GS01C25F05 01en
GS01C25F05 01en
GS01C25F05 01en
STANDARD SPECIFICATIONS
Refer to GS 01C25T02-01EN for Fieldbus
communication type and GS 01C25T04-01EN for □ PERFORMANCE SPECIFICATIONS
PROFIBUS PA communication type for the items
Zero-based calibrated span, linear output, wetted
marked with “◊.”
parts material code ‘S’ and silicone oil, unless
otherwise mentioned.
□ SPAN AND RANGE LIMITS For Fieldbus and PROFIBUS PA communication
(For EJX610A, values are in absolute pressure types, use calibrated range instead of span in the
and lower range limits are 0.) following specifications.
Specification Conformance
Measurement psi bar kgf/cm2 EJX series ensures specification conformance to at
MPa
Span/Range (/D1) (/D3) (/D4) least ±3σ.
2 to 200 Reference Accuracy of Calibrated Span
Span 0.3 to 29 0.02 to 2 0.02 to 2
kPa
A (includes the effects of terminal-based linearity,
−100 to −14.5 to hysteresis, and repeatability)
Range −1 to 2 −1 to 2
200 kPa 29
Span 0.01 to 2 1.5 to 290 0.1 to 20 0.1 to 20 Measurement span A
B −14.5 to Reference Span≥X ±0.04% of Span
Range −0.1 to 2 −1 to 20 −1 to 20
290 accuracy Span<X ±(0.02+0.007 URL/span)% of Span
7.3 to X 70 kPa (10.2 psi)
Span 0.05 to 10 0.5 to 100 0.5 to 100
1450
C URL
−14.5 to 200 kPa (29 psi)
Range −0.1 to 10 −1 to 100 −1 to 100 (upper range limit)
1450
50.8 to
Span 0.35 to 70 3.5 to 700 3.5 to 700 Measurement span B C D
10150
D
−14.5 to Reference Span≥X ±0.04% of Span
Range −0.1 to 70 −1 to 700 −1 to 700
10150 accuracy Span<X ±(0.005+0.0035 URL/span)% of Span
0.2 MPa 1 MPa 7 MPa
X
(29 psi) (145 psi) (1015 psi)
URL 2 MPa 10 MPa 70 MPa
(upper range limit) (290 psi) (1450 psi) (10150 psi)
Mode
Ambient Temperature Effects per 28°C (50°F) Burnout Fall back Off
Change 110%,
A capsule: ±(0.04% of Span + 0.075% of URL) Standard
21.6mA or more
B capsule: ±(0.04% of Span + 0.018% of URL) Holds to a
-2.5%, specified value
C and D capsule: ±(0.04% of Span + 0.009% of URL) /C1
3.6mA or less within the
Stability (All normal operating condition) Normal output
Option -1.25%, output range
EJX630A: ±0.1% of URL for 15 years /C2
Code 3.8mA or less from 3.6mA to
EJX610A: ±0.2% of URL for 15 years 21.6mA
103.1%,
Power Supply Effects /C3
20.5mA or more
±0.005 % per Volt (from 21.6 to 32 V DC, 350Ω)
Vibration Effects Damping Time Constant (1st order)
Amplifier housing code 1 and 3: Amplifier’s damping time constant is adjustable from
Less than 0.1% of URL when tested per the 0.00 to 100.00 s by software and added to response
requirements of IEC60770-1 field or pipeline time.
with high vibration level (10-60 Hz, 0.21 mm Note: For BRAIN protocol type, when the software
displacement/60-2000 Hz 3 g) damping is set to less than 0.5 s, communication
Amplifier housing code 2: may occasionally be unavailble during the
Less than ±0.1% of URL when tested per the operation, especially while output changes
requirements of IEC60770-1 field with general dynamically. The default setting of damping
ensures stable communication.
application or pipeline with low vibration level (10-60
Hz 0.15mm displacement /60-500 Hz 2g) Update Period “◊”
Pressure: 45 ms
Mounting Position Effects
Rotation in diaphragm plane has no effect. Tilting Zero Adjustment Limits
up to 90 degree will cause zero shift up to 0.21 kPa Zero can be fully elevated or suppressed, within the
(0.84 inH2O) which can be corrected by the zero lower and upper range limits of the capsule.
adjustment. External Zero Adjustment
Response Time (All capsules) “◊” External zero is continuously adjustable with
90 ms 0.01% incremental resolution of span. Re-range
When software damping is set to zero and including can be done locally using the digital indicator with
dead time of 45 ms (nominal) rangesetting switch.
Integral Indicator (LCD display, optional) “◊”
□ FUNCTIONAL SPECIFICATIONS 5-digit numerical display, 6-digit unit display and bar
graph.
Output The indicator is configurable to display one or up to
Two wire 4 to 20 mA DC output with digital three of the following variables periodically; pressure
communications, linear or square root programmable. in %, scaled pressure, measured pressure.
BRAIN or HART FSK protocol are superimposed on See also “Factory Settings.”
the 4 to 20 mA signal.
Local Parameter Setting (Output signal code D, E,
Output range: 3.6 mA to 21.6 mA
and J)
Output limits conforming to NAMUR NE43 can be
Parameter configuration by the external zero
pre-set by option code C2 or C3.
adjustment screw and push button (Integral indicator
code E) offers easy and quick setup for parameters
of Loop test, Tag number, Unit, LRV, URV, Damping,
Output mode (linear/square root), Display out 1, and
Re-range by applying actual pressure (LRV/URV)
and Device Information.
Burst Pressure
A, B and C capsule: 50 MPa
D capsule: 182 MPa
Self Diagnostics
CPU failure, hardware failure, configuration error,
process alarm for pressure or capsule temperature.
User-configurable process high/low alarm for
pressure is also available, and its status can be
output when optional status output is specified.
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN Nov. 20, 2019-00
<<Contents>> <<Index>> 3
Advanced Diagnostics (optional) “◊” Working Pressure Limits (Silicone oil)
Applicable for Output signal code E, J and F. Maximum Pressure Limits
• Impulse line blockage detection Values are in absolute pressure for EJX610A.
The impulse line condition can be calculated and
detected by extracting the fluctuation component Capsule Pressure
from the static pressure signal. A 200 kPa (29 psi)
• Heat trace monitoring B 2 MPa (290 psi)
The change of the process connection temperature C 10 MPa (1450 psi)
calculated by using the two temperature sensors D 70 MPa (10150 psi)
built in the EJX enables to detect the heat trace
breakage or the abnormal temperature due to the
Minimum Pressure Limit
failure.
See graph below
Signal Characterizer (Output signal code D, E
and J) 100(750)
User-configurable 10-segment signal characterizer
for 4 to 20 mA output.
Status Output (optional, output signal code D, E
and J)
One transistor contact output (sink type) to output 10(75)
the status of user configurable high/low alarm for
pressure.
Rating: 30 V DC, 120 mA DC max. 2.7(20)
Note: A check meter cannot be connected when status Working
output option (/AL) is specified. pressure Applicable range
Refer to ‘Wiring Example for Analog Output and kPa abs 1(7.5)
Status Output.’ (mmHg abs)
SIL Certification
EJX series transmitters except Fieldbus and 0.13(1)
PROFIBUS PA communication types are certified in 0.1(0.75)
compliance with the following standards;
IEC 61508: 2010;
Functional Safety of Electrical/electronic/
programmable electronic related systems; SIL 2
capability for single transmitter use, SIL 3 capability 0.013(0.1)
for dual transmitter use. 0.01(0.075) -40 0 40 80 120 (248)
Reliability Data different depending on hardware and (-40) (32) (104) (176) 85
software revision. (185)
For details, refer to Functional Safety Data Sheet. Process temperature °C (°F)
(Document number: TI 01C25A05-01EN or TI F01E.ai
01C25A05-21EN for option code SLT ) Figure 1-1. Working Pressure and Process
The document can be downloaded from the website Temperature [For EJX610A]
of Yokogawa.
(Website address: https://www.yokogawa.com/
solutions/products-platforms/field-instruments/)
100(14.5) Atmospheric
□ NORMAL OPERATING CONDITION pressure
Working
(Selected features may affect limits.) pressure
Ambient Temperature Limits kPa abs
−40 to 85°C (−40 to 185°F) (psia)
−30 to 80°C (−22 to 176°F) with LCD display Applicable range
Process Temperature Limits 10(1.4)
−40 to 120°C (−40 to 248°F)
Ambient Humidity Limits
0 to 100% RH
2.7(0.38)
Maximum Over Pressure
Values are in absolute pressure for EJX610A.
1(0.14)
Capsule Pressure -40 0 40 80 120
(-40) (32) (104) (176) (248)
A 4 MPa (580 psi)
B 16 MPa (2300 psi) Process temperature °C (°F)
C 25 MPa (3600 psi) F02E.ai
D 105 MPa (15200 psi)
Figure 1-2. Working Pressure and Process
Temperature [For EJX630A]
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN Nov. 20, 2019-00
<<Contents>> <<Index>> 4
Supply & Load Requirements With option code /PE3 (for D capsule)
(Output signal code D, E and J. Optional Category III, Module H, Type of Equipment: Pressure
features or approval codes may affect electrical Accessory-Vessel, Type of Fluid: Liquid and Gas,
requirements.) Group of Fluid: 1 and 2
With 24 V DC supply, up to a 550Ω load can be
used. See graph below. EU RoHS Directive
EN IEC 63000
Safety Requirement Standards
EN 61010-1, C22.2 No.61010-1
• Installation category: I
600
(Anticipated transient overvoltage 330 V)
E-10.5 • Pollution degree: 2
R=
External 0.0244 Digital • Indoor/Outdoor use
load Communication
resistance range □ PHYSICAL SPECIFICATIONS
BRAIN and HART
Wetted Parts Materials
250
Diaphragm, process connector
Refer to “MODEL AND SUFFIX CODES.”
R (Ω)
Non-wetted Parts Materials
Housing
10.5 16.6 25.2 42 • Low copper cast aluminum alloy
Power supply voltage E (V DC) • Low copper cast aluminum alloy with corrosion
F03E.ai resistance properties (copper content ≤ 0.03%,
iron content ≤ 0.15%) (optional)
Figure 2. Relationship Between Power Supply
• ASTM CF-8M Stainless steel (optional)
Voltage and External Load Resistance
Coating of housing
Supply Voltage “◊” [for aluminum housing]
10.5 to 42 V DC for general use and flameproof type. Polyester resin powder coating
10.5 to 32 V DC for lightning protector Mint-green paint (Munsell 5.6BG 3.3/2.9 or its
(option code A). equivalent)
10.5 to 30 V DC for intrinsically safe, type n, non- [for option code /P or /X2]
incendive or non-sparking type. Epoxy and polyurethane resin solvent coating
Minimum voltage limited at 16.6 V DC for digital Degrees of protection
communications, BRAIN and HART IP66/IP67, Type 4X
Load (Output signal code D, E and J) Pipe
0 to 1290Ω for operation Polypropylene
250 to 600Ω for digital communication
Cover O-rings
Communication Requirements “◊” Buna-N, fluoro-rubber (optional)
(Approval codes may affect electrical requirements.)
Name plate and tag
BRAIN 316 SST
Communication distance Fill fluid
Up to 2 km (1.25 miles) when using CEV Silicone, Fluorinated oil (optional)
polyethylene-insulated PVC-sheathed cables.
Weight
Communication distance varies depending on type of
Capsule A, B and C: 1.2 kg (2.6 lb)*
cable used.
Capsule D: 1.4 kg (3.1 lb)*
Load capacitance *: Without integral indicator and mounting bracket.
0.22 µF or less Add 1.5 kg (3.3 lb) for Amplifier housing code 2.
Load inductance Connections
3.3 mH or less Refer to “MODEL AND SUFFIX CODES.”
Input impedance of communicating device
10 kΩ or more at 2.4 kHz. < Related Instruments>
EMC Conformity Standards FieldMate Versatile Device Management Wizard:
EN 61326-1 Class A, Table2 Refer to GS 01R01A01-01E.
EN 61326-2-3 BRAIN TERMINAL: Refer to GS 01C00A11-00E
EN 61326-2-5 (for fieldbus) Power Distributor: Refer to GS 01B04T01-02E or
European Pressure Equipment Directive GS 01B04T02-02E
2014/68/EU
Sound Engineering Practice (for all capsules)
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN July 1, 2021-00
<<Contents>> <<Index>> 5
< Reference >
1. is a registered trademark of Yokogawa
Electric Corporation.
2. FieldMate; Trademark of Yokogawa Electric
Corporation.
3. Hastelloy; Trademark of Haynes International Inc.
4. HART®: Registered trademark of the FieldComm
Group.
5. FOUNDATION Fieldbus; Trademark of the
FieldComm Group.
6. PROFIBUS; Registered trademark of Profibus
Nutzerorganisation e.v., Karlsruhe, Germany.
Other company names and product names used in
this material are registered trademarks or trademarks
of their respective owners.
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN Mar. 25, 2019-00
<<Contents>> <<Index>> 6
MODEL AND SUFFIX CODES
For EJXC40A Remote Digital Sensor, please refer to GS 01C25W05-01EN.
The “►” marks indicates the most typical selection for each specification. Example: EJX630A-DAS4N-012NN/.
*1: Hastelloy C-276 or ASTM N10276.
*2: Users must consider the characteristics of selected wetted parts material and the influence of process fluids. The use of
inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage
to plant facilities. It is also possible that the diaphragm itself can be damaged and that material from the broken diaphragm
and the fill fluid can contaminate the user’s process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium
hypochlorite, and high-temperature steam (150°C [302°F] or above). Contact Yokogawa for detailed information of the
wetted parts material.
*3: Not applicable for combination of capsule code D and wetted parts material code H. Threads are based on the withdrawn
DIN 16 288.
*4: Not applicable for electrical connection code 0, 5, 7, 9 and A.
*5: Not applicable for electrical connection code 0, 5, 7 or 9.
*6: Material of a blind plug; aluminum alloy for code 5 and 9, and SUS304 for code 7.
*7: Not applicable for output signal code G.
*8: Not applicable for output signal code F.
*9: Intergranular corrosion test passed according to ASTM A262 Practice E.
The ‘#’marks indicate the construction materials conform to NACE material recommendations per MR0175/ISO 15156.
Please refer to the latest standards for details. Selected materials also conform to NACE MR0103.
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN July 1, 2021-00
<<Contents>> <<Index>> 7
OPTIONAL SPECIFICATIONS (For Explosion Protected type) “◊”
For other agency approvals and marine approvals, please refer to GS 01C25A20-01EN.
Please select appropriate equipment in accordance with the laws and regulations of the relevant country/region, when
it is used in a location where explosive atmospheres may be present.
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN July 1, 2021-00
<<Contents>> <<Index>> 8
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN Feb. 1, 2022-00
<<Contents>> <<Index>> 9
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN Feb. 1, 2022-00
<<Contents>> <<Index>> 10
OPTIONAL SPECIFICATIONS
Item Description Code
High Accuracy type *19 Reference accuracy: ±0.025% of Span HAC
Painting Color change Amplifier cover only*13 P
Amplifier cover and terminal cover, Munsell 7.5 R4/14 PR
Coating change Anti-corrosion coating*1 X2
316 SST exterior parts 316 SST zero-adjustment screw and setscrews*15 HC
Fluoro-rubber O-ring All O-rings of amplifier housing. Lower limit of ambient temperature: –15°C (5°F) HE
Lightning protector Transmitter power supply voltage: 10.5 to 32 V DC ( 10.5 to 30 V DC for intrinsically safe type,
9 to 32 V DC for Fieldbus communication type.)
A
Allowable current: Max. 6000 A ( 1×40 µs ), Repeating 1000 A ( 1×40 µs ) 100 times
Applicable Standards: IEC 61000-4-4, IEC 61000-4-5
Status output*2 Transistor output (sink type)
AL
Rating: 30 V DC, 120 mA DC (max) Low level: 0 to 2 V DC
Oil-prohibited use Degrease cleansing treatment K1
Degrease cleansing treatment with fluorinated oilfilled capsule.
K2
Operating temperature −20 to 80°C (−4 to 176°F)
Oil-prohibited use with Degrease cleansing and dehydrating treatment K5
dehydrating treatment Degrease cleansing and dehydrating treatment with fluorinated oilfilled capsule.
K6
Operating temperature −20 to 80°C ( −4 to 176°F)
Capsule fill fluid Flourinated oil filled in capsule
K3
Operating temperature −20 to 80°C (−4 to 176°F)
Calibration units*3 P calibration (psi unit) D1
bar calibration (bar unit) (See Table for Span and Range Limits.) D3
M calibration (kgf/cm2 unit) D4
Output limits and failure Failure alarm down-scale : Output status at CPU failure and hardware error is −5%, 3.2mA DC
C1
operation*4 or less.
Failure alarm down-scale: Output status at CPU
NAMUR NE43 Compliant C2
failure and hardware error is −5%, 3.2 mA DC or less.
Output signal limits:
3.8 mA to 20.5 mA Failure alarm up-scale: Output status at CPU
C3
failure and hardware error is 110%, 21.6 mA or more.
Gold-plated diaphragm *18 Surface of isolating diaphragms are gold plated, effective for hydrogen permeation. A1
Wired tag plate 316 SST tag plate wired onto transmitter N4
Data configuration at factory*5 Software damping, Descriptor,
Data configuration for HART communication type CA
Message
Data configuration for BRAIN communication type Software damping CB
Advanced diagnostics*16 Multi-sensing process monitoring
• Impulse line blockage detection *17 DG6
• Heat trace monitoring
European Pressure PED 2014/68/EU
Equipment Directive*14 Category: III, Module: H, Type of Equipment: Pressure Accessory-Vessel, PE3
Type of Fluid: Liquid and Gas, Group of Fluid: 1 and 2
Material certificate*6 Process Connector M15
Process connector, Diaphragm, Capsule body MA2
Pressure test/ Test Pressure: 200 kPa (29 psi) *7 T05
Leak test certificate*12 Test Pressure: 2 MPa (290 psi) *8 Nitrogen Gas or Water*11 T06
Test Pressure: 10 MPa (1450 psi) *9 Retention time: one minute T07
Test Pressure: 70 MPa (10150 psi) *10 T15
Parameter list*22 List of setting and adjustment parameters YP
Functional safety(SIL)*20*21 Low temperature expansion of functional safety Amb.Temp.: −55 to 85°C SLT
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN Mar. 30, 2020-00
<<Contents>> <<Index>> 11
*1: Not applicable with color change option. Not applicable for amplifier housing code 2.
*2: Check/External indicator terminals cannot be used when this option code is specified. Not applicable for output signal code F
and G.
*3: The unit of MWP (Max. working pressure) on the name plate of a housing is the same unit as specified by option codes D1,
D3, and D4.
*4: Applicable for output signal codes D, E and J. The hardware error indicates faulty amplifier or capsule.
*5: Also see ‘Ordering Information’.
*6: Material traceability certification, per EN 10204 3.1 B.
*7: Applicable for capsule code A.
*8: Applicable for capsule code B.
*9: Applicable for capsule code C.
*10: Applicable for capsule code D.
*11: Dry nitrogen gas or pure water is used for oil-prohibited use (option codes K1 and K2).
*12: The unit on the certificate is always kPa/MPa regardless of selection of option code D1, D3 and D4.
*13: Not applicable for amplifier housing code 2 and 3.
*14: Applicable for measurement span code D. If compliance with category III is needed, specify this option code.
*15: 316 or 316L SST. The specification is included in amplifier code 2.
*16: Applicable only for output signal code E and J.
*17: The change of pressure fluctuation is monitored and then detects the impulse line blockage. See TI 01C25A31-01E for
detailed technical information required for using this function.
*18: Applicable for wetted parts material code S.
*19: Refer to “PERFORMANCE SPECIFICATIONS”. Applicable only for EJX630A.
Not applicable for the combination of measurement span code A and wetted parts material code H. Not applicable for option
code K2, K3 or A1. When specified range values include negative value for A capsule, the accuracy shall be the standard
accuracy even if high accuracy option (HAC) is specified.
*20: Not applicable for EJX610A.
*21: Not applicable for output signal code F, G, P and S.
*22: Applicable only for output signal code D, E and J.
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN Mar. 30, 2020-00
<<Contents>> <<Index>> 12
DIMENSIONS
Unit: mm (approx.inch)
Model EJX610A and EJX630A
● With process connections code 7
95(3.74) 188(7.40)
60(2.36)*2 Integral indicator
(optional) Zero 110(4.33)
Extenal indicator adjustment 12 39
Conduit connection Conduit connection
(0.47) (1.54)
(optional)
ø78(3.07)
(2.76)
ø70
126(4.96)
54
110(4.33)
159(6.26)
(2.13)
(6.50)
6
165
(0.24)
Ground terminal
2-inch pipe
Mounting bracket 12*3 (O.D. 60.5 mm) 129(5.08)
(optional) (0.47)
(6.77)
172
ø6
5(0.20)
(0.24)
F04E.ai
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN Nov. 20, 2019-00
<<Contents>> <<Index>> 13
Connection Description
Analog output EJX electrical terminal
Distributor
+ + 24V DC
SUPPLY
– 250Ω
*1
CHECK A *1: Either A or + –
– – 250Ω
If shield cable is not
*1
used, communication ALARM A
is not possible.
*1: Either A or +
Magnetic
External power supply valve
30V DC, 120mA max
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN June 1, 2017-00
<<Contents>> <<Index>> 14
< Ordering Information > “◊” < Factory Setting > “◊”
Specify the following when ordering
Tag number As specified in order
For output signal code –J, refer to GS 01C25T01-
01EN. Software
‘2.00 s’ or as specified in order
damping *1
1. Model, suffix codes, and option codes
2. Calibration range and units Calibration range
As specified in order
lower range value
1) Calibration range can be specified with range
value specifications up to 5 digits (excluding any Calibration range
decimal point) for low or high range limits within upper range As specified in order
the range of -32000 to 32000. When reverse value
range is designated, specify Lower Range [EJX630A]
Value (LRV) as greater than Upper Range Value Selected from mmH2O, mmH2O(68°F),
(URV). When specified range value includes mmAq*2, mmWG*2, mmHg, Pa, hPa*2,
kPa, MPa, mbar, bar, gf/cm2, kgf/cm2,
minus value for A capsule, the accuracy shall
inH2O, inH2O(68°F), inHg, ftH2O,
be the standard accuracy even if high accuracy ftH2O(68°F) or psi. (Only one unit can be
option (/HAC) is specified. specified)
Calibration range
2) Specify only one unit from the table, ‘Factory
units [EJX610A]
Settings’ when shipped.’
3. Display scale and units (for transmitters equipped Torr, Pa abs, hPa abs*2, kPa abs,
MPa abs, mbar abs, bar abs, kgf/cm2 abs,
with integral indicator only)
mmH2O abs, mmH2O abs(68°F),
Specify either 0 to 100 % or engineering unit scale mmHg abs, inH2O abs, inH2O abs(68°F),
and ‘Range and Unit’ for engineering units scale: inHg abs, ftH2O abs, ftH2O abs(68°F),
Scale range can be specified with range limit psia, atm.
specifications up to 5 digits (excluding any decimal Designated value specified in order.
point) for low or high range limits within the range Display setting
(%, or user scaled value.)
of -32000 to 32000. The unit display consists of
6-digit, therefore, if the specified unit is longer than *1: To specify these items at factroy, /CA or /CB
7 characters excluding ‘/’, the first 6 characters will option is required.
be displayed on the unit display. *2: Not available for HART protocol type.
4. Tag Number (if required)
Specified characters (up to 16 characters for < Material Cross Reference >
BRAIN, 22 characters for HART, or 16 characters
for /N4 tag) are engraved on the stainless steel tag ASTM JIS
plate fixed on the housing. grade 316 SUS316
5. SOFTWARE TAG (for HART only. If required) grade 316L SUS316L
Specified characters (up to 32 characters) are set
grade 304 SUS304
as “Tag” (the first 8 characters) and “Long tag”*1
(32 characters) in the amplifier memory. Use
alphanumeric capital letters. <Information on EU WEEE Directive>
When the “SOFTWARE TAG” is not specified, EU WEEE (Waste Electrical and Electronic Equipment)
specified “TAG NO” is set as “Tag” (the first 8 Directive is only valid in the EU.
characters) and “Long tag”*1 (22 characters) in the
amplifier memory. This instrument is intended to be sold and used only
*1: applicable only when HART 7 is selected. as a part of equipment which is excluded from WEEE
6. Other factory configurations (if required) Directive, such as large-scale stationary industrial
Specifying option code CA or CB will allow further tools, a large-scale fixed installation and so on, and,
configuration at factory. Following are configurable therefore, subjected to the exclusion from the scope
items and setting range. of the WEEE Directive. The instrument should be
[/CA : For HART communication type] disposed of in accordance with local and national
1) Descriptor (up to 16 characters) legislation/regulations.
2) Message (up to 30 characters)
3) Software damping in second (0.00 to 100.00)
[/CB : For BRAIN communication type]
1) Software damping in second (0.00 to 100.00)
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 01C25F05-01EN Sep. 27, 2018-00
Subject to change without notice.