• Wastewater
• Drinking Water
Technical Data*
Model NT3100sc UV Nitrate Sensor
Path Length 1 mm 2 mm 5 mm
Measuring range1 0.1 - 90 mg/L NO3-N 0.05 - 50 mg/L NO3-N 0.02 - 25 mg/L NO3-N
Lower Limit of
0.1 mg/L NO3-N 0.05 mg/L NO3-N 0.02 mg/L NO3-N
Detection (LOD)
± 4% of reading ± 0.1 mg/L NO3-N ± 3% of reading ± 0.1 mg/L NO3-N
from 0.05 - 22 mg/L, from 0.02 - 13 mg/L,
Accuracy2 ± 5% of reading ± 0.1 mg/L NO3-N
± 5% of reading ± 0.1 mg/L NO3-N ± 5% of reading ± 0.1 mg/L NO3-N
from 22 - 50 mg/L from 13 - 25 mg/L
Model NT3200sc UV Nitrate and Nitrite Sensor
Path Length 1 mm 2 mm 5 mm
0.3 - 70 mg/L NO3-N; 0.15 - 34 mg/L NO3-N; 0.07 - 17 mg/L NO3-N;
Measuring range1,2
0.4 - 92 mg/L NO2-N 0.15 - 61 mg/L NO2-N 0.06 - 23 mg/L NO2-N
Lower Limit of
0.3 mg/L NO3-N; 0.4 mg/L NO2-N 0.15 mg/L NO3-N; 0.15 mg/L NO2-N 0.07 mg/L NO3-N; 0.06 mg/L NO2-N
Detection (LOD)2
± 4% of reading ± 0.6 mg/L NO3-N, ± 3% of reading ± 0.3 mg/L NO3-N, ± 3% of reading ± 0.1 mg/L NO3-N,
Accuracy2
± 4% of reading ± 0.6 mg/L NO2-N ± 3% of reading ± 0.2 mg/L NO2-N ± 5% of reading ± 0.24 mg/L NO2-N
1) Not expressed for mixed NOx concentration. Measured with NO3-N for NT3100sc and NO3-N/NO2-N for NT3200sc standard solutions.
2) Measured in lab conditions.
Principle of Operation
Nitrate dissolved in water absorbs UV light with wavelengths below 250 nm. The nitrate and nitrite absorption of UV light makes it
possible to photometrically determine the nitrate and nitrite dissolved without reagents. The sensor is put directly in the medium.
The color of the medium does not have an effect on the measurement because the measuring principle is based on the analysis of
invisible UV light.
Dimensions
[mm]
Sensor
Sensor mounting
4 NT3 Series UV Nitrate and Nitrite Sensors
Applications
NT3100sc: General recommendation. Each application needs to be assessed for NOx concentrations and turbidity.
NT3200sc: When both nitrate and nitrite are present, the accuracy of NO2-N and maximum NOx-N concentrations are reduced.
The max NO3-N value can vary depending on the UV absorbance of the matrix. Examples of parameters impacting absorbance:
solids, organics, humic acids, and salts (e.g., fluoride). This table is used to help choose the best path length for your conditions.
Contact your local Sales Representative to discuss your specific application.
✔ Suitable application
● Limited accuracy for nitrite
1 Limited accuracy for nitrite with NO3-N values >10 mg/L
2 Limited accuracy for nitrite with NO3 values >80 mg/L
3 Only in low (~0 mAbs/mm) UV absorbance matrix
5 NT3 Series UV Nitrate and Nitrite Sensors
Order Information
Sensors
LXV448.99.11001 NT3100sc UV Nitrate Sensor, 1 mm path length
LXV448.99.21001 NT3100sc UV Nitrate Sensor, 2 mm path length
LXV448.99.51001 NT3100sc UV Nitrate Sensor, 5 mm path length
LXV448.99.12001 NT3200sc UV Nitrate and Nitrite Sensor, 1 mm path length
LXV448.99.22001 NT3200sc UV Nitrate and Nitrite Sensor, 2 mm path length
LXV448.99.52001 NT3200sc UV Nitrate and Nitrite Sensor, 5 mm path length
A Hach SC controller is required to operate the NT3100sc and NT3200sc sensors, controller must be purchased separately.
This instrument connects to Claros, Hach’s innovative Water With Hach Service, you have a global partner who understands
Intelligence System. Claros allows you to seamlessly connect your needs and cares about delivering timely, high-quality service
and manage instruments, data, and process – anywhere, you can trust. Our Service Team brings unique expertise to help
anytime. The result is greater confidence in your data and you maximize instrument uptime, ensure data integrity, maintain
improved efficiencies in your operations. To unlock the full operational stability, and reduce compliance risk.
potential of Claros, insist on Claros Enabled instruments.