Aptima BV
Aptima BV
Aptima BV
Aptima BV Assay ®
General Information
Intended Use
The Aptima ® BV assay is an in vitro nucleic acid amplification test that utilizes real time
transcription-mediated amplification (TMA) for detection and quantitation of ribosomal RNA
from bacteria associated with bacterial vaginosis (BV), including Lactobacillus (L. gasseri, L.
crispatus, and L. jensenii), Gardnerella vaginalis, and Atopobium vaginae. The assay reports
a qualitative result for BV and does not report results for individual organisms. The assay is
intended to aid in the diagnosis of BV on the automated Panther® system using
clinician-collected and patient-collected vaginal swab specimens from females with a clinical
presentation consistent with vaginitis and/or vaginosis.
C. To reduce the risk of invalid results, carefully read the entire package insert and the
Panther/Panther Fusion System Operator’s Manual prior to performing this assay.
D. Only personnel adequately trained in the use of the Aptima BV assay and in handling
potentially infectious materials should perform this procedure. If a spill occurs,
immediately disinfect following appropriate site procedures.
E. For additional specific warnings and precautions, refer to the Panther/Panther Fusion
System Operator’s Manual.
Laboratory Related
G. Use routine laboratory precautions. Do not pipette by mouth. Do not eat, drink or smoke
in designated work areas. Wear disposable, powderless gloves, protective eye wear, and
laboratory coats when handling specimens and kit reagents. Wash hands thoroughly after
handling specimens and kit reagents.
H. Work surfaces, pipettes, and other equipment must be regularly decontaminated with
2.5% to 3.5% (0.35 M to 0.5 M) sodium hypochlorite solution. Thoroughly clean and
disinfect all work surfaces.
I. Dispose of all materials that have come in contact with specimens and reagents in
accordance with applicable national, international, and regional regulations (14, 15, 16).
Thoroughly clean and disinfect all work surfaces.
Specimen Related
J. Expiration dates for the collection kits pertain to the collection of specimens and not to the
specimen testing. Samples collected any time prior to the expiration date of the collection
kit, and transported and stored in accordance with the package insert, are valid for testing
even if the expiration date on the collection tube has passed.
K. Specimens may be infectious. Use Universal Precautions when performing this assay
(14,15). Proper handling and disposal methods should be established according to local
regulations (16). Only personnel adequately trained in the use of the Aptima BV assay
and trained in handling infectious materials should perform this procedure.
L. Maintain proper storage conditions during specimen shipping to ensure the integrity of the
specimen. Specimen stability under shipping conditions other than those recommended
has not been evaluated.
M. Avoid cross-contamination during the specimen handling steps. Specimens can contain
extremely high levels of organisms. Ensure that specimen containers do not contact one
another, and discard used materials without passing over any open container. Change
gloves if they come in contact with a specimen.
N. Upon piercing, liquid can discharge from Aptima transfer tube caps under certain
conditions. Refer to the Panther System Test Procedure for more information.
O. If the lab receives an Aptima Multitest Swab Specimen Collection Kit transport tube with
no swab, two swabs, a cleaning swab, or a swab not supplied by Hologic, the specimen
must be rejected.
Assay Related
P. Do not interchange, mix, or combine assay reagents from kits with different master lot
numbers. Controls, the calibrator, and assay fluids may be interchanged.
Q. Cap and store reagents at the specified temperatures. The performance of the assay may
be affected by use of improperly stored reagents. See Reagent Storage and Handling
Requirements and Panther System Test Procedure for more information.
R. Do not combine any assay reagents or fluids without specific instruction. Do not top off
reagents or fluids. The Panther system verifies reagent levels.
T. Do not use the reagent, control, or calibrator kits after the expiration date.
U. Some reagents used with the Aptima BV assay are labeled with risk and safety symbols.
Note: Hazard Communication information for labeling of globally marketed products reflects the US
and EU Safety Data Sheets (SDS) classifications. For hazard communication information specific to
your region, refer to the region specific SDS on the Safety Data Sheet Library at www.hologicsds.com.
For more information on the symbols, refer to the symbol legend on www.hologic.com/package-inserts.
EU Hazard Information
Promoter Reagent
MAGNESIUM CHLORIDE 35 - 40%
—
H412 - Harmful to aquatic life with long lasting effects.
P273 - Avoid release to the environment
P280 - Wear eye protection/ face protection.
2
Storage condition for the working Target Capture Reagent (Target Capture Reagent with Internal Control added).
B. Discard any unused reconstituted reagents and working Target Capture Reagent (wTCR)
after 30 days or after the Master Lot expiration date, whichever comes first.
C. Reagents stored on the Panther system have 120 hours of onboard stability. Reagents
can be loaded onto the Panther system up to 8 times. The system logs each time the
reagents are loaded.
D. The Promoter Reagent and reconstituted Promoter Reagent are photosensitive. Protect
these reagents from light during storage and preparation for use.
E. Avoid cross-contamination during reagent handling and storage. Recap all reconstituted
reagents with new reagent caps prior to storage.
A. Specimen collection
Refer to the appropriate specimen collection kit package insert for specific collection
instructions.
Note: Specimens must be shipped in accordance with applicable national, international, and regional
transportation regulations.
Panther System
Reagents for the Aptima BV assay are listed below for the Panther system. Reagent
identification symbols are also listed next to the reagent name.
c. Check the lot numbers on the Master Lot Barcode Sheet to ensure that the
appropriate reagents are paired.
d. Open the lyophilized reagent vial and firmly insert the notched end of the
reconstitution collar into the vial opening (Figure 1, Step 1).
e. Open the matching reconstitution solution bottle, and set the cap on a clean,
covered work surface.
f. While holding the reconstitution solution bottle on the bench, firmly insert the other
end of the reconstitution collar into the bottle opening (Figure 1, Step 2).
g. Slowly invert the assembled bottles. Allow the solution to drain from the bottle into
the glass vial (Figure 1, Step 3).
h. Gently swirl the solution in the bottle to mix. Avoid creating foam while swirling the
bottle (Figure 1, Step 4).
i. Wait at least 15 minutes for the lyophilized reagent to go into solution, then invert the
assembled bottles again, tilting at a 45° angle to minimize foaming (Figure 1, Step
5). Allow all of the liquid to drain back into the plastic bottle.
j. Remove the reconstitution collar and glass vial (Figure 1, Step 6).
k. Recap the plastic bottle. Record operator initials and reconstitution date on the label
(Figure 1, Step 7).
l. Discard the reconstitution collar and glass vial (Figure 1, Step 8).
Option: Additional mixing of the Amplification, Enzyme, and Promoter Reagents
using a tube rocker is allowed. The reagents may be mixed by placing the
recapped plastic bottle on a tube rocker set to 20 RPM (or equivalent) for a
minimum of 5 minutes.
Warning: Avoid creating foam when reconstituting reagents. Foam compromises the
level-sensing in the Panther system.
d. Open the bottle of IC and pour the entire contents into the bottle of TCR. Expect a
small amount of liquid to remain in the IC bottle.
e. Cap the bottle and gently swirl the solution to mix the contents. Avoid creating foam
during this step.
f. Record operator initials and the current date on the label.
g. Discard the IC bottle and cap.
D. Specimen Handling
1. Allow the specimens to reach room temperature prior to processing.
2. Do not vortex specimens.
3. Visually confirm that each specimen tube meets the following criteria:
a. The presence of a single pink Aptima collection swab in a swab specimen
transport tube.
4. Inspect specimen tubes before loading into rack:
a. If a specimen tube contains bubbles in the space between the liquid and the cap,
centrifuge the tube for 5 minutes at 420 RCF to eliminate the bubbles.
b. If a specimen tube has a lower volume than typically observed when collection
instructions have been followed, centrifuge the tube for 5 minutes at 420 RCF to
ensure that no liquid is in the cap.
Note: Failure to follow Steps 4a-4b may result in liquid discharge from the specimen tube cap.
Note: Up to 4 separate aliquots can be tested from each specimen tube. Attempts to pipette
more than 4 aliquots from the specimen tube can lead to processing errors.
E. System Preparation
1. Set up the system according to the instructions in the Panther/Panther Fusion System
Operator’s Manual and Procedural Notes. Make sure that the appropriately sized
reagent racks and TCR adapters are used.
Procedural Notes
A. Calibrator and Controls
Allow the calibrator and controls to reach room temperature prior to processing.
1. The positive calibrator, positive control and negative control tubes can be loaded in
any rack position or in any Sample Bay Lane on the Panther system. Specimen
pipetting will begin when one of the following 2 conditions has been met:
a. The calibrator and controls are currently being processed by the system.
b. Valid results for the calibrator and controls are registered on the system.
2. Once the calibrator and control tubes have been pipetted and are processing for a
specific reagent kit, patient specimens can be tested with the associated kit up to 24
hours unless:
a. The calibrator result or control results are invalid.
b. The associated assay reagent kit is removed from the system.
c. The associated assay reagent kit has exceeded stability limits.
3. Each calibrator or each control tube can be used once. Attempts to use more than once
can lead to processing errors.
B. Temperature
Room temperature is defined as 15°C to 30°C.
C. Glove Powder
As in any reagent system, excess powder on some gloves may cause contamination of
opened tubes. Powderless gloves are recommended.
Quality Control
An operator may invalidate an individual specimen or an entire run if it was observed and
documented that a procedural, technical, or instrument-related error occurred while
performing the assay.
Assay Calibration
To generate valid results, an assay calibration must be completed. The calibrator is run in
triplicate each time a reagent kit is loaded on the Panther system. Once established, the
calibration is valid for up to 24 hours. Software on the Panther system alerts the operator
when a calibration is required. The operator scans the calibration coefficients found on the
Master Lot Barcode Sheet provided with each reagent kit.
During processing, criteria for acceptance of the calibrator is automatically verified by the
software on the Panther system. If less than two of the calibrator replicates are valid, the
software automatically invalidates the run. Samples in an invalidated run must be retested
using a freshly prepared calibrator and freshly prepared controls.
Internal Control
An IC is added to each sample with the wTCR. During processing, IC acceptance criteria are
automatically verified by the Panther system software. If an IC result is invalid, the sample
result is invalidated. Every sample with an invalid IC result must be retested to obtain a valid
result.
The Panther system software is designed to accurately verify processes when procedures
are performed following the instructions provided in this package insert and the Panther/
Panther Fusion System Operator's Manual.
Test Interpretation
Test results are automatically determined by the assay software. The table below shows the
possible results reported in a valid run and result interpretations. Samples with invalid test
results should be retested.
Table 1: Result Interpretation
BV Result Interpretation
Positive Positive for BV
BV Result Interpretation
Negative Negative for BV
Invalid Invalid test
Limitations
A. Use of this assay is limited to personnel who have been trained in the procedure. Failure
to follow the instructions given in this package insert may result in erroneous results.
B. The effects of other potential variables such as vaginal discharge, use of tampons, and
specimen collection variables have not been determined.
C. Performance with specimen types other than vaginal swab specimens has not been
evaluated.
D. Reliable results are dependent on adequate specimen collection, transport, storage, and
processing. Failure to observe proper procedures in any one of these steps can lead to
incorrect results. Because the transport system used for this assay does not permit
microscopic assessment of specimen adequacy, training of clinicians in proper specimen
collection techniques is necessary. See Specimen Collection and Storage for instructions.
For detailed information, refer to the appropriate instructions for use.
E. Therapeutic failure or success cannot be determined with the Aptima BV assay since
nucleic acid may persist following appropriate antimicrobial therapy.
F. Bacterial species targeted by the Aptima BV assay may comprise part of the normal
microbiome for a significant number of women; a BV positive result should be interpreted
in conjunction with other clinical data available to the clinician.
G. A negative result does not preclude a possible infection. Test results may be affected by
improper specimen collection, technical error, or specimen mix-up.
I. Performance of the assay has not been evaluated in individuals less than 14 years of
age.
K. The Aptima BV assay has not been evaluated for use with specimens collected by
patients at home.
L. Collection and testing of patient-collected vaginal swab specimens with the Aptima BV
assay is not intended to replace clinical examination.
O. Interference with the Aptima BV assay was observed in the presence of the following
substances: Mucus (1.5% V/V), Vaginal Moisturizing Gel (0.5% W/V) and Tioconazole
(5% W/V).
P. Cross-reactivity was observed with the Aptima BV assay in the presence of Lactobacillus
acidophilus 1x104 CFU/mL).
Q. A positive test result does not necessarily indicate the presence of viable organisms. A
positive result is indicative of the presence of target RNA.
Reproducibility
Aptima BV assay reproducibility was evaluated on the Panther system at three US sites
using seven panel members. Two operators performed testing at each site. Each operator
performed one run per day over six days using one reagent lot over the course of testing.
Each run had three replicates of each panel member.
The panel members were made using a simulated vaginal swab matrix (‘SVSM’, which
contains specimen transport media (STM) spiked with simulated vaginal fluid) negative for
Lactobacillus species, G. vaginalis, and A. vaginae. Six panel members contained cell lysates
of at least 1 of the following organisms: L. crispatus, L. jensenii, G. vaginalis, or A. vaginae;
different bacterial combinations were prepared to represent the variety of targeted BV
organism combinations present in vaginal specimens. One negative panel member contained
only the matrix with no added target analytes.
The agreement with expected results was 100% for all panel members.
Signal variability of the Aptima BV assay was calculated for each target in analyte positive
panel members. Only samples with valid results were included in the analyses. Variability,
calculated between sites, between operators, between days, between runs, within run, and
overall, is shown in Table 3 to Table 5 for Lactobacillus, G. vaginalis, and A. vaginae positive
panel members, respectively.
L. jensenii
108 24.31 0.00 0.00 0.77 3.16 0.00 0.00 0.80 3.28 0.15 0.62 1.12 4.60
BV Low Positive2
G. vaginalis
108 14.33 0.30 2.07 0.37 2.58 0.00 0.00 0.35 2.41 0.14 0.98 0.60 4.21
Moderate Positive
A. vaginae
108 14.95 0.38 2.52 0.41 2.75 0.00 0.00 0.39 2.61 0.14 0.93 0.69 4.64
Low Positive
A. vaginae
108 14.94 0.41 2.76 0.37 2.51 0.00 0.00 0.37 2.45 0.17 1.13 0.69 4.60
BV Low Positive2
A. vaginae
108 13.99 0.29 2.08 0.36 2.60 0.03 0.18 0.39 2.82 0.14 1.00 0.63 4.48
Moderate Positive
Note: In the event that variability from some factors is numerically negative, SD and CV are shown as 0.00.
Characteristics Total
Total, N N 1417
Age (years) Mean ± SD 34.7 ± 11.11
Median 33.0
Range 14-75
Age category (years), n (%) 14-17 4 (0.3)
18-29 537 (37.9)
30-39 469 (33.1)
40-49 235 (16.6)
>50 172 (12.1)
Race/Ethnicity, n (%) Asian 67 (4.7)
Black or African American 731 (51.6)
White (Hispanic or Latino) 248 (17.5)
White (Not Hispanic or Latino) 307 (21.7)
Other1 64 (4.5)
1 Includes patient-reported other, mixed, and unknown races.
For the 1417 evaluable subjects, 1413 clinician-collected vaginal swab samples and 1405
patient-collected vaginal swab samples were included in the analyses. The sensitivity and
specificity of the Aptima BV assay for the detection of BV are shown for both sample types
overall and by site in Table 7. Assay performance is shown stratified by race/ethnicity in
Table 8, and by clinical condition in Table 9.
White
43.9% (18/41)
(Hispanic/Latino)
White
15.9% (7/44)
(Not Hispanic/Latino)
Invalid Rates
A total of 3175 clinician- and patient-collected samples from symptomatic and asymptomatic
subjects were processed in valid Aptima BV runs to establish clinical performance. Of these,
0.7% had initial invalid results. Upon retest, 0.1% remained invalid and were excluded from
all analyses.
Analytical Sensitivity
The analytical sensitivity (Limit of Detection or LoD) and BV positivity limits of the Aptima BV
assay were determined by testing a series of panels consisting of L. crispatus, L. gasseri,
L. jensenii, G. vaginalis, or A. vaginae cell lysates diluted into simulated vaginal swab matrix
(SVSM). A minimum of 20 replicates of each panel member were tested with each of two
reagent lots for a minimum of 40 replicates per panel member. The predicted detection limits
for each organism calculated using Probit analysis are shown in Table 11.
1 Predicted BV Positivity Limits (C ) for A. vaginae and G. vaginalis in the Aptima BV assay are
95
approximately 5.10 log CFU/mL and 4.86 log CFU/mL, respectively.
Analytical Inclusivity
Five strains of each target organism were tested using lysate targeting 3X C95 for G. vaginalis
and A. vaginae, and 3X LoD for Lactobacillus species (L. crispatus, L. gasseri, and
L. jensenii) in SVSM. The Aptima BV Assay was BV positive for all five strains of G.
vaginalis and A. vaginae at 3X C95. All five strains of L. crispatus and L. gasseri were
detected at 3X LoD. Three of the five strains of L. jensenii were detected at 3X LoD, and the
remaining two strains at 10X LoD.
CFU = Colony Forming Units; IFU = Inclusion Forming Units; TCID50 = Median Tissue Culture Infectious Dose.
1 In Vitro Transcript tested.
2 Lactobacillus acidophilus affects BV positivity at 1x104 CFU/mL or higher.
Interference
Potentially interfering substances were tested in the Aptima BV assay. Panels were built in
SVSM and evaluated for potential effects on assay sensitivity and specificity. Sensitivity
performance was evaluated separately for L. crispatus by spiking lysate at 3X LoD, and for
G. vaginalis and A. vaginae by spiking lysate at 3X C95. Negative panels containing each
substance were also evaluated for specificity.
No interference was observed in the presence of the following exogenous and endogenous
substances tested at the concentrations listed in Table 13.
0
SVSM 0/168 0%
(0.0-1.6)
L. crispatus, A. vaginae 0
0 /168 <5%
BV Negative (0.0-1.6)
L. crispatus, G. vaginalis 45.2
76 /168 20-80%
BV High Negative (37.9-52.8)
L. crispatus, G. vaginalis, A. vaginae 79.4
131/1651 20-80%
BV High Negative (72.6-84.9)
G. vaginalis 100
168/168 ≥95%
BV Low Positive (98.4-100.0)
A. vaginae 100
168/168 ≥95%
BV Low Positive (98.4-100.0)
L. jensenii, A. vaginae 100
168/168 ≥95%
BV Low Positive (98.4-100.0)
G. vaginalis, A. vaginae 100
168/168 ≥95%
BV Low Positive (98.4-100.0)
L. crispatus, G. vaginalis, A. vaginae 100
168/168 ≥95%
BV Low Positive (98.4-100.0)
G. vaginalis 100
168/168 100%
BV Mod Positive (98.4-100.0)
A. vaginae 100
168/168 100%
BV Mod Positive (98.4-100.0)
1
Three invalid results were excluded from the analysis.
Panel Mean
N SD CV (%) SD CV (%) SD CV (%) SD CV (%) SD CV (%) SD CV (%) SD CV (%)
Description TTime1
L. crispatus
168 19.87 0.10 0.49 0.16 0.80 0.14 0.71 1.03 5.18 0.17 0.09 0.18 0.93 1.08 5.46
BV Negative2
L. crispatus
168 23.95 0.11 0.47 0.12 0.52 0.19 0.79 1.22 5.11 0.18 0.77 0.28 1.15 1.29 5.40
BV High Negative2
L. crispatus
1654 22.40 0.09 0.40 0.17 0.74 0.20 0.87 1.22 5.47 0.09 0.39 0.27 1.21 1.29 5.74
BV High Negative3
L. jensenii
168 24.80 0.10 0.38 0.14 0.57 0.14 0.57 1.33 5.35 0.17 0.69 0.25 1.01 1.38 5.56
BV Low Positive2
L. crispatus
168 23.51 0.15 0.63 0.09 0.40 0.17 0.73 1.36 5.77 0.10 0.44 0.31 1.31 1.42 6.02
BV Low Positive3
CV = Coefficient of variation.
1 TTime is shown for Lactobacillus only.
2 Panel member contains 2 different organisms: results are shown for only the Lactobacillus component.
3 Panel member contains 3 different organisms: results are shown for only the Lactobacillus component.
4 Three invalid results were excluded from the analysis.
Note: Variability from some factors may be numerically negative. This can occur if the variability due to those factors is very small.
In these cases, SD and CV are shown as 0.00.
Panel Mean
N SD CV (%) SD CV (%) SD CV (%) SD CV (%) SD CV (%) SD CV (%) SD CV (%)
Description TTime1
G. vaginalis
168 17.11 0.00 0.00 0.18 1.08 0.17 0.99 0.47 2.75 0.17 0.96 0.16 0.94 0.58 3.39
BV High Negative2
G. vaginalis
1654 15.71 0.00 0.00 0.19 1.19 0.18 1.12 0.48 3.05 0.11 0.72 0.12 0.79 0.57 3.62
BV High Negative3
G. vaginalis
168 15.80 0.00 0.00 0.16 1.00 0.14 0.89 0.43 2.70 0.15 0.97 0.15 0.92 0.52 3.30
BV Low Positive
G. vaginalis
168 14.46 0.00 0.00 0.17 1.18 0.05 0.35 0.38 2.63 0.16 1.09 0.18 1.25 0.48 3.35
BV Mod Positive
G. vaginalis
168 15.01 0.00 0.00 0.14 0.93 0.14 0.91 0.40 2.67 0.16 1.08 0.13 0.86 0.49 3.28
BV Low Positive2
G. vaginalis
168 14.06 0.00 0.00 0.16 1.11 0.15 1.09 0.39 2.75 0.14 0.99 0.16 1.16 0.49 3.51
BV Low Positive3
Panel Mean
N SD CV (%) SD CV (%) SD CV (%) SD CV (%) SD CV (%) SD CV (%) SD CV (%)
Description TTime1
A. vaginae
168 18.20 0.02 0.11 0.25 1.36 0.15 0.84 0.58 3.17 0.19 1.02 0.19 1.05 0.70 3.84
BV Negative2
A. vaginae
1654 16.56 0.00 0.00 0.25 1.53 0.18 1.11 0.56 3.38 0.13 0.79 0.12 0.70 0.67 4.02
BV High Negative3
A. vaginae
168 15.11 0.00 0.00 0.19 1.25 0.15 0.97 0.51 3.40 0.12 0.82 0.12 0.78 0.59 3.92
BV Low Positive
A. vaginae
168 15.13 0.00 0.00 0.20 1.30 0.12 0.80 0.51 3.34 0.14 0.89 0.16 1.07 0.59 3.92
BV Low Positive2
A. vaginae
168 14.13 0.08 0.54 0.21 1.50 0.17 1.21 0.51 3.63 0.08 0.57 0.20 1.40 0.62 4.41
BV Mod Positive
A. vaginae
168 15.78 0.03 0.16 0.17 1.09 0.10 0.65 0.50 3.17 0.16 1.00 0.12 0.75 0.57 3.64
BV Low Positive2
A. vaginae
168 15.61 0.00 0.00 0.23 1.47 0.15 0.94 0.51 3.29 0.10 0.66 0.18 1.15 0.62 3.95
BV Low Positive3
Bibliography
1. Hainer BL, Gibson MV. Vaginitis: Diagnosis and Treatment. Am Fam Physician. 2011 Apr 1;83(7):807-815.
2. Granato PA. Vaginitis: Clinical and Laboratory Aspects for Diagnosis. Clin Microbiol Newsletter. 2010 Aug 1,(15): 111-116.
3. Onderdonk AB, Delaney ML, Fichorova RN. The Human Microbiome during Bacterial Vaginosis. Clin Microbiol Rev. 2016
Apr;29(2):223-38. doi: 10.1128/CMR.00075-15.
4. Haggerty CL, Hillier SL, Bass DC, Ness RB; PID Evaluation and Clinical Health study investigators. Bacterial vaginosis and
anaerobic bacteria are associated with endometritis. Clin Infect Dis. 2004 Oct 1;39(7):990-5. Epub 2004 Sep 2.
5. Marrazzo JM, Wiesenfeld HC, Murray PJ, Busse B, Meyn L, Krohn M, Hillier SL. Risk factors for cervicitis among women with
bacterial vaginosis. J Infect Dis. 2006 Mar 1;193(5):617-624. Epub 2006 Feb 2.
6. Bautista CT, Wurapa EK, Sateren WB, Morris SM, Hollingsworth BP, Sanchez JL. Association of Bacterial Vaginosis with
Chlamydia and Gonorrhea Among Women in the U.S. Army. Am J Prev Med. 2017;52(5):632-639. doi: 10.1016/
j.amepre.2016.09.016.
7. Cherpes TL, Meyn LA, Krohn MA, Lurie JG, Hillier SL. Association between acquisition of herpes simplex virus type 2 in
women and bacterial vaginosis. Clin Infect Dis. 2003 Aug 1;37(3):319-325.
8. Cohen CR, Lingappa JR, Baeten JM, et al. Bacterial vaginosis associated with increased risk of female-to-male HIV-1
transmission: a prospective cohort analysis among African couples. PLoS Med. 2012;9(6):e1001251. doi: 10.1371/
journal.pmed.1001251.
9. Işik G, Demirezen, Dönmez HG, Beksaç MS. Bacterial vaginosis in association with spontaneous abortion and recurrent
pregnancy losses. J Cytol. 2016 Jul-Sep;33(3):135-140.
10. Donders GG, Van Calsteren K, Bellen G, et al. Predictive value for preterm birth of abnormal vaginal flora, bacterial vaginosis
and aerobic vaginitis during the first trimester of pregnancy. BJOG. 2009 Sep;116(10):1315-24.
11. Amsel R, Totten PA, Spiegel CA, Chen KCS, Eschenbach DA, Holmes KK. Nonspecific vaginitis: diagnostic criteria and
epidemiologic associations. Am J Med. 1983 74:14–22.
12. Nugent RP, Krohn MA, Hillier SL. Reliability of Diagnosing Bacterial Vaginosis Is Improved by a Standardized Method of Gram
Stain Interpretation. J Clin Microbiol. Feb 1991, 29(2): 297-301.
13. Plummer EL, Garland SM, Bradshaw CS, et al. Molecular diagnosis of bacterial vaginosis: Does adjustment for total bacterial
load or human cellular content improve diagnostic performance? J Microbiol Methods. 2017 Feb;133:66-68. doi: 10.1016/
j.mimet.2016.12.024. Epub 2016 Dec 29.
14. Clinical and Laboratory Standards Institute. 2006. Collection, Transport, Preparation, and Storage of Specimens for Molecular
Methods; Approved Guideline. CLSI Document MM13. Wayne, PA.
15. Centers for Disease Control and Prevention/National Institutes of Health. 2009. Biosafety in Microbiological and Biomedical
Laboratories (BMBL).
16. Clinical and Laboratory Standards Institute. 2011. Clinical Laboratory Waste Management. CLSI Document GP5. Villanova,
PA.
17. Centers for Disease Control and Prevention. 2015. United States Morbidity and Mortality Weekly Report. Recommendations
and Reports, Vol. 64, No. 3.
For country-specific Technical Support and Customer Service email address and telephone number, visit www.hologic.com/
support.
Serious incidents occurring in relation to the device in the European Union should be reported to the manufacturer and the
competent authority of the Member State in which the user and/or the patient is established.
Hologic, Aptima, TMA, Panther, and associated logos are trademarks and/or registered trademarks of Hologic, Inc. and/or its
subsidiaries in the United States and/or other countries.
All other trademarks, registered trademarks, and product names that may appear in this package insert are the property of their
respective owners.
This product may be covered by one or more U.S. patents identified at www.hologic.com/patents.
©2022 Hologic, Inc. All rights reserved.
AW-23712-001 Rev. 001
2022-08