Ensure that the drip chamber is located between the infusion bottle/
bag and the pump, while the roller clamp is located between the
patient and the pump. The infusion tubing should be inserted in
accordance with the infusion direction indicator on the pump.
1. Hang the infusion bottle (bag) filled with medication on the IV
pole, close the flow regulator, and make sure the drip chamber is
about half filled.
2. Lift the pump lock handle to open the pump door
Lock handle
3. Position the infusion tubing in the power box slot, lift up the fluid
clamp, and press the infusion tubing firmly into the ultrasonic
sensor slot.
Power box
Fluid clamp
Flow regulator
4. Lift the lock handle to make sure the retractor clasps the door
pin, then press the lock handle down to close the pump door. At
this time, the surface of the lock handle should be flush with the
pump’s outer case.
Program a Continuous Infusion
the air has been expelled and fluid begins to drip from the end
of the tubing.
Note: Priming volume is not calculated in the cumulative volume
Calibration of brand and serial number of
infusion tubing
The infusion tubing brand should be updated and calibrated.
If you’re using other brands, calibration should be done after
selecting a serial number at the first use. The calibration methods
are as follows:
1. At the stop status, press the move cursor key
and select key SELECT at the same time to enter the rate
calibration screen
2. Load the infusion tubing
3. Set the serial number, rate and test volume (e.g. No. 02 Lifeline,
calibration rate 200 mL/h, calibration volume 20 mL)
4. Press the Start key to start the calibration. Once it reaches
to the test volume, press the Stop key immediately. The first
calibration is completed .
5. Repeat steps 3 and 4 to calibrate the tubing again
6. After finishing the second calibration, “OK” is shown on the
screen, if successful; move the cursor to “OK” by pressing the
Move cursor key .
7. If the same brand of infusion tube is used after calibration,
only the brand or serial number will have to be selected before
use. In consideration of the manufacturer’s accuracy factor for
the infusion set, it is advised to calibrate all brands of infusion
tubing before use. For more details on the calibration procedure,
please refer to the Instruction Manual.
1. Set up the infusion pump and make sure the infusion tubing has
been primed
2. Connect the infusion tubing to the patient and open the
flow regulator
3. Press the green Start key and the infusion will begin
4. The water droplet symbol flowing on the left of the LCD display
screen indicates the infusion pump is working. After 15 seconds
in the screensaver mode, the pump screen displays the infusion
rate values, the infusion rate unit, infusion time left and the
dynamic water droplet symbol, as shown in Figure 1.
Infusion
rate unit
Infusion Rate
ml/h 100 Infusion rate value
If the infusion pump alarms, listen for the beeps and check the
display screen for alarm messages or symbols.