Technische Aspecten GAITrite Walkway System
Technische Aspecten GAITrite Walkway System
Technische Aspecten GAITrite Walkway System
L Page 1 of 50
CONTENTS
The GAITRite System is an electronic walkway utilized to measure the temporal (timing) and spatial (two
dimension geometric position) parameters of its pressure activated sensors. The GAITRite system’s intent is to
be utilized as a measuring device for the events occurring during biped and quadruped locomotion.
Inferred parameters are easily obtained by applying common physics and math formulas to the directly
measured temporal and spatial data i.e calculate velocity, relationships between spatial and temporal events
etc.
GAITRite
Walkway
!
AC Adaptor
Master Label
! !
GAITRite
connection Computer USB Port
connection
• Make sure you have read and understood all the information listed in this document and
the User’s Manual. If you require additional information and/or training, please contact
our Customer Service.
• The company assumes that the user is trained in relevant professional fields or is familiar
with the measurements provided by the GAITRite system.
• The company assumes that you have read and understood the intended use of the
GAITRite system and the safety instructions.
• Before allowing subjects to ambulate across the GAITRite walkway, the subjects must be
given adequate support, based on an assessment of their gait and balance abilities.
• Barefoot testing is not recommended. If the protocol requires you to test barefoot, then
you will have to setup your own quality procedure that follows the disinfection rules per
our cleaning instructions and accounts for allergy/irritation produced, on the bare feet,
from the walkway’s top surface.
• The GAITRite equipment should be installed and set up in the test area prior to the
arrival of any Subjects.
• Cables must be routed in a safe manner to avoid any risk of tripping Subjects and users.
WARNING Adhesive tape can be utilized to affix cables on the floor.
• For safety, prepare the subjects by pointing out the tripping hazards posed by the
perimeter height of the walkway and the routed cables.
• For safety, Subjects should stand on the “dead” area of the walkway before beginning a
test, thus avoiding to trip at the entry point of the walkway.
• The test area must be large enough for a fully unrolled walkway and also provide ample
space at the beginning and end of the walkway for the Subjects to safely start and finish
their walk.
• When lifting, packing or relocating the GAITRite walkway, make sure that you observe
proper lifting for heavy items, 60-120 Lbs.
• The floor of the test area must be clean and completely dry prior to unrolling the
walkway.
• Before testing Subjects, make sure their shoes do not have points or sharp edges that
can damage the surface of the walkway.
Do not use power adapters and/or cables other than those supplied with your
CAUTION GAITRite equipment.
If the walkway is not working properly follow these steps to determine the area of the problem:
1. Check the power light on the Interface Module, if it is ON=lit go to step 2. If the light is not on check the
wall outlet, if determined that it is working properly then contact our Customer service department.
2. Check the Green power light, located on the Master cover of the GAITRite Electronic Walkway, if it is
not ON=lit then be sure the cables are securely attached. If you determined that all cables are secured
then the problem is a failure of the Interface Cable contact our Customer service department.
The GAITRite does not contain any serviceable parts. Please, contact our Technical Support department.
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 9 of 50
• Before cleaning ensure that the GAITRite equipment is disconnected from the mains
(power outlets) and any peripheral devices.
• The walkway may be cleaned with a damp cloth using any proprietary solution of mild
detergent or disinfectant. Do not use solvents or abrasives.
WARNING • Afterwards, moisten a clean cloth with clear, cold water and wipe the walkway surface.
• Do not place the walkway near any flame or ignition sources, and do not drop lighted
cigarettes/matches on or near the walkway.
• Always lay the container on its side. As indicated by the labels do not stand it on end.
• Do not fold, crease or puncture the walkway, as this will void the warranty.
• Never move the walkway without first rolling it up on the supplied drum.
• If you decide to allow the walkway to permanently lay on the floor then make sure you
WARNING provide adequate protection from rolling heavy equipment over the walkway.
• If you decide to allow the walkway to permanently lay on the floor then make sure you
provide adequate signs to warn the floor cleaning personnel to keep a distance while
mopping the floors.
• If you decide to allow the walkway to permanently lay on the floor then make sure that
the area will not flood during storms. If uncertain, please remove the AC Adapter from
the mains, when the walkway is not in use.
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 10 of 50
Condition Specification
-29 to 50 degrees C
(-20 to 122 degrees F)
Storage & Transportation Temperature/Humidity
Relative humidity up to 80% (up to 31 degrees C)
decreasing linearly to 50% at 40 degrees C
10 to 40 degrees C
Operating (40 to 104 degrees F)
Temperature/Humidity Relative humidity up to 80% (up to 31 degrees C)
decreasing linearly to 50% at 40 degrees C
WARNING
General Use
INDOOR USE ONLY. Not suitable for wet locations.
WARNING
NOT SUITABLE FOR USE IN THE PRESENCE OF
FLAMMABLE ANESTHETIC MIXTURE WITH AIR,
Flammable Gases
OXYGEN, NITROUS OXIDE, WHERE SUCH GASES
MAY ACCUMULATE IN CONCETRATION (CLOSED
SPACE)
Active Area
24” (±.015) x 240” (±.15”) (60.96cm x 609.6cm)
(for 10 pad walkway)
Walkway-to-Walkway
±.5” (±1.27cm)
Spatial Accuracy
Walkway-to-Walkway
±1 sample
Temporal Accuracy
Walkway-to-Walkway
±.5 switching level
Switching Level Accuracy
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Gaitrite
Walkway System requires continued operation during power mains interruptions, it is recommended that the
Gaitrite Walkway System be powered from an uninterruptible power supply or battery. Certifications for the
specified AC Adaptor can be furnished upon request.
Rating Specification
WARNING
UTILIZE ONLY THE AC ADAPTOR SUPPLIED WITH YOUR EQUIPMENT.
AC Adapter 100-240V- .8-.4 A
Input Voltage
AC Adapter 9 VDC 3.0A
Output Voltage
The GAITRite System has been tested and found to comply with the limits for medical devices in IEC 60601-1-
2. These are designed to provide reasonable protection against harmful interference in a typical medical
installation. In the event of interference, power devices from separate mains supplies and/or increase physical
distance between devices. Contact CIR Systems Inc. Customer Service if you have any questions.
Voltage dips, short >95% dip 0.5 cycle Complies Mains power quality should be that of a
interrupts and voltage typical commercial or hospital environment.
variations on power 60% dip 5 cycles If the user of the Gaitrite Walkway System
supply input lines requires continued operation during power
IEC 61000-4-11 70% dip 25 cycles mains interruptions, it is recommended that
the Gaitrite Walkway System be powered
95% dip 5 sec. from an uninterruptible power supply or
battery.
Disposal of the walkway should be done according to the requirements of the local authorities. The materials
included in the walkway are listed in the “GAITRite Materials Specifications” section of this document. Local
Authorities may require separation of the electronic devices and recycling during disposal.
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 14 of 50
This document describes the use and specifications for the GAITRite Interface Module. The Interface Module
is not an accessory; it is an integral part of the GAITRite System. It supplies power to the GAITRite walkway,
provides the communications cabling between the GAITRite walkway and the host personal computer and
provides means to interface and synchronize with external systems.
The GAITRite Interface Module includes standard utilities which are hard-wired, always enabled, and available
to communicate with all GAITRite software modules and versions of the GAITRite Electronic Walkway. The
standard utilities are:
The GAITRite Interface Module includes optional utilities. Optional utilities must be purchased separately and
are enabled from the GAITRite Software. The optional utilities are:
This input connection is designed to accept an external transformer. The AC Adaptor provides power to the
Interface Module and to the GAITRite walkway and was selected with very specific characteristics.
Do not use power adapters and/or cables other than those supplied with your
WARNING GAITRite equipment.
If AC Adaptor failure is suspected, please call our Technical Support Department for a replacement.
To apply power, first connect the AC plug of the external transformer to the AC wall outlet (mains) and then
insert the other end, 2.1 mm plug, to the Interface Module Power Input connector. When these connections
are done correctly the green light marked “POWER” will be lit (ON) and power is available for the GAITRite
Interface Module and the GAITRite electronic walkway. Specifications and certifications for the specified AC
Adaptor can be furnished upon request.
Antenna (Optional)
Internal
SPEAKER
METRONOME STATUS
POWER INPUT
Connector
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 16 of 50
A cable, six feet long, has one end hard-wired with the GAITRite Interface Module; the other end has a USB
connector ready to plug to a standard USB 2.0 port of a host computer.
Note An extension cable can be utilized to extend the range of this connection. The total
length of this connection should not exceed 12 feet.
With power applied to the GAITRite Interface Module connect this cable to the USB port of your GAITRite
computer. The red light associated with this connection, marked “COMPUTER”, will be not lit (OFF) if the
computer has successfully communicated with GAITRite Interface Module. The red light will be lit (ON) if the
computer can not successfully communicate with the GAITRite Interface Module.
Antenna (Optional)
Internal
SPEAKER
METRONOME STATUS
POWER INPUT
Connector
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 17 of 50
A cable, nine feet long, has one end hard-wired with the GAITRite Interface Module; the other end has a
“Firewire” connector ready to plug to a Platinum Gaitrite System.
An approved extension cable can be utilized to extend the range of this connection.
Note The total length should not exceed 48 feet. Approved cables are available from our
Technical Support Department.
With power applied to the GAITRite Interface Module connect this cable to the walkway. The red light
associated with this connection, marked “GAITRite”, indicates the status of the connection between the
Interface Module and the walkway. The light will not be lit (OFF) if the connection is successful. The light will
be lit (ON) if the connector is not plugged into the walkway or the Interface Module can not communicate with
the walkway.
Antenna (Optional)
Internal
SPEAKER
METRONOME STATUS
POWER INPUT
Connector
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 18 of 50
A yellow light, marked “STATUS”, is utilized to display errors. It blinks when a problem is encountered. The
blinking rate and the number of blinks are utilized by our Technical Support Department to identify the
problem. You are encouraged to call our Technical Support Department if you observe a blinking status light
after you start a test.
Antenna (Optional)
Internal
SPEAKER
METRONOME STATUS
POWER INPUT
Connector
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 19 of 50
The SYNC OUTPUT connection is utilized to synchronize external systems with the GAITRite walkway such
as Electromyography equipment etc.
Antenna (Optional)
Internal
SPEAKER
METRONOME STATUS
POWER INPUT
Connector
A Yellow light, marked “SYNC OUTPUT” is employed as a visual indicator for the functions associated with
this connection.
Logic transitions for the SYNC OUTPUT signal can be selected from a list included in the GAITRite Software.
The options are shown below:
Signal at
SYNC OUTPUT connector Description
Start with SYNCOUT high. During the When this option is selected the SYNCOUT starts at level high,
first scan change SYNCOUT form High- SYNCOUT light=ON.
to-Low. At end of test change During the first scan the SYNCOUT signal changes state to level
SYNCOUT back to high. low, the light turns OFF, not lit. It continues on this state until an
end of test; the SYNCOUT signal will go back to level high and
the light will turn ON.
Start with SYNCOUT low. Upon When this option is selected SYNCOUT starts at level low,
detecting first sensor activity change SYNCOUT light=OFF.
SYNCOUT form Low-to-High. At end of The SYNCOUT will remain low until the first sensor change is
test change SYNCOUT back to low. detected: during this scan the SYNCOUT signal will change
state to level high, the light will turn ON and will continue on this
state until an end of test: the output signal will go back to level
low and the light will turn OFF.
Start with SYNCOUT high. Upon When this option is selected SYNCOUT starts at level high,
detecting first sensor activity change SYNCOUT light=ON.
SYNCOUT form High-to-Low. At end of The SYNCOUT will remain high until the first sensor change is
test change SYNCOUT back to high. detected: during this scan the SYNCOUT signal will change
state to level low, the light will turn OFF and will continue on this
state until an end of test: the output signal will go back to level
high and the light will turn ON.
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 21 of 50
The diagram below provides a detailed view of the connections required at the SYNC OUTPUT connector.
Users can make their own interface cable to use between two systems or our company can provide you one.
Please call our Technical Support Department.
METRONOME STATUS
SYNC OUTPUT
connection
Errata: 04/20/2011
This section applies to Interface Devices with “Orange” label and version numbers “F” or greater. For
all other versions, please contact our Support Department.
The SYNC INPUT connection is utilized to capture the logic state of an external signal identified as an event
by the GAITRite system. Events are inserted in the data stream of the GAITRite walkway as they occur (in
real time). Later these events will appear as time marks in the temporal parameters review.
Antenna (Optional)
Internal
SPEAKER
METRONOME STATUS
POWER INPUT
Connector
A Yellow light, marked “SYNC INPUT” is employed as a visual indicator for the functions associated with this
connection.
The functionality of the SYNC INPUT port is configured and enabled from your GAITRite Software. The SYNC
input port can be configured for the following:
ADDITIONAL
HARDWARE
FUNCTION REQUIRED RESULTS
Mark Event GAITRite Event When the switch gets pressed, the yellow light will
Switch with wires or go ON and a mark will be placed on the current
wireless frame of the GAITRite data.
Mark Event with audible GAITRite Event When the switch gets pressed, the yellow light will
tone Switch with wires or go ON, a mark will be placed on the current frame
wireless of the GAITRite data and an audible tone will be
generated by the GAITRite Interface Module.
External System Trigger External system To synchronize the GAITRite System with an
compatible with the external system the signal of the external system
electrical must meet the electrical characteristics described
characteristics of the below.
Sync Input Port
detailed below. On every logical transition (signal edge) of the
signal at the SYNC INPUT, the GAITRite walkway
will insert a mark in the GAITRite data stream.
The SYNC INPUT signal includes an internal pull-up resistor (2.2KOhm), and has the following
electrical characteristics:
The diagram below provides a detailed view of the connections required at the SYNC INPUT connector.
Users can make their own interface cable to use between two systems or our company can provide you one.
Please call our Technical Support Department.
METRONOME STATUS
SYNC INPUT
connection
SYNC
INPUT SYNC
signal OUTPUT
GND
SYNC
INPUT
The following diagram represents the reporting activities of the Sync Input port.
The state of the SYNC INPUT, A and A1, gets reported on the First Frame always. Subsequent
changes of the SYNC INPUT get reported within one frame of occurrence.
SYNC INPUT
FIRST FRAME STATE
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
FRAME COUNTER
SYNC INPUT
STATE A
C E
B D
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
FRAME COUNTER
SYNC INPUT
STATE A1
B
B1 D1
C1 E1
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 27 of 50
3.8 Metronome
The Metronome function of the GAITRite Interface Module provides an audible tone at pre-selected rates
(beats/minute). A yellow light labeled “METRONOME” is associated with this function and blinks at the rate
(beats/minute). The audible tone is generated by an internal speaker located above the metronome yellow
light. Please avoid covering this area of the Interface Module enclosure.
You can enable the metronome and select the beat rate through the GAITRite Software. By default the
metronome function is disabled.
Antenna (Optional)
Internal
SPEAKER
METRONOME STATUS
POWER INPUT
Connector
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 28 of 50
Encapsulated within the electronic walkway are sensor pads. Each sensor pad has an active area of 24
inches square (61cm square) and contains 2,304 sensors arranged in (48x48) grid pattern. The sensors are
placed on .5 inch (1.27cm) centers. Multiple sensor pads are connected to form the desired length of the
walkway. Measurements are provided using x,y coordinates of sensors, with distance converted to cm and
time in seconds up to an accuracy of 6dp. Distance may be either the distance travelled along the plane of the
walkway or the calculated distance between locations on the walkway.
1
The method for defining the quadrilateral of a footprint was first developed by Shores and later improved upon
2
by Lisa Selby-Silverstein , while an automated improved version of the algorithm has been implemented by the
GAITRite system. The following steps refer to Figure 3.
5. The formed quadrilateral (ALRG), in this case a trapezoid, encloses the footprint efficiently.
Refer to Figure 3 and identify points (C) and (E); these two points divide the medial line (AG)
into three equal spaces.
From point (C), draw a line perpendicular to the medial line and extended it in order to intercept
the lateral line at point (N). Similarly, from point (E), draw a line perpendicular to the medial
line and extended it in order to intercept the lateral line at point (P).
Quadrilateral (ALNC) contains the Heel area sensors, (CNPE) contains the mid foot area
sensors and (EPRG) contains the toe area sensors.
1. Shores M. Footprint analysis in gait documentation: an instructional sheet format, Phys Ther 60:1163, 1980.
2. Silverstein LS: the effect of neutral position foot orthoses on gait of children with down syndrome. Doctoral Thesis.
PA, 1993, Hahnemann University.
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 30 of 50
As illustrated in Figure 6, in this specific footprint the quadrilaterals form trapezoids, but in other cases the
quadrilaterals could form rectangles. Point (C1), the centroid of trapezoid (ALNC) represents the heel center of
the footprint. Point (C2), the centroid of trapezoid (EPRG) represents the toe/metatarsal center point. This
footprint is geometrically represented by twelve trapezoids; six medial and six lateral area. Quadrilaterals are
formed in the two dimensional representation of the footprint in order to isolate the sensors and later perform
calculations based on the isolated sensors of each quadrilateral.
The line that connects point C1 to point C2 is the midline of the footprint.
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 32 of 50
The walkway does not only sense the geometry of the activating footprints but also the relative arrangement
between them in a two dimensional space. Figure 7, illustrates three footprints.
It is calculated by multiplying the midline of the footprint (connecting point C1 to point C2 in Figure 6), by
1.5. The unit of measure is cm.
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 34 of 50
Figure 9
4.4.1 First Contact
It is the time that the first sensor appears in any quadrilateral. It is expressed in seconds (sec).
4.4.9 Velocity
It is obtained after dividing the Distance Traveled by the Ambulation time. It is expressed in centimeters per
second (cm/sec)
.
4.4.10 Mean Normalized Velocity
It is obtained after dividing the Velocity by the Average Leg Length and it is expressed in leg length per
second (LL/sec). The average Leg Length is computed (left leg length + right leg length)/2.
Heel only is that time that only the heel is on the floor:
Heel On - Heel On
Mid section is the time when the foot mid section is on the floor, excludes heel/toe only:
Mid section = Mid Off - Mid On
Toe only is the time that only the toe is on the floor:
Toe Off - Mid Off
Percentage values can only be calculated correctly for walks where the heel contact is the earliest foot event
and the toe off the last event.
The % Coefficient of variance is the standard deviation divided by the mean multiplied by 100.
SD calculations of foot width or foot length by Left & Right feet are not implemented.
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 39 of 50
The GAITRite walkway’s unique and patented sensor avoids false peripheral activation. Each sensor has
been constructed with two flexible elements riding on a pivot point. When pressure is applied to the sensor
both elements must flex around the pivot point to initiate activation, otherwise the pivot point will toggle the
sensor in either side without activation. After activation, the sensor begins to change its value linearly with
the vertical component of pressure exerted upon it.
The walkway contains thousands of sensors,
therefore calibration of each and every sensor
Color Switching level is not impossible but cost prohibitive. Pressure
values are normalized and expressed as a
Dark Gray 1=lowest percent of the maximum pressure and then
divided into levels. Currently there are seven
Light Gray 2 switching levels, illustrated in Table 1.
The walkway does not only sense the
Cyan 3 geometry of the activating footprints but also
the relative arrangement between them in a
Yellow 4 two dimensional plane and the relative vertical
component of pressure exerted by each
Magenta 5 footprint.
The pressure is represented by a switching
Red 6 level. As discussed earlier, the division of the
footprint provides us with twelve sections; six
Blue 7=highest sections are located in the medial side of the
footprint and six in the lateral. Each section
contains a number of activated sensors
Table 1. Switching Levels, Color Assignment. enclosed within a trapezoid. The GAITRite
algorithms utilize the sectional division to
identify the activated sensors included within
each quadrilateral and then perform
calculations to objectively describe the
behavior of the section.
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 40 of 50
As illustrated in Figure 11, medial trapezoid (AHIB) includes the following sensors: 1,2,3,4,6 and 7.
Lateral trapezoid (HLMI) includes the sensors: 5,10,11,15 and 16. A sensor can be claimed by only one
trapezoid and not shared by others. Sensor 4, has most of its surface area within trapezoid (AHIB) and
will be claimed only by this trapezoid. If the surface area of a sensor is equally divided among two or
more quadrilaterals, i.e. sensor 12, then the sensor will be claimed by the first trapezoid in the algorithm.
The results of the switching parameters are presented in the format illustrated in Table 2.
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 41 of 50
1. P*t for a section, is the sectional integrated pressure over time expressed as a percent of the overall
integrated pressure over time. The overall P*t can be found by summing the P*t of each sensor,
sectional P*t can be found by summing the P*t of the sensors included in the particular section also
expressed as a percent of the overall P*t. As illustrated in Figure 12, P*t for sensor 1 is the total area
under the curve calculated by:
2. (P*t)s1=(.020-.010)*1+(.030-.020)*2+ ......+(.070-.060)*6+.........+(.140-.130)*1
3. Peak Time for a section is the first time point that one or more sensors within a section was at the
maximum switching level. Assume that a section included only the two sensors, illustrated in Figure
12; the peak time is at .060 seconds. Time count begins from first contact of the footprint.
2
4. Area of a section is expressed in centimeters squared (cm ) and represents the sum of the active
sensor areas within a section. Each sensor has an area of .5 in x .5 in or (1.27 cm x 1.27 cm).
5. Peak P(ressure) for a section, is the maximum sectional switching level expressed as a percent of the
overall maximum switching level. Sectional switching level occurs at the peak time of the section.
Assume that a section included only the two sensors illustrated in Figure 12, then peak time is at .060
seconds and a peak switching level of 6, then expressed as a percent of the overall maximum
switching level.
5.1 Software
The software is licensed to the end user for use with the GAITRite System. It may be installed on any number
of computers, within the purchasing organization, for this application without any additional fee. Copies of the
software may not be resold. No transfer of copyright is granted or implied by this agreement.
5.4 WARRANTY
This product is warranted for 12 (twelve) months from date of shipment, for manufacturing defects and/or
component failures. Defective equipment will be replaced or repaired at the discretion of CIR Systems Inc.
Warranty will be voided by the limitations listed in this document.
Source Data
Field Name set Description
Pt Record # Patient GAITRite Patient Record number
Patient Id Number Patient Patient Id Number
Medical Record # Patient Medical Record Number
Last Name Patient Last Name
First Name Patient First Name
Middle Initial Patient Middle Initial
Gender Patient Gender
Birthdate Patient Birth date
Age Patient Age
Height Patient Height
Weight Patient Weight
Primary Doctor Id Patient Primary Doctor Id
Primary Doctor Name Patient Primary Doctor Name
Secondary Dr Code Patient Secondary Dr Code
Secondary Dr Name Patient Secondary Dr Name
Leg Length Left Patient Leg Length Left
Leg Length Right Patient Leg Length Right
Problem Id Code Patient Problem Id Code
Problem Description Patient Problem Description
Registered on Date Patient Registered on Date
Last Tested On Date Patient Last Tested On Date
Last FAP score Patient Last FAP score
Prime Contact Name Patient Prime Contact Name
Phone number Patient Phone number
Fax Number Patient Fax Number
Address line 1 Patient Address line 1
Address Line 2 Patient Address Line 2
City Patient City
State Patient State
Zip Patient Zip
Country Patient Country
Location Patient Location
Discharge Date Patient Discharge Date
Discharge Flag Patient Discharge Flag
Footfall Id Pattern Code Patient Footfall Id Pattern Code
Test Record # GAITRite Test GAITRite Test Record number
Pt Record # GAITRite Test GAITRite Patient Record number
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 46 of 50
Source Data
Field Name set Description
Date / Time of Test GAITRite Test Date / Time of Test
Distance GAITRite Test Distance
Ambulation Time GAITRite Test Ambulation Time
Velocity GAITRite Test Velocity
Step Count GAITRite Test Step Count
Cadence GAITRite Test Cadence
Step Time(sec) L GAITRite Test Step Time(seconds) Left
Step Length(cm) L GAITRite Test Step Length(cm) Left
Step Time(sec) R GAITRite Test Step Time(sec) Right
Step Length(cm) R GAITRite Test Step Length(cm) Right
Step Extremity(ratio) L GAITRite Test Step Extremity(ratio) Left
Step Extremity(ratio) R GAITRite Test Step Extremity(ratio) Right
Cycle Time(sec) L GAITRite Test Cycle Time(seconds) Left
Cycle Time(sec) R GAITRite Test Cycle Time(seconds) Right
Stride Length(cm) L GAITRite Test Stride Length(cm) Left
Stride Length(cm) R GAITRite Test Stride Length(cm) Right
HH Base Support (cm) L GAITRite Test Heel Heel Base Support (cm) Left
HH Base Support (cm) R GAITRite Test Heel Heel Base Support (cm) Right
Swing % of Cycle L GAITRite Test Swing percent of Cycle Left
Swing % of Cycle R GAITRite Test Swing percent of Cycle Right
Swing Time(sec) L GAITRite Test Swing Time(seconds) Left
Swing Time(sec) R GAITRite Test Swing Time(seconds) Right
Stance % of Cycle L GAITRite Test Stance percent of Cycle Left
Stance % of Cycle R GAITRite Test Stance percent of Cycle Right
Stance Time(sec) L GAITRite Test Stance Time(seconds) Left
Stance Time(sec) R GAITRite Test Stance Time(seconds) Right
Single Supp % Cycle L GAITRite Test Single Support percent Cycle Left
Single Supp % Cycle R GAITRite Test Single Support percent Cycle Right
Single Supp. Time(sec) L GAITRite Test Single Support Time(sec) Left
Single Supp. Time(sec) R GAITRite Test Single Support Time(sec) Right
Double Supp % Cycle L GAITRite Test Double Support percent Cycle Left
Double Supp % Cycle R GAITRite Test Double Support percent Cycle Right
Double Supp. Time(sec) L GAITRite Test Double Support Time(sec) Left
Double Supp. Time(sec) R GAITRite Test Double Support Time(sec) Right
Step Time Differential GAITRite Test Left: Right foot Step time ratio
Step Length Differential GAITRite Test Left: Right foot Step Length ratio
Cycle Time Differential GAITRite Test Left: Right foot Cycle time ratio
Functional Amb. Profile GAITRite Test FAP score
Treatment Count GAITRite Test Treatment Count
Footfalls GAITRite Test Footfalls
Comments GAITRite Test Comments
Normalized Velocity GAITRite Test Velocity normalized using leg length
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 47 of 50
Source Data
Field Name set Description
Leg Length (cm) L GAITRite Test Leg Length (cm) Left
Leg Length (cm) R GAITRite Test Leg Length (cm) Right
Number of Passes GAITRite Test Number of Passes
Ambulatory Devices GAITRite Test User defined
Ambulatory Aids GAITRite Test User defined
User GAITRite Test User
Toe In / Out L GAITRite Test Angle of toe
Toe In / Out R GAITRite Test Angle of toe
Heel Off On Time L GAITRite Test Heel Off On Time Left
Heel Off On Time R GAITRite Test Heel Off On Time Right
Heel Off On Perc L GAITRite Test Heel Off On Percent Left
Heel Off On Perc R GAITRite Test Heel Off On Percent Right
Double Support Load Time L GAITRite Test Double Support Load Time Left
Double Support Load Time R GAITRite Test Double Support Load Time Right
Double Support Load %GC L GAITRite Test Double Support Load Percent Gait Cycle Left
Double Support Load %GC R GAITRite Test Double Support Load Percent Gait Cycle Right
Double Support Unload Time L GAITRite Test Double Support Unload Time Left
Double Support Unload Time R GAITRite Test Double Support Unload Time Right
Double Support Unload %GC L GAITRite Test Double Support Unload Percent Gait Cycle Left
Double Support Unload %GC R GAITRite Test Double Support Unload Percent Gait Cycle Right
Asst, Ortho, etc. 1 GAITRite Test User defined
Asst, Ortho, etc. 2 GAITRite Test User defined
Asst, Ortho, etc. 3 GAITRite Test User defined
Asst, Ortho, etc. 4 GAITRite Test User defined
Asst, Ortho, etc. 5 GAITRite Test User defined
Type of Test GAITRite Test User defined
Software Version GAITRite Test Software Version
Number of Pads GAITRite Test Number of Pads
Normal Group Code GAITRite Test Normal Group Code
Standard Walk Steps GAITRite Test Standard Walk Steps
Standard Amb Time GAITRite Test Standard Ambulation Time
Stride Velocity Left GAITRite Test Stride Velocity Left
Stride Velocity Right GAITRite Test Stride Velocity Right
Shoe Flag GAITRite Test Shoe Flag
Expected Foot Length GAITRite Test Expected Foot Length
Expected Foot Width GAITRite Test Expected Foot Width
Foot Length Left GAITRite Test Foot Length Left
Foot Width Left GAITRite Test Foot Width Left
Foot Length Right GAITRite Test Foot Length Right
Foot Width Right GAITRite Test Foot Width Right
Step Len Std Dev Left GAITRite Test Step Length Standard Deviation Left
Step Len Std Dev Right GAITRite Test Step Length Standard Deviation Right
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 48 of 50
Source Data
Field Name set Description
Step Time Std Dev Left GAITRite Test Step Time Standard Deviation Left
Step Time Std Dev Right GAITRite Test Step Time Standard Deviation Right
Stride Length Std Dev Left GAITRite Test Stride Length Standard Deviation Left
Stride Length Std Dev Right GAITRite Test Stride Length Standard Deviation Right
Stride Time Std Dev Left GAITRite Test Stride Time Standard Deviation Left
Stride Time Std Dev Right GAITRite Test Stride Time Standard Deviation Right
Swing Time Std Dev Left GAITRite Test Swing Time Standard Deviation Left
Swing Time Std Dev Right! GAITRite Test Swing Time Standard Deviation Right
Stance Time Std Dev Left GAITRite Test Stance Time Standard Deviation Left
Stance Time Std Dev Right GAITRite Test Stance Time Standard Deviation Right
Stride Velocity Std Dev Left GAITRite Test Stride Velocity Standard Deviation Left
Stride Velocity Std Dev Right GAITRite Test Stride Velocity Standard Deviation Right
Single Supp Time Std Dev Left GAITRite Test Single Support Time Standard Deviation Left
Single Supp Time Std Dev Right GAITRite Test Single Support Time Standard Deviation Right
Double Supp Time Std Dev Left GAITRite Test Double Support Time Standard Deviation Left
Double Supp Time Std Dev Right GAITRite Test Double Support Time Standard Deviation Right
Heel Off On Std Dev Left GAITRite Test Heel Off On Standard Deviation Left
Heel Off On Std Dev Right GAITRite Test Heel Off On Standard Deviation Right
Supp Base Std Dev Left GAITRite Test Support Base Standard Deviation Left
Supp Base Std Dev Right GAITRite Test Support Base Standard Deviation Right
Foot Length Std Dev Left GAITRite Test Foot Length Standard Deviation Left
Foot Length Std Dev Right GAITRite Test Foot Length Standard Deviation Right
Foot Width Std Dev Left GAITRite Test Foot Width Standard Deviation Left
Foot Width Std Dev Right GAITRite Test Foot Width Standard Deviation Right
Computer Time (MSec) GAITRite Test Computer Time (MSec)
Trigger 1 0 GAITRite Test Metronome first beep time
Trigger 1 1 GAITRite Test Not used
Trigger 2 0 GAITRite Test Switch first time pressed
Trigger 2 1 GAITRite Test Not used
Trigger 1 E GAITRite Test Metronome last beep time
Trigger 1 E 1 GAITRite Test Not used
Trigger 2 E GAITRite Test Switch last time pressed
Trigger 2 E 1 GAITRite Test Not used
Inclinometer1 GAITRite Test Tray inclinometer
Inclin_Avg_Tilt_LR GAITRite Test Tray Inclinometer Average Tilt Left to Right
Inclin_Avg_Tilt_FB GAITRite Test Tray Inclinometer Average Tilt Front to Back
Inclin_Min_Tilt_LR GAITRite Test Tray Inclinometer Minimum Tilt Left to Right
Inclin_Max_Tilt_LR GAITRite Test Tray Inclinometer Maximum Tilt Left to Right
Inclin_Max_Tilt_FB GAITRite Test Tray Inclinometer Maximum Tilt Front to Back
Inclin_Min_Tilt_FB GAITRite Test Tray Inclinometer Minimum Tilt Front to Back
Inclin_Veloc_Tilt_LR GAITRite Test Tray Inclinometer Velocity Tilt Left to Right
Inclin_Veloc_Tilt_FB GAITRite Test Tray Inclinometer Velocity Tilt Front to Back
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 49 of 50
Source Data
Field Name set Description
Inclin_SD_Tilt_LR GAITRite Test Tray Inclinometer Standard Deviation Tilt Left to Right
Inclin_SD_Tilt_FB GAITRite Test Tray Inclinometer Standard Deviation Tilt Front to Back
Inclinometer2 GAITRite Test Cup inclinometer
Inclin2_Avg_Tilt_LR GAITRite Test Cup Inclinometer Average Tilt Left to Right
Inclin2_Avg_Tilt_FB GAITRite Test Cup Inclinometer Average Tilt Front to Back
Inclin2_Min_Tilt_LR GAITRite Test Cup Inclinometer Minimum Tilt Left to Right
Inclin2_Max_Tilt_LR GAITRite Test Cup Inclinometer Maximum Tilt Left to Right
Record # Detail GAITRite Record Number
Test Record # Detail GAITRite Test Record Number
Pt Record # Detail GAITRite Patient Record Number
FootFall Object # Detail Foot Fall Object Number
Date / Time Stamp Detail Date / Time Stamp
Heel X Detail Sensor Location
Heel Y Detail Sensor Location
Toe X Detail Sensor Location
Toe Y Detail Sensor Location relative to pressure
Lowest X Detail Sensor Location relative to pressure
Highest X Detail Sensor Location relative to pressure
Lowest Y Detail Sensor Location relative to pressure
Highest Y Detail Sensor Location relative to pressure
First Contact Time Detail First Contact Time
Last Contact Time Detail Last Contact Time
Foot Flat Time Detail Foot Flat Time
Begin Time Detail Begin Time
End Time Detail End Time
Left/Right Foot Detail Left/Right Foot
Step Length Detail Step Length
Stride Length Detail Stride Length
Base of Support Detail Base of Support
Step Time Detail Step Time
Stride Time Detail Stride Time
Swing Time Detail Swing Time
Stance Time Detail Stance Time
Single Support Time (sec) Detail Single Support Time (sec)
Double Support Time (sec) Detail Double Support Time (sec)
Stride Velocity Detail Stride Velocity
Real Foot Flag Detail Real Foot Flag
Pass Number Detail Pass Number
Toe In / Out Detail Toe In / Out
Step Width Detail Step Width
Stride Width Detail Stride Width
Heel On Detail Heel On
(47DevM2) GAITRite Electronic Walkway Technical Reference (WI-02-15) Rev.L Page 50 of 50
Source Data
Field Name set Description
Heel Off Detail Heel Off
Mid On Detail Mid On
Mid Off Detail Mid Off
Toe On Detail Toe On
Toe Off Detail Toe Off
Heel off/on Detail Heel off/on
Double Support Loading Detail Double Support Loading
Double Support Unloading Detail Double Support Unloading