Graseby™ 1200

Large Volume Infusion Pump


Quick Reference Guide

Read the entire Instruction Manual before operating the Graseby™


1200 Infusion Pump . Failure to properly follow warnings, cautions
and instructions could result in death or serious injury to the patient.
The Graseby™ 1200 is a volumetric infusion pump that provides
accurate and reliable infusions at a rate of 1 - 1200 mL/h

Start the pump

Before starting, the infusion pump must be placed horizontally for


use, and should be placed on a stable surface or attached to an IV
pole. Please keep a safe distance between the infusion pump and the
patient to help prevent the IV pole from falling over onto the patient.
1. Connect the AC power supply plug to the pump
2. Press the Power key for two seconds
3. The pump switches on and automatically performs a self-check
4. After the self-check, continue installing the infusion tubing if
there isn’t an alarm. Make sure the infusion tubing is in good
shape before starting the pump.
Install the infusion tubing

Ensure that the drip chamber is located between the infusion bottle/
bag and the pump, while the roller clamp is located between the
patient and the pump. The infusion tubing should be inserted in
accordance with the infusion direction indicator on the pump.
1. Hang the infusion bottle (bag) filled with medication on the IV
pole, close the flow regulator, and make sure the drip chamber is
about half filled.
2. Lift the pump lock handle to open the pump door

Lock handle

3. Position the infusion tubing in the power box slot, lift up the fluid
clamp, and press the infusion tubing firmly into the ultrasonic
sensor slot.

Power box

Fluid clamp

Flow regulator

4. Lift the lock handle to make sure the retractor clasps the door
pin, then press the lock handle down to close the pump door. At
this time, the surface of the lock handle should be flush with the
pump’s outer case.
Program a Continuous Infusion

The Graseby™ 1200 infusion pump can be programmed to infuse in


two ways:
1. Set the infusion volume and rate, and the infusion time will
automatically be calculated
2. Set the infusion volume and infusion time, and the infusion rate
will automatically be calculated
After installing the infusion tubing, set up the infusion pump
1. Use to input the infusion rate in mL/h. The rate unit is
only altered at a configuration interface
2. Move cursor below using the key
3. Use to input the infusion volume in mL. The pump will
automatically calculate the infusion time based on the infusion
rate and infusion volume
4. The brand of the infusion tubing can only be set in the
configuration screen. It is recommended to set the parameters
such as brand, occlusion grade, rate unit and KVO rate before
setting the rate, infusion, and other parameters.

Priming the Infusion Set

After the infusion tubing is installed and the pump is programmed,


prime the line to remove any remaining air in the tubing before
connecting to the patient.
While priming, ensure that the tubing is not connected to
the patient.
1. Open the flow regulator between the patient and the
infusion pump
2. Press the Priming/Bolus key twice and hold at the second
PRIMING
/TUBING

time to prime the infusion tube at the rate of 1000 mL/h


3. Release the Priming/Bolus key when you observe that
PRIMING
/TUBING

the air has been expelled and fluid begins to drip from the end
of the tubing.
Note: Priming volume is not calculated in the cumulative volume
Calibration of brand and serial number of
infusion tubing
The infusion tubing brand should be updated and calibrated.
If you’re using other brands, calibration should be done after
selecting a serial number at the first use. The calibration methods
are as follows:
1. At the stop status, press the move cursor key
and select key SELECT at the same time to enter the rate
calibration screen
2. Load the infusion tubing
3. Set the serial number, rate and test volume (e.g. No. 02 Lifeline,
calibration rate 200 mL/h, calibration volume 20 mL)
4. Press the Start key to start the calibration. Once it reaches
to the test volume, press the Stop key immediately. The first
calibration is completed .
5. Repeat steps 3 and 4 to calibrate the tubing again
6. After finishing the second calibration, “OK” is shown on the
screen, if successful; move the cursor to “OK” by pressing the
Move cursor key .
7. If the same brand of infusion tube is used after calibration,
only the brand or serial number will have to be selected before
use. In consideration of the manufacturer’s accuracy factor for
the infusion set, it is advised to calibrate all brands of infusion
tubing before use. For more details on the calibration procedure,
please refer to the Instruction Manual.

Start the infusion

1. Set up the infusion pump and make sure the infusion tubing has
been primed
2. Connect the infusion tubing to the patient and open the
flow regulator
3. Press the green Start key and the infusion will begin
4. The water droplet symbol flowing on the left of the LCD display
screen indicates the infusion pump is working. After 15 seconds
in the screensaver mode, the pump screen displays the infusion
rate values, the infusion rate unit, infusion time left and the
dynamic water droplet symbol, as shown in Figure 1.

High: 120 kPa; Middle: 66.7 kPa; Low:10 kPa

Infusion
rate unit
Infusion Rate
ml/h 100 Infusion rate value

Dynamic drop Infusion


symbol Time Left 09h59 Infusion time left

Figure 1: Pump screensaver


Infusion program setting

The Graseby™ 1200 infusion pump can be used for consecutive


infusions through two different infusion programs. Before setting up
the infusion pump,first install the infusion tubing.
1. In the programming mode, press the Select key until the screen
is changed to the infusion program setting screen
2. To set Infusion Program 1: Use to input the infusion
rate and press to move the cursor to the infusion volume.
Use to input the infusion volume and press to move
the cursor to the infusion rate for Infusion Program 2.
3. To set Infusion Program 2: Use the same method from Infusion
Program 1 to set the infusion rate and infusion volume for
Infusion Program 2.
4. After the correct infusion program is displayed, press to
move the cursor to “OK”, then press the Select key to confirm
and save the infusion program setting and return to the main
screen. On the Infusion Program Setting screen, press the
Stop key and then the Select key concurrently to clear the
infusion program.
Note: When infusion programs are applied, the infusion pump
will perform Infusion Program 2 automatically after the end of
Infusion Program 1.
Alarms

If the infusion pump alarms, listen for the beeps and check the
display screen for alarm messages or symbols.

Alarms Tips/reason Operation


The battery capacity Press the Mute key ,
Battery low is low/infusion pump connect the pump to an
Beep-beep-beep could work about AC power supply as soon
30 minutes as possible
Battery depleted The battery capacity Connect the infusion pump
is empty, the infusion to an AC power supply
Beep pump stops immediately
Nearly completed The infusion is Press the Mute key ,
infusion about to complete prepare to stop
Beep-beep-beep in 3 minutes the infusion
Completed The infusion is
infusion completed, the Press the Stop key to
pump will operate stop the infusion
Beep-beep-beep in the KVO rate
Turn off the flow regulator
and press the Stop key
Tubing route . After the occlusion is
occlusion The infusion tubing or
eliminated, open the flow
cannula is blocked
Beep-beep-beep regulator and press the
Start key to continue
the infusion or start over.
Press the Mute key to
Forget to operate The pump has been
clear the alarm and start
left idle for 2 mins
Beep-beep-beep the infusion or close the
without any operation
infusion pump

The door isn’t closed, Press the Stop key


The door to stop infusion or
isn’t closed the infusion stops
or the infusion close the door and
Beep-beep-beep cannot begin repress. Start key
to continue infusion.

PRODUCT(S) DESCRIBED MAY NOT BE LICENSED OR


AVAILABLE FOR SALE IN CANADA AND OTHER COUNTRIES

Smith Medical EC Authorized MZHCCA-0009


International Representative
1500 Eureka Park, Smiths Medical
Lower Pemberton International Ltd.
Ashford Kent, TN25 4BF TN25 4BF, UK
Tel: +44 (0)845 850 0445 Tel: +44 (0) 1233 722100
www.smiths-medical.com

Find your local contact information at:


www.smiths-medical.com/customer-support
Smiths Medical is part of the global technology business Smiths Group plc. Please see
the Instructions for Use/Operator’s Manual for a complete listing of the indications,
contraindications, warnings and precautions. Graseby and the Smiths Medical design mark are
trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U.S.
Patent and Trademark Office and certain other countries. All other names and marks mentioned
are the trademarks or service marks of their respective owners. ©2017 Smiths Medical.
All rights reserved. IN193909GB-052017

You might also like