Manual - PHASE3 - Ingl
Manual - PHASE3 - Ingl
Manual - PHASE3 - Ingl
Operating Manual
Phase 3
Defibrillator and Transcutaneous Pacemaker Analyzer
Operating Manual
Copyright
Datrend System s Inc. (“DSI”) agrees to a lim ited copyright release that allows you to reproduce m anuals and other
printed m aterials for use in service training program s and other technical publications. If you would like other
reproductions or distributions, subm it a written request to Datrend System s Inc.
Claim s
Our routine m ethod of shipm ent is via com m on carrier. Upon delivery, if physical dam age is found, retain all packing
m aterials in their original condition and contact the carrier im m ediately to file a claim .
If the instrum ent is delivered in good physical condition but does not operate within specifications, or if there are any
other problem s not caused by shipping dam age, please contact your local sales representative or DSI im m ediately.
Return Procedure
Every product returned for refund/credit m ust be accom panied by a Return Material Authorization (RMA) num ber,
obtained from Datrend Custom er Service. All item s being returned m ust be sent prepaid (freight, duty, brokerage,
and taxes) to our factory location.
Restocking Charges
Products returned within 30 days of original purchase are subject to a m inim um restocking fee of 15%. Products
returned in excess of 30 days after purchase, but prior to 90 days, are subject to a m inim um restocking fee of 20%.
Additional charges for damage and/or m issing parts and accessories will be applied to all returns. Products which
are not in “as new” and resalable condition, are not eligible for credit return and will be returned to the custom er at
their expense.
Certification
This instrum ent was thoroughly tested and inspected and found to m eet DSI’s m anufacturing specifications when it
was shipped from the factory. Calibration m easurem ents are traceable to the National Research Council of Canada
(NRC) and/or the National Institute of Standards and Technology (NIST). Devices for which there are no NRC/NIST
calibration standards are m easured against in-house perform ance standards using accepted test procedures.
Page i
W arranty
No warranty shall apply when dam age is caused by any of the following:
! Discharge of a defibrillator into any term inal, interface connection or port, other than the specified Defibrillator
Input term inals of the instrum ent,
! Discharge of a transcutaneous pacem aker into any term inal, interface connection or port, other than the specified
Pacer Input term inals of the instrum ent,
! Use of an AC power supply adapter other than the AC adapter supplied with the instrum ent,
! Power failure, surges, or spikes,
! Dam age in transit or when m oving the instrum ent,
! Im proper power supply such as low voltage, incorrect voltage, defective wiring or inadequate fuses,
! Accident, alteration, abuse or m isuse of the instrum ent,
! Fire, water dam age, theft, war, riot, hostility, acts of God, such as hurricanes, floods, etc.
Only serialized products (those item s bearing a distinct serial num ber tag) and their accessory item s are covered
under this warranty. PHYSICAL DAMAGE CAUSED BY MISUSE OR PHYSICAL ABUSE IS NOT COVERED
UNDER THE W ARRANTY. Item s such as cables and non-serialized m odules are not covered under this warranty.
This warranty gives you specific legal rights and you m ay have other rights, which vary from province to province,
state to state, or country to country. This warranty is lim ited to repairing the instrum ent to DSI's specifications.
W hen you return an instrum ent to DSI for service, repair or calibration, we recom m end shipm ent using the original
shipping foam and container. If the original packing m aterials are not available, we recom m end the following guide
for repackaging:
! Use a double-walled carton of sufficient strength for the weight being shipped.
! Use heavy paper or cardboard to protect all instrum ent surfaces. Use nonabrasive m aterial around all projecting
parts.
! Use at least four inches of tightly packed, industrial-approved, shock-absorbent m aterial all around the
instrum ent.
DSI will not be responsible for lost shipm ents or instrum ents received in dam aged condition due to im proper
packaging or handling. All warranty claim shipm ents m ust be m ade on a prepaid basis (freight, duty, brokerage, and
taxes). No returns will be accepted without a Return Materials Authorization ("RMA) num ber. Please contact
Datrend (see Chapter 7) to obtain an RMA num ber and receive help with shipping/custom s docum entation.
Recalibration of instrum ents, which have a recom m ended annual calibration frequency, is not covered under the
warranty.
W arranty Disclaimer
Should you elect to have your instrum ent serviced and/or calibrated by som eone other than Datrend System s,
please be advised that the original warranty covering your product becom es void when the tam per-resistant Quality
Seal is rem oved or broken without proper factory authorization. W e strongly recom m end, therefore, that you send
your instrum ent to Datrend System s for service and calibration, especially during the original warranty period.
In all cases, breaking the tam per-resistant Quality Seal should be avoided at all cost, as this seal is the key to your
original instrum ent warranty. In the event that the seal m ust be broken to gain internal access to the instrum ent (e.g.,
in the case of a custom er-installed firm ware upgrade), you m ust first contact Datrend System s (see Chapter 7). You
will be required to provide us with the serial num ber for your instrum ent as well as a valid reason for breaking the
Quality Seal. You should break this seal only after you have received factory authorization. Do not break the Quality
Seal before you have contacted us! Following these steps will help ensure that you will retain the original warranty on
your instrum ent without interruption.
Page ii
W ARNING
Unauthorized user m odifications or application beyond the published specifications m ay result in electrical shock
hazards or im proper operation. Datrend System s will not be responsible for any injuries sustained due to
unauthorized equipm ent m odifications.
DSI DISCLAIM S ALL OTHER W ARRANTIES, EXPRESSED OR IM PLIED, INCLUDING ANY W ARRANTY OF
M ERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OR APPLICATION.
Page iii
EC Declaration of Conformity
Classification
While the Phase 3 may be used in a clinical environment, the Phase 3 is not to
be used to verify the performance of any device which is connected to a
patient.
The Phase 3 from Datrend Systems Inc. has been designed to conform to the
following standards:
EN55022:1994 Class B
EN61000-4-3 Level A
IEC 801-2 Level B
IEC 801-4 Level B
EN61010-1
IEC 61010-1
Page iv
Phase 3 OPERATING MANUAL
Table of Contents
1 SPECIFICATIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Tests Performed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Performance Specifications.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2.1 Energy Measurement, General.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2.2 Defibrillator High Range Energy Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.3 Defibrillator Low Range Energy Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.4 Defibrillator Charge Time Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2.5 Defibrillator Cardioversion Test.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2.6 AED Performance Test.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2.7 Pacemaker Pulse Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2.8 Pacemaker Noise Immunity Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2.9 Pacemaker Sensitivity Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2.10 Pacemaker Refractory Period Test.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 ECG Simulator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3.1 Performance Test Waveforms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3.2 Normal Sinus Rhythm.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3.3 Cardioversion, Shock Advisory and AED Test Waveforms. . . . . . . . . . . . . . . . . . . 5
1.3.4 Arrhythmia Simulations.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3.5 Pacemaker Test Waveforms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3.6 Performance Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.4 Non-Volatile Memory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.4.1 Memory Type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.4.2 Data Capacity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.4.3 Test Record Content. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.4.4 Waveform Record Content. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.5 Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5.1 User Interface.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5.2 Defibrillator Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5.3 Pacemaker Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5.4 ECG Simulator Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5.5 Defibrillator/Pacer Waveform Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5.6 High-Level ECG Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5.7 USB Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5.8 Serial (RS-232) Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5.9 Keyboard Port.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5.10 Power Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.6 Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.7 Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.8 Accessories.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.8.1 Standard Accessories.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.8.2 Optional Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2 OVERVIEW.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1 General Description.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1.1 Test Capabilities. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1.2 Data Log.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.1.3 Automated Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.1.4 Connectivity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.1.5 Viewing and Saving Defibrillator Waveforms. . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.1.6 Principle of Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.2 Controls and Interface Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3 SETUP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1 Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.2 Optional Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2.1 High-Level ECG Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2.2 Waveform Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2.3 Serial (RS-232) Input/Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.2.4 Keyboard Port.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.2.5 USB Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.3 Setting Up a HyperTerminal Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Hz Hertz
Definitions # Page xi
Phase 3 OPERATING MANUAL
Test Record Test result data which has been saved to the non-volatile
Data Log as a record of all tests performed on a particular
defibrillator or pacemaker, as identified by I.D. or control
number. A Test Record may be recalled from the Data Log
for viewing on the LCD or printing. Test Records can be
downloaded in bulk to a PC using Phase3pc software.
Transcutaneous Pacemaker A pacemaker which stimulates cardiac muscle via
electrodes applied to the outside of the body, usually the
chest. Many defibrillators incorporate a transcutaneous
pacemaker.
V1-V6 Precordial ECG leads
The following symbols appear on the labelling of Phase 3, and throughout this manual:
The operating manual provides valuable information on the proper use of Phase 3.
It is highly recommended that the operator read the instructions thoroughly before
operating the device. It is possible to damage the Device Under Test (DUT) and/or
cause harm to the operator if Phase 3 is used incorrectly
These warnings appear on the serial number label of Phase 3. Phase 3 measures hazardous
electrical voltages which MAY BE present when the covers are removed. The operator should read
the Operating Manual to determine what actions to take in the event of failure of Phase 3. Refer to
the Troubleshooting and Maintenance section of the Operating Manual.
In order to ensure ongoing performance accuracy, a TEST mode is provided on the Main Menu
(refer to section 4.3.1). It is recommended that this feature be used on a routine basis, or any time
a test result is in question.
CAUTION: This device measures high voltage signals. For protection of the
operator, inspect the Paddle Adapter and the interconnecting Paddle Adapter Cables
on a regular basis for nicks, cracks or other damage that may compromise the
insulating properties of the housing, wire or connectors. Replace any damaged or
worn components immediately.
Add Test Text which appears on the PC screen when running Phase3pc software, or
text which describes the action of an icon located on the PC screen.
1 SPECIFICATIONS
• Defibrillator Energy
• Defibrillator Charge Time
• Defibrillator Cardioversion
• Automated External Defibrillator (AED) Performance
• ECG Monitor Performance
• Pacer Pulse Characteristics
• Pacer Noise Immunity
• Pacer Sensitivity
• Pacer Refractory Periods
Specifications/Chapter 1 # Page 1
Phase 3 OPERATING MANUAL
Specifications/Chapter 1 # Page 2
Phase 3 OPERATING MANUAL
Specifications/Chapter 1 # Page 3
Phase 3 OPERATING MANUAL
Specifications/Chapter 1 # Page 4
Phase 3 OPERATING MANUAL
Specifications/Chapter 1 # Page 5
Phase 3 OPERATING MANUAL
Specifications/Chapter 1 # Page 6
Phase 3 OPERATING MANUAL
1.5 Interface
Specifications/Chapter 1 # Page 7
Phase 3 OPERATING MANUAL
1.6 Environment
• 15EC to 40EC
• 10% to 90% RH
• Altitude: 2000m max.
• Indoor Use Only
• Category II
• Pollution Degree 2
1.7 Dimensions
• 9.5" (24cm) W x 8" (20cm) H x 5.5" (14cm) D
• Weight 3 lbs. (1.4 kg)
1.8 Accessories
Specifications/Chapter 1 # Page 8
Phase 3 OPERATING MANUAL
CAN/CSA-C22.2 No. 61010-1-(2nd Edition) - Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use, Part 1: General Requirements
UL Std. No. 61010-1 (2nd Edition) - Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use - Part 1: General Requirements
IEC/EN 61010-1 (Ed.2.0) - Safety requirements for electrical equipment for measurement, control,
and laboratory use - Part 1: General requirement
Following manufacture, all units are subjected to, and have passed the requirements of the ‘Routine
Tests’ defined in Annex F of the above Standard(s).
Specifications/Chapter 1 # Page 9
Phase 3 OPERATING MANUAL
Specifications/Chapter 1 # Page 10
Phase 3 OPERATING MANUAL
2 OVERVIEW
Overview/Chapter 2 # Page 11
Phase 3 OPERATING MANUAL
Phase 3 incorporates one fixed 50 ohm load for defibrillator testing, and a variable load for pacer
testing that may be set from 50 to 1600 ohms in 50 ohm steps. To accommodate requirements of
the IEC60601-2-4 standard for defibrillators, the Variable Load Module (VLM) may be added to
Phase 3 as an optional accessory, allowing the user to perform energy tests with an alternative
defibrillator load ranging from 25 to 175 ohms in 25 ohm steps.
Phase 3 may be powered from its internal NiCad battery or from AC power using the accessory
adapter provided with the unit. The rechargeable battery can provide up to 24 hours of continuous
operation.
Phase 3 is small and light-weight, allowing you to easily carry the analyzer to institutional areas
where defibrillators are located. As you perform each test, results can be saved to the non-volatile
Data Log of Phase 3, for later review on the LCD, output to an attached printer, or download to a
personal computer.
Phase 3 incorporates a non-volatile memory or "data log" which is used to save test results or Test
Records. For each defibrillator tested, Phase 3 can create a Test Record containing energy, charge
time and cardioversion test results, which may be combined with results from ECG and
transcutaneous pacemaker tests. The non-volatile memory of Phase 3 provides storage for up to 80
such Test Records, each Record being identified by the control number of the defibrillator tested.
The non-volatile memory can also be used to store up to 10 defibrillator discharge waveforms as
Wave Records. Test Records and Wave Records can be downloaded to a personal computer by
means of Phase3pc software, included on CDROM with the analyzer.
Overview/Chapter 2 # Page 12
Phase 3 OPERATING MANUAL
For fully automated testing, Phase 3 provides 32 AutoSequences which may be user-programmed
by means of Phase3pc software. An AutoSequence test may include steps for measuring
defibrillator energy, charge time, and synchronization delay; ECG performance; and pacemaker
pulse characteristics, noise immunity, sensitivity, and refractory periods. Phase 3 comes equipped
with pre-programmed AutoSequences loaded into its non-volatile memory, which can be modified
or supplemented as desired via Phase3pc software.
For each test or measurement performed in an AutoSequence, the expected value of the result or
target is specified, so that Phase 3 will automatically pass or fail the result depending on an
acceptance range derived from the target and a selected tolerance. In manual tests, target and
tolerance parameters may be optionally entered "on the fly", while measurements are being taken.
2.1.4 Connectivity
Phase 3 has two communication ports: Serial I/O and USB. The Serial I/O interface is an RS-232
Port that may be connected to a personal computer, accessory barcode reader, serial printer, or
automated testing system such as the ES601Plus Automated Safety Analyzer. The USB Port is
intended solely for rapid data transfer to a personal computer using the accessory Phase3pc
software. To permit rapid data entry of equipment control numbers, Phase 3 also provides a PS/2
Port which may be connected to a standard PS/2 keyboard, or to a PS/2 barcode reader.
Show Waveform After discharging a defibrillator into Phase 3, view the captured
waveform on the Phase 3 graphical LCD. Three selections for both
vertical and horizontal axes are provided, allowing you examine the
waveform in greater detail.
Playback Play back the captured defibrillator waveform into the ECG monitor
or chart recorder of the defibrillator, with 200:1 timebase expansion.
Phase 3 may be configured to automatically playback the waveform
following detection of a valid defibrillator discharge.
Overview/Chapter 2 # Page 13
Phase 3 OPERATING MANUAL
Wave Records Save the captured waveform as a Wave Record, for later download
and graphical plotting by the accessory Phase3pc software
(included). To allow a greater number of Wave Records to be stored,
the signal data is compressed by Phase 3 before saving.
Wave Records are meant to provide a qualitative "snap shot" of the defibrillator signal, but for some
types of discharges, data resolution of the compressed waveform may be insufficient for detailed
numerical analysis such as might be required in a laboratory investigation or research. As an
alternative to Wave Records, Phase 3 in conjunction with Phase3pc software can provide real-time
waveform capture of a discharge. With real-time capture, Phase3pc can produce a graphic chart of
the waveform data on the PC screen immediately after a discharge, accompanied by the analysis
results such as energy, voltage, current and pulse duration. High-resolution waveform data can then
be saved as a file on the PC for later recall, or for export to other data analysis programs such as a
spreadsheet.
Measurements are accomplished by sampling the defibrillator signal from the defib pads or from the
pacer terminals at a high speed. Sampling is triggered by either the rising or falling edge of the input
signal, ensuring the waveform will be captured even if the operator connects the defib or pacer to
Phase 3 with reversed polarity.
Once Phase 3 is triggered, the signal is digitized with 13 bits of precision and approximately 58
milliseconds of signal data having a sample interval of 11.4 microseconds are then stored. After
acquiring the signal data, Phase 3 calculates test results based on the samples. As each measurement
result is derived (energy; voltage; current; and so on), the result is temporarily stored in the Clip
Record which is located in volatile memory (RAM). Once all tests have been completed, the user
can save the test results by entering an equipment control number for identification and then
transferring the contents of the Clip Record to a Test Record in the non-volatile Data Log.
Overview/Chapter 2 # Page 14
Phase 3 OPERATING MANUAL
Figure 1 illustrates the user controls and electrical interfaces of Phase 3. The electrical interfaces are
described in Chapter 3 of this manual. Operation of Phase 3 via the keypad and LCD is described in
Chapter 4.
1. Liquid Crystal Display (LCD) 8. ECG Outputs (RA; RL; LA; LL; V1)
2. Keypad 9. ECG Outputs (V2; V3; V4; V5; V6)
3. Defibrillator Input 10. Pacemaker Input
4. Power ON/OFF Switch 11. USB Port
5. DC Power Input 12. PS/2 Keyboard Port
6. High-Level ECG Output (1V QRS) 13. Serial I/O Port (RS-232)
7. Waveform Output (to oscilloscope) 14. Carrying Handle
Overview/Chapter 2 # Page 15
Phase 3 OPERATING MANUAL
Overview/Chapter 2 # Page 16
Phase 3 OPERATING MANUAL
3 SETUP
3.1 Connections
a. Place Phase 3 on a stable surface and insert the plug of the provided AC adapter into
the power input jack located on the right side panel (Figure 1). Plug the adapter into
an AC outlet.
IMPORTANT:
Setup/Chapter 3 # Page 17
Phase 3 OPERATING MANUAL
b. If you wish to test a defibrillator having external paddles, connect the Adult Paddle
Adapter included with Phase 3 to the Defibrillator Input using the Paddle Adapter
Cable (Figure 2). At each end of the cable, depress the catch on the plug, fit the plug all
the way into the mating receptacle, then release the catch to lock the cable securely in
place.
Alternatively, for hands-free testing, you may connect the output of the defibrillator
directly to the Defibrillator Input of Phase 3 by means of a suitable adapter cable. An
unterminated defibrillator adapter cable is available as an optional accessory (Datrend
P/N 7200-445).
Setup/Chapter 3 # Page 18
Phase 3 OPERATING MANUAL
NOTE:
An adapter set for testing internal paddles is also available as an optional accessory
(Datrend P/N 7400-443). Refer to the information sheet supplied with the Internal
Paddle Adapter Set for instructions on how to fit these adapters to the pads on the
Adult Paddle Adapter.
c. Connect ECG leads from the defibrillator under test (DUT) to the ECG simulator
outputs located on left and right side panels of Phase 3 (Figure 1). Connections for
standard 3- or 5-wire ECG patient cables (RA; RL; LA; LL; V) are provided on the
right side panel of Phase 3; optional chest connections (V2-V6) for diagnostic ECG are
located on the left side panel. Phase 3 accepts standard or safety-style banana plugs for
the ECG connections. An Adapter Set which can interface Phase 3 to snap-style ECG
patient leads is available as an optional accessory (Datrend P/N 7500-425).
Setup/Chapter 3 # Page 19
Phase 3 OPERATING MANUAL
IMPORTANT:
Setup/Chapter 3 # Page 20
Phase 3 OPERATING MANUAL
Refer to Figure 1 to locate the optional interface connections described in this section.
A high-level analog output signal from the Phase 3 ECG simulator is available at the 1V ECG jack
located on the right side panel. This is a 1/8" monaural phono jack having the signal on the tip
connection and ground on the sleeve. This output is capable of driving a load of 10k ohms or
higher.
The signal amplitude observed at the 1V ECG jack depends on the SCALE setting of the ECG
simulator. At this jack, the QRS complex of the ECG will have 1 volt peak amplitude when the
simulator is set to "x1 scale"; 2 volt amplitude on "x2 scale"; and 0.5 volt amplitude on "x0.5 scale".
The WAVEFORM output is located on the right side panel of Phase 3 and is intended for
connection of an oscilloscope or similar instrument. The WAVEFORM output is a 1/8" monaural
phono jack having the signal on the tip connection and ground on the sleeve.
When performing defibrillator tests, the WAVEFORM output will provide a scaled analog signal
representative of the defibrillator discharge. When LOW RANGE energy measurement is selected,
the WAVEFORM output provides 1 volt output per 200 volts at the Phase 3 Defibrillator Input
(200:1 attenuation). When HIGH RANGE energy measurement is selected, the WAVEFORM
output provides 1 volt output per 1000 volts at the Defibrillator Input (1000:1 attenuation).
When performing pacemaker tests, the WAVEFORM output provides a scaled analog signal
representative of the pacemaker pulse train. In this case, the WAVEFORM output provides 1 volt
output per 50mA at the Phase 3 PACER INPUT. The WAVEFORM scale remains constant
regardless of the pacemaker load selected for the test.
Setup/Chapter 3 # Page 21
Phase 3 OPERATING MANUAL
The Serial I/O Port of Phase 3 can be used to interface to any one of the following RS-232
peripheral devices.
The multifunction Serial Port is realized through a DB9 male connector whose pinout is shown in
Figure 4. The protocol of the Serial Port is 9600 baud, N, 8, 1.
Barcode readers which are compatible with the Phase 3 RS-232 Port are available as optional
accessories (RS-232 Pen Reader: Datrend P/N 7050-050; RS-232 CCD Scanner: Datrend P/N
7600-051).
Optional accessory cables are also available to connect other peripheral devices to Phase 3:
• Connection from Phase 3 to a DB9 PC COM port is achieved using a DB9F to DB9F NULL
MODEM cable (Datrend P/N 3140-400).
Setup/Chapter 3 # Page 22
Phase 3 OPERATING MANUAL
• Connection from Phase 3 to a DB25 COM port is achieved using a NULL MODEM cable,
noted above, in combination with a DB9M to DB25F adapter (Datrend P/Ns 3140-400 +
3140-401)
• Connection from Phase 3 to the accessory Citizen serial printer (Datrend P/N 7050-055) is
achieved using a NULL MODEM cable, noted above, in combination with a DB9M to
DB25M adapter (Datrend P/Ns 3140-400 + 3140-402).
Alternatively, cables made to the specifications listed in Table 3-1 may be used to connect Phase 3
to a peripheral device. Connectors specified in the table are for the cables, and shaded cells in the
table indicate no connection.
IMPORTANT:
Setup/Chapter 3 # Page 23
Phase 3 OPERATING MANUAL
The Keyboard Port of Phase 3 is intended for optional connection of a PS/2 personal computer
keyboard, or alternatively, a PS/2-compatible Barcode Scanner (Datrend P/N 7600-052; optional
accessory).
Connection of a PS/2 keyboard or barcode scanner to Phase 3 is optional but simplifies entry of an
ID, such as an equipment control number, for the defibrillator under test (DUT). Alternatively, the
keypad of Phase 3 may be used to enter a DUT ID if an external keyboard or barcode scanner is
not available.
IMPORTANT:
The USB Port of Phase 3 is intended to work specifically with Phase3pc software, supplied with
Phase 3 on CD ROM as a standard accessory (Datrend P/N 6950-004). A USB cable is also
provided with Phase 3 as a standard accessory (Datrend P/N 3140-440).
Phase 3 may be "hot-plugged" to the PC using the supplied USB cable at any time while the
instrument is running. Once Phase3pc successfully detects Phase 3, data may be transferred to or
from the PC via the high-speed USB connection.
• Create, modify, save, and print automated sequences of defibrillator and pacemaker tests
("AutoSequences"). AutoSequences created with Phase3pc can be uploaded into the non-
volatile memory of Phase 3. AutoSequences stored in Phase 3 can be downloaded to
Phase3pc for editing and re-uploading to Phase 3.
• Download Test Records stored in the non-volatile Data Log of Phase 3. Phase3pc generates
Test Report documents from the Test Records, which you can then print, save to disk, or
export to equipment management software.
Setup/Chapter 3 # Page 24
Phase 3 OPERATING MANUAL
• Download Wave Records stored in the non-volatile memory of Phase 3. Phase3pc generates
graphical plots of the waveform data which you can then print or save to disk.
Refer to Chapter 6 of this manual for more information on installing and using Phase3pc software.
Setup/Chapter 3 # Page 25
Phase 3 OPERATING MANUAL
As described in Section 3.2.3, the Serial I/O Port of Phase 3 may be connected to the serial
communications port (COM port) of a personal computer. A terminal program (i.e., Windows
98/2000/XP HyperTerminal or equivalent) running on the PC may then be used to receive Phase 3
Test Reports and save them to a text file. Follow the instructions below to set up HyperTerminal
for use with Phase 3.
a. Open HyperTerminal: In Windows XP, click Start, point to All Programs, point to
Accessories, point to Communications and click on HyperTerminal.
b. You will be prompted to enter a name for the HyperTerminal connection and select an
icon associated with this new connection (Figure 5). Enter a suitable name, select the
desired icon, and click on OK. Next time, click on this icon from Explorer or the Start
menu to avoid having to set up a new connection every time you wish to use
HyperTerminal.
c. A dialogue box will appear asking which port to connect to. Select the appropriate
COM port and click OK.
d. Another dialogue box appears to configure the port settings. Configure the port
settings as shown in Figure 6 and click OK.
Setup/Chapter 3 # Page 26
Phase 3 OPERATING MANUAL
In the HyperTerminal program window, click on Transfer in the menu bar and click on Capture
Text to save all Test Records to a file. In the File box, type a descriptive name for the file, and then
click Start.
e. HyperTerminal is now ready to receive data sent from Phase 3 to the PC. Refer to
Section 4.9.2.2 for instructions on how to initiate RS-232 transmission of Test Reports
from Phase 3.
f. Once all Test Reports have been transferred, stop capturing text. In the Transfer
menu, point to Capture Text and click on Stop.
Setup/Chapter 3 # Page 27
Phase 3 OPERATING MANUAL
Setup/Chapter 3 # Page 28
Phase 3 OPERATING MANUAL
The keypad and display of Phase 3 are shown in Figure 7. The keypad comprises four general-
purpose keys that have functions defined by the Phase 3 software, as represented by labels which
are shown on the display immediately above each key. In this manual, these keys are referred to as
"softkeys". Four additional keys labelled with arrows take specific functions depending on the state
of the display.
Operation/Chapter 4 # Page 29
Phase 3 OPERATING MANUAL
As defibrillator and pacemaker tests are performed with Phase 3, the individual test results are
collected in a temporary area of memory which is designated the Clip Record. When testing is
complete and results are to be saved, the contents of the Clip Record are then copied to a Test
Record, which resides in non-volatile memory. When a Test Record is recalled from non-volatile
memory for viewing or printing, the Record data is placed in the Clip Record.
The Clip Record is in volatile memory and any data stored in it will be lost if Phase 3 power is
interrupted before the Clip Record data is saved as a Test Record. Test Records saved in non-
volatile memory are retained when the power to Phase 3 is switched off.
As Phase 3 is used to perform individual measurements and tests, the Clip Record will progressively
accumulate the following data:
Data accumulated in the Clip Record may be reviewed at any point during the manual testing
process by pressing the VIEW button on the MANUAL TEST menu (see Section 4.4).
Operation/Chapter 4 # Page 30
Phase 3 OPERATING MANUAL
CANCEL is a special softkey that is used throughout the operator interface of Phase 3. CANCEL
allows the user to cancel a measurement in progress, or a series of previously-acquired
measurements, and return the operator interface to a higher level menu from which various types of
measurements are selected.
Pressing CANCEL can lead to erasure of selected test data in the Clip Record. The amount and
type of data erased will depend on which user interface menu is being displayed when CANCEL is
pressed. CANCEL has increasing impact on the Clip Record data as the level of the menu in the
user interface increases.
At the lowest level, if CANCEL is pressed immediately after a measurement is taken, then the
result of that single measurement is discarded without confirmation, and will not be copied to the
Clip Record. For example, if you measure a defibrillator energy, you may press CANCEL
immediately after the measurement is displayed to discard that single result. You may then repeat the
energy measurement. If a softkey other than CANCEL is pressed following an energy
measurement, then the energy result is copied to a designated location in the Clip Record.
If CANCEL is pressed before a measurement is taken, but after a number of measurements of the
same type have been performed, then all measurements of that type will be deleted from the Clip
Record, but only if you confirm the cancellation when prompted by Phase 3 (see Figure 8).
CANCEL TESTING
YES NO
Figure 8 - Confirmation prompt for CANCEL keypress.
For example, if you measure a series of defibrillator energies and then press CANCEL before
completing the next energy measurement, then all energy measurement data accumulated in the Clip
Record will be deleted if YES is pressed in response to the prompt of Figure 8.
Operation/Chapter 4 # Page 31
Phase 3 OPERATING MANUAL
Various types of defibrillator tests, including energy, charge time and cardioversion, are selected via
the DEFIBRILLATOR TEST menu which is described in Section 4.5 of this manual. If CANCEL
is pressed on the DEFIBRILLATOR TEST menu, then all energy, charge time and cardioversion
results being held in the Clip Record will be deleted, but only if you confirm the cancellation by
pressing YES when prompted by Phase 3 (Figure 8). However, only the defibrillator test data will
be affected; ECG and pacemaker results that may also be present in the Clip Record are not
disturbed if CANCEL on the DEFIBRILLATOR TEST menu is pressed and confirmed. The
CANCEL softkey has similar deletion scope on the ECG and PACER TEST menus, as described
later in this manual.
If CANCEL is pressed and confirmed at the highest level of menu in the user interface, for example
the MANUAL TEST menu (Section 4.4), then all test data currently stored in the Clip Record will
be deleted.
If CANCEL is pressed on a menu which is used for entering data or setting a test parameter, the
effect of CANCEL is to discard any changes that may have been made to the parameter being
edited. Examples include menus for the ECG simulator, data entry of target and tolerance, and ID
(control number) entry. In these cases, pressing CANCEL has no effect on the contents of the Clip
Record.
The deletion scope of the CANCEL softkey is explicitly specified in each Section of this manual
which discusses a user interface screen having a cancel function. When in doubt, refer to the
pertinent Section to determine exactly what potential effect CANCEL may have on the Clip
Record.
Certain parameters or settings in the Phase 3 are meant to be set to the operator's test preference.
These parameters, once set, will remain at that setting until specifically changed by the operator.
These parameters will retain their setting after the power is turned off, and will be in effect the next
time power is applied. The parameters that exhibit this behaviour are:
Operation/Chapter 4 # Page 32
Phase 3 OPERATING MANUAL
4.2 Powering Up
a. Set the power switch on the left side panel of Phase 3 to the ON position. The Phase 3
LCD will display a power-up screen while performing a number of internal self-tests
(Figure 9). The firmware version of the instrument will be shown on this screen while
the self-test is in progress.
b. If required, adjust the LCD contrast control on the right side panel of Phase 3 using a
trimpot adjustment tool (see Figure 1).
c. Upon completion of the self-test, the screen illustrated in Figure 10 will appear,
reporting the self-test results as follows:
SELF-TEST RESULTS
SELF-TEST: Pass
BATTERY: Charging
CLOCK: JUN 15 2006 13:04:25
TEST RECORDS: (0/80) used
WAVE RECORDS: (0/10) used
OK
Figure 10 - Self-test results.
Operation/Chapter 4 # Page 33
Phase 3 OPERATING MANUAL
d. The label displayed on the LCD above each front panel key corresponds to the
function of the key for the currently-viewed screen. In this manual, these keys are
referred to as "softkeys".
e. Upon successful conclusion of the self-test, press the OK softkey to proceed to the
main operating menu of Phase 3.
Operation/Chapter 4 # Page 34
Phase 3 OPERATING MANUAL
Referring to Figure 11, functions available on the MAIN MENU of Phase 3 are:
MAIN MENU
AUTO Press to view a list of automated test sequences ("AutoSequences") stored in the
non-volatile memory of Phase 3. See Section 4.10 for further information on
automated testing.
MANUAL Press to enter the MANUAL TEST menu, from which manually-conducted
defibrillator, ECG, or pacemaker tests may be selected. See Section 4.4 for
further information on manual testing.
DATA LOG Press to view a list of Test Records currently stored in the Data Log. See Section
4.9 for information related to Test Record viewing and printing.
REMOTE Press to activate Remote Mode. In this mode, Phase 3 may be controlled via
RS-232 from a PC, or from a safety analyzer such as ES601Plus, to perform
various types of measurements and acquire the data from such measurements.
This mode is also used to configure AutoSequences, via a USB connection, with
the assistance of Phase3pc software supplied with Phase 3. Refer to Chapter 5
for more information on Remote Mode, and Chapter 6 for information on
installation and use of Phase3pc.
< Press to access additional softkey functions as shown in Figure 12, and as
described below.
Operation/Chapter 4 # Page 35
Phase 3 OPERATING MANUAL
TEST Press to trigger a self-test of the data acquisition hardware. Refer to Section 4.3.1.
SET CLK Press to set local time and date. Refer to Section 4.3.3.
When selected from the MAIN MENU, the TEST PULSE function automatically triggers a series
of internal test pulses which are captured by the data acquisition system and are then analyzed,
thereby simulating the measurement of a defibrillator or pacemaker discharge. This automated self-
check procedure requires a few seconds to complete. While the procedure is in progress, the
CANCEL softkey may be pressed to abort the test.
Once the test pulses have been analyzed, the results are displayed and compared to 'expected values'
which are determined at the time of manufacture of your Phase 3 unit (Figure 13).
TESTING COMPLETED
OK
Figure 13 - Test Pulse results.
A result within the acceptance range is indicated as "PASS" on the display of Figure 13. Any result
outside the acceptance range is indicated as "FAIL". A "FAIL" result is suggestive of a
corresponding hardware defect inside the instrument. In the event of a "FAIL", refer to Chapter 7
of this manual.
It is recommended that this test be performed on a regular basis, or any time a test result is in
question. Press OK to continue with the use of the tester.
Operation/Chapter 4 # Page 36
Phase 3 OPERATING MANUAL
NOTE:
4.3.2 Phase 3 ID
Pressing ID on the MAIN MENU allows you to enter information, such as a control number, for
uniquely identifying Phase 3 within an equipment management system. After downloading Test
Record data from Phase 3 into Phase3pc software, this ID will then appear in Test Report
documents that are automatically generated by the software.
Phase 3 ID
Enter a unique ID / control number
using the ? > = <, or by means of an >
attached barcode reader or PS/2 kbd.
Referring to Figure 14, the softkeys operate as follows on the ID entry screen:
CLEAR Clear the displayed ID and show ------------------ . Changes to ENTER if any
characters are input.
Operation/Chapter 4 # Page 37
Phase 3 OPERATING MANUAL
The > ? keys to the right of the LCD display are used to increment/decrement the character
shown above the ^ cursor.
If a RS-232 or PS/2 barcode reader (wand or scanner) is plugged into the appropriate port of
Phase 3, then power will be applied to the reader when this menu is entered. A barcode may be
scanned which will then appear on this display. Press ENTER to store the displayed information as
the ID, See Sections 3.2.3 and 3.2.4 for barcode reader connection instructions.
In Manual Entry mode, the ID is input by placing the ^ character below a - using the ² ÷ keys and
scrolling through available characters using the v w keys. The ID displayed on the screen may be
cleared by pressing CLEAR, accepted by pressing ENTER, or CANCEL to discard the data and
return to the MAIN MENU.
If a barcode scanner is attached and is used to scan a barcode representing the ID, the screen of
4.1.2 will be displayed and the ID will be automatically accepted when the AUTO-ENTER timer
reaches zero.
Phase 3 ID
Enter a unique ID / control number
using the ? > = <, or by means of an >
attached barcode reader or PS/2 kbd.
PHASE 3 ID : XCV16652------- ?
AUTO-ENTER in 6 sec.
ENTER CANCEL
If the Phase 3 is attached to a PS/2 keyboard, the entire softkey row will be blanked after you start
typing on the keyboard. When data entry is complete, press ENTER or ESC on the keyboard
(replaces ENTER and CANCEL on the softkey row) to complete the action.
Operation/Chapter 4 # Page 38
Phase 3 OPERATING MANUAL
Pressing SET CLK on the MAIN MENU allows you to set the time and date of the real-time clock.
The clock is maintained by a lithium backup battery when the Phase 3 power is switched off.
>
JUN 15 2006 11:23:02
^^^
?
Referring to Figure 16, the softkeys operate as follows on the SET CLK screen:
ENTER Set the clock of Phase 3 to the time and date shown on the display.
CANCEL Do not change the current setting of the clock, discard any changes.
The > ? keys to the right of the LCD display are used to increment/decrement the item shown
above the ^ cursor.
Operation/Chapter 4 # Page 39
Phase 3 OPERATING MANUAL
Pressing the MANUAL softkey on the MAIN MENU causes Phase 3 to display the MANUAL
TEST menu (Figure 17). Functions available on the MANUAL TEST menu are:
MANUAL TESTS
DEFIB Press to enter a menu for manually testing defibrillator energy output, charge
time to maximum energy, and synchronization of the discharge to the ECG.
Refer to Section 4.5.
ECG Press to enter a menu for manually controlling the ECG simulator, to allow the
monitor of a defibrillator to be tested. Refer to Section 4.6.
PACER Press to enter a menu for manually testing a defibrillator's transcutaneous
pacemaker, including measurement of pacer pulse characteristics and refractory
intervals. Refer to Section 4.7.
AED Press to enter a menu for manually testing output energy and ECG pattern
analysis of an Automated External Defibrillator (AED). Refer to Section 4.8.
< Press to access additional softkey functions as shown in Figure 18, and as
described below.
VIEW Press to view measurement data currently stored in the Clip Record. Refer to
Section 4.9.
Operation/Chapter 4 # Page 40
Phase 3 OPERATING MANUAL
STORE Press to copy the current contents of the Clip Record to a Test Record in the
non-volatile memory. If an ID for the defibrillator under test has not been
entered when STORE is pressed, you will be prompted to enter an ID per
Section 4.4.1.
I.D. At any time during the manual test process, press I.D. to input identification
information, such as a control number, for the defibrillator or pacemaker under
test. Refer to Section 4.4.1.
CANCEL Cancel the manual test and return to the MAIN MENU. All measurement data
in the Clip Record will be deleted if you press YES in response to the
cancellation confirmation which will be displayed. For additional information on
CANCEL, see Section 4.1.3.
4.4.1 DUT ID
Pressing I.D. on the MANUAL TEST menu allows you to enter information, such as a control
number, for uniquely identifying the defibrillator under test (DUT). Test Records stored in the non-
volatile Data Log of Phase 3 are identified by DUT ID, as are the Test Report documents which are
generated by Phase3pc software when the Test Record data is downloaded from Phase 3 to the PC.
DEFIBRILLATOR ID
Enter a device ID / control number
using the ? > = <, or by means of an v
attached barcode reader or PS/2 kbd.
Operation/Chapter 4 # Page 41
Phase 3 OPERATING MANUAL
Referring to Figure 19, Figure 20, the softkeys operate as follows on the ID entry screen:
CLEAR Clear the displayed ID and show ------------------ . Changes to ENTER if any
characters are input.
ENTER Save the displayed ID in the Clip Record.
The > ? keys to the right of the LCD display are used to increment/decrement the character
shown above the ^ cursor.
In Manual Entry mode, the ID is input by placing the ^ character below a - using the ² ÷ keys and
scrolling through available characters using the v w keys. The ID displayed on the screen may be
cleared by pressing CLEAR, accepted by pressing ENTER, or CANCEL to discard the data and
return to the MAIN MENU.
If a barcode scanner is attached and is used to scan a barcode representing the ID, the screen of
Figure 20 will be displayed and the ID will be automatically accepted when the AUTO-ENTER
timer reaches zero.
DEFIBRILLATOR ID
Enter a device ID / control number
using the ? > = <, or by means of an v
attached barcode reader or PS/2 kbd.
If the Phase 3 is attached to a PS/2 keyboard, the entire softkey row will be blanked after you start
typing on the keyboard. When data entry is complete, press ENTER or ESC on the keyboard
(replaces ENTER and CANCEL on the softkey row) to complete the action.
Operation/Chapter 4 # Page 42
Phase 3 OPERATING MANUAL
If you select DEFIB, ECG, PACER or AED from the MANUAL TEST menu, and the Clip
Record already contains measurement data corresponding to that type of test, Phase 3 will prompt
you to confirm overwriting the existing test results with new results from the test you have selected
(Figure 21).
YES NO
Figure 21 - Confirmation for restarting a test.
Press YES to proceed with the selected manual test, or NO to retain the current contents of the
Clip Record and return to the MANUAL TEST menu.
Once you have started testing a defibrillator or pacemaker and have collected measurement data in
the Clip Record for these types of DUT, you cannot then change to testing an Automated External
Defibrillator (AED). This is because AED measurement data is incompatible with the measurement
data obtained from testing a conventional defibrillator/pacemaker. If AED is selected from the
MANUAL TEST menu, and the Clip Record already contains measurement data from defibrillator
and/or pacemaker tests, then the error message of Figure 22 is displayed.
OK
Figure 22 - Data Incompatibility Error.
The message of Figure 22 will also be displayed if you have collected AED measurement data in
the Clip Record, and then try to initiate defibrillator or pacemaker tests by pressing DEFIB or
PACER respectively.
Operation/Chapter 4 # Page 43
Phase 3 OPERATING MANUAL
If the display of Figure 22 appears, press OK to acknowledge the error and return to the
MANUAL TEST menu. Save the current contents of the Clip Record by pressing STORE. Then,
initiate a new test by pressing the desired softkey on the MANUAL TEST menu.
Press the DEFIB softkey on the MANUAL TEST menu to enter the DEFIBRILLATOR TEST
menu (Figure 23). Functions available on this menu are:
DEFIBRILLATOR TESTS
ENERGY Press to measure delivered energy, peak voltage and current, and discharge pulse
width. Refer to Section 4.5.1.
CHG TIME Press to measure charge time at maximum output energy. Refer to Section 4.5.2.
< Press to access additional softkey functions as shown in Figure 24, and as
described below.
< CANCEL
Figure 24 - Alternate softkeys of Defibrillator Test menu.
Operation/Chapter 4 # Page 44
Phase 3 OPERATING MANUAL
CANCEL Cancel the defibrillator test and return to the MAIN MENU. Defibrillator
measurement data in the Clip Record will be deleted if you press YES in
response to the cancellation confirmation which will be displayed. All other data
in the Clip Record will be retained. For additional information on CANCEL, see
Section 4.1.3.
Press the ENERGY softkey on the DEFIBRILLATOR TEST menu to enter the DEFIB
ENERGY test menu (Figure 25). Functions available on this menu are:
RANGE Press to toggle between LOW and HIGH range energy measurement. The range
currently in use is indicated on the display as either RNG: LO or RNG: HI
respectively. Select the range based on the energy setting of the defibrillator under
test (DUT), as follows:
Operation/Chapter 4 # Page 45
Phase 3 OPERATING MANUAL
ECG Press to enter a menu for setting up the ECG simulator. Note the waveform
setting and the amplitude scale of the simulator are shown on the display, to the
right of ECG Wave. According to Figure 25, the default setting of the simulator
for energy tests is coarse ventricular fibrillation (VFIB1) with amplitude set to x1
scale (1mV QRS on ECG Lead II). There are several options available for the
operation of the tester after an energy pulse has been measured. Refer to Section
4.5.1.2 for instructions describing operation of the ECG simulator and the settings
available.
TARGET Press to enter optional energy target and tolerance settings which will allow
Phase 3 to automatically pass or fail the energy measured when the defibrillator
is discharged. The current settings of the energy target (in joules) and the tolerance
about the target (in per-cent) are shown on the display as TARGET and TOL
respectively. According to Figure 25, by default the target and tolerance are
disabled until values are entered by the user, which then enables the automatic
pass/fail feature of the ENERGY TEST menu. Refer to Section 4.5.1.3 for
instructions on how to input a target and tolerance.
DONE Press to return to the DEFIBRILLATOR TEST menu. Energy measurement data
in the Clip Record are retained.
< Press to access additional softkey functions as shown in Figure 26, and as
described below.
< CANCEL
CANCEL Cancel the energy test and return to the DEFIBRILLATOR TEST menu. Energy
measurement data in the Clip Record will be deleted if you press YES in response
to the cancellation confirmation which will be displayed. All other data in the Clip
Record will be retained. For additional information on CANCEL, see Section
4.1.3.
Operation/Chapter 4 # Page 46
Phase 3 OPERATING MANUAL
Pressing the ECG softkey on the ENERGY TEST menu causes Phase 3 to display a menu for
setting up the ECG simulator (Figure 27, page 50). The ECG simulator provides a total of 55
Waveforms which are organised in seven Wave Banks. Depending on the test being performed,
certain banks may be disabled as the waveforms contained within those banks are not applicable to
the test.
The display of Figure 27 shows the WAVE BANK and the full title of the WAVEFORM currently in
effect, as well as an abbreviated waveform title which is shown on the display immediately below the
full title. The abbreviated waveform title (e.g. VFIB1) may be used in place of the full title on some
displays of the Phase 3 operator interface. Table 4-1 lists all of the available Banks and Waveforms,
and whether or not the waveform is available on the ECG SETUP menu for specified defibrillator,
pacemaker and AED tests.
Operation/Chapter 4 # Page 47
Phase 3 OPERATING MANUAL
Asystole ASYS1 ! ! ! !
Operation/Chapter 4 # Page 48
Phase 3 OPERATING MANUAL
Wave Bank Waveform Short Title Defib Defib Defib ECG AED Pacer
Energy Chg Sync Test Test Tests
Test Test Test
Sinewave, 10 Hz SIN10 !
Sinewave, 20 Hz SIN20 !
Sinewave, 40 Hz SIN40 !
Sinewave, 50 Hz SIN50 !
Sinewave, 60 Hz SIN60 !
Sinewave, 70 Hz SIN70 !
Bigeminy BIGEM !
Trigeminy TRGEM !
Operation/Chapter 4 # Page 49
Phase 3 OPERATING MANUAL
Referring to the ECG SETUP menu shown in Figure 27, functions available on the menu are:
OK Press to accept the ECG simulator settings as shown on the display and return
to the previous menu. Note: The ECG waveform does not change until OK is
pressed.
BANK Press to modify the Wave Bank of the ECG simulator. Refer to Table 4-1 to
determine which Wave Banks will be enabled for the type of test being
performed.
WAVEFORM Press to modify the Waveform for the selected Bank. Refer to Table 4-1 to
determine which Waveforms will be enabled for the type of test being
performed.
SCALE Press to set the simulator's amplitude scale to one of these choices:
< Press to access additional softkey functions as shown in Figure 28, and as
described below.
Operation/Chapter 4 # Page 50
Phase 3 OPERATING MANUAL
MODE Press to select the response of the ECG simulator to a discharge from the
defibrillator under test, as follows:
See Figure 29
Operation/Chapter 4 # Page 51
Phase 3 OPERATING MANUAL
PLAYBACK Press to toggle between the two available choices for simulator playback of an
acquired defibrillator waveform:
See Figure 29
CANCEL Press to discard any changes made to ECG simulator settings and return to
the previous menu.
Note: The MODE and PLAYBACK selections are User Preference Settings that will remain in
effect until changed by the operator. When operating the Phase 3 in manual test mode, the
changes made here will apply to the defibrillator test currently in progress as well as to future
tests, even if the Phase 3 power is switched off and then back on.
SIMULATOR MODE:
Revert to sinus rhythm at 60 BPM
after shock detected.
PLAYBACK MODE:
1/3 waveform (4 sec play time)
< OK MODE PLAYBACK CANCEL
Figure 29 - Simulator Mode settings.
Phase 3 allows you to optionally specify a target or expected value for a measurement to be made,
and a tolerance or acceptance range which spans either side of the target. If a target and tolerance
have been entered, Phase 3 will automatically compare the measurement result to the target when
the measurement is performed. If the difference between the measurement result and the target lies
within the tolerance range, Phase 3 will automatically assign a "PASS" to the result. If the difference
lies outside the tolerance, Phase 3 will automatically assign a "FAIL" to the result.
Operation/Chapter 4 # Page 52
Phase 3 OPERATING MANUAL
Entering tolerances and targets for manual measurements is always optional. If you do not wish to
enter a target and tolerance, you can simply "PASS" or "FAIL" any test result yourself, via
corresponding softkeys provided on the Phase 3 keypad, and thereby manually store the result with
your pass/fail assessment in the Clip Record.
Press the TARGET softkey on the ENERGY TEST menu to display a menu for inputting an
energy target and tolerance (Figure 30).
>? Increment/decrement the digit shown above the ^ cursor. Phase 3 will not allow
you to set the target value outside the permitted range which is shown closer to
the bottom of the display.
ENTER Save the displayed target value and tolerance.
±5%
±10%
±15%
±20%
OFF (disables automatic pass/fail assessment)
< Press to access additional softkey functions as shown in Figure 31, and described
as follows:
Operation/Chapter 4 # Page 53
Phase 3 OPERATING MANUAL
CANCEL Press to discard any changes made to target and tolerance settings and return to
the previous menu.
The purpose of the energy test is to verify that a defibrillator accurately delivers the amount of
energy which has been selected by the operator. In addition to delivered energy, the test reports
other parameters of the defibrillator discharge, such as voltage, current and pulse width, so that
conformance to manufacturer's specifications can be confirmed. Normally, a number of different
energy settings will be tested, for example: 10, 20, 50, 100, 200, 300 and 360 joules.
a. Connect the defibrillator under test (DUT) to Phase 3 according to Section 3.1.
Operation/Chapter 4 # Page 54
Phase 3 OPERATING MANUAL
f. Optionally, if you would like to change the waveform of the ECG simulator, press ECG
to view the ECG setup menu. Refer to Section 4.5.1.2.
g. Optionally, if you would like to have Phase 3 automatically pass or fail the measurement
you are about to take, press TARGET, enter a target energy, and select a tolerance.
Refer to Section 4.5.1.3.
h. If using external or internal paddles, apply the paddles firmly to the Phase 3 Paddle
Adapter.
j. Discharge the DUT into Phase 3. Peak current and voltage readings will appear on the
display immediately after the discharge. Phase 3 will calculate the energy and display
the result in Joules. The energy calculation will require approximately four seconds to
complete. Depending on the ECG MODE settings (see section 4.5.1.2), Phase 3 may
automatically play back the defibrillator pulse waveform to the ECG terminals and
paddle plates, then change the ECG output to normal sinus rhythm ("NSR60").
k. If the measurement is being performed on LOW range, and Phase 3 detects distortion
(clipping) in the acquired waveform data, Phase 3 will produce an alarm message on the
display (Figure 33). The measurement results are invalid if this alarm is displayed.
Acknowledge the alarm by pressing the OK softkey, then repeat the measurement after
selecting HIGH range on the menu of Figure 32.
OK
Figure 33 - Warning if signal distortion detected.
Operation/Chapter 4 # Page 55
Phase 3 OPERATING MANUAL
l. If the measurement is being performed on HIGH range, and Phase 3 detects distortion
(clipping) in the acquired waveform data, Phase 3 will produce an alarm message on the
display. The alarm title will be “HIGH RANGE ALARM”, and the warning message
is:
m. If the acquired signal data is valid, but a target has not been entered, Phase 3 will
display the results according to Figure 34.
PASS Save the displayed measurement data (energy, current, voltage, and
pulse widths) to the Clip Record with a "PASS" assessment. The test
counter shown at the top right corner of the display will increment,
until a maximum of 10 energy tests have been accumulated in the
Clip Record. If PASS is pressed after 10 energy results have been
stored, then each new result will overwrite the Clip Record location
reserved for energy test #10.
FAIL As described above for the PASS softkey, except that a "FAIL"
assessment will be saved in the Clip Record with the data, rather
than a "PASS".
SAVE Simply save the measurement results to the Clip Record, with
neither a "PASS" nor a "FAIL". The test counter displayed at top
right is incremented, as described above for PASS.
SHOW WAV Press to view the captured defibrillator waveform on the LCD. Refer
to Section 4.5.1.5.
Operation/Chapter 4 # Page 56
Phase 3 OPERATING MANUAL
n. If a target and tolerance have been entered per Section 4.5.1.3, Phase 3 will display the
results according to Figure 36.
Operation/Chapter 4 # Page 57
Phase 3 OPERATING MANUAL
o. While showing the results of an energy discharge, Phase 3 is ready to process another
discharge without any further operator intervention. The display screen ( Figure 34 or
Figure 36) will be indicating "WAITING FOR DISCHARGE". If the defibrillator is
discharged at this time, without pressing a PASS, FAIL, SAVE, or OK softkey,
Phase 3 will automatically SAVE the measurement results currently being displayed,
and will then proceed to process the new discharge. This allows a series of manual
energy tests to be performed without the need to press any keys on Phase 3 .
p. Repeat steps (d) to (o) as required until the desired number of DUT energy levels have
been tested, up to a maximum of 10 energies. More than 10 energy levels can be tested,
however, if such results are saved to the Clip Record via PASS, FAIL, SAVE, or OK
softkeys, then the Clip Record location reserved for energy test #10 will be overwritten
with each new result.
q. When all tests have been completed, press DONE (see Figure 32). Energy results
accumulated in the Clip Record are retained and may be supplemented with data from
other defibrillator, ECG and pacemaker tests.
Operation/Chapter 4 # Page 58
Phase 3 OPERATING MANUAL
Phase 3 provides unique features for viewing and saving the waveform of a defibrillator discharge.
These features are enabled after completion of each energy measurement, via the PLAYBACK,
SHOW WAV and SAVE WAV softkeys described in the preceding Section.
PLAYBACK After an energy measurement, press this softkey to play back the captured
defibrillator waveform into the ECG monitor or chart recorder of the DUT, via the
ECG simulator of Phase 3. The display will appear as shown in Figure 37 as the
playback progresses. To stop a playback in progress, press the CANCEL softkey at
any time.
DEFIB ENERGY TEST(01/10)
CANCEL
Figure 37 - Playback of captured waveform via ECG simulator.
If the energy measurement was performed on HIGH range, an ECG monitor set to
Lead II will show 1mV for each 1000V of defibrillator signal, or 1mV for each
2000V if the signal is being observed at the contact pads of the Phase 3 Paddle
Adapter. If the measurement was performed on LOW range, an ECG monitor on
Lead II will show 1mV per 200V of defib signal, or 1mV per 400V at the Paddle
Adapter. Each second of signal shown on the ECG monitor or chart recorder will
correspond to 5 milliseconds of the defibrillator waveform (200:1 timebase
expansion).
The length of the play back will be determined by the PLAYBACK MODE setting
(refer to section 4.1.5.2). Play back of the entire waveform requires approximately
12 seconds. The alternative PLAYBACK MODE will play back the first 1/3 of the
waveform data, which will require approximately 4 seconds. In most cases, this will
display the complete signal of interest.
Operation/Chapter 4 # Page 59
Phase 3 OPERATING MANUAL
SHOW WAV After an energy measurement, press this softkey to view the captured defibrillator
waveform on the Phase 3 display (Figure 38).
When viewing a defibrillator waveform with the SHOW WAV function, the
Phase 3 keypad operates as follows:
?> Contract/expand the vertical scale. The available choices depend on the
energy measurement range, as follows:
• 0 to 20 milliseconds
• 0 to 40 milliseconds
• 0 to 60 milliseconds (default)
Press any key other than ?>=< to return the display to the previous menu.
Operation/Chapter 4 # Page 60
Phase 3 OPERATING MANUAL
SAVE WAV After an energy measurement, press this softkey to save the defibrillator waveform
in the non-volatile memory as a Wave Record. If you have not yet entered an ID
for the defibrillator under test (DUT) when SAVE WAV is pressed, you will be
prompted to enter an ID (see Section 4.4.1). Once a DUT ID has been provided
to Phase 3, the screen of Figure 39 will be shown.
Operation/Chapter 4 # Page 61
Phase 3 OPERATING MANUAL
NOTE:
Once Wave Records have been saved to the non-volatile memory, the Records may later be
downloaded into Phase3pc software for graphical plotting and printing. Refer to Chapter 6 for
details.
Phase 3 with Phase3pc software provides a feature for capturing high-resolution waveform data on
a personal computer. The data can be acquired after each defibrillator energy measurement is
completed, while the display is in the state shown in Figure 34 or Figure 36.
Clicking the Waveform Capture icon in Phase3pc causes the detailed waveform data to be
rapidly transferred to the PC via the USB connection. After a waveform graph is plotted, Phase3pc
allows the user to print the graph or save the detailed data to a file for further numerical analysis.
Refer to Section 6.5.3 for additional details.
Defibrillator waveform capture may also be performed via RS-232 when Phase 3 is operating in
remote control mode. Refer to Chapter 5 for details.
Operation/Chapter 4 # Page 62
Phase 3 OPERATING MANUAL
Press the CHG TIME softkey on the DEFIBRILLATOR TEST menu to enter the CHARGE
TIME menu (Figure 40). Functions available on this menu are:
The intent of this test is to determine if the battery of the defibrillator is in reasonable condition and
will be able to charge the defibrillator, even when under some stress from clinical use. The
defibrillator should first be unplugged from the AC power. Prior to beginning the actual charge
time test, the defibrillator may now be discharged into Phase 3 several times. The number of pre-
test discharges will be tracked and reported in the upper right corner of the screen. These pre-test
discharges are counted, but they are not analyzed. When the requisite number of pre-test discharges
have been made, the actual charge time measurement can be initiated by pressing the START
softkey. The number of pre-test discharges will be reported in the final test report.
ECG Press to enter a menu for setting up the ECG simulator. Refer to Section 4.5.1.2 for
instructions describing operation of the ECG simulator.
MAX Press to input an upper limit for the charge time determined by this test. Choices
TIME available are: 5, 10, 15, 20, 25, 30, 35, 40 and 99.9 seconds. If an upper limit less
than 99.9 seconds is selected, the charge time obtained by the test will be compared
to the limit and automatically passed or failed by Phase 3 accordingly.
TARGET Press to enter optional energy target and tolerance settings which will allow Phase 3
to automatically pass or fail the energy measured when the defibrillator is
discharged. Refer to Section 4.5.1.3 for instructions on how to input a target and
tolerance.
Operation/Chapter 4 # Page 63
Phase 3 OPERATING MANUAL
< Press to access additional softkey functions as shown in Figure 41, and as described
below.
< CANCEL
Figure 41 - Alternate softkeys of Defibrillator Charge Time menu.
CANCEL Cancel the charge time test and return to the DEFIBRILLATOR TEST menu.
Charge time data in the Clip Record will be deleted if you press YES in response to
the cancellation confirmation which will be displayed. All other data in the Clip
Record will be retained. For additional information on CANCEL, see Section 4.1.3.
The purpose of a charge time test is to determine the length of time a defibrillator requires to charge
up to its maximum energy. To simulate a worst-case operating condition, the test is normally done
with the defibrillator powered by its internal battery.
Note that real-time waveform capture via USB is inhibited in this test mode. Use the energy test
described in Section 4.5.1 to capture defibrillator waveform data in real-time.
a. Connect the defibrillator under test (DUT) to Phase 3 according to Section 3.1.
b. On the DEFIBRILLATOR TEST menu, press CHG TIME. The display of Figure 40
will appear.
d. Disconnect the DUT from the AC line so that the DUT is operating from its internal
battery supply. This simulates a worst-case operating condition for the DUT.
e. On Phase 3, optionally press ECG to modify the settings of the ECG Simulator as
desired. Refer to Section 4.5.1.2.
f. Optionally, if you would like to have Phase 3 automatically pass or fail the measured
charge time, press MAX TIME to select the desired upper limit.
Operation/Chapter 4 # Page 64
Phase 3 OPERATING MANUAL
g. Optionally, if you would like to have Phase 3 automatically pass or fail the measured
energy, press TARGET, enter a target value for the maximum energy, and select a
tolerance. Refer to Section 4.5.1.3.
h. Optionally, discharge the DUT several times into Phase 3 before starting the actual
charge time test to exercise the battery. The number of pre-test discharges performed
will appear at the top right corner of the LCD display.
i. Press the START softkey. A timer will be displayed which will start counting down
from 5 seconds.
j. Apply the paddles firmly to the Phase 3 Paddle Adapter. You have 5 seconds to get
ready.
k. Charge the DUT when the countdown timer reaches zero and Phase 3 emits a beep
tone. Phase 3 will also display: CHARGE THEN DISCHARGE as an additional prompt
to you.
l. Discharge the DUT as soon as it reaches full charge. The measured charge time will be
displayed, after which Phase 3 will calculate and display the energy delivered.
m. If a MAX TIME has not been entered and an energy TARGET has not been entered,
Phase 3 will display the results according to Figure 42.
CHARGE TIME
TEST RESULTS:
n. Refer to Section 4.5.1.4 for an explanation of the softkey functions appearing on this
display.
Operation/Chapter 4 # Page 65
Phase 3 OPERATING MANUAL
o. If either a MAX TIME or an energy TARGET has been entered, Phase 3 will display
the results according to Figure 43. If both targets are specified, charge time and energy
results must both lie within their respective acceptance ranges in order to obtain a
"PASS", otherwise the overall result assigned by Phase 3 will be a "FAIL".
CHARGE TIME
TEST RESULTS:
< CANCEL
Figure 43 - Charge time results: Target and MAX TIME specified.
p. Refer to Section 4.5.1.4 for an explanation of the softkey functions appearing on the
display of Figure 43.
Operation/Chapter 4 # Page 66
Phase 3 OPERATING MANUAL
Press the SYNC softkey on the DEFIBRILLATOR TEST menu to enter the cardioversion or
"SYNC TEST" menu (Figure 44). Functions available on this menu are:
RANGE Press to select either LOW or HIGH energy measurement range as required. Refer
to Section 4.5.1.
ECG Press to enter a menu for setting up the ECG simulator. Refer to Section 4.5.1.2
for instructions describing operation of the ECG simulator. According to
Figure 44, note the default simulator settings for this test will cause the ECG to
automatically switch from coarse atrial fibrillation to normal sinus rhythm at 60
BPM when Phase 3 detects the defibrillator discharge.
SYNC TO Press to select the time reference with respect to which the defibrillator
synchronization delay will be measured. Options available are:
• Q wave of simulated ECG (default)
• R wave of simulated ECG
DONE Press to return to the DEFIBRILLATOR TEST menu. Cardioversion test results
in the Clip Record are retained.
< Press to access additional softkey functions as shown in Figure 45, and as
described below.
+-----TARGET-------+
< ENERGY DELAY CANCEL
Figure 45 - Alternate softkeys of Defibrillator Cardioversion Test menu.
Operation/Chapter 4 # Page 67
Phase 3 OPERATING MANUAL
TARGET Press to enter optional energy target and tolerance settings which will allow
ENERGY Phase 3 to automatically pass or fail the energy measured when the defibrillator is
discharged. Refer to Section 4.5.1.3 for instructions on how to input a target and
tolerance.
TARGET Press to enable or disable comparison of the measured sync delay to a predefined
DELAY acceptance range of +20 to +65 milliseconds. This acceptance range is derived
from specifications appearing in Section 104 of the IEC60601-2-4 standard for
defibrillators. If the delay target is enabled, Phase 3 will automatically pass or fail
the measured sync delay accordingly.
CANCEL Cancel the cardioversion test and return to the DEFIBRILLATOR TEST menu.
Cardioversion test data in the Clip Record will be deleted if you press YES in
response to the cancellation confirmation which will be displayed. All other data in
the Clip Record will be retained. For additional information on CANCEL, see
Section 4.1.3.
The purpose of a cardioversion test is to verify that a defibrillator can correctly synchronize its
discharge to the electrocardiogram. Phase 3 measures synchronization delay from a specified point
on the ECG QRS complex to the beginning of the defibrillator discharge. Phase 3 provides space in
the Clip Record for up to four cardioversion tests, allowing you to record the results for different
types of ECG waveform, or at various energy levels.
Note that real-time waveform capture via USB is inhibited in this test mode. Use the energy test
described in Section 4.5.1 to capture defibrillator waveform data in real-time.
a. Connect the defibrillator under test (DUT) to Phase 3 according to Section 3.1.
b. On the DEFIBRILLATOR TEST menu, press SYNC. The display of Figure 44 will
appear.
Operation/Chapter 4 # Page 68
Phase 3 OPERATING MANUAL
d. Activate "SYNC" operation or the equivalent functional mode on the DUT. The ECG
monitor of the DUT may now show indicator marks on the ECG trace, marking each
detected QRS in the ECG.
f. Optionally, if you would like to change the waveform or mode of the ECG simulator,
press ECG to view the ECG setup menu. Refer to Section 4.5.1.2.
g. Press SYNC TO if required to choose the timing reference on the simulated ECG for
measurement of the delay time.
h. Optionally, if you would like to have Phase 3 automatically pass or fail the measured
energy, press TARGET ENERGY, enter a target value for the energy, and select a
tolerance. Refer to Section 4.5.1.3.
i. Optionally, if you would like to have Phase 3 automatically pass or fail the measured sync
delay, press TARGET DELAY to enable the sync acceptance range.
j. If using external or internal paddles, apply the paddles firmly to the Phase 3 Paddle
Adapter.
l. Discharge the DUT into Phase 3. The timing of the discharge with respect to the
simulated QRS will be displayed immediately, and the ECG simulator will change
automatically to a normal sinus rhythm at 60 BPM. Phase 3 will then calculate the
energy and display the result.
m. If the measurement is being performed on LOW range, and Phase 3 detects distortion
(clipping) in the acquired waveform data, Phase 3 will produce an alarm message on the
display (Figure 46). The measurement results are invalid if this alarm is displayed.
Acknowledge the alarm by pressing the OK softkey, then repeat the measurement after
selecting HIGH range via the menu of Figure 44.
Operation/Chapter 4 # Page 69
Phase 3 OPERATING MANUAL
OK
Figure 46 - Warning if signal distortion detected.
n. If the delay target has not been enabled and an energy target has not been entered,
Phase 3 will display the results according to Figure 47. Refer to Section 4.5.1.4 for an
explanation of the softkey functions appearing on this display.
o. If either a delay target or an energy target has been entered, Phase 3 will display the
results according to Figure 48. If both targets are specified, sync delay and energy
results must both lie within their respective acceptance ranges in order to obtain a
"PASS", otherwise the overall result assigned by Phase 3 will be a "FAIL". Refer to
Section 4.5.1.4 for an explanation of the softkey functions appearing on the display of
Figure 48.
Operation/Chapter 4 # Page 70
Phase 3 OPERATING MANUAL
p. Similarly to the ENERGY test, while showing the results of an SYNC energy discharge,
Phase 3 is ready to process another discharge without any further operator intervention.
The display screen (Figure 47 or Figure 48) will be indicating "WAITING FOR
DISCHARGE". If the defibrillator is discharged at this time, without pressing a PASS,
FAIL, SAVE, or OK softkey, Phase 3 will automatically SAVE the measurement
results currently being displayed, and will then proceed to process the new discharge.
This allows a series of manual SYNC tests to be performed without the need to press
any keys on Phase 3 .
q. Repeat steps (c) to (o) as required until the desired number of cardioversion tests have
been performed, up to a maximum of 4. More than 4 tests can be done, however, if
such results are saved to the Clip Record via PASS, FAIL, SAVE, or OK softkeys,
then the Clip Record location reserved for cardioversion test #4 will be overwritten
with each new result.
r. When all tests have been completed, press DONE (see Figure 44). Cardioversion test
results accumulated in the Clip Record are retained and may be supplemented with data
from other defibrillator, ECG and pacemaker tests.
Operation/Chapter 4 # Page 71
Phase 3 OPERATING MANUAL
Press the ECG softkey on the MANUAL TEST menu to enter the ECG TEST menu (Figure 49).
The ECG simulator provides a total of 55 Waveforms which are organised in seven Wave Banks.
The display of Figure 49 shows the WAVE BANK and the full title of the WAVEFORM currently in
effect, as well as an abbreviated waveform title which is shown on the display immediately below the
full title. Refer to Table 4-1 of Section 4.5.1.2 for a listing of Banks and Waveforms provided by the
ECG simulator of Phase 3.
BANK Press to modify the Wave Bank of the ECG simulator. Refer to Table 4-1 of
Section 4.5.1.2.
WAVEFORM Press to modify the Waveform for the selected Bank. Refer to Table 4-1 of
Section 4.5.1.2.
SCALE Press to set the simulator's amplitude scale to one of these choices:
< Press to access additional softkey functions as shown in Figure 50, and as
described below.
Operation/Chapter 4 # Page 72
Phase 3 OPERATING MANUAL
PASS Save the ECG waveform to the Clip Record along with a "PASS" mark. The
test counter shown at the top right corner of the display will increment, until a
maximum of 12 ECG tests have been accumulated in the Clip Record. If
PASS is pressed after 12 ECG results have been stored, then the new result
will overwrite the Clip Record location reserved for ECG test #12.
FAIL As described above for PASS, except save a "FAIL" mark to the Clip Record
with the ECG waveform.
CANCEL Cancel the ECG test and return to the MANUAL TEST menu. ECG results
in the Clip Record will be deleted if you press YES in response to the
cancellation confirmation which will be displayed. All other data in the Clip
Record will be retained. For additional information on CANCEL, see Section
4.1.3.
Operation/Chapter 4 # Page 73
Phase 3 OPERATING MANUAL
The purpose of an ECG test is to verify that the ECG monitor and/or strip chart recorder of a
defibrillator is functioning according to the manufacturer's performance specifications. Tests
typically performed on a defibrillator ECG monitor could include those listed in Table 4-2:
Wideband frequency response Sine Wave, 0.1 - 100 Hz (Performance Wave Bank)
Rate accuracy and R-wave recognition All waveforms of Sinus Rhythm Bank, 30-300 BPM
a. Connect the defibrillator under test (DUT) to Phase 3 according to Section 3.1.
b. On the MANUAL TEST menu, press ECG. The display of Figure 49 will appear.
d. Observe the response of the monitor and/or chart recorder and press PASS or FAIL
as appropriate to record the waveform tested and the test result.
e. Repeat steps (c) and (d) as required until the desired number of ECG tests have been
performed, up to a maximum of 12. More than 12 tests can be done, however, if such
results are saved to the Clip Record via PASS or FAIL softkeys, then the Clip Record
location reserved for ECG test #12 will be overwritten with each new result.
f. When all tests have been completed, press DONE. ECG test results accumulated in the
Clip Record are retained and may be supplemented with data from manual defibrillator
and pacemaker tests.
Operation/Chapter 4 # Page 74
Phase 3 OPERATING MANUAL
Press the PACER softkey on the MANUAL TEST menu to enter the PACER TEST menu
(Figure 51). Functions available on this menu are:
PACER TESTS
LOAD Press to set the load resistor that will be used in all pacemaker tests. Refer to
Section 4.7.1.
PULSE Press to measure pacemaker pulse characteristics, such as amplitude, rate and
pulse width. Refer to Section 4.7.2.
REFRACT Press to measure the pacemaker's refractory periods. Refer to Section 4.7.3.
DONE Press to return to the MANUAL TEST menu. Pacemaker measurement data in
the Clip Record are retained.
< Press to access additional softkey functions as shown in Figure 52, and as
described below.
Operation/Chapter 4 # Page 75
Phase 3 OPERATING MANUAL
CANCEL Cancel the pacemaker test and return to the MAIN MENU. Pacemaker
measurement data in the Clip Record will be deleted if you press YES in
response to the cancellation confirmation which will be displayed. All other data
in the Clip Record will be retained. For additional information on CANCEL, see
Section 4.1.3.
Press the LOAD softkey on the PACER TEST menu to view or change the pacemaker test load
(Figure 53). Functions available on this menu are:
LOAD: 50 ohms
SELECT Press to set the load to the displayed resistance and return to the PACER TEST
menu.
LOAD+50 Press to increase the load by 50 ohms. Adjustment is limited to the range of 50 to
1600 ohms, as shown on the display.
LOAD-50 Press to decrease the load by 50 ohms.
CANCEL Press to return to the PACER TEST menu without changing the load.
Operation/Chapter 4 # Page 76
Phase 3 OPERATING MANUAL
< Press to access additional softkey functions as shown in Figure 54, and as
described below.
Note: The pacer load selection is a User Preference Setting that will remain in effect until changed
by the operator. When operating the Phase 3 in manual test mode, the changes made here
will apply to the pacer test currently in progress as well as to future tests, even if the
Phase 3 power is switched off and then back on.
NOTE:
All pacemaker tests must be performed with the same load
resistor. Always choose the load before beginning a series of
pacer tests.
If the Clip Record contains pacemaker test data when the LOAD
softkey on the PACER TEST menu is pressed, Phase 3 will warn
you that changing the load will cause pacemaker results in the
Clip Record to be erased (see Figure 55).
Operation/Chapter 4 # Page 77
Phase 3 OPERATING MANUAL
Press the PULSE softkey on the PACER TEST menu to enter a menu for determining the
characteristics of the pacemaker pulse (Figure 56). The selected test load will appear on the top line
of the display as shown in the figure. Functions available on the PACER PULSE menu are:
Operation/Chapter 4 # Page 78
Phase 3 OPERATING MANUAL
RATE Press to enter optional target and tolerance settings for the pulse rate, which
will allow Phase 3 to automatically pass or fail the measured rate when the
pacemaker pulse train is detected. Entering a pacer rate target is very similar to
entering an energy target during a defibrillator test. Refer to Section 4.5.1.3 for
instructions on how to input a target and tolerance.
DONE Press to return to the PACER TEST menu. Pulse test results in the Clip
Record are retained.
< Press to access additional softkey functions as shown in Figure 57, and as
described below.
WIDTH Press to enter optional target and tolerance settings for the pulse width, which
will allow Phase 3 to automatically pass or fail the measured width when the
pacemaker pulse train is detected. Entering a pacer width target is very similar
to entering an energy target during a defibrillator test. Refer to Section 4.5.1.3
for instructions on how to input a target and tolerance.
METHOD Press to select the method to be applied in determining the pulse amplitude.
The choices available are:
Note: The selection for amplitude METHOD is a User Preference Setting that will remain in
effect until changed by the operator. When operating the Phase 3 in manual test mode, the
changes made here will apply to the pacer test currently in progress as well as to future
tests, even if the Phase 3 power is switched off and then back on.
Operation/Chapter 4 # Page 79
Phase 3 OPERATING MANUAL
The purpose of a pacemaker pulse test is to verify that the pacemaker output signal is in
conformance with manufacturer's specifications for a given load resistance. Phase 3 provides space
in the Clip Record for up to 10 pulse tests, allowing you to record the results for different settings of
pacemaker output current and pulse rate.
a. Connect the output from the transcutaneous pacemaker of the defibrillator under test
(DUT) to Phase 3 according to Section 3.1.
b. Select the pacemaker load (Section 4.7.1) according to testing requirements specified by
the DUT manufacturer. Typically, the load corresponds to the largest resistance into
which the pacemaker can deliver its maximum rated output current. Refer to the
manufacturer's service manual for the DUT to determine the required load.
c. On the PACER TEST menu, press PULSE. The display of Figure 56 will appear.
e. Set the output current and pulse rate on the DUT to the desired values.
f. Optionally, if you would like Phase 3 to automatically pass or fail the measured pulse
amplitude, press AMPL and enter a target value and tolerance.
g. Optionally, if you would like Phase 3 to automatically pass or fail the measured rate,
press RATE and enter a target value and tolerance.
h. Optionally, if you would like Phase 3 to automatically pass or fail the measured pulse
width, press WIDTH and enter a target value and tolerance.
i. Select the amplitude measurement METHOD. The required method may be specified
by the manufacturer in the DUT service manual. If the manufacturer does not specify a
method, or if the DUT service manual is not available, choose "Average".
j. Press START to measure the pacemaker pulse. The START softkey will be changed
to a STOP softkey (Figure 58). The first measurement will be completed and the
results displayed after two successive pacemaker pulses are detected. Phase 3 will
continue to acquire and analyze pacer pulses until the STOP softkey is pressed. Each
time the results are displayed on the screen, Phase 3 will emit an audible 'click'.
Operation/Chapter 4 # Page 80
Phase 3 OPERATING MANUAL
NOTE: the audible 'click' is not an indication of the pacer rate, only an indication that new
data is on the display. Generally, the 'click' will occur every 2 to 3 pacer pulses, depending
on the rate. For a given rate, the 'click' will be periodic and a missing 'click' may indicate
Phase 3 encountered a pulse that was difficult to analyze. Frequent missing 'clicks' may be
an indication that the pacer pulses are of an abnormal shape or inconsistent rate.
STOP CANCEL
Figure 58 - Pacemaker Pulse Test in progress: no targets specified.
k. If no targets (amplitude, rate or width) have been entered when STOP is pressed,
Phase 3 will display the results for the last pacer pulse acquired according toFigure 59.
Refer to Section 4.5.1.4 for an explanation of the softkey functions appearing on this
display.
l. If any of the three available targets has been entered (amplitude, rate or width), Phase 3
will display the results according to Figure 60. If more than one target has been
specified, then all corresponding results must lie within their respective acceptance
ranges in order to obtain a "PASS", otherwise the overall result assigned by Phase 3 will
be a "FAIL". Refer to Section 4.5.1.4 for an explanation of the softkey functions
appearing on the display of Figure 60.
Operation/Chapter 4 # Page 81
Phase 3 OPERATING MANUAL
m. Repeat steps (e) to (l) as required until the desired number of pulse tests have been
performed, up to a maximum of 10. More than 10 tests can be done, however, if such
results are saved to the Clip Record via PASS, FAIL, SAVE, or OK softkeys, then the
Clip Record location reserved for pacemaker pulse test #10 will be overwritten with
each new result.
n. When all tests have been completed, press DONE (see Figure 56). Pacemaker pulse
test results accumulated in the Clip Record are retained and may be supplemented with
data from other pacemaker, ECG and defibrillator tests.
Phase 3 with Phase3pc software provides a feature for capturing high-resolution waveform data on
a personal computer. The data can be acquired after the STOP softkey has been pressed, while the
display is in the state shown in Figure 58 or Figure 59.
Clicking the Waveform Capture icon in Phase3pc causes the detailed waveform data to be
rapidly transferred to the PC via the USB connection. After a waveform graph is plotted, Phase3pc
allows the user to print the graph or save the detailed data to a file for further numerical analysis.
Refer to Section 6.5.3 for additional details.
Pacemaker waveform capture may also be performed via RS-232 when Phase 3 is operating in
remote control mode. Refer to Chapter 5 for details.
Operation/Chapter 4 # Page 82
Phase 3 OPERATING MANUAL
Press the REFRACT softkey on the PACER TEST menu to enter a menu for determining
refractory intervals of a pacemaker (Figure 61). The selected test load will appear on the top line of
the display as shown in the figure. Functions available on the PACER REFRACTORY menu are:
START Press to start the test. The pacemaker should be running in synchronous or demand
mode before this key is pressed.
ECG Press to enter a menu for setting up the ECG simulator. Refer to Section 4.5.1.2 for
instructions describing operation of the ECG simulator.
DONE Press to return to the PACER TEST menu. Refractory measurement data in the
Clip Record are retained.
< Press to access additional softkey functions as shown in Figure 62, and as described
below.
MIN Press to enter an optional minimum permitted value for the refractory periods,
which will allow Phase 3 to automatically pass or fail the measured periods at the
end of the test. Both the paced and sensed refractory periods will be compared to
this minimum limit if a limit is entered. Entering a minimum limit is very similar to
entering an energy target during a defibrillator test. Refer to Section 4.5.1.3 for
instructions on how to input a target. The MIN limit range is 50 - 750 msec.
Operation/Chapter 4 # Page 83
Phase 3 OPERATING MANUAL
MAX Press to enter an optional maximum permitted value for the refractory periods,
which will allow Phase 3 to automatically pass or fail the measured periods at the
end of the test. Both the paced and sensed refractory periods will be compared to
this maximum limit if a limit is entered. Entering a maximum limit is very similar to
entering an energy target during a defibrillator test. Refer to Section 4.5.1.3 for
instructions on how to input a target. The MAX limit range is 50 - 750 msec.
CANCEL Cancel the refractory period test and return to the PACER TEST menu. Refractory
period data in the Clip Record will be deleted if you press YES in response to the
cancellation confirmation which will be displayed. All other data in the Clip Record
will be retained. For additional information on CANCEL, see Section 4.1.3.
Note: The "ECG WAVE" and "Scale" for the refractory test, set by means of the ECG menu, are
User Preference Settings that will remain in effect until changed by the operator. When
operating the Phase 3 in manual test mode, the changes made here will apply to the
refractory test currently in progress as well as to future tests, even if the Phase 3 power is
switched off and then back on.
The purpose of a pacemaker refractory period test is to verify that paced and sensed refractory
periods are per manufacturer's specifications for the DUT. These intervals are defined as follows:
Operation/Chapter 4 # Page 84
Phase 3 OPERATING MANUAL
Phase 3 provides space in the Clip Record for up to 2 refractory period tests, allowing you to record
results for different rate settings on the pacemaker, or for different trigger waveforms on the
Phase 3 ECG simulator.
a. Connect the output from the transcutaneous pacemaker of the defibrillator under test
(DUT) to Phase 3 according to Section 3.1.
b. If no pacemaker tests have yet been performed, select the pacemaker load (Section
4.7.1) according to testing requirements specified by the DUT manufacturer. Do not
change the load if the Clip Record contains data from previous pacemaker tests.
c. On the PACER TEST menu, press REFRACT. The display of Figure 61 will appear.
e. Set the output current and pulse rate on the DUT. The output current should be set to
approximately mid-range. A pacer rate ranging from 60 to 120 PPM is recommended.
Rates in excess of 120 PPM should be avoided.
f. On Phase 3, optionally press ECG to modify the settings of the ECG simulator as
desired. Refer to Section 4.5.1.2.
g. Optionally, if you would like Phase 3 to automatically pass or fail the measured refractory
intervals, press MIN and enter a minimum allowed refractory period, then press MAX
and enter the maximum allowed period.
h. Press START to start the test. The test proceeds as a number of successive pacemaker
pulses are acquired. First, Phase 3 determines the pacemaker's pulse rate, which is
displayed on the LCD as PACER RATE. Then, Phase 3 will determine the PRP by
outputting ECG trigger pulses after each pacing pulse is detected. Following
determination and display of PRP, Phase 3 then proceeds to determine the SRP.
During the test, Phase 3 will display a progress indicator just above the softkey row. At
a pacer rate of 90 PPM, the entire test will require about 30 seconds to complete.
i. If neither a minimum limit nor a maximum limit has been entered, Phase 3 will display
the results according to Figure 63. Refer to Section 4.5.1.4 for an explanation of the
softkey functions appearing on this display.
Operation/Chapter 4 # Page 85
Phase 3 OPERATING MANUAL
j. If either limit has been entered, Phase 3 will display the results according to Figure 64.
If limits have been specified, then both refractory periods must lie within the limits in
order to obtain a "PASS", otherwise the overall result assigned by Phase 3 will be a
"FAIL". Refer to Section 4.5.1.4 for an explanation of the softkey functions appearing
on the display of Figure 64.
k. Repeat steps (e) to (j) if desired. More than two tests can be done, however, if such
results are saved to the Clip Record via PASS, FAIL, SAVE, or OK softkeys, then the
Clip Record location reserved for pacemaker refractory test #2 will be overwritten with
each new result.
l. When all tests have been completed, press DONE (see Figure 61). Refractory period
results accumulated in the Clip Record are retained and may be supplemented with data
from other pacemaker, ECG and defibrillator tests.
Operation/Chapter 4 # Page 86
Phase 3 OPERATING MANUAL
Press the IMMUNITY softkey on the PACER TEST menu to enter a menu for determining
immunity of a pacemaker to line-frequency noise (Figure 65). The selected test load will appear on
the top line of the display as shown in the figure. Functions available on the PACER NOISE
IMMUNITY menu are:
START Press to start the test. The pacemaker should be running in synchronous or
demand mode before this key is pressed.
NOISE Press to toggle the noise frequency between 50 Hz and 60 Hz.
CANCEL Cancel the noise test and return to the PACER TEST menu. Noise Test data in
the Clip Record will be deleted if you press YES in response to the cancellation
confirmation which will be displayed. All other data in the Clip Record will be
retained. For additional information on CANCEL, see Section 4.1.3.
Note: The NOISE frequency is a User Preference Setting that will remain in effect until changed
by the operator. When operating the Phase 3 in manual test mode, the changes made here
will apply to the immunity test currently in progress as well as to future tests, even if the
Phase 3 power is switched off and then back on.
Operation/Chapter 4 # Page 87
Phase 3 OPERATING MANUAL
The purpose of the noise immunity test is to verify that a pacemaker operating in demand mode
does not respond to line frequency noise at its ECG input.
In this test, Phase 3 outputs a line frequency sine wave via the ECG simulator and observes the
effect on the pacemaker. The test automatically passes if the pacemaker continues to pace at its set
rate when the output of the ECG simulator is set to maximum (6 millivolts peak-to-peak). The test
automatically fails if the pacemaker ceases pacing for a sine wave amplitude less than 6 millivolts.
a. Connect the output from the transcutaneous pacemaker of the defibrillator under test
(DUT) to Phase 3 according to Section 3.1.
b. If no pacemaker tests have yet been performed, select the pacemaker load (Section
4.7.1) according to testing requirements specified by the DUT manufacturer. Do not
change the load if the Clip Record contains data from previous pacemaker tests.
c. On the PACER TEST menu, press IMMUNITY. The display of Figure 65 will appear.
e. Set the output current and pulse rate on the DUT. The output current should be set to
approximately mid-range. A pacer rate ranging from 60 to 120 PPM is recommended.
Rates in excess of 120 PPM should be avoided.
g. Press START to start the test. The test proceeds as a number of successive pacemaker
pulses are acquired. First, Phase 3 determines the pacemaker's pulse rate. After
measuring the rate, Phase 3 will then begin to output a sine wave to the pacemaker via
the ECG simulator.
h. As each pacer pulse is detected, Phase 3 introduces an audible 'click' and increases the
noise signal amplitude by 0.0845 mV. The test will pass if the pacemaker does not react
to a 6 millivolt sine wave (Figure 66). If there is no activity from the pacer for 4
seconds for any noise level less than 6 mV, the test will fail. At 90PPM, the test will
require about 50 seconds to complete if the pacemaker passes, or less if it fails.
i. Press OK to save the immunity result in the Clip Record, or CANCEL to discard the
result.
Operation/Chapter 4 # Page 88
Phase 3 OPERATING MANUAL
Press the SENS softkey on the PACER TEST menu to enter a menu for determining sensitivity of
a pacemaker to the ECG amplitude (Figure 67). The selected test load will appear on the top line of
the display as shown in the figure. Functions available on the PACER SENSITIVITY menu are:
START Press to start the test. The pacemaker should be running in synchronous or
demand mode before this key is pressed.
LIMIT Press to enter an optional maximum permitted value for the sensitivity, which will
allow Phase 3 to automatically pass the AMPL result at the end of the test,
provided the AMPL result is less than the limit. Entering this limit is very similar to
entering an energy target during a defibrillator test. Refer to Section 4.5.1.3 for
instructions on how to input a target.
DONE Press to return to the PACER TEST menu. Sensitivity measurement data in the
Clip Record are retained.
Operation/Chapter 4 # Page 89
Phase 3 OPERATING MANUAL
< Press to access additional softkey functions as shown in Figure 68, and as
described below.
< CANCEL
Figure 68 - Alternate softkeys of Pacer Sensitivity Test menu.
CANCEL Cancel the sensitivity test and return to the PACER TEST menu. Sensitivity Test
data in the Clip Record will be deleted if you press YES in response to the
cancellation confirmation which will be displayed. All other data in the Clip
Record will be retained. For additional information on CANCEL, see
Section 4.1.3.
The purpose of the sensitivity test is to determine the smallest ECG signal that will reliably trigger a
pacemaker, and to verify this signal level is in accordance with manufacturer's specifications.
Phase 3 provides space in the Clip Record for up to 2 sensitivity tests, allowing you to record results
for different trigger waveforms on the Phase 3 ECG simulator.
a. Connect the output from the transcutaneous pacemaker of the defibrillator under test
(DUT) to Phase 3 according to Section 3.1.
b. If no pacemaker tests have yet been performed, select the pacemaker load
(Section 4.7.1) according to testing requirements specified by the DUT manufacturer.
Do not change the load if the Clip Record contains data from previous pacemaker tests.
c. On the PACER TEST menu, press SENS. The display of Figure 67 will appear.
e. Set the output current and pulse rate on the DUT. The output current should be set to
approximately mid-range. A pacer rate ranging from 60 to 120 PPM is recommended.
Rates in excess of 120 PPM should be avoided.
Operation/Chapter 4 # Page 90
Phase 3 OPERATING MANUAL
f. On Phase 3, optionally press ECG to modify the settings of the ECG simulator as
desired. Refer to Section 4.5.1.2.
g. Optionally, if you would like Phase 3 to automatically pass or fail the measured
sensitivity, press LIMIT and enter a maximum permitted value for the test result.
h. Press START to start the test. The test proceeds as a number of successive pacemaker
pulses are acquired. First, Phase 3 determines the pacemaker's pulse rate, which is
displayed on the LCD as PACER RATE. Then, Phase 3 will determine the sensitivity by
outputting ECG trigger pulses of variable amplitude after each pacing pulse is detected.
During the test, Phase 3 will display a progress indicator (a rotating line) just above the
softkey row.
i. As each pacer pulse is detected, Phase 3 will emit an audible 'click'. For each detected
pacer pulse, Phase 3 will increase the ECG signal level in steps of 0.045mV until the
pacer is inhibited. After the pacer is inhibited, the ECG level is reduced and then
increased in steps of 0.015mV until the pacer is again inhibited. The test will stop when
the pacer is inhibited by the ECG signal, or a maximum ECG level of 3 mV is reached.
For a pacer having a sensitivity of nominally 1 mV, and at a rate of 90PPM, the
sensitivity test will require about 20 seconds.
j. If a sensitivity limit has not been entered, Phase 3 will display the results according to
Figure 69. Refer to Section 4.5.1.4 for an explanation of the softkey functions
appearing on this display.
k. If a sensitivity limit has been entered, Phase 3 will display the results according to
Figure 70. Refer to Section 4.5.1.4 for an explanation of the softkey functions
appearing on this display.
Operation/Chapter 4 # Page 91
Phase 3 OPERATING MANUAL
l. Repeat steps (e) to (j) if desired. More than two tests can be done, however, if such
results are saved to the Clip Record via PASS, FAIL, SAVE, or OK softkeys, then the
Clip Record location reserved for sensitivity test #2 will be overwritten with each new
result.
m. When all tests have been completed, press DONE (see Figure 67). Sensitivity results
accumulated in the Clip Record are retained and may be supplemented with data from
other pacemaker, ECG and defibrillator tests.
Operation/Chapter 4 # Page 92
Phase 3 OPERATING MANUAL
Press the AED softkey on the MANUAL TEST menu to enter a menu for testing Automated
External Defibrillators or "AEDs" (Figure 71). Functions available on this menu are:
NO SHOCK Press this softkey if the AED audibly and/or visually advises the user that the
AED cannot or should not be discharged.
ECG Press to enter a menu for setting up the ECG simulator. Refer to Section
4.5.1.2 for instructions describing operation of the ECG simulator.
AUTO-NSR Press this softkey to enable or disable automatic response of the ECG simulator
to a defibrillator discharge. If AUTO-NSR is enabled, "> NSR60" will appear
on the LCD to the right of the ECG WAVE, in which case a defibrillator
discharge will cause the simulator to automatically change to a normal sinus
rhythm at 60BPM. Once triggered by a discharge, the NSR60 simulation will
remain in effect for subsequent tests, until changed via the ECG softkey
described above.
DONE Press to return to the MANUAL TEST menu. AED measurement data in the
Clip Record are retained.
< Press to access additional softkey functions as shown in Figure 72, and as
described below.
< CANCEL
Figure 72 - Alternate softkeys of AED Test menu.
Operation/Chapter 4 # Page 93
Phase 3 OPERATING MANUAL
CANCEL Cancel the AED test and return to the MANUAL TEST menu. AED Test data
in the Clip Record will be deleted if you press YES in response to the
cancellation confirmation which will be displayed. All other data in the Clip
Record will be retained. For additional information on CANCEL, see Section
4.1.3.
The primary purpose of the AED test is to determine whether an AED can analyze and correctly
classify an input ECG signal as shockable or not shockable. A second purpose of the test is to
record the energy of the discharge when the AED advises the user to administer a shock, and a
shock is then triggered. Phase 3 provides space in the Clip Record for up to 32 AED test
measurements, allowing you to record results for different ECG waveforms, AED shock series, and
AED energy levels.
In an AED test, the Phase 3 ECG simulator is limited to the waveforms listed in Table 4-1 of
Section 4.5.1.2. These ECG waveforms are classified according to AAMI guidelines1 as follows:
Operation/Chapter 4 # Page 94
Phase 3 OPERATING MANUAL
According the preceding table, an AED which operates according to AAMI guidelines1 for
sensitivity and specificity should advise a shock only if the ECG clearly indicates ventricular
fibrillation (> 100 ìV), or a wide complex ventricular tachycardia in excess of 150 BPM.
b. On the MANUAL TEST menu, press AED. The display of Figure 71 will appear.
c. Optionally, if you would like to change the ECG waveform, press ECG to modify the
settings of the ECG simulator. Refer to Section 4.5.1.2.
d. If necessary, press the "ANALYZE" or equivalent button on the AED, or set the AED
to automatically analyze and report on the ECG waveform.
e. If the AED analyzes the waveform and advises you not to shock, press the NO
SHOCK softkey on Phase 3. If this is the correct AED response given the type of
ECG waveform being simulated, Phase 3 will automatically assign a "PASS" to the test
result (Figure 73). The result of the test will be displayed for 3 seconds, a zero energy
result will be saved in the Clip Rec ord, and then the Phase 3 will go back to the screen
of Figure 71 . If the AED' s response is not correct for the selected ECG waveform,
Phase 3 will automatically assign a "FAIL" to the result.
ENERGY: ---.-J
ECG WAVE: AFIB1 (x1 Scale)
ECG TYPE: UNSHOCKABLE
***************** PASS *****************
f. If the AED advises you 'to shock', discharge the AED into Phase 3. If this is the
correct response for the type of ECG waveform being simulated, Phase 3 will assign a
"PASS" to the test result (Figure 74). If AUTO-NSR is enabled, the ECG simulator
will output a normal sinus rhythm at 60BPM. The results of the test will be displayed
for 5 seconds, the energy result will be saved in the Clip Record, and then the system
will go back to the screen of Figure 71. If the AED' s response is not correct for the
selected ECG waveform, Phase 3 will assign a "FAIL" to the result.
Operation/Chapter 4 # Page 95
Phase 3 OPERATING MANUAL
ENERGY: 150.8J
ECG WAVE: VFIB1 (x1 Scale)
ECG TYPE: SHOCKABLE
***************** PASS *****************
g. If the AED advises a shock, but this conflicts with the ECG TYPE displayed on
Phase 3, discharge the AED into Phase 3 regardless. Phase 3 will automatically assign a
"FAIL" to this type of test result (Figure 75).
ENERGY: 149.6J
ECG WAVE: VTACH1 (x2 Scale)
ECG TYPE: UNSHOCKABLE
***************** FAIL *****************
h. Repeat steps (c) to (g) as required until the desired number of AED tests have been
performed, up to a maximum of 32.
i. When all tests have been completed, press DONE (see Figure 71). AED test results
accumulated in the Clip Record are retained and may be supplemented with data from
additional ECG tests. However, additional defibrillator and pacemaker tests cannot be
done since the Clip Record now contains AED results (see Section 4.4.2). The Clip
Record contents must be saved to the non-volatile memory before new defibrillator
and/or pacemaker tests can be initiated.
Operation/Chapter 4 # Page 96
Phase 3 OPERATING MANUAL
As defibrillator, ECG and pacemaker tests are executed from the MANUAL TEST menu (see
Section 4.4), measurement data produced by these tests are accumulated in the Clip Record. The
alternate softkey set of the MANUAL TEST menu (Figure 76) provides means for viewing the
current contents of the Clip Record, or saving the Clip Record to a Test Record in the non-volatile
Data Log.
MANUAL TESTS
VIEW Press to view measurement data currently stored in the Clip Record. Phase 3 will
format the measurement data into a viewable TEST REPORT document, and will
place the TEST REPORT on the display. The display will then appear similar to the
example shown in Figure 77.
TEST REPORT
DATE: June 15 2006 TIME: 13:45:13
DEFIB ID: CN1234
Operation/Chapter 4 # Page 97
Phase 3 OPERATING MANUAL
Press the ? > keys to scroll down/up through the TEST REPORT which is
shown on the display. Press any other key to exit the Clip Record viewer and return
to the MANUAL TEST menu.
STORE Press to copy the current contents of the Clip Record to a Test Record in the non-
volatile memory. If an ID for the defibrillator under test has not been entered when
STORE is pressed, you will be prompted to enter an ID per Section 4.4.1.
I.D. At any time during the manual test process, press I.D. to input identification
information, such as a control number, for the defibrillator or pacemaker under test.
Refer to Section 4.4.1.
CANCEL Cancel the manual test and return to the MAIN MENU. All measurement data in
the Clip Record will be deleted if you press YES in response to the cancellation
confirmation which will be displayed. For additional information on CANCEL, see
Section 4.1.3.
The non-volatile Data Log is accessed from the MAIN MENU of Phase 3 (see Section 4.3). The
display will appear as shown in Figure 78 when DATA LOG on the MAIN MENU is pressed. As
shown in the figure, the number of Test Records currently stored in the Data Log is indicated at the
top right corner. The VIEW and PRINT softkeys will not be visible unless a Clip Record has been
loaded into memory.
Operation/Chapter 4 # Page 98
Phase 3 OPERATING MANUAL
LOAD Press to select a Test Record from the Data Log, and then load the Test Record
into the Clip Record for viewing or printing. Refer to Section 4.9.2.1.
VIEW Press to view the current contents of the Clip Record. This softkey will not be
visible until a Test Record has been loaded from the Data Log. Phase 3 will
format the measurement data into a TEST REPORT document, and will place the
TEST REPORT on the display. The display will then appear similar to the
example shown in Figure 77. Press the ? > keys to scroll down/up through the
TEST REPORT which is shown on the display. Press any other key to exit the
Clip Record viewer and return to the MANUAL TEST menu.
PRINT Press to print a TEST REPORT document via the Serial I/O Port of Phase 3.
Refer to Section 4.9.2.2. This softkey will not be visible until a Test Record has
been loaded from the Data Log.
DELETE Press to delete the last Test Record that was saved to the Data Log. A prompt
requiring you to confirm the deletion will be displayed if you press the DELETE
key.
< Press to access additional softkey functions as shown in Figure 79, and as
described below:
DEL WAV Press to delete all Waveform Records in the Data Log. A prompt requiring you to
confirm the deletion will be displayed if you press the DEL WAV key.
DEL ALL Press to delete all Test Records and Wave Records currently stored in the Data
Log. A prompt requiring you to confirm the deletion will be displayed if you press
the DEL ALL key.
CANCEL Press to return to the MAIN MENU.
Operation/Chapter 4 # Page 99
Phase 3 OPERATING MANUAL
Press the LOAD softkey on the DATA LOG menu to enter a menu for selecting a Test Record for
viewing or printing (Figure 80). Press the ? > keys to scroll down/up through the list of Test
Records one record at a time. The Test Records are displayed on the screen with the most recent
test first. Other softkey functions provided on this menu are:
PAGE DN Press to move the displayed list down by four Test Records.
SELECT Press to select the Test Record highlighted at the top of the list. The Test Record
will be loaded into the Clip Record, and can then be viewed or printed
respectively via VIEW or PRINT softkeys on the DATA LOG menu.
CANCEL Do not select a Test Record; return to the DATA LOG menu.
Phase 3 can transmit a TEST REPORT document from its DB9 Serial I/O Port to receiving
devices having a RS-232 serial port, such as a serial printer or a personal computer. Before
attempting to print, connect the receiving device to the Serial I/O Port of Phase 3 according to the
instructions in Section 3.2.3. Note the serial protocol of Phase 3 is 9600 baud, No parity, 8 data bits,
1 stop bit (9600, N, 8, 1).
After loading a Test Record from the Data Log, press the PRINT softkey on the DATA LOG
menu to print the TEST REPORT document. The TEST REPORT is 40 characters wide and will
comprise a variable number of lines, depending on what tests were performed. While the document
is being printed, the Phase 3 display will appear as shown in Figure 81.
Printing Report...
CANCEL
Figure 81 - Test Report printout in progress.
CANCEL may be pressed to abort the printout in progress. When the printout is completed, the
display will automatically return to the DATA LOG menu.
If Phase 3 detects that the receiving device is off-line, an alarm message will be shown (Figure 82).
Correct the problem with the receiving device, then press OK. The display will return to the DATA
LOG menu, from which a new printout can be initiated via the PRINT softkey.
PRINTER ERROR
OK
Figure 82 - Off-line error during printout.
For fully automated testing, Phase 3 provides 32 AutoSequences which may be user-programmed
by means of the accessory Phase3pc software (included on the CD ROM supplied with your
Phase 3). An AutoSequence test may include steps for measuring defibrillator energy, charge time,
and synchronization delay; ECG performance; and pacemaker pulse characteristics, noise immunity,
sensitivity, and refractory periods. Phase 3 comes equipped with pre-programmed AutoSequences
loaded into its non-volatile memory, which are provided as examples, and should be modified or
supplemented to meet the users needs via Phase3pc software.
For each test or measurement performed in an AutoSequence, the expected value of the result or
target is specified, so that Phase 3 will automatically pass or fail the result depending on an
acceptance range derived from the target and a selected tolerance.
There are two basic types of AutoSequence provided by Phase 3: Defib/Pacer and AED. In a
Defib/Pacer AutoSequence, the order of test execution is:
• Defib Energy
• Defib Charge Time
• Defib Cardioversion (Sync)
• ECG Tests
• Pacer Pulse
• Pacer Noise Immunity
• Pacer Sensitivity
• Pacer Refractory
In an AED AutoSequence, the only type of test that can be performed is the "shock/no shock"
energy test (see Section 4.8). Charge time, cardioversion, and pacemaker tests cannot be performed
if the "DUT Type" of the AutoSequence has been set, via Phase3pc software, to "AED".
When an AutoSequence is selected from the AutoSequence menu of Phase 3, the Clip Record is
erased. As each step of the AutoSequence is completed, the measurement data is accumulated in the
Clip Record. At the end of the AutoSequence, you will be prompted to enter an ID for the
defibrillator under test (DUT), after which a Test Record containing the results will then be saved to
the non-volatile Data Log.
Press the AUTO softkey on the Phase 3 MAIN MENU (Section 4.3) to enter a menu for selecting
an AutoSequence (Figure 83). Press the ? > keys to scroll down/up through the list of
AutoSequences one sequence at a time.
AUTOSEQUENCE TESTS
---------- SELECT TEST ----------
Test# Title >
>001 HP CODEMASTER
002 LAERDAL 2000
003 MARQUETTE 1500 ?
004 PHYSIO LP9
PAGE DN PAGE UP SELECT CANCEL
Figure 83 - Selecting an AutoSequence.
SELECT Press to select the AutoSequence indicated by the > cursor. The Clip Record will
be erased, then the AutoSequence will begin execution.
CANCEL Do not select an AutoSequence; return to the MAIN MENU.
NOTE:
If the Test Record portion of the Data Log is full when the AUTO softkey on the MAIN
MENU is pressed, an error message is displayed (Figure 84).
The Data Log must not be full when attempting to run an AutoSequence. Pressing
OK on the display of Figure 84 will return Phase 3 to the MAIN MENU. The Data Log
contents may then be downloaded and saved via Phase3pc software (see Chapter
6). After downloading, the Data Log may be partially or completely erased via the
DATA LOG menu (see Section 4.9).
OK
Figure 84 - AutoSequence execution prevented due to full Data Log.
When running an AutoSequence for a defib/pacer, Phase 3 will display test setup instructions for
each step of the sequence, similar to the example of Figure 85 for a defibrillator energy test. The
type of test being performed at any step in the sequence is indicated at the top left of the display
(DEFIB ENERGY in the example of Figure 85). The test count (01) and number of tests to be
performed in the sequence (in this example, 07) are shown at top right.
Any Phase 3 Autosequence can be programmed by the user to have an automatic timeout/advance
feature. At each test step, the Auto-Test Timer of Phase 3 will move the sequence forward from
the user prompt to the measurement automatically, provided there is no operator input before a
user-set delay time expires. After each measurement is completed, the Auto-Test Timer causes the
user prompt for the next step of the sequence to be displayed. The Auto-Test Timer of a
Autosequence may be set to "OFF", or it may be programmed for a 3, 5 or 9-second delay by means
of the Phase3pc software. For example, Figure 85 shows the first energy test of an Autosequence
for which the Auto-Test Timer has been enabled with a delay of 5 seconds. The operator may press
TEST to run the indicated test, or the user can wait for the Auto-Test Timer to count down to zero
at which point the energy measurement screen, Figure 86, will be displayed automatically.
TEST Press this softkey after setting the DUT as instructed, and proceed with the test
measurement indicated at top left on the display.
SKIP Press to skip the test indicated on the display and increment Phase 3 to the next step
of the sequence. Pressing this softkey will not affect any data stored in the Clip
Record for the step number indicated at the top right corner of the display.
BACK Press to make Phase 3 go back to the immediately previous step of the sequence.
Pressing this softkey will not affect any data stored in the Clip Record for the step
number indicated at the top right corner of the display.
ABORT Press to abort the sequence in progress. A prompt will be displayed asking you to
confirm the abort. Pressing YES in response to this prompt will abort the sequence
and return Phase 3 to the MAIN MENU.
If TEST of Figure 85 is pressed, Phase 3 will then display a screen corresponding to the type of
measurement which is to be performed. For the case of a defib energy test, the display will appear
similar to the example of Figure 86. On this display, the BACK softkey may be pressed to restore
the instruction screen shown in Figure 85.
BACK
Figure 86 - Energy Test display during an AutoSequence.
When a defibrillator discharge is detected, Phase 3 will calculate and display the results, including a
pass/fail assessment which is based on the target and tolerance which have been defined for this
step of the sequence. If the Auto-Test Timer is enabled, wait for approximately 4 seconds for
Phase 3 to automatically save the results and then advance to the next test step. The softkeys will be
blanked if the Auto-Test Timer is enabled.
If the Auto-Test Timer is disabled, Phase 3 will then provide the softkeys shown in Figure 87.
Press OK to save the displayed result to the Clip Record and advance the sequence to the next step.
Refer to Section 4.5.1.4 for a description of the other softkey functions appearing on the display of
Figure 87.
For charge time and cardioversion tests, instructions similar to Figure 85 are shown immediately
prior to each test step. The appearance of the display following acknowledgement of the instructions
will then be similar to the displays shown for manual charge time and cardioversion tests (see
Sections 4.5.2 and 4.5.3 respectively). After the defibrillator discharge, the results are displayed and
the softkeys will appear as shown in Figure 87.
Charge Time and Cardioversion (SYNC) tests operate in a similar manner, subject to the
descriptions in sections 4.5.2 and 4.5.3 respectively.
If the AutoSequence includes ECG tests, then this part of the sequence will be preceded by the
instruction display shown in Figure 88. If the Auto-Test Timer has been enabled, the delay
countdown will appear on the LCD per Figure 88.
ECG TEST
SKIP Press this softkey to skip over ALL ECG tests that have been specified in the
AutoSequence. If pacemaker tests have been enabled, the next display that will be
shown will be instructions for the pacer tests.
BACK Press this softkey to decrement the sequencer to the immediately previous defib test
(cardioversion, charge time, or energy - depending on what has been enabled in the
AutoSequence) and show the instruction screen appropriate for that measurement.
ABORT Press to abort the sequence in progress. A prompt will be displayed asking you to
confirm the abort. Pressing YES in response to this prompt will abort the sequence
and return Phase 3 to the MAIN MENU.
If TEST of Figure 88 is pressed, Phase 3 will then set the ECG simulator to the first waveform
which has been specified in the sequence (Figure 89). The operator must press one of the softkeys
to proceed, the auto-timer has no effect in this screen. The softkeys shown in the figure will then
operate as follows:
PASS Record a "PASS" in the Clip Record as the result for the displayed waveform, and
increment the sequence to the next ECG waveform, or alternatively to the first of
the pacer measurements if the current ECG waveform is the last waveform in the
sequence.
FAIL Record a "FAIL" in the Clip Record as the result, and increment the sequence to the
next ECG waveform.
BACK Return the sequence to the screen of Figure 88. Current contents of the Clip
Record are not affected.
If the AutoSequence includes pacemaker pulse tests, then this part of the sequence will be preceded
by an instruction display which will be similar to the example shown in Figure 90. As indicated in
the figure, the pacer test load which has been specified in the AutoSequence setup is indicated on
the top line of the display. Softkeys appearing on the display of Figure 90 provide the same
functions as described previously for defibrillator energy tests (see Section 4.10.3.1). The Auto-Test
Timer, if enabled, will not appear on the display of Figure 90 , but the timer will cause the sequence
to automatically advance to the next measurement after a 4-second delay once the pacer pulse test
has been stopped.
Press TEST after the pacemaker has been set up as instructed on the display of Figure 90. The
display will then appear as shown in Figure 91 below. The test will begin immediately after the
TEST key is pressed. Press BACK to review the instructions of Figure 90, or alternatively press
STOP to halt the measurement.
STOP BACK
Figure 91 - Start of pacer pulse test during an AutoSequence.
When STOP is pressed, the measurement will be halted and the results displayed in a manner
similar to the example shown in Figure 92. If the Auto-Test Timer is enabled, wait approximately 4
seconds for Phase 3 to automatically save the results and advance to the next test step. The softkeys
will all be blanked if the Auto-Test Timer is enabled. If the Auto-Test Timer is disabled, press the
OK softkey to save the data to the Clip Record and advance Phase 3 to the next step of the
sequence.
Instructional displays for pacer noise immunity, sensitivity and refractory tests will appear similar to
the screen of Figure 93. The softkeys on this display operate as previously described for the pacer
pulse test.
The pacer test procedure will be performed and the results displayed in a manner similar to the case
of a manual test performed with targets (see Figure 94). Note that noise immunity, sensitivity, and
refractory tests will each require some time to complete, ranging from about 10 seconds to 60
seconds, depending on the test and the rate setting of the pacemaker. When the test is complete and
the result is displayed, press the OK softkey to save the data to the Clip Record and advance
Phase 3 to the next step of the sequence, or wait for an automatic save if the Auto-Test Timer is
enabled. If the Auto-Test Timer is enabled, the softkeys will be blanked when the measurement is
completed.
When running an AutoSequence for an AED, Phase 3 will operate similarly as described in
Section 4.8 for a manual AED test. The primary difference between a manual AED test and an
AutoSequence AED test is that the ECG waveforms are defined in the setup of an AED
AutoSequence. The Auto-Test Timer cannot be enabled in AED Autosequences.
As in the manual mode, the ECG waveform may be set to revert to NSR60 after a discharge has
been detected. This setting may be toggled using the AUTO-NSR key. After the test result has
been accepted with the OK key, the ECG waveform will change to the setting dictated by the next
step in the AutoSequence.
Upon selecting an AED AutoSequence, the display of Phase 3 will appear as shown in Figure 95.
The softkeys will then assume the following functions:
NO SHOCK Press this softkey if the AED audibly and/or visually advises the user that the
AED cannot or should not be discharged. For further details, see Section 4.8.
AUTO-NSR Press this softkey to toggle between changing to NSR60 after a discharge is
detected, and, retaining the current ECG WAVE after a discharge is detected.
The ECG WAVE will change to the next waveform in the sequence when the
results are accepted with the OK key.
BACK Press this softkey to decrement the sequencer to the immediately previous AED
test step.
ABORT Press to abort the sequence in progress. A prompt will be displayed asking you
to confirm the abort. Pressing YES in response to this prompt will abort the
sequence and return Phase 3 to the MAIN MENU.
If NO SHOCK is pressed, or if a discharge is detected, Phase 3 will automatically assess the result
as shown in the example display of Figure 96. When the result is displayed, press OK to save the
result to the Clip Record and advance to the next step of the sequence.
ENERGY: 150.8J
ECG WAVE: VTACH2 (x1 Scale)
ECG TYPE: SHOCKABLE
***************** PASS *****************
OK
Figure 96 - AED test result during an AutoSequence.
If the AED has delivered a shock, and if the AED has been configured to automatically execute a
"shock series", you should press the OK key before the AED announces that the ECG is being
analyzed. For a shockable ECG waveform, enabling the AUTO-NSR feature of Phase 3 before
activating the "ANALYZE" function on the AED should inhibit the AED's "shock series" after the
first discharge, in which case you will then have more time to respond to the test result with the OK
key. Alternatively, when a shock is delivered be prepared to press OK as soon as the energy result is
displayed; this will depend on the particular shock protocol developed by the manufacturer of the
AED.
After all measurements in the AutoSequence have been completed, the operator will be prompted to
save the results to a Test Record via the menus of Figure 97, Figure 98.
SAVE RECORD
SAVE Press to save the measurement data acquired during the AutoSequence to a Test
Record in the non-volatile Data Log. You will be prompted to enter an ID
(e.g. control number) for the defibrillator under test. Refer to Section 4.4.1 for
instructions regarding ID entry.
BACK Press this softkey to return Phase 3 to the last test performed in the AutoSequence.
ABORT Press to abort the sequence and discard all data acquired. A prompt will be displayed
asking you to confirm the abort. Pressing YES in response to this prompt will abort
the sequence and return Phase 3 to the MAIN MENU.
After the results have been saved, the screen of Figure 99 will be displayed. Press OK to exit the
AutoSequence.
AUTOSEQUENCE TEST RESULTS SAVED
Press OK to continue.
OK
Figure 98 - Prompt displayed at end of AutoSequence.
When Phase 3 is powered up, it will conduct a self-test as described in Section 4.2. Normally the
self-test will pass, allowing the user to access the MAIN MENU of Phase 3.
Failures encountered during the self-test are reported via the display, accompanied by an alarm tone.
Critical failures will prevent further operation of the instrument. For non-critical self-test failures,
Phase 3 will report the fault but will remain operable after the alarm is acknowledged by the user.
BATTERY LOW WARNING Phase 3 is being operated from the internal NiCad battery, but
the battery has less than 20 minutes of usage remaining, before
Phase 3 will automatically shut itself off. The AC adapter
should be connected to Phase 3 as soon as possible. Phase 3
may be operated while the battery is being recharged.
CLOCK WARNING The time/date setting of the internal real-time clock was found
to be invalid. After this alarm is acknowledged, the user will be
prompted to set the clock. This self-test alarm may indicate the
lithium battery which powers the clock should be replaced. The
clock backup battery is not a user-serviceable item. Refer to
Chapter 7 of this manual.
5 REMOTE CONTROL
The Serial I/O Port provides a RS-232 interface for controlling Phase 3 remotely using an external
device such as a personal computer or automated test system. Phase 3 will recognize a predefined
set of commands issued to its Serial I/O Port allowing the external device to start a test, stop a test,
or retrieve test results automatically.
Remote commands allow the creation of simple programs on a personal computer that will
automatically perform and document a defibrillator and/or pacemaker test according to a desired
protocol. The command set is also compatible with automated test instrumentation, such as the
ES601Plus Automated Safety Analyzer. Refer to the Operating Manual for your automated test
instrument, especially with regard to port settings, cable wiring, data transmission format and data
packet length. Note the serial protocol of Phase 3 is 9600 Baud, 8 Data Bits, No Parity, 1 Stop Bit
(9600, N, 8, 1).
The following sections present the remote control command set of Phase 3. When using these
commands, it is assumed the user has connected the DUT to Phase 3 prior to enabling the
controlling device, and the controlling device is connected to Phase 3 using the appropriate interface
cable as described in Section 3.2.3. If using a program on a personal computer to command Phase 3,
allow a delay of at least 100 milliseconds between consecutive remote commands to provide enough
time for Phase 3 to process and respond to each received command.
In order for Phase 3 to respond to remote control commands, it must first be placed in Remote Mode.
Phase 3 will ignore all commands sent unless the instrument is running the Remote Mode interface.
To enter Remote Mode, press the REMOTE softkey on the MAIN MENU of Phase 3. The display
will then appear as shown in Figure 99. Commands received via the Serial I/O Port are displayed
on the LCD, as are the data which are returned in response to the commands. Phase 3 also displays
a counter which increments each time a valid command is received and processed. Between
commands, CANCEL may be pressed to exit Remote Mode to the MAIN MENU of Phase 3.
CANCEL
Figure 99 - Remote Mode menu.
There are 16 commands in the remote control command set of Phase 3. Each command consists of
a string of upper-case ASCII characters and numerals, terminated by a carriage return character
(<CR> or \r).
If Phase 3 receives a valid RS-232 command, it will execute the command, and return either an
asterisk character followed by a carriage return (*<CR>) or a packet of data back to the controlling
device depending on the type of command issued. With many of the measurement-related
commands, a substantial amount of time may elapse before any data, for example energy test results,
are returned to the controlling device.
Description: The ECG configuration command sets the waveform, amplitude scale and discharge
response mode of the ECG simulator. Typically, this command would be sent prior
to other commands which invoke defibrillator or pacemaker measurements.
Syntax: ECG<code><scale><mode><CR>
where: <code> is a two letter code designating the Wave Bank and Waveform of
the simulator. See table below for the Bank and Waveform letter codes.
Notes: To stop the simulation corresponding to waveform codes AF-AL, press the STOP
softkey which will be displayed once the simulation has been started, or alternatively
transmit an ASCII <Escape> character (0x1B) to Phase 3 via the communications
connection to abort the simulation and permit Phase 3 to receive and process
subsequent commands.
DC Pulse, 4sec AA
Square Wave, 2 Hz AB
Triangle, 2 Hz AC
Sinewave, 0.1 Hz AD
Sinewave, 0.5 Hz AE
Sinewave, 10 Hz AF
Sinewave, 20 Hz AG
Sinewave, 40 Hz AH
Sinewave, 50 Hz AI
Sinewave, 60 Hz AJ
Sinewave, 70 Hz AK
Sinewave, 100 Hz AL
Multifocal PVC CF
Run of 5 PVC CG
Bigeminy CH
Trigeminy CI
Asystole (noise) DC
Asystole (flatline) DD
SupraVent. Tachycardia DE
Shock Advisory/AED
Ventricular Tachycardia, 140BPM DF
Coarse VFIB DJ
Fine VFIB DK
10msec / 60BPM EA
25msec / 60BPM EB
100msec / 60BPM ED
200msec / 60BPM EE
10msec / 60BPM FA
25msec / 60BPM FB
100msec / 60BPM FD
200msec / 60BPM FE
10msec / 60BPM GA
25msec / 60BPM GB
100msec / 60BPM GD
200msec / 60BPM GE
Description: The defib energy command configures Phase 3 to perform an energy test. Phase 3
will return data in one of two possible formats after a discharge is acquired. See
Section 4.5.1 for further information on the defib energy test.
Syntax: DE<range><return_data_format><CR>
<return_data_format> is:
C - for comma-separated variables (CSV) format
T - for "text" format
Example: "DEHT\r"
Run defib energy test; high range; return data in text format.
Description: The defib charge time command configures Phase 3 to perform a charge time test.
Receipt of this command will initiate the 5-second countdown timer shown on the
display. Phase 3 will return data in one two possible formats, after a discharge is
acquired. See Section 4.5.2 for further information on the defib charge time test.
Syntax: DCH<return_data_format><CR>
Example: "DCHC\r"
Notes: Charge Time measurements are normally preceded by one or more discharges of the
defibrillator with the AC power disconnected. This portion of the test must be
simulated in the remote mode by performing regular energy discharges. The
programmer must keep track of the number of pre-test discharges performed.
Description: The defib sync command configures Phase 3 to perform a discharge sync/
cardioversion test. Phase 3 will return data in one of two possible formats after a
discharge is acquired. See Section 4.5.3 for further information on the defib
cardioversion test. If the current ECG WAVE setting is not appropriate for this
test, the ECG WAVE will be set to AFIB x1 automatically.
Syntax: DS<range><sync><return_data_format><CR>
<return_data_format> is:
C - for comma-separated variables (CSV) format
T - for "text" format
Example: "DSHQT\r"
Run defib sync test; high range; Q-wave sync; return data in "text" format
Description: The pacer load select command sets the load resistor for pacemaker tests. Typically,
this command would be sent prior to other commands which invoke pacemaker
measurements.
Syntax: PLD<code><CR>
where: <code> is a single-letter code designating the load resistor. See table below
for the resistor letter codes.
Example: "PLDd\r"
Description: The pacer pulse command configures Phase 3 to perform a pacer pulse test. Phase 3
will return data in one of two possible formats after at least two consecutive pacer
pulses have been acquired. See Section 4.7.2 for further information on the pacer
pulse test.
Syntax: PP<method><return_data_format><CR>
<return_data_format> is:
C - for comma-separated variables (CSV) format
T - for "text" format
Example: "PPLC\r" Run pacer pulse test; leading edge method; return data in
CSV format.
Description: The pacer refractory command configures Phase 3 to determine both paced and
sensed refractory intervals (PRP and SRP). Phase 3 will return data in one of two
possible formats, after both the PRP and SRP have been measured; this will require
about 40 seconds at a pacer rate of 90 PPM. See Section 4.7.3 for further
information on the pacer refractory test. If the current ECG WAVE setting is not
appropriate for this test, the ECG WAVE will be set to TRIP40 x2 automatically.
Syntax: PRF<return_data_format><CR>
Example: "PRFT\r" Run pacer refractory test; return data in text format.
Description: The pacer noise immunity command configures Phase 3 to perform a noise
immunity test. Phase 3 will return data in one of two possible formats after the test is
completed; if the test passes, this will require about 60 seconds at a pacer rate of 90
PPM. See Section 4.7.4 for further information on the pacer noise immunity test.
Syntax: PN<freq><return_data_format><CR>
<return_data_format> is:
C - for comma-separated variables (CSV) format
T - for "text" format
Example: "PNAT\r"
Run pacer noise immunity test; noise frequency = 60 Hz; return data in text format.
Description: The pacer sensitivity command configures Phase 3 to perform a pacer sensitivity
test. Phase 3 will return data in one two possible formats, after the test is completed;
this will require about 20 seconds at a pacer rate of 90 PPM. See Section 4.7.5 for
further information on the pacer sensitivity test.. If the current ECG WAVE setting
is not appropriate for this test, the ECG WAVE will be set to TRIP40 automatically
(scale not required).
Syntax: PSN<return_data_format><CR>
Example: "PSNT\r"
Description: The ID command causes Phase 3 to return its name to the controlling device. This
command can be used to check for connection of Phase 3 to the controlling device,
and to verify that Phase 3 is operating in Remote Mode.
Syntax: IDENT<CR>
Example: "IDENT\r"
Description: The CLT command causes Phase 3 to delete the last test record stored in EEPROM
Syntax: CLT<CR>
Example: "CLT\r"
Description: The CWV command causes Phase 3 to delete all Wave Records stored in EEPROM.
Syntax: CWV<CR>
Example: "CWV\r"
Description: The CTR command causes Phase 3 to delete ALL Test and Wave Records stored in
EEPROM.
Syntax: CTR<CR>
Example: "CTR\r"
Description: The WAV command causes Phase 3 to return the waveform data currently held in
the data acquisition buffer via the RS-232 port.
This command works ONLY after completion of a successful “DE” (defib energy
test) or “PP” (pacer pulse test) command. The data returned in response to “WAV”
will represent a defibrillator discharge waveform after a “DE” command, or a single
pacer pulse after a “PP” command.
Syntax: WAV<length><sample_rate><format><CR>
Example: "WAV3HA\r"
Returns: The returned signal sample data are signed values. For a defib waveform, the data
correspond to current in units of amperes x 100 (e.g. a value of -1579 corresponds to
a signal sample equal to -15.79 amperes). For a pacer waveform, the data correspond
to current in units of milliamps x 10 (e.g. a value of +453 corresponds to +45.3
milliamps).
(1) For ASCII return format, each signal sample is represented as a string of
ASCII digits, preceded by a ‘+’ or ‘-’ character to convey the sign as
appropriate, and terminated with a carriage return character (‘\r’). After all
requested signal data have been returned, a form feed character (‘\f’) is then
sent and can be used to detect conclusion of the waveform download in
ASCII mode.
(2) For binary return format, each data value is sent as a signed short (16-bit)
integer, sent in low byte/high byte order. The requesting host can detect
end-of-download via a timeout mechanism, or by waiting for the expected
number of data values to be received. The number of data points conveyed
in the download depends on the command parameters and is listed in the
following table.
<sample_rate>
<length> ‘L’ ‘M’ ‘H’
‘1’ 220 440 880
‘2’ 440 880 1760
‘3’ 660 1320 2640
‘4’ 880 1760 3520
‘5’ 1100 2200 4400
‘6’ 1320 2640 5280
Description: The MEM command causes Phase 3 to return the number of Test Records currently
stored in the EEPROM memory.
Syntax: MEM<CR>
Example: "MEM\r"
Returns: (1) "nnn\r" where nnn is " 0" to " 80" (note leading whitespace).
(2) "?\r" Syntax error/invalid command
Description: The STR command causes Phase 3 to return the Test Record requested, in the
format of a printed Test Report (see section 4.9.2.2).
Example: "STR16\r"
Returns: (1) "ASCII printed report" There is no 'end-of-file ' marker. User must
monitor for a specific text string (such as
Date:), or time out to end the transmission.
(2) "e\r" Record does not exist.
(3) "?\r" Syntax error/invalid command
6 Phase3pc SOFTWARE
Datrend Phase3pc software provides a communication utility for Phase 3 which allows the user to
configure AutoSequences, download Test Records and Wave Records, and capture waveform data
in real-time. Phase3pc is a self-contained Windows® application which is supplied as a standard
accessory on CD ROM.
6.1 Installation
To install Phase3pc to your computer, place the Phase3pc CDROM into your computer's
CDROM drive and close the drive. The installation application will begin automatically
(Figure 100), unless your computer is set to ignore autorun applications.
In the event the installation program does not start automatically, use the file explorer utility on your
computer to locate the CDROM drive. Browse this drive to locate the file called
Phase3pcNNSetup.exe . Double-click or otherwise execute this file and the installer will guide you
through the installation process. We recommend that all default installation locations be left intact
so future upgrades and enhancements will run smoothly.
At the end of the short installation process, you will be presented with a notification that the
software has been installed, at which time you may either select or deselect the checkbox which asks
if you would like to run Phase3pc upon closing the dialogue box. The computer does not have to
be restarted before running Phase3pc
When installation of Phase3pc is complete, connect Phase 3 to your PC using the USB cable
provided with your instrument. When Phase 3 is connected to the PC for the first time, Windows
should detect the connection and then automatically launch the New Hardware Wizard which will
prompt you for the USB driver (Figure 101).
The USB drivers for Phase 3 are provided with Phase3pc software. When the Hardware Wizard
asks for the files, direct the Hardware Wizard to your CD drive, or to the location where the files
were downloaded. The Wizard should then find and install the required driver files.
NOTE:
Figure 102 - Click "continue anyway". Figure 103 - Driver installation in progress.
In order for Phase 3 to work with Phase3pc, it must first be placed in Remote Mode. To enter
Remote Mode, press the REMOTE softkey on the MAIN MENU of Phase 3. The display will then
appear as shown in Figure 104. CANCEL may be pressed at any time to exit Remote Mode to the
MAIN MENU of Phase 3.
CANCEL
Figure 104 - Remote Mode menu of Phase 3.
When Phase3pc is run for the first time, the program starts up by displaying a library of
AutoSequences which are stored locally on the computer (Figure 105). Each row in the grid
represents one AutoSequence, as identified by a title and the type of device to be tested
(defibrillator/pacer, or AED).
To copy AutoSequences, select any number of AutoSequences in the grid by dragging the mouse
over them. Then, click Copy AutoSequences, or press b +C on the keyboard. If the
resulting number of AutoSequences is 32 or less, the copied AutoSequences will appear at the
bottom of the grid.
To delete AutoSequences, select any number of AutoSequences in the grid by dragging the mouse
over them. Then, click Delete Autosequences, or press c on the keyboard. A warning will
appear requiring you to confirm the action. Click Yes to delete the AutoSequences.
AutoSequences can be transferred to Phase 3 via a USB connection. Before attempting to upload or
download, make sure Phase 3 is connected to the PC and that Phase 3 is running in Remote Mode
(see Section 6.2).
To upload AutoSequences into Phase 3, click the Upload Autosequences button, or press
b +U on the keyboard. To download AutoSequences stored in Phase 3 to the PC, click the
Download Autosequences button, or press b +D on the keyboard. In both instances, a
communications progress window will appear. When the transfer has completed, click “OK”.
Phase3pc allows you to print all the AutoSequences, or a selected number of AutoSequences. To
print one AutoSequence, highlight it in the grid. To print a range of AutoSequences, select the
desired sequences by dragging the mouse over them.
Click Print Autosequences, or press b +P on the keyboard. The Print window will appear.
Choose to print either the selected AutoSequence(s) or all of the AutoSequences, via the radio
buttons which are provided on the Print dialog.
Some test types, such as defibrillator energy, can be repeated a number of times with different
configurations for each test. The AutoSequence Editor provides each test type with its own tab. If
the test is a repeatable type, a list box will appear on the left-hand side of the Editor window, with
Add and Delete buttons located below the list (Figure 107). The test options will appear
on the right-hand side of the Editor window.
To add a new test to the list, press the Add test button below the list box. Select a test in the
list box by clicking on it. The parameters of the selected test will then appear in the text boxes on
the right. Delete a test from the list by selecting the test in the list box and clicking the Delete test
button, or pressing c on the keyboard.
Edit the title of the AutoSequence in the Autosequence Title text field on the top of the
Autosequence Editor. Normally, you would choose a title that represents the make and model of the
defibrillator. The title can be a maximum of 16 characters.
To test either a defibrillator and/or transcutaneous pacemaker, select Defibrillator in the Device
Under Test combo box. To test an automated external defibrillator (AED), select AED in the
Device Under Test combo box.
A defibrillator energy test measures the energy delivered by a defibrillator when it is discharged. The
test is normally repeated several times, with the defibrillator set to different energy levels for each
test. In an AutoSequence, each measured energy is compared a target value corresponding to the
energy setting on the defibrillator. The tolerance of the test defines a range of acceptance for the
measured energy. If the measured energy falls within the acceptance range defined by the
AutoSequence settings for target and tolerance, Phase 3 automatically passes the measured energy.
To begin editing defibrillator tests, select the Defib Energy tab (Figure 107). The maximum
number of energy tests than can be performed in an AutoSequence is 10. Add tests if required, via
the button as previously described. Select a test from the list on the left. Enter the Target
Energy for this test (i.e., the energy setting on the defibrillator), and choose a Tolerance. Set up
the Phase 3 ECG simulator for this test by first choosing the simulator Wave Bank, and then
selecting the Wave (waveform). The amplitude scale of the Wave can also be changed by selecting a
value in the Scale combo box.
A charge time test measures the time required for a defibrillator to charge up to a specified energy
level, which is typically the maximum energy provided by the device.
To begin editing the charge time test, select the Charge Time tab on the AutoSequence Editor.
An AutoSequence provides only one charge time test. To enable this test, click the Enable
Charge Time Test checkbox. Enter the Target Energy for this test (i.e., the maximum energy
setting on the defibrillator), and choose a Tolerance. Enter an upper limit for the measured charge
time in the Measured Charge Time box. Set up the Phase 3 ECG simulator for this test by first
choosing the simulator Wave Bank, and then selecting the Wave (waveform). The amplitude
scale of the Wave can also be changed by selecting a value in the Scale combo box.
A cardioversion test verifies the ability of a defibrillator to synchronize its discharge to the QRS
complex of the ECG.
To begin editing cardioversion tests, click the Cardioversion tab of the AutoSequence Editor.
When the first cardioversion test is enabled, the ECG simulator will be configured so that it reverts
to normal sinus rhythm when a discharge is detected. If desired, disable this option under the
Advanced tab of the AutoSequence Editor.
The maximum number of cardioversion tests than can be performed in an AutoSequence is 4. Add
tests if required, via the button as previously described. Select a test from the list on the left.
Click the Enable Delay Target checkbox if you want Phase 3 to pass or fail the measured sync
delay. Enter the Target Energy for this test (i.e., the energy setting on the defibrillator), and
choose a Tolerance. Set up the Phase 3 ECG simulator for this test by first choosing the simulator
Wave Bank, and then selecting the Wave (waveform). The amplitude scale of the Wave can also
be changed by selecting a value in the Scale combo box. In the Sync To combo box, select the
timing reference that will be used to measure the synchronization delay (Q or R wave).
An ECG test verifies operation and performance of the monitor and/or chart recorder of a
defibrillator. To begin editing ECG performance tests, click the ECG Performance tab of the
AutoSequence Editor (Figure 108).
The maximum number of ECG tests that can be performed in an AutoSequence is 12. Select the
ECG signal amplitude via the Scale combo box; the same amplitude will be used for all waveforms
in the ECG sequence. Set the ECG simulator for each step of the sequence by first choosing the
simulator Wave Bank, and then selecting the Wave (waveform).
To begin editing pacer tests, click the Pacer tab of the AutoSequence Editor. The pacer sub menu
will appear below the main tab menu (Figure 109). Click on the buttons of the pacer sub menu to
navigate the various pacer tests described in the following sections.
The pacer load is the load value which will be used by Phase 3 in all pacer tests of an AutoSequence.
Click the up or down arrow to respectively increase or decrease the resistance in 50 ohm increments.
A pacer pulse test measures amplitude, rate and width of the pulses generated by a pacemaker
(Figure 109). The maximum number of pulse tests that can be performed in an AutoSequence is 10.
To edit pacer pulse tests, click the "Pacer" tab and then click the "Pulse" button on the pacer sub
menu. Select a test in the list on the left by clicking on it. For the selected test, enter a "Target
Current" and Pulse Rate in the text fields provided on the right. A target for pulse width is
optional, and can be enabled by clicking the checkbox next to the Target Pulse Width text field.
Select the measuring method for pulse amplitude via the Measuring Method combo box. For
more information on amplitude measuring methods, see Section 4.7.2.
A noise immunity test verifies that a pacer will not respond to line frequency noise injected into its
ECG input. An AutoSequence can include only one noise immunity test.
To edit the noise immunity test, click the Pacer tab and then click the Noise Immunity button
on the pacer sub menu. To enable the test, click the Enable Noise Immunity checkbox. Select
the frequency of the noise signal via the Test Type combo box.
A demand sensitivity test determines the smallest ECG signal that will reliably trigger a pacer when
the pacer is operating in demand mode. The maximum number of sensitivity tests that can be
performed in an AutoSequence is two.
To edit the sensitivity test, click the Pacer tab and then click the Demand Sensitivity button on
the pacer sub menu. Select a test in the list on the left by clicking on it. For the selected test, enter a
Sensitivity Limit in the text field provided on the right. This limit is the largest ECG signal that
can be accepted to result in a passing test. Set up the Phase 3 ECG simulator for this test by first
choosing the simulator Wave Bank, and then selecting the Wave (waveform).
A refractory period test determines the two refractory periods of the pacemaker under test. The
paced refractory period (PRP) is the time interval following a pacer pulse during which ECG QRS
complexes are ignored. The sensed refractory period (SRP) is the time interval following a detected
QRS where a subsequent QRS is ignored. The maximum number of refractory tests that can be
performed in an AutoSequence is two.
To edit the refractory test, click the Pacer tab and then click the Refractory Period button on
the pacer sub menu. Select a test in the list on the left by clicking on it. For the selected test, enter
maximum and minimum limits for both PRP and SRP in the text fields provided. Set up the Phase 3
ECG simulator for this test by first choosing the simulator Wave Bank, and then selecting the
Wave (waveform).
The ECG Trigger Mode listbox allows selection of one of four possible responses of the ECG
simulator to a defibrillator discharge. Phase 3 may automatically change its waveform to normal
sinus rhythm when a discharge is detected from the defibrillator under test. Alternatively or in
addition, Phase 3 can play back the discharge waveform into the paddles or ECG leads of the
defibrillator. See section 4.5.1.2 for more details on the MODE settings.
The Auto-Test Timer listbox allows selection of the length of time Phase 3 will wait before
proceeding to the next step in an AutoSequence, where appropriate. The Auto-Test Timer allows
the operator time to read or respond to data or messages on the display screen, and may be turned
OFF if full manual control of the Autosequence is desired. See section (ref 4.10.3.1) for more details
on the Auto-Test Timer.
If the device under test is an AED, select "AED" in the Device Under Test combo box at the
top of the AutoSequence Editor. The AED Energy test options will appear, while defibrillator and
pacer test options will be hidden (Figure 110).
The maximum number of energy measurements that can be performed in an AED AutoSequence
is 32. Select a test in the list on the left by clicking on it. Set up the Phase 3 ECG simulator for this
test by first choosing the simulator Wave Bank, and then selecting the Wave (waveform). The
amplitude scale of the Wave can also be changed by selecting a value in the Scale combo box. The
Shockable field displays the expected outcome of the test according to AAMI guidelines (see
Section 4.8).
After a test is completed with Phase 3, results of the test can be stored as a Test Record in the non-
volatile Data Log. Later, the test data can be downloaded into Phase3pc and viewed as Test Report
documents. Test Reports can be saved in text, CSV, and Sentinel32 MUP formats.
Start working with test reports by clicking the Test Reports button on the main menu of
Phase3pc. This will activate the Test Report toolbar (Figure 111).
Test Report data can be transferred from Phase 3 via a USB connection. Before attempting to
download, make sure Phase 3 is connected to the PC and that Phase 3 is running in Remote Mode
(see Section 6.2). To download Test Reports, click the Download Test Reports button, or press
b +D on the keyboard. A communications progress window will appear. When the transfer has
completed, click "OK". The downloaded Reports will appear in a grid display below the Test Report
toolbar (Figure 112).
To view one or more Test Reports, simply select the desired Test Reports in the main Test Report
grid (Figure 112) by dragging the mouse over them. Click the View Test Reports button. The
Test Report Viewer will appear showing the test data from the selected Test Reports.
Double-clicking on a single Test Report will also open the Test Report Viewer, which will then
show only the selected Report (Figure 113 .
In the Test Report Viewer, Test Reports can be saved in both TXT (Text) and CSV (Comma
Separated Values) formats. To save the viewed Test Report(s) in text format, click the "Save To
Text" button of the Test Report Viewer. A directory browser will appear. Select the directory where
the Test Reports should be saved. A file for each individual Test Report in the Viewer will now be
created, where each file will be automatically assigned a filename according to:
DeviceID.YYYY_MM_DD.hh_mm_ss.txt
Above, DeviceID is the defibrillator ID (equipment control number or other ID), YYYY is the
year, MM is the month, DD is the day, hh is the hour, mm is the minute, and ss is the second in
which the Test Report file was created on the PC.
Saving Test Reports in CSV format produces the same result as saving to text format, except a CSV
file is also created listing each device and a path to the Test Report text file for that device. Each line
of the CSV file thus consists of:
"Device ID", "Test Date and Time", "Test Type", "Device Under Test","Overall Pass",
"Filename and Path of Test Report"
Before creating the CSV file, a save file dialog will appear. Select the desired directory and specify a
name for the CSV file. The CSV file and all corresponding Test Report text files will be saved in the
chosen directory.
To save the viewed Test Reports in Sentinel32 MUP format, select the Save To MUP option
under the File menu of the Test Report Viewer. Select a location and either accept the default
filename which is generated automatically, or enter an alternative filename on the Save dialog. By
default, Phase3pc automatically generates a "medTester-style" filename which is intended to
facilitate direct import into Sentinel32 through the medTester Communication dialog of that
program. The format of the default MUP filename is:
MMDDhhmm.mup
where MM is the month, DD is the day, hh is the hour and mm is the minute in which the MUP file
was created on the PC.
In the main Test Report window of Phase3pc, there is a search text box in the upper right hand
corner (Figure 114). Use this to filter the downloaded Test Reports by Device ID, Device Under
Test, Date, and Test Type. The search is real-time, so the results will display as the characters are
typed into the search box. To cancel a search, click the Cancel Search button to the right of the
search box, or press the | key on the keyboard.
To open a previously-saved Test Report for viewing, click the Open Test Report Button, or
press b +O on the keyboard. Browse for the Test Report file to view.
When viewing Test Report(s) in the Test Report Viewer, click the "Print" button to show the print
dialog. Select the printer and click "OK" to print all Test Reports currently being shown in the
Viewer. Clicking "Cancel" will close the print dialogue without printing.
Wave Records can be saved by Phase 3 while testing a defibrillator (for more information on Wave
Records, see Section 4.5.1.5). A total of 10 low-resolution waveforms can be stored as individual
Wave Records in the non-volatile Data Log of Phase 3. Later, these waveforms can be downloaded
to Phase3pc for plotting as graphs.
Start working with waveforms by clicking the Waveforms button on the main menu of Phase3pc.
This will activate the Waveform toolbar (Figure 115).
Wave Record data can be transferred from Phase 3 via a USB connection. Before attempting to
download, make sure Phase 3 is connected to the PC and that Phase 3 is running in Remote Mode
(see Section 6.2). To download Wave Records, click the Download Test Reports button, or
press b +D on the keyboard. A communications progress window will appear. When the
transfer has completed, click "OK". The downloaded Wave Records will appear in a grid display
below the Waveform toolbar (Figure 116).
To view a Wave Record, double-click on an entry in the main waveform grid. The Waveform
Viewer will then appear (Figure 117). See Using The Waveform Viewer (Section 6.5.5) for more
information.
After making a Defibrillator Energy measurement or Pacer Pulse measurement with Phase 3
according to Section 4.5.1.4, the waveform from the test can then be captured by Phase3pc.
Phase 3 does not need to be running in Remote Mode for this to work. Simply make sure Phase 3 is
connected to the PC via USB, then click the Capture Waveform button, or press b +W on
the keyboard.
Note: be sure to use a high quality, shielded USB 2.0 type cable (i.e. Datrend P/N 3140-440)
connected to a USB port on the computer which is not on an extension cable from the main board.
Defibrillator discharges may cause a disconnection of the USB device driver if the shield and
grounding are not adequate. If disconnection occurs, Exit the program, unplug and re-plug the USB
cable and re-start Phase3pc .
After the data transfer is complete, the Waveform Viewer will appear immediately with the resulting
defibrillator discharge or pacer pulse waveform. Based on the content of the data received by
Phase3pc, the display will format the results and graph to match a defibrillator energy discharge or a
pacer pulse. See Using The Waveform Viewer (Section 6.5.5) for more information on
manipulating the graph characteristics.
To open a previously saved waveform for viewing, click the Open Waveform Button, or press
b +O on the keyboard. Browse for the waveform file to view.
Opening a Wave Record or waveform data file, or capturing a high-resolution waveform, activates
the Waveform Viewer toolbar of Phase3pc (Figure 118). Waveform viewing and printing features
are described in this section.
Zoom in by clicking the Zoom In button, or by pressing b += on the keyboard. Zoom out by
clicking the Zoom Out button, or by pressing b +- on the keyboard. Zoom into a particular
region of the chart by clicking the Zoom Rectangle button. The mouse cursor will change into a
cross symbol. Select a zoom area by holding the left mouse button down and dragging, creating a
rectangular area on the chart. Then, release the mouse button to see the changes. Note: The zoom
area rectangle must not go left of the Y-Axis Zero Position.
Click the Zoom Rectangle button if this button is not already depressed. Pan the chart by
holding down the left mouse button and dragging inside the chart area. The Zoom Rectangle
button must be depressed to enable the drag-panning feature.
Click the Undo Zoom button, or press b +Z on the keyboard to restore the chart to its
original view.
Click the Show Data Points button, or press b +T on the keyboard to activate or de-
activate markers on the individual data points which form the waveform trace.
To explicitly set one or both chart axes, select Set Chart Axis from the Edit pull-down menu
above the toolbar, or press b +H on the keyboard. Alternatively, you can right-click the chart
and then select Set Chart Axis. This will activate the Set Chart Axis form (Figure 119). Type
minimum and maximum values for the time axis and the current axis in the text boxes provided,
then click "OK". The chart will be re-drawn with the axes you have selected.
A waveform being viewed can be saved in CSV format by clicking the Save button on the
Waveform Viewer toolbar, or by pressing b +S on the keyboard. A Save Dialog will appear.
Select the directory, enter a filename and click "OK". Each data point in the CSV file will be
represented on its own line, in the following format:
[time],[current]
A waveform being viewed can be printed. To print the waveform, click the Print button on the
Viewer toolbar, or press b +P on the keyboard. A print dialog will appear. Select the printer
and click "OK". The waveform will be printed as it appears in the chart, i.e., with the existing
zoom/axis settings.
The waveform can be printed either with or without the test data shown at left in the Wave
Information Panel of the Viewer. To toggle the information panel on or off, click Show Wave
Info on the View pull-down menu above the toolbar, or press b +I on the keyboard.
After capturing a high-resolution waveform, you may wish to input a device ID to identify the
device which produced the waveform. To edit the device ID, select the Device ID item under the
Edit pull-down menu above the toolbar. Alternatively, click on the Device ID field in the Wave
Information Panel. The Device ID field will become an editable field. Type the new device ID into
the field, then press e on the keyboard.
b +C Copy
b +D Download
b +F Search / Find
b +N New
b +O Open
b +P Print
b +S Save
b +U Upload
b +W Capture Waveform
b +Z Undo Zoom
b += Zoom In
b +- Zoom Out
c Delete
| Cancel Search
In the event of battery discharge, connect Phase 3 to the AC adapter/charger supplied with your
instrument. Refer to Section 3.1 for further information on the AC adapter/charger. A fully
depleted battery will require a minimum of 14 hours to recharge. Phase 3 may be operated from the
AC adapter while the battery is being charged.
Phase 3 contains no user-serviceable parts. Opening the case of Phase 3 for any
reason will void the warranty and may compromise user safety. Refer servicing to
qualified personnel.
NOTE:
Protection of the operator may be compromised if the
instructions in this manual are not followed, or if Phase 3 is
used for a purpose not specified in this manual.
For service assistance or calibration, contact Datrend Systems for a Return Materials Authorization
(RMA) number and the location of the nearest Service Facility.
Phase 3 should be cleaned with a soft, lint-free, damp cloth using a mild detergent. Use of other
cleaning agents is prohibited.
CAUTION: This device measures high voltage signals. For protection of the
operator, inspect the Paddle Adapter and the interconnecting Paddle Adapter
Cables on a regular basis for nicks, cracks or other damage that may compromise
the insulating properties of the housing, wire or connectors. Replace any
damaged or worn components immediately.
Phase 3 comes equipped with AutoSequences stored in its non-volatile memory to provide
examples of various test parameters configured by an AutoSequence. Default AutoSequences are
installed with Phase3pc when the application is installed. Phase3pc permits the default
AutoSequences to be modified or supplemented as desired; refer to Chapter 6 for instructions. The
following AutoSequences are examples only to highlight some of the possible settings, and may not
match those installed in your Phase 3 . They may be modified at any time by Datrend in future
releases of Phase 3 or Phase3pc . These AutoSequences should be reviewed by the end user to
determine if they are suitable for their purposes. Datrend takes no responsibility for the
inappropriate use of these examples.
Cardioversion Test 2
Target Energy: 100 J
Tolerance: +/-15%
Waveform: Atrial Fibrillation, Coarse (AFIB1) - x1
ECG Test
1. Square Wave, 2 Hz (SQR2HZ) - x1
2. DC Pulse, 4 sec (PULSE) - x1
3. Triangle Wave, 2 Hz (TRI2HZ) - x1
4. Sinewave, 10 Hz (SIN10) - x1
5. Sinewave, 40 Hz (SIN40) - x1
6. Sinewave, 60 Hz (SIN60) - x1
7. Sinewave, 100 Hz (SIN100) - x1
8. ECG 60 BPM (ECG60) - x1
9. ECG 30 BPM (ECG30) - x1
10. ECG 120 BPM (ECG120) - x1
11. ECG 180 BPM (ECG180) - x1
12. ECG 240 BPM (ECG240) - x1
Pacer Pulse Test 1
Target Current: 55 mA
Tolerance On Current: +/-10%
Target Pulse Rate: 40 PPM
Tolerance On Rate: +/-10%
Measuring Method: Peak Current
Pacer Pulse Test 2
Target Current: 55 mA
Tolerance On Current: +/-10%
Target Pulse Rate: 60 PPM
Tolerance On Rate: +/-10%
Measuring Method: Peak Current
Pacer Pulse Test 3
Target Current: 55 mA
Tolerance On Current: +/-10%
Target Pulse Rate: 100 PPM
Tolerance On Rate: +/-10%
Measuring Method: Peak Current
Pacer Pulse Test 4
Target Current: 55 mA
Tolerance On Current: +/-10%
Target Pulse Rate: 140 PPM
Tolerance On Rate: +/-10%
Measuring Method: Peak Current
Pacer Pulse Test 5
Target Current: 55 mA
Tolerance On Current: +/-10%
Target Pulse Rate: 170 PPM
Tolerance On Rate: +/-10%
Measuring Method: Peak Current
Pacer Pulse Test 6
Target Current: 100 mA
Tolerance On Current: +/-10%
Target Pulse Rate: 80 PPM
Tolerance On Rate: +/-10%
Target Pulse Width: 20 msec
Tolerance On Width: +/-10%
Measuring Method: Peak Current
Pacer Pulse Test 7
Target Current: 130 mA
Tolerance On Current: +/-10%
Target Pulse Rate: 80 PPM
Below is an example of a Test Report document created by Phase3pc. Phase 3 itself has the capability of generating a
simplified report document containing similar information, which can be transmitted via the Serial I/O Port to a serial
printer or personal computer. Refer to Section 4.9.2.2 for instructions on how to output a Test Report via the Serial I/O
Port of Phase 3.
The results in the Report shown below were obtained from an AutoSequence performed on a Medtronic (nee Physio-
Control) LifePak model 9P. The Report incorporates results from defibrillator energy, charge time, and cardioversion
tests, as well as pacemaker pulse and refractory tests.
********************************************************************************
1.1 Energy
Pre-test Discharges: 5
********************************************************************************
TEST TEST PULSE TARGET AMP PULSE TARGET WIDTH PULSE TARGET RATE RESULT
METHOD AMPL. AMPL. TOL WIDTH WIDTH TOL RATE RATE TOL
-------------(mA)---(mA)--(%)---(msec)-(msec)--(%)---(PPM)--(PPM)-(%)---------
1 AVG. 18.7 20 10 19.93 --.-- -- 60.5 60 10 PASS
1 AVG. 47.1 50 10 19.92 --.-- -- 60.6 60 10 PASS
1 AVG. 94.2 100 10 19.92 --.-- -- 60.6 60 10 PASS
1 AVG. 47.2 50 10 19.93 --.-- -- 30.2 30 10 PASS
1 AVG. 47.1 50 10 19.92 --.-- -- 120.9 120 10 PASS
1 AVG. 47.2 50 10 19.93 --.-- -- 181.4 180 10 PASS