In the seca 769 electronic personal scale, In addition to the conventional determina-
you have purchased a high-precision yet tion of weight, the seca 769 also has a
sturdy piece of equipment. function for determining body mass index.
seca has been putting its experience at For this purpose, height is entered on the
the service of health for over 150 years keypad and the body mass index associ-
now, and as market leader in many coun- ated with the weight value is output.
tries of the world, is always setting new The scale is of a very sturdy construction
standards with its innovative develop- and will give you long and loyal service. It
ments for weighing and measuring. is easy to operate and the large display is
In accordance with national regulations, the easy to read.
seca 769 electronic personal scale is used The seca 769 can be moved on castors,
primarily in hospitals, doctors’ surgeries and its low power consumption means
and in-patient nursing facilities. that it can be used on a mobile basis with
Weight is determined within a few sec- batteries for long periods.
onds.
2. Safety
Before using the new scale, please take a little time to read the following safety instruc-
tions.
• Follow the instructions in the instruc- • Use only the type of battery stated (see
tions for use. “Changing the batteries” on page 19).
• Keep the operating instructions and the • Have scale serviced and re-calibrated
declaration of conformity in a safe on a regular basis (see chapter entitled
place. “Maintenance” on page 19).
• Do not drop the scale or subject it to • Have repairs carried out only by author-
violent shocks. ised persons.
12
Power supply
The scale is generally powered by batteries.
The LC display has a low power consumption – 10,000 weighing operations can be per-
formed with one set of batteries.
– Carefully turn over the scale so that the
base of the scale is accessible.
– Push the battery compartment closure
towards “Battery” and open the cover.
– Connect the cable to the holder and
batteries.
– Place the holder in the battery compart-
ment.
– Close the cover and then turn the scale
back up the right way.
Note: the scale is prepared for use with a seca mains unit which is available as
an accessory.
Model 769 GB 13
4. Operation
Controls and displays
The controls and the display are on the display head on the column.
14
Correct weighing
– Press the Start key with no load on the
scale.
se(A, 8.8.8.8.8. and 0.0 appear conse-
cutively in the display. The scale is then
automatically set to zero and ready to
use.
– If necessary, switch the weight display
(see page 15).
– Stand on the scale and keep still.
– Read the weighing result off the digital
display.
– If a load greater than 200 kg or 450 lbs
is placed on the scale, St0P appears in
the display.
– To switch off the scale, press the Start
key again.
Model 769 GB 15
Taring weight display (TARE)
The tare function is very useful, for example for determining the weight of a child, animal
or other object placed on the scale in addition to yourself.
Proceed as follows:
– Press the Start key with no load on the
scale.
– First stand on the scales yourself with-
out the additional weight and activate
the Tare function by depressing and
holding the HOLD/TARE key.
The display is reset to zero. 0.0 appears
in the display. The note “NET” appears.
– Now take the additional load onto the
scale. You can leave the scale to do so,
in which case ---- appears.
The scale determines the weight of the
additional load.
You can now perform as many weig-
hing operations as you like.
– Hold the key down again to deactivate
the Tare function; the scale is then back
in normal weighing mode and the tare
value has been erased.
– The tare value is likewise erased by the
scale being switched off.
16
Body Mass Index (BMI)
Body Mass Index puts height and weight in relationship to one another, providing more
accurate information than ideal Broca weight, for example. A tolerance range is given
which is considered ideal for health.
– Press the Start key with no load on the
scale.
se(A, 8.8.8.8.8. and 0.0 appear con-
secutively in the display. The scale is
then automatically set to zero and ready for operation.
– Press the FUNC key.
On the display, you will see the last
height entered.
– You can change the value in 1 cm incre-
ments using the arrow keys.
Model 769 GB 17
5. Cleaning
Clean the rubber surface and the housing as required using a domestic cleaning agent or
commercially-available disinfectant. Follow the manufacturer’s instructions.
Under no circumstances use abrasive or acid cleaners, white spirit, benzene or the like
for cleaning. Such substances can damage the high-quality surfaces.
6. What do I do if…
… no weight display comes on?
– Is the scale switched on?
– Check the batteries.
– Is the plug contact on the scale properly connected??
… 0.0 does not appear before the weighing operation?
– Press the Start key again – there must not be any load on the scale
… ---- appears instead of 0.0 before weighing?
– Press the Start key again – there must not be any load on the scale.
… one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.
… the display shows ?
– Battery voltage is running low. Change the batteries in the next few days.
… bAtt appears in the display?
– Batteries are empty. Put in new batteries.
… St0p appears in the display?
– Maximum load has been exceeded.
… the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-
tected a steady weight value. The display will then stop flashing.
– If there has been no activation of a function beforehand, remove the load from the scale
and wait until 0.0 is displayed, then weigh again.
… the display E and XX.XX appear?
– Switch off the scale with the Start key and start the scale again. The scale will then
work normally again.
If this is not the case, disconnect the power supply by briefly removing the batteries. If
this measure is equally unsuccessful, inform the Service department.
18
7. Maintenance
Your seca scale leaves the factory with an accuracy greater than ± 0.15 %. To ensure that
this accuracy is maintained, the product must be carefully set up and regularly maintained.
Depending on the frequency of use, we recommend servicing every 3 to 5 years. Please
ensure that a qualified service agent is used. If in doubt, contact seca Customer Service for
further information.
Model 769 GB 19
8. Technical data
Dimensions
width: 292 mm
height: 830 mm
depth: 390 mm
Height of figures 20 mm
Weight approx. 6.8 kg
Temperature range +10 °C to +40 °C
Power supply Battery/mains unit (option)
Batteries 6 type AA 1.5 V batteries
EN 60 601-1:
Insulated equipment,
Protection Class II:
medical electrical equipment type B:
Maximum load 200 kg/450 lbs/32 sts
Graduations 100 g/0.2 lbs
Accuracy greater than ± 0.15 %
Functions Switch between units, Tare, Hold
Body Mass Index
20
10. Disposal
Disposing of the device Batteries
Do not dispose of the device in Do not throw used batteries away in do-
domestic waste. The device must mestic waste. Dispose of batteries at col-
be disposed of properly as elec- lection points in the vicinity. When buying
tronic waste. Follow the national regula- new batteries, select those low in harmful
tions which apply in your case. For further substances and containing no mercury
information, contact our service depart- (Hg), cadmium (Cd) or lead (Pb).
ment at:
11. Warranty
A two-year warranty from date of delivery honoured only if the complete original pa-
applies to defects attributable to poor ma- kkaging was used for any transport and
terials or workmanship. All moveable parts the scale secured and attached in that pa-
- batteries, cables, mains units, recharge- kkaging just as it was when originally pa-
able batteries etc. - are excluded. Defects kked. All the packaging should therefore
which come under warranty will be made be retained.
good for the customer at no charge on A claim under warranty will not be hon-
production of the receipt. No further oured if the equipment is opened by per-
claims can be entertained. The costs of sons not expressly authorised by seca to
transport in both directions will be borne do so.
by the customer should the equipment be We would ask our customers abroad to
located anywhere other than the custom- contact their local sales agent in the event
er’s premises. In the event of transport of a warranty matter.
damage, claims under warranty can be
Model 769 GB 21