Removing the Flat Panel 5 Repair and Disassembly

Removing the Flat Panel


1 Separate the front and back half of the monitor as described above.
2 Remove the touch controller board.

Touch Controller Board

Touch Sensor Cable

Reassembly Note: Plug the touch sensor cable into the touch controller board first, then re-insert the
touch board

121
5 Repair and Disassembly Removing the Flat Panel

3 Remove the two screws securing the power switch and the LED board and pull out the board. The
board and the screws are part of the flex panel adapter cable.
Reassembly Note: The screws securing this board are self-cutting - turn in opposite direction first
when re-inserting them
.

Flex Panel Adapter Cable

4 Unplug the backlight tube cables and release them from the strain relief cable holders.

5 Unscrew the screw holding the metal sheet under touch controller board and remove the metal
sheet.

122
Removing the Flat Panel 5 Repair and Disassembly

6 Slip a screwdriver under the end of the flex panel adapter cable and lift upwards to remove the
connector to the display.

7 Remove the eight screws and carefully lift off the display assembly by releasing the two snaps in the
monitor bezel.

Snaps

N O TE This procedure must be performed in a dust free environment. The display assembly may be hard to
remove from the touch panel due to the gasket (adhesive strips) holding the two together.

123
5 Repair and Disassembly Removing the Backlight Inverter Board

8 Turn the display unit around and remove the four remaining screws to separate the display from
the metal chassis.

Removing the Backlight Inverter Board


1 Separate the front and back half of the monitor as described above.
2 Unplug the backlight tube cables (1) and the connector to the flex panel cable (2).
3 Remove the two screws (3) securing the Backlight Inverter Board and pull off the board.

(1)

(2)

(3)
(1) (3)

N O TE The backlight tube cable must not be unplugged for display exchange

124
Removing the Silicon Mat for the Power Switch and the LEDs 5 Repair and Disassembly

Removing the Silicon Mat for the Power Switch and


the LEDs
1 Separate the front and back half of the monitor as described above.
2 Remove the two screws securing the power switch and the LED board and pull out the board. The
board and the screws are part of the flex panel adapter cable.
Reassembly Note: The screws securing this board are self-cutting - turn in opposite direction first
when re-inserting them
.

Flex Panel Adapter Cable

3 Pull out the silicon mat (for power switch and LEDs).

125
5 Repair and Disassembly Removing the Navigation Point Assembly

Removing the Navigation Point Assembly


1 Separate the front and back half of the monitor as described above.
2 Remove the two screws at the back of the navigation point assembly and remove the connector
from the main board by pulling it towards front. Then pull off the navigation point assembly.

N O TE To remove the flat ribbon cables with red connectors (e.g. Navigation Point cable) simply pull directly
on the cable.

126
Removing the Navigation Point Assembly 5 Repair and Disassembly

Reassembly Note: When reinserting the connector insert the protruding nose at the top

Reassembly Note: When re-inserting the input device, make sure the spring is inserted into the hole in
the plastic housing.

N O TE The screws securing the Navigation Point Assembly are self-cutting - turn in opposite direction first
when re-inserting them.

127
5 Repair and Disassembly Removing the ECG Out/Alarm LED Board

Removing the ECG Out/Alarm LED Board


1 Separate the back and front half of the monitor as described above and lay the back half of the
monitor on its back.
2 Remove the flat ribbon connector of the ECG Out/Alarm LED board from the main board and
remove the 3 screws that hold the board and lightpipes.

N O TE To remove the flat ribbon cables with red connectors (e.g. Navigation Point cable) simply pull directly
on the cable.

3 Lift the board out to the right so the ECG out connector is pulled out of the monitor housing.

N O TE The connector must be unplugged from the main board.

128
Removing the Handle 5 Repair and Disassembly

4 Pull off the lightpipe holder if necessary. It is snapped in on both sides of the board.

Feet

Snaps

Reassembly Note: Make sure the snaps and feet are properly hooked into the ECG Out board when
reattaching the lightpipe holder.

Removing the Handle


1 Separate the front and back half of the monitor as described above.
2 Remove the ECG Out board as described above.

129
5 Repair and Disassembly Removing the Handle

3 Remove the two screws securing the handle and remove the handle.