Nothing Special   »   [go: up one dir, main page]

Kane 458 Operating Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 64

KANE458

Flue Gas Analyser with direct CO2


measurement and CO sensor
protection

Stock No: 19902 May 2017


© Kane International Ltd
Page 2
CONTENTS

Page No:

KANE458 Overview 5-6


ANALYSER LAYOUT & FEATURES 7-8
1. BATTERIES 9

2. BEFORE USING THE ANALYSER EVERY TIME 10-11


2.1 FRESH AIR PURGE 10
2.2 STATUS DISPLAY 11

3. USING THE FOUR FUNCTION BUTTONS 12-13

4. USING THE ANALYSER 14-27


4.1 COMBUSTION TEST 14-16
4.2 COMMISSIONING TEST 17
4.3 PRESSURE/TEMPERATURE TESTS 18-19
4.4 LET BY AND TIGHTNESS 20-21
4.5 ROOM CO TEST 22-24
4.6 KANE458 PRINTOUTS 25-27

5. USING THE MENU 28-31

6. USING THE KANE458 AS A THERMOMETER


OR PRESSURE METER 32-33

7. MEASURING FLUE GASES 34-35


CO Protection Pump Operation

8. ANALYSER PROBLEM SOLVING 36

9. FREQUENTLY ASKED QUESTIONS 37

Page 3
10. KANE DASHBOARD 38-39

11. ONLINE BOOKING PROCEDURE 40-42

12. ANALYSER ANNUAL SERVICE & RECERTIFY 43-46


12.1 SERVICE-CALIBRATE-RECERTIFY 43-44
12.2 RETURNING YOUR ANALYSER TO KANE 45
12.3 WHERE TO SEND YOUR ANALYSER 46

13. COLD WEATHER PRECAUTIONS 46

14. ANALYSER SPECIFICATION 47-48

15. ELECTROMAGNETIC COMPATIBILITY 49

16. END OF LIFE DISPOSAL 49

17. EN50379 REGULATED INSTRUCTIONS 50-52

APPENDIX 1 - MAIN PARAMETERS 53-55


ADDENDUM - OPTIONAL NITRIC OXIDE
(NO) SENSOR 56-59
PRODUCT REGISTRATION 61-62

Page 4
KANE458 OVERVIEW
The KANE 458 is based on the KANE456 with the addition of a dilution
pump which helps to protect the CO sensor from high concentration that
otherwise might damage the sensor.

The KANE458 Combustion Analyser measures carbon dioxide (CO2),


carbon monoxide (CO), differential temperature and differential pressure.
The direct measurement of CO2 is achieved using a Kane designed infra-
red sensing system. Below 1% CO2 the readings of CO2 are displayed to
two decimal places

CO2 is set to zero in fresh air automatically after the initial countdown.
The countdown varies between 90 and 60 seconds dependent on
ambient temperature.

If “RESET GAS ZERO” is indicated ensure that the unit is in outside


fresh air before pressing the button with an “Enter” symbol.

It is important that re-zeroing is done in outside fresh air as indoor CO2


levels are affected by human breath.

It calculates oxygen (O2), CO/CO2 ratio, losses, combustion efficiency


(Net or Condensing Gross).

The KANE458 Combustion Analyser can also measure CO levels in


ambient air - useful when a CO Alarm is triggered. It can also perform a
Room CO Test for up to 30 minutes duration.

A structured Commissioning Test has been included for the installation of


boilers.

The analyser has a protective rubber cover with a magnet for “hands–
free” operation and is supplied with a flue probe with integral temperature
sensor.

A low flow detection system warns of low flow and switches the pump off.
This also helps to prevent water ingress from overfilled water traps.
Its LCD display is protected with a toughened screen.

The large display shows 6 readings at a time and all data can be printed
via an optional infrared printer. The printed data can be 'live' data or
‘stored’ data.

Page 5
The memory can store up to:

60 combustion tests
20 AUX tests
20 let-by/tightness tests
20 temperature & pressure tests
20 room CO tests
20 commissioning Tests

Two lines of 20 characters can be added to the header of printouts.


Printouts can be made on the optional Kane IRP printers with ‘fast print’
capability using the IRP2 printer. Alternatively the analyser can be
equipped with optional wireless communications to either Android or
Apple devices.

The analyser is controlled using 4 function buttons and a rotary dial.

The four buttons (from left to right) switch on and off the analyser, switch
on and off the torch light, switch on and off the pump and send data to a
printer or to the memory. The buttons with UP, DOWN and ENTER
arrows also change settings such as date, time, fuel, etc. when in MENU
mode.

Page 6
ANALYSER LAYOUT & FEATURES

Tasklight and infra-red emitter

Battery
Compartment
(behind rubber
cover)

Menu controls
Function Scroll up/down
buttons x 4

Enter

Particle Filter

Rotary Switch

Water Trap
“Battery Charging”

Flue Gas Inlet

Page 7
Page 8
1. BATTERIES
Battery Type
This analyser has been designed for use with disposable alkaline batteries or
rechargeable Nickel Metal Hydride (NiMH) batteries. No other battery types are
recommended.

WARNING
The battery charger unit must only be used when NiMH batteries are fitted. Do
not mix NiMH cells of different capacities or from different manufacturers. All
four cells must be identical

Replacing Batteries
Turn over the analyser, remove its protective rubber cover and fit 4 “AA” batteries in
the battery compartment. Take great care to ensure they are fitted with the
correct battery polarity. Replace the battery cover and protective rubber cover.

Switch the analyser on and check that the analyser’s time and date are correct. To
reset see USING THE MENU, Section 5.

Charging NiMH Batteries


Ensure that you use the correct charger. The part number is 19278.

To fully charge NiMH batteries:


The charger must be connected and switched on.
When charging, the red Battery Charging Indicator will illuminate.
After a few seconds, the display will show “ CHARGING BATTERY” if they
need extra charge. If they are already fully charged this message will not
appear.

The first charge should be for 12 hours continuously. NiMH batteries are suitable for
top up charging at any time, even for short periods.

An in-vehicle charger can be used to top up the analyser's batteries from a 12 volt
vehicle battery. The part number is 18342.

Battery Disposal
Always dispose of depleted batteries using approved disposal methods that protect
the environment.

Page 9
2. BEFORE USING THE ANALYSER EVERY TIME:
Check the water trap is empty and the particle filter is not dirty:

- To empty water trap, unscrew the red screw plug and re-tighten once it is
empty.

- To change the particle filter, remove protective rubber cover, slide the water
trap unit from the analyser, remove the particle filter from its spigot and
replace. Reconnect the water trap unit and rubber protective cover.

Connect the flue probe hose to the analyser’s flue gas inlet and connect the flue
probe’s temperature plug to the T1 socket – check the plug’s orientation is correct -
see Page 6.

2.1 FRESH AIR PURGE


Position the flue probe in outside fresh air, then press / . The
analyser’s pump starts and the analyser auto-calibrates. When complete:

Select “Ratio” on the dial. In fresh air the CO reading should be zero.
Select “O2/Eff” on the dial. In fresh air the O2 reading should be 20.9% ± 0.3%.

FRESH AIR
PURGE
REMOVE
PROBE
FROM FLUE
PRESS 

WARNING
This message indicates that the analyser needs to be reset in fresh air. To do so,

ensure that the analyser is in outside fresh air and press / .


To perform a manual ‘Gas Zero’, select ‘Ratio’ on the dial, hold down the
key and you will see the message above.

Page 10
2.2 STATUS DISPLAY
Select “Status” on the dial to view the following:

NAT GAS → Current fuel selection. Press up/down to


change the fuel selection.

14:56:29 → Current time. Can be re-set via the “Menu”

11/03/06 → Current date. Can be re-set via the “Menu”

Ta 23.8C → Ambient temperature of the analyser


Shows number of days until next calibration is
CAL 283 → due

BAT → Shows the charge level of the batteries

SAFETY WARNING

This analyser extracts combustion gases that may be toxic in relatively low
concentrations. These gases are exhausted from the back of the instrument. This
analyser must only be used in well-ventilated locations by trained and
competent persons after due consideration of all the potential hazards.

Users of portable gas detectors are recommended to conduct a “bump” check before
relying on the unit to verify an atmosphere is free from hazard.

A “bump” test is a means of verifying that an instrument is working within acceptable


limits by briefly exposing to a known gas mixture formulated to change the output of
all the sensors present. (This is different from a calibration where the instrument is
also exposed to a known gas mixture but is allowed to settle to a steady figure and
the reading adjusted to the stated gas concentration of the test gas).

Page 11
3. USING THE FOUR FUNCTION BUTTONS:

Switching ON Rotate the dial to the mode you want to use before switching on.
the Analyser This may eliminate the need for a full countdown in some of the
modes and save you time.

Press / button to switch the unit ON. This must be


done in fresh air to ensure that the analyser auto calibrates its
sensors properly.
When switched on, the analyser beeps and briefly displays
software version, date and time. Its bottom line counts down until
the sensors are ready to use. If the analyser will not auto calibrate,
its sensors need to be replaced or recalibrated by an authorised
repair centre.
If an inlet temperature probe (optional) is connected into the T2
socket during its countdown, the measured temperature from the
inlet probe will be used as the inlet temperature.
If an inlet temperature probe is not connected to the analyser
during countdown the measured temperature from the flue probe
will be used as the inlet temperature.
If neither probe is connected during countdown the analyser’s
internal ambient temperature will be used as the inlet temperature.

Switching OFF Press / & hold for 2 seconds to switch the analyser
the Analyser OFF. The display counts down from 30 or less with the pump on to
clear the sensors with fresh air – If the probe is still connected,
make sure analyser and probe are in fresh air.

Press / if you want to stop the countdown and return


to making measurements.
Note: The analyser will not switch off unless the CO
reading is below 40ppm.

Torch Light Press / to switch the torch light on and off.


NOTE: Use of the torch light significantly increases the
current drain on the batteries.

Page 12
Switching The analyser normally operates with the pump on.
PUMP on / off
Press / to switch the pump off and on.

When the pump is switched off “-PO-" is displayed instead of the


O2, CO & CO2 readings. The analyser also displays "PUMP OFF"
on the top line approx every 30 seconds.

NOTE:
The pump will automatically switch itself off when the rotary
switch is set to Menu, Status, Pressure, Tightness or
Differential Temperature.

Zeroing the To re-zero the pressure sensor when “Prs/Temp” is selected on the
pressure
sensor dial, press and hold / until the top line display
shows CAL ZERO.
Always disconnect the pressure hose before zeroing.

Printing Data Press and quickly release / to start the analyser


printing. The analyser displays a series of bars until this is
completed. Press and release the key again to abort printing.

Make sure the printer is switched on, ready to accept data and its
infrared receiver is in line with the analyser’s emitter (on top of the
analyser).

Storing a set Press and hold / for approx. 2 seconds.


of readings
The top line briefly displays the log number.

Note: This STORE function is inhibited in normal operation if


the pump is switched off.

Using / The function buttons below the symbols / / are used


to navigate through the menu when the rotary switch is set to
/ MENU – See USING THE MENU, Section 5.
Buttons

Page 13
4. USING THE ANALYSER:

4.1 COMBUSTION TESTS:


Insert the tip of the flue probe into the centre of the flue. The readings will stabilise
within 60 seconds assuming the boiler conditions are stable.

The rotary switch can be used to display the following information:

RATIO Display

NAT GAS → Fuel type can be changed via “Menu” or


STATUS setting.

Ra 0.0008 → CO/CO2 ratio.

CO 52p → Carbon monoxide (ppm).

CO2 6.3% → Carbon dioxide (%).

XAIR 88.7 → Excess air %

PRS 0.01m → Pressure reading

Press / to print a full combustion test, (or send to PC via optional


Wireless module).

Hold / for 2+ seconds to log a full combustion report.

Page 14
O2/EFF display

NAT GAS → Fuel type


Oxygen (%) left after combustion. Should be
O2 9.8% → 20.9% ± 0.3% in fresh air.

Efg 81.1% → Gross efficiency

Tf 145.1C → Flue temperature (oC).


Inlet temperature (oC). Normally set by flue
Ta 5.4C → probe during fresh air purge.

ΔT 139.7C → Differential temperature

Press / to print a full combustion test, (or send to PC via optional


Wireless module).

Hold / for 2+ seconds to log a full combustion report.

AUX display

The AUX (auxillary) display can be


PRS 0.00m → customised via MENU / SCREEN / AUX.
The parameters displayed on lines 1, 2, 3, 4,
Ra 0.0008 → 5 and 6 can be set by the user.
They remain the AUX parameters until
CO 52p → changed by the user.

CO2 6.3% →

NO -N/F- → NO sensor not fitted

NOx -N/F- → NO sensor not fitted

Press / to print a full combustion test, (or send to PC via optional


Wireless module).

Hold / for 2+ seconds to log a full combustion report.

Page 15
Viewing / printing overview
The side lights on the display point to the active line.

Use or to change the pointer.

Press / to select a line. The side lights now flash.

Use or to scroll or change the selected line.

Press / to exit a line.

To view / print a logged report


Select MENU / REPORT / COMBUSTION / VIEW.

The side lights will point to the top line.

Press / to select this line. The side lights will flash.

Use or to scroll or change the Log No. ( If only one report is logged,
number will not change).

Press / to confirm a Log No. The side lights will stop flashing.

To view logged data press or to move the pointer to another line.

Press / . Sidelights will flash on that line.

Use or to scroll through data.

To finish, press / . Sidelights stop flashing.

Use or to scroll down to “PRINT”

Press / to print.

Viewing / printing a logged combustion test

Press / to print the test, (or send to PC via optional Wireless module).

Page 16
4.2 COMMISSIONING TEST
The Commissioning Test is based on TB143

Rotate the dial to COM TEST position and follow the instructions on the screen

TEST 1 check the boiler at Max Gas rate.

The boiler is switched on at Max rate.

The analyser is first zeroed in outside fresh air.

Once the boiler is stable at max gas flow rate the probe is inserted into the air inlet of
the flue and the CO2 level is measure. The reading needs to be stable and less than
or equal to 0.20%.

TEST 2

The probe is then inserted into the exhaust outlet of the boiler and the RATIO, CO
and CO2 levels are measured. These levels must be as per manufacturers
instruction. Where manufacturers instructions are not available the CO must be less
than 350 ppm and the RATIO must be less than 0.0040.

TEST 3 checks the boiler at minimum gas flow rate where this is possible.

With the boiler operating stably at minimum gas rate the RATIO, CO and CO2 levels
are measured.

These levels must be as per manufacturers instruction. Where manufacturers’


instructions are not available the CO must be less than 350 ppm and the RATIO
must be less than 0.0040.

TEST 4 Measures Flow and Return Temperatures from the boiler

All the measured readings are logged and can be printed or transmitted to PC if an
optional wireless module is fitted.

Page 17
4.3 PRESSURE/TEMPERATURE TESTING

WARNING

NEVER ATTEMPT TO TAKE A PRESSURE READING WITHOUT KNOWING THE


MAXIMUM PRESSURE THAT MIGHT BE PRESENT. THIS INSTRUMENT’S PRESSURE
TRANSDUCER IS RATED AT 80 MBAR WITH A MAXIMUM OVER RANGE OF 400 MBAR.

Select “Prs/Temp”. The pump stops automatically. Press / to auto-


zero the pressure sensor. Using the black connectors and manometer hose, connect
to P1 for single pressure or P1 and P2 for differential pressure.
PRS/TEMP display

Normal response or smoothed (damped)


PRS 0.01m → response can be selected via “Menu”.
‘High’ or ‘Low’ resolution readings can be
m= mbar → selected via “Menu”.

→ Pressure units can be selected via “Menu”.

T1 75.1C → Eg Flow Temp

T2 40.2C → Eg Return Temp

ΔT 34.9C → Differential Temp

Press / to print a full pressure test, (or send to PC via optional


Wireless module).

Hold / for 2+ seconds to log a pressure report.

Viewing / printing a logged pressure/temp test


Select MENU / REPORT / PRS-TEMP / VIEW

Use or to select the log number to be printed.

Press / to print the test, (or send to PC via optional Wireless module).

Page 18
WARNING
Before using the KANE458 to measure the pressure of a gas/air ratio valve, read the
boiler manufacturer’s instructions thoroughly. If in doubt contact the boiler
manufacturer.

After adjusting a gas/air ratio valve it is essential that the CO, CO2 and CO/CO2 ratio
readings are within the boiler manufacturer’s specified limits.

If using larger bore tubing when performing pressure tests:


Push ‘orange’ tube over the rim of the spigot to ensure a gas tight seal.


This may not produce a gas tight seal.

Page 19
4.4 LET-BY & TIGHTNESS TESTING

Select “Tightness”. The pump stops automatically. Press / to auto-


zero the pressure sensor. Connect from the test point to P1 using a black connector
and manometer hose.

The display shows “LET BY?”. Use , and to select YES or NO.

If YES is selected set the let-by pressure then press to start the let-by test. The
display shows:

The let-by test is automatically stored in the


LET BY → memory.

PR1 10.15m → Pressure at start of let-by test.

PR2 10.15m → Real time pressure reading.


Let-by default time is 1 minute. Can be
TIME 59 → changed via “Menu”.

If the let-by test fails simply move the rotary switch to any position other than
“tightness” to abort the test.

If the let-by test passes adjust the gas pressure for the tightness test and press
to start the stabilisation test. The display shows:

STABIL’N →

PR1 20.01m → Real time pressure reading.

Stabilisation default time is 1 minute. Can be


TIME 59 → changed via “Menu”.

When complete press to start the tightness test:

Page 20
TIGHTNESS →

PR1 20.01m → Pressure at start of tightness test.

PR2 20.01m → Real time pressure reading.

Tightness default time is 2 minutes. Can be


TIME 119 → changed via “Menu”.

When complete the display will show:

The tightness test is automatically stored in


LOG 01 → the memory.

PR1 20.01m → Pressure at start of stabilisation test.

PR2 19.98m → Pressure at end of stabilisation test.

PR1 19.98m → Pressure at start of tightness test

PR2 19.97m → Pressure at end of tightness test

PRINT  → Press to print the complete test.

Viewing / printing a logged Let-by and Tightness test


Select MENU / REPORT / TIGHTNESS / VIEW

Use or to select the log number to be printed.

Press / to print the test, (or send to PC via optional Wireless module).

Note: The analyser’s memory can store up to 20 tightness tests. Tightness tests
are logged automatically therefore the tightness section of the memory will
be full after the 20th tightness test is complete. Before the 21st tightness
test can be performed the tightness section of the memory must be
cleared. To do this select MENU / REPORT / TIGHTNESS / DEL ALL /
YES then press .

Page 21
4.5 ROOM CO TESTING
Select “Room CO” to measure and record CO readings for up to 30 minutes.

SELECT →

ROOM CO →

TEST TYPE →

GENERAL → Use /
the following:
to select the test type from

TEST TYPES

GENERAL: 15 minute test with results LIMIT = 10ppm


stored every minute ALARM = 30 ppm

SWEEP TEST: 2 minute test with max LIMIT = 10ppm


reading stored at end ALARM = 30 ppm

MIGRATION TEST: 15 minute test with results LIMIT = 10ppm


stored every minute ALARM = 30 ppm

TYPE C SEALED APPLIANCE: 15 minute test with results LIMIT = 10ppm


stored every minute ALARM = 30 ppm

TYPE B BOILER OPEN FLUE: 15 minute test with results LIMIT = 10ppm
stored every minute ALARM = 30 ppm

TYPE A COOKER: 30 minute test with results LIMIT = 30ppm


stored every minute ALARM = 90 ppm

TYPE A WATER HEATER: 5 minute test with results LIMIT = 10ppm


stored every minute ALARM = 30 ppm

TYPE A SPACE HEATER: 30 minute test with results LIMIT = 10ppm


stored every 1 minute ALARM = 30 ppm

Press / to start Room CO testing.

Page 22
ROOM CO display

CO readings are recorded every minute for


ROOM CO → up to 30 minutes.
Test 00 = initial CO test in series.
TEST 1/15 → Test 30 = maximum of 30 tests in series.

INT. 31s → Test Interval Time

CO 0p → Real time CO reading (ppm).

LIMIT 10p
ALARM 30p
The user can stop the Room CO test at any time by pressing / .

If not stopped earlier, the Room CO test will automatically end after the designated
time.

The CO test series is automatically stored in the memory as a log number.

When completed the log can be printed immediately by pressing .

Viewing / printing a logged Room CO test


Select MENU / REPORT / ROOM CO / VIEW

When LEDs are not flashing

Use the / keys to change line.

Press to cause the LEDs on that line to flash.

With the LEDs flashing, press / to allow the parameter on that line to be
changed.

Press to select that change.

The LEDs stop flashing and /

Use / to change the line again.

Page 23
ROOM CO →

LOG 01 →

TEST 3 →

CO 0p →

GENERAL → TEST TYPE

PRINT  →

With no LEDs flashing

Use the / keys to move the lit LEDs to the line you want.

You can change the LOG number and the TEST number so that you can view
individual test results.

Press / to select the line you want and the LEDs will start to flash.

Now use the / keys to change the number (the TEST number or the LOG
number)

Press / when you are happy with the changes. The LEDs will stop
flashing. Now use the / keys to move the LEDs to the PRINT line.

Sending to the printer or wireless device will only occur when you move the LEDs to

the print line and press / .

Press / to print the test, (or send to PC via optional Wireless module).

Page 24
4.6 PRINTOUTS
COMBUSTION COMMISSION TEST PRS/TEMP

KANE458 SW19392 V1.00 KANE458 SW19392 V1.00 KANE458 SW19392 V1.00

YOUR COMPANY NAME & YOUR COMPANY NAME & YOUR COMPANY NAME &
PHONE NUMBER HERE PHONE NUMBER HERE PHONE NUMBER HERE

SERIAL NO. 9876543210 SERIAL NO. 9876543210 SERIAL NO. 9876543210

LOG NO. 01 LOG NO. 08 LOG NO. 20

DATE 19/01/16 DATE 19/01/16 DATE 19/01/16


TIME 11:06:09 TIME 11:50:04 TIME 12:23:59

CAL DUE ON 18/12/16 CAL DUE ON 18/12/16 CAL DUE ON 18/12/16

COMBUSTION COMMISSION TEST PRS/TEMP


FUEL TYPE NAT GAS PRS mbar 18.01
o
ANALYSER ZERO T1 C 75.5
o
CO2 % 9.0 CO2 % 0.00 T2 C 65.2
o
O2 % 5.1 CO ppm 0 ΔT C 10.3
CO ppm 50
NO ppm -N/F- FLUE INTEGRITY CUSTOMER
NOx ppm -N/F- CO2 % 0.00
o
FLUE C 65.2 MAX GAS FLOW
o
INLET C 17.2 CO2 % 9.1 APPLIANCE
o
NETT C 48.0 CO ppm 50
CO/CO2 0.0005
CO/CO2 0.0005
MIN GAS FLOW REFERENCE
NET % 97.9 CO2 % 9.0
LOSS % 2.1 CO ppm 48
XAIR % 32 CO/CO2 0.0005

PRS mbar 0.21 FLOW & RETURN


o
T1 C 65.5
o
CUSTOMER T2 C 48.2
o
ΔT C 17.3

CUSTOMER
APPLIANCE

APPLIANCE
REFERENCE

REFERENCE

Page 25
SWEEP TEST TYPE C SEALED TIGHTNESS TEST
APPLIANCE

KANE458 SW19392 V1.00 KANE458 SW19392 V1.00 KANE458 SW19392 V1.00

YOUR COMPANY NAME & YOUR COMPANY NAME & YOUR COMPANY NAME &
PHONE NUMBER HERE PHONE NUMBER HERE PHONE NUMBER HERE

SERIAL NO. 9876543210 SERIAL NO. 9876543210 SERIAL NO. 9876543210

LOG NO. 06 LOG NO. 03 DATE 19/01/16


TIME 11:09:16
DATE 19/01/16 DATE 19/01/16
TIME 10:11:11 TIME 12:25:27 CAL DUE ON 18/12/16

CAL DUE ON 18/12/16 CAL DUE ON 18/12/16 LET BY TEST


PRS 1 mbar 10.80
ROOM CO ROOM CO PRS 2 mbar 10.78
SWEEP TEST TYPE C LET BY MINS 1:00
SEALED
LIMIT 10ppm APPLIANCE TIGHTNESS TEST
ALARM 30ppm PRS 1 mbar 20.14
TESTS 1 LIMIT 10ppm PRS 2 mbar 20.13
ALARM 30ppm ΔP mbar -0.01
M CO ppm TESTS 15
01 0 STABILIS’N MINS 1:00
MAXIMUM CO ppm 0 TEST CO ppm TIGHTNESS MINS 2:00
01 0
02 0 CUSTOMER
CUSTOMER 03 0
04 0
05 0
06 0 APPLIANCE
APPLIANCE 07 0
08 0
09 0
10 0 REFERENCE
REFERENCE 11 0
12 0
13 0
14 0
15 0
MAXIMUM CO ppm 0

CUSTOMER

APPLIANCE

REFERENCE

Page 26
AUX

KANE458 SW19392 V1.00

YOUR COMPANY NAME &


PHONE NUMBER HERE

SERIAL NO. 9876543210

LOG NO. 01

DATE 19/01/16
TIME 09:46:53

CAL DUE ON 18/12/16

AUX
FUEL TYPE LIGHT OIL

CO2 % 0.53
CO(n) ppm 02++
CO/CO2 0.0000
O2 % 20.2
o
FLUE C -N/F-
o
INLET C -N/F-

CUSTOMER

APPLIANCE

REFERENCE

Page 27
5. USING THE MENU
Select “Menu” on the rotary switch and navigate using the function buttons:

 = Scroll up  = Scroll down = Enter

NOTE: To EXIT the MENU at any time simply move the rotary switch to any
position other than “Menu”. Any changes that have not been
“entered” will be ignored.

MENU →
SETUP →
UNITS →
SCREEN →
REPORT →
SERVICE →

As you scroll up or down the side LEDs illuminate to point to the active line

MAIN MENU SUB MENU OPTIONS / COMMENTS

SETUP Language English

HH:MM:SS format
SET TIME
e.g. 7 am = 07:00:00, 7pm = 19:00:00

SET DATE DD/MM/YY format

KM IRP
KANE IRP-2
PRINTER
WIRELESS
SERIAL

PASSKEY 1111 ( wait 5 secs after entering last digit)

BACK

As you scroll up or down the side LEDs illuminate to point to the active line

Page 28
NOTE: To EXIT the MENU at any time simply move the rotary switch to any
position other than “Menu”. Any changes that have not been
“entered” will be ignored.

SUB
MENU OPTIONS / COMMENTS
MENU
NAT GAS, TOWN GAS, COKE GAS, PROPANE,
UNITS Fuel Type BUTANE, LPG, LIGHT OIL, BIO OIL, WOOD
PELLETS, BIO GAS, NAT GAS 1 to 5

UK, FRANCE, SPAIN, N AMERICA, BELGIUM,


Fuel Origin
NETHERLAND

EFFICIENCY
GROSS, NET, GROSS COND, NET COND

PRESSURE See next table below

GAS ppm, ppm(n), mg/m3, mg/m3(n), mg/kWh,


mg/kWh(n)
TEMP C,F

O2 REF Up/down to set value (3% default)

NOx CALC Up/down to set value (5% default)

BACK

As you scroll up or down the side LEDs illuminate to point to the active line

NOTE: To EXIT the MENU at any time simply move the rotary switch to any
position other than “Menu”. Any changes that have not been
“entered” will be ignored.

Page 29
MENU SUB MENU OPTIONS / COMMENTS

OFF = normal response.


PRESSURE FILTER
ON = slower (damped) response

LOW = e.g. 0.01mbar resolution.


RESOLUTION
HIGH = displays to an extra decimal place

UNITS mbar, Pa, PSI, mmHg, hPa, inH2O, mmH20, kPa,


psi
LET BY = Set duration of let-by test in minutes.
Default = 1 minute
STABIL’N = Set duration of stabilisation in minutes.
TIME
Default = 1 minute
TIGHTN’S = Set duration of tightness test in
minutes. Default = 2 minute

BACK

As you scroll up or down the side LEDs illuminate to point to the active line

NOTE: To EXIT the MENU at any time simply move the rotary switch to any
position other than “Menu”. Any changes that have not been
“entered” will be ignored.

MENU SUB MENU OPTIONS / COMMENTS

SCREEN CONTRAST Factory setting is 14

BACKLIGHT 0 to 300 secs

Enables users to customise the parameters on the


AUX AUX display: LINE 1, LINE 2, LINE 3, LINE 4,
LINE 5, LINE 6, BACK

BACK

As you scroll up or down the side LEDs illuminate to point to the active line

NOTE: To EXIT the MENU at any time simply move the rotary switch to any
position other than “Menu”. Any changes that have not been
“entered” will be ignored.

Page 30
MENU SUB MENU OPTIONS / COMMENTS

Stored AUX tests


REPORT AUX
VIEW, DEL ALL, BACK

Stored combustion tests:


COMBUSTION
VIEW, DEL ALL, BACK

Stored commission tests:


COMMISSION
VIEW, DEL ALL, BACK

Stored pressure tests:


PRS/TEMP
VIEW, DEL ALL, BACK

Stored tightness tests:


TIGHTN’S
VIEW, DEL ALL, BACK

Stored room CO tests:


ROOM CO
VIEW, DEL ALL, BACK

LINE 1
HEADER LINE 2
BACK

BACK

As you scroll up or down the side LEDs illuminate to point to the active line

NOTE: To EXIT the MENU at any time simply move the rotary switch to any
position other than “Menu”. Any changes that have not been
“entered” will be ignored.

MENU SUB MENU OPTIONS / COMMENTS

Password protected for authorised service agents


SERVICE CODE
only. Leave set to 000000.

As you scroll up or down the side LEDs illuminate to point to the active line

NOTE: To EXIT the MENU at any time simply move the rotary switch to any
position other than “Menu”. Any changes that have not been
“entered” will be ignored.

Page 31
6. USING THE KANE458 AS A THERMOMETER OR
PRESSURE METER
Rotate the dial to the PRS/TEMP position.

The display will show:

P 0.00m → Real time pressure reading.

m= mbar
Use the T1 connection for the flow
T1 21.3C → temperature sensor.
Use the T2 connection for the return
T2 21.3C → temperature sensor.

∆T 0.0C → Real time temperature difference.

The standard printout for this mode is as follows:

KANE458 SW19392 V1.00

YOUR COMPANY NAME &


PHONE NUMBER HERE

SERIAL NO. 9876543210

LOG NO. 20

DATE 19/01/16
TIME 12:23:59

CAL DUE ON 18/12/16

PRS/TEMP
PRS mbar 0.01
o
T1 C 17.6
o
T2 C 17.4
o
ΔT C 0.2

CUSTOMER

APPLIANCE

REFERENCE

Page 32
If using larger bore tubing when performing pressure tests:


Push ‘orange’ tube over the rim of the spigot to ensure a gas tight seal.


This may not produce a gas tight seal.

Page 33
7. MEASURING FLUE GASES
After the countdown is finished and the analyser is correctly set up, put its flue probe
into the appliance’s sampling point. The tip of the probe should be at the centre of the
flue. Use the flue probe’s depth stop cone to set the position.

With balanced flues, make sure the probe is positioned far enough into the flue so no
air can ‘back flush’ into the probe.

NOTE: Ensure that the flue probe handle does not get hot!

Make sure you do not exceed the analyser’s operating specifications. In particular:

 Do not exceed the flue probe’s maximum temperature (600oC)


 Do not exceed the analyser’s internal temperature operating range
 Do not put the analyser on a hot surface
 Do not exceed the water trap’s levels
 Do not let the analyser’s particle filter become dirty and blocked

View the displayed data to ensure that stable operating conditions have been
achieved and the readings are within the expected range.

Press and quickly release / to start the analyser printing. The analyser
displays a series of bars until this is completed. Press and release the key again to
abort printing.

Make sure the printer is switched on, ready to accept data and its infrared receiver is
in line with the analyser’s emitter (on top of the analyser).

CO PROTECTION PUMP OPERATION

CO Protection pump operation is totally automatic. When the analyser measures a


CO concentration of 2000ppm the pump is switched on. The main pump is
switched off and the display shows P-OFF until the gas in the sensor is below
2000ppm.

Page 34
.

WOOD PELLS →Fuel type can be changed via “Menu” or


STATUS

R 0.0448 →CO/CO2 ratio.

CO P-OFF →Pump Off pump operating

CO2 P-OFF →Carbon dioxide (%).

XAIR 88.7 →Excess air %

PRS 0.01m →Pressure reading

DILUT flashes on the screen to indicate that the CO measurement has been diluted.

The time taken for the CO sensor to return to zero takes much longer after it has
been exposed to high levels of CO.

The CO high warning still operates whenever the CO reading is above 400 ppm until
the dilution pump operates..

Page 35
8. ANALYSER PROBLEM SOLVING
If any problems are not solved with these solutions, contact us or an authorised
repair center.

Fault symptom Causes / Solutions

 Oxygen too high  Air leaking into probe, tubing, water trap,
 CO2 too low connectors or internal to analyser.

 Batteries not holding charge  Batteries exhausted.


 Analyser not running on mains  AC charger not giving correct output.
adapter.  No fuse
 Analyser does not respond to flue  Particle filter blocked.
gas  Probe or tubing blocked.
 Pump not working or damaged with
contaminants.
 Net temperature or Efficiency  Ambient temperature set wrong during
calculation incorrect. Automatic Calibration.
 Flue temperature readings erratic  Temperature plug reversed in socket.
 Faulty connection or break in cable or
plug.
 T flue or ΔT displays (-N/F-)  Probe not connected.
 Faulty connection or break in cable or
plug.
EFF or X-Air displays (- O2++-)  CO2 reading is below 2%.
 02 > 18%
 Analyser just continually beeps  Turn dial back to MENU and press ENTER
 Turn dial back to Tightness and press
ENTER

Page 36
9. FREQUENTLY ASKED QUESTIONS
Q: What is the countdown time on a KANE458

A: There are two levels of countdown (aka fresh air purge) on a KANE458.
From first switch on if ‘cold’ (more than 5oC from the temperature at which
calibrated) = 90 secs.
From first switch on if ‘warm’ (within 5oC of the temperature at which
calibrated) = 60 secs

Page 37
10. KANE DASHBOARD

We really want you to register your Kane analyser at www.kane.co.uk


to create your dashboard - your personal online analyser portal

Benefits include:

An automatic reminder when it's time to Service - Calibrate –


Recertify

Our new simple online booking in procedure ensures you


receive the fastest turnaround times possible

Easy access to your analyser's calibration certificates & service


history

A record of your model & serial number with ability to report your
analyser stolen

Product specific information & resources all in one place

Fixed service prices are always visible

Page 38
An out of calibration analyser may mean your combustion
reports are invalid.

Page 39
11. ONLINE BOOKING PROCEDURE @ www.kane.co.uk
To organise your Kane analyser’s Service & Recertification online it must
be registered.

To register, click

If you have already registered, click to sign in

From your dashboard, click for the analyser you wish to

organise Service & Recertification, then click

& follow our simple online booking process:

1. This page confirms if a service/recertification is required & how


much it will cost

Page 40
2. Book your service/recertification & order any additional spare parts
you need

3. Once happy with your order just confirm & pay

Page 41
4. See your order & print out the return to Kane label

Don’t forget to put the label on your package and send it to us

Page 42
12. ANALYSER ANNUAL SERVICE & RECERTIFY
Although sensor life is typically more than five years, the analyser should
be serviced and re-certified annually to counter any long-term sensor or
electronics drift or accidental damage.

Local regulations may require more frequent re-certification.

Kane International has service facilities at Atherton near Manchester Tel:


01942-873434 (the primary service centre for UK customers) and at
Welwyn Garden City in Hertfordshire Tel: 01707-375550 (the primary
service centre for non-UK customers).

By sending your analyser back to Kane for an annual fixed price service
(check www.kane.co.uk for details) you have the opportunity to extend
the warranty on your analyser to 5 years.

12.1 SERVICE - CALIBRATE - RECERTIFY


YOU CAN COUNT THE REASONS WHY ON ONE HAND!

1. Fixed prices ensure you know the full cost of ownership before you
buy.

2. Flue Gas Analysers will drift out of calibration over time – only Kane
can adjust them back to the manufacturers' specification & return
them as accurate as when they were purchased.

3. Analysers always seem to require a service when you need them


most! Kane's two UK Service Centres pride themselves on their fast
service turnaround times.

4. Every Kane analyser is returned with a fully traceable calibration


certificate & another 12 months warranty. After 6 years, warranty
becomes limited to replacement parts only.

5. In some cases an out-of-calibration analyser means any combustion


reports you produce are invalid, so it is very important you get your
analyser recertified

Page 43
KANE IS NOW THE UK'S LARGEST, FASTEST & MOST
RELIABLE FGA AFTER SALES SERVICE PROVIDER

 Our Northern & Southern Service Centres are the UK's only
locations authorized to Service - Calibrate - Recertify & Repair
Kane analysers

 You can organise your Kane analyser’s Service & Recertification


online via your Dashboard on www.kane.co.uk, send it or call us to
arrange a pre-booked while-you-wait service, saving you time &
potential lost revenue

We pride ourselves on our fast turnaround for a fixed price, which


includes:
- Inspecting your analyser & accessories, testing its functionality &
adjusting it to our original specification
- Free software upgrades, where applicable
- Parts & labour on these replaceable parts as required: CO &
CO2/O2 sensors, pump & battery
- Parts not included are optionally fitted toxic cells, cases & probes
- All internal tubing, water trap, particle & chemical filters
- Recertifying to national standards & issuing a fully traceable
calibration certificate to confirm your analyser's performance
- A complimentary Service Pack comprising PF400/5 (pack of five)
replacement filter elements, Thermal Printer Roll, Water Trap Plug
& Probe Connector (RRP for filters & roll = £17.40)
- Return carriage via Next Day delivery service (UK mainland only)

 Kane is the only UK manufacturer of hand-held gas analysers


granted UKAS accreditation to ISO / IEC 17025:2005 for its Welwyn
Garden City calibration laboratory, covering a range of gases,
temperature & pressure. Kane is certified to ISO 9001, ISO 14001 &
BS OHSAS 18001

 Our Northern Service Centre offers service & repair for your other
instruments, including: thermometers, pressure meters &
multimeters

Page 44
12.2 RETURNING YOUR ANALYSER TO KANE
Before returning your analyser to Kane, please ensure that you enclose:

- RMA label if you have used our new simple online booking in process
- Your full contact details
- A daytime telephone number
- Details of faults you might have experienced
- Any relevant accessories (eg. probe, printer, adaptor & leak detectors).
Any accessories that are returned will be checked

PACKING YOUR ANALYSER:

When returning an analyser with its probe, please send them back in
their carry bag. The bag should be put into a suitably sized box (approx.
45cm x 20cm x 23cm)

When only returning an analyser, use a container as big as a shoe box &
pack out empty space with newspaper

Before sealing your package, please ensure you followed our procedure
above & have clearly marked your box for the Kane Service Team

If you do not have an account with a courier company, take your


package to your local Post Office – we recommend Special Delivery so
that it is insured & traceable in transit

WHEN WE RECEIVE YOUR ANALYSER:

Our Service Engineers will inspect the analyser & accessories. If you
haven't booked in & paid online, they will confirm the total service cost.

Once accepted, the work will be carried out & on completion, returned to
you by Next Day delivery service (UK mainland only)

Page 45
12.3 WHERE TO SEND YOUR ANALYSER
Northern Customer Service Southern & International Customer Service
Kane International Ltd Kane International Ltd
Gibfield Park Avenue Kane House, 11 Bessemer Road
Atherton, Welwyn Garden City
Manchester Hertfordshire
M46 0SY, UK AL7 1GF, UK
t: +44 (0) 1942 873434 t: +44 (0) 1707 384834
f: +44 (0) 1942 873558 f: +44 (0) 1707 384833
e: nservice@kane.co.uk e: sservice@kane.co.uk

13. COLD WEATHER PRECAUTIONS


It is important you keep your flue gas analyser in a warm place overnight

Electronic devices that become really cold, by being left in a vehicle


overnight, suffer when taken into a warm room the next morning.
Condensation may form which can affect the analyser’s performance &
cause permanent damage

Electrochemical sensors used in flue gas analysers can be affected by


condensation or water being sucked into the analyser, as the small
apertures on top of sensors can become blocked with water, stopping
sensors seeing flue gas. When this happens, oxygen or carbon dioxide
reading will display as "—" & sensors may be permanently damaged

If you think that your analyser is affected by condensation or water


ingress, it may be possible to rectify the problem yourself. Simple leave
the analyser running in a warm place, with the pump 'ON' sampling fresh
air for a few hours (use mains adapter/battery charger if needed). If,
after doing this, you still experience problems please contact our Service
Centres.

Page 46
14. ANALYSER SPECIFICATION
(NOTE: MAY BE SUBJECT TO CHANGE)

Parameter Range Resolution Accuracy

Temp Measurement
Flue Temperature 0-600oC 0.1oC +2.0oC
+0.3% reading
Inlet Temperature 0-50oC 0.1oC +1.0oC
(Internal sensor) +0.3% reading
Inlet Temperature 0-600oC 0.1oC +2.0oC
(External sensor) +0.3% reading

Flue Gas Measurement


Oxygen*2 0-21% 0.1% +0.3%
*1
Carbon monoxide 0-20ppm 1ppm +3ppm
21-2,000ppm +5% reading

Above 2000ppm
Purge pump unspecified
operates
Carbon dioxide *1 0-20% 0.1% +0.3% volume
Efficiency (Net or Gross) *2 0-99.9% 0.1% +1.0% reading
Efficiency High (C) *2 0-119.9% 0.1% +1.0% reading
*2
Excess Air 0-250% 0.1% +0.2% reading
CO/CO2 ratio *2 0-0.999 0.0001 +5% reading
Pressure (differential)
Nominal range +80mbar +0.2 mbar Maximum +0.005 mbar
Maximum over range without 0.001 mbar
damage to sensor is +1 mbar <25mbar +0.03 mbar
+400mbar +80 mbar +3% of reading
Pre-programmed Fuels Natural gas, Propane, Butane, LPG, Light Oils (28/35 sec),
Wood Pellets, Town Gas, Coke Gas, Bio Oil, Bio Gas
User programmed Fuels 5 user defined fuels

Storage Capacity 60 Combustion tests


20 Pressure & Temperature tests
20 Tightness tests
20 Temperature tests
20 Room CO tests
20 Commissioning tests
*1
Using dry gases at STP
*2
Calculated

Page 47
Carbon Dioxide resolution is 0.01% below 1% measured value.

Ambient Operating Range 0oC to +45oC 10% to 90% RH non-condensing


Battery Type / Life 4 AA cells
>8 hours using Alkaline AA cells
Chargers (optional) 100-240v charger, for NiMH batteries only
12v in vehicle charger, for NiMH batteries only
Dimensions
Weight: 0.8kg handset with protective rubber cover
Handset: 200 x 45 x 90mm
Probe: 300mm long including handle.
6mm diameter x 240mm long stainless steel shaft with
2m long neoprene hose. Type K thermocouple
CO Protection Pump: Operates at 2000ppm measured CO.

Page 48
15. ELECTROMAGNETIC COMPATIBILITY

European Council Directive 89/336/EEC requires electronic equipment not to


generate electromagnetic disturbances exceeding defined levels and have adequate
immunity levels for normal operation. Specific standards applicable to this analyser
are stated below.
As there are electrical products in use pre-dating this Directive, they may emit excess
electromagnetic radiation levels and, occasionally, it may be appropriate to check the
analyser before use by:
Use the normal start up sequence in the location where the analyser will be
used.
Switch on all localized electrical equipment capable of causing interference.
Check all readings are as expected. A level of disturbance is acceptable.
If not acceptable, adjust the analyser’s position to minimize interference or
switch off, if possible, the offending equipment during your test.
At the time of writing this manual (January 2017) Kane International Ltd are not
aware of any field based situation where such interference has occurred and this
advice is only given to satisfy the requirements of the Directive.

This product has been tested for compliance


with the following generic standards:

EN 61000-6-3 : 2011
EN 61000-6-1 : 2007

and is certified to be compliant

Specification EC/EMC/KI/KANE458 details


the specific test configuration, performance
and conditions of use.

16. END OF LIFE DISPOSAL


The Waste Electrical or Electronic Equipment (WEEE) Directive requires countries in
the EU to maximise collection and environrmentally responsible processing of these
items.

Products are now labelled with a crossed out wheeled bin symbol to remind you that
they can be recycled.

Please Note: Batteries used in this instrument should be disposed of in accordance


with current legislation and local guidelines.

Page 49
17. EN 50379 REGULATED INSTRUCTIONS
EN 50379 Section 4.3.2 “Instructions” defines a number of specific points that must
be included in the relevant instruction manuals. The paragraph numbering below
relates to that section of EN 50379.

a) The KANE458 is compliant the EN 50379 Part 2 and Part 3.

b) The KANE458 is intended to be used with the following fuels:


Natural gas
Light oil (28/35 sec)
Propane
LPG
Wood pellets
Butane

c) The KANE458 is designed for use with either non-rechargeable alkaline AA


cells or rechargeable NiMH AA cells. Four cells are needed. Types cannot be
mixed. Under no circumstances should any attempt be made to recharge
alkaline cells.

The battery charger supplied with the KANE458 is rated for indoor use only. Its
voltage input must be in the range 100 – 240 V ac at 50 – 60 Hz with a current
capability of 0.3 A. The chargers output voltage is 9 V dc at a maximum of
0.66A.

The charger has no user serviceable components.

Only a correctly specified and rated charger must be used with the KANE458.

d) The KANE458 is not designed for continuous use and is not suitable for use as
a fixed safety alarm.

e) An explanation of all the symbols used on the analyser’s display is given in


Appendix 1 of this manual.

f) The recommended minimum time required to perform one complete


measurement cycle and achieve correct indication of the measured values in
EN 50379 Part 2 is 110 seconds. This is based on the T90 times defined in the
standard, always assuming that parameters being measured have reached
stability. This time is the summation of the times for a draught test (10 secs)
and a combustion test (90 secs) plus the time to move the hose connection
from the pressure input to the water trap (10 secs)

Page 50
g) The recommended minimum time required to perform one checking procedure
in EN 50379 Part 3 is 110 seconds as described in section f) above.

h) Some commonly occurring materials, vapour or gases may affect the operation
of the KANE458 in the long or the short term though in normal use Kane
International Ltd is not aware of any specific issues that have affected the
product. The following list is included to satisfy the stated requirements of
EN 50379:
Solvents
Cleaning fluids
Polishes
Paints
Petrochemicals
Corrosive gases

i) The KANE458 is fitted with an electrochemical CO sensor and an infra-red CO2


sensor which have an expected life of more than 5 years. The calibration of
these sensors must be confirmed on an annual basis.

The batteries have an expected operational life of more than 500 re-charge
cycles.

j) The KANE458 is designed to operate at ambient temperatures in the range 0oC


to +45oC with relative humidity of 10% to 90% non-condensing. Whilst it is
recommended that the analyser is given the protection of a carry case during
transportation it is not required for normal operation.

k) The KANE458 has an initial start up delay following switch on of between 90


and 60 seconds dependent on ambient temperature. There is no additional
delay after battery replacement.

l) Most sensors used in combustion analysers give a zero output when they fail
and it is widely recommended that analysers are regularly checked (also known
as a bump test) using either a can of test gas or a known source of combustion
products.

The KANE458 must have its calibration checked on an annual basis and be
issued with a traceable Certificate of Calibration.

The sensor within the KANE458 can only be replaced by Kane International Ltd
or one of its trained and approved service partners.

Page 51
The water trap should be checked on a regular basis whilst the analyser is in
use (every few minutes) as the amount of condensate generated varies with the
fuel type, atmospheric conditions and the appliances operating characteristics.

The particle filter should be checked at least on a daily basis when using ‘clean’
fuels and more often when using liquid or solid fuels.

Detailed instructions regarding the changing of the filter and the emptying of the
water trap are given in Section 2 of this manual.

m) WARNING!
When using a KANE458 to test an appliance a full visual inspection of the
appliance, in accordance with its manufacturer’s instructions, must also be
carried out.

Page 52
Appendix 1 - Main Parameter:
Here are the legends used and what they mean:

O2 : Oxygen (Calculated) reading in percentage (%)

CO : Carbon monoxide (Measured) reading displayed in ppm (parts per


million). If ‘- - - -’ is displayed there is a fault with the CO sensor or
the instrument has not set to zero correctly. Switch off instrument
and try again.

COn carbon monoxide normalised

CO2 : Carbon dioxide (Measured) reading in percentage (%).

Ra: CO to CO2 ratio

Tf : Temperature measured by the flue gas probe in centigrade (oC). It


displays ‘- OC -’ if the flue probe is disconnected or faulty.

Ti : If an inlet temperature probe (optional) is connected into the T2


socket during its countdown, the measured temperature from the
inlet probe will be used as the inlet temperature.

If an inlet temperature probe is not connected to the analyser during


countdown the measured temperature from the flue probe will be
used as the inlet temperature.

If neither probe is connected during countdown the analyser’s


internal ambient temperature will be used as the inlet temperature.

ΔT : Nett temperature calculated by deducting the INLET temperature


from the measured FLUE temperature. It displays ‘- OC -’ if the flue
probe is not connected or broken.

EFF : Combustion efficiency calculation displayed in percentage either as


Gross Ef(G) or Nett Ef(N) or Condensing Nett Ef(C) - Use MENU to
change. The calculation is determined by fuel type and uses the
calculation in British Standard BS845. The efficiency is displayed
during a combustion test, ‘- - - -’ is displayed while in fresh air.

LOSS : Losses calculated from oxygen and type of fuel. Displays reading
during a combustion test. ‘- - - -’ is displayed while in fresh air.

X - AIR : Excess air calculated from the calculated oxygen and type of fuel.
Displays reading during a combustion test. ‘- - - -’ is displayed while
in fresh air.

CO/CO2: CO/CO2 Ratio: measured CO (ppm) divided by (CO2 (%) x 10,000).

Page 53
PRS : Pressure reading, either single point or differential.

BAT : Displays the Battery power available.


Readings may be affected if used with low power batteries.

DATE : Date shown as day, month and year, DD/MM/YY. Date is recorded
when each combustion test is printed or stored.

TIME : The time shown is expressed in “Military” time HH:MM:SS. Time is


recorded when each test is printed or stored.

Note! When changing the batteries on the instrument the


memory will store the date and time for up to one minute, if
outside this time it may be necessary to re-enter the details.
Date and time may also need to be reset if re-chargeable
batteries are allowed to totally discharge.

FULL : The maximum number of tests have been stored in the memory. To
delete the stored memory, Select Reports then select the tests to be
deleted (see Page 23).

Pressure units:

m: millibar
s: psi
h: hPa
P: Pa
g: mmHg
i: inH2O
w: mmH20
k: kPa

Page 54
SYMBOLS used on the display
PRS Pressure

Ra CO/CO2

XAIR Excess Air

Tf Flue temperature

Ta Inlet temperature

∆T Nett temperature / Differential temperature

EfG Gross efficiency

EfCG Gross condensing Efficiency

EfN Nett efficiency

EfCN Nett condensing efficiency

-P-OFF Pump off

‘O2++% Calculated oxygen greater than 18% so calculation is disabled

N/F Temperature input not fitted

CAL Number of days left before recalibration is due

BAT Battery level symbol

N/F Not fitted.

INT Interval in seconds

ppm parts per million

p parts per million

ppm(n) parts per million normalised

O2ref reference level in % for normalisation calculation

mg/m3 milligrams per metre cubed

mg/m3(n) milligrams per metre cubed normalised

mg/kWh milligrams per kilowatt hour

mg/kWh(n) milligrams per kilowatt hour normalised


Page 55
ADDENDUM

Instructions for
KANE458 analysers
fitted with optional
Nitric Oxide (NO)
sensors

Page 56
DISPLAYING THE NO READING
Select “Menu” on the rotary switch and navigate using the function buttons:

 = Scroll up  = Scroll down = Enter

Select SCREEN and then select AUX

Choose a line to display the required readings as below

AUX display

The AUX (auxillary) display can be


PRS 0.00m → customised via MENU / SCREEN / AUX.
The parameters displayed on lines 1, 2, 3, 4,
Ra 0.0008 → 5 and 6 can be set by the user.
They remain the AUX parameters until
CO 52p → changed by the user.

CO2 6.3% →

NO 100p →

NOx 105p →

NOTE: To EXIT the MENU at any time simply move the rotary switch to any
position other than MENU. Any changes that have not been “entered” will
be ignored.

Use or to navigate to the main menu option SCREEN.

Press .

Use or to navigate to the sub menu option AUX.

Press .

Page 57
The display will show

AUX
LINE 1

Press and a third line will appear.

Use or to navigate to the desired parameter to be displayed on line 1.

Press to select the parameter for Line 1 and repeat the process to select the
display parameter for all four lines and then EXIT

Rotate the dial from MENU to AUX to display all your chosen settings.

PRINTING and STORING


The NO reading are printed and stored in the same way as the other combustion gas
readings. On the printouts the NO readings appear directly below the flue CO
readings.

Note the rotor needs to be in the AUX, O2/Eff or Ratio positions to print or store flue
combustion readings

Page 58
NITRIC OXIDE SENSOR SPECIFICATION

Gas Measurement Resolution Accuracy Range

Nitric Oxide (NO) 1ppm + 2ppm <30ppm*1 0 to 100 ppm


+5ppm <100ppm*1

Overrange to
+5% reading >100ppm 1500 ppm

*1 Using dry gases at STP

NOTE: When the dilution pump is operating to protect the CO sensor the NO
readings are also affected by an unspecified amount. The screen will typically show

P 0.00 →

R 0.0008 →

CO P-OFF →Pump OFF

CO2 P-OFF →Pump OFF

NO P-OFF →Pump OFF

NOx P-OFF →Pump OFF

Page 59
Page 60
PRODUCT REGISTRATION
Please register your Kane analyser at www.kane.co.uk
or complete, detach and return to: Kane International Ltd, Kane House,
11 Bessemer Road, Welwyn Garden City, Hertfordshire, AL7 1GF

Your Details
Name:

Job Title:

Company Name:

Company Address 1:

Address 2:

Town/City:

County:

Postcode:

Country:

Phone Number:

Fax Number:

Mobile Number:

Email Address:

Product Details
Note: Proof of Purchase may be required for warranty claims.

Date Purchased:
as numbers (28.01.14):

Purchased From:

Model Number: KANE458

Product Serial Number:


located on the rear product label
beneath the protective rubber sleeve

Page 61
Why did you buy a Kane Product? (Select all that apply)

 Kane Brand  Previous Owner


 Dealer Recommendation  Our Fixed Price Servicing Programme
 Made in the UK  Value for Money
 Not your Decision  Other:

What brand was your previous analyser?

 1st time purchase  Kane


 Testo  Anton
 Telegan  TPI
 Other:

Which magazines do you read?

Registered Gas Engineer 


Gas Installer 
P.H.P.I. 
P.H.A.M. News 
Heating Ventilating & Plumbing 
Heating & Plumbing Monthly 

Your feedback is important to us, please add any additional comments you
would like to make with regard to your recent Kane purchase:

Thank you for completing this survey.


All the information we have collected is confidential.
We do not sell or share data with any other company or organisation.

Page 62
Page 63
Thank you for buying this
analyser.
Before use, please register on
our website
www.kane.co.uk

Scan the QR code to go directly


to Register your Product on-line
or complete, detach and return
the Product Registration form in
this manual.
Page 64

You might also like