Q-Exactive-GC Preinstallation Guide
Q-Exactive-GC Preinstallation Guide
Q-Exactive-GC Preinstallation Guide
Mass Spectrometer
Preinstallation Requirements Guide
The following are registered trademarks in the United States and possibly other countries:
ShockWatch is a registered trademark of Shockwatch, Inc. Sigma and Sigma-Aldrich are registered
trademarks of Sigma-Aldrich Biotechnology. Swagelok is a registered trademark of Swagelok
Company. Teflon is a registered trademark of E. I. du Pont de Nemours & Co. Tygon is a registered
trademark of Norton Company. Oerlikon Leybold Vacuum is a registered trademark of OC Oerlikon
Corporation AG. SOGEVAC is a registered trademark of Oerlikon Leybold Vacuum. Dranetz is a
registered trademark of Dranetz Technologies, Inc.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use
in the product operation. This document is copyright protected and any reproduction of the whole or
any part of this document is strictly prohibited, except with the written authorization of Thermo
Fisher Scientific Inc.
Release History: Revision A released in June 2015. Revision B released in July 2015
1. Yes No All laboratory remodeling has been completed and the space available is sufficient to meet the minimum requirements for
the configuration ordered? The floor is certified to meet the load requirements of the system?
2. Yes No Your instrument has been delivered and is either in the laboratory or can be delivered immediately on the arrival of the
installation engineer?
3. Yes No The key operator will be available during the installation period. The person with the authority to accept the instrument at
the end of the installation will also be available to sign the required acceptance document?
Please provide the names of these individuals: ___________________________________________________________
4. Yes No The entrance to the laboratory and the route from the loading dock are at least 90 cm (36 in.) wide with additional space at
corners?
5. Yes No Sufficient bench space is available for all of the equipment? List the following:
Width: ________________________, Depth: ________________________, Height: _______________________
6. Yes No Workbench can support the load of the system including optional equipment and is free from vibration?
7. Yes No Lighting is adequate?
8. Yes No Floor vibrations and electromagnetic interferences are below the specified levels?
9. Yes No Main power is installed and in compliance with local electrical codes?
10. Yes No The power outlets are of the correct configuration?
11. Yes No The electrical power has been measured?
Please note voltages: _______________ Volts AC input to ground.
Please note voltages: _______________ Volts neutral to ground.
Please note voltages: _______________ Volts AC input to neutral.
12. Yes No Power is free from fluctuations due to slow changes in the average voltage or changes due to surges, sags, or transients?
13. Yes No Air conditioning is adequate for temperature, humidity, and particulate matter control? The laboratory can be maintained
at a constant temperature, between 15 and 26 °C (59 and 78 °F)?
14. Yes No The relative humidity is between 40% and 70%, with no condensation?
15. Yes No The system work area is free from magnetic disruption and electrostatic discharge?
16. Yes No All gases required are on site, gas lines are installed, and appropriate gas regulators are available?
List gases and purity: _________________________________________________________________
17. Yes No There is a functional telephone close to the system? Phone number ______________________________
18. Yes No All relevant local safety regulations have been met and the equipment installed will not affect compliance?
19. Yes No All required chemicals and equipment for installing the system are on site?
20. Yes No Have any special acceptance specifications been agreed within the contract?
If YES, please attach full details of specification.
21. Yes No Is there any additional equipment that needs to be interfaced for the system?
If YES, please supply details.
I, the undersigned, confirm that the site requirements as stated above have been accomplished and the laboratory is prepared for the
installation of the Thermo Scientific Exactive Series instrument. I understand that I may be liable for a Field Service Representatives’
travel or lodging expenses if they are unable to carry out the installation on the pre-scheduled date due to insufficient lab preparation.
If circumstances warrants, Thermo Fisher Scientific will make every effort to reschedule an installation as soon as possible with the
next available representative.
Signed: _______________________________________ Print Name: ___________________________________________
Company name: ________________________________ Email: _______________________________________________
Date: _________________________________________ Phone: _______________________________________________
Fax to: Attn: Local Service Engineer
Note After we receive this checklist, your local Field Service Representative will contact you to schedule installation. ▲
Assistance
For technical support and ordering information, please visit:
gc-gcms-customersupport.com/WebPage/Share/Default.aspx
www.unitylabservice.com
www.thermoscientific.com
Technical Documentation
Thermo Fisher Scientific
2215 Grand Avenue Parkway
United States
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) iii
Contents
Index......................................................................................... I-1
Contents
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 1-1
Using This Manual
About This Manual
1-2 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Using This Manual
Typographical Conventions
Typographical Conventions
This section describes typographical conventions that have been
established for Thermo Fisher Scientific manuals.
Signal Word
Make sure you follow the precautionary statements presented in this
manual. The special notices appear different from the main flow of text:
Viewpoint Orientation
The expressions left and right used in this manual always refer to the
viewpoint of a person that is facing the front side of the instrument.
Data Input
Throughout this manual, the following conventions indicate data input
and output via the computer:
• For brevity, expressions such as “choose File > Directories” are used
rather than “pull down the File menu and choose Directories.”
• Any button that you click on the screen is represented in bold face
letters. For example, “click Close”.
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 1-3
Using This Manual
Typographical Conventions
Topic Headings
The following headings are used to show the organization of topics
within a chapter:
1-4 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Using This Manual
Reference Documentation
Reference Documentation
Reference documentation for the Q Exactive GC mass spectrometer
includes the following:
You can access PDF files of the documents listed above and of this
manual from the data system computer. The software also provides
Help.
Go to Start > All Programs > Thermo Exactive Series > Manuals.
• Forepump
• Turbomolecular pump
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 1-5
Chapter 2 Site Preparation
Contents
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 2-1
Site Preparation
Entrance Requirements
Entrance Requirements
This section lists data for packed units of typical Exactive Series systems.
The instrument (basic unit) is shipped in the largest container. Other
modules such as the data system, gas chromatograph, and accessories are
shipped in their own containers. Their dimensions and weight are less
than that of the container for the basic unit.
NOTICE Some chemicals that are needed for installation will be shipped
in a separate package. See “Performance Test Samples for Q Exactive
GC Systems” on page 7-2 for details. ▲
2-2 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Site Preparation
Entrance Requirements
Shipping Containers
Your Q Exactive GC GC-MS instrument is shipped in a container, the
smallest dimension of which is 90 cm (36 in.). To allow moving a packed
instrument, the entrance to your facility and the width of all hallways,
elevators, etc., should have a minimum width of 91 cm (36 in.)
Dimensions and weights of the shipping containers for Q Exactive GC
GC-MS systems are given in Table 2-1.
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 2-3
Site Preparation
Entrance Requirements
a
The computer, keyboard, monitor, foreline pump, and Q Exactive GC Installation Kit are
included in the box with the system.
b
Optional.
2-4 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Site Preparation
Space and Load Requirements
Monitor
Ethernet
communication
cables Wall outlet for data system components
(connected to 120 or 230 V AC
the mass
spectrometer)
1
A printer is not a standard part of the data system.
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 2-5
Site Preparation
Space and Load Requirements
See Table 2-2 and Figure 2-1 for the space requirements and weights of
the typical data system hardware components.
Table 2-2. Space and load requirements of typical data system hardware
components
Module Height Width Length Weight
cm in. cm in. cm in. kg lb
Monitor 36 14 41 16 18 7 6 13
Minitower computer 48 19 18 7 43 17 14 30
Laser printera 20 8 41 16 46 18 7 16
a
Not a standard part of the data system. The actual values depend upon your equipment.
The workbench for the GC/MS system must stand in a secure and level
position. Note that only workbenches with at least four legs provide
sufficient stability for the instrument. The workbench top must be dry
and clean (free of grease). Thermo Fisher Scientific recommends using a
workbench with a skid proof top.
Ensure you have adequate workbench space for the system. Refer to the
Table 2-3 for exact measurements of each component. Use the
information in the table below to configure the workbench.
2-6 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Site Preparation
Space and Load Requirements
Table 2-3. Space and load requirements of the Q Exactive GC GC/MS system
Module Height Width Length Weight
cm in. cm in. cm in. kg lb
Computerb 48 19 20 8 43 17 12 27
Monitorb 16 7 46 18 32 13 4 8
Keyboardb 23 9 46 18 5 2 1 2
AS 1310 Autosamplerd 28 11 41 16 40 19 6 13
a
This item is placed on the floor under the system.
b
Dimensions vary per manufacturer.
c
Add 0.8 kg (1.8 lbs) for each injector/detector module.
d
Mounts on top of the TRACE 1300/TRACE 1310.
e
About 20 cm (8 in.) of the orthogonal crossrail (Y-axis) protrudes from the rear of the GC.
f
Weight without accessories.
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 2-7
Site Preparation
Space and Load Requirements
95
Top View
148
91
Minimum Clearance
Allow at least 15 cm (6 in.) of clear space behind the system for proper
air circulation and for clearance of the gas lines and electrical
connections. This also provides sufficient space for accessing the fan
filters on the rear side of the MS. Also, allow at least 92 cm (36 in.) of
vertical clearance between the top of the mass spectrometer and any
shelves above it. Be sure to leave 46 cm (18 in.) of extra space to the
right of the Q Exactive GC system for performing maintenance.
To allow shutting off the mass spectrometer in an emergency, free access
to the power panel on the left side and to the power column on the rear
side of the instrument must be possible at any time.
2-8 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Site Preparation
Space and Load Requirements
Because of a weight of about 180 kg, Exactive Series instruments are too
heavy for one person alone to handle safely. Lifting and moving an
instrument requires the effort of at least six persons. The instrument is
shipped with two lifting devices, each providing sufficient space for
three persons that are standing side by side. Refer to the Q Exactive GC
Operating Manual for instructions about using the lifting devices.
An instrument with attached lifting devices does not fit through a door
with a width of 80 cm (32 in.), in contrast to the unpacked instrument
alone. Furthermore, the six persons that carry the instrument require
considerable space for maneuvering. Therefore, Thermo Fisher
Scientific recommends using a pallet jack to move the instrument and to
lift it to the height of the workbench.
The rear pair of the four support points for the instrument consists of
wheels. Thus, only two persons are necessary for moving the instrument
into its final position on a bench, while holding the two front handles.
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 2-9
Chapter 3 Operating Environment
Contents
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 3-1
Operating Environment
Temperature
Temperature
The laboratory room temperature must be maintained between 15 and
26 °C (59 and 78 °F).
There must be a good flow of room air around the system, and the air
conditioning system must be capable of maintaining a constant
temperature (within the temperature specification given above) in the
immediate vicinity of the system.
We recommend the installation of an air conditioner, if the specified
limits will be exceeded due to unfavorable climatic conditions.
Preferably, the air conditioner should be equipped with a flow controller
valve and PID microprocessor control (available e.g. from Landis &
Gyr, Polygyr RWX…, see www.landisgyr.com). This ensures
temperature drifts within the limits given above.
NOTICE Do not put the instrument under an air duct, near windows,
or near heating and cooling sources. Temperature fluctuations of 1°C or
more over a 10 minutes period can affect performance. ▲
3-2 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Operating Environment
Temperature
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 3-3
Operating Environment
Humidity
Humidity
The relative humidity of the operating environment must be between
40 and 70%, with no condensation. It is recommended that your
laboratory be equipped with a temperature/humidity monitor to ensure
that your laboratory is always within the required temperature and
humidity specifications.
Vibration
Floors must be free of vibration caused, for example, by equipment in
adjoining locations.
Radio Frequencies
Q Exactive GC instruments are able to withstand electromagnetic fields
of 1 V/m in the frequency range 26 MHz to 1 GHz without any
influence to operation.
3-4 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Operating Environment
Altitude
Altitude
Q Exactive GC systems are designed for indoor use at an altitude of up
to 2000 m (6500 ft) above sea level. Do not use the systems above
2000 m.
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 3-5
Chapter 4 Line Power
Contents
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 4-1
Line Power
Available Outlets
Available Outlets
Exactive Series instruments operate at a nominal voltage of 230 V AC,
50/60 Hz. Line voltages can vary between a minimum of 207 V AC and
a maximum of 253 V AC.
The mass spectrometer must be properly grounded. For this reason, the
power cable of the MS must be connected to a wall outlet with
grounding pin. Make sure that the grounding pin of this power cable is
connected to earth ground, even when a buck/boost transformer or
other upstream equipment (for example, UPS) is used.
The interconnected power outlets for the system are to have a common
point to one ground connector. If there are two such points, each of
which is connected to separate external ground, they will cause noise
current to flow through the ground system via the ground loop that is
formed.
Voltage between ground and neutral line should be less that 1 Volt.
4-2 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Line Power
Available Outlets
Table 4-1. System power requirements for the Q Exactive GC GC-MS system
Equipment Wall Outlets Max Current (A) at Maximum Power (W)
230 VAC (± 10%)
Q Exactive GC MSa 1 15 3450 (700 W effective)
Foreline pump N/A 3.4 550
TRACE 1310a 1 10 2000
Computerb Additional (as needed) 2.6 600
Monitorb Additional (as needed) 1 240
TriPlus RSHc Additional (as needed) 1.7 200 with one power
Sampling system module
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 4-3
Line Power
Available Outlets
The Q Exactive GC GC-MS system requires two power cables and wall
receptacles: one for the mass spectrometer and one for the TRACE 1310
GC. Two receptacles (P/N 2105500) are provided by Thermo Fisher
Scientific as part of the Preinstallation Kit. Two power cables (P/N
2112490) are provided by Thermo Fisher Scientific as part of the
Installation Kit.
The power cable for the forevacuum pump is shipped as part of the
forevacuum pump package. The foreline pump for the Q Exactive GC
system is plugged into the MS with the same cable used to connect the
power strip.
The cables for personal computer, monitor, and Ethernet switch are
provided by Thermo Fisher Scientific. They are approximately 2 m
(6 ft) long. Local codes in your area might require the installation of
another type of plug and receptacle. The Thermo Fisher Scientific field
service engineer for your country provides the appropriate power plugs.
4-4 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Line Power
Connecting Mass Spectrometer and Modules to Wall Outlets
The electrical wall outlet for the main power of the mass spectrometer
should be located at the wall near the intended location of the
instrument.
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 4-5
Line Power
Connecting Mass Spectrometer and Modules to Wall Outlets
See the sample laboratory setup in Table 4-2 for the recommended
number of outlets.
Table 4-2. Sample laboratory setup
Item Outlets
GC system Autosampler 1 or 2
GC 1 (230 V)
MS system Mass spectrometer 1 (230 V)
Forepumpa —
Data System Data system computer 1
Monitor 1
Printer (optional) 1
Ethernet switch 1
Optional High intensity lamp (for help in 1
instrument maintenance)
Laboratory stereoscope (for inspecting 1
fused-silica parts)
Nitrogen generator 1
Total outlets required for this configuration 6–14
a
Power is supplied by MS system
4-6 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Chapter 5 Consumables
Your instrument requires gases and solvents that must meet defined
purity specifications. The Thermo Fisher Scientific field service engineer
might also require certain solvents for the installation verification of
your system.
Contents
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 5-1
Consumables
Fittings and Parts
additional length.
Connection for the opposite end of the Teflon hose to Not provided in kit. You supply these parts.
the nitrogen gas source
T-piece (P/N 1128140) provided
NOTICE Your connections and gas delivery system might vary, and it is
your responsibility to supply any fittings or connections necessary
during installation.
If the pressure regulator of the laboratory gas supply has an 1/8 inch
NPT outlet, examples1 of suitable 1/8-in.-to-6-mm adapters are
Swagelok™ part numbers B-6M0-7-2 (female) and B-6M0-1-2
(male). ▲
1
Thermo Fisher Scientific does not endorse any manufacturer, nor does it endorse products
other than its own. Companies and products listed in this guide are given as examples only.
5-2 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Consumables
Gases
Gases
Q Exactive GC instruments require nitrogen gas for the C-Trap bath
gas, and HCD collision gas. Your system can use large amounts of gas
during daily operations. It is essential that the gas be delivered with the
necessary pressure and purity. This section provides information on the
purity and pressure that your system requires.
NOTICE Do not store gas cylinders where they can damage cables or gas
lines, and secure them in accordance with standard safety practices. ▲
Nitrogen Gas
The nitrogen gas should be of ultra high purity (≥99.999%). Particulate
filters can be a source of contamination, they are not recommended.
The required gas pressure is 800 ± 30 kPa (8.0 ± 0.3 bar, 116 ± 4 psi).
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 5-3
Consumables
Gases
5-4 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Consumables
Gases
• Some CI gases, such as ammonia, are toxic. Make sure these gases
are properly exhausted and all gas fittings on the system are
leak-free. Consult your local Environmental and Safety Regulations
for information about how to properly exhaust fumes from your
laboratory.
• Do not exceed 240 kPa (35 psi) or you could damage the CI reagent
gas flow module. ▲
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 5-5
Consumables
Gases
TriPlus SPME Nitrogen 99.999% 200 kPa Dual-stage brass regulator with CGA-580
Headspace (30 psi) stainless steel diaphragm
a
Connectors will vary with cylinder size. Confirm that your regulator will work with your gas tank. All connections to the GC/MS are 1/8 in.
Swagelok fittings.
b
Pure, particle and oil free, and contained in one tank.
Obtain the proper gas line filters, which help prevent impurities and
contaminants from entering your system. Water, oxygen, and total
hydrocarbons should be less than 1 ppm to avoid high background noise
and prevent contamination. The GC is equipped with two intake filters
that trap moisture, oxygen, and hydrocarbons.
Store gas tanks and bottles properly so they will not damage cables or gas
lines. Ensure they are secured in accordance with standard safety
practices.
5-6 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Consumables
Cleaning Agents
Cleaning Agents
We recommend having the following cleaning agents available:
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 5-7
Chapter 6 Exhaust
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 6-1
Exhaust
Exhaust System
Exhaust System
It is mandatory to connect the forepump to a fume exhaust system. The
forepump eventually exhausts much of what is introduced into the mass
spectrometer, including the small amount of oil vapor that mechanical
pumps can emit. It is your responsibility to provide an adequate exhaust
system.
Supply a 1-in. i.d. hose to the building exhaust air system, an oil mist
filter, or other scrubbing device. The pump exhaust contains carrier gas,
solvents, analytes, and a small amount of oil vapor. These materials may
be flammable, poisonous, or corrosive. Do not allow the exhaust from
the foreline pump, which includes your analytes, to accumulate to
unsafe levels in your laboratory. Consult your local Environmental and
Safety Regulations for information about how to properly exhaust fumes
from your laboratory.
NOTICE Do not route exhaust tubing from the pump vertically toward
the ceiling. To maintain pump integrity, route the tubing from the
exhaust port down to the floor. ▲
NOTICE The exhaust hose should travel at floor level for a minimum of
two meters (78.5 in.) before it reaches the external exhaust system. This
tubing acts as a trap for exhaust fumes that would otherwise recondense
in the forepump oil. ▲
Ventilation
Most of the nitrogen that is introduced into the Q Exactive GC escapes
into the laboratory atmosphere. Therefore, provide for good air
exchange to prevent accumulation of gaseous nitrogen in the laboratory.
6-2 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Chapter 7 Installation
When your lab site preparation is completed and the system is delivered,
please call your Thermo Fisher Scientific office to arrange for an
installation date.
Contents
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 7-1
Installation
Performance Test Samples for Q Exactive GC Systems
The installation will not begin until the arrival of all chemicals listed in
Table 7-1!
7-2 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Chapter 8 General Preinstallation Information
Contents
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 8-1
General Preinstallation Information
Instrument Arrival
Instrument Arrival
When your lab site preparation is completed, and the system is
delivered, call your local Thermo Fisher Scientific office to arrange for
an installation date.
Transportation Risk
Transportation risk depends on the terms of delivery agreed. The terms
of shipment determine who has responsibility for filing a claim against
the carrier if the system is damaged in transit.
8-2 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
General Preinstallation Information
Installation
Installation
It is the policy of Thermo Fisher Scientific that the customer should not
unpack the system or accessory items prior to installation of the system.
NOTICE Consumables sent with the system are intended for use by the
Thermo Fisher Scientific field service engineer during the installation. ▲
Key Operator
Experience has shown that the maximum benefit can be derived from a
scientific instrument if there is one person, a key operator, who has
major responsibility for that instrument. It is recommended that you
designate a key operator to oversee the operation and maintenance of
the system in your laboratory. The key operator should be available to
the installing engineer throughout the installation. This person will also
be the key figure in the communication between your laboratory and
Thermo Fisher Scientific.
NOTICE Do not plan to use your new system for sample analysis until
the installation is complete and the Acceptance Form has been signed. ▲
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 8-3
General Preinstallation Information
Installation
Germany
Preventive Maintenance
Routine and preventive maintenance of mass spectrometer is in the
responsibility of the customer. Included in this category are replacement
of the cone, exchange of pump oil, replacement of filters, etc. on a
regular basis. Refer also to the manufacturers manuals delivered with the
instrument—especially for the maintenance of mechanical pumps and
turbomolecular pumps.
8-4 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
General Preinstallation Information
Operating Environment
Operating Environment
These general specifications for the operating environment help
ensuring continued high performance of the system.
Lighting
Good lighting makes any work area more enjoyable. Since a lot of work
is done on the computer terminal, it may be convenient to have a
dimmer switch on the lights to reduce eyestrain. A small, high-intensity
lamp is recommended for cleaning mass spectrometer components,
source inspection, and manipulation of small components.
Particulate Matter
Particulate matter might contaminate the samples and the ion source
and may limit the background level of the instrument.
The air in your laboratory must not contain excessive dust, smoke, or
other particulate matter. For reference, the air should contain fewer than
35×106 particles per cubic meter (1×106 particles per cubic foot) in
excess of 5 µm.
Dust can clog the air filters, causing a reduction in air flow around
electronic components. Dust will also form a layer on electronic
components that will act as an insulating blanket and thus reduce the
heat transfer from the components to the surrounding air.
Quality of Power
The quality of power supplied to your system is very important for its
performance. The quality of line voltage must be stable and within the
specifications listed in this manual. The line voltage must be free voltage
surges, sags, or transients.
• Sags and surges are sudden changes in average RMS voltage level,
with typical durations between 50 µs and 2 s.
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 8-5
General Preinstallation Information
Operating Environment
• Slow average
• Transient
In the first two cases, the duration as well as the amplitude of the
disturbance is indicated by time interval recording.
1
Thermo Fisher Scientific does not endorse any manufacturer, nor does it endorse products
other than its own. Companies and products listed in this guide are given as examples only.
8-6 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
General Preinstallation Information
Operating Environment
Various line voltage conditioning devices are available that can correct
your line voltage problem. If you have good regulation but the power
line disturbance analyzer shows transient voltages, then an
isolation/noise suppression transformer should be adequate to resolve
the problem. If there are both transient and regulation problems, then
you should consider power conditioners, which can control both of
these problems.
When nominal voltage is free from voltage sags, surges, and impulses
but different from 230 V AC line to ground, the supply voltage can be
lowered (bucked) or raised (boosted) using a buck/boost transformer.
Buck/boost transformers are also available from Thermo Fisher
Scientific.
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 8-7
General Preinstallation Information
Operating Environment
Technical Assistance
1
Thermo Fisher Scientific does not endorse any manufacturer, nor does it endorse products
other than its own. Companies and products listed in this guide are given as examples only.
8-8 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
General Preinstallation Information
Operating Environment
Electrostatic Discharge
Electrostatic discharge (ESD) can damage the electronic components of
your MS system. Thermo Scientific instruments are designed to
withstand electrostatic discharges (ESD) up to 4 kV (air discharge) and
4 kV (contact discharge) with all panels in place. However, if the panels
are removed and the PCBs are handled without proper precautions, the
electronic components might be damaged or fail prematurely. Static
electricity can develop in a variety of ways. A few examples of how
electrostatic charge can develop are as follows:
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) 8-9
Legal Documents
This chapter provides legal documents.
Contents
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) D-1
Legal Documents
FCC Compliance Statement
D-2 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Legal Documents
WEEE Compliance
WEEE Compliance
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment
(WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each
European Union (EU) Member State, and these companies should dispose of or recycle this product. See
www.thermoscientific.com/rohsweee for further information on Thermo Fisher Scientific’s compliance
with these Directives and the recyclers in your country.
WEEE Konformität
Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2002/96/EC
erfüllen. Das Produkt ist durch folgendes Symbol gekennzeichnet:
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) D-3
Legal Documents
WEEE Compliance
Conformité DEEE
Ce produit doit être conforme à la directive européenne (2002/96/EC) des Déchets d'Equipements
Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant:
Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état
membre de l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage
d'informations sur la conformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre pays
et les informations sur les produits Thermo Fisher Scientific qui peuvent aider la détection des substances
sujettes à la directive RoHS sont disponibles sur www.thermoscientific.com/rohsweee.
D-4 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Index
A wall outlets 4-6
workbench 2-5
acceptance form 8-3
delivery terms 8-2
advanced training courses 8-4
detergents 5-7
air
circulation 2-8 dimensions, of instrument 2-8
duct 3-2 dimensions, of packed units 2-4, 2-6, 5-3–5-5
exchange 6-2 distilled water 5-7
flow 8-5 documentation survey 0-i
air conditioning doorways 2-2
humidity 3-4 dust 8-5
load 3-2
temperature 3-2
altitude 3-5 E
earth ground 4-2
electromagnetic fields 3-4
B electronic components 3-4, 8-5, 8-9
buck/boost transformer 4-2, 8-7 electrostatic discharge (ESD)
damage 8-9
precautions 8-9
C emergency shutdown 2-8
entrance requirements 2-2–2-3
calibration compounds 7-2
Ethernet
cell phones 3-4 cables 2-5–2-6
certification, of power conditioning devices 8-7 switch 2-5, 4-4
chemicals exhaust
for installation 2-2 hose 6-2
cleaning agents 5-7 system 6-2
collision gas, for HCD 5-3
compliance, WEEE D-3
connecting
wall outlets 4-5
F
connector, for mains supply 4-4 failures, repair 1-2
container fan filters 2-8
dimensions 2-2 fittings 5-2
weights 2-2 floor covering 8-9
contamination forepump
gas lines 5-3 location 3-4
C-Trap, bath gas 5-3 power supply 4-4–4-5
vibration 3-4
freight insurance 8-2
fume exhaust system 6-2
D fuses 4-5
damage
documenting 8-2
to equipment 8-2
data system G
dimensions 2-6, 5-3–5-5 gas
placing 2-5 connection hardware 5-2
shipping container 2-2 contaminants 5-3
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) I-1
Index: H–R
I O
instrument oil vapor 6-2
damage 5-3 overheat, avoiding 2-9
dimensions 2-8
failures 1-2
lifting 2-9 P
moving 2-9
performance 3-4 packing materials 8-9
repair 1-2 pallet jack 2-2, 2-9
storing after arrival 8-2 particulate filters, for gases 5-3
wall outlet 4-6 particulate matter 8-5
weight 2-5, 2-9 performance, of instrument 3-4
insulating blanket 8-5 placing, the instrument 3-2
isolation/noise suppression transformer 8-7 polystyrene foam cups 8-9
power 4-3
cables 4-4
conditioning 8-7–8-8
K connector 4-4
key operator 8-3–8-4 consumption 3-2
line 8-6
monitoring 8-6
L outlets 4-2, 4-6
quality 8-5
laboratory 8-5 sags and surges 8-5
atmosphere 6-2 slow average 8-5
chairs 8-9 supply quality 8-6
coats 8-9 transients 8-5
exhaust system 6-2
power supply
floor 2-5 data system 4-5
lighting 8-5
peripheral devices 4-5
setup 4-6
Preinstallation Kit 1-5, 7-2
ventilation 6-2
pressure, of nitrogen gas 5-3
lifting devices 2-9
preventive maintenance 8-4
lifting, the instrument 2-9
printer 2-5
lighting, in the laboratory 8-5
M R
radio frequencies 3-4
maintenance techniques 8-4
receptacles 4-4
mobile phones 3-4
I-2 Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) Thermo Scientific
Index: S–W
removing the instrument, from shipping container 2-2 transportation risk 8-2
room temperature 3-2
U
S uninterruptible power supply (UPS) 4-2, 4-6, 8-7
shipment terms 8-2 user’s responsibilities 1-2, 2-1, 3-1, 4-1, 5-1–5-2, 6-1–6-2,
short circuits 3-4 8-3–8-4, 8-8
shutting off, power 4-2 using, your new instrument 8-3
smoke 8-5
solvents
for cleaning 5-7 V
waste 6-2
vacuum
space requirements 2-2, 2-5
hose 2-9
static electricity 3-4, 8-9
ventilation slots 2-9
storing
ventilation, in the laboratory 6-2
gas cylinders 5-3
vertical clearance 2-8
instrument and installation equipment 7-1
vibration 3-4
support points, of instrument 2-9
voltage
survey link 0-i
disturbances 8-5–8-6
system reliability 3-2
sags 4-2
T W
technical assistance 8-8
wall outlets 2-5, 4-4–4-5
temperature
WEEE compliance D-3
fluctuations 3-2
monitor 3-4 weights
equipment 2-4, 2-6, 5-3–5-5
third-party manuals 1-5
instrument 2-5
tilt angle, of instrument 2-9
options 2-5
training courses 8-4
wheels, of the instrument 2-9
transformers workbenches
buck/boost 8-7 data system 2-6
installation 8-3 dimensions 2-6
isolation/noise 8-7 MS system 2-6
transient voltages 8-7 top 2-6
transport pallet 2-9
Thermo Scientific Q Exactive GC Preinstallation Requirements Guide (P/N 1R120706-0001, Revision B) I-3