AM1660/1640/1620 Precision Industrial PRTs

OVERVIEW

The AM1660/1640/1620 series


Precision Industrial PRTs (IPRTs) are
rugged probes with excellent
accuracy and stability. These IPRTs
cover a wide range of temperatures
from -200°C to 670°C, with an
FEATURES amazing accuracy of ±0.035°C and a
Temperature range: -200°C to 670°C short term stability of ±0.01°C.
Accuracy: <0.035°C
Long term drift: <0.03°C To reach the best performance in
Short term stability: <0.01°C stability and repeatability, the wire-
Durable and shock resistant
wound sensing element has been
Temperature Coefficient 0.00385 Ω/Ω/°C
specially designed to protect the
Follows DIN/IEC-751 curve precisely
Inconel™ sheath on AM1660/1640 models platinum sensing wire from
contamination at high
temperatures. A unique supporting
structure and filling material
MULTIPLE SIZE OPTIONS AVAILABLE provide the best balance among the
Model Length OD Inconel™ hysteresis effect, mechanical shock
and thermal shock performance.
1660 12in 0.25in ✓
This probe conforms to the
1640-12 12in 0.25in ✓
standard 385 curve so the
1640-9 9in 0.187in ✓
resistance ratio of the PRT follows
1620-12 12in 0.25in
the DIN/IEC-751 curve precisely.
1620-9 9in 0.187in

Address: 90 N William Dillard Drive C-107, Gilbert, AZ 85233 For Technical Help: info@accumac.com
Phone: 480-634-0603 For Sales: sales@accumac.com
Web: www.accumac.com
© 2020 AccuMac Technology Inc, All Rights Reserved
AM1660/1640/1620 Precision Industrial PRTs

SPECIFICATIONS
1660-12 1640-12 1640-9 1620-12 1620-9

Temperature Range -200°C to 670°C -200°C to 420°C -60°C to 300°C

Nominal Resistance at 0°C 100 Ω

Temperature Coefficient 0.00385 Ω/Ω/°C

Accuracy <0.035°C at 0°C

Long Term Drift at 0.01°C* <0.03°C at 0°C after 100 hours at maximum temperature

Short Term Stability <0.01°C

Thermal Shock <0.007°C after 10 thermal cycles from minimum to maximum temperatures

Hysteresis ≤0.01°C

Self-heating 50 mW/°C

Response Time** 5 seconds

Measurement Current 0.5 mA or 1 mA

Sensor Length 32 mm

Sensor Location 5 mm from tip

Insulation Resistance >1000 MΩ at room temperature

Sheath Material Inconel™ Stainless Steel

Sheath Dimensions 0.25in (OD) x 12in (L) 0.25in (OD) x 12in (L) 0.187in (OD) x 9in (L) 0.25in (OD) x 12in (L) 0.187in (OD) x 9in (L)

External Leads Teflon™ insulated copper wire, 4 leads, 2 meters

Handle Dimension 15mm (OD) x 65mm (L) 15mm (OD) x 65mm (L) 10mm (OD) x 50mm (L) 15mm (OD) x 65mm (L) 10mm (OD) x 50mm (L)

Handle Temperature Range -50°C to 180°C

*Long-term drift rate is for reference only. It could be affected by such facts as handling, application, and maintenance, etc.
**for 63% response to step change in water moving at 3 feet per second

ORDERING OPTIONS
Model Description

9001 Wooden carrying case for 12” and 9” probe

5007 NIST traceable calibration

5033 ISO 17025 accredited calibration

Termination Bare Wire (default) Spade 8010 Connector ADT LEMO Connector
Options

Address: 90 N William Dillard Drive C-107, Gilbert, AZ 85233 For Technical Help: info@accumac.com
Phone: 480-634-0603 For Sales: sales@accumac.com
Web: www.accumac.com
© 2020 AccuMac Technology Inc, All Rights Reserved

You might also like