SOP LNFillStation
SOP LNFillStation
SOP LNFillStation
________________________________________________________
Read the EH&S Standard Operating Procedures Fact Sheet before filling out this form.
Print out the completed form and keep a readily accessible hard copy in the lab (also
keeping an electronic copy is highly recommended).
______________________________________________________
Section 1 – Process
This SOP outlines a safe procedure for the filling of liquid nitrogen dewars from the bulk liquid
nitrogen storage tanks behind Cory Hall via the fill station and is required to be completed before
access to the fill station is granted.
Burns - Direct contact of skin and cryogenic liquids can cause cold burns and frostbite. Prolonged
contact may result in blood clots.
Adhesion - The cold surface of equipment and piping containing cryogenic liquid can cause the
skin to stick to the surface, which will then tear as you attempt to remove it. Even non-metallic
materials are dangerous to touch at such low temperatures.
Boiling and Splashing - Cryogenic liquids can boil and splash when first added to a warm
container.
Oxygen Deficiency and Asphyxiation - Cryogenic liquids have the potential to create an oxygen
deficient environment because of their large liquid-to-gas volume displacement ratios, typically
about 700:1.
Pressure and Explosions - Large liquid-to-gas ratios can lead to rapid pressure changes as
cryogenic liquids vaporize. All cryogens can condense sufficient moisture from the air
subsequently freezing and blocking the opening of storage vessels. This can lead to an explosion
from the buildup of trapped gases in the container; for instance, cryotubes stored in liquid nitrogen
may explode when removed from the dewar.
Flammability and Explosions - Nitrogen and helium are considered non-reactive and
non-flammable; however, liquid nitrogen and liquid helium can condense oxygen out of air.
Liquid oxygen is VERY reactive and hazardous. Combustible substances exposed to liquid
oxygen become more likely to ignite, will burn more vigorously, and may potentially explode.
Materials usually considered non-flammable can burn vigorously in an oxygen enriched
environment. Organic materials that can react violently with liquid oxygen include oil, grease,
kerosene, tar, cloth, and asphalt.
Use of the liquid nitrogen fill station requires authorization by the NanoLab. Authorization
requires the completion of a short online training (approximately 20 minutes) on the Safe Use of
Cryogenic Materials. Go to the UC Learning Center at https://blu.berkeley.edu and type
“cryogens” in the search field. Non-employees/non-students or anyone without a Calnet ID can
access the training via:
http://ehs.berkeley.edu//sites/default/files/lines-of-services/training/AccessProblemSolver/story.html
Bring your certificate of completion when you apply for access to the fill station and sign off on this SOP.
The fill station is behind Cory Hall, attached to the cage surrounding the bulk liquid nitrogen
storage tanks.
Section 6 – Special Handling Procedures and Storage Requirements
Preparation
1. Be familiar with hazards associated with cryogen use.
2. Examine containers and pressure relief valves for signs of defect. Never use a container that
has defects. Ask cryogenic vendor for assistance with questions on cryogenic equipment and
pressure relief valves.
3. Ensure that all equipment and containers are free of oil, grease, dirt, or other materials which
may lead to flammability hazard upon contact with liquid oxygen.
4. Select working materials carefully. Cold cryogenic liquids may alter the physical
characteristics of many materials, make them brittle and fail.
5. Verify there is pressure relief for any place that there can be a pressure build-up. If unsure,
have your dewar reviewed by NanoLab staff.
6. Schedule dewar fills during normal business hours, when staff are available to assist. After
hours access is not available. Checked out keys must be returned to the NanoLab office before the
end of the business day.
7. Sometimes the dewar pressure relief valves fail and allow improper or frequent venting. Safely
move the damaged dewar to a well-ventilated area, post a warning sign, and contact the dewar
manufacturer or Campus cryogen vendor for assistance.
Storage
1. Store cryogens in well-ventilated areas to prevent oxygen deficiency.
2. Avoid contact of moisture with storage containers to prevent ice plugs in relief devices.
3. Periodically check container necks for ice plugs; core out ice plugs if present.
4. Keep all heat sources away from cryogenic liquids.
Section 7 – Personal Protective Equipment
Hand Protection
• Wear loose fitting gloves made for cryogenic work (blue cryogenic gloves) or smooth leather
welding type gloves without gauntlets. Loose fitting gloves can be thrown off if some cryogen
leaks or is spilled into them.
• Rubber gloves should not be used because they will harden instantly - if your hand is bent, you
may not be able to remove your hand.
• A thin gas barrier forms between the skin and the cryogenic liquid when it is spilled on the skin.
This will protect you unless the liquid hits you under force. This gas barrier is very cold and can
also burn you.
• Use non-metallic tongs to add or remove materials from cryogenic liquids.
Eye Protection
• Face shields and goggles provide the best protection for the eyes and face. Safety glasses will
not protect your face, and cold liquids can hit your face and run under the glasses into your eyes.
Safety goggles will keep liquid out of eyes but leave face exposed.
• When filling dewars or transferring cryogenic liquids from one container to another, face
shields must be worn.
Clothing
• Closed toe shoes are required when handling cryogenic liquids. Leather will shed the spilled
liquid. Cuff-less pants should cover the shoe top. Sneakers are typically made with absorbent
materials which could draw liquid toward your skin.
• Long sleeve shirts made of non-absorbent material are best.
• An apron made of leather or other non-absorbent material should be used when working with
liquid cryogens. Most clothing material will absorb spilled liquid cryogens, bringing the liquid
close to the skin.
The fill station is outdoors so requires no special attention to ventilation. There is an oxygen
deficiency monitor mounted inside the fill station cabinet. If this alarm should ever go off during
a fill, evacuate the area and contact NanoLab staff immediately. Area is safe to return to once
alarm turns off.
Section 9 – Spill and Accident Procedures
If a spill of liquid nitrogen occurs, make sure you do no come in contact with the liquid. It will
evaporate on its own and requires no clean up. If you are in an interior space with little or no
ventilation, evacuate the area since large amounts of evaporating liquid nitrogen will quickly
deplete the area of oxygen.
• If skin comes into contact with a cryogen, run the area under cool or warm water for fifteen
minutes. Never use hot or cold water. The re-warming, or thawing, of affected area(s) should be
done gradually. It may take up to 60 minutes to thaw the affected area(s) and bring back the
natural color of the skin.
• If your finger is burned, do not put it in your mouth. This could burn your mouth or tongue.
• Do not rub a burned area: rubbing can cause further tissue damage.
• Always seek medical attention for frostbite injuries. You should obtain medical assistance as
soon as possible when cryogens contact your skin. Immediately upon exposure, the frozen skin
appears waxy and yellow and the burn usually is not painful. Then it painfully swells and blisters
while the skin defrosts.
• Always push dewars if they need to be moved. Never pull on dewars - they are very heavy and
can tip and crush you. Large dewars can lead to injuries (back injuries, crushed foot, crushed
hand).
In case of a pressure explosion, evacuate the area and contact NanoLab staff immediately.
N/A
Section 11 – Decontamination
N/A
Section 12 – Process Steps