FCR PRIMA Console
c. Service Operation Guide
1. Outline
This document describes basic knowledge necessary for operation works such as outline of service
operation of FCR PRIMA Console, how to see the service manual, how to use service tools, and other
notices. By reading this document, we hope that you can perform necessary works without hesitation
by referring the service manual.
NOTE In this document, if not otherwise specified, "CR-IR 391CL" and "CR-IR 391V CL" are
called "PRIMA CL" as a whole.
2. Contents
1. Outline 1
2. Contents 1
3. Service Operation overview 3
3.1 Service Manual overview 3
3.2 Installation 3
3.3 Additional installation 9
3.4 Reinstalling the OS 9
3.5 Maintenance 9
3.6 Troubleshooting 9
4. How to use the installer 10
4.1 Server Application Software 11
4.2 Option keys and Application Key 11
5. Restrictions 12
5.1 DICOM Storage 12
5.2 Others 12
6. Supplementation of utility functions 14
6.1 Service Utility 14
6.1.1 How to start service utility 14
6.1.2 List of the service utility function 14
6.1.3 Simplified setup wizard 15
6.1.4 Service settings 16
6.1.5 Maintenance tool 17
6.2 User Utility 18
6.2.1 List of the user utility function 18
6.2.2 How to start User utility 20
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
c. Service Operation Guide
2009.07.17
6.3 Routine Screen 21
6.3.1 How to login for servicing 21
6.4 Backup Utility 22
6.4.1 List of the backup utility function 22
6.4.2 How to start Backup utility 23
6.5 Disk copy tool 24
6.5.1 List of the disk copy tool function 24
6.5.2 How to start Disk copy tool 24
6.6 Client Utility 25
6.6.1 List of the client utility function 25
6.6.2 How to start Client utility 26
6.7 Indicator 27
6.7.1 List of the indicator function 27
6.7.2 How to start Indicator 28
6.8 Disk Format Tool 29
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
c. Service Operation Guide
2009.07.17
3. Service Operation overview
3.1 Service Manual overview
The service manual aims to improve usability by making chapter style the same as that of Reader Unit.
Followings are the descriptions of each chapter.
Chapter Title
Safety Safety Precautions
Describe precautions related to safety.
Spec Product Specifications
Describe abbreviation of product name, size/weight, environment and discard
conditions.
MD Machine Description
Describe system configuration, examples of connection, function guideline, as well
as relation between FCR and RU.
MT Machine Troubleshooting
Describe failure analysis and its troubleshooting. Detailed information about
hardware failure is incorporated.
MC Checks, Replacement and Adjustment of Parts
Describe procedures how to exchange parts. Also describe procedures for
reinstalling the OS and the application. Procedure of the application upgrade is
also included.
MU Maintenance Utility
Describe necessary information to customize settings in response to users'
request. Explain options and meanings of each item.
SP Service Parts
Describe the service parts list and the locations where they are used in figure and
chart format.
PM Preventive Maintenance
Describe items to be inspected on periodic check. Hardware cleaning, which is
conducted in Japan, is described as an example with photographs.
IN Installation
PC Performance Check
3.2 Installation
Install by IN part. Installation of equipment below is available by this chapter.
z PRIMA CL main unit
z Reader Unit
z DRYPIX (Option)
z DICOM CR Storage (Option)
Configurations for connection are possible for Reader Unit and DRYPIX.
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
c. Service Operation Guide
2009.07.17
For installation of Reader Unit and DRYPIX, see the latest service manual and ECN of each device.
Basically, installation completes by placing the hardware, installing option key software and the
application key software, and performing the simplified setup wizard.
For checking operation and recording installation for each site, use the checklist of PC part.
For further customization as described below, see an appropriate page of the service manual.
Customization Described part
Changing language IN "4.2 Changing the Displayed Language and Replacing the Menu
Database"
Changing menu/parameter IN "4.2 Changing the Displayed Language and Replacing the Menu
file Database"
Setup automatic output MU Appendix 4. Settings for Automatic Output
Setup automatic verification MU Appendix 5. Setup for Auto-verify
TagReplace, TagLookup MU Appendix 6 DICOM Tag Mapping Function
Changing host name Changing the Computer Name of PRIMA CL is not supported at
present. If needed, perform MC 1. Reinstalling the OS and the
Application.
Changing IP address MC Appendix 5. Procedures for Changing the CR-IR 391CL, RU, and
Printer IP Addresses
Voltage setting of PC IN 1.4 Power Requirements
Note:The PCs are shipped from Japan with the voltage setting of 100V.
z Items specific to each site (setup items available only in SYSTEM CONFIG)
No. Name Option
(up to 64 characters)
ultrasound device.
1 : Metric
output * 1 : Type1
2 : Type2
(year/month/day) + “_”
number” +
Date of birth
1 : R-reverse
mark.
* 2 : AP or PA
12,13,14,15,16,
17,18,19,20,21,
22,23,24,25,26,
27,28,29,30,31
12-33 Setting for validity/invalidity of Yes : Reduces the image to 18×24cm/8×10 inch
was made.
storage No
* 2 : Final writing
* No: No compression
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
c. Service Operation Guide
2009.07.17
Image transmission to DICOM Storage IN Appendix 1 Settings for Saving Images into DICOM Storage Equipment
device
Changing display LUT IN Appendix 1 Settings for Saving Images into DICOM Storage Equipment
When restoring images or image database from the storage disks, be sure to perform reconstruction
beginning from the oldest dated disk.
See "1.14 Reconstructing the Image Database" of MC part for details.
3.5 Maintenance
What can be done by MU part:
MU part supplies detailed information for each setting value. While IN part introduces necessary works
step-by-step, MU part focuses on customization of each function.
See the document of workflow customization and the operation manual for customization by user
functions (such as User Utility and Client Utility).
3.6 Troubleshooting
MT part describes:
For hardware trouble - Troubleshooting for each PC type.
For software trouble - The list of error messages.
The flowchart and its links to replacing procedures (MC part) are described for each PC part. Follow
the replacing procedure of each types of PC. Be careful especially when replacing the motherboard
because reinstallation of option key software is necessary after the replacement. If it is not done, option
functions cannot be used.
See "2.7 motherboard" of MC part for details.
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
c. Service Operation Guide
2009.07.17
4. How to use the installer
The installation launcher can proceed the installation work interactively. It is easy to check the works to
be executed visually because each operation button has different colors.
(1) Green: Installation is possible.
(2) Red: Already installed.
(3) Gray: Installation is impossible (not satisfying the conditions).
(1)
(2)
(3)
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
c. Service Operation Guide
2009.07.17
4.1 Server Application Software
Application Software is preinstalled on the Server PC.
Top
Tool
Reinstall configuration files.
* Configuration will be initialized.
Upgrade parameters.
* Menu parameters will be carried on.
Supplementation:
PRIMA CL can register storage devices up to two, but the same LUT is applied for each device.
Transmission is possible to only one unit at a time.
As the folder, where the LUT file is located, is defined as a hidden folder, see "Appx IN 1: 2.4
Changing the Display Gradation Correction LUT File" of FCR PRIMA Console Service Manual to
edit this file. In case of creating LUT file, see the operation manual of "MONITOR QC*1".
*1
FCR Diagnostic Image Display Gradation Correction Procedure "MONITOR QC"
5.2 Others
(1) Do not connect to the Internet
PRIMA CL cannot be connected to the Internet. PRIMA CL works with Internet Explorer (I.E.) so I.E.
and Windows OS should not be changed their version.
(1)
(2) (3)
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
d. Service Operation Guide
2009.07.17
Item Function description
(1) Simplified settings Simplified setup wizard Execute necessary settings for PRIMA CL
etc.)
it.
etc.) is changeable.
(3) Maintenance tool Database Utility Initialize the current database in case of its
corruption or inconsistency.
Since all data in the database will be lost after
Acquiring EDR backup data Acquire and store the EDR backup data which is
Verification.
<NOTE>
Date and time setup is also done by this wizard. Specify the same format for "Date/time format" setting
of System Setting and date and time setting of the OS. When the display setting and the format of the
system time acquired from the OS is different, such information as study date is not displayed correctly.
For details, see "MU: Maintenance Utility 4. Regional and Language Options Settings" of "FCR PRIMA
Console Service Manual".
y System
Followings are available from this configuration.
Item Description
1 Localize setting Set the date format (commonly applied to all date expressions
including birth date input, display, and film output) and the time
display format.
5 Setting for image collection Settings related to receiving images from RU.
6 Setting for image browsing Settings related to image display (display unit of
measurement result, etc.) on the viewer screen.
7 Setting for general-purpose file output Setting related to general-purpose file output
8 Linkage with the receipt computer Following functions are used only in Japan.
9 Setting for image link Following functions are used only in Japan.
11 Image incidental information Settings related to the image process change flag.
12 Print format Settings related to film output such as film margin or
trimming volume.
13 Film character setting Settings for character information to be output on film such
as image processing conditions or menu name.
14 Setting for output function Whether or not re-try has been executed after an abnormal
output was made. Status acquisition timing of DICOM
printer.
15 Operation (low frequency) setting The maximum registration number of patient/study
information.
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
d. Service Operation Guide
2009.07.17
Item Description
19 Setting for analysis Settings related to log acquisition for analysis. 19-1, 19-6,
19-7 are used for whether or not DICOM log is acquired,
and whether or not detailed log is acquired.
"19-8 Collect image data for maintenance logs" (available
when deleting personal information) is linked to the option
when maintenance logs are acquired from the backup
utility. The larger this value becomes, the more it takes for
acquiring logs on the personal information deletion
process.
20 Settings for endoscopic image import Following functions are used only in Japan.
21 Electrocardiographic data import Not used.
settings
y Modality Settings
No need to change this setting. This setting is to display images with applying adequate Modality LUT
but PRIMA CL does not support importing images of other modalities.
(3) (10)
(4) (11)
(5) (12)
(6) (13)
(7)
Function Description
(1) Menu Settings Create, edit or delete an Exposure Menu and Study Menu.
(2) Output Destination Settings Set the parameters to store images to a storage disk, print images on
(3) Image Deletion Settings You can change the time that images read from a disk for reference
elapsed.
(4) Client Application Startup You can set the Client Application so that it automatically starts after
(5) Image-incidental Info Display Additional image information such as menu names and
You can select the incidental image information you want to have
displayed.
(6) Media Consistency Check Check the consistency of the disk, for example, whether the
When the storage disk is full, first carry out an consistency check and
(8) Character String Info Annotated Film character string refers to the character strings annotated on a fi
on Film lm. The film character string is set in each Exposure Menu.
Verified.
(10) Backup Notification Settings If you have not made a backup for a certain period of time, you can set
This message will appear when you start up the PRIMA CL.
(11) Prefix Display Settings Set to add a prefix after the patient names displayed on the PRIMA
CL.
(12) Patient Database import Use an import file to import patient information to the PRIMA CL
database.
import file.
(13) History-info Consistency Check Check that the history of processes such as PRIMA CL login/logoff,
<Note>
After the user utility is terminated, the routine screen opens. Termination with holding down the Shift
key displays the Windows desktop screen. When the user utility terminates, automatic transition to
the routine screen (restating the application) is performed.
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
d. Service Operation Guide
2009.07.17
6.3 Routine Screen
6.3.1 How to login for servicing
Click 4 corners below slowly and the special login screen only for Service Engineer will be displayed.
Enter password “fieldengineer” and you can login.
Click
Click Click
Click
(1)
(2)
(3)
Item Description
(1) Backup Acquire and storage patient information, setting information and Log
information.
(2) Restore In the event of unit failure, use the backup to restore the patient
information and system setting information.
In this case, the patient information and system setting information
can be restored to the point of backup, but changes made after
backup are not reflected.
(3) Collecting logs for Acquire logs for maintenance purpose or configuration information.
The function to delete personal information from log files is available. After pressing (3), the dialog to
select type of log is displayed. On this dialog, it is available to select [Deleting personal information off
the logs]. After performing log acquisition with selecting deleting personal info. the “Study.txt” will
created on the desktop. You can check the deleted personal info. from the log files by “Study.txt”.
22/29
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
d. Service Operation Guide
2009.07.17
When you collect log files of DICOM communication error, change the DICOM log mode to the delail
mode and reproduce DICOM communication error before you collect logs. To change DICOM log
mode, edit System Settings 19-1, 19-6, 19-7, 19-9 by using Service Utility.
23/29
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
d. Service Operation Guide
2009.07.17
6.5 Disk copy tool
6.5.1 List of the disk copy tool function
Item Description
Disk Copy Duplicate image storage disks to provide against possible failures to
24/29
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
d. Service Operation Guide
2009.07.17
6.6 Client Utility
6.6.1 List of the client utility function
(1) (5)
(2) (6
(6)
(3) (7)
(4)
Item Description
(1) Patient Info Maintenance Register new Patient information in the PRIMA CL database.
(3) User Settings(for Administrator) Register, edit or delete information about the users who use PRIMA CL.
You can also assign operational authorities to user groups and set
authority.
(4) User Settings(for General Users) The user who is currently logged in to PRIMA CL can edit its own
(5) Explanatory Info Maintenance Registers, edits or deletes explanatory information that will be printed on
paper.
25/29
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
d. Service Operation Guide
2009.07.17
Item Description
(6) History Info Confirm the following four kinds of history information:
and modification
> History of sessions of backup and restoration carried out using the
backup utility
(7) Disk Operation You can perform the following operations on disks:
> Restore a storage disk or work disk (at a disk restoration prompt)
> Restore the database with information from a storage disk or work disk
26/29
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
d. Service Operation Guide
2009.07.17
6.7 Indicator
6.7.1 List of the indicator function
Indicators provide visual indications of the status of Reader Unit, disk, printer (option), data transfer
and events (errors, warnings and information).
Item Description
(1) Reader Confirm the status of Reader Unit connected to PRIMA CL. You can also
perform scanner cleaning.
(2) Disk Confirm disk status. You can also perform the following operations on disks:
• Replace disks
• Configure a storage disk as a write-enabled or read-only disk
• Restore a storage disk or work disk (at a disk restoration prompt)
• Restore the database with information from a storage disk or work disk
(3) Printer Confirm the status of the film printer connected to PRIMA CL system.
(4) Output Confirm the output status of images as they are printed on a film (option), saved
to a disk, saved in a in a general-purpose file format (some options) or
transmitted to DICOM server (option). Images whose output is canceled or ailed
due to an error output can be output again.
(5) Event Confirm events (errors, warnings or information) occurring on PRIMA CL
system.
If the event indicator is blinking, it indicates that there are unconfirmed events.
Confirm the events and respond to them appropriately.
27/29
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
d. Service Operation Guide
2009.07.17
28/29
CR-IR 391CL V1.0 / CR-IR 391V CL V1.0
d. Service Operation Guide
2009.07.17
6.8 Disk Format Tool
To start handling DVD media, it should be formatted (UDF 1.5). In the routine workflow, user exchange
DVD media by using “Client Utility - Disk operation” or “Indicator - Disk tab”. At that time, if the DVD
media has not been formatted, formatting process will automatically start.
To start formatting process “FFUdfFmtTool” manually, shut down PRIMA CL server application and
open “My Computer” using Explorer.exe and right click DVD drive and select “Format”.
For precisely, refer Operation manual.
29/29