V3 Wiring Diagrams Rev06 PDF
V3 Wiring Diagrams Rev06 PDF
V3 Wiring Diagrams Rev06 PDF
A
B
E
C
D
H
G
A2
8
8
- - + +
24VDC (240W)
0 BL
1 BR
13 BK
44 RD
7
7
EARTH 1 YW/GN
LIVE 2 1 BR
NEUTRAL 3 0 BL
4
5
L-PUMP 6 1 BR
N-PUMP 7 0 BL
6
6
44 RD ++24VDC 7 44 RD
Potensiostat Board
AG
44 RD ++24VDC 8 44 RD
-VS
+VS
REF
CNT
GND
13 BK C-24VDC 9 13 BK
A1
A0
SP
CR
13 BK C-24VDC 10 13 BK
24V
+VS
GND
GND
GND
GND
GND
5
5
51 PR
MEAS
MEAS
50 GR
49 WH
EARTH 47 GN
VALVE-1 48 18 YW
49 WH ++12VDC 11 49 WH
VALVE-2 49 19 YW
49 WH ++12VDC 12 49 WH
VALVE-3 50 20 YW
51 PR C-12VDC 13 51 PR
VALVE-4 51 21 YW
51 PR C-12VDC 14 51 PR
VALVE-5 52 22 YW
VALVE-6 53 23 YW
VALVE-7 54 24 YW
C-24VDC 55 13 BK
+CELL-LEVEL 56 38 PK
4
5
0
1
2
REV
- CELL-LEVEL 57
DOS-PUMP-LEVEL 58 60 WH
DOS-PUMP-ERROR 59 61 WH
+DOS-PUMP 60 4 YW
4
4
+DOS-PUMP AUX 61 73 YW
0 BL
1 BR
51 PR
50 GR
49 WH
- DOS-PUMP 62 5 BK
L
E
N
+24VDC 63 44 RD + -
COM
LEAK 64 59 WH
100-240VAC -/+ 12VDC
GND
TEMP 1 65 12 PK
-/+12VDC (36W)
+VS 49 WH
OUT
DESCRIPTION
C-24VDC 66 13 BK
+pH 67 71 RD -VS 50 GR
AG Titration module
REVISIONS
- pH 68 41 PK
DATE
+ 12 VDC 74 49 WH
2008/07/20
2008/07/24
2013/02/28
2015/05/28
2009/02/09
2011/06/23
C-12VDC 75 51 GR
3
3
BY
AM
HSS
HSS
MRR
MRR
MRR
MINTEK.
THE WRITTEN PERMISSION OF
DESIGN OR INSTALLATION WITOUTH
NOT BE USED IN THE MANUFACTURE,
IS THE PROPERTY OF MINTEK. IT MAY
THIS DRAWING IS CONFIDENTIAL AND
2
2
SCALE
SIZE
1:2
FSCM NO
NOTES:
LEGEND:
MINTEK
SHEET
DWG NO
STANDARD ETHERNET CABLE
1
1
MAC_CYNV3_WD_R1
WAD CYANIDE ANALYSERS (CLIENT: HERRADURA)
1 OF 8
5
REV
F
A
B
E
C
D
H
G
8 7 6 5 4 3 2 1
A2 NOTES:
H H
G G
69 PK
63 PK
64 PK
55 PK
13 BK
38 PK
59 PK
14 PK
61 PK
60 PK
+24
+24
CPU 1100 (1763-I16BWA)
IV1(+)
COM
IV(+)
COM
COM
+24
+24
I/0
I/1
I/2
I/3
I/4
I/5
I/6
I/7
I/8
I/9
SLOT 0
+24VDC
-24VDC
F F
VACDC
VACDC
VACDC
VACDC
VACDC
VACDC
O/0
O/1
O/2
O/3
O/4
O/5
E
E E
D D
73 YW
25 YW
68 YW
24 YW
44 RD
44 RD
43 BR
14 PK
13 BK
13 BK
63 PK
64 PK
55 PK
56 BK
60 PK
61 PK
59 PK
12 PK
13 BK
38 PK
69 PK
2 YW
6 YW
8 YW
4 YW
3 BK
7 BK
9 BK
5 BK
GN
GN
BL
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
29
30
62
63
64
65
57
54
55
56
58
59
60
61
28
PUMP_FLOAT SWITCH
DOS-PUMP-ERROR
- ALARM-WARNING
+ALARM-WARNING
- ALARM-CRITICAL
+ALARM-CRITICAL
DOS-PUMP-LEVEL
BOOSTER-PUMP
+DOS-PUMP LEX
-DOS-PUMP AUX
+REN STREAM 1
+REN STREAM 2
N-ALARM-LIGHT
L-ALARM-LIGHT
FLUSH_VALVE
- CELL-LEVEL
+CELL-LEVEL
C C
- MAIN-PUMP
+MAIN-PUMP
-DOS-PUMP
C-STREAM
C-24VDC
C-24VDC
C-24VDC
VALVE-7
24VDC+
+24VDC
EARTH
EARTH
TEMP1
LEAK
AI1
LEGEND:
C
B B
TO PLANT PLC
A
2
1
ISSUED FOR APPROVAL & COMMENTS
2008/07/24
AM
HSS
THE WRITTEN PERMISSION OF
MINTEK. MINTEK
PRIVATE BAG X3015 RANDBURG 2125 A
GAUTENG SOUTH AFRICA +27 11 709 4111
0 ISSUED FOR APPROVAL & COMMENTS 2008/07/20 HSS
SIZE FSCM NO DWG NO REV
8 7 6 5 4 3 2 1
F
A
B
E
C
D
H
G
A2
8
8
L
E
N
- - + +
-24VDC +24 +24
12VDC (180W)
E COM 69 PK
0 BL
1 BR
51 GR
51 GR
49 WH
49 WH
I/0 63 PK
2 YW VACDC I/1 64 PK
3 BK O/0 I/2
4 YW VACDC I/3
5 BK O/1 COM 13 BK
6 YW VACDC I/4 14 PK
8 YW VACDC 59 PK
7
7
EARTH 11 GN I/6
N-ALARM-LIGHT 12 BL 9 BK O/3 I/7
L-ALARM-LIGHT 13 43 BR 25 YW VACDC I/8 61 PK
FLUSH_VALVE 18 68 YW
CR
-VS
+VS
REF
+VS
CNT
GND
MEAS
24VDC+ 19 44 RD
A1
A0
SP
C-24VDC 20 13 BK
24V
GND
GND
GND
GND
GND
51 PR
MEAS
50 GR
70 YW
58 YW
49 WH
C-24VDC 21 13 BK
+ALARM-CRITICAL 22 6 YW
- ALARM-CRITICAL 23 7 BK
+ALARM-WARNING 24 8 YW
- ALARM-WARNING 25 9 BK
6
6
+REN STREAM 1 26 YW
+REN STREAM 2 27 YW
0 BL N 1 0 BL
0 BL N 2 0 BL
0 BL N 3 0 BL
EARTH 47 GN
VALVE-1 48 18 YW
1 BR L 4 1 BR
VALVE-2 49 19 YW
1 BR L 5 1 BR
VALVE-3 50 20 YW
1 BR L 6 1 BR
VALVE-4 51 21 YW
VALVE-5 52 22 YW
5
5
44 RD ++24VDC 7 44 RD
VALVE-6 53 23 YW
44 RD ++24VDC 8 44 RD
VALVE-7 54 24 YW
13 BK C-24VDC 9 13 BK
C-24VDC 55 BK
44 RD
15 YW
68 YW
24 YW
23 YW
22 YW
21 YW
20 YW
19 YW
18 YW
70 YW
17 YW
58 YW
66 YW
72 YW
16 YW
49 WH
13 BK C-24VDC 10 13 BK
+CELL-LEVEL 56 38 WH
49 WH ++12VDC 11 49 WH
- CELL-LEVEL 57
0
9
8
7
6
5
4
3
2
1
15
14
13
12
11
10
L1
49 WH ++12VDC 12 49 WH
+ DC
SLOT 2
OUTPUT
DOS-PUMP-LEVEL 58 60 WH
51 PR C-12VDC 13 51 PR
(1762-OW16)
DOS-PUMP-ERROR 59 61 WH
VAC/VDC RELAY
51 PR C-12VDC 14 51 PR
+DOS-PUMP LEX 60 4 YW
-DOS-PUMP AUX 61 73 YW
4
5
0
1
2
- DOS-PUMP 62 5 BK
REV
+24VDC 63 44 RD
LEAK 64 59 WH
RELAY 1
TEMP 1 65 12 PK
51 PR 8 4
C-24VDC 66 13 BK GND
66 YW 1 BR
4
4
OUT
+VS 7 3
+pH 67 71 RD
6 2
- pH 68 41 PK
Titration module -VS 43 BR 5 1
69 45 PR AG Titration module
70 46 BK
ON 72 YW
RELAY 2
71 62 YW REF
51 PR
DESCRIPTION
8 4
+STIRRER 72 67 GR
15 YW 7 3
-STIRRER 73 51 PR
REVISIONS
6 2
RELAY 3
51 PR 8 4
16 YW 7 3
DATE
62 YW 6 2
2008/07/20
2008/07/24
2013/02/28
2015/05/28
2009/02/09
2011/06/23
48 GR 5 1
3
3
BY
AM
HSS
HSS
MRR
MRR
MRR
12 – 5 VDC REGULATOR
MINTEK.
67 GR +5 V +12 V 17 YW
51 PR COM COM 51 PR
THE WRITTEN PERMISSION OF
DESIGN OR INSTALLATION WITOUTH
NOT BE USED IN THE MANUFACTURE,
IS THE PROPERTY OF MINTEK. IT MAY
THIS DRAWING IS CONFIDENTIAL AND
2
2
SCALE
SIZE
1:2
FSCM NO
NOTES:
LEGEND:
MINTEK
SHEET
DWG NO
STANDARD ETHERNET CABLE
1
1
MAC_CYNV3_WD_R1
WAD CYANIDE ANALYSERS (CLIENT: HERRADURA)
3 OF 8
5
REV
F
A
B
E
C
D
H
G
8 7 6 5 4 3 2 1
A2 NOTES:
H H
MEAS 57 PK
GND 56 PR
49 WH
49 WH
51 GR
51 GR
44 RD
44 RD
13 BK
13 BK
1 BR
1 BR
1 BR
0 BL
0 BL
0 BL
A0
GND
10
11
12
13
14
1
2
3
4
5
6
7
8
9
A1
GND 56 BK
G GND
G
Potensiostat Board
SP
GND
MEAS 58 YW
CR 70 YW
++24VDC
++24VDC
++12VDC
++12VDC
C-24VDC
C-24VDC
C-12VDC
C-12VDC
+VS
N
N
N
L
L
L
CNT
49 GN
49 GN
44 RD
44 RD
13 BK
13 BK
51 BK
51 BK
1 BR
1 BR
1 BR
0 BL
0 BL
0 BL
REF
AG
F -VS 50 GR
F
GND 51 PR
+VS 49 WH
ANALOG INPUT
(1762-IF4)
+VS
-VS
ON
E SLOT 1 E
71 RD
41 PK
57 PK IN 0(+)
module
I+
10 PK IN 1(+)
I-
11 BK IN 1(-)
module
12 PK IN 2(+)
Titration
13 BK IN 2(-)
GND
74 PK IN 3(+)
Titration
O+H
O+
O-
N/-
56 BK IN 3(-)
L
COM
D D
44 RD
10 PK
11 BK
13 BK
COM
OUT
REF
OUT
REF
AG
AG
74 PK
71 RD
12 PK
13 BK
41 PK
C C
65
66
67
68
LEGEND:
STANDARD ETHERNET CABLE
B B
C-24VDC
TEMP 1
- pH
+pH
A
2
1
ISSUED FOR APPROVAL & COMMENTS
2008/07/24
AM
HSS
THE WRITTEN PERMISSION OF
MINTEK. MINTEK
PRIVATE BAG X3015 RANDBURG 2125 A
GAUTENG SOUTH AFRICA +27 11 709 4111
0 ISSUED FOR APPROVAL & COMMENTS 2008/07/20 HSS
SIZE FSCM NO DWG NO REV
8 7 6 5 4 3 2 1
8 7 6 5 4 3 2 1
A2 NOTES:
H H
26 OR
27 OR
28 OR
29 OR
32 OR
33 OR
34 OR
35 OR
30 BK
30 BK
31 BK
31 BK
36 BK
36 BK
37 BK
37 BK
27 OR I OUT 1 33 OR I OUT 1
28 OR I OUT 2 34 OR I OUT 2
ISOLATOR CARDS
1+
3+
1+
3+
1+
3+
1+
3+
2-
4-
2-
4-
2-
4-
2-
4-
29 OR I OUT 3 35 OR I OUT 3
V OUT 0 V OUT 0
4
V OUT 1 V OUT 1
V OUT 2 V OUT 2
5+
7+
5+
7+
5+
7+
6-
8-
6-
8-
6-
8-
6-
8-
7+
5+
V OUT 3 V OUT 3
30A1 BK
31A1 BK
36A1 BK
37A1 BK
F F
26A OR
27A OR
28A OR
29A OR
32A OR
33A OR
34A OR
35A OR
30A BK
31A BK
36A BK
37A BK
30 BK COM 36 BK COM
31 BK COM 37 BK COM
E E
30A1BK
36A1BK
37A1BK
26A OR
27A OR
28A OR
29A OR
32A OR
33A OR
34A OR
35A OR
30A BK
31A BK
31A BK
36A BK
37A BK
63 WH
64 WH
69 BK
55 PK
56 BK
D D
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
TERMINATION BLOCK
BOTTOM
+REN STREAM 1
+REN STREAM 2
C-STREAM
C C
AO1
AO2
AO3
AO4
AO5
AO6
AO7
AO8
AI1
C
LEGEND:
STANDARD ETHERNET CABLE
B B
A
2
1
ISSUED FOR APPROVAL & COMMENTS
2008/07/24
AM
HSS
THE WRITTEN PERMISSION OF
MINTEK. MINTEK
PRIVATE BAG X3015 RANDBURG 2125 A
GAUTENG SOUTH AFRICA +27 11 709 4111
0 ISSUED FOR APPROVAL & COMMENTS 2008/07/20 HSS
SIZE FSCM NO DWG NO REV
8 7 6 5 4 3 2 1
8 7 6 5 4 3 2 1
A2 NOTES:
H H
-/+12VDC (36W)
MEAS 57 PK
-/+ 12VDC
GND 56 PR
A0 50 GR
-
GND 56 BK 51 PR COM
G 49 WH
G
+
A1
GND 56 BK
100-240VAC
24V 0 BL
GND 1 BR L
Potensiostat Board
SP E
GND
MEAS 58 YW
CR 70 YW
+VS
F CNT 54 PK
F
REF 53 PK
AG 52 PK
-VS 50 GR
VAC/VDC RELAY
GND 51 PR
ANALOG INPUT OUTPUT
+VS 49 WH
(1762-IF4) (1762-OW16)
SLOT 1 SLOT 2
+ DC
+VS
-VS
57 PK IN 0(+)
ON
56 PR IN 0(-) 15 YW 0
10 PK IN 1(+) 16 YW 1
E 11 BK IN 1(-) 2 E
Titration module 12 PK IN 2(+) 66 YW 3
13 BK IN 2(-) 58 YW 4
IN 3(+) 17 YW 5
IN 3(-) 70 YW 6
56 BK
GND
COM 7
COM
44 RD L1
18 YW 8
19 YW 9
D 20 YW 10 D
OUT
REF
AG
21 YW 11
22 YW 12
53 PK
52PK
23 YW 13
24 YW 14
68 YW 15
49 WH
67 GR
51 PR
51 BK
53 PK
52 PK
54 PK
C C
73
74
75
77
72
76
78
BOTTOM TERMINATION BLOCK
LEGEND:
STANDARD ETHERNET CABLE
+STIRRER
-STIRRER
B B
+ 12 VDC
C-12VDC
CN_AUX
CN_REF
CN_AG
A
2
1
ISSUED FOR APPROVAL & COMMENTS
2008/07/24
AM
HSS
THE WRITTEN PERMISSION OF
MINTEK. MINTEK
PRIVATE BAG X3015 RANDBURG 2125 A
GAUTENG SOUTH AFRICA +27 11 709 4111
0 ISSUED FOR APPROVAL & COMMENTS 2008/07/20 HSS
SIZE FSCM NO DWG NO REV
8 7 6 5 4 3 2 1
8 7 6 5 4 3 2 1
A2 NOTES:
H H
G G
GN EARTH 47 EARTH 1 GN
18 YW VALVE-1 48 +CELL-LEVEL 2 38 BK
19 YW VALVE-2 49 -CELL-LEVEL 3
20 YW VALVE-3 50 DOS-PUMP-LEVEL 4 60 BL
21 YW VALVE-4 51 DOS-PUMP-ERROR 5 61 BK
22 YW VALVE-5 52 +DOS-PUMP LEX 6 4 BK
23 YW VALVE-6 53 +DOS-PUMP AUX 7 4 BK
24 YW VALVE-7 54 - DOS-PUMP 8 5 WH/BR
F 13 BK C-24VDC 55 +24VDC 9 44 RD F
38 PK +CELL-LEVEL 56 +24VDC 10 44 RD
- CELL-LEVEL 57 +24VDC 11 44 RD
60 WH DOS-PUMP-LEVEL 58 +24VDC 12 44 RD
61 WH DOS-PUMP-ERROR 59 +24VDC 13 44 RD
4 YW +DOS-PUMP LEX 60 LEAK 14 59 BK
73 YW +DOS-PUMP AUX 61 TEMP 1 15 12 PK
5 BK - DOS-PUMP 62 C-24VDC 16 13 BK
44 RD +24VDC 63 C-24VDC 17 13 BK
59 WH LEAK 64 C-24VDC-FAN 18 13 BK
E 12 PK TEMP 1 65 - pH 19 41 PK E
13 BK C-24VDC 66 + pH 20 71 RD
71 RD +pH 67 +12 VDCBOX-FAN 21 49 RD
41 PK - pH 68 +TEC-PAD 22 45 OR
45 PR +TEC-PAD 69 - TEC-PAD 23 46 BK
46 BK - TEC-PAD 70 TEC-FAN 24 62 RD
62 YW TEC-FAN 71 +STIRRER 25 67 BR
67 PR + STIRRER 72 -STIRRER 26 51 GN
51 PR -STIRRER 73 C-12VDC-TEC-FAN 27 51 GR
49 WH +12VDC 74 C-12VDC-BOX-FAN 28 51 GR
C VALVE-4 39 21 RD C
VALVE-5 40 22 RD
VALVE-6 41 23 RD
VALVE-7 42 24 RD
LEGEND:
EARTH-VALVES 43 GN
STANDARD ETHERNET CABLE
EARTH-VALVES 44 GN
EARTH-VALVES 45 GN
EARTH-VALVES 46 GN
A
2
1
ISSUED FOR APPROVAL & COMMENTS
2008/07/24
AM
HSS
THE WRITTEN PERMISSION OF
MINTEK. MINTEK
PRIVATE BAG X3015 RANDBURG 2125 A
GAUTENG SOUTH AFRICA +27 11 709 4111
0 ISSUED FOR APPROVAL & COMMENTS 2008/07/20 HSS
SIZE FSCM NO DWG NO REV
8 7 6 5 4 3 2 1
8 7 6 5 4 3 2 1
A2 NOTES:
H H
38
39
60
61
73
44
44
59
12
13
13
41
71
49
67
51
51
51
52
53
54
13
13
18
19
20
21
22
23
24
4
5
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
PH - TRANSMITTER
0
1
2
3
4
5
6
7
8
9
TEMP - TRANSMITTER
RTD
-
PK
BOX-FAN
41 +
RD - + -
+
71
G G
44 RD
12 PK
DOS-PUMP-ERROR
B+
-
B-
+
A
DOS-PUMP-LEVEL
BK_2 +DOS-PUMP_AUX
13 BK C-24VDC-VALVES
13 BK C-24VDC-VALVES
13 BK C-24VDC-VALVES
13 BK C-24VDC-VALVES
BK_1 +DOS-PUMP_LEX
EARTH-VALVES
EARTH-VALVES
EARTH-VALVES
EARTH-VALVES
BR_D - DOS-PUMP_C
49 RD
51 BL
38 BK +CELL-LEVEL
39 BK -CELL-LEVEL
49 RD + BOX FAN
67 BR +STIRRER
51 GR -STIRRER
13 BL C-24VDC
13 BL C-24VDC
51 BL C-12VDC
51 BL C-12VDC
18 RD VALVE-1
19 RD VALVE-2
20 RD VALVE-3
21 RD VALVE-4
22 RD VALVE-5
23 RD VALVE-6
24 RD VALVE-7
54WH CN-AUX
+24VDC
+24VDC
+24VDC
+24VDC
+24VDC
CN-REF
12 PK TEMP 1
EARTH
52 BK CN-AG
59 BK LEAK
71 RD + pH
41 PK -pH
GN
BK
BL
BL
44
44
44
44
44
E
E
E
E
WET BOX ENCLOSURE
F F
VALVE 1
LEVEL SENSOR 18 RD +
E
51 GR
67 BR
E 1 2 13 BK - E
STIRRER
-
44 BR
13 BL
VALVE 2
DOSING-PUMP_AUX
pH – CALIBRATION
19 RD +
44 BR + 24 VDC
38 BK E
SOCKET 3
PROBE
44 WH + 24 VDC
LEVEL / 3 13 BK -
60 BL Level
ERROR
61 BK Error
VALVE 3
5 BR GND SOCKET 4
20 RD +
73 BK Contact CONTACT
E
44 BR
+
24 VDC
D 13 BL - POWER
13 BK - D
SUPPLY
GN
VALVE 4
CN-AG CN-AUX 21 RD +
52 BK 54 WH E
13 BK -
TERMO- VALVE 5
COUPLE 22 RD +
DOSING-PUMP_LEX E
C 44 BR + 24 VDC + 13 BK - C
44 WH + 24 VDC SOCKET 3 -
LEVEL /
60 BL Level VALVE 6
ERROR
61 BK Error 23 RD +
LEGEND:
5 BR GND SOCKET 4 E
CONTACT STANDARD ETHERNET CABLE
73 BK Contact 13 BK -
44 BR
+
24 VDC
13 BL - POWER VALVE 7
SUPPLY
GN 24 RD +
E
B 13 BK - B
44 BR
59 BK
13 BL
LEAK DETECTOR
-
+
A
2
1
ISSUED FOR APPROVAL & COMMENTS
2008/07/24
AM
HSS
THE WRITTEN PERMISSION OF
MINTEK. MINTEK
PRIVATE BAG X3015 RANDBURG 2125 A
GAUTENG SOUTH AFRICA +27 11 709 4111
0 ISSUED FOR APPROVAL & COMMENTS 2008/07/20 HSS
SIZE FSCM NO DWG NO REV
8 7 6 5 4 3 2 1