Finecare™ FIA Meter Plus (Model No.

: FS-113)

FinecareTM FIA Meter Plus

Operation Manual

Model No.: FS-113

Version: 10/03/2016 Guangzhou Wondfo Biotech Co., Ltd.


Finecare™ FIA Meter Plus (Model No.: FS-113)

TABLE OF CONTENTS

Section Content Page

Ⅰ Introduction and Principle 4

Ⅱ Composition of System 5

Ⅲ Specifications and Operation Conditions 8

Ⅳ Initial Installation Procedure 9

Ⅴ Operation Instruction 10

Ⅵ Instruction for connection with LIS system and PC 21

Ⅶ Quality Control 31

Ⅷ Warning, Precautions and Limitations 33

Ⅸ Service, Maintenance and Disposal 35

Ⅹ Trouble Shooting 37

Ⅺ Contact Information 38

Version: 10/03/2016 Guangzhou Wondfo Biotech Co., Ltd.

-1-
Finecare™ FIA Meter Plus (Model No.: FS-113)

Index of Symbols

Read all contents of this Manual carefully prior to use.


If the instrument is used in a manner not specified by the manufacturer, the protection provided by the
instrument may be impaired.

The following symbols are used on the Finecare™ FIA Meter Plus, related components and
accessories, labels or in the text of this user manual:

Biohazard

Caution

Consult Operator’s Manual

WEEE (Waste Electrical and Electronic Equipment)

Keep Dry

Operation Temperature Limitation

Storage and Transportation Temperature Limitation

Keep away from Sunlight

For in vitro Diagnostic Use only

Manufacturing Date

Serial Number

Version: 10/03/2016 Guangzhou Wondfo Biotech Co., Ltd.

-2-
Manufacturer

Authorized Representative

Direct current

ON (Power)

○ Off (Power)

-3-
Finecare™ FIA Meter Plus (Model No.: FS-113)

Section Ⅰ Introduction and Principle

This manual contains the instructions for the operation of the Finecare™ FIA Meter Plus and for the
sample testing procedures. For specific tests’ procedure please refer to the Instruction for Use of specific
products.

Finecare™ FIA Meter Plus manufactured by Guangzhou Wondfo Biotech Co., Ltd., is a portable
instrument for fluorescence detection to quantify concentration of various kind of analytes in human
blood or urine. The Test Device and instrument are for in vitro diagnostic use only. Finecare™ FIA
Meter Plus is intended for professional use.

Finecare™ FIA Meter Plus uses an LED as the Excitation Light Source. The emitted light from the
fluorescence dye is collected and converted into an electrical signal. The signals are closely related to
the amount of fluoresceing dye molecules present on the spot under examination.

After a buffer-mixed sample is applied to the Test Device, the Test Device is inserted into Finecare™
FIA Meter Plus and the concentration of the analyte is calculated by a pre-programmed calibration
process. Finecare™ FIA Meter Plus can only accept Test Devices that are designed specifically for use
with this instrument.

Finecare™ FIA Meter Plus is equipped with a built-in Test Device Holder and does not require an
external holder to place the Test Device in. The Test Device Holder will appear through the opening on
the front surface of the instrument.

The Power ON/OFF Switch, located on the left of the instrument, which powers the Finecare™ FIA
Meter Plus. The instrument has a built-in printer.

Figure 1 Working Principle

Version: 10/03/2016 Guangzhou Wondfo Biotech Co., Ltd.

-4-
Finecare™ FIA Meter Plus (Model No.: FS-113)

Section Ⅱ Composition of System

Unpacking

After opening the package box, compare the contents with the list below.

If any item is missing, please contact your sales distributor or Guangzhou Wondfo Biotech Co., Ltd.
Detailed contact information is mentioned in Section Ⅹ Ⅰ.

Contents

1. Standard contents:
No. Name Quantity

1 Finecare™ FIA Meter Plus 1

Power adapter for instrument (including the power


2 1
cord)

3 Quality Control Card 1

4 Ethernet cable 1

5 Printer paper (roll) 1

6 Operation manual 1

7 Quick operation instruction for use 1

8 Warranty Card 1

9 Packing List 1

Version: 10/03/2016 Guangzhou Wondfo Biotech Co., Ltd.

-5-
Fig. 2.1 Structure Chart of Finecare™ FIA Meter Plus- Front View

Printer

Touch Screen

Power and Communication


interface

Test Device Holder

ID Chip Port

Fig. 2.2 Finecare™ FIA Meter Plus Features - Left View


Power switch, Ethernet Port, USB Interface, Serial Interface, Power interface, Selector switch for
operating mode, which are used in Finecare™ FIA Meter Plus’s startup, shutdown, select the operating
mode and connection with PC and Ethernet.

Power ON/OFF

External Power Interface Selector for operation


mode: select operation
shipped in instrument
USB Interface or PC mode.
Serial Port

Ethernet Interface

-6-
Fig. 2.3 Power Adapter

Consumables of Reagent Kit

Fig. 2.4 Test Device Features


(Remark: You shall buy the Test Devices separately)

(B) (A)

Note: This accessory is Test regent’s consumable

Fig. 2.5 Test Device ID Chip

(The ID Chip contains the information of test item and lot number)
Note: This accessory is Test regent’s consumable

Fig. 2.6 Quality Control Card

-7-
Section Ⅲ Specifications and Operation Conditions

1. Basic Specification
Model No. FS-113
Excitation light source LED
Excitation spectrum Center wavelength λ0 = 470nm
Absorption spectrum Center wavelength λ1 = 525nm
Test Specimen Whole blood, serum / plasma and urine
Software system English and Chinese input method, customized intelligent
management module
Interface RS232, USB, Ethernet network
Built-in Printer Thermal
Size 270*238*146mm
Weight About 2800g

Operating Environment
Temperature 10℃~30℃(50℉-86℉)
Relative Humidity 10%~80%
Atmospheric pressure 86kPa~106kPa
Power Input: 100-240Vac, 50-60Hz, 1.4-0.7A
Adapter
Supply Output: 12V 5.0A 60W MAX
Power Input: 12V ,45W
Location Dry, clean, flat, horizontal surface away from direct sunlight,
mechanical vibration, and strong electromagnetic interferences.

Storage and Transportation Environment


Temperature -10℃~+50℃
Relative Humidity ≤85%

Performance
Repeatability CV ≤ 15%
Stability CV ≤ 8%

-8-
Section Ⅳ Initial Installation Procedure

1. Power Requirements
If you are uncertain of your power line, consult with your utility manager. The FinecareTM FIA Meter
Plus can only be used with the power adapter (including the power cord) provided. If the replacement
for Power Adaptor is required, please consult with your sales distributor or Guangzhou Wondfo Biotech
Co., Ltd.

2. Installation
Please be noted that Finecare™ FIA Meter Plus should only be used in 10℃~30℃ (The operational
temperature of Test Devices should refer to relevant Instruction for Use) and Relative Humidity of
10%~80% conditions (Detailed information is mentioned in Section Ⅲ).

①. Place Finecare™ FIA Meter Plus on a dry, clean, flat, horizontal surface, with the front surface at
least 10 cm inside of the table edge.
②. Connect the Power Cord with Finecare™ FIA Meter Plus to the wall power outlet.
Note: The meter shall be connected to a mains socket outlet with a protective earthing connection.
The mains outlet shall be located in a user accessible position for easy unplugging in emergency.
The mains plug of the power cord set is a disconnect device of the instrument. In order to completely
disconnect the instrument from the supply mains, the mains plug of the power cord set should be
unplugged from the mains outlet.
③. Turn on the Power ON/OFF Switch on the left of Finecare™ FIA Meter Plus to carry out power
startup and shutdown.( See Fig. 4. 1)
Fig. 4.1

Power ON/OFF
Selector for operation mode
External Power Interface

*Caution:
Keep the power cord for the Power Adapter in suitable place. Power cord without good maintenance
could render hazardous to connect/disconnect the Power Adapter.

-9-
Finecare™ FIA Meter Plus (Model No.: FS-113)

Section Ⅴ Operation Instruction

FinecareTM FIA Meter Plus provides two test modes, including Standard Test and Quick Test. Under
the standard test mode, the instrument as a incubation place reaction of test item to provide required
standard countdown. Under the quick test mode, the test item shoud be completed incubation prior, the
user can directly conduct testing.

FinecareTM FIA Meter Plus can be operated through LCD screen with finger touch or touch pen, also
can be operated through with a mouse and keyboard.

V.1 Start
V.1.1 Turn on the Power ON/OFF on the left of Finecare™ FIA Meter Plus.
Fig 5.1

V.1.2 The instrument will load the test system.


V.1.3 Waiting for 1 minute, the instrument will show the initial screen, and slot pop-up the Test
Device holder automatically.
V.1.4 Quick Test interface will be showed by default when you turn on Finecare™ FIA Meter
Plus at the first time, after then boot screen displays the initial screen that the user once left off. (See
Fig. 5.2)
Fig 5.2 Quick Test interface

Version: 10/03/2016 Guangzhou Wondfo Biotech Co., Ltd.

- 10 -
A Function Buttons - Touch the button to achieve the corresponding function.
B Information Input - Input Serial Number, Sample number and Sample type.
C Results Display - Display test serial number, item name and test results.
D Quick Set - Reset the slot, slot out, send test result to LIS system, print and print automatically.

【Temperature Display】Display the temperature, the bald thermometer will turn to red by default
when the temperature is above 30℃, the default temperature can be set in the system setting.
【Test】To carry out testing of test item.
【History】Inquire historical record.
【Account】Account the test data.
【QC】FinecareTM FIA Meter Plus selftesting of quality.
【Set Up】Set instrument parameters.
【Test Mode】Choose test mode.
【Start Test】 Start to test.

*Caution:
If the initial interface doesn’t show after more than 3 minutes please turn off the Power ON/OFF
Switch and restart the instrument. If failure continues, please contact your sales distributor or
Guangzhou Wondfo Biotech Co., Ltd.

V.2 Test Operation Example


Please insert the ID chip of corrsponding test item before clicking “Test”, Under the initial interface of
Finecare™ FIA Meter Plus, the user can operate instrument by clicking the corresponding function
button for invoking different procedures.

Finecare™ FIA Meter Plus enter the initial test interface after booting, you can freely choose
“Standard Test” mode or “Quick Test” mode according to different needs. (See Fig. 5.3)
Fig. 5.3

- 11 -
*Caution:
1. It is recommended to insert ID chip after successful boot!
2. Please make sure the “selector switch for operating mode” on the left side of the instrument to
be appropriated for "MAC”, when you choose the test mode of Finecare™ FIA Meter Plus’ system.

2.1 Choose Test mode (Standard Test or Quick Test).


2.2 Input basic patient medical information, including serial number, sample number and sample type. (The
number will be automatically generated by system if you don’t input it.) .Click” Detailed information”
to input more detailed information.
2.3 Add sample to the buffer, mix well.
*Caution:
Please carefully read relevant Instruction for Use of specific Test Device before adding sample to test,
and operate the test according to requirements.
2.4 Prepare a Test Device and apply sample-buffer mixture onto the sample well of Test Device (Refer
to the Instruction for Use of specific Test Device).
2.4.1 If choose Quick Test at step 2.1, insert the Test Device that is completely reacted into the Test
Device Holder of Finecare™ FIA Meter Plus, and click “StartTest”.
2.4.2 If choose Standard Test at step 2.1, immediately insert the Test Device that is into the Test Device
Holder of Finecare™ FIA Meter Plus, and click “StartTest”.
2.5 Finecare™ FIA Meter Plus will scan the Test Device automatically, the screen shows “Testing" and
testing progress. (See Fig. 5.4)
Fig. 5.4

- 12 -
*Caution:
If the lot number of the ID Chip does not match with the lot number of the Test Device when use the
specific Test Device at the first time, the instrument will show “insert the corresponding ID chip”.
Please check the lot number of ID Chip and Test Device if they are mismatched, please click "
Change reagent and test again " , Finecare™ FIA Meter Plus will pop-up Test Device holder and
show " Cancel test " on the screen. If the lot number of ID Chip and Test Device are matched, please
click "OK", Finecare™ FIA Meter PLUS will automatically recheck ID chip and Test Device.

2.6 The test result will be showed under the “Result Display” of the screen. (See Fig. 5.5)
Fig. 5.5

2.7 Click “Print” to print the test result.


*Caution:
The test result will be printed automatically after ticking off “Print automatically”.

2.8 Take out the Test Device. Insert the next Test Device that is completely reacted for the next testing.
2.9 Click "Read ID card ", Finecare™ FIA Meter Plus will read and keep the ID card information.
2.10 When Finecare™ FIA Meter Plus is not used, click “Reset the slot”, then turn off the power
switch and unplug the power source of the instrument.
2.11 Click “Reset the slot” after completing the test.

V.3 History Operation Example


“History” function provides the user with a function of conditional filter, view the history of
test records. There are Query, Delete and Print buttons to carry out operation of query, delete
and print test results under the interface of “History”.

- 13 -
3.1 Click “History” on the top of the instrument to enter the query interface, the test results of
intraday will be listed. (See Fig. 5.6)
Fig. 5.6

3.2 Click “Query” under the “History” interface, then input the query conditions, Finecare™ FIA
Meter Plus will filter out the test results that match the query conditions to conduct fuzzy query.
(See Fig. 5.7)
Fig. 5.7

3.3 Tick all or specific data record, click “Delete” under the “History” interface, then click “Yes” to
delete the selected records. (See Fig. 5.8)
Fig. 5.8

- 14 -
3.4 Tick all or specific data record, click “Print” under the “History” interface to print a selected
records. When the selected data is greater than 10, only 10 data can be printed. (See Fig. 5.9)
Fig. 5.9

3.5 In the case of the LIS system was connected, select all or specific data record, click “LIS” under
the “History” interface to send corresponding test result to LIS system. (See Fig. 5.10)
Fig. 5.10

- 15 -
V.4 Account Operation Example
Click “Account” on the top of the instrument to enter the statistics interface. “Account”
function provides the user with “Testing Account” and “Workload Account” function.
Click “Testing Account” and “Workload Account” to show statistical results according
to statistic conditions.

4.1 Testing Account: Count the number according to items or samples. (See Fig. 5.11)
Fig. 5.11

4.2 Workload Account: Workload can be accounted according to submitting doctor, submitting
division and inspecting doctor. (See Fig. 5.12)
Fig. 5.12

- 16 -
V.5 Set Up Operation Example
“Set Up” function provides the user with setting function for the machine function parameters,
there are two mode for users, including general users and R & D users, they can enter the appropriate
settings mode by using different password.

Click “Set Up” to see the Software version,input the password in the "Authorization code "
field , then click "Yes" to enter the setting interface.

*Caution:
The users who no password permissions do not have permission about parameter settings.

- 17 -
5.1 General user can only enter the "System Setting" interface to set LIS connection, bald temperature
parameters.

5.2 R & D users interface, in addition to the "System Setting" function, also increase the "Parameter
Setting" function.

5.2.1 ”Item Parameters” under the "Parameter Setting" interface, the users can add and
modify test items and the reference range.

5.2.2 Click “Add” to add items.

- 18 -
5.2.3 Click “Edit” to modify the existing items’ parameters.

5.2.4 Click “Reset” to restore item parameters to the start parameter, then click ”OK” to
complete the reset:

5.2.5 ”Device Parameters” under the "Parameter Setting" interface, the users can set “AD
magnification times”, “Total length of the slot in running”, “Barcode recognition deadline” and

- 19 -
“Sampling starting point”.

AD magnification times: Magnification of fluorescence signal.


Total length of the slot in running: The maximum distance of running forward when motor start
with photoelectric switch.
Barcode recognition deadline: The timeout period of barcode recognition.
Sampling starting point: The starting position of fluorescence data collection.

*Caution:
Please note that this setting must be executed by authorized technical service personnel, unauthorized
users shall not operate.

- 20 -
SectionⅥ Instruction for connection with LIS System and PC

Ⅵintroduce the operational


This section Connection withofLIS
instruction system
FinecareTM and PC
FIA Meter Plus connect with LIS and
PC.


Transfer Protocol Instruction: Connection with LIS system and PC
FinecareTM FIA Meter Plus sends data to outside by UDP, click “LIS” when the test is completed
every time, the current test data will be sent out.

Preset initial IP address of FinecareTM FIA Meter Plus: 192.168.0.3


Preset initial subnet mask of FinecareTM FIA Meter Plus: 255.255.255.0
Preset initial default gateway of FinecareTM FIA Meter Plus: 192.168.0.1
Preset initial receiving IP of PC: 192.168.0.88
Subnet mask of PC: 255.255.255.0
Default gateway of PC: 192.168.0.1
PC port: 20035
Communication mode: UDP

PC Data receiving format:


For CRP: FF&test serial number&CRP test result&hs CRP test result&test time&item name&EE
Example:FF&1&<0.5&0.93&2012-07-30 13:59:33& Whole Range CRP&EE

TM
Ⅵ.1 Methods of Finecare FIA Meter Plus connect with LIS \ PC.
Method 1: When FinecareTM FIA Meter Plus connects to exclusive computer (see Fig. 6.1)

1) Use a standard network cable that can make normal PC access the Internet, one end of the network
cable connects FinecareTM FIA Meter Plus’s Ethernet port, the other end connects the PC 's Ethernet
port.

Fig. 6.1

Ethernet Port

Ethernet Port
Network cable

2) Set the IP address of PC according to protocol, click “Start” -> “Routine”-> “Attachment”->
“Communication”-> “Network”; double click “local connection” and enter to “Local Area Connection
Status”.( See Fig. 6.2)

- 21 -
Fig. 6.2

3) Click “Properties” and enter to “Local Area Connection Properties”. (See Fig. 6.3)
Fig. 6.3

4) Double click “Internet Protocol Version 4 (TCP/IPv4)”. Enter to “Internet Protocol (TCP/IP)
Properties”. Set IP address as 192.168.0.88,click “OK” after finishing. (See Fig. 6.4)

Fig. 6.4

- 22 -
Method 2: Multiple Network Information Card
Install one more NIC (Network Information Card) and corresponding driver in computer to work
properly. Then use a standard network cable that can make normal PC access the Internet, one end of the
network cable connects FinecareTM FIA Meter Plus’s Ethernet port, the other end connects PC 's
Ethernet port of new Installed NIC. Alter the new NIC IP address in computer, the method as same as
Method 1. (See Fig. 6.5)

Fig. 6.5

Router, Switchboard
Ethernet Port
Ethernet Port1:IP1

Network cable

Ethernet Port
Ethernet Port 2:IP2

Network cable

Method 3: LAN (Local Area Network)


Use a standard network cable that can make normal PC access the Internet, one end of the network cable
connects FinecareTM FIA Meter Plus’s Ethernet port, the other end connects Laboratory LAN(such as
router and switchboard etc. ). (See Fig. 6.6)

Fig. 6.6

- 23 -
Router, Switchboard
Ethernet Port Ethernet Port

Network cable

Ethernet Port Ethernet Port

Network cable

Click “Start” -> “Routine”-> “Attachment”-> “Communication”-> “Network”; double click “local
connection” and enter to “Local Area Connection Status”, click “Properties” and enter to “Local Area
Connection Properties”, double click “Internet Protocol Version 4 (TCP/IPv4)” and enter to “Internet
Protocol Version 4 (TCP/IPv4) Properties”.
Don’t change primary IP address of PC. Example:Primary IP address:192.168.1.6.
1) Cick “Advanced”,enter to “Advanced TCP/IP settings” (See Fig. 6.7)
Fig. 6.7

- 24 -
2)Click “Add” under “IP address”, enter to“TCP/IP address”. (See Fig. 6.8)
Fig. 6.8

3)Add IP address:192.168.0.88,Subnet mask:255.255.255.0 (See Fig. 6.9)


Fig. 6.9

4)Then click “Add” to return to “Advanced TCP/IP settings” (See Fig. 6.10)
Fig. 6.10

- 25 -
5)Click “Add” under “Default gateway”,enter to “TCP/IP Gateway Address”. (See Fig. 6.11)
Fig. 6.11

6)Input Gateway:192.168.0.1, click “Add” after finishing, return to “Advanced TCP/IP settings”. (See
Fig. 6.12)
Fig. 6.12

- 26 -
7) Click “OK”, return to IP address setting menu, then click “OK” to exit IP setting, run the software of
ReceiveFromCRPTemp.exe after setting successful.
Note:FinecareTM FIA Meter Plus will send the current test results to computer automatically after
test(in the case of LIS system is connected). Click “Record”, selected historical record can also be sent
by clicking “Send”.
Run the software of ReceiveFromCRPTemp.exe, FinecareTM FIA Meter Plus will send the testing
results as follows. (In an example of CRP, see Fig. 6.13)
Fig. 6.13

Ⅵ.1.2 Instrument terminal IP setup steps


1) Enter "System Settings" interface, click "Settings", input the corresponding IP address with computer :

- 27 -
"192.168.0.88" .

2) Choose “Ethernet configuration” to enter Network Settings interface.

3) Click “Turn on Ethernet”, pop-up setting box where you can set the IP address of instrument terminal.

- 28 -
4) ) Set the network port IP address of instrument for LIS transmission and ensure consistency with IP
address of the computer terminal.

5) Turn on LIS data receiving software of the computer, select "network port" and click “Start”.
Note: LIS serial transmission function has been canceled, not verify here.

6) The instrument can test normally, the test results will be sent automatically to the PC reception
software. If you did not input the details of test item, such as name and sample types, computer terminal
will be processed as blank at this option.

- 29 -
7) Or check specific items and click “LIS” to upload the selected results to PC in the current result
display interface.

- 30 -
Section Ⅶ Quality Control

When the FinecareTM FIA Meter Plus is used for the first time or each time, user can use the Quality
Control Card to check the internal systems of the meter (including Optical System, Transmission System,
Circuit System), to ensure instrument is functioning properly before testing through quality inspection of
instrument. Please follow the operation steps as below:

Step 1 Click “QC” on the top of the instrument to enter the quality control interface, it is used to
make sure FinecareTM FIA Meter Plus to work properly.

Step2 Push the Quality Control Card all the way into the Test Device Holder until the card comes
to a stop (Please click "Out" if Text device Holder did not come out.). Use only the push of
your thumb or the index finger. Don’t use excessive force on the Test Device, or it will
result in a mechanical failure.

Step 3 Click “Start” to start quality inspection. (See Fig. 7.1)


Fig. 7.1

*Caution:
When Test Device Holder is not come out, click “Out”, the Test Device Holder will move out of the
main body of the instrument.)

Step4 Click “Print” after completing the quality inspection to print the quality inspection result.
(See Fig. 7.2)

Fig. 7.2

- 31 -
Step 5 C lick “Reset the solt” if you do not need to use.

- 32 -
Section Ⅷ Warning, Precautions and Limitations

Caution in Handling the Test Device

● The test sample should be applied to the Sample Well B, as shown in Fig. 2.8 of Section Ⅱ.

● When inserting the Test Device into the Test Device Holder, orient the Test Device in such a way
that the Sample Well B should be toward Finecare™ FIA Meter Plus. This direction is also marked
with an arrow on the Test Device.

● Push the Test Device all the way into the Test Device Holder until the Test Device comes to a stop.
Use only the push of your thumb or the index finger. Don’t use excessive force on the Test Device,
or it will result in a mechanical failure.

● The used Test Device should be treated as a potential bio-hazard and should be disposed of
according to the standard procedures and relevant regulations observed by microbiological hazard
materials. We recommended high temperature autoclaving by incineration.

● Gloves, masks and other protective measures should be used when handling potential infectious
materials.

Caution in Use with the Finecare™ FIA Meter Plus

● Do not insert anything other than a Test Device provided by the manufacturer into the Test Device
Holder.

● Do not spill any liquid on Finecare™ FIA Meter Plus. This may disable the system.

● Do not drop or crash Finecare™ FIA Meter Plus.

● Do not subject Finecare™ FIA Meter Plus to mechanical shocks heavier than light taps with hands.

● Do not dismantle Finecare™ FIA Meter Plus without written authorization from Guangzhou
Wondfo Biotech Co., Ltd. or its representative.

● Do not place heavy objects on Finecare™ FIA Meter Plus. This may damage the optical alignment
to result in degraded performance or in mechanical damage.

● The mains outlet for Finecare™ FIA Meter Plus shall be located in a user accessible position for
easy unplugging in emergency.

● Please make sure power outlet properly grounded. Improper grounding may cause electrical shock

- 33 -
and systematic damage.

● Use only the Power Adapter (including the power cord) provided with Finecare™ FIA Meter Plus.

● Use Finecare™ FIA Meter Plus only in environment as specified in Section Ⅲ.

● Finecare™ FIA Meter Plus malfunction due to maintenance, processing or transporting damage,
please contact customer service representative.

● Use the test result as a pre-screening guide. The result should be interpreted by trained medical
personnel only.

The FinecareTM FIA Meter Plus shall be operated by professional person.

● If the Finecare™ FIA Meter Plus is used in a manner not specified by the manufacturer, the
protection provided by the instrument may be impaired.

*Cautions should be taken to use Finecare™ FIA Meter Plus in Settings meeting the
requirements specified in Section Ⅲ of this manual and in observance of the warning in Section
Ⅶ. Otherwise, the built-in safety features of Finecare™ FIA Meter Plus may be compromised to
present severe electrical, mechanical or biological hazards to the user.

- 34 -
Finecare™ FIA Meter Plus (Model No.: FS-113)

Section Ⅸ Service, Maintenance and Disposal

If any service or maintenance were to be required, FinecareTM FIA Meter Plus should be sent to
Guangzhou Wondfo Biotech Co., Ltd. No Special maintenance other than paper replacement and
periodic cleaning is required for the instrument. Occasional cleansing of the exterior with a mild-dry
cloth is sufficient to assure the instrument operation. The instrument shall be separated from mains
supply prior to cleaning.

External cleaning
Use a damp cloth with 0.5% bleach, 70% isopropyl alcohol or 70% ethanol to clean the external surface
of instrument. To prevent damages of screen or other surface device, strong bleach (bleach concentration
more than 0.5%), oxidizing substances and solvents are prohibited.

Decontamination
If the instrument needs maintenance or replacement after clinical use, decontamination and disinfection
before repacking and transporting should be done. Use disinfectant (e.g. bleach concentration lower than
0.1%) and a cloth to scrub external surface of instrument completely. To prevent damages of inner
devices, spray-washing or clean any internal parts and inner surface with disinfectant is prohibited.

*Caution:
Instrument does not contain any serviceable components by operator. To avoid electric shock,
regular maintenance must be executed by authorized technical service personnel.

Maintenance
In addition to regularly cleaning or replace paper, FinecareTM FIA Meter Plus generally does not
require special maintenance, with a soft, dry cloth to wipe the external surface of the instrument to
ensure the normal operation of FinecareTM FIA Meter Plus.

Power Supply
Use only the AC/DC power supply provided with the instrument. In case the power adapter needs
replacement, you should contact Guangzhou Wondfo Biotech Co., Ltd. or its designated representative.
Guangzhou Wondfo Biotech Co., Ltd. CE certified Class 2 power supplies, rated at +12V 5A.

Return Procedure
If FinecareTM FIA Meter Plus happened to malfunction, please contact Guangzhou Wondfo Biotech
Co., Ltd. or the local distributor at first. If it were determined that the unit will be returned to the
manufacturer, a return authorization number will be issued, then Guangzhou Wondfo Biotech Co., Ltd.
will send a FinecareTM FIA Meter Plus for replacing. The user is expected to utilize the packaging
supplies accompanying the replacement to ship the malfunctioning unit in. Verify the return
authorization number on the package and send the unit to Guangzhou Wondfo Biotech Co., Ltd. upon
receiving the replacement FinecareTM FIA Meter Plus.

Version: 10/03/2016 Guangzhou Wondfo Biotech Co., Ltd.

- 35 -
Transporting and storing
The original shipping container should be used to ship or transport FinecareTM FIA Meter Plus. The
original shipping container is also recommended for storing FinecareTM FIA Meter Plus over an
extended period. When transporting or storing FinecareTM FIA Meter Plus, keep it dry in upright
position and protect it from sun, mechanical shocks, transportation requirements according to an order
under the contract.
FinecareTM FIA Meter Plus should be stored in a well-ventilated room that has storage temperature of
-10℃~50℃, relative humidity is less than 85% and no corrosive gas.

Disposal
Any spent Test Device should be disposed according to local ordinances regarding the disposal of
bio-hazardous materials.

WEEE (Waste Electrical and Electronic Instrument)

This symbol for the marking of electrical and electronics devices is according to the Directive
2002/96/EC. The FinecareTM FIA Meter Plus, accessories and the packaging have to be disposed of
waste correctly at the end of usage. Please follow Local Ordinances or Regulations for disposal.

- 36 -
Section Ⅹ Trouble Shooting

Phenomena Probable Cause Recommended Measures


Power Failure Check the Power Plug.
Poor Connection between Pull out the Cord and Re-connect
TM
Finecare FIA Meter Plus the Power Adapter and the firmly.
does not response Cord.
Main switch is off. Turn on the main switch.
Power Adapter is broken. Call Customer Service.
Excessive Computation Wait till the computation is
Run finished but no result Load finished.
returned Computational Abnormality Turn off the power and start fresh.
Mechanical Failure Call Customer Service.
Test Device Holder does not (Buzzing Noise)
come out Software Abnormality. Turn off the power and start fresh.
Clicking noise when the Test Normal Mechanical Setting No User Action required.
Device Holder moves
Electrostatic Influence Discharge any static electricity,
Display Screen does not reboot the system
function properly Electric Circuit Failure. Call Customer Service.
Message: Reset reagent and Test device or ID chip was Insert the test device or ID chip.
test again! not inserted.
ID chip is not inserted or the Insert the right ID chip.
Message: Insert the
lot of Test Device is
corresponding ID chip
incompatible.
Message: “Optical System Internal hardware failure Please Contact Wondfo Customer
Fault”, “Transmission System Service Center.
Fault”, and/or “Circuit System
Fault”

- 37 -
Finecare™ FIA Meter Plus (Model No.: FS-113)

Section Ⅺ Contact Information

Guangzhou Wondfo Biotech Co., Ltd. ’s expressed and implied warranties are conditioned upon full
observance of manufacturer’s published direction with respect to the use of Guangzhou Wondfo
Biotech Co., Ltd. ’s products. Under no circumstance whatsoever shall Guangzhou Wondfo Biotech
Co., Ltd. be held liable for any indirect or consequential damages.

For technical assistance, call or send e-mail to us:

Guangzhou Wondfo Biotech Co., Ltd.


No. 8 Lizhishan Road, Science City,
Luogang District,
510663 Guangzhou, P.R.China
Manufacturer TEL: 0086-20-3229-6083 FAX: 0086-20-3229-6063
Website: www.wondfo.com.cn E-mail: sales@wondfo.com.cn

Qarad b.v.b.a.
Cipalstraat 3
B-2440 Geel, Belgium

Version: 10/03/2016 Guangzhou Wondfo Biotech Co., Ltd.

- 38 -