Vitalograph 6300 Micro Service Manual
Vitalograph 6300 Micro Service Manual
Vitalograph 6300 Micro Service Manual
VITALOGRAPH micro
MODEL 6300
SERVICE MANUAL
Vitalograph (Ireland) Ltd., Gort Road Business Park, Ennis, Co. Clare, Ireland
Phone: +353 (0)65 6864100 Fax: +353 (0)65 6829289
e-mail: sales@vitalograph.ie
www.vitalograph.com
CHAPTER 1: INTRODUCTION........................................................................................................................................ 4
Vitalograph micro Display............................................................................................................................................... 5
Vitalograph micro Label.................................................................................................................................................. 5
CHAPTER 2: PRINCIPLE OF OPERATION..................................................................................................................... 6
CUSTOMER SERVICE.................................................................................................................................................... 19
The Vitalograph micro is a portable and easy to use precision spirometer. It offers total confidence in rapid and
accurate spirometry measurements as a part of your routine physical examination.
The Vitalograph micro Service Manual should be used in conjunction with the Vitalograph micro User Manual (07056).
Flowhead
Down Arrow
Button
Flowhead
Release
Button
Up Arrow
Button
LCD Display
(see below)
On/Off
Button
Low Battery
Blow Now Symbol, indicates the
Symbol (see
unit is ready to do a test
Thumbs Up/Down Symbol, indicates table below)
successful/unsuccessful accuracy
check/calibration update
Weight
Gender
Age
Height
The Vitalograph micro measures a subject’s lung ventilation by using a flowhead containing a small Fleisch-type
Pneumotachograph capable of giving linear signals throughout the entire physiological range.
During testing, the airflow through the Pneumotach produces a very low-pressure differential. Internally this pressure is
applied to a pressure transducer, which produces an electric signal. This signal is converted into digital form so that the
unit can perform calculations and the results be displayed.
The Vitalograph micro retains last best-test even when switched off (not overwritten until a blow occurs after switch-on)
• Flow-Volume graph
• Volume-Time graph
Note: If an inspiratory section of the flow-volume graph is printed, this is only an indication that an Inspiratory test
has occurred and does not accurately represent the volume inspired.
The Vitalograph micro is not supplied with its power source (9V PP3 type alkaline battery - recommended capacity of at
least 550mAh).
The battery should be fitted by attaching it to the connector in the battery compartment and locating the battery in the
slot provided. It is necessary to feed the excess connector cable back into the hole from which it exits to ensure that
the battery fits correctly. When the battery is fitted, the unit will power up immediately.
Note that it is recommended to remove the battery if you do not intend using the Vitalograph micro for several weeks.
Setting Up
Follow these instructions in sequence to set up the Vitalograph micro before you use it for the first time, or to change settings
later.
1. Press .
Note that the Vitalograph micro turns itself off if not used within two minutes.
2. Press to continue
1. To select the correct hour press or . Note that pressing down continuously on the button changes the number
quickly.
2. Press
4. Press
2. Press
4. Press
Copyright Vitalograph 1999, 2003, 2005 Issue 3 7
5. To select the correct day press or .
After completion of the above, the Vitalograph micro is ready for testing. The time and the ‘blow now’ icon appears on
the screen, accompanied by a double beep.
The display shows the last test, but the best test is held in memory. Blows may be repeated when the ‘blow icon’ is
showing.
On completion of testing the user can attach a printer and print the current session. The best test in the session will be
printed.
Immediately after a test session is completed, plug the jack plug into the Vitalograph micro jack socket, which is located
on the side of the device.
The Vitalograph micro will take about two minutes to download all the test data during which time it should not be
interfered with. After two minutes transmission, the device will ‘beep’ and briefly display ‘PRN’ again.
At this time the creation of the spirometry report begins in the PC Print Utility.
1
Depending on your settings, a PDF will be created and/or a print command will be sent to your designated printer .
Note: The best test in the session is printed, not necessarily the last test which is still showing on the LCD
screen.
To cancel at any time press and hold the switch for 15 seconds approx.
1
If you installed the Print Utility with default settings, no print command will be sent but a PDF file will be created in the
following location: C:\Program Files\Vitalograph\VitalPrint\Data
Copyright Vitalograph 1999, 2003, 2005 Issue 3 8
Note: When the Vitalograph micro switches itself off, the best test from the last session is retained until the next blow.
This stored test may be printed at any time as long as no new blow is detected by the device. Switch ON, press Enter
repeatedly until the Blow Icon and time are displayed, then press and together to print.
Accuracy Check
Your Vitalograph micro is extremely stable and needs no calibration adjustment between the annual service intervals.
However, international guidelines such as ATS/ERS recommend that the accuracy of spirometers be checked ’at least
daily’ using a 3-L Precision Syringe.
The Vitalograph micro can be checked for accuracy easily using the following procedure, using a 3-L Precision
Syringe. Ensure that the syringe has been in the same temperature environment as the unit for at least an hour.
1. After switching the unit on, press the upper recess / Accuracy Check button with a suitably pointed object. The
display will look like …..
2. Press Up and Down to change the syringe stroke volume if you are not using a 3-L syringe. The ‘03’
represents the stroke volume of your precision syringe, not the total volume that will be delivered. It is very
important to set this up correctly, otherwise, the accuracy check will not function properly.
3. Press Enter. The display will look like …….
4. Ensure the syringe piston is fully withdrawn. At this point, connect the Vitalograph micro to the syringe using the
adaptor supplied. Leave the Vitalograph micro in a stable position - do not move the unit during this procedure.
5. With an even stroke, pump the air from the syringe through the device. The stroke must be at least one second in duration.
6. Wait for a double beep, withdraw the syringe piston fully and repeat step 5 several times until ‘five beeps’ occur in rapid
succession. It is very important to wait for the double beep before withdrawing the piston each time.
7. If there is a Thumbs Up on the LCD, then the unit is within the +/- 3% accuracy tolerance …
Note: with reference to step 6 above, the user will be prompted by means of a double beep for another stroke of the syringe when the
volume difference of the present stroke is not within 3% of the previous stroke. If the maximum of five strokes is executed, then the
five beeps will occur, signifying the end of the accuracy check.
1. End Of Procedure
If there is a Thumbs Down on the LCD, then the unit is outside the +/- 3% accuracy requirements. Check for
faulty assembly or procedure, then recheck accuracy. If still wrong contact Technical Support.
IMPORTANT: If your Vitalograph micro has a current Certificate of Calibration Traceability, entering the calibration
adjustment procedure will invalidate this certificate. Any applied calibration traceability stickers should also be removed.
The calibration of the Vitalograph micro can be adjusted easily using the following procedure. A 3-L precision syringe
is recommended.
1. After switching the unit on, press the Calibration Adjust / lower recess button with a suitably pointed object.
The display will look like…
2. Press Up and Down only if your Precision Syringe is NOT a 3-L stroke volume. It is very important to set this to
the stroke volume of the syringe otherwise the calibration adjustment procedure will not function properly.
3. Press Enter. The display will look like…
4. Ensure the syringe piston is fully withdrawn. At this point, connect the unit to the syringe and leave it in a stable
position - do not move the unit during this procedure.
5. With an even stroke, pump the air from the syringe into the unit. The stroke must be at least one second in duration.
6. Wait for a double beep, withdraw the syringe piston fully and repeat step 5 until five beeps occurs. It is very important to wait for
the double beep before withdrawing the piston each time.
7. If there is a Thumbs Up on the LCD, then the unit has adjusted its calibration successfully…..
8. If there is a Thumbs Down on the LCD, then a problem was encountered while trying to adjust calibration.……
SERIAL # ……………
YOUR UNIT’S ACCURACY CHECK HISTORY LOG
1. Turn on the unit; scroll to a height of 0 and press ‘Enter’. This will allow you to get to the ‘ready’ screen without
having to enter all the demographic data.
2. The time and the “Blow” symbol should appear after a few seconds.
3. Use a syringe to blow air through the flowhead. Check that a volume registers. (It may not be the same as the
displaced volume because BTPS correction will be applied in test mode.
4. If this seems to function properly, perform an accuracy check as detailed in Section E – Accuracy Check &
Calibration.
Fault condition: If the instrument is faulty, contact your Vitalograph service centre for advice,
service or repair.
General Recommendation
®
To reduce the risk of cross infection the use of SafeTway mouthpieces is recommended.
A Risk Assessment should be carried out to assess the risks presented to both operator and subject, and an action
plan devised by the facility to minimise the chance of cross infection occurring, particularly where known-infectious or
immuno-deficient subjects are being tested.
An assessment should be made of methods of decontamination available to the operator, and their effectiveness
against potential risks - a table of materials used in the Vitalograph micro is provided below to assist in this.
It is recommended that in cases of high risk with no effective disinfection methods available, that the contaminated
parts are disposed of. For this device, the Flowhead Assembly: FLOWHEAD COMPLETE.
Routine Practice
A new mouthpiece (SafeTway mouthpiece) should be used for each subject. A period of at least 5 minutes should be
allowed between subjects to allow settling of aerosolised particles.
It is recommended that the Flowhead Assembly be regularly cleaned according to the guidelines of the user’s facility.
In the event of visible contamination to the FLOWHEAD - it should be cleaned and disinfected or replaced. It should
also be replaced in the event of damage.
It is recommended to replace the complete: FLOWHEAD as a part of the routine annual service.
All parts of the Vitalograph micro require cleaning, i.e. the removal of visible particulate contamination. The parts of
the Vitalograph micro that make up the flowhead also require disinfecting.
HYGIENE
It is vital for the user to set guidelines for protective hygiene measures whilst performing spirometry testing. There are
three main potential sources of cross contamination, skin contact; aerosolised particles and saliva/body fluids. By far
the most important is the last item – a minimum requirement is to use a new disposable mouthpiece for each subject
tested. We recommend using the SafeTway mouthpiece. We also recommend that a period of at least 5 minutes be
allowed between subjects. This allows aerosolised organisms to be removed by gravitational sedimentation between
tests. (Am J Respir Crit Care Med Vol 159. pp 610-612, 1999)
Cleaning Instructions
All parts of the Vitalograph micro require regular cleaning, i.e. the removal of visible particulate contamination. The
parts of the Vitalograph micro that make up the flowhead also require disinfecting. Vitalograph recommends that
cleaning and disinfecting should normally be carried out on a weekly basis. In addition to a weekly cleaning and
disinfection cycle, Vitalograph recommends cleaning and disinfection of equipment to be carried out after use on
infected subjects or prior to use on immuno-compromised subjects. The user must determine what level of disinfection
is appropriate and acceptable in any particular circumstance.
Flowhead
Hydrophilic Filters
Flowhead release
button
Vitalograph Trained Service Technician: I hereby Certify this instrumentation to be safe, clinically
clean and in proper working order.
Name:
Calibration of Spirometer & Syringe Certified as above.
Signature: Date:
Flow
Sensor
Random
PRESSURE Access Memory
EEPROM Transduc
er
REAL
TIME
Micro Processor Clock
with 64K Read Only
9V PP3 Memory.
BATTERY Keypad
CUSTOMISED LIQUID
SERIAL Crystal Display
PORT BUZZER
For the names and addresses of approved Vitalograph Service Agents, please contact any of the following offices or
your local Vitalograph distributor.
Vitalograph (Irl.) Ltd, Gort Road Business Park, Ennis, Co. Clare, Ireland
Phone: (065) 6829611 Fax: (065) 6829289
e-mail: sales@vitalograph.ie
Internet: www.vitalograph.com