Quality Control of CT Image Using American College of Radiology (ACR) nd/4.0/)
Quality Control of CT Image Using American College of Radiology (ACR) nd/4.0/)
Quality Control of CT Image Using American College of Radiology (ACR) nd/4.0/)
net/publication/308679907
CITATIONS READS
2 1,354
5 authors, including:
SEE PROFILE
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Tarek A El-Diasty on 27 September 2016.
Original Article
a r t i c l e i n f o a b s t r a c t
Article history: Objective: To study image quality of CT scanner using the American College of Radiology
Received 19 June 2016 (ACR) phantom.
Accepted 21 August 2016 Material and methods: A multidetector CT scanner was used to measure parameters of
Available online xxxx
image quality using ACR phantom. The phantom included four modules to measure these
parameters. We obtained CT image for each module to measure these parameters for the
Keywords: brain protocol.
CT
Results: The acceptable levels of image quality were obtained for the positioning, CT num-
ACR phantom
Image quality
ber accuracy, slice thickness, low contrast resolution, uniformity and high contrast resolu-
tion tests that represent the parameters of image quality. In positioning test, the three BBs
were visible. In CT number accuracy, the CT number of three materials was in the range of
tolerance values unlike the bone value which was 846 HU. In low contrast resolution test
the smallest contrast groups were seen. In High contrast resolution test the 5 lp/cm was
visible. All these tests of image quality were accepted because they were within the toler-
ance values, so the quality of Philips CT scanner was improved. Ring artifact was indicated
also, which is a type of scanner performance errors.
Conclusions: Image quality tests are very important tests in acceptance of any CT scanner
after installation and maintenance to approve that the image parameters are acceptable by
using ACR phantom.
Ó 2016 The Egyptian Society of Radiology and Nuclear Medicine. Production and hosting by
Elsevier. This is an open access article under the CC BY-NC-ND license (http://creativecom-
mons.org/licenses/by-nc-nd/4.0/).
http://dx.doi.org/10.1016/j.ejrnm.2016.08.016
0378-603X/Ó 2016 The Egyptian Society of Radiology and Nuclear Medicine. Production and hosting by Elsevier.
This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
Please cite this article in press as: Mansour Z et al. Quality control of CT image using American College of Radiology (ACR) phantom. Egypt J
Radiol Nucl Med (2016), http://dx.doi.org/10.1016/j.ejrnm.2016.08.016
2 Z. Mansour et al. / The Egyptian Journal of Radiology and Nuclear Medicine xxx (2016) xxx–xxx
Fig. 1. (a) Photograph of correctly aligned ACR phantom and centering of For CT number calibration, the alignment light was cen-
the phantom in the axial (z-axis, cranial/caudal), coronal (y-axis, ante- tered again over module 1. We used the same scan param-
rior/posterior), and sagittal (x-axis, left/right) directions. (b) Four modules eters that are used in positioning test with varying
of ACR phantom indicating the scanner parameters. Module 1 measures
alignment, CT number accuracy and slice thickness, Module 2 measures
following parameters SFOV = 220 mm, Ww = 400 HU,
low contrast resolution, module 3 measures uniformity, noise and ssp. Wl = 0 HU, thickness = 5 mm and scan time = 1.5 m. Acicu-
and Module 4 measures high contrast resolution. lar region of interest (ROI) of approximately 200 mm2
Please cite this article in press as: Mansour Z et al. Quality control of CT image using American College of Radiology (ACR) phantom. Egypt J
Radiol Nucl Med (2016), http://dx.doi.org/10.1016/j.ejrnm.2016.08.016
Z. Mansour et al. / The Egyptian Journal of Radiology and Nuclear Medicine xxx (2016) xxx–xxx 3
Fig. 2. (a) Module 1 shows five cylindrical rods: Water, bone, polyethylene, air, and acrylic. (b) Module 2 shows different sized cylindrical rods. (c) Module 3
consists of tissue equivalent materials. (d) Module 4 shows eight different spatial frequency bar patterns.
Table 1
Scan parameters used in alignment of ACR phantom in axial brain protocol.
within each cylinder was placed, and on the data sheet, the An image at this location was acquired. We viewed the
mean CT numbers for each material were recorded [18]. image located at the center of module 2, and noted that
For slice thickness determination, we used sets of slice there were four cylinders for each of following diameters:
thickness to measure their thickness. We used slice thick- 2, 3, 4, 5, and 6 mm.
ness = 4 mm and filmed this image. To determine this We determined which all four cylinders were visual-
thickness, we counted the number of wires seen in the ized. The diameters of these cylinders were recorded on
top or the bottom and divided by two. This step was the data sheet. We were able to see the 6-mm rod.
repeated for the following slice thicknesses 1.3, 3.8,
2.5 mm. We recorded the measured slice thickness on data
sheet. 3.5. Noise determination
Please cite this article in press as: Mansour Z et al. Quality control of CT image using American College of Radiology (ACR) phantom. Egypt J
Radiol Nucl Med (2016), http://dx.doi.org/10.1016/j.ejrnm.2016.08.016
4 Z. Mansour et al. / The Egyptian Journal of Radiology and Nuclear Medicine xxx (2016) xxx–xxx
rod. This image was filmed with the two ROIs. All four rods values as indicated in Table 2 and Fig. 4. Then we com-
(of a given diameter) must be seen. We used the following pared the actual and the measured CT numbers of these
formula for measuring the value of contrast to noise ratio. materials in this protocol as shown in Fig. 5. All these
CNR = |A B|/SD [16,17]. materials were nearly the same CT number of the actual
values except bone.
3.6. Uniformity and plane distance accuracy test
4.1.3. Slice thickness determination
The table was moved to center the light over module 3, Table 3 indicates the measured thickness and indicates
and we used ROI = 400 mm, Ww = 100 HU, Wl = 0, thick- that the tests of slice thickness determination were
ness = 5 mm, and collimation = 2 0.625 mm for this test. accepted for all the selected slice thicknesses. In Fig. 6 slice
The image at this location was filmed, and we placed an thickness equaled 4.
ROI of approximately 400 mm2 at the center of the image
(A) and the four edge positions. We recorded the mean
4.2. Module 2: Low contrast resolution measurements of ACR
CT numbers for all five ROIs for our records, then recorded
axial brain
the standard deviation of the center ROI. The CT numbers
for all five ROIs must be within ±5 HU of the center ROI
For module 2, we enabled to see group 25 in this test
mean value. Finally we calculated and recorded the unifor-
which indicated the smallest contrast group, and this
mity value (center mean CT number – edge mean) on the
data sheet [17].
4. Results
4.1. Module 1
Please cite this article in press as: Mansour Z et al. Quality control of CT image using American College of Radiology (ACR) phantom. Egypt J
Radiol Nucl Med (2016), http://dx.doi.org/10.1016/j.ejrnm.2016.08.016
Z. Mansour et al. / The Egyptian Journal of Radiology and Nuclear Medicine xxx (2016) xxx–xxx 5
Table 2
The measured CT numbers of different materials in ACR axial brain.
1000 4.5. High contrast resolution (HCR) for ACR axial brain
-1500
5. Discussion
Fig. 5. Diagram illustrated the measured and actual CT numbers of the
different materials in ACR axial scan. All these materials are nearly the The provided figures indicated acceptable levels of
same CT number except bone. image quality, as well as a type of image artifacts (ring arti-
fact) and scanner performance errors, so biomedical engi-
neer was required to calibrate the detectors. For module
means that the test was accepted as shown in Fig. 7. It can
1, the test of alignments was accepted because the three
be concluded from Fig. 7 finding of ring artifacts, and it was
BBs that were visible were within the tolerance value.
due to the fact that one of the detectors was out of calibra-
The test of slice thickness determination was accepted
tion on this scanner; the detector gave a consistently erro-
also, because for slice thickness 1.3 mm, the number of
neous reading at each angular position, resulting in a
wires seen in the top or the bottom of the image and
circular artifact, so medical engineer was required [19].
divided by two, equaled (2). For 2.5 mm, the number of
Table 3
All the choosing thicknesses passed the test.
Thickness set No. of wires seen in top or bottom Thickness measured (mm) Tolerance value (mm) Results
1.3 2 1.0 ±1.5 Pass
2.5 4 2 Pass
3.8 7 3.5 Pass
4 8 4 Pass
Please cite this article in press as: Mansour Z et al. Quality control of CT image using American College of Radiology (ACR) phantom. Egypt J
Radiol Nucl Med (2016), http://dx.doi.org/10.1016/j.ejrnm.2016.08.016
6 Z. Mansour et al. / The Egyptian Journal of Radiology and Nuclear Medicine xxx (2016) xxx–xxx
Fig. 8. Module 3 in ACR axial brain protocol showing that the CT numbers
Fig. 6. Slice thickness determination in ACR axial brain equaled the
in left, right, top, bottom and center are not exceeded the tolerance value
number of wires seen in the top or the bottom divided by two.
as the same time they are nearly the same values and BBs dis-
tance = 99.9 mm, but there are ring artifacts.
Fig. 7. Module 2 indicating low contrast group in ACR axial. There are Fig. 9. Module 4 in ACR axial brain protocol indicating eight different
four cylinders with 6 mm diameter. There is ring artifact. spatial frequency bar patterns. The visibility of 4 lp/cm and 5 lp/cm was
easy while 6,7,8,9,10,12 lp/cm were harder.
wires that were seen was 4. For 3.8 and 4 mm the numbers
of wires seen were 7 and 8. All the previous slice thick- 6 mm. The 6 mm cylinder was also seen, and the four cylin-
nesses that were selected were within tolerance value ders of it were visible, so the test was passed. This result is
(±1.5 mm). Our results are in agreement with those of Hob- in agreement with that of Nookala et al. [20].
son et al. The test of CT number accuracy was accepted For module 3 the uniformity test was also accepted,
also, because the measured CT number of polyethylene, because the measured values of the mean CT numbers of
acrylic, water and Air was within the range of the tolerance all four edges were (0) HU that were ±5 HU of the mean
values and they are in agreement with those of McCul- value of the center where the ±5 HU is the tolerance value.
lough et al. The bone value was not accepted, because it In module 4, the high contrast resolution test was
was out of the range of the tolerance values. It is in dis- accepted, since the tolerance values of the brain protocol
agreement with McCullough et al. for this test were 5 lp/cm and the 5 lp/cm which were vis-
We obtained this result of the bone value, because the ible. The result of uniformity and high contrast resolution
number of applied protocols was limited [9,10,19]. test is in agreement with that of Nookala et al. also [20].
For module 2, we know that this module has different
sized sets of cylindrical rods. We must see the smallest 6. Conclusion
low contrast group. The low contrast resolution test was
accepted, because we had seen the large cylinders To the best of our knowledge, this is the first study
(25 mm) that were necessary to be seen. The tolerance is carried out in Egypt.
Please cite this article in press as: Mansour Z et al. Quality control of CT image using American College of Radiology (ACR) phantom. Egypt J
Radiol Nucl Med (2016), http://dx.doi.org/10.1016/j.ejrnm.2016.08.016
Z. Mansour et al. / The Egyptian Journal of Radiology and Nuclear Medicine xxx (2016) xxx–xxx 7
All the previous studies were in Europe only, so we rec- [8] ACR. CT Accreditation phantom program. Reston (VA): American
College of Radiology; 2015.
ommend that image quality test for CT must be performed
[9] McCollough CH, McNitt-Gray MF, Ruckdeschel T. The phantom
in other Egyptian and African centers, because image qual- portion of the American College of Radiology (ACR) Computed
ity tests are very important tests in acceptance of any CT Tomography (CT) accreditation program. Practical tips, artifact,
scanner after installation and during maintenance to example, and pitfalls to avoid. Med Phys 2004;31(9):2423–42.
[10] Hobson MA, Soisson AT, Davis SD. Using the ACR CT accreditation
approve that the image parameters are accepted by using phantom for routine image quality assurance on both CT and CBCT
ACR phantom. imaging systems in a radiotherapy environment. J Appl Med Phys
2014;15(4):226–39.
[11] Khoury HJ, Andrade ME, Kramer R, et al. Evaluation of CT dose and
Conflict of interest image quality in Recife, Brazil; 2008. p. 1–10.
[12] Gammex rim: Quality Control for Diagnostic Radiology, 1-800
GAMMEX 1; 2003. Available at <www.gammex.com>.
There is no conflict of interest.
[13] ACR and AAPM: ACR–AAPM technical standard for diagnostic
medical physics performance monitoring of computed tomography
References (CT) equipment; 2012:4.
[14] Winslow J F. Construction and application of anthropomorphic
[1] International Marketing Ventures, Rockville, MD; 2008. September phantoms for use in CT dose studies 2009;30–38.
22, 2010). Available at <http://www.imvinfo.com>. [15] Greene-Donnelly 1 KA, Ogden KM. Evaluation of commercial
[2] Brenner DJ, Hall EJ. Computed tomography – an increasing source of extension plates for the ACR-CT accreditation phantom. J Appl Clin
radiation exposure. N Engl J Med 2007;357:2277–84. Med Phys 2016;17(1):416–20.
[3] Hall EJ, Brenner DJ. Cancer risks from diagnostic radiology. Br J Radiol [16] ACRIN. CQIE MOP Part B, Version 3.2; March 2013. p. 1–7.
2008;81:362–78. [17] ACR. CT accreditation phantom instructions Available from:
[4] Barrington de González A, Mahesh M, Kim KP, et al. Projected Cancer Available from: <http://www.acr.org/~/media/ACR/Documents/
risks from computed tomographic scans performed in the United Accreditation/CT/PhantomTestingInstruction.pdf>2013.
States in 2007. Arch Int Med 2009;169:2071–7. [18] Bongartz G, Golding SJ, Jurik AG, et al. European guidelines of quality
[5] European guidelines on quality for Computed Tomography. Report criteria for computed tomography. European Commission; 2004.
EUR 16262 EN. [19] Barrett JF, Keat N. Artifacts in CT. Recognition and avoidance
[6] Zeman RK. Why seek accreditation of your CT programAvailable 2004;24(6):1679–91.
from: <www.Image wisely.org>2010. [20] Nookala PK, et al. Modification of CT quality assurance phantom for
[7] Schuyler D. Quality control report, software for scoring ACR CT PET/CT alignment and PET resolution; 2005.
phantoms aids in quality controlAvailable from: <www.
radimage.com>2010.
Please cite this article in press as: Mansour Z et al. Quality control of CT image using American College of Radiology (ACR) phantom. Egypt J
Radiol Nucl Med (2016), http://dx.doi.org/10.1016/j.ejrnm.2016.08.016