FE1E04200C74E0479F
FE1E04200C74E0479F
FE1E04200C74E0479F
UserManual
Page1of86
Asymbolsglossarycanbefoundatquidel.com/glossary.
TableofContents
TableofContents............................................................................................................................... .............2
GeneralInformation............................................................................................................................... .............5
QuidelContactInformation............................................................................................................................. 5
MainMenuStructure............................................................................................................................... .......6
LabelsandSymbols............................................................................................................................... ..........7
Warnings,PrecautionsandLimitations...........................................................................................................9
Introduction............................................................................................................................... ........................10
HowtoUseThisGuide............................................................................................................................... ...10
Description............................................................................................................................... .....................10
WhatistheQuidelTriageMeterPro?.......................................................................................................10
Unpacking............................................................................................................................... ...................11
Installation............................................................................................................................... ..........................15
UnpackMeter............................................................................................................................... .................15
PowerMeterOn............................................................................................................................... .........15
InstallPaper............................................................................................................................... ................16
InsertSupervisorCodeChipModule.........................................................................................................17
ToChangeMeterSettings......................................................................................................................... 17
ChangeClock............................................................................................................................... ..............19
SetUserID............................................................................................................................... ..................20
ChangeReferenceRanges/ThresholdsforTests:......................................................................................20
ChangeTestSettings............................................................................................................................... ..20
ChangeQCParameters.............................................................................................................................. 21
ChangeBypassSettings............................................................................................................................. 21
SetParameters............................................................................................................................... ...............23
Purpose............................................................................................................................... .......................23
ListofProgrammableParameters.............................................................................................................23
SupervisorAccess............................................................................................................................... .......23
HowtoSetParameters:BasicInstructions...................................................................................................24
MeterSettings............................................................................................................................... ............25
IDSettings............................................................................................................................... ...................25
Page2of86
InstallCodeChip............................................................................................................................... .............37
Purpose............................................................................................................................... .......................37
TypesofCODECHIPmodules....................................................................................................................37
InstallationofReagent,QCSampleorQCDeviceCODECHIPModules...................................................37
Operation............................................................................................................................... ...........................39
AccesstotheQuidelTriageMeterPro..........................................................................................................39
RunTest............................................................................................................................... ..........................39
Purpose............................................................................................................................... .......................39
Options............................................................................................................................... .......................39
RecallResults............................................................................................................................... ..................49
Purpose............................................................................................................................... .......................49
Function............................................................................................................................... ......................49
LastRecord............................................................................................................................... .................49
PatientResults............................................................................................................................... ............50
PatientID............................................................................................................................... ....................50
TestDeviceLotNumber(L/N)...................................................................................................................51
UserID............................................................................................................................... ........................52
FromDate............................................................................................................................... ...................52
TillDate............................................................................................................................... .......................53
QCResults............................................................................................................................... ...................54
QCSampleLotNumber(L/N)....................................................................................................................55
DeviceLotNumber(L/N)........................................................................................................................... 55
UserID............................................................................................................................... ........................56
FromDate............................................................................................................................... ...................57
TillDate............................................................................................................................... .......................57
ReagentLots–QC............................................................................................................................... ......58
ArchiveLastTest............................................................................................................................... .........58
Misc.TestResults............................................................................................................................... .......59
Misc.TestID............................................................................................................................... ...............59
TestDeviceLotNumber(L/N)...................................................................................................................60
UserID............................................................................................................................... ........................61
FromDate............................................................................................................................... ...................61
TillDate............................................................................................................................... .......................62
PrintAllResults............................................................................................................................... ...........63
PatientResults............................................................................................................................... ............63
QCSampleResults............................................................................................................................... ......63
Page3of86
QCDeviceResults............................................................................................................................... .......64
Misc.TestResults............................................................................................................................... .......64
UploadResultstoLIS............................................................................................................................... ..64
DeleteResults............................................................................................................................... .................64
Purpose............................................................................................................................... .......................64
TheQuidelTriageMeterProMemory.......................................................................................................65
DeletePatientResult............................................................................................................................... ..65
DisplayBacklight............................................................................................................................... .............66
AlphaNumericMode............................................................................................................................... .....67
ServiceandMaintenanceProcedures...............................................................................................................68
PaperReplacement............................................................................................................................... ........68
ChangingPaperBetweenFunctions..........................................................................................................68
ChangingPaperWhilePrinting..................................................................................................................69
Cleaning............................................................................................................................... ..........................70
CheckBatteryLevel............................................................................................................................... ........70
BatteryReplacement............................................................................................................................... ..70
TotalQualityAssurance............................................................................................................................... ..71
Introduction............................................................................................................................... ................71
QualityAssuranceFunctions.....................................................................................................................72
InternalDeviceQualityAssurance............................................................................................................72
MeterQualityAssurance........................................................................................................................... 73
SoftwareQualityAssurance...................................................................................................................... 73
QualityControlSamples............................................................................................................................ 74
NotesRegardingSoftwareControls..........................................................................................................74
Appendices............................................................................................................................... .........................76
AppendixASpecifications............................................................................................................................. 76
MeterSymptoms............................................................................................................................... ........77
AppendixBTroubleshooting......................................................................................................................... 78
MeterMessageorSymptom.....................................................................................................................78
AppendixCReturnandDisposalProcedures................................................................................................80
ReturnProcedure............................................................................................................................... .......80
DisposalProcedure............................................................................................................................... .....80
AppendixDGlossary............................................................................................................................... .......80
AppendixESampleLogSheets...................................................................................................................... 82
Page4of86
GeneralInformation
QuidelContactInformation
ContactQuidelTechnicalSupportfrom7:00a.m.to5:00p.m.PacificTime
Tel: 800.874.1517(intheU.S.);
858.552.1100(outsidetheU.S.)
Fax: 858.552.7905
Email: technicalsupport@quidel.comorcontactyourlocaldistributor
Website: quidel.com
IfoutsidetheU.S.contactlocaldistributororoneoftheTechnicalSupportcenterslistedbelow.Youmay
alsocontactusatquidel.com.
Page5of86
MainMenuStructure
*SupervisorCODECHIPOnly
x NewUserID x QCDeviceFreq.
Misc. Upload
Last QC Patient PrintAll x UpdateUserID x QCSampleFreq.
Test Results
Record Results Samples Results x DeleteUserID x NumberofControls
Results toLIS
x UserIDList x NumQuantStdDevs
x QCSampleResults
x QCDeviceResults ID Display Comm
x ReagentLots–QC Settings Settings Settings
x ArchiveLastTest
x CharactersUserID x Language
x x BaudRate
CharactersPat.ID x PrintMode
x x Pat.ResultApproval
CharactersAux.ID x AutoPowerͲOff
x x LISEnable/Disable
Aux.IDEnable/Disable x DisplayContrast
x x AutoUpload
CharactersMisc.ID x PrinterContrast
Page6of86
LabelsandSymbols
Label/Symbol Description
Manufacturer
Invitrodiagnosticmedicaldevice
Consultinstructionsforuse
AuthorizedrepresentativeintheEuropeanCommunity
Temperaturelimitation
Humiditylimitation
Wasteelectricalandelectronicequipment(WEEE)
SerialNumber
CatalogNumber
Warning/Caution
Important
Page7of86
CEmark Batchcode Dateofmanufacture
UseͲbydate Intendeduse Upperlimitoftemperature
DonotreͲuse ChinaRoHS Patientnumber
Testdevice Control CalibrationVerification
Prescriptionuseonly Contents/contains Expectedvalues
Biologicalrisks Thaw Transferpipette
Mean Standarddeviation CODECHIPmodule
PrinterPaper Peelopenhere Addsamplehere
UseEDTAwholebloodorplasma
UseEDTAplasmasampleonly Useurinesampleonly sampleonly
Addsampleimmediatelyafter
openingfoilpouch
Page8of86
Warnings,PrecautionsandLimitations
TheQuidelTriageMeterPromustbeusedwithintheoperatingtemperaturerangerequiredbythe
specifictestdeviceproductbeingused.Refertothetestdevicepackageinsertfordetails.
OperatetheQuidelTriageMeterProonalevel,drysurfaceawayfromdirectsunlight.
TheQCDeviceislightͲsensitiveandshouldbestoredinitsblackopaquecasewhennotinuse.
DoNOTwriteorplacealabelanywhereontheQCdevice,asitmayinterferewiththedevice
functionality.
DonotmovetheQuidelTriageMeterProwhileatestisinprogress.
Ensureallsamplefluidshaveabsorbedintothetestdevicepriortorunningthetesttopreventinternal
contaminationofthemeter.
UseonlytheAC/DCpoweradapterprovidedwiththeQuidelTriageMeterPro.Anidentifyingtaghas
beenplacedonthecordforquickidentification.
DonotdropQuidelTriageMeterPro.
DonotplaceobjectsontheQuidelTriageMeterPro.
DonotimmersetheQuidelTriageMeterProinwateroranyliquids.
Page9of86
Introduction
HowtoUseThisGuide
Thismanualcontains:
InstructionsfortheoperationandmaintenanceoftheQuidelTriageMeterPro,and;
Basicinstructionsfortestingsamples
NOTE:Torunspecifictests,suchastheQuidelTriageCardiacPanel,youalsowillneeddetailed
informationcontainedintheproductͲspecificPackageInsert,includedinthetestkit.
TheQuidelTriageMeterProsoftwarehasbeendesignedtoprovidetheuserstepbystepinstructions.The
meterdisplaystestresultsandmenuoptionsintheupperhalfofthemeterscreen.Insmallerlettersatthe
bottomofeachscreen,themeterdisplaysinstructionspertainingtothetaskbeingperformed.
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODECHIP
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
TheQuidelTriageMeterProUserManualisdividedintosixsections.
Section1– GeneralInformation:Contactinformation,MenuStructurefortheQuidelTriageMeterPro,
Symbols,WarningsandPrecautions
Section2–Introduction:ProvidesanoverviewoftheQuidelTriageMeterPro.
Section3– Installation:Guidestheuserthroughfirsttimesetup.
Section4– Operation:Guidestheuserthroughrunningtests,recallingresultsanddeletingresults.
Section5– ServiceandMaintenanceProcedures:PaperReplacement,Cleaning,Batteries,TotalQuality
Assurance,andWarranty.
Section6– Appendix:Specifications,Troubleshooting,ReturnandDisposal,Glossary,SampleLogSheets
Description
WhatistheQuidelTriageMeterPro?
TheQuidelTriageMeterProisaportablefluorescenceinstrumentusedtomeasuretheresultsoftests
manufacturedbyQuidelCardiovascularInc.TheQuidelTriageMeterProcanbeusedinalaboratoryorina
pointͲofͲcaresetting.
TheQuidelTriageMeterProusesalaserasalightsource.Lightfromthelaserhitsatestdevicethathas
beeninsertedinthemeter.Thiscausesthefluorescentdyeinthetestdevicetogiveoffenergy.Themore
energythefluorescentdyegivesoff,thestrongerthesignal.
Page10of86
TriageTestDevice.Atestdeviceholdsthe
sampleandalsocontainsthereagentsand
othermaterialsneededforthetest.
Overview:RunningaTest
Afteratestsample(forexample,blood)fromapatientisaddedtothetestdevice,thetestdeviceisinserted
intheQuidelTriageMeterPro.Themetermeasureshowmuchofthesubstance(aparticularprotein
marker)ispresent,basedonstandardsthathavebeenpreͲprogrammedintothemeter.Thepatient
identification,theUserI.D.andthetestresultscanbeprintedoutonapapertape.
Inadditiontotestingsamplesfrompatients,theQuidelTriageMeterProcontainspreͲprogrammedQuality
Control(QC)functionsthatallowthepersonoperatingthemetertomonitorqualitycontrol.
NOTE:TheQuidelTriageMeterProacceptstestdevicesthataredesignedspecificallyforusewiththe
QuidelTriageMeterPro.PleaserefertoproductͲspecificPackageInsertsincludedineachtestkitformore
information.
Unpacking
TheQuidelTriageMeterProandtheitemsthatcomewithitareprovidedinasinglebox.
QuidelTriagetestdevicesarepackagedseparatelyandincludeinstructionsforrunningspecifictests.Ifyou
haveanyquestions,contactQuidel(refertotheAssistancesection).
Page11of86
Contents
QuidelTriageMeterPro
QCDeviceandCODECHIPmodule(intheQCDevicebox)
AC/DCPowerConverter
AABatteries
RollsofPrinterPaper(additionalrollsaresuppliedineachboxoftestdevices)
SupervisorCODECHIPmoduleandCODECHIPmoduleBox
Printerpaper
locatedbelow
QCDevicebox.
Page12of86
PartsoftheQuidelTriageMeterPro
Page13of86
On/OffKey UsedtoturntheQuidelTriageMeterProonandoff.
(Powerbutton)
PrintKey Usedtoprintthetestresultsordisplayscreen.
DeleteKey Usedtodeletedata.
Allowstheoperatortoselectmenuitemsandacknowledgealarm
EnterKey
conditions.
ExitKey Usedtoexitthedisplayedmenuorejectthetestdevice.
BrightnessKey Usedtosetscreenbrightness
Keypad Usedtoenteridentificationnumbers,letters,andspecialsymbols.
Usedtotogglethroughmenuitems.
Up
Down
ArrowKeys
Left
Right
ShiftKey
SymbolKey UsedwithShiftKey
ThermalPrinter Printsthetestresultonpaper.
Connectionpointforthe(optional)BarCodeScanner,allowingthe
operatortoscanpatientoruseridentificationdatainsteadofmanually
DataPort
enteringnumbersorlettersonthekeypad.Mayalsobeusedtoconnect
toadatamanagementsystem/LIS.
PaperFeedKey Usedtofeedpaper
CODECHIP InsertionpointforCODECHIPmoduleswhichcontainlot/devicespecific
ModulePort dataforuseinprovidingtestresults.
PowerSupplyPort ConnectionpointforthesuppliedAC/DCPowerConverter.
PrinterCover Thecover,whichwhenpulledstraightup,revealsthepaperroll.
TheLCD(LiquidCrystalDisplay)Screenshowsthemenuofpossibletests
LCDScreen
andtasksandpromptstheoperatortotakethenextstep.
ScreencommandsandoptionsareindicatedbyaBOLDEDFONT.
Page14of86
Installation
UnpackMeter
a. Removethemeterfromtheboxandprotectiveplasticbag.
PowerMeterOn
Batteries
a. Turnthemeterover.
b. Removebatterycover.
c. Install4AAbatteries,payingattentiontobatteryorientationinthe
batterycompartment.
d. Replacebatterycoverandturnmeterrightsideup.
8:23AM01•20•18
LowPowerIndicator
Timeanddate
RUN Menuitems
TEST
RECALLRESULTS
INSTALLNEWCODECHIP
SELECTWITHvOR^ANDPRESSENTER
UserInstructions
PRESSEXITTOEJECTDEVICE
TheMainMenu
NOTE:TheLowPowerIndicatormonitorstheactivepowersource.Toverifyadequatebatterpower:
NavigatetotheMainMenu
RemovetheACpowersupply
PresstheEXITkey(themeterattemptstoejectthedevice)
ObservetheMainMenuonthemeter’sdisplay.
Replacethebatteriesifthepowericonisvisible.
Page15of86
ACPowerSupply
a. Removethepowersupplyfromthebox.
b. PlugoneendintoanACoutlet.
c. Plugtheoppositeendintotheroundholeinthebackofthemeter.
InstallPaper
a. Removetheprintercoverbypullinguponthecoverasindicatedbythearrowonthebackofthecover.
FormoredetailsaboutInstallingPaperseetheCareandMaintenancesection.
b. Tearorcutaclean,straightedgetofeedintotheprinter.Donotcutpaperatanangle,astheprinter
mustsensetheedgeofthepaperalongthefeedpath.
NOTE:Theprintercontainsapapersensorandwillfeedthenewpaperrollonlywhenpaperwithaclean
straightlineispressedintothepaperroller.
c. Positionthepapersuchthatthepaperwillfeedfromundertheroll(asopposedtooverthetopofthe
roll,seepicture).
Page16of86
d. Insertthepaperedgeunderthepaperroller(platen)untilitfirmlyseatsorresistanceisfelt.
e. Pressthe keytwice,paperwillfeedthroughtheprinterandextendoutthemeter.
f. Placetherollofpaperintothepapercompartment.
g. Replacetheprintercoverandcontinueoperation.
InsertSupervisorCodeChipModule
a. RemovetheSupervisorCODECHIPmoduleBoxfromthemeterbox.
b. RemovetheSupervisorCODECHIPmodulefromitsbox.
c. InserttheSupervisorCODECHIPmoduleintotheCODECHIPmodulePort.Theportislocatedontheleft
sideofthemeter,towardsthebottomfrontcorner.NotetheorientationoftheCODECHIPmodulein
thedrawingbelow.
d. OncetheSupervisorCODECHIPmoduleisinstalled,verifythattwoadditionalmenuitemsappearonthe
displayscreen.
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODECHIP
Twoadditionalmenuoptions SETPARAMETERS
accessiblewhenSupervisor
CODECHIPmoduleisinserted. DELETERESULTS
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
TheSupervisor’sMainMenu
FormoredetailsaboutCODECHIPmoduleseepage22.
ToChangeMeterSettings
a. SelectSETPARAMETERSusingthe keysandpressthe key.
Page17of86
c. SelectIDSETTINGS,DISPLAYSETTINGSorCOMMSETTINGSusing
byusingthe keys.
IDSettings Default Options NewSettings
CHARACTERSUSERID Min1 1Ͳ16
Max16 1Ͳ16
CHARACTERSPAT.ID Min1 1Ͳ20
Max20 1Ͳ20
CHARACTERSAUXID Min1 1Ͳ12
Max12 1Ͳ12
CHARACTERSMISCID Min1 1Ͳ20
Max20 1Ͳ20
AUX.IDENABLE/DISABLE Disabled Disabled,Enabled
Forreference,recordyoursettingsinthespaceprovided.
e. Pressthe keytosavechangestoIDSettings.
Page18of86
f. Press tosavechangestoDisplaySettings.
LISPASSWORD
k. SelectMMͲDDͲYY,DDͲMMͲYYorYYͲMMͲDDusingthe
keys.ThepreviouslyenteredDATEchangesas
theDATEFORMATchanges.
Page19of86
Setting Default Options NewSettings
TIMEFORMAT AM/PM AM/PM;24HR.
DATEFORMAT MMͲDDͲYY MMͲDDͲYY,
DD.MM.YY,
YYͲMMͲDD
SetUserID
TheQuidelTriageMeterPronormallyrequiresaUserIDtobeenteredpriortorunningapatienttest,QC
test,orMisctest,andpriortorecallingpatientresults.Refertopage31forinstructions.Alternately,this
featuremaybebypassed.Seepage21.
ChangeReferenceRanges/ThresholdsforTests:
a. SelectSETPARAMETERSusingthe keysandpressthe key.
NOTE:ThelowestvalueoftheRangeorThresholdmaybedisabledforsometestpanels.Refertothe
appropriateProductInsertforspecificinformation.
FormoredetailsaboutRangesSettingsseepage33.
ChangeTestSettings
a. SelectSETPARAMETERSusingthe keysandpressthe key.
NOTE:Theabilitytochangetestsettingsmaybedisabledforsometestsinspecificpanels.Refertothe
appropriateProductInsertforspecificinformation.
FormoredetailsaboutTestSettingsseepage34.
Page20of86
ChangeQCParameters
a. SelectSETPARAMETERSusingthe keysandpress .
NOTE:IfthebypassisON,theuserIDfeatureisdisabledandanyonemayrunatest.IfthebypassisOFF,theuserID
featureisactiveandonlyauthorizedusersmayrunatest.ToInstallUserID's,seepage31.
FormoredetailsaboutBypassSettingsseepage36.
Priortocommencingpatienttesting,removetheSupervisorCODECHIPmodulefromthemeter,returnit
tothestorageboxandplaceinasafeplaceforfutureuse.
ThemeterisnowreadytorunQCTestsinpreparationforPatientTesting.
QCTestingshouldbeperformedinthefollowingmanner:
1. RunQCDEVICE(seepage40).
2. RunQCSAMPLEforeachlotofdevices(seepage42).
3. Ifappropriate,runCALIBRATIONVERIFICATIONSETasaMiscTestsampleforeachtestpaneltypetobe
used.(RefertotheapplicableProcedureManualandCalibrationVerificationSetPackageInsertfor
detailedinstructions).
Page21of86
Location
TheQuidelTriageMeterProshouldbeplacedonadry,clean,flat,horizontalsurfaceawayfromdirect
sunlight.
Power
TheQuidelTriageMeterProrequires6ͲvoltsDCsuppliedbyeitherACorDC.Themeterwillrun
approximately100testsonbatteries.Itisrecommendedtousebatteriesasabackupintheeventofa
poweroutage.
NOTE:TheQuidelTriageMeterProwillautomaticallyturnoffifleftunusedfor2hours,unless
programmedotherwisebythesupervisor.Optionalsettingsallowautomatedturnoffafter30minutes,
1hour,2hours,4hoursorNONE.
CAUTION:Useonlythepowersupplyprovidedwiththemeter,failuretodosomaycausepermanentdamageto
themeter.
SelfͲTest
TheQuidelTriageMeterProchecksthesystemtobesureitisoperatingproperlybyrunningselfͲtestswhen
poweredonandpriortoeachtest.
PowerOn–includeslaseroperation,internalstandardrange,batterypower,andsoftwareverification.
EachTest–verifiesthelaseroperation,internalstandardrangeandbatterypower.
CODECHIPModules
MeterdataisupdatedviaadisposableCODECHIPmodule.TheCODECHIPmodulecontainsmicrochip
circuitryembeddedintoaplastichousing.WheninsertedintothemeterCODECHIPmoduleportand
activated,theinformationistransferredintotheQuidelTriageMeterPro’smemory.ACODECHIPmodule
typicallyneedstobeinstalledonlyonceandremainsinthemetermemory.ACODECHIPmoduledoesnot
needtoremaininthemeterwhileperformingtests.
TherearefivetypesofCODECHIPmoduleswhichcanbedistinguishedbyshapeorcolor.
StandardCODECHIPmodule
ReagentTestCODECHIPmodule–includedineachboxoftestdevices;containscalibration,expirationdate
andotherdataaboutthedevicelot.(Colorvariesbytesttype.)
QCSampleCODECHIPmodule–includedineachboxofQCSample;containsexpirationdateandother
dataabouttheQCSamplelotincludingacceptablerange.(White.)
QCDeviceCODECHIPmodule–includedintheblackQCDevicebox(comeswiththeQuidelTriage
MeterPro);containsdataabouthowthemetershouldreaditsmatchingQCDevice.(Darkgray.)
SupervisorAccessCODECHIPmodule–includedwitheachQuidelTriageMeterPro;containscodeto
activateSupervisorfunctions.(Lightgray.)
Page22of86
SoftwareUpgradeCODECHIPmodule–providedwhenQuidelTriageMeterProsoftwareupgradesare
available;containsnewsoftwareforthemeter.
SetParameters
Purpose
TheSETPARAMETERSfunctionallowsthesupervisortoselectanumberofsettings.Accesstothesesettings
iscontrolledusingtheSupervisorCODECHIPmodule.
ListofProgrammableParameters
Asupervisorcansetthefollowingparameters:
Heading Parameters
IDSETTINGS NumberofCharactersinUserID,NumberofCharactersinPatient
ID,NumberofCharactersinAuxiliaryID,EnableorDisable
AuxiliaryID,NumberofCharactersinMiscTestID
DISPLAYSETTINGS Language,PrinterMode,AutoPowerͲOff,DisplayContrast,
PrinterContrast
COMMUNICATIONS BaudRate,EnableorDisableLIS,EnableorDisableAutoUpload,
EnableorDisablePatientResultApproval
CLOCK Time,Date,andDisplayFormat
USERID AddNewUserID,UpdateUserID,DeleteUserIDorUserIDList
RANGES TestCutoffs
TESTSETTINGS BlockAnalytes
QCPARAMETERS MinimumFrequencyforQCTests,NumberofControls,Number
ofStandardDeviations(QuantitativeTests)
BYPASS DisableUserIDrequirement
SupervisorAccess
ASupervisorCODECHIPmoduleisshippedwitheachmeter.Installing
theCODECHIPmodulegivesthesupervisoraccesstotheSET
PARAMETERSandDELETERESULTSfunction.
Page23of86
ToinserttheSupervisorCODECHIPmoduleintotheQuidelTriageMeterPro,slidethechiplabeled"SPR"
intothemeter’sCODECHIPmoduleportasshowninthepicture.
HowtoSetParameters:BasicInstructions
Allinstructionsassume:
Themeterison.
Themeterscreenisdisplayingthemainmenu.
TheSupervisorCODECHIPmoduleisinstalled.
1.
SelectSETPARAMETERSusingthe keys.
2. Pressthe key.
3.
Youwillseealistofprogrammableparameters.
METERSETTINGS
CLOCK
USERID
RANGES
TESTSETTINGS
QCPARAMETERS
BYPASS
SELECTWITHvOR^ANDPRESSENTER
ORPRESSEXIT
TheSetParametersMenu
4. Usethe keystoselecttheparameteryouwanttoprogram:
METERSETTINGS,CLOCK,USERID,RANGES,TESTSETTINGS,QCPARAMETERS,orBYPASS.
5. Pressthe key.
6. Followthedirectionsforsettingtheselectedparameter(seethenextsection).
NOTE:Tosavechanges,presstheENTERkey.Tocancelchanges,presstheEXITkey.Wheneitherthe
ENTERortheEXITkeysarepressed,themeterperformstheappropriatefunctionandreturnstothe
previousmenu.
Page24of86
MeterSettings MainMenu
Thismenuwillallowyoutosetbasicoperationalcriteriaforthemeter:language, SetParameters
displaycontrast,maximumlengthoftheuser,patient,auxiliaryandmisc.ID's, MeterSettings
printermodeandautomaticpoweroff. IDSettings
IDSettings MainMenu
CharactersUserID SetParameters
Thissettingwillsettheminimumandmaximumnumberofcharactersin MeterSettings
theUserID. IDSettings
ThissettingwillsettheminimumandmaximumnumberofcharactersinthePatientID.
1. Fromalistofprogrammableparameters,chooseINSTRUMENTSETTINGSusingthe keys.
2. Pressthe key.
3. Usethe keystoselectIDSETTINGS.
4. Pressthe key.
5. Usethe keystoselectthenumbertotherightofCHARACTERSPAT.IDundertheMINcolumn.
6. Usethe keystochoosethesmallestnumberofcharactersthatwillappearinthePatientID.
Theminimumnumberofcharacterspossibleis1.
7. Usethe keytoselectthenumbertotherightofCHARACTERSPAT.IDundertheMAXcolumn.
8. Usethe keystochoosethelargestnumberofcharactersthatwillappearinthePatientID.The
maximumnumberofcharacterspossibleis20.
Page25of86
9. Pressthe keytosavechanges.
CharactersAux.ID MainMenu
ThissettingwillsettheminimumandmaximumnumberofcharactersintheAuxID. SetParameters
IfthesettingofAUX.IDENABLE/DISABLEisENABLED,anAuxiliaryIDismandatoryfor MeterSettings
eachpatienttest. IDSettings
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthe keys.
2.
Pressthe key.
Aux.IDEnable/Disable MainMenu
ThissettingwillturnonorturnofftheAuxiliaryIDfunction.TheAuxiliaryIDmaybe SetParameters
usedforrecordingatestordernumber,physicianIDorotherspecializedidentifying MeterSettings
information. IDSettings
Page26of86
CharactersMisc.ID
ThissettingwillsettheminimumandmaximumnumberofcharactersintheMisc.TestID.
Fromalistofprogrammableparameters,chooseINSTRUMENTSETTINGSusingthe keys.
Page27of86
3. Usethe keystoselectDISPLAYSETTINGS.
4. Pressthe key.
5. Usethe keystoselecthighlightthewordtotherightofPRINTMODE.
6. Usethe keystochooseAUTOMATICorMANUAL.IfAUTOMATICisselected,themeterwill
immediatelyprintouttheresultsaftereachtest(patient,QCorMiscTest).IfMANUALisselected,the
usermustusethe buttononthekeypadtoprintoutresults.
7. Pressthe keytosavechanges.
MainMenu
AutoPower–Off
SetParameters
Thissettingwilldeterminethelengthoftimebeforeaninactivemeter
MeterSettings
willpoweroff.
DisplaySettings
NOTE:Whenpoweredonlybybatteries,select1/2hourtoconservebatterylife.Certainscreens,whendisplayed,
willnotautomaticallypoweroff.TheNewResultsscreenisanexample.
.
DisplayContrast
Thissettingwilladjustthedisplaycontrast.
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthe keys.
2. Pressthe key.
3. Usethe keystoselectDISPLAYSETTINGS.
4. Pressthe key.
5. Usethe keystoselectthenumbertotherightofContrast.
6. Usethe keystochangethecontrastofthemeter’sscreen.Therangeisfrom0–8(8beingthe
greatestcontrast).
7. Pressthe keytosavechanges.
Page28of86
PrinterContrast MainMenu
Thissettingwilladjusttheprintercontrastformeterswithaserialnumber SetParameters
29901orgreater. MeterSettings
DisplaySettings
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthe keys.
2. Pressthe key.
3. Usethe keystoselectPRINTERCONTRAST.
4. Pressthe key.
5. Usethe keystoselectthenumbertotherightofContrast.
6. Usethe keystochangethecontrastofthemeter’sscreen.Therangeisfrom0Ͳ8(8beingthe
greatestcontrast).
7. Pressthe keytosavechanges.
CommunicationSettings
MainMenu
PriortoactivatingtheLISfeature,obtainthemeter’sserialnumberlocated
SetParameters
underneaththemeterandcontactQuidel(RefertotheAssistancesection).
MeterSettings
TheQuidelRepresentativewillprovideyouwithapasswordthat,whenentered,
CommSettings
willallowdatatobetransmitted.
BaudRate
1.
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthe keys.
1.
2.
1. Pressthe
key.
3. Usethe keystoselectCOMMSETTINGS.
4. Pressthe key.
5. Usethe keystoselectBAUDRATE.
6. Usethe keystochoose9600or38400.
7. Pressthe key.
PatientResultApproval
Someinstitutionsmaywishforpatienttestresultstobeverifiedbytheuser MainMenu
beforeprintingorsendingtotheLIS.Activationofthisfeaturedisplaystheresult, SetParameters
thenrequirestheusertoacceptorrejecttheresultbeforetherecordisprinted MeterSettings
andsaved.Ifthetestisrejected,themeterwillpermanentlyflagitasrejected CommSettings
bytheuser.
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthe keys.
2. Pressthe key.
Page29of86
3. Usethe keystoselectCOMMSETTINGS.
Whenenabled,afterapatientresultisavailableanddisplayedonthemeterscreen,theusermustacceptor
rejecttheresultbeforeprinting,uploadingorperforminganyothermeterfunctions.
ThemeterpromptstheusertoPRESS1TOACCEPTOR0TOREJECT.Oncethetesthasbeenacceptedor
rejected,themeteroperatesnormally.
Ifthetesthasbeenrejected,thedisplayandprintoutwillbeflaggedwithPAT.RESULTREJECTEDor
RESULTSREJECTEDBYUSER.RecalledresultswilldisplayanRbeforethepatientIDintherejectedrecord.
LISEnable MainMenu
1. Fromalistofprogrammableparameters,chooseMETERSETTINGS SetParameters
usingthe keys. MeterSettings
2. Pressthe key. CommSettings
3. Usethe keystoselectCOMMSETTINGS.
4. Pressthe key.
5. Usethe keystoselectLIS.
6. Usethe keystochooseDISABLEDorENABLED.
7. Pressthe key.
8. IfDISABLEDwasselected,ascreenshowingaLISDISABLEDwillappear.Press
the key.
9. IfENABLEDwasselected,ascreenshowingaENTERLISPASSWORDwillappear.Enterthepassword
providedbytheQuidelRepresentative.Pressthe key.
10.AcknowledgetheLISENABLEDmessagebypressingthe key.
AutoUpload
(ThisfeatureisvisibleaftertheLIShasbeenenabled)
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthe keys.
2. Pressthe key.
3. Usethe keystoselectCOMMSETTINGS.
4. Pressthe key.
5. Usethe keystoselectAUTOUPLOAD.
6. Usethe keystochooseENABLEDorDISABLED.
7. Pressthe key.
Page30of86
Clock MainMenu
Theclockmenusetsthetimeanddateofthemeterandtheformatfordisplaying SetParameters
both.Ifthemeterhaslostpower,thetimeanddatewillblinkontheMainMenu. Clock
1. Fromthelistofprogrammableparameters,chooseCLOCKusingthe keys.
NOTE:IfyouhaveselectedAM/PM,AMorPMappearsafterthetime,asappropriate.
UserID
MainMenu
Upto600UserID'swith1–16charactersmaybeprogrammedintothemeter.
SetParameters
TheUserIDmayalternatelybeenteredusingtheexternalBarCodeScanner
UserID
(optional).
AnadditionalsecurityfeatureoftheUserIDisthepartialmaskingoftheIDondisplaysandprintouts.This
preventsunauthorizedindividualsfromdetectingandusinganexistingUserID.WhentheSupervisorCODE
CHIPmoduleisinstalledinthemeter,thisfeatureisoverriddenandallUserIDcharactersaredisplayed.
NewUserID
1. Fromthelistofprogrammableparameters,chooseUSERIDusingthe keys.
2. Pressthe key.
3. SelectNEWUSERIDusingthe keys.
4. Pressthe key.
Page31of86
5. TypeinaUserIDwithanumberofcharacterswithinMINCHARACTERSUSERIDandMAXCHARACTERS
USERID.Tocorrect,press tocleartheentireID,orselecttheincorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
6. Pressthe key.
7. Onthenextscreen,usethe keystoselectanexpirationdatetotherightofUSERIDSTATUS:6
MONTHS,12MONTHS,VALIDorEXPIRED.
8. Pressthe keytosavechanges.
NOTE:UserID9999999999(10Ͳ9’s)isadefaultsettingdisplayedwhentheUserIDBypassisonorwhena
testisrunwiththeSupervisorCODECHIPmoduleinstalled.
NOTE:IfyoutypeinaUserIDthathasalreadybeenassigned,themessageonthescreenstates:
USER###ALREADYEXISTS.PressENTERtogobackandtypeinadifferentID
TheQuidelTriageMeterProrecognizes0’sasauniquecharacter.Therefore,‘4341’willberecognized
separatelyfrom‘04341’.
Usingthebarcodescanneroralphamode(shiftkey)willallowtheQuidelTriageMeterProtoacceptalpha
charactersandthefollowingsymbols:#()*Ͳ./\
UpdateUserID MainMenu
1. Fromthelistofprogrammableparameters,chooseUSERIDusingthe SetParameters
keys. UserID
2. Pressthe key.
3. SelectUPDATEUSERIDusingthe keys.
4. Pressthe key.
5. TypeintheuserIDyouwanttoupdate.Tocorrect,press tocleartheentireID,orselectthe
incorrectcharacterbyusingthe keys.ThentypeinthecorrectID.
6. Pressthe key.
7. TheuserIDisdisplayedwiththeexpirationdate.Youcanchangetheexpirationdateto6MONTHS,12
MONTHS,EXPIREDorVALIDbyusingthe keys.VALIDreconfirmsthepreviouslyselected
interval.
8. Pressthe keytosavechanges.
NOTE:IfyoutypeinaUserIDthatdoesnotexist,themessageonthescreenstates:USERID###DOES
NOTEXIST.Ifthishappens,pressENTERtoreturntothepreviousscreenandcorrecttheID,orpresstheEXITkey.
DeleteUserID
Fromthelistofprogrammableparameters,chooseUSERIDusingthe keys.
Page32of86
SelectDELETEUSERIDusingthe keys.
Pressthe key.
TypeintheuserIDtobedeleted.Tocorrect,press tocleartheentireID,orselecttheincorrect
characterbyusingthe keys.ThentypeinthecorrectID.
Pressthe key.Themessageonthescreenstates:USERID###WILLBEDELETED.
Pressthe keytodelete.AmessageonthescreentellsyouthattheIDhasbeendeleted:USERID
###DELETED.
NOTE:
IfyoutypeinaUserIDthatdoesnotexist,themessageonthescreenstates:USERID###DOESNOT
EXIST.Ifthishappens,pressENTERtoreturntothepreviousscreenandcorrecttheID,orpresstheEXITkey.
UserIDList MainMenu
1. Fromthelistofprogrammableparameters,chooseUSERIDusing SetParameters
the keys. UserID
2. Pressthe key.
3. SelectUSERIDLISTusingthe keys.
4. Pressthe key.
5. Alistisdisplayed.
6. Usethe keystoselectthetypeofIDlistyouwant:USERID,FROMEXPIRATIONDATE,TILL
EXPIRATIONDATE.
ForalistofallID's,selectUSERID.
ForasingleID,selectUSERIDandtypeintheID.Tocorrect,press tocleartheentireID,or
selecttheincorrectcharacterbyusingthe keys.ThentypeinthecorrectID.
ForarangeofID's,entertheFROMand/ortheTOdates.Thedatesareenteredbytypingthemonth
number,thedaynumberandtheyearnumberintheformatyouhavechosen.Thenumbersyou
typewillappearfromtherightandmoveleftasyoucontinuetotype.Tocorrect,press toclear
theentiredate.Thentypeinthecorrectdate.
Page33of86
theproductpackageinsertorcontactQuidelforclarification.Thesevalueswillbeinreversevideoon
thepatientresults.
Analyteswithacomma–Threerangespossible.Ifthelowernumberis0.0,themetertreatsthe
analyteashavingonlytworanges.Ifthelowernumberisnot0.0,thenormalrangeisanyvalueless
thanthelowernumber.Thefirstabnormalrangeisanyvaluebetweenthetwonumbersandisinclusive
ofthesenumbers.Thesevalueswillbeboxedonpatientresults.Thesecondabnormalrangeisanyvalue
greaterthantheuppernumber.Thesevalueswillbeinreversevideoonpatientresults.Consultthe
productpackageinsertorcontactQuidelforclarification.
1. Fromthelistofprogrammableparameters,chooseRANGESusingthe keys.
2. Pressthe key.
3. Selectthetestpaneltypeusingthe keys.
4. Selecttheanalyteandlevelusingthe keys.TheHIGHandLOWvalueforeachoftheanalytes
canbechangedusingthekeypadtotypeinthevalues.Tocorrect,press tocleartheentirenumber.
Thentypeinthecorrectnumber.
NOTE:Thefieldwillbedisabledifthepanelsettingisfixed.Consultthespecificpanelpackageinsertto
determineifthesettingmaybechanged.
TestSettings
Incertaincases,thelabsupervisormaywishtocontrolwhichtestsareavailable MainMenu
toberunatthetimeofpatienttesting.Theoptionsare: SetParameters
1. ACTIVE(ACTIVONSCREEN):Thetestisalwaysrunandcannotbedeselected.
2. INACTIVE(INACTONSCREEN):Thetestisneverrunandisnotdisplayed.
3. USERINACTIVE(USRIONSCREEN):Thetestcanbeselectedbytheuseratthetimeofrunningthetest.
4. USERACTIVE(USRAONSCREEN):Thetestcanbedeselectedbytheuseratthetimeofrunningthetest
Iftheseoptionsarenotavailableforatest,themeterskipsoverthesettingasthesupervisorscrollsthrough
theoptions.Seepage45formoredetails.
1. Fromthelistofprogrammableparameters,chooseTESTSETTINGSusingthe
1. keys.
1.
2.
1. Pressthe
key.
3.
1. Selectthetestpaneltypeusingthe keys.
1.
4. Selecttheanalyteusingthe keys.
1.
5. Selectthedesiredmodeusingthe
1. keys.
1. ACTIVE(ACTIVONSCREEN)–Thetestwillrunoneverypatientspecimen.
1. INACTIVE(INACTONSCREEN)–Thetestwillneverrunonapatientspecimenandisdroppedfrom
1. alldisplaysandprintouts.
1.
1. USERINACTIVE(USRIONSCREEN)–Atthetimeofpatienttesting,theuserhastheabilitytoselect
1. thetest.
1.
1.
Page34of86
USERACTIVE(USRAONSCREEN)–Atthetimeofpatienttesting,theuserhastheabilityto
deselectthetest.
NOTE:Whentheanalyteisdeselected,thevalueisnotcalculatednorstored
QCTestsWithTestSettingsActivated
AnytestwhichtheSupervisorhassettoACTIVE,USERACTIVEorUSERINACTIVEwillrunontheQCSample
panel.Themeterwilldisplayandprintthetest’sQCsampleresultandupdatetheQCSampletimerforthe
test.IftheSupervisorhassetthetesttoINACTIVE,thetestwillhavetheQCSampleturnedOFF.TheQC
Sampleresultwon’tshowforthattestandtheQCSampletimerwon’tbeupdatedforit.
IftheSupervisorchangesthetestfromINACTIVEtoACTIVE,toUSERACTIVEortoUSERINACTIVEthattest
maybeoutofQCSampledate.AsalwaysforatestwithanexpiredQCSample,apassingQCSampletestwill
needtorunbeforeapatientsamplecanreportaresultforthattest.
QCParameters
TheQCParametersmenuprovidestheusertheoptionsfordeterminingwhatthe
MainMenu
maximumfrequencyisforrunningtheQCDeviceandQCSample,whether1or2
SetParameters
levelsofcontrolarerequiredfortheQCSample,andwhethertheallowedQC
QCParameters
SampleRangeofquantitativetestsis2or3standarddeviations.
QCDeviceFrequency
TheQCDEVICEFREQUENCYsettingdeterminesthemaximumintervalbetweenQCDeviceteststhatauser
isallowedtorunpatienttests.Whentheintervalhaslapsed,allusersarelockedoutoftheRunTestmenu
untilaQCDevicehasbeensuccessfullyrun.
ItisrecommendedthattheQCDeviceberundailywhenperformingpatienttesting.
Page35of86
1. Fromthelistofprogrammableparameters,chooseQCPARAMETERSusingthe keys.
2. Pressthe key.
3. Usethe keystoselecttheinformationtotherightofQCSAMPLEFREQ.
4. Usingthe keys,selectoneoftheseoptions:NONE,8HOURS,DAILY,WEEKLY,MONTHLY.
5. Pressthe keytosavechanges.
NumberofControls MainMenu
Fromthelistofprogrammableparameters,chooseQCPARAMETERS SetParameters
usingthe keys. QCParameters
NumberofControlStandardDeviationsforQuantitativeTests
1. Fromthelistofprogrammableparameters,chooseQCPARAMETERSusingthe keys.
BypassUserID MainMenu
Thisfunctionallowsthesupervisortoselectivelybypasstherequirementtoenter SetParameters
auserIDbeforeperforminganytest. Bypass
1. Fromthelistofprogrammableparameters,chooseBYPASSusingthe keys.
2. Pressthe key.
3. SelectUSERIDusingthe keys.
Page36of86
InstallCodeChip
Purpose
TheQuidelTriageMeterProwillprompttheusertoinstallaCODECHIPmodulewhenrequired.However,
theINSTALLCODECHIPfunctionmaybeusedasanalternatemethodtotransferinformationfromaCODE
CHIPmoduleintothemeter’smemory.ThemeterwilldirecttheusertoinstallaCODECHIPmoduleifitis
attemptingtorunatestanddoesnothavethedatainitsmemory.
TypesofCODECHIPmodules
TherearefivetypesofCODECHIPmodules
ReagentCODECHIPmodule–includedineachboxoftest
devices.LotnumberbeginswithaW.
QCSampleCODECHIPmodule–includedineachboxof
controls.LotnumberbeginswithaC.
QCDeviceCODECHIPmodule–includedintheblackQC
Devicebox(comeswiththeQuidelTriageMeterPro)andis
labeledwiththeserialnumberoftheQCDevice.
SupervisorAccessCODECHIPmodule–includedwitheach
QuidelTriageMeterPro.LotNumberbeginswithSPR.
SoftwareUpgradeCODECHIPmodule–sentseparately
whenQuidelTriageMeterProsoftwareupgradesare
available.
CAUTION:RunningtestswiththeSupervisorCODECHIPmoduleinstalleddisablesallQC
Lockouts.RemovetheSupervisorCODECHIPmodulepriortorunningtests.
InstallationofReagent,QCSampleorQCDeviceCODECHIPModules
InstallCodeChip
1. FromtheMainMenuselectINSTALLCODECHIPusingthe keys. MainMenu
2. Pressthe key.
Page37of86
3. SlidetheCODECHIPmoduleintotheCODECHIPmodulePortintheQuidelTriageMeterPro,asshownin
thepicture.
INSTALLATIONOFSUPERVISORCODECHIPMODULE
TheSupervisorCODECHIPmoduleneedsonlytobeinsertedintotheCODECHIPmodulePorttoactivatethe
supervisorfunctions.ThisisanautomaticfunctionandstandardCODECHIPmoduleinstallationisnot
necessary.
Page38of86
Operation
AccesstotheQuidelTriageMeterPro
Function Use Restrictedto
RUNTEST PatientTesting OperatorswithvalidUserID's*
QualityControlTesting OperatorswithvalidUserID's*
Misc.TestTesting OperatorswithvalidUserID's*
RECALLRESULTS LastRecord Iflasttestrunisapatientsample:
OperatorswithvalidUserID's*
Otherwise:Norestriction
PatientTestResults OperatorswithvalidUserID's*
QualityControlResults NoRestriction
Misc.TestResults NoRestriction
INSTALLCODECHIPS TypesofCODECHIP IfCODECHIPmoduleissoftware
modules upgrade:
Supervisor
InstallationProcedure Otherwise:Norestriction
SETPARAMETERS AccessControl Supervisor
ProgrammableSettings Supervisor
DELETERESULTS PatientTestResults Supervisor
*UnlessUserIDisbypassedortheSupervisorCODECHIPmoduleisinstalled.
IMPORTANT:Thelasttwofunctions,SETPARAMETERS,andDELETERESULTS,appearonthemain
menuonlywhentheSupervisorCODECHIPmoduleisinsertedintotheCODECHIPmodulePort.The
SupervisorCODECHIPmoduleshouldberemovedfromthemeterduringroutinepatient,quality
control,andMisc.testing.
NOTE:Accesstovariouspartsofthesoftwareisuser(institution)defined.
RunTest
Purpose
RunTestisusedfortheanalysisofatest.
Options
TheRuntestfunctionallowsyoutotest:
QCDevice
Page39of86
QCSample
PatientSample
Misc.TestSample
Inordertorunapatientsample,themeterrequires:
1. AvalidUserIDhasbeenentered(unlessthisfeatureisbypassedoraSupervisorCODECHIPmoduleis
installed).Seepage31.
2. AQCDevicehaspassedwithinthespecifiedtimeperiod.Seepage41.
3. AQCSampleforthetestdevicelothaspassedwithinthespecifiedtimeperiod.Seepage42.
Ifanyoftheseconditionsdonotexist,themeterwillnotcontinueandwillprompttheusertoperformthe
requiredsteps.
IMPORTANT:TheQuidelTriageMeterProwillalerttheuserifQChasnotbeenperformedforthe
devicelotbeingtested.
NOTE:TheSupervisorCODECHIPmoduleprovidesaccesstofunctionsthatmayoverridetheseLockouts.
NOTE:IftheQuidelTriageMeterPro’sinternalqualitycontrol(QC)resultsareunacceptable,exclamation
points(!)willbedisplayedonthescreenintheplaceofthepatientresults.Toobtainresultsrepeatthe
testusinganewdevice.
QCDevice
TheQCDeviceshouldberunoneachdayofpatienttesting.
ToruntheQCDevicetest: MainMenu
RunTest
1. Pressthe keytopowertheQuidelTriageMeterProon.
NOTE:IftheUserIDbypassison,themeterwillnotaskyoutoenteraUserID.
Page40of86
5. Pressthe key.
6. SelectQCDeviceusingthe keys.
PATIENTSAMPLE “EXP”indicatesthatthe
QCDevicereͲtest
QCDEVICE EXP frequencyhasexpired.
QCSAMPLE EXP
MISC.TEST
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
MenuscreenasitlookswhenoperatorselectsQCDevice.
Pressthe key.
8. GentlyinserttheQCDeviceintotheQuidelTriageMeterPrountilyoufeeltheQCDevicecatchonthe
pinandhearanaudible‘click.’
Pressthe key.
NOTE:ThemeterwillprompttheusertoinstalltheQCDeviceCODECHIPmoduleiftheQCDevicehasnot
beenrunbefore.
10.TheQuidelTriageMeterPropullsintheQCDeviceandscansit.Thetest
devicemaypartiallymoveinandoutofthemeterseveraltimes.
11.Whenthetestiscomplete,themeterwillbeep,ejectthedeviceanddisplaytheresultsonthe
meter’sscreen.
12.Pressthe keytomakeaprintedcopyoftheresults.
13.RemovetheQCDevicefromtheMeterandplaceintheQCDeviceBox.DONOTDISCARDTHE
QCDEVICE.
IMPORTANT:Importantinstructionsforalltests:
a. GentlyinsertthetestdeviceintotheQuidelTriageMeterPrountilyoufeelthedevicecatchonthe
pin.Youwillhearanaudible‘click’thattellsyouthetestdevicehasbeeninsertedproperly.
b. Pressthe keytostartthetest.
c. TheQuidelTriageMeterPropullsinthetestdeviceandscansit.
d. Whenthetestiscomplete,themeterwillbeepanddisplaytheresultsonthemeter’sscreen.
e. Pressthe keytomakeaprintedcopyoftheresults.
CAUTION:Afterthetestdevicehasbeeninserted,donotpushthedeviceinfurtherorattempt
topullitout.ThedevicemaybeejectedbyreturningtotheMainMenuandpressingtheEXIT
key.
Page41of86
IMPORTANT:ImportantInstructionsforQCDevicetests:
Dust,lint,fibersandothersmallparticlesmayinterferewiththeQCDevice.KeeptheQCDevicefreeof
contaminants.
TheQCDeviceislightͲsensitiveandshouldbestoredinitsblackopaquecasewhennotinuse.
IftheQCDevicetestsfail,wipetheQCDevicecleanwithalintfreeclothtoremoveanyoils,dust,
fibers,orfingerprints.DonotapplyanyliquidtotheQCDevice.Aftercleaningthedevice,repeatthe
QCDevicetest.
IftheQCDevicetestfailsafteryouhavecleanedthedevice,contactQuidel.
QCSample MainMenu
ToruntheQualityControlsample:
RunTest
1. Pressthe keytopowertheQuidelTriageMeterProon.
2. SelectRUNTESTusingthe keys.
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODECHIP
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
TheMainMenu
3. Pressthe key.
4. Typeinyouridentification(UserID).Tocorrect,pressthe keytocleartheentireID,orselectthe
incorrectcharacterbyusingthe keys.ThentypeinthecorrectID.
NOTE:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID.
5. Pressthe key.
6. SelectQCSampleusingthe keys.
PATIENTSAMPLE
QCDEVICE
QCSAMPLE
MISC.TEST
SELECTWITHvOR^ANDPRESSENTER
ORPRESSEXIT TOEJECTDEVICE
MenuscreenasitlookswhenoperatorselectsQCSample.
Page42of86
7. Pressthe key.
8. EntertheQClotnumberfromthelabelonthesideofthebottlevialcontainingtheQCSample.
NOTE:OnlyenterthefourͲdigitnumericvalueoftheQClotnumber.Donotentertheprecedingalpha
character.
9. Tocorrect,pressthe keytocleartheentirenumber.
Thentypeinthecorrectnumber.
NOTE:IfyouenteraQClotnumberforwhichthereisnodataintheQuidelTriageMeterPro’smemory,you
willseeamessageonthemeter’sscreen:NOQCSampledatainmemory.
Tocorrectthis:InstalltheQCSampleCODECHIP(seepictureonpage37).
PressENTERtocontinue.
10.Pressthe key.
11.Performthetestaccordingtotheinstructionsprovidedinthepackageoftestdevicesyouareusing.
IMPORTANT:Importantinstructionsforalltests:
a. GentlyinsertthetestdeviceintotheQuidelTriageMeterPrountilyoufeelthedevicecatchonthe
pin.Youwillhearanaudible‘click’thattellsyouthetestdevicehasbeeninsertedproperly.
b. Pressthe keytostartthetest.
c. TheQuidelTriageMeterPropullsinthetestdeviceandscansit.
d. Whenthetestiscomplete,themeterwillbeepanddisplaytheresultsonthemeter’sscreen.
e. Pressthe keytomakeaprintedcopyoftheresults.
CAUTION:Afterthetestdevicehasbeeninserted,donotpushthedeviceinfurtherorattempt
topullitout.ThedevicemaybeejectedbyreturningtotheMainMenuandpressingEXIT.
12.Repeatstepsforeachqualitycontrolsample.
IMPORTANT:ImportantinstructionsforQCSampleTests
D Ifthetestdevice’sinternalqualitycontrol(QC)resultsareunacceptable,theresultsforthe
affectedanalyte(s)willnotappearonthescreen(insteadwillbeflaggedwitha!symbol).
b. IfanyoftheQCSampleresultsareoutofrange,theresultsforthatparticularanalytewillbein
reversevideo(lighttextonadarkbackground).
c. Apanelwithmultipleanalyteswillstillbeabletoreportpatientresultsonthoseanalyteswhich
passedQC.AnanalytethatfailedQCwillnotbereportedonpatienttests(insteadwillbeflagged
witha#symbol).
Page43of86
d. Ifanindividualanalyteisoutsidethespecifiedrange,theQuidelTriageMeterProallowstheusertorerunonlythefailed
analyteonthenextQCSampleTest.Whenallanalyteshavepassed,theQCSampletimerissettothedateofthefirst
passinganalyte.
PatientSample MainMenu
1. Pressthe keytopowertheQuidelTriageMeterProon. RunTest
NOTE:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID.
NOTE:IftheBarCodeScannerisbeingused,simplypointthescanneratthebarcode,pressthebuttonon
thescannerhandleandwaitforthebeep.TheIDwillappearonthedisplay
Page44of86
PatientSample
MainMenu
10.Iftheinformationiscorrect,pressthe keytoconfirmthePatientID.
RunTest
Iftheinformationisincorrect:
SelectCORRECTPATIENTIDusingthe keys.
Pressthe key.
Pressthe keytocleartheentireID,orselecttheincorrectcharacterusingthe keys.
ThentypeinthecorrectID.
Pressthe key.
11.IftheAUXILIARYIDfeatureisactivated,typeorscanintheID.Tocorrect,pressthe keytoclearthe
entireID,orselecttheincorrectcharacterusingthe keys.ThentypeinthecorrectID.
12.Pressthe key.
13.Preparethetestsampleandperformthetestinaccordancetotheinstructionsprovidedinthepackage
oftestdevicesyouareusing.
14.Ifthetestingfacilityhasauthorizedananalytetobeselectedordeselectedatthetimeofrunninga
patienttest,theuserwillbepromptedtoselectdesiredanalytesordeselectundesiredanalytes.After
thetestdevicehasbeenpulledintothemeter,themeterwilldisplayamessagesimilartothefollowing:
SELECTTESTS
CARDIAC
CKMB2 MYO3 TNI
PRESS^TOPAUSEOR0Ͳ9FORTEST
PRESSENTERTOCONTINUEOREXIT
Acheckmarkindicatestestswhichareselected.Example:TNI
Theabsenceofacheckmarkindicateswhichtestshavenotbeenselected.Example:MYO
NOTE:Atleastoneanalytemustbeselectedinordertocontinuetesting.
Anumbertotheleftofthetestindicatestestswhichcanbeselectedordeselected.
Example:2MYOindicatesthatMYOiscurrentlynotselected,butcanbeselectedbypressing2onthe
numerickeypadtodisplay2MYO.
Example:3TNIindicatesthatTNIiscurrentlyselected,butcanbedeselectedbypressing3onthe
numerickeypadtodisplay3TNI.
Pressingthenumberagaintogglesthetestbetweenselectanddeselect.
Nonumbertotheleftofthetestindicatestestswhichcannotbealtered.
ExampleCKMBindicatesthatCKMBisalwaysselected,andcannotbedeselected.
f Pressthe key.
f Iftheinformationiscorrect,pressthe keytocontinue.
Page45of86
f Iftheinformationisincorrect,selectCORRECTTESTusingthe keysandpressthe key
toreturntothelistsoftests.
NOTE:IfnokeyispressedontheSELECTTESTSscreen,themeterwaits30secondsthenproceeds
usingonlytheselecteddefaulttests.Then,ifnokeyispressedontheCONFIRMATIONSCREEN,the
testproceedsafterwaiting30seconds.
IMPORTANT: Importantinstructionsforalltests:
a. GentlyinsertthetestdeviceintotheQuidelTriageMeteruntilyoufeelthedevicecatchonthe
pin.Youwillhearanaudible‘click’thattellsyouthetestdevicehasbeeninsertedproperly.
b. Pressthe keytostartthetest.
c. TheQuidelTriageMeterPropullsinthetestdeviceandscansit.
d. Whenthetestiscomplete,themeterwillbeepanddisplaytheresultsonthemeter’sscreen.
CAUTION: Afterthetestdevicehasbeeninserted,donotpushthedeviceinfurtherorattempt
topullitout.ThedevicemaybeejectedbyreturningtotheMainMenuandpressingEXIT.
IMPORTANT:Importantinstructionsforpatienttests:
a. Ifthetestdevice’sInternalQualityControl(QC)resultsareunacceptable,theresultsforthe
affectedanalyte(s)willnotappearonthescreen(insteadwillbeflaggedwitha!symbol).
b. IfanyoftheQCSampleresultsforthelotbeingtestedwereoutofrange,theresultsforthat
particularanalytewillnotappearonthescreen(insteadwillbeflaggedwitha#symbol).
PAT.ID89547326 PAT.ID89547326
BNP 15 pg/mL BNP 650 pg/mL
PAT.RESULTABNORMAL
PRESS1TOACCEPTOR0TOREJECT PRESS1TOACCEPTOR0TOREJECT
Screendisplayingpatientsresult Screendisplayingpatientsresultabovethe
belowthecutofflevelandacceptor cutofflevels.
rejectinstructions.
Page46of86
Misc.TestSample
1. Pressthe keytopowertheQuidelTriageMeterProon. MainMenu
RunTest
2. SelectRUNTESTusingthe keys.
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODE CHIP
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
NOTE:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID.
PATIENTSAMPLE
QCDEVICE
QCSAMPLE
MISCTEST
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
NOTE:IftheBarCodeScannerisbeingused,simplypointthescanneratthebarcode,pressthebutton
onthescannerhandleandwaitforthebeep.TheIDwillappearonthedisplay.
Page47of86
Pressthe keytocleartheentireID,orselecttheincorrectcharacterusingthe keys.
ThentypeinthecorrectID.
Pressthe key.
11.Preparethetestsampleandperformthetestinaccordancetotheinstructions
providedinthepackageoftestdevicesyouareusing.
IMPORTANT:Importantinstructionsforalltests:
a. GentlyinsertthetestdeviceintotheQuidelTriageMeteruntilyoufeelthedevicecatchonthepin.
Youwillhearanaudible‘click’thattellsyouthetestdevicehasbeeninsertedproperly.
b. Pressthe keytostartthetest.
c. TheQuidelTriageMeterPropullsinthetestdeviceandscansit.
d. Whenthetestiscomplete,themeterwillbeepanddisplaytheresultsonthemeter’sscreen.
e. Pressthe keytomakeaprintedcopyoftheresults.
CAUTION: Afterthetestdevicehasbeeninserted,donotpushthedeviceinfurtherorattempt
topullitout.ThedevicemaybeejectedbyreturningtotheMainMenuandpressingEXIT.
IMPORTANT:Importantinstructionsformisc.tests:
a. Ifthetestdevice’sInternalQualityControl(QC)resultsareunacceptable,theresultsforthe
affectedanalyte(s)willnotappearonthescreen(insteadwillbeflaggedwitha!symbol).
Page48of86
RecallResults
Purpose
RecallResultsisusedtoretrieveresultsstoredintheQuidelTriageMeterPro’smemory.
Function
TheRecallResultsfunctioncanretrievetheseresults:
LastRecord QCResults
PatientResults f QCSampleResults
f PatientID f QCDeviceResults
f PanelType f ReagentLots–QC
f TestDeviceLotNumber(L/N) f ArchiveLastTest
f UserID Misc.TestResults
f FromDate PrintallResults
f TillDate f PatientResults
f QCSampleResults
f QCDeviceResults
f Misc.TestResults
UploadResultstoLIS
NOTE:Ifaresultwasrejectedbytheuser,an“R”willappearnexttotherecalledresult.
IMPORTANT:TosafeguardtheUserIDLockout,displayedandprintedUserID'sarepartially
masked.TheSupervisorCODECHIPmoduleisrequiredtobeinstalledtoviewthefullUserID.
LastRecord MainMenu
Purpose:LastRecordisusedtoretrievethetestresultsforthelasttestrun. RecallResults
Toretrievethelastrecord:
1. Pressthe keytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselectRECALLRESULTSusingthe keys.
3. Pressthe key.
4. SelectLASTRECORDusingthe keys.
5. Pressthe key.
6. Ifthelasttestrunisapatientresult,typeorscaninyouridentification(UserID).Tocorrect,press
tocleartheentireID,orselecttheincorrectcharacterbyusingthe keys.Thentypeinthe
correctID.Pressthe key.
7. Theresultsfromthelasttestrunappearonthescreen.
8. Usethe keystoviewtheentirerecord.
9. Pressthe keytomakeaprintedcopy.
NOTE:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID
Page49of86
PatientResults
Options:Patientresultscanberetrievedbasedonavarietyofcriteria: MainMenu
PatientID RecallResults
TestDeviceLotNumber(L/N)
UserID
FromDate
TillDate
MainMenu
PatientID
RecallResults
ToretrieveresultsbasedonthepatientID:
PatientResults
1. Pressthe keytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselectRECALLRESULTSusingthe keys.
3. Pressthe key.
4. SelectPATIENTRESULTSusingthe keys.
5. Pressthe key.
6. Typeorscaninyouridentification(UserID).Tocorrect,press tocleartheentireID,orselectthe
incorrectcharacterbyusingthe keys.ThentypeinthecorrectID.
NOTE:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID
NOTE:Youmaychoosemorethanoneoftheoptionsforrecallingresultsinordertodefinemoreprecisely
theinformationyouwanttorecallfromtheQuidelTriageMeterPro’smemory.Whenyouchoosemultiple
options,firstinputvaluesinallthecategoriesyouhavechosen.ThenpresstheENTERkey.
Page50of86
TestDeviceLotNumber(L/N)
Toretrieveresultsbasedonthelotnumberofthetestdevicesused:
MainMenu
RecallResults
PatientResults
1. Pressthe keytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselectRECALLRESULTSusingthe keys.
3. Pressthe key.
4. SelectPATIENTRESULTSusingthe keys.
NOTE:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID.
NOTE:OnlyenterthefiveͲdigitnumericvalueofthelotnumber.
13.Pressthe key.
14.Allresultsfromtestsrunontherequestedlotnumberfortheselectedpanelaredisplayed.
15.Usethe keystoviewtheentirerecord.
16.Pressthe keytomakeaprintedcopy.
17.Pressthe keytouploadtheselectedresultstoLIS.
Page51of86
UserID MainMenu
ToretrieveresultsfromtestsrunontheQuidelTriageMeterProstartingwith RecallResults
theuseryouselect: PatientResults
orselecttheincorrectcharacterbyusingthe keys.ThentypeinthecorrectID.
NOTE:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID
13. Pressthe key.
14. Allresultsfromtestsrunbytherequesteduserfortheselectedpanelaredisplayed.
15. Usethe keystoviewtheentirerecord.
16. Pressthe keytomakeaprintedcopy.
17. Pressthe keytouploadtheselectedresultstoLIS.
FromDate MainMenu
ToretrieveresultsfromtestsrunontheQuidelTriageMeterProstartingwith RecallResults
adateyouselect: PatientResults
1. Pressthe keytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselectRECALLRESULTSusingthe keys.
3. Pressthe key.
Page52of86
4. SelectPATIENTRESULTSusingthe keys.
5. Pressthe key.
6. Typeorscaninyouridentification(UserID).Tocorrect,press tocleartheentireID,orselectthe
incorrectcharacterbyusingthe keys.ThentypeinthecorrectID.
NOTE:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID.
7. Pressthe key.
8. SelectPANELTYPEusingthe keys.
9. Usethe keystoselectthedesiredpanel.
10. SelectFROMDATEusingthe keys.
11. Todisplayallresultsfortheselectedpanel,pressthe key.
12. Todisplayresultsspecifictoonedate,typeinthedate.Tocorrect,pressthe keytocleartheentire
number.Thentypeinthecorrectnumber.
13. Pressthe key.
14. Allresultsfromtestsrunfromthedateselectedfortheselectedpanelaredisplayed.
15. Usethe keystoviewtheentirerecord.
16. Pressthe keytomakeaprintedcopy.
17. Pressthe keytouploadtheselectedresultstoLIS.
TillDate
MainMenu
ToretrieveresultsfromtestsrunontheQuidelTriageMeterProbeforeandincluding
RecallResults
adateyouselect:
PatientResults
1. Pressthe keytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselectRECALLRESULTSusingthe keys.
3. Pressthe key.
4. SelectPATIENTRESULTSusingthe keys.
5. Pressthe key.
6. Typeorscaninyouridentification(UserID).Tocorrect,press tocleartheentireID,orselectthe
incorrectcharacterbyusingthe keys.ThentypeinthecorrectID.
NOTE:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID
7. Pressthe key.
Page53of86
8. Select3$1(/7<3(usingthe keys.
9. Usethe keystoselectthedesiredpanel.
10. Select7,//'$7(usingthe keys.
11. Todisplayallresultsfortheselectedpanel,pressthe key.
12. Todisplayresultsspecifictoonedate,typeinthedate.Tocorrect,pressthe keytocleartheentire
number.Thentypeinthecorrectnumber.
13. Pressthe key.
14. Allresultsfromtestsrunbeforeandincludingthedateyouselectedfortheselectedpanelare
displayed.
15. Usethe keystoviewtheentirerecord.
16. Pressthe keytomakeaprintedcopy.
17. Pressthe keytouploadtheselectedresultstoLIS.
QCResults
Options:Theresultsofthesequalitycontrol(QC)testscanberetrievedfromthe MainMenu
QuidelTriageMeterPro’smemory: RecallResults
QualityControl(QCtest) Resultscanberetrievedby:
QCSampleResults PanelType
QCSampleLotNumber(L/N)
DeviceLotNumber(L/N)
UserID
FromDate
TillDate
ViewAll(Default)
QCDeviceResults UserID
FromDate
TillDate
ViewAll(Default)
ReagentLots–QC (DefaultstoViewAll)
ToretrieveQCResultsfortheQCSampleandtheQCDevice,selecttheoptiondesiredandfollowthe
directions.
NOTE:Youmaychoosemorethanoneoftheoptionsforrecallingresultsinordertomoreprecisely
definetheinformationyouwanttorecallfromtheQuidelTriageMeterPro’smemory.Whenyouchoose
multipleoptions,firstmakeselectionsandinputvaluesinallthecategoriesyouhavechosen.Thenpress
theENTERkey.
Page54of86
QCSampleLotNumber(L/N) MainMenu
ToretrieveresultsbasedonthelotnumberoftheQCSampleused: RecallResults
1. Pressthe keytopowertheQuidelTriageMeterProon. QCResults
QCSampleResults
2. FromtheMAINMENUselectRECALLRESULTSusingthe keys.
3. Pressthe key.
4. SelectQCRESULTSusingthe keys.
5. Pressthe key.
6. SelectQCSAMPLERESULTSusingthe keys.
7. Pressthe key.
8. SelectPANELTYPEusingthe keys.
9. Usethe keystoselectthedesiredpanel.
NOTE:Todisplayallresultsfortheselectedpanel,leavetheQCSampleL/Nblank.
DeviceLotNumber(L/N) MainMenu
ToretrieveQCSampleresultsbasedonthelotnumberofthetestdevicesused: RecallResults
QCResults
QCSampleResults
1. Pressthe keytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUSELECTRECALLRESULTSusingthe keys.
Page55of86
3. Pressthe key.
4. SelectQCRESULTSusingthe keys.
5. Pressthe key.
6. SelectQCSAMPLERESULTSusingthe keys.
7. Pressthe key.
8. SelectPANELTYPEusingthe keys.
9. Usethe keystoselectthedesiredpanel.
10. SelectDEVICEL/Nusingthe keys.
11. Typeinthedevicelotnumberfromthesideoftheboxoftestdevices.Tocorrect,pressthe keyto
cleartheentirenumber.Thentypeinthecorrectnumber.
12. Pressthe key.
13. AllQCresultsthathavebeenrunontherequestedlotnumberfortheselectedpanelaredisplayed.
14. Usethe keystoviewtheentirerecord.
15. Pressthe keytomakeaprintedcopy.
16. Pressthe keytouploadtheselectedresultstoLIS.
MainMenu
UserID RecallResults
ToretrieveQCresultsfromtestsrunontheQuidelTriageMeterProstarting QCResults
withtheuseryouselect: QCSampleResults
1. Pressthe keytopowertheQuidelTriageMeterProon. QCDeviceResults
FromtheMAINMENUselectRECALLRESULTSusingthe
keys.
Page56of86
15.Usethe keystoviewtheentirerecord.
16.Pressthe keytomakeaprintedcopy.
17.Pressthe keytouploadtheselectedresultstoLIS.
FromDate MainMenu
ToretrieveQCresultsfromtestsrunontheQuidelTriageMeterProstarting RecallResults
withadateyouselect: QCResults
1. Pressthe keytopowertheQuidelTriageMeterProon. QCSampleResults
QCDeviceResults
2. FromtheMAINMENUselectRECALLRESULTSusingthe keys.
3. Pressthe key.
4. SelectQCRESULTSusingthe keys.
5. Pressthe key.
6. Selecttheappropriatemenu,QCSAMPLERESULTSorQCDEVICERESULTSusingthe keys.
7. Pressthe key.
8. IfQCSAMPLERESULTS,selectPANELTYPEusingthe keys.Usethe keystoselectthe
desiredpanel.
9. SelectFROMDATEusingthe keys.
10. Pressthe key.
11. Typeinthedate.Tocorrect,pressthe keytocleartheentirenumber.Thentypeinthecorrect
number.
12. Pressthe key.
13. Allresultsfromtestsrunfromthedateselected,fortheselectedpanelifQCSample,aredisplayed.
14. Usethe keystoviewtheentirerecord.
15. Pressthe keytomakeaprintedcopy.
16. Pressthe keytouploadtheselectedresultstoLIS.
TillDate
ToretrieveQCresultsfromtestsrunontheQuidelTriageMeterProbeforeand
includingadateyouselect: MainMenu
RecallResults
1. Pressthe keytopowertheQuidelTriageMeterProon. QCResults
2. FromtheMAINMENUselectRECALLRESULTSusingthe keys. QCSampleResults
QCDeviceResults
3. Pressthe key.
4. SelectQCRESULTSusingthe keys.
5. Pressthe key.
6. Selecttheappropriatemenu,QCSAMPLERESULTSorQCDEVICERESULTSusingthe keys.
7. Pressthe key.
8. IfQCSAMPLERESULTS,selectPANELTYPEusingthe keys.Usethe keystoselectthe
desiredpanel.
Page57of86
9. SelectTILLDATEusingthe keys.
10. Pressthe key.
11. Typeinthedate.Tocorrect,pressthe keytocleartheentirenumber.Thentypeinthecorrect
number.
12. Pressthe key.
13. Allresultsfromtestsrunbeforeandincludingthedateselected,fortheselectedpanelifQCSample,are
displayed.
14. Usethe keystoviewtheentirerecord.
15. Pressthe keytomakeaprintedcopy.
16. Pressthe keytouploadtheselectedresultstoLIS.
ReagentLots–QC MainMenu
AlistofReagentCODECHIPmodulesinmemory,alongwithcorresponding RecallResults
expirationdatesforaQCSamplecanberecalledfrommemory. QCResults
ReagentLots–QC
ArchiveLastTest
Thelistwillcontain:LotNumber,PanelType,ExpirationDate
IftheQCexpirationdatehaspassed(expired),thedatewillbeinreversevideo(lighttextonadark
background).
IftheQCforthelothasfailed,thewordFAILEDwillappearinplaceofanexpirationdate.
IfaCODECHIPhasbeeninstalledforanewlot,butQChasyettobeestablished,thewordsNOTRUN
willappearinplaceoftheexpirationdate.
IfQCfrequencyissettonone,thereagentlotexpirationdatewillbedisplayed.
REAGENTLOTSͲQC
DLN PANEL EXP
47067 CARDIAC 12Ͳ16Ͳ17
48741 CARDIAC 09Ͳ04Ͳ18
48197 BNP NOTRUN
47827 DRUGSCREEN FAILED
PRESSPRINTORPRESSENTER
TheReagentLots–QCScreen
ArchiveLastTest
ThisfunctionwillsavedetailedinformationregardingthelastperformedtestforanalysisbyQuidel.Usethis
functionwhenrequestedbyQuidel.
Thisfunctionshouldbeusedwhenaveryunusualtestresultisreceivedandwhenthelaboratorywishesto
havethemeterevaluatedfurther.ARCHIVELASTTESTwillonlybeabletosavedatafromthemostrecent
testrun.
Page58of86
Misc.TestResults MainMenu
Options: Misc.Testresultscanberetrievedbasedonavarietyofcriteria: RecallResults
Misc.TestID
TestDeviceLotNumber(L/N)
UserID
FromDate
TillDate
Misc.TestID MainMenu
ToretrieveresultsbasedontheMisc.TestID: RecallResults
MiscTestResults
1. Pressthe keytopowertheQuidelTriageMeterProon.
NOTE:Youmaychoosemorethanoneoftheoptionsforrecallingresultsinordertodefinemoreprecisely
theinformationyouwanttorecallfromtheQuidelTriageMeterPro’smemory.Whenyouchoosemultiple
options,firstinputvaluesinallthecategoriesyouhavechosen.ThenpresstheENTERkey.
IfyouwanttoseeallMisc.Testresults,pressthe key.AlltheMisc.TestresultsintheQuidelTriage
MeterPro’smemoryfortheselectedpanelwillbedisplayed.
ForindividualMisc.Testresults:
a. SelectMISC.TESTIDusingthe keys.
Page59of86
TestDeviceLotNumber(L/N) MainMenu
Toretrieveresultsbasedonthelotnumberofthetestdevicesused: RecallResults
Pressthe keytopowertheQuidelTriageMeterProon. MiscTestResults
10. Todisplayresultsspecifictoonelot,typeinthelotnumberofthetestdevice.Tocorrect,pressthe
keytocleartheentirenumber.Thentypeinthecorrectnumber.
NOTE:OnlyenterthefiveͲdigitnumericvalueoftheQClotnumber.
Page60of86
UserID
ToretrieveresultsfromtestrunontheQuidelTriageMeterProstartingwith MainMenu
theuseryouselect: RecallResults
MiscTestResults
1. Pressthe keytopowertheQuidelTriageMeterProon.
FromDate MainMenu
ToretrieveresultsfromtestrunontheQuidelTriageMeterProstartingwithadate RecallResults
youselect: MiscTestResults
Pressthe keytopowertheQuidelTriageMeterProon.
Page61of86
9. Todisplayallresultsfortheselectedpanel,pressthe key.
TillDate MainMenu
ToretrieveresultsfromtestrunontheQuidelTriageMeterProbefoeandincluding RecallResults
adateyouselect: MiscTestResults
Page62of86
PrintAllResults MainMenu
Options:AllstoredresultscanbeprintedfromtheQuidelTriageMeterPro’s RecallResults
memory. MiscTestResults
PatientResults
QCSampleResults
QCDeviceResults
Misc.TestResults
PatientResults MainMenu
ToprintallPatientResults: RecallResults
PrintAllResults
1. Pressthe keytopowertheQuidelTriageMeterProon.
NOTE:Whenprintinglargequantitiesofpatientresults,itisrecommendedtofirstinsertafullrollof
paper.
NOTE:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID.
QCSampleResults
ToprintallQCSampleResults:
1. Pressthe keytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselectRECALLRESULTSusingthe keys.
3. Pressthe key.
4. SelectPRINTALLRESULTSusingthe keys.
5. Pressthe key.
6. SelectQCSAMPLERESULTSusingthe keys.
7. Pressthe keytomakeaprintedcopy.
Page63of86
QCDeviceResults MainMenu
ToprintallQCDeviceResults: RecallResults
PrintAllResults
1. Pressthe keytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselectRECALLRESULTSusingthe keys.
3. Pressthe key.
4. SelectPRINTALLRESULTSusingthe keys.
5. Pressthe key.
6. SelectQCDEVICERESULTSusingthe keys.
7. Pressthe keytomakeaprintedcopy.
Misc.TestResults
ToprintallMisc.TestResults:
1. Pressthe keytopowertheQuidelTriageMeterProon.
UploadResultstoLIS MainMenu
WheninterfaceddirectlytoanLIS,selectingthisfeaturewillsendalltestresults RecallResults
totheLIS.Priortothefirstupload,theLISfeaturemustbeactivatedfromthe UploadResultstoLIS
SetParametermenuitemusingtheSupervisorCODECHIPmodule.
DeleteResults
Purpose
TheDeletefunctionisusedtoremoveallpatienttestinformationfromtheQuidelTriageMeterPromemory.
NOTE:ThedeletefunctionisavailableonlywhentheSupervisorCODECHIPmoduleisinsertedinthe
QuidelTriageMeterProCODECHIPmodulePort.
Page64of86
THEQuidelTriageMeterProMemory
WithintheQuidelTriageMeterPro,PatientResultmemorycanhold750datasets,QCSamplecanhold200
datasets,QCDevicecanhold70datasets,andMisc.Testcanhold250datasets.Whenthememoryisfull,
theoldestresultisautomaticallyoverwrittenwhenanewresultisadded.Themeterdisplayalertstheuser
asthepatientmemorybecomesfull.
WARNING
PATIENTMEMORYHAS
SPACEFORXXXMORE
RECORDS
PRESSENTER
TheMemoryAlertScreen
IMPORTANT:TheDeletefunctionpermanentlyremovesallpatientdatafromthemetermemory.
Beforecontinuing,youmaywishtofirstprintresults(oruploadtoyourdatamanagement
system/LIS).QCorMisc.TestResultscannotbemanuallydeleted.
DeletePatientResult MainMenu
Thisfunctionwilldeleteallpatientresultsinmetermemory.Sincethisactionisnot DeleteResults
reversible,themeterwillrequiretwoconfirmationsofintenttodeleteresults.
1. InserttheSupervisorCODECHIPmoduleintheCODECHIPmodulePort
2. FromtheMAINMENU,selectDELETERESULTSusingthe keys.
Page65of86
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODECHIP
SETPARAMETERS
DELETERESULTS
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
TheSupervisor’sMainMenu
NOTE:PresstheEXITkeytocancelatanytimebeforestep7.
Page66of86
AlphaNumericMode
TheQuidelTriageMeterProhasthecapabilityfortheoperatortoenterbothnumericandalphacharacters
fromthekeypad.Thekeypadoperatesinnumericmodeuntilalphamodeisactivatedbypressingthe
key.Themeterstaysinalphamode,indicatedbytheletters"ABC"shownatthetopofthedisplay,untilthe
keyispressedagain.Wheninalphamode,pressinganykeycausesthealphacharactertobe
displayed.
1 Ͳ/#()*.\ 7 PQRS
2 ABC 8 TUV
3 DEF 9 WXYZ
4 GHI 0 noalphacharacterisdisplayed
5 JKL . noalphacharacterisdisplayed
6 MNO
Thefirstpressofabuttoncausesthefirstlettertobedisplayed.Thesecondpressofthesamebuttonwithin
threesecondscausesthesecondlettertobedisplayed.Thethirdpressofthesamebuttonwithinthree
secondscausesthethirdlettertobedisplayed.
Afterthelastcharacterinasequenceisdisplayed,thenextbuttonpressofthesamebuttonwithin3
secondsdisplaysthefirstcharacterandthesequenceisrepeated.
Ifmorethanthreesecondspasswithouttheoperatorpressingthebutton,thenextpressofthatbutton
causesthefirstcharactertobedisplayed.Ifanewbuttonispressed,thefirstcharacterofthenewbuttonis
displayedwithsubsequentpressesrepeatingthesequence.
Page67of86
ServiceandMaintenanceProcedures
Ifanyserviceormaintenanceisrequired,theQuidelTriageMeterProshouldbesenttothemanufacturer.
Nomaintenanceotherthanpaper/batteryreplacementandperiodicexternalcleaningisrequiredofthe
operator.
PaperReplacement
ChangingPaperBetweenFunctions
1. TearoffanyexcesspaperstickingoutoftheQuidelTriageMeterPro.
2. Removethepapercompartmentcoverbypullinguponthecoverasindicatebythearrowontheback
ofthecover.
3. Removeunusedpaperortheemptypaperspindlefrompapercompartment.
4. Tearorcutaclean,straightedgetofeedintotheprinter.Donotcutpaperatanangle,astheprinter
mustsensetheedgeofthepaperalongthefeedpath.
NOTE:Theprintercontainsapapersensorandwillfeedthenewpaperrollonlywhenpaperwithaclean
straightlineispressedintothepaperroller.
5. Insertthenewrollofpaperintothepapercompartment.
6. Positionthepapersuchthatthepaperwillfeedfromundertheroll(asopposedtooverthetopofthe
roll,seepicture,above).
7. Insertthepaperedgeunderthepaperroller(platen)untilitfirmlyseatsorresistanceisfelt.
8. Pressthe key.
NOTE:IfPAPERFEEDbuttonispressedbeforetheprintersensesthepaper,theprinterwillnotrespond
tothecommand.
9. Replacethecoveroftheprinterandcontinueoperation.
NOTE:Theprinterutilizesthermalpaper,iftheprintedpaperisblankafterreplacingaroll,verifythatthe
paperrollisnotupsidedown.
Page68of86
ChangingPaperWhilePrinting
IfthemeterrunsoutofpaperwhileinthemiddleofaprintingtaskandisdisplayingthemessagePRINTER
FAILURE,CHECKPRINTERORREPLACEPAPER,thepapermaybechangedandtheprintjobcontinued
withoutlossofdataorreprintingfromthebeginning.Performthefollowingstepspriortopressinganynew
keys:
1. Removethepapercompartmentcoverbypullinguponthecoverasindicatebythearrowontheback
ofthecover.
2. Gentlypullremainingpaperslowlyupthroughtheprinter.
3. Removepaperspindlefromthepapercompartment.
4. Tearorcutaclean,straightedgetofeedintotheprinter.Donotcutpaperatanangle,astheprinter
mustsensetheedgeofthepaperalongthefeedpath.
NOTE:Theprintercontainsapapersensorandwillfeedthenewpaperrollonlywhenpaperwithaclean
straightlineispressedintothepaperroller.
5. Insertthenewrollofpaperintothepapercompartment.
6. Positionthepapersuchthatthepaperwillfeedfromundertheroll(asopposedtooverthetopofthe
roll,seepicture,above).
7. Insertthepaperedgeunderthepaperroller(platen)untilitfirmlyseatsoraresistanceisfelt.
8. Pressthe key.
NOTE:IfPAPERFEEDbuttonispressedbeforetheprintersensesthepaper,theprinterwillnotrespond
tothecommand.
9. Whentheprinterbeginstopullthepaperin,releasethepaper.TheprinterwillfeedafewlinesandreͲ
commencetheprintjob.Toensurenodataislost,theprinterwillreprintuptotheprevious5lines.
10.Afterthepaperhasappearedabovetheprinter,replacetheprintercover,takingcaretoensurethe
paperdoesnotgetcaughtinsidethemeter.
NOTE:Theprinterutilizesthermalpaper,iftheprintedpaperisblankafterreplacingaroll,verifythatthe
paperrollisnotupsidedown.
Page69of86
Cleaning
TheQuidelTriageMeterProrequiresminimalmaintenance.Occasionalcleaningoftheexteriorwithmild
soapandwatersolutionissufficient.Afterusingadamp,notwet,spongeorclothtoapplyamildsoapand
watersolutionontheoutsideofthemeter,wipethemeterdryusingasoftclothorabsorbenttissue.Donot
allowwatertoseepintotheprinter.Donotimmersethemeterinwaterorotherliquids.
Ifbloodorotherfluidsarenotallowedenoughtimetofullyabsorbintotestdevices,thedevicetrackdoor
mayoccasionallyrequirecleaning.Usingacottonswabdampenedwithisopropylalcoholandapairof
tweezers,carefullyliftthedoorandcleanbothfrontandbackofthedoor.
NOTE:Neverinsertanytools,swabs,orcleaningmaterialsintothedevicetrack.
CheckBatteryLevel
TheQuidelTriageMeterProcheckstheactivepowersourceafterejectingadeviceorprinting.Tocheckthe
batterypower,navigatetotheMAINMENU,disconnecttheACpowersupply,pressthe key(themeter
attemptstoejectthedevice),andlookfortheLowPowerIndicatorontheMainMenuofthemeterdisplay.
Iftheflashingindicatorappears,replacethebatteries.
NOTE:Themeterisdesignedtoperformandprintatleast100testsbeforenewbatteriesarerequiredtobe
replaced.Whenthemeterisnotinuse,itshouldbeturnedoff.Topreservebatterylife,ensuretheAUTO
POWERͲOFFfunctionissetto1/2HOUR.
BatteryReplacement
Turnoffthemeter.
Removethebatterycoverfromthebottomofthemeterbyslidingthecoverinthedirectionofthe
arrow.
Page70of86
Removethebatteries.Somecountriesrequirethatbatteriesbedisposedofinaccordancewithspecific
governmentalregulations(Directive2006/66/EC).Disposeofthebatteriesincompliancewithlocal
regulations.
InsertfoursizeAA1.5Ͳvoltbatteriesintothebatterycompartment.Besurethattheplus(+)andminus
(Ͳ)signsofthebatteriesmatchthesamesignsembossedinthebatterycompartment
Replacethecovertothebatterycompartment.
NOTE:Ifthemeterdoesnotpoweronafterreplacingbatteries,verifythebatteriesarelinedupaccording
tothesymbolsinthebatterycompartment.
NOTE:Rechargeablebatteriesareauthorizedforuse,howeverduetothenatureofthesetypesof
batteries,itisexpectedtheywillrequirerechargingmorefrequently.TheQuidelTriageMeterProdoesnot
containabuiltͲinbatterycharger,rechargeablebatteriesmustberechargedoutsideofthemeter.
TotalQualityAssurance
Introduction
NOTE:Formoredetailedinformation,contactQuidel.
Qualityassuranceprogramscontrolandmonitoranalyticalmethods.Thequalitycontrolofanalytical
methodsachievestwobasicgoals:
1. Ensuresthatthereagentsandinstrumentareworkingproperly
2. Ensuresthattheprocedurewasperformedcorrectly
Themechanismbywhichanalyticalmethodsaretraditionallycontrolledisthroughtheuseofliquidcontrol
specimenscontainingknownamountsofanalytes.
Overthelast15yearsadvancesintechnologyhavechangedthemannerinwhichthediagnosticreagents
areconfigured.AssaysystemsforthePointͲofͲCarearegenerallyunitizedtestingdevicesthataredistinct
fromtraditionallaboratoryͲbasedanalyzers.TheQuidelTriagetestsareunitizedteststhatdonotrequire
Page71of86
additionalreagents.Eachdevicecontainsafullcomplementofreagentsthatarerequiredtoperformthe
immunoassay.
TheQuidelTriagetestshavebeendesignedtoincludeassaycontrolsineachtestdevicethatmonitorthe
assayprocedureandthereagentintegrity.Theassaycontrolspresentauniqueandvaluableextensionto
thetraditionalapproachtoqualitycontrolusingliquidreagents.Iftheproductisstoredunderspecified
conditions,properfunctioningoftheassaycontrolsindicatesthatthetestdeviceisvalidandproper
adherencetoassayprotocolhasbeenfollowed.TheassaycontrolsintheQuidelTriagetestsrequirethe
needtoreͲthinkthefrequencywithwhichexternalqualitycontrolspecimensshouldbetested.
TheQuidelTriagetestshavebeendesignedtoyieldlongͲtermreagentstability.Inaddition,thesystem
providesqualitycontrolparametersinthreeseparateanddistinctcategories:
Internaldevicecontrols
Meterelectroniccontrols
Softwarecontrols.
Thecombinationofthesequalitycontrolfeaturesprovidesmoreinformationaboutthevalidityofeach
assaythancanbeprovidedbythemoretraditionalmethodsofreagentandassayprocedurevalidation.
QualityAssuranceFunctions
1. InternalDeviceQualityAssurance
InternalPositiveControlZone
BaselineControlZone
TimingControlZone
ZoneQualityRequirements(TraceQC)
2. MeterQualityAssurance
SelfͲTestMode
InternalCalibrationChip
ElectronicQCDevice
BarCode
3. SoftwareQualityAssurance
SupervisorCODECHIPmodule
UserID#
PatientID#
Resultmemorystorage
QCFrequency
QCLockouts
4. QualityControlSamples
InternalDeviceQualityAssurance
TheInternalControlZonesarebuiltͲinfunctionalimmunoassays.Thiscontrolverifiesthatthedevice
functionedproperlyandthattheassaywascorrectlyperformed.TheInternalPositiveControlZoneensures
thattheantibodies,thereagentreconstitution,thetimegateandthedeviceflowarefunctioningproperly.
Themanufacturersetsacceptablerangesforthecontrolzonesforeachlotofreagents.
TheBaselineControlZonemonitorsthepresenceofinterferingsubstancesinthepatientsamplethatmay
altertheimmunoassayprocess.Ifaspecimencontainsasubstancethathasamajoreffectonthehighand
Page72of86
lowcontrolsandthebaselinecontrolandtherebyontheimmunoassay,theresultsarenotdisplayedand
erroneousresultsarenotreported.
TheTimingControlZonemonitorsassaycompletion.Ifaninadequateamountofspecimenisaddedtothe
testdeviceorifaspecimenclotsinthedevice,thiscontrolwillpreventthedisplayoftheresults.
ATraceQCalgorithmhasbeenprogrammedwhichevaluatesthesize,locationandsignaltonoiseratioof
eachzoneonthedevice.Aberrationsinzonequalityexceedingpresetlimitswillcauserejectionofoneor
moreoftheassayresults.
MeterQualityAssurance
TheSelfͲTestModeisinitiatedeachtimethemeteristurnedON.IntheselfͲtestmodethemeterscansan
InternalCalibrationChip.Eachcalibrationchipscanisusedtovalidateandadjust,ifnecessary,themeter
calibration.IntheselfͲtestmode,theavailablememoryforpatientresultsischecked.Whenthecapacityis
belowapreͲsetlimit,theremainingcapacityisrelatedviaascreenmessage.Thesoftwareprogramis
evaluatedandisterminatedifcorrupted.SoftwaretestresultsandCODECHIPmoduledataareverifiedfor
integritybeforeeachuse.
AQCDeviceisprovidedwitheachmetertoallowtheusertofurtherchecktheintegrityofthemeter.The
simulatorcontainssixfluorescentzonesofvaryingintensitythataremeasuredbythemeter.TherearepreͲ
programmedacceptablerangesforthesixzones.Inaddition,thezonesallowthemetertocheckfor
horizontalandverticalalignmentofthezonemeasurementtoverifylaseralignmentanddevicetransport.
TheQCDevicealsoconfirmslaserstability.
ABarCodeisprintedonthebottomofeverytestdevicetoidentifythereagentlotnumberandtoprevent
theuseofexpiredreagents.
SoftwareQualityAssurance
ThesoftwareQAiscontrolledthroughavarietyofsoftwarefeatures,someofwhichcanbeoptional
dependingontheenvironmentinwhichtheassayisperformed.InaPointͲofͲCarelocation,adesignated
supervisorcanselectprogramparameterstoensureadherencetoqualitystandards.Inacentrallaboratory
location,selectedparameterscanbebypassedifdesired.
ASupervisorCODECHIPmoduleisprovidedwitheachmeter.ThisCODECHIPmoduleallowsaccessto
meterfunctionsnotavailabletotheroutineuser.Theadditionalfunctionsareselectionofparameter
settingsanddownloadinganddeletionofresultsfromthemetermemory.
AUserIDcanbeassignedinthesoftwaretolimitaccesstothepatienttestingmodetoonlythose
individualswhohavebeenproperlytrainedandhavedemonstratedtestingproficiency.AUserIDand
expirationdateareprogrammedintothemetermemoryandaresubsequentlyassociatedwithallpatient
resultsgeneratedbytheuser.Thisallowsthesupervisortoreviewusers’testingfrequency,andintheevent
offrequenttestfailures,toidentifypotentialproblemsintraining.
AnadditionalsecurityfeatureoftheUserIDisthepartialmaskingoftheIDondisplaysandprintouts.This
preventsunauthorizedindividualsfromdetectingandusinganexistingUserID.Thisfeatureisoverridden
whentheSupervisorCODECHIPmoduleisinstalled.
Page73of86
APatientIDisrequiredforallpatienttestsperformed.AllQCsampleandpatientresultsarestoredinthe
meter’smemoryandcanbedownloadedtoalaboratoryinformationsystem(LIS)oranexternaldata
managementsoftwareprogramforfurtherviewingandprintingofqualitycontrolreports.
TheSupervisorcanselecttheQCFrequencyforperformingexternalliquidcontrolsamples.Whenitistime
toperformexternalQCsamples,theusermusttesttheexternalliquidcontrolsandobtainacceptable
resultsbeforepatientsamplescanbetested.WhenQCsampleshavebeensuccessfullyperformed,theuser
isallowedtoperformtestsonpatientsamples.IfQCsamplesarenotwithinanacceptablerange,aQC
Lockoutfunctionpreventstestingofpatientsamples.Ifdesired,thesupervisorinacentrallabsettingcan
bypassthisfunction.Iftestdevicecontrolsareoutofrange,aQCLockoutfunctionisactivatedandpatient
testresultsareblockedfromviewingandprinting.Thispreventsunacceptableresultstobecommunicated
toandusedbythephysician.
Additionally,iftestdevicesareexpired,aQCLockoutfunctionisactivatedandpreventstestingusing
expiredtestdevices.
QualityControlSamples
Itisstillvaluabletoapplythetraditionalapproachtoqualitycontrolbytestingqualitycontrolsamples.
Thesecontrolswillcheckthetotalintegrityofthesystem.Theintervalforanalyzingthesecontrols,
however,canbeextendedduetothemanyotherTotalQualityAssurancefeaturesinherentintheQuidel
Triagetests.
TheQuidelTriagetestshavebeendesignedtomaximizeTotalQualityAssuranceinanytestingenvironment.
ThecombinationoftheQCfeaturesoftheQuidelTriagetestsreducetheimpactofproceduralerrors,
ensurereagentintegrity,andassurethatpatientresultsareaccurateeachandeverytimeatestis
performed.BasedonthecurrentCLIAguidelinesandotherregulatingbodies,Quidelmakesthefollowing
QCrecommendations:
RuntwolevelsorPOSandNEGexternalliquidcontrolsamplesasappropriatewitheachnewlotof
reagentsandonceeverythirtydayswithcontinueduseofthesamereagentlotnumber
RuntheQCDevicedaily.
NotesRegardingSoftwareControls
1. UserIDAccess
Topreventuntrainedpersonsfromperformingatest,avalidUserID(1Ͳ16charactersinlength)
mustbeenteredbeforeaccessisgiventotheRUNTESTfunctioniftheUserIDbypassisOFF.Ifa
UserIDBypassisON,thenanyonecananalyzeatest.
Topreventunauthorizedpersonsfromaccessingpatientinformation,avalidUserIDmustbe
enteredbeforeaccessisgiventotheRECALLPATIENTRESULTSfunctionsiftheUserIDbypassis
OFF.IfaUserIDBypassisON,thenanyonecanrecallpatientdata.RECALLPATIENTRESULTS
functionsincluderecallingpatientresults,printingpatientresults,andrecallingthelasttestrunon
themeterwhenthelasttestisapatientsample,
TheUserIDBypassiscontrolledbytheSETPARAMETERSfunction,whichrequiresaSupervisor
CODECHIPmoduleforaccess.
AccessisavailabletoanyonewantingtoRECALLNONͲPATIENTRESULTSfrommemoryorINSTALL
CODECHIP.
2. OutͲofͲRangeFlags
Page74of86
Ifapatientisoutsidethenormalrangeforanyoftheanalytes,thentheresultwillbeflaggedby
reversevideo,thatis,thebackgroundoftheresultwillbeblackandthevalueinwhitenumbers.A
warningwillbedisplayedbelowtheresults:
PATIENTRESULTABNORMAL
IftheInternalQCzonesforthedeviceoraspecificanalyteareunacceptable,theanalyte(s)in
questiondisplayanexclamationpoint(!)inplaceofavalueandawarningisdisplayed:
!INTERNALQCOUTOFRANGE.
IftheQCSampleforaspecificanalyteonthatdevicelotisunacceptable,theanalyteinquestionwill
displayapoundsign(#)inplaceofavalueandawarningisdisplayed:
#QCSAMPLEOUTOFRANGE.
IfbothQCSampleandInternalQCerrorsarepresentforaspecificanalyte,bothapoundsignandan
exclamationpoint(#!)willbedisplayedinplaceofavalueandbothwarningsaredisplayed:
!INTERNALQCOUTOFRANGEand#QCSAMPLEOUTOFRANGE.
3. BlockedResults
WhenQCerrorconditionsexistonatestdevice,whetherforaspecificanalyteortheentiretestpanel,
theanalytesinquestionareblockedfromviewandreplacedwithanOutofRangeFlag(!and/or#).The
errorconditionmustberemediedandthetestrepeatedtoobtainaresult.
4. QCSampleFrequency
Howoftenanexternalliquidcontrol,QCSample,shouldberuncanbeselectedunderSET
PARAMETERS.Ifthetimeintervalhasexpired,QCSamplesmustberunandacceptableresults
obtainedbeforeapatientsamplecanbetested.
AlistoftestdevicelotsandthedatethattheQCSampleexpirescanbeviewedandprintedby
selectingREAGENTLOTS–QCundertheRECALLRESULTSmenu.
5. SupervisorAccess
Topreventmisuse,theSupervisorCODECHIPmoduleisrequiredforaccesstotheSETPARAMETERSand
DELETERESULTSfunctionsofthesoftware.TheSupervisorCODECHIPisuniversalandmaybeusedonany
QuidelTriageMeterPro.
LimitedWarranty.FORTHEAPPLICABLEWARRANTYPERIOD,QUIDELWARRANTSTHATEACHPRODUCT
SHALLBE(I)OFGOODQUALITYANDFREEOFMATERIALDEFECTS,(II)FUNCTIONINACCORDANCEWITHTHE
MATERIALSPECIFICATIONSREFERENCEDINTHEPRODUCTMANUAL,AND(III)APPROVEDBYTHEPROPER
GOVERNMENTALAGENCIESREQUIREDFORTHESALEOFPRODUCTSFORTHEIRINTENDEDUSE
(the“LIMITEDWARRANTY”).IFTHEPRODUCTFAILSTOMEETTHEREQUIREMENTSOFTHELIMITED
WARRANTY,THENASCUSTOMER’SSOLEREMEDY,QUIDELSHALLEITHERREPAIRORREPLACE,ATQUIDEL'S
DISCRETION,THEPRODUCT.EXCEPTFORTHELIMITEDWARRANTYSTATEDINTHISSECTION,QUIDEL
DISCLAIMSANYANDALLWARRANTIES,EXPRESSORIMPLIED,INCLUDINGBUTNOTLIMITEDTO,ANY
WARRANTYOFMERCHANTABILITY,FITNESSFORAPARTICULARPURPOSEANDNONͲINFRINGEMENT
REGARDINGTHEPRODUCT.QUIDEL'SMAXIMUMLIABILITYWITHANYCUSTOMERCLAIMSHALLNOT
EXCEEDTHENETPRODUCTPRICEPAIDBYCUSTOMER.NEITHERPARTYSHALLBELIABLETOTHEOTHER
PARTYFORSPECIAL,INCIDENTALORCONSEQUENTIALDAMAGES,INCLUDING,WITHOUTLIMITATION,LOSS
OFBUSINESS,PROFITS,DATAORREVENUE,EVENIFAPARTYRECEIVESNOTICEINADVANCETHATTHESE
KINDSOFDAMAGESMIGHTRESULT.
TheLimitedWarrantyaboveshallnotapplyiftheCustomerhassubjectedtheProducttophysicalabuse,
misuse,abnormaluse,useinconsistentwiththeProductManualorInsert,fraud,tampering,unusual
physicalstress,negligenceoraccidents.AnywarrantyclaimbyCustomerpursuanttotheLimitedWarranty
shallbemadeinwritingwithintheapplicableLimitedWarrantyperiod.
Page75of86
Appendices
AppendixASpecifications
Physical
Size 8.5”x6.25“x2.75”(22.5cmx19cmx7cm)DxWxH
Weight 1.5pounds(0.7kg)withoutbatteries
Electrical 6vDCat1amp–suppliedvia4AAbatteriesorAC/DCConverter
Keypad Numericwithspecialfunctionkeys(22total)
Environmental
Temperature 15ºC–30ºC
Humidity 10%–85%
Location Dry,clean,flathorizontalsurfaceawayfromdirectsunlight
Optical
LightSource LaserDiode–<1milliwatt
LaserClassification 1
Detector SiliconPhotodiode
MemoryCapacity
750PatientRecords 32ReagentLotCODECHIPmodules
200QCSampleResults 200QCSampleCODECHIPmodules
70QCDeviceResults 4QCDeviceCODECHIPmodules
250MiscTestResults
600UserID's
Miscellaneous
RSͲ232computerinterfaceport
ThermalPrinter
LCDscreendisplay
Options
ExternalBarCodeReader(rapidentryofUser,Patient,AuxiliaryorMiscTestID's).
QuidelTriageCensusDataManagementSoftware.
QuidelTriageCensusDataManagementSoftwarewithLISConnectivity.
TestDeviceSpecifications
Testdevicespecificinformationisprovidedintheapplicableproductinsert.Theinstrumentanalyzes
immunoassaytestdevicesmanufacturedbyQuidelCardiovascularInc.
Page76of86
AppendixBTroubleshooting
MeterSymptoms
Symptom ProbableCause CorrectiveAction
a. Inappropriatebodyfluidor a. Retestsampleusinganewtestdevice
InvalidResults
improperlyanticoagulatedsample. withaproperspecimen.
b. Retestsampleonanewtestdevice.
b. Samplesizeincorrect—toolittle Makesuretheentiretubeisfull.Be
sample.Testdevice’sinternalQC certainthelowerbulbonthepipette
controlswillbeoutofrange. tubecontainsasmallamountfluid
beforeyoupressthetopbulb.
c. Iflessthan30minuteshaspassedsince
c. Meterwaspickeduporcarried preparingthetestdevice,placemeter
verticallywhiletestwasrunning. onflatsurfaceandreinsertthetest
Thetestdevice’sinternalQC device.Ifmorethan30minuteshas
controlswillbeoutofrange. passed,retestsampleonnewtest
device.
Ifmeterrepeatedlyfailstopullthedevice
Meterwillnotpull Hookthatpullstestdeviceintothe
in,themetermayneedtobereplaced.
testdevicein. metermaybebroken.
ContactQuidel.
CODECHIPmodule a. CODECHIPmodulemayhavebeen a. ReͲinsertCODECHIPmodulerightside
willnotfitinmeter. insertedupsidedown. up.
b. TrynewCODECHIPmodule,if
b. CODECHIPmoduleorMeter
available.Ifstillnotagoodfit,contact
contactscouldbebent.
Quidel.
Printerworksbutno
printappearson Papermaybeinstalledbackwards. Installpaperreversetocurrentposition.
paper.
Carefullyremoveoldpaperoutofthe
Papermayhavebeenfedintothe
Printerjam. printer.Donotinsertanytoolsintothe
meterimproperly.
printer.
Printerstops
Replacebatteriesorverifypowercordis
printing Batterypowerlow.
installedinthemeterandwalloutlet.
orskipsaline.
Meterwillnot
a. Verifybatteriesareinstalledand
poweronormeter a. Batteriesnotinstalled,drainedor
correctlyaligned(+)and(Ͳ);replace
powersoffwhen incorrectlyaligned.
batteriesifnecessary.
runningatest.
b. Verifypowercordisinstalledinthe
b. Powersourcenotpluggedin.
meterandwalloutlet.
Blinkingbattery
Lowbatterypower. Replacebatteries.
icon.
Metercontainsno
batteriesor
AlldataissavedintheMeterexceptdate
batteriesarelow
andtime.Resetdateandtime.
whenfacilityhas
lossofpower.
Page77of86
AppendixBTroubleshooting
MeterMessageorSymptom
Message ProbableCause CorrectiveAction
Patientsampleconcentrationisabove Seeinstructionsthatcameintheboxof
>XXX.X
thedynamicrange. testdevicesorcontactQuidel.
Patientresultisbelowthedynamic
rangeorisaconcentrationbelowthe Forsomeparameters,a0.0ng/mlpatient
<XX.X
statisticallydeterminedlowesttest resultmaybeappropriate.
concentration.
Replacewith4ͲAAbatteriesorusetheAC
BatteryLow. Batteriesneedreplacement.
poweradapterasapowersource.
Damagedbarcodeormeterhookmay Lookforparticlesonthebarcodeofthe
CannotReadBar
bebroken.(Themeterhookpullsthe testdevice.wipeoffwithadrycloth.
Code
testdeviceintothemeter.) Repeatthetest.
a. • EnsureCODECHIPmoduleis
completely inserted.
• TryareagentCODECHIPmodulefrom
adifferentboxofdevicesofthesamelot
a. CODECHIPmoduleerror.
number.
• IfanewCODECHIPmoduleeliminates
CannotRead
themessage,theoriginalCODECHIPisat
CodeChip.
fault,discardCODECHIPmodule.
b. IfthenewCODECHIPmoduledoesnot
work,installaknownworkingCODE
b. Metererror CHIPmoduletoverifyproperoperation.
IftheCODECHIPmodulefailstowork,
contactQuidel.
Powermeteroffandwait15seconds.
OpticsDetectordidnotproperlypower
DetectorFailure Powermeteron.Ifmessagefailstoclear,
on.
contactQuidel.
a. Inappropriatebodyfluidor a. Retestsampleusingnewdevicewitha
improperlyanticoagulatedsample. properspecimen.
b. Retestsampleonnewdevice.Becertain
b. Samplesizeincorrect–toolittle thelowerbulbonpipetcontainsasmall
sample. amountoffluidandtheentiretubeisfull
beforedispensingcontents.
c. Retestsampleandleavemeteronbench
Measurement c. Meterpickeduporcarriedvertically topwhileperformingassay.Ifmorethan
Failure. whiletestisrunning. 30minutessinceinoculation,useanew
device.
d. Possiblyduetocolddevicesorsample,
highhematocritwithawholeblood
sample,ormechanicalissuewithdevice.
d. Samplerunningslowly
Havecustomerimmediatelyreinsert
deviceintometer(ifwithin30minutes
ofsampleaddition).
Page78of86
Message ProbableCause CorrectiveAction
Motorceasestofunction.
Ejectandinspectthedeviceforanysticky
• Metermaysloworstopduetoobject
substances,labelsontopofthedeviceor
lodgedontheMetertrack.
MotorFailure. otheritemsthatmayinterferewiththe
meter.Ifproblempersists,contact
• Testdevicemaystickonthemeter
Quidel.
track.
InstalltheCODECHIPmoduleincludedin
NoDeviceDatain ReagentCODECHIPmodulehasnotbeen
theboxoftestdevices(theReagent
Memory. installed.
CODECHIPmodule).
a. Powermeteroffthenonagain.If
OPTICFAILUREmessageclears,run
a. Internalstandardisoutof
theQCDevicetoverifyproper
specification.
operation.Ifmessagefailstoclear,
OpticFailure.
contactQuidel.
b. Powermeteroff.Allowmetertositat
b. Meterexposedtoextreme
roomtemperaturefor1–2hours.
temperatures.
Powermeterbackon.
ProgramCheckOK.
PressENTERtostart MessageappearswhentheON/OFF
existingprogramor buttonispressedwhenthemeterhas PresstheENTERkey.
pressEXITtoloada beenwithoutpower.
newprogram.
QCDevice— a. CleanQCDevicewithalintͲfreecloth
a. Maybeduetolintordustaffecting
calibration, orspraywithcannedairandrepeat
reading.
alignmentorlaser thetest.
failed. b. QCDeviceshouldberunregularly
Messages: evenifpatientsamplesarenotrun.
CalibPass(orFail) b. QCDevicewasnotrunregularly.
LaserPass(orFail) c. Ifproblempersists,contactQuidel.
AlignPass(orFail)
Resettimeanddateifrequired.Check
Time/Dateisblinking Metertemporarilylostallpower. theAC/DCpowerconverterplugand
batteries.
Warningindicatesthattheinternal
qualitycontrolonthetestdevicewasout
Warning:InternalQC Repeatsampleusinganewtestdevice.If
ofrange.
OutofRange problempersists,contactQuidel.
Thismaybesamplespecific.
Thismaybedevicerelated.
Options:
Messageisdisplayedwhenmeteris
Warning:Patient a. Printallresults.
turnedonifthereisspaceleftforless
MemoryhasSpace b. Deletepatientresults.
than100patientrecordsorspaceleftfor
for###More c. Noactionisrequiredasthemeter
lessthan20patientrecordswhen
Records. willdeletetheoldestresultoncethe
runningpatientsample.
memoryisfull.
Page79of86
AppendixCReturnandDisposalProcedures
ReturnProcedure
Shouldamalfunctionoccur,contactQuidel.IfQuideldeterminesthatthemetershouldbereturned,youwill
beassignedareturnauthorizationnumberandbeprovidedshippinginstructions.
Immediatelyafterthereturnhasbeenauthorized,QuidelwillsendoutareplacementQuidelTriage
MeterPro.ReturnthemalfunctioningmetertoQuidel.Notethereturnauthorizationnumberonboththe
shippingboxandtheairbillandsendthemeterbacktoQuidelassoonaspossiblefollowingreceiptofthe
replacementmeter.
DisposalProcedure
Theproductmaycomeintocontactwithbloodduringtesting.Usedproductsthereforecarryariskof
infection.Whendisposingofameterthathasbeenutilized,pleasedosoinaccordancetotheregulations
applicableinyourcountry.Forinformationaboutcorrectdisposal,pleasecontactyourlocalcouncilor
authority.TheproductfallsoutsidethescopeoftheEuropeanDirective2002/96/EC(directiveonWaste
ElectricalandElectronicEquipment(WEEE)).
AppendixDGlossary
Analyte:Chemicalsubstancebeingmeasured.
Assay:Test.
Calibration:Comparisonofresulttoareferencestandard.
Character:Onepartofalongerstringoflettersandnumbers;forexample,inthestring"ABC",theA,B,
andCareeachacharacter.
CODECHIPModule:Anyofseveralkindsofchipsbearinginformationthatcanbedownloadedinthe
QuidelTriageMeterPro,includingReagent,QCSample,QCDevice,SupervisorandProgram.
CODECHIPModulePort:ThesmallslotontheunderneathsideoftheQuidelTriageMeterProwherea
CODECHIPmodulecanbeinserted.
Dateformats:DDͲMMͲYY:DayͲMonthͲYear
MMͲDDͲYY:MonthͲDayͲYear
YYͲMMͲDD:YearͲMonthͲDay
NOTE:AllpartsofdatesshouldbegivenastwoͲdigitnumbers.Forexample,themonthofMayis05.
DeviceL/N:Lotnumberofatestdevicecanbefoundonthesideoftheboxthatcontainsthetestdevices
andoneachtestdevice.ThedeviceL/NisafiveͲdigitnumberthatmaybeprecededbyoneormoreletters.
Themeterdisregardsallletters.
Digit:Onepartofalongernumber;forexample,inthenumber231,the2,3,and1areeachadigit.
Fluorescence:Thecharacteristicofachemicalsubstancethatenablesittogiveofflightwhenstimulated.
Page80of86
ID:Identification.
Immunoassay:Atestthatusesantibodiestomeasuresubstances.
InternalQC:QualitycontrolsinplaceaspartoftheQuidelTriageMeterPro’ssoftwareorzonesbuiltinto
thetestdevice.
Invitro:Madetooccurinalaboratoryvesselorothercontrolledexperimentalenvironmentratherthanin
alivingorganism(literally,inglass).
Invitrodiagnosticuse:Foruseinacontrolledsetting.
Laserdiode:LightsourceusedintheQuidelTriageMeterPro.
LCD:Liquidcrystaldisplay;thescreenontheQuidelTriageMeterPro.Showsthemenuofpossibletestsand
proceduresandpromptstheoperatortotakethenextstep.
Lotnumber:L/N;identificationnumbergiventoagroupoftestdevices.
L/N:Lotnumber;thenumberthatidentifiesabatchof
testdevices.
Menu:ListofchoicesdisplayedontheQuidelTriage
MeterPro’sscreen.
MMX:TheMultiMarkerIndexvalue,isanumber
calculatedfromthemeasuredvaluesoftheanalytesin
aTestDevice.NotallQuidelTriagetestpanelshave
theMMXfeature.
Parameters:Options;functions
Photodiode:DetectsthelightemittedfromthefluorescentdyeinthetestdevicesusedintheQuidelTriage
MeterPro.
PointofCare(POC):Notinalaboratory;attheplacewherecareisbeinggiventoapatient.
QC:QualityControl.
QCSample:Externalcontrols;aliquidsolutioncontainingchemicalsthat,whendispensedintoatest
device,reactwiththetestreagents.Usedtoverifyproperperformanceoftestdevicesaftershipmentor
longperiodsofstorage.
QCSampleCODECHIPmodule:CODECHIPmodulethatcanbedownloadedintotheQuidelTriage
MeterPromemorytogivethemeterinformationneededaboutthespecificlotofQCSample.
QCDevice(previouslycalledQCSimulator):
Testdevicecontaining6independentzonesthatverifiesproperoperationofthe
QuidelTriageMeterPro’sabilitytoproperlyreadpatienttestsacrossvaryinganalyte
concentrations.
QCDeviceCODECHIP:CODECHIPmodulethatcanbedownloadedintotheQuidel
TriageMeterPro’smemorytogivethemeterinformationneededaboutthespecificQCDevice’s
performanceintheassignedmeter.
Page81of86
Quantify,quantifying:Togivetheresultsofatestasnumber.
Reagent:Substancethatcauseschemicalreactions;usedinanalysis.
ReagentCODECHIPmodule:CODECHIPTMmoduleloadedwithinformationaboutaspecifictestandlot
number.
Retrieve:Find,bringback.
ReverseVideo:Lighttextdisplayedonadarkbackground.
Screen:ThepartoftheQuidelTriageMeterProthatdisplayswordsandnumbers.
Scrollto:Usearrowkeys
togototheinformationyouwantonthescreen.
SetParameters:AfunctionoftheQuidelTriageMeterPro;requirestheuseoftheSupervisorCODECHIP
module.
SoftwareUpgradeCODECHIP:
CODECHIPmoduleusedtodownloadnewsoftwareontotheQuidelTriageMeterPro.
SupervisorCODECHIPmodule:TheCODECHIPmodulethatallowsasupervisortoset
parametersanddeleteresults.
Testdevice:Holdsthesample(forexample,blood)fromthepatientsothatitcanbeinserted
intotheQuidelTriageMeterProfortesting.Containsreagentzonesthatcausereactionsto
occurthatallowthemetertodetermineanalytelevelsinthepatient.
Testpanel:AlternatenameforTestDevice,usuallyindicativeofaTestDevicewithmultipleanalytes.
AppendixESampleLogSheets
Thefollowinglogsheetsareprovidedasalternatemethodsoftrackingtestresults.
TheQuidelTriageMeterProiscapableoftrackingandprintingouttesthistoryandprovidingQCLockouts
topreventpatienttestingwhenoutsideQCrequirements.
Page82of86
Page83of86
Page84of86
Page85of86
55070–QuidelTriageMeterPro
55071–QuidelTriageMeterPro
MDSSGmBH
Schiffgraben41
30175Hannover,
Germany
QuidelCardiovascularInc.
9975SummersRidgeDrive ENSRC26611enA
SanDiego,CA92121USA PN:26611enRev.A2018/03
quidel.com
RevisionChanges:
InitialreleaseforQuidelCardiovascularInc.