the utility menu, or 3/UTIL to
exit and return to “live” mode.
Selects the units - in/lbf, mm/kN, or mm/Kgf. To select a units press , then or .
Then press to return to the utility menu. Press to advance through the utility menu,
or 3/UTIL to exit and return to “live” mode.
- 8 - Forney Incorporated
Auto-ID press , then (on) or (off). Then press to return to the utility menu.
Press to advance through the utilities menu, or 3/UTIL to exit and return to “live” mode.
ON. To set Auto-Save press , then (on) or (off). Then press to return to
the utilities menu. Press to advance through the utilities menu, or 3/UTIL to exit and
return to “live” mode.
Set the clock date. To enter the date press , then enter the date with the numeric keys.
The format is mm-dd-yy. Do NOT press again! This will clear the date and you will
have to re-enter it. Press to advance through the utilities menu, or 3/UTIL to exit and
return to “live” mode.
Set the clock time. To enter the time press , then enter the time with the numeric keys.
The format is hh:mm:ss. Do NOT press again! This will clear the time and you will
have to re-enter it. Press to advance through the utilities menu, or 3/UTIL to exit and
return to “live” mode.
To adjust display resolution press . Press to scroll through the transducer options,
then press to make your selection.
Press and to choose a resolution, then press to return to the transducer options
menu. You must select EXIT from the transducer options menu to return to the
RESOLUTION + menu. Then press to advance through the utilities menu, or 3/UTIL to
exit and return to “live” mode.
Forney Incorporated - 9
To select a sample rate press , and or . Then press to return to the utilities
menu. Press to advance through the utilities menu, or 3/UTIL to exit and return to “live”
mode.
Note: When the live force display is unstable, choose a low sample rate and the highest
possible resolution as shown in paragraph 5.2.7, to stabilize the display. The AC electrical
power feeding the testing machine may also cause display instability. The AC electrical
power should be observed using an AC voltmeter while the machine is in operation to
observe the electrical power stability.
To access the FILES + menu press . Press to advance through menu options, then
to make your selection.
If SEND TO FILE is selected the transfer begins immediately. Upon completion CO-PILOT
returns to the FILES + menu.
Hit the start button to close this window. Hyper Terminal is now ready to receive data from
the Co-Pilot.
Select “Send to File” from the Files menu of the Co-Pilot.
Note the Hyper Terminal screen is full of data – if the screen remains blank, the download
was unsuccessful.
From the Hyper Terminal pull-downs, go into transfer – capture text – stop.
At this point, the Hyper Terminal window can be closed or minimized.
Open a session of Microsoft® Excel. Open the file just created above, which should have a
.CSV extension on it. Note the data is imported, and should look similar to the example
below:
Test Results
If ERASE ALL + is selected a prompt ‘Erase All? No’ is displayed, where ‘No’ is the
default option. To abandon this action press or . To clear all stored tests press to
change the option from ‘No’ to ‘Yes’, then press or . This process may take a few
seconds.
When the operation is complete the message ‘Done. Press key’ is displayed. Press or
to return to the FILES + menu. Press to advance through the utilities menu, or
3/UTIL to exit and return to “live” mode.
Calibrations are password protected. At the ‘Passwrd’ prompt, Enter the password (call
Forney tech support for the password), then press . If you enter an incorrect password
CO-PILOT returns to the CALIBRATION menu.
WARNING: Entering the calibration function without the proper calibration
equipment, procedures and training will corrupt the existing calibration rendering the
CO-PILOT out of calibration and not capable of meeting ASTM testing requirements.
2. At the prompt “Cap FS”. Press , enter the capacity (full scale range) of the
device in Lbf, kN or Kg as indicated, then press again. Press to advance to
the next step.
5. At the prompt “Value?” press to accept the default value of 0.00000. This is the
applied force associated with electrical zero stored in the previous step.
6. At the prompt “At Point2?” apply load to the proving device to obtain the first
calibration force, then press . The display will automatically advance.
7. At the prompt “Value?” press , enter the applied force obtained on the proving
device, then press again. Note: if you need to re-enter this value just press
again and repeat this step. Press to advance to the next step.
8. Repeat steps 6 to 7 until you have entered all calibration points.
9. At the prompt “Done. Save? Yes”, press to store the calibration, or press and
change your option to “No”, then press to abandon the calibration and restore the
previous one. CO-PILOT returns to the CALIBRATION + menu.
10. Press to advance to the next transducer or to EXIT, then press . Press to
advance through the utilities menu, or 3/UTIL to exit and return to “live” mode.
Exit the UTILITIES menu. Press to exit the UTILITIES menu and return to live display
mode or to advance through the utilities menu. You may also press 3/UTIL to exit and
return to “live” mode.
6. Proceed with applying a load to the specimen per ASTM C39. When the load
exceeds 0.25% of capacity, data logging and break detection automatically engage.
7. Continue loading the specimen until break is detected (load decreases to 50% of
peak).
store the results, or then to proceed without saving. After making your
selection the Peak Load indicator lights, and peak load is displayed.
9. Note: you may store results at any time up to the start of a new test or until the CO-
PILOT is switched off. To store test results at a later time Press 5/STORE.
11. If Auto-Save is set to on, results (peak load) are automatically calculated and stored,
the Peak Load indicator lights, and peak load is displayed. Press 9/Peak to return to
the live display mode.
Forney Incorporated - 13
7 Specifications
Part Number TA-1252
Load Channels
No. of channels 2
Input type +/- 2mV/V to +/-4mV/V , 350 to1000 Ohm, Full Bridge
Display Range (Lbf) +9,999,950 / -999,950
Accuracy +/-.5% of reading from 1% of Full Scale to Full Scale
Linear Correction 5 (0 + 4) points
Resolution, A/D Counts 200,000 (note 1)
Display Resolution Selectivity 1 – 5 – 10 Least digit; User selectable
Load Units, Load Units Always displayed
Load Units, Lb Yes
Load Units, Kn Yes
Load Units, Kg Yes
Load Rate Units, Load Rate Units Toggle for Display
Load Rate Units, Kg/sec Yes
Load Rate Units, Lb/sec Yes
Load Rate Units, Kn/sec Yes
Data Options
Send stored tests (RS232) Yes
Display Date/Time Yes
Display Peak Load Yes
Auto Store Yes (optional)
Tests Stored in Memory 500
Specimen I.D. Yes, 6-digit numeric
Auto-Increment Specimen I.D. Yes (optional)
Sample Break/Peak Sensitivity Preset to 50% of Peak Load
Auto Data Log Start Sensitivity 0.25% of Channel Capacity
Date/Time Stamp Tests Yes
Negative Load Display Yes
Capacity Range 7 Digits
Calibration Units All Available
Languages English + Spanish
Function Indicator Lights Yes- 4 indicators
Password Protection Yes
Water/dust Resistant Case Yes
Simple Operation/Calibration Yes
Easy Channel Selection Yes
- 14 - Forney Incorporated
8 Menu Tree
Ø UTIILITIES+
Ø TEST FRAME Ø TEST FRAME1
Ø TEST FRAME2
Ø in, lbf
Ø UNITS Ø mm,kN
Ø mm,Kgf
Ø On
Ø AUTO-ID Ø Off
Ø On
Ø AUTO SAVE Ø Off
Ø 01/01/05
Ø DATE
Ø 08:45:00
Ø TIME
Ø LOAD1 Ø 5, 10 or 50
Ø RESOLUTION+ Ø LOAD2
Ø EXIT Ø Exit
Ø 10
Ø SAMPLES/SEC Ø 20
Ø 30
Ø 60
Ø SEND TO FILE
Ø FILES+ Ø ERASE ALL
Ø EXIT
Ø LOAD1 Ø Capacity:
Ø CALIBRATION+ Ø LOAD2 Ø Points(2-5)?
Ø EXIT Ø At zero- #1?
Ø Value?
Ø At Point 2?
Ø Value?
Ø At Point 3,4?
Ø Value?
Ø Save?
Ø Yes
Ø No
Ø English
Ø ENG/ESP Ø Espanol
Ø EXIT