Piccolo Xpress Chemistry Analyser SOP
Piccolo Xpress Chemistry Analyser SOP
Piccolo Xpress Chemistry Analyser SOP
1
No. Issue Date 09/03/2018
1. TABLE OF CONTENTS
1. TABLE OF CONTENTS..................................................................................2
2. HAZARDS AND PRECAUTIONS....................................................................3
2.1. Procedure Risk Assessment.................................................................3
2.2. Chemical...............................................................................................3
2.3. Biological Hazard..................................................................................3
2.4. Physical Hazard....................................................................................4
2.5. Room Risk Assessment (if required).....................................................4
3. CONTENT.......................................................................................................5
A) PURPOSE OF THE EXAMINATION......................................................................5
B) PRINCIPLE AND METHOD OF THE PROCEDURE USED FOR EXAMINATIONS
................................................................................................................................... 5
C) PERFORMANCE CHARACTERISTICS (SEE 5.5.1.2 AND 5.5.1.3)......................5
D) TYPE OF SAMPLE (E.G. PLASMA, SERUM, URINE)..........................................5
E) PATIENT PREPARATION.....................................................................................5
F) TYPE OF CONTAINER AND ADDITIVES..............................................................5
G) REQUIRED EQUIPMENT AND REAGENTS.........................................................5
H) ENVIRONMENTAL AND SAFETY CONTROLS....................................................5
I) CALIBRATION PROCEDURES (METROLOGICAL TRACEABILITY).....................5
J) PROCEDURAL STEPS..........................................................................................5
K) QUALITY CONTROL PROCEDURES...................................................................5
L) INTERFERENCES (E.G. LIPAEMIA, HAEMOLYSIS, BILIRUBINEMIA, DRUGS)
AND CROSS REACTIONS.........................................................................................5
M) PRINCIPLE OF PROCEDURE FOR CALCULATING RESULTS INCLUDING,
WHERE RELEVANT, THE MEASUREMENT UNCERTAINTY OF MEASURED
QUANTITY VALUES..................................................................................................6
N) BIOLOGICAL REFERENCE INTERVALS OR CLINICAL DECISION VALUES.....6
O) REPORTABLE INTERVAL OF EXAMINATION RESULTS....................................6
P) INSTRUCTIONS FOR DETERMINING QUANTITATIVE RESULTS WHEN A
RESULT IS NOT WITHIN THE MEASUREMENT INTERVAL....................................6
Q) ALERT/CRITICAL VALUES, WHERE APPROPRIATE........................................6
R) LABORATORY CLINICAL INTERPRETATION.....................................................6
S) POTENTIAL SOURCES OF VARIATION...............................................................6
T) REFERENCES.......................................................................................................6
2.2. Chemical
Please note – only chemicals which present a risk when used in this procedure need to be
included in this table.
Chemical Risk Precautions Storage and Discard 1st Aid Measures MSDS reference
Requirements (Q-pulse)
Wear gloves when Store in fridge and Fush exposed skin POCT-SDS4
handling rotor discard into a with copious amounts
sharpsafe™ of water for at least 15
Skin irritant minutes
lyophylised chemical
beads enclosed in a
plastic rotor
Chemicals used in
rotor:
D-Mannitol
Polyethylene glycol,
8000
Dextran, 70 USP
Tris(hydroxymethyl)
amino
Polyethylene glycol,
3400
Polyethylene glycol,
2000
Sodium Chloride
POPSO, Disodium
salt
Sodium
Thiocyanate
L-Aspartic Acid
Lithium Hydroxide,
Tris, HCL
Please note – this includes any manual handling and VDU risk assessments required and
need to be included in this table.
Please note – this includes any room which had special requirements for example dark
rooms, CL3 laboratories.
3. CONTENT
2. Albumin
Bromocresol purple (BCP) when bound with albumin, changes colour from a
yellow to blue colour. The absorbance maximum changes with the colour shift.
4. Amylase (AMY)
The substrate, 2-chloro-p-nitrophenly-α-D-maltotrioside (CNPG3), reacts with α-
amylase in the patient sample, releasing 2-chloro-p-nitrophenol (CNP). The
release of CNP creates a change in colour.
6. Calcium (Ca)
Calcium in the patient sample binds with arsenaso 111 to form a calcium dye
complex
The endpoint reaction is monitored at 405nm, 467nm and 600nm. The amount of
calcium in the sample is proportional to the absorbence
7. Creatinine (CRE)
Creatinine amidohydrolase hydrolyses creatinine to creatine. A second enzyme,
creatine amidinohydrolsae, catalyses the formation of sarcosine from creatine.
Sarcosine oxidase causes the oxidation of sarcosine to glycine, formaldehyde
and hydrogen peroxide (H2O2). Peroxidase catalyses the reaction among
hydrogen peroxide, 2,4,6-tribromo-3-hydroxybenzoic acid (TBHBA) and 4-
aminoantipyrine (4-AAP) into a red quinoneimine dye. Sodium ferrocyanide and
ascorbate oxidase are added to the reaction mixture to minimise the potential
interference of bilirubin and ascorbic acid, respectively.
Two cuvettes are used to determine the concentration of creatinine in the sample.
Endogenous creatine is measured in the blank cuvette which is subtracted from
the combined endogenous creatine and the creatine formed from the enzyme
reactions in the test cuvette. Once the endogenous creatine is eliminated from
the calculations, the concentration of creatinine is proportional to the intensity of
the red colour produces. The endpoint reaction is measured as the difference in
absorbance between 550nm and 600 nm.
9. Glucose (GLU)
The reaction of glucose with adenosine triphosphate (ATP), catalysed br
hexokinase (HK) produces glucose-6-phosphate (G-6-P) and adenosine
diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PDH) catalyses
the reaction of 6-G-P into 6-phosphogluconate and the reduction of nicotinamide
adenine dinucleotide (NAD+) to NADH.
The rate of change of the absorbance difference between 340nm and 405nm is
caused by the conversion of NADH to NAD+ and is directly proportional to the
amount of urea present in the sample.
E) PATIENT PREPARATION
J) PROCEDURAL STEPS
Running a patient sample
Turn on the analyser by pressing the Power button on the front of the analyser
The analyser turns on then performs a self test
If the analyser needs time to warm the disc chamber to operating temperature the
display shows ‘warming’
The reagent discs can be used straight from the fridge. The discs can remain in the
sealed pouch at room temperature for a cummulative period of up to 48 hours
Check for tears and punctures in unopened foil pouch, use disk within 20 minutes of
opening
4. Slowly release the button to pick up the sample, pause then remove the
pipette from the sample tube.
5. Make sure there are no air bubbles or air gaps in the pipette tip
6. Place the pipette tip into the discs sample chamber at 45° so that the
entire sample flows into the sample chamber, the tip should touch the
sample chamber
7. Push the plunger down with a slow, continuous motion. Take care not to
overfill the sample chamber
8. Keep the the pipette plunger pressed down until the pipette tip is removed
from the sample chamber
9. Discard the pipette tip into a sharpsafe™
10. Carry the prepared disc to the analyser. Hold the disc by it’s edges and keep
level to avoid spills
11. Place the disc in the recessed area in the drawer
12. Select close. The analyser closes the drawer
13. Select the sample type from patient or control
14. Enter an ID number for the sample
15. Then select done
Authority For Issue: Jennifer Brown Page 11 of 16
Document printed from Q-pulse 00/00/0000 00:00:00 by
This is a controlled document: This copy is valid on day of print only, after which the user must ensure that this is the
correct version by comparing against the current document details in Q-Pulse
Document POCT-SOP7 Version No. 1
No. Issue Date 09/03/2018
16. The analyser checks the disc type and begins processing the sample
17. When the sample processing is complete the results will be printed
automatically
18. Select open to open the disc drawer
19. Remove the disc from the drawer and discard into a sharpsafe™
20. When finished select close to close the drawer and return the analyser to
standby mode
.
The control material used is Randox Abaxis chemistry controls level 1 and 2, which
are supplied lyophilised.
To reconstitute:
3. Allow to stand for 30 minutes before use. Ensure contents are completely
dissolved by swirling gently. Avoid formation of foam. Do not shake.
Quality controls (level 1 and 2) are run before the instrument is used at the donor
hospital
To run a QC
1. In the home screen, select ‘analyse’ to open the disc drawer
2. Select the control type to use by using the up and down arrow keys
Once complete check the printout to make sure that the results are within the target
range
Any results that are affected by >10% interference from haemolysis, lipaemia or
icterus are suppressed and HEM, LIP or ICT respectively are printed on the printout
in place of the result.
For a complete list of interfering substances, please refer to the General Chemistry
13 package insert POCT-EXT6
Not applicable
Albumin 35 – 55 g/L
Alkaline Phosphatase 42 – 141 U/L
Alanine Aminotransferase 10 – 47 U/L
Aspartate Aminotransferase 11- 38 U/L
Amylase 14 – 97 U/L
Urea 2.5 – 7.9 mmol/L
Calcium 2.00 – 2.58 mmol/L
Creatinine 53 – 106 µmol/L
Gamma Glutamyltransferase 5 – 65 U/L
Glucose 4.1 – 6.6
Bilirubin 3 – 27 µmol/L
Total Protein 64 – 81 g/L
Uric Acid 2.2 – 8.0 mg/DL
Albumin 10 – 65 g/L
Alkaline Phosphatase 5 – 2400 U/L
Alanine Aminotransferase 5 – 2000 U/L
Aspartate Aminotransferase 5 – 2000 U/L
Amylase 5 – 4000 U/L
Urea 0.7 – 64.3 mmol/L
Calcium 1 – 4 mmol/L
Creatinine 18 – 1768 µmol/L
Gamma Glutamyltransferase 5 – 3000 U/L
Glucose 0.56 – 38.9 mmol/L
Bilirubin 1.7 – 513 µmol/L
Total Protein 20 – 140 g/L
Uric Acid 1-15 mg/dL
When a result is outwith the measurement interval, a sample must be sent to the
laboratory for confirmation
Not applicable
Incorrect sample collection or collection into a tube with EDTA, fluoride oxalate or
citrate as an anticoagulant will interfere with test results
T) REFERENCES