AAMI - Emergency Use Ventilator Design Guidance
AAMI - Emergency Use Ventilator Design Guidance
AAMI - Emergency Use Ventilator Design Guidance
Revisions are expected to be made to this document as the COVID-19 situation evolves. Please go to
https://www.aami.org/covid_cr to find the most current version as well as past versions. This document is
freely available and may be shared with all interested stakeholders. Contact celliott@aami.org with any
comments or questions.
Abstract: Provides targeted design constraints to enable rapid development of emergency use ventilators
(EUV) to treat patients with COVID-19 respiratory failure. Also intended to guide the review of an
EUV by an authority having jurisdiction.
Keywords: COVID-19
AAMI Consensus Report
A Consensus Report (CR) is a publication of the Association for the Advancement of Medical Instrumentation (AAMI)
developed to provide concise, prompt and practical guidance on narrowly focused topics of high importance to the
health technology community. A Consensus Report is intended provide initial consensus guidance in response to an
urgent/immediate need for guidance in the following instances:
A Consensus Report is not subject to the same formal process as a standard and while similar in nature to a technical
information report (TIR), a CR is based on the collective knowledge and experience of a selected group of stakeholders
and has not undergone the wider reviews of a TIR or standard and offers an even greater response time.
CAUTION NOTICE: This AAMI CR may be revised or withdrawn at any time. Because it addresses a rapidly evolving
field or technology, readers are cautioned to ensure that they have also considered information that may be more recent
than this document.
All standards, technical information reports, consensus reports and other types of technical documents developed by
AAMI are voluntary, and their application is solely within the discretion and professional judgment of the user of the
document. Occasionally, voluntary technical documents are adopted by government regulatory agencies or
procurement authorities, in which case the adopting agency is responsible for enforcement of its rules and regulations.
Comments on this technical information report are invited and should be sent to AAMI, Attn: Standards Department,
901 N. Glebe Rd, Suite 300, Arlington, VA 22203-1633.
Published by
Publication, reproduction, photocopying, storage, or transmission, electronically or otherwise, of all or any part of this
document without the prior written permission of the Association for the Advancement of Medical Instrumentation is
strictly prohibited by law. It is illegal under federal law (17 U.S.C. § 101, et seq.) to make copies of all or any part of this
document (whether internally or externally) without the prior written permission of the Association for the Advancement
of Medical Instrumentation. Violators risk legal action, including civil and criminal penalties, and damages of $100,000
per offense. For permission regarding the use of all or any part of this document, visit the Copyright Clearing Center
ISBN 978-1-57020-748-8
Contents Page
This AAMI Consensus Report (CR) was developed by a task group under the auspices of the AAMI COVID-19
Response Team.
The AAMI COVID-19 Response Team had the following members:
NOTE—Participation by federal agency representatives in the development of this document does not constitute
endorsement by the federal government or any of its agencies.
AAMI gratefully acknowledges the writing team members, Julian Goldman, Dave Osborn, Anthony Ciccarello and
Sandy Weininger for their outstanding and expeditious work in preparing these drafts for committee review and
approval.
4 The goals of this document are to provide targeted design constraints to enable rapid development of
5 emergency use ventilators (EUV) to treat patients with COVID-19 respiratory failure. This document is also
6 intended to guide the review of an EUV by an authority having jurisdiction.
7 It is recognized that the surge in COVID-19 is requiring extraordinary measures to provide mechanical
8 ventilatory support to keep pace with clinical need. This global community of clinicians, engineers,
9 manufacturers, regulators, and others are responding to this need by designing and producing, inexpensive,
10 and often open-source, ventilators of varying complexity and capabilities for rapid deployment. This
11 document identifies clinical, engineering and test requirements appropriate to support safe operation. The
12 document identifies requirements that are required for non-EUVs but might not be required for EUVs that
13 have appropriate disclosures. Therefore, ventilators complying with the requirements of this document need
14 not provide a level of performance equivalent to that of critical care ventilators (ISO 80601-2-12 1) or life-
15 supporting homecare ventilators (ISO 80601-2-722).
16 NOTE This document is intended to be used in conjunction with AAMI CR502:2020, End user disclosures for
17 emergency use ventilators (EUVs).
18 Introduction
19 The requirements outlined in this paper are modeled on ISO 80601-2-80:20183 presuming usage in
20 traditional healthcare facilities (e.g. hospitals, assisted living facilities, nursing homes) as well as spaces
21 converted for the care of large numbers of COVID-19 patients (e.g. convention centers, university
22 dormitories, motels). This paper presumes that the operators of the EUV are all trained professional
23 healthcare providers and not lay persons. Hence the requirements of ISO 80601-2-80:2018 specifically for
24 lay operators or the home healthcare environment are considered not applicable to an EUV intended for
25 the treatment of COVID-19 patients.
26 Fundamentally, the EUV needs to provide ventilation at the patient-connection port within the alarm limits
27 set by the operator or inform the operator via an alarm condition that ventilation within the alarm limits is
28 not occurring. Such alarm conditions need to include:
1 ISO 80601-2-12, Medical electrical equipment — Part 2-12: Particular requirements for basic safety and
essential performance of critical care ventilators
2 ISO 80601-2-72, Medical electrical equipment — Part 2-72: Particular requirements for basic safety and
essential performance of home healthcare environment ventilators for ventilator-dependent patients
3 ISO 80601-2-80, Medical electrical equipment — Part 2-80: Particular requirements for basic safety and
essential performance of ventilatory support equipment for ventilatory insufficiency
32 • Inspiratory and PEEP pressure not achieved (equivalent to disconnection alarm condition).
34 The ventilatory support needs of a COVID-19 patient can range from simple BIPAP (bilevel positive airway
35 pressure) for patients that are breathing spontaneously to mandatory ventilation in either a pressure-support
36 or volume control mode. Additionally, these patients are very likely to require inspired oxygen
37 concentrations (FiO2) in excess of the 21 % contained in room air.
38 To properly manage a COVID-19 patient, the EUV needs to indicate to the operator at a minimum:
39 • The current settings (e.g., inspiratory pressure, tidal volume, frequency, PEEP, FiO2, ventilation
40 mode).
41 • The current delivery (e.g., inspiratory pressure, tidal volume, respiratory rate, PEEP, and FiO2 at
42 the patient-connection port).
43 To properly manage a COVID-19 patient, the operator needs to be able to control the EUV at a minimum:
44 • FiO2 over the range of 21 % (ambient) to 95 % of the source oxygen concentration input to the EUV
45 in no more than 10 % steps
46 NOTE When oxygen is provided by an oxygen concentrator, the input concentration is not 99.5%, but can
47 vary from 90% to 96% in which case the upper limit of FiO2 would be 90 %.
48 • Set PEEP (i.e. BAP) (5 to 20) cmH2O in no more than 5 cmH2O steps
49 • I:E ratio (ratio of inspiratory to expiratory time) of 1:2 preferably adjustable from 1:1 to 1:3
50 • For mandatory modes, respiratory rate from (10 to 30) inflations/min preferably adjustable in steps
51 of no more than 2 inflations/min
52 • Tidal volume (350 to 450) ml ±10 % in no more than steps of 50 ml, preferably a lower range of
53 250 ml and an upper range of 600 ml or 800 ml
54 • Where applicable, inspiratory pressure limit (15 to 40) cmH2O preferably adjustable in steps of no
55 more than 5 cmH2O
56 To help prevent contaminating the environment (and particularly the clinicians), filters need to be placed in
57 the expiratory pathways. Particular attention needs to be placed on the exhaust port.
59 NOTE Any subclause marked with an asterisk (*) means that further guidance for this requirement is available in
60 Annex A of the standard.
67 NOTE Words written in SMALL CAPS are not ‘normal English’. They are defined terms and have specific, defined
68 meanings. See Clause 3 in the GS and 201.3 in ISO 80601-2-80 for their definitions.
70 Fully required.
71 These are the requirements for an EUV intended to connect to either an air or oxygen pipeline.
74 201.5.101 Additional requirements for the general requirements for testing of ME EQUIPMENT
75 Fully required.
76 This Clause explains how to interpret and perform tests as well as how to indicate specifications.
80 Unless there are electrical connections to the PATIENT (e.g. monitoring ACCESSORIES) or heated breathing
81 tubes or electrically powered ACCESSORIES (e.g. expiratory valves located proximal to the PATIENT), the
82 plastic breathing tubes provide adequate floating electrical isolation.
83 Protection from the ingress of water: IP21 is required and IP22 is recommended. Body fluids and IV bags
84 are an expected normal part of the environment of use.
85 Since the EUV is expected to handle gas with an oxygen concentration in excess of the ambient 25 %, the
86 considerations for an OXYGEN RICH ENVIRONMENT (see IEC 60601-1, 11.2.2) are fully applicable.
4 IEC 60601-1, Medical electrical equipment — Part 1: General requirements for basic safety and essential
performance
5 IEC 60601-1-2, Medical electrical equipment — Part 1-2: General requirements for basic safety and essential
performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
6 IEC 60601-1-6, Medical electrical equipment — Part 1-6: General requirements for basic safety and essential
performance - Collateral standard: Usability
7 IEC 60601-1-8, Medical electrical equipment — Part 1-8: General requirements for basic safety and essential
performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical
equipment and medical electrical systems
8 ANSI, https://webstore.ansi.org/
9 AAMI, https://my.aami.org/store/
88 This subclause is recommended but not required. An EUV need not be TRANSIT-OPERABLE.
89 Rationale: For pandemic treatment, a tabletop (i.e. somewhat large) EUV is acceptable.
91 7.1 General
93 Rationale: Although ensuring that the EUV can be read both over the indicated illumination level and the
94 indicated cone of visibility is recommended, in this pandemic situation it is not considered mandatory. It is
95 noted that operators are likely wearing PPE and will have reduced visual acuity. Consideration should be
96 given to doubling the distance of the observer.
123 These subclauses are required except for the portions of these subclauses relating to LAY OPERATORS that
124 are not required.
127 Elements e) and g) are not required as they are not relevant in this situation.
134 NOTE 1 Unless there are electrical connections to the PATIENT (e.g. monitoring ACCESSORIES) or heated breathing tubes
135 or electrically powered ACCESSORIES (e.g. expiratory valves located proximal to the PATIENT), the plastic breathing tubes
136 provide adequate floating electrical isolation for PATIENT LEAKAGE CURRENT.
137 NOTE 2 Commercially available ITC (information technology communications) power supplies can be used, but
138 electrical safety criteria (e.g. ENCLOSURE TOUCH CURRENTS and dielectric withstand) are likely to exceed IEC 60601-1
139 limits. This can be mitigated in several ways such as:
140 • use of a low leakage SEPARATION DEVICE (isolation transformer) (see 16.5 of the GS);
141 • a second PERMANENTLY INSTALLED PROTECTIVE EARTH CONNECTION (see 16.6 of the GS);
142 • instructing the OPERATOR to not touch the EUV and the PATIENT at the same time.
144 This Clause of the GS is recommended but not required, expect for 9.3 that is required.
145 201.9.4.3.101 Additional requirements for instability from unwanted lateral movement
147 Rationale: This requirement is for equipment intended to be used while moving in e.g. a car.
150 Rationale: This requirement is intended to make it easy to move the equipment around between uses. That
151 is not crucial for use during a pandemic.
154 Rationale: This test is hard to perform and takes expensive equipment to perform. It only provides
155 information for disclosure that is not crucial for use during a pandemic.
161 This subclause is only applicable if a heated humidifier is utilized. See ISO 80601-2-74.
166 The chosen materials for the GAS PATHWAYS need to be reasonably pure and simple in nature (minimize the
167 use of additives where possible). Avoid Polyvinyl chloride (PVC) in the GAS PATHWAYS. When possible,
168 efforts should be taken to use materials which have a long history of safe use in currently marketed medical
169 devices. Care is needed to ensure that gas pathways are free of foreign material (e.g. oil, particles, volatile
170 organic compounds, mold release agents should be avoided in the GAS PATHWAYS). Care is needed to
171 ensure that gas pathways do not contain toxic compounds (e.g., formaldehyde), and do not release noxious
172 gases (e.g., ozone, carbon monoxide) and fumes. The ACCOMPANYING DOCUMENTS should include
173 cautionary statement for any BIOCOMPATIBILITY identified RISK.
174 Rationale: The tests of ISO 18562 (series)10 are very expensive, time consuming to perform and require
175 very specialized test equipment. Requiring these tests for an EUV would so delay their availability such that
176 new designs would not be available when needed.
177 201.11.8.101 Additional requirements for interruption of the power supply/SUPPLY MAINS to ME
178 EQUIPMENT ALARM CONDITION
10 ISO 18562 (series), Biocompatibility evaluation of breathing gas pathways in healthcare applications
181 Rationale: The power back up and appropriate notification of power loss is what is important. It need not
182 be integrated into the EUV.
186 Clause 12 Accuracy of controls and instruments and protection against hazardous outputs
190 Rationale: These requirements are intended for home use by LAY OPERATORS.
204 201.13.2.102 * Independence of ventilation control function and related RISK CONTROL measures
212 Rationale: These requirements are directed to the needs of a LAY OPERATOR.
229 Rationale: These early warning maintenance-related requirements are not absolutely necessary in a
230 pandemic situation.
239 Rationale: These requirements are needed when there are LAY OPERATORS.
242 Rationale: The tests of IEC 60601-1-2 are time consuming and expensive set of tests that take very
243 specialized equipment. Requiring these tests for an EUV would delay availability such that new designs
244 might not be available when needed. Disclosure that these tests have not been performed and that other
245 equipment must be kept at a distance should be considered sufficient.
248 Rationale: USABILITY as described in IEC 60601-1-6 ensures safety by proscribing a design PROCESS. A
249 proper USABILITY evaluation is extremely time consuming and requires subject matter experts. A hard to use
250 EUV can be better than no EUV.
251 208 General requirements, tests and guidance for alarm systems in medical electrical equipment
252 and medical electrical systems
254 Rationale: Full compliance with IEC 60601-1-8 would be helpful to the OPERATORS as they would more
255 readily understand the operation of the EUV ALARM SYSTEM. Care needs to be taken with auditory ALARM
256 SIGNALS to ensure that they are not too obtrusive, appropriately priority encoded (so that more urgent
257 problems are more highlighted) and there must be a means to inactivate any auditory ALARM SIGNAL. The
258 ALARM SYSTEM, ALARM LIMITS, and ALARM CONDITION priorities are complex areas to optimize for USABILITY.
259 Annex A of IEC 60601-1-8 provides a great deal of guidance.
260 211 Requirements for medical electrical equipment and medical electrical systems used in the
261 home healthcare environment