50 Series F/L Flow Meter

world class water monitoring series 50 for open channels


Data Sheet

MAIN CHARACTERISTICS
• Flow and temperature measurement on open channels with ultrasound or radar or piezo
4-20mA probe (Temperature only with ultrasonic sensors)
• Programming key pad with 5 keys
• "CAL" Function Key to direct access to the calibration menu
• "GRAPH" Function Key to direct access to the graphs of measure
• “USB” Function Key for data download on USB support
• “MODE” Function Key for self-recognition probes
• LCD Graphic color display 480(R.G.B.) (W) x 272(H)
• Internal Data Logger (flash 32 Mbit) with the possibility of graphic and table visualisation
of measurement trends
• PID adjustment
• Serial output RS485 MOD BUS RTU
• Serial output PROFI BUS (Optional)
• Serial output ETHERNET Modbus TCP/IP (Optional)
• ETHERNET IP (on request)
• Data download on USB support
• 4 Programmable Analogical Outlets
• 4 Relay Outlets for intervention thresholds
• 1 Relay Outlet for Instrument Anomaly Alarm
• 1 Digital Input for disabling of doses

50Series F/L – technical data Sheet Page 1


50 Series F/L Flow Meter
world class water monitoring series 50 for open channels
Data Sheet

 Main hardware characteristics of the electronic device


The hardware structure of this periphery is based on the adoption of extremely new CPU
CMOS with 32 bits developed specifically for the execution of the so-called “embedded”
applications.
The card uses flash memories to store archives of historical data and LOG files of events.
The Card has 1 RS485 gate for sensors and 1 RS485 gate (opto-isolated) for local networks
used for connections with local communication devices (configuration computer, terminals
and remote controls etc). As optional it is possible to install a PROFI BUS or ETHERNET
serial port for connection with the communication devices PROFI BUS and ETHERNET
respectively.
The card integrates a Real Time Clock (clock with date) that allows the software to storage
figures in a chronological order.

 The device has been designed to be fitted onto a panel, and is built with IP66
protection panel.

 Controller maximum capability

Power supply

Characteristics of the Measure


3
Measurement Ranges 0-99999m /h
(depending on the For resolution and accuracy see the characteristics of the connected
connected sensor) sensor
Temperature Automatic – only for ULTRASONIC probes
Compensation
Simultaneous flow rate, resettable and non-resettable totalizer,
Visualization
temperature (with ultrasound probes)

50Series F/L – technical data Sheet Page 2


50 Series F/L Flow Meter
world class water monitoring series 50 for open channels
Data Sheet

Software features and functions


Internal Flash 32Mbit Memory (near to 128000 records).
Records interval: 01:00 ÷ 99:99 min
Type: Circular (F.I.F.O.) or Filling
Storage of:
- instant flow
Data storage
- totalizers
- alarms
With minimum storage interval of 1 min.
Possibility of visualization of the stored data in tabular and graphic form,
with indication of max, min and average values of the selected period.
Zoom function
4 Analogue Outputs 1 for each measure with possibility of PID management
4 Uscite digitali di Set Point ON - OFF: setting of the working range (hysteresis / direction)
Comando and pulses for totalizer
Max 2 uscite per misura

Uscita digitale di Allarme / Reporting: Instrumental anomalies, lack of ultrasound echo-probe


Digital input To disable dosages or activate washing cycle
For set-up and real-time data acquisition from remote or for stored data
RS485 Serial output
download (using a dedicate-SW) .
MODBUS RTU communication protocol
Possibility to simulate all the analogue and digital outputs using the
Manual controls
keyboard
Rectangular – Thomson – Bazin – Venturi – Palmer Bowlus – 30 points
Possible open weirs
table
Totalizers Nr 2 1 resettable, 1 non-resettable , totalizer has 10 digits

Hardware Characteristics
Visualization LCD Graphic color display 480(R.G.B.) (W) x 272(H)
Programming 5 bubble-Keys keyboard
Data Logger Flash 32Mbit Memory (near to 128000 records).
0 / 4.00 ÷ 20.00 mA
Galvanic separation: 1KV Optoisolator
Analogue Outputs
Maximum load 500 Ohm
Second Alarm output: NAMUR 2.4 mA (with 4/20mA Range)
Switching Relays Usable as NO contact
Digital Outputs
Max resistive load 3A a 230Vac
Active and already supplied
Digital Input
Possibility to link with a 3 wires - inductive sensor
Analogue Input 4-20mA Analogue Input freely programmable
RS485 with 1200÷38400 Baud Rate programmable speed
Serial Output
MOD BUS RTU Protocol
Operating temperature -20÷65°C
Operating conditions Storage and transport -25÷65°C
Humidity 10-95% (non-condensing)
Power supply 100÷240Vac/dc 50- 60 Hz – (Optional 24Vac/dc)
- Sensors power supply:
- 2-wire: 12Vdc
Power Supply/ - 4-wire: 24Vdc
Electrical protections – Transformer isolation4KV
– Absorbing average < 7W
– Electrical Protection: EMI / RFI CEI-EN55011 – 05/99
– According to UL, not connect to relay outputs a voltage exceeding 115V

50Series F/L – technical data Sheet Page 3


50 Series F/L Flow Meter
world class water monitoring series 50 for open channels
Data Sheet

 Mechanical Dimensions

Mechanical Dimensions 50Series Flow Wall IP66


Dimensions (L x H x P) 144x144x122,5mm
Mounting thickness 122,5mm
Material Grey ABS RAL 7045
Mounting Wall
Weight 1 Kg
Front Panel UV resistant Polycarbonate

50Series Flow Panel


Mechanical Dimensions
144x144
Dimensions (L x H x P) 144x144x86,5mm
Mounting thickness 61mm
Material Grey ABS RAL 7045
Mounting Panel
Weight 0,7 Kg
Front Panel UV resistant Polycarbonate

50Series F/L – technical data Sheet Page 4


50 Series F/L Flow Meter
world class water monitoring series 50 for open channels
Data Sheet

 Controls, indicators and connections

5
6

2
Front panel, wall mounting version Access to terminal box

1. LCD Display
2. DOWN
3. UP
4. ENTER
5. ESC
6. GRAPH-USB

 Modbus-RTU serial interface

• Real-time data communication


• Download of the stored data on a PC via SW or directly on an USB support

50Series F/L – technical data Sheet Page 5


50 Series F/L Flow Meter
world class water monitoring series 50 for open channels
Data Sheet

 Order codes for 50Series F/L Analyzers

97P0610000A 50Series F/L Dig. plug&play analyser Flow Modbus Panel 220V

97P0610003A 50Series F/L Dig. plug&play analyser Flow Modbus Panel 24Vac

97P0610010A 50Series F/L Dig. plug&play analyser flow Modbus Wall 220V

97P0610013A 50Series F/L Dig. plug&play analyser flow Modbus Wall 24Vac

97P0610100A 50Series F/L Dig. plug&play analyser Flow Profibus Panel 220V

97P0610103A 50Series F/L Dig. plug&play analyser Flow Profibus Panel 24Vac

97P0610110A 50Series F/L Dig. plug&play analyser Flow Profibus Wall 220V

97P0610113A 50Series F/L Dig. plug&play analyser Flow Profibus Wall 24Vac

97P0610210A 50Series F/L Dig. plug&play analyser Flow Ethernet Wall 220V

CHEMITEC s.r.l.
Via Isaac Newton 28 - 50018 Scandicci (FI)
Tel. +39 055 7576801 fax +39 055 756697
Web site: www.chemitec.it E-mail: sales@chemitec.it

50Series F/L – technical data Sheet Page 6