Biological Safety Cabinet Maintenance

Manual

Thank you for Choosing Biolab products. Please read the “Operating Instructions” and
“Warranty” before operating this unit to assure proper operation.
Biological Safety Cabinet Maintenance Manual

Index
01 Device Debugging.........................................................................................................03
02 Failure and Maintenance..............................................................................................07
03 The Fan and Filter Replacement................................................................................23
04 Wiring Diagram................................................................................................................47

2 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

01 Device Debugging
Standard values: inflow 0.53m/s, downflow 0.33m/s
Adjustment of actual wind speed

Attention:
Safety cabinets before leaving the factory, wind speeds have already adjusted to standard
values.
Install and run the device, if wind speeds in screen are inconsistent with standard values,
please test the actual wind speed values with anemograph(Test down flow) and weather
gauge(Test inflow)
If you do not have the weather gauge, you could also use the anemometer instead.(But
accuracy is not guaranteed)

anemometer

If test wind speed and standard values are consistent, following the Display pressure and wind
speed adjustment in part2 Failure and Maintenance,fault 12.

If the two are inconsistent, need to adjust the speed gears of fan, you need following the The
actual wind speed adjustment in part1 first, and then follow the part 2 Failure and
Maintenance, fault 12. the steps are as follows :

www.biolabscientific.com 3
Biological Safety Cabinet Maintenance Manual

The actual wind speed adjustment of the new control board A2 type safety cabinet

Steps:
The wind speed adjustment is divided into two situations(A or B):
A: If the inflow and down flow wind speed are both greater than or less than the nominal
value, adjust the setting of the fan gear:

1.Turn on the power to turn on the power lock,Enter the standby interface.
(Standby state refers to the situation where the power lock is open but the power
button is not pressed. The display screen is not light up)

2. Press The fan key which is shown in picture 1 for about 5 seconds to enter the
adjustment interface of the fan speed.

Picture 1

3. After entering the fan speed adjustment interface, you can select to adjust the wind speed
of the external exhaust fan (the left side Number)or inner fan (the right side number) by
pressing the mute (sound) button

For example , there are 60(left) and 50(right) Numbers in picture 2 will blink .

The 60 number is the speed adjustment of the external exhaust fan, which is the inflow of
wind speed adjustment.

4 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

The 50 number is the internal fan speed adjustment, which is the downflow wind speed
adjustment.

Value of these two Numbers are for the inner and outer fan gears, The larger the value, the
larger the flow velocity, and the smaller the numerical value, the smaller the flow velocity, and
by adjusting the two values, the wind speed can reach the nominal value.

Picture 2

4. After adjusting (inflow 0.53m/s, downflow 0.33m/s, which is showed on the anemograph),
press the fan key for 5 seconds to save.

B: When the inflow of wind speed and the down flow wind speed do not match, the inflow
wind speed is less than the nominal value and the down flow wind speed is greater than the
nominal value, or the inflow of wind speed is greater than the nominal value and the down
flow wind speed is less than the nominal value:

1. Loosen the retaining bolt of the top adjusting plate, as shown in figure 3, adjust the
adjusting plate at the top:

2. Pull out the adjustment plate, the inlet air flow velocity increases, the down flow velocity
decreases;

www.biolabscientific.com 5
Biological Safety Cabinet Maintenance Manual

3. When the adjustment plate is pushed in, the inlet air flow velocity decreases and the down
flow velocity increases.

4. To match the inflow wind speed and the down flow wind speed, after adjusting, tighten the
retaining bolt at the top.

Picture 3

Next, you need to adjust the display speed on the display screen. Please refer to Part 2

Attention:
All operations of Part 2 are built on the premise that the actual wind speed has reached the
nominal value through the anemometer and the weather gauge.

6 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

02 Failure and Maintenance


CAUTION: All maintenance actions on this equipment must be performed by a qualified
technician who is familiar with the proper maintenance procedures required for this
equipment. This includes both certification as well as repair.

Fault 1 : Equipment no electricity


After electrifying equipment and turn on power lock, equipment did not respond (no alarm,
display is not bright, buttons no reaction).
The step deal with this type of problems as follows:
1.1 Ensure the power supply has electric and electrical parameters consistent with equipment
requirements. Using a multimeter to test the device power supply is normal.
1.2 Equipment equipped with a fire line fuse. You can remove the fuse with a screwdriver and
make sure the fuse is undamaged. If the fuse is damaged, you should replace with the fuse of
corresponding specification.

www.biolabscientific.com 7
Biological Safety Cabinet Maintenance Manual

Picture 4

1.3 Confirm the above no problem, open the operation panel (as shown in picture5),using the
support frame (fixed the operation panel as picture) hold up operation panel in a safe
position, using a multimeter to test input and output voltage with power lock opening
(voltage should be the power supply voltage).

8 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

Picture 5

1.4 Confirm the above no problem, use multimeter testing input voltage of transformer
(voltage values should be the supply voltage of the device) and the output voltage. First use
multimeter test two lines at CN9(as shown in picture6),check whether there have voltage, this
voltage is the supply voltage. Then test the transformer out put voltage: AC test with a
multimeter, if CN10 voltage is between 12V-14V , indicating it is normal; If there is no voltage,
indicating that transformer is damaged, please replace the transformer.

www.biolabscientific.com 9
Biological Safety Cabinet Maintenance Manual

Picture 6

10 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

Picture 7

1.4 Confirm the above no problem, check the cable connection between CN11(in picture6)
and CN1(in picture7).
1.5After confirming the above no problems, please replace control board.

Fault 2: The display is not bright (Normal alarm when power on)
After electrifying equipment and turn on power lock, button functions are normal, the display
is not bright. The step deal with this type of problems as follows:
2.1 Check whether the screen is broken, if it is damaged, replace display panels.
2.2 Open the operation panel (picture5), check the cable between display panel and control
panel is correctly connected.
2.3 Confirm the above no problem, please replace control board.

Fault 3: Button no response and display is not bright (Normal alarm when power on)
After electrifying equipment and turn on power lock, button functions no response, the
display is not bright The step deal with this type of problems consult fault 2

Fault 4 : Power lock is not responding


Power lock is not responding, please replace power lock. Method for removing power lock as
shown in picture 8.
( BCBS-201/ BCBS-204 the power lock type is AR22JR-2A20A;
BCBS-202/ BCBS-203 the power lock type is LA42Y2P-20B)

www.biolabscientific.com 11
Biological Safety Cabinet Maintenance Manual

Picture 8

Fault 5: No alarm when the front window is not in the safe height
NO alarm when the front window is not in the safe height (Normally higher or below the safe
height(200mm) the device will alarm; when front window completely at the bottom, it won’t
alarm)), the front window will automatically stop and stop alarm when it reaches the safe
height during the up and down.
The step deal with this type of problems as follows:
5.1When the current window is not in the safe height, confirm whether" "on the screen is
flashing, if the icon is flashing and not in the mute mode, but the device don’t alarm,
indicating control board needs to be replaced.
5.2 If isn’t flashing, open the operation panel (see picture5). Using a multimeter to test
whether micro-switches are damaged or short circuit(Test H, L,CN10- Double travel
switch/CN10-Single travel switch,see picture7), make sure travel switch lever contact with
glass and the switch can be closed.
When front window is in safety height, high limit travel switch of double switch bracket is
disconnected, low switch of double switch bracket and travel witch of single switch bracket is
closed (as shown in Picture9).
When front window is not in safe height, loose bolts, adjusting travel switch to make double
switch bracket of high switch in disconnect mode, low switch and travel switch of single switch
bracket is closed (as shown in Picture10).
This method is also applicable the front window reach safe height does not automatically
stop.

12 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

Picture 9

Check whether the welding wiring harness of two micro switch on the double micro switch
bracket is firmly connected, check whether the micro switch is damaged, and whether the
switch action can be made when the glass is moving.

www.biolabscientific.com 13
Biological Safety Cabinet Maintenance Manual

Picture 10

5.3 Check wiring connection in ‘CN8’ (as shown in picture7), high switch connect to ‘CN8
H’,low switch connect to the ‘CN8 L’. Single-switch connect to ‘CN10’ (as shown in picture7).
5.4 Confirm the above no problem, please replace control board.

Fault 6 : Front window failure to up and down


The front window is not response when press up and down keys. The step deal with this type
of problems as follows:
6.1Operation panel’s buttons no response
Turn on the device, open operation panel (see 1.3,picture5),hold down the ‘up’ button, using
multimeter test voltage(voltage should be the power supply voltage) of CN1(between
blue(white) line and brown(red) line) at strong electric board (as shown in picture7), hold
down the ‘down’ button, using multimeter test voltage(voltage should be the power supply
voltage) of CN1(between blue(white) line and black line) at strong electric board (as shown in
picture7), if there is no voltage, control panel is damaged, replace the control board.
If above is normal, connected motor to the power supply directly, test whether the motor is
working,.If the glass door motor is not working, indicating the motor is damaged, replace the
glass door motor (glass door motor position as shown in picture11).

14 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

Picture 11

6.2 Remote control button does not respond


Confirm the remote control have battery . If the remote control have changed battery,
device still no response, remote control is damaged,please replace remote control.If using a
new remote control still no function,see 6.1.
6.3 Foot switch does not respond
Using the multimeter to test whether the foot switch is damaged.( Remove the foot switch
and press the two pedal, separately identify the terminals 1 and 2, the 2 and the 3 switches
are respectively connected.If the foot switch wire is broken, the foot switch is damaged and
please replace the foot switch).
Open the operation panel, check CN8 and CN7(as shown in Picture7), the blue line (GND)
and the red line (UP), the black line (DOWN) are good contact on the control panel.If they are
good contact,see 6.1.

Foot switch terminal

www.biolabscientific.com 15
Biological Safety Cabinet Maintenance Manual

Fault 7 : Remote control failure


Remote control failure refer to the device no function when press remote buttons.
Confirm the remote control have battery. If the remote control have changed battery, device
still no response, remote control is damaged, please replace remote control. If using a new
remote control still no function, indicating the circuit board is damaged, please replace the
circuit board.

Fault 8 : The socket(operating area)has no electricity


Sockets(operating area)have no electricity when press socket button. The step deal with
this type of problems as follows:
8.1 Check fire line fuse (as shown in picture4), if the fuse is broken, please replace it.
8.2 Remove sockets and check whether the socket is damaged. Press(hold)socket button,
using a multimeter test socket voltage between null line and firing line. If the voltage is
normal (voltage should be the power supply voltage of the instrument), the Waterproof socket
will be damaged, please replace sockets.

Picture 12

8.3 Confirm the above no problem, open the operation panel(see 1.3 picture5). Use
multimeter to check whether the line is disconnected, if it is, please connect it again. Socket
line connect to CN4 in strong electric board (Note: Check circuits under cut off the power).
8.4 If above no problem, electrifying equipment and turn on power lock, press the power
button and socket button, observe whether the socket icon is bright, if it is not bright, control
board or display board is damaged, replace the circuit board.
8.5 If above no problem, press the power button and the socket button, using multimeter to

16 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

test CN4 voltage(voltage should be the power supply voltage of the instrument) in the control
board (as shown in Picture7), if no voltage, control board is damaged, please replace the
control board.

Fault 9 : Fluorescent lamp doesn’t work


Fluorescent lamp doesn’t work when press ‘light’ key . The step deal with this type of
problems as follows:
9.1 Check the contact between lamp and lamp holder is good,.Check whether the lamp is
damaged(Confirm whether the tube body on both sides of the lamp is black),if the lamp is
damaged,please replace the same specification lamp.

9.2 If the above no problem, open the operation panel(see 1.3 picture5), using multimeter to
test CN5 voltage(voltage should be the power supply voltage of the instrument) in the control
board (as shown in Picture7), if not ,control board is damaged, please replace the control
board.
9.3 If above no problem, check whether the connection between lamp and ballast is loose ,if
loose, reconnect the wires .

www.biolabscientific.com 17
Biological Safety Cabinet Maintenance Manual

Picture 13

9.4 If above no problem, ballast is damaged, replace the new same specification lamp
ballast.

Fault 10 : UV light does not work


UV light doesn’t work when press ‘UV’ key.
Note: Only in the front window on the bottom ,the UV light is on.The step deal with this
type of problems as follows:
10.1 Check the contact between lamp and lamp holder is good,.Check whether the lamp
is damaged,if the lamp is damaged,please replace the same specification lamp.After
replacement,press UV botton 5s In standby mode(refer to page2) to reset the UV use time.

10.2 If above no problem,open the operation panel(see 1.3 picture5),check whether control
UV interlock switch is in the off state and the wire is disconnected, connect to the " CN10".
10.3 If above no problem,to power the device and turn on the power lock, press the power

18 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

button and the UV key in turn .Using multimeter to test CN6 voltage (voltage should be the
power supply voltage of the instrument)in the control board (as shown in Picture7), if not
control board is damaged, please replace the control board.

Picture 14

10.3 If above no problem, check whether the connection between UV lamp and UV ballast
is loose ;check whether the connection between control board and UV ballast is loose,If loose,
reconnect the wires .
10.4 If above no problem, ballast is damaged, replace same specification UV ballast

Fault 11 : Fan does not work


Fan does not work when press fan key, the fan will not start.
Note: when the glass is drop to the bottom of the instrument, the fan will not start
The step deal with this type of problems as follows:
11.1 Electrifying equipment and turn on power lock, press fan button,check whether the
fan icon is bright。If it isn’t bright, open the operation panel(see 1.3 picture5), Check whether
the UV lamp interlock microswitch is in a closed state or whether the connection wire is
damaged, connect to the " CN10"(picture6).
If bright, using multimeter test voltage(voltage should be in 120V-200V ).) of CN2.
Or, check whether the fan insurance tube fuse, and insurance tube seat harness out
position (see figure 1), if the fuse break down please replace the same type of insurance.
11.2 If above no problem, connected the fan to the power supply directly and check
whether the fan starts, if it is not started, fan is damaged, replace the same model fan, fan
wire is connected to the " CN2 ".
11.3 If above no problem, check whether wire connection between the fan and circuit

www.biolabscientific.com 19
Biological Safety Cabinet Maintenance Manual

board are loose, if it is loose connections, reconnect it.

Fault 12: Display pressure and wind speed adjustment


12.1 Before adjusting the wind speed, first check whether the gas-guide tube is twist or
bent

12.2 Setting:
Pressure regulation
The pressure of the filter (air supply filter and exhaust filter) of A2 type safety cabinet is
between 80 and 110Pa.
If the pressure is in the range, there is no need to adjust the pressure.
If you do not need to adjust the pressure, simply click the fan key and wait for 5
minutes to start regulating the wind speed.
If the pressure is off, it needs to adjust. The method is as follows:

Steps:
1. Press the fan key to open the fan.

2.Wait for the fan to work for about 5 minutes, Let the wind speed reach a stable state. long

press the socket key about 5 seconds, after the alarm sound, enter the adjustment
pressure state.

3.First, adjust the pressure of the downflow filter (the number on the right side of the screen
flashing, for example "50"), and adjust the key to the nominal value through the upper
and lower keys;

4.And then press the mute(sound) button, switch to the exhaust filter pressure
adjustment (the number on the left side of the screen flashing, for example "60"), through the
upper and lower keys adjust the minor to the nominal value.

5.After adjustment, press socket key about 5 seconds to save.

Wind speed adjustment


Steps:
If you do not need to adjust the pressure, simply click the fan key and wait for 5 minutes to
start regulating the wind speed.

1. After the pressure adjustment has been completed, the fan has been in a running state.
Then adjust the wind speed, press the UV key for about 5 seconds, then enter the wind
speed adjustment state after hearing the alarm sound.

20 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

2.First, adjust the downflow wind speed (the number on the right side of the screen flashing,
for example "50"), and adjust to the nominal value through the upper and lower
keys;

3.Then press the mute(sound) button to switch to the inflow speed adjustment (the
number on the left side of the screen flashing, for example "60"), and adjust to the nominal

value through the upper and lower keys

After adjusting, press the UV key for about 5 seconds to save

Fault 13 Filter pressure differential high alarm, replace the filter

When barcodes is full on the screen, the device will alarm and the filter need to be replaced.

After replacing the filter, pressing the ‘light’ button 5 seconds(In standby mode, refer to

page2) to reset the using time. Steps of filter replace see Part3.The fan and filter replacement.

www.biolabscientific.com 21
Biological Safety Cabinet Maintenance Manual

Fault 14 : Front window broken, replace the front window glassOpen the operation panel (see

1.3 picture5), remove right glass decoration strip, remove right PVC glass guide (keep upper

screws are not removed),take off the glass locking buckle, remove the glass; install a same

specification front window.

The fault 15 : Adjust display time


electrifying equipment and turn on power lock, in the standby state(refer to Page2) . Press the

22 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

light button first, then press the power button for about 5 seconds. After hearing the alarm,
enter the clock adjustment state.
At the beginning,“minutes” flashing,through up and down keys to adjust to the current
time;
press the mute button to switch to the “hour”, by up and down keys to adjust to the current
time.
After the clock has been adjusted, Press the light button first, then press the power button for
about 5 seconds to save the adjustment.

Picture 20

Fault 16 : Glass door motor replacement


First disconnect the belt, remove the spring as shown, lift the motor of the left, remove the
motor. Install the new glass door motor.

www.biolabscientific.com 23
Biological Safety Cabinet Maintenance Manual

Fault 17 : Temperature and humidity display incorrectly


Right corner of the screen displays don’t fit the actual temperature and humidity. The step
deal with this type of problems as follows:
1.Check whether the temperature and humidity sensor wiring is connected, as shown in
picture7 " CN11 "“H、GND、VCC、T”.
2.Check whether temperature and humidity sensor is broken. Replace the same type sensor, as
shown in picture 23.

24 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

Picture 23

3.If above no problems, indicating circuit board is broken, need to replace the circuit board.

Fault 18: Gas-guide tube replacement instructions


According to the original connection way to reconnect.

Notice: The gas-guide tube should not be bent or twisted


Gas-guide tube installation instructions, FIG. A, B, C (three styles):
Black 1 2 3 4 represent the front Gas-guide tube position on the upper right corner of the
instrument (after the operation panel of the instrument is opened).

www.biolabscientific.com 25
Biological Safety Cabinet Maintenance Manual

FIG A BCBS-201

26 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

Picture24(Sensor Gas-guide tube position)

Fault 19 : Inflow velocity and down velocity are zero


When the inflow velocity is zero, first determine whether gas-guide ducts are twisted or bent,
and open the operation panel(see 1.3 picture5), recovery ducts;Then test whether the fan is
working, If the fan doesn’t work, according to the fault 11 to deal with.

03 The Fan and Filter Replacement


3.1 Replace filter and the old filter processing
Before replacing the filter, have to fumigate and sterilize the cabinet.
In order to ensure the safety of personnel ,the operating personnel must wear masks, latex
gloves, safety glasses and positive pressure helmet, suit and shoes when replacement.
The old filter should be placed in a special bag for medical waste, sealed and labeled with
"hazardous waste".

www.biolabscientific.com 27
Biological Safety Cabinet Maintenance Manual

Picture 25

3.2 BCBS-203 and BCBS-204 Safety Cabinet replace fan and filter steps:
a) Place the glass door to the bottom, cut off the power, open the operation panel.
b) Take off the glass lock button, loosen the conveyor belt,remove travel switch bracket,
remove weld screws for securing front panel, remove the front panel.

28 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

c) Remove press bar nut

d) Remove compressive bar

e) Remove internal front panel, pull out the ducts, record the position.

www.biolabscientific.com 29
Biological Safety Cabinet Maintenance Manual

f) Remove the fan installation plate, pay attention to the safety of the bottom glass

g) Remove the filter and replace the filter with the same specification

h) Remove filter press bar bolt

i) Remove ducts, take the filter tray and the filter, pay attention to the arrow direction

30 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

j) Replace the same type of upper filter, adjusting tray and filter, reconnect ducts,

tighten screw nuts

k) Remove the fan bolts, remove the fan box

www.biolabscientific.com 31
Biological Safety Cabinet Maintenance Manual

l) Cut white porcelain between the fan and fan plate with a blade.

M) Remove the fan and mounting plate connecting piece

n) Replace the same type of fan,install fan mounting plate connector and gluing.

32 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

o) Install the fan and the fan box

p) Sealant coating

p) Put in the below filter, pay attention to the direction of the arrow is down

q) Put in the below filter, pay attention to the direction of the arrow is down

www.biolabscientific.com 33
Biological Safety Cabinet Maintenance Manual

r) Install the fan installation board,reconnet ducts.

s) Air duct connection part gluing treatment

t) Reconnect wiring lines.

u) Install inner front panel

34 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

v) Install press bars, tighten nuts.

w) Install the front panel and travel switch bracket.

www.biolabscientific.com 35
Biological Safety Cabinet Maintenance Manual

x) Install glass lock button.

3.3BCBS-201 Safety Cabinet replace fan and filter steps:


a) Place the glass door to the bottom, cut off the power, open the operation panel.
b) Take off the glass lock button, loosen the conveyor belt,remove travel switch bracket,
remove weld screws for securing front panel, remove the front panel.

c) Remove press bar nut

36 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

d) Remove air duct connecting plate

www.biolabscientific.com 37
Biological Safety Cabinet Maintenance Manual

e) Disconnect fan connection

f) Remove press bar and bolt.

38 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

g) Cut white porcelain between the fan and fan plate with a blade

www.biolabscientific.com 39
Biological Safety Cabinet Maintenance Manual

h) Remove the fan installation plate, remove below filter.

40 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

i) Remove upper press bar, remove the upper filter and plate

Uniform net
Fan

j) After replacing the filter, recover them.

Replace the fan steps: take the fan out of the safety cabinet,According to the

following figure, remove the uniform net, unloading fan, replace the same type

of fan, reinstall the disassembled components to the device.

www.biolabscientific.com 41
Biological Safety Cabinet Maintenance Manual

3.4 BCBS-202 Safety Cabinet replace fan and filter steps:


a) Place the glass door to the bottom, cut off the power, open the operation panel.
b) Take off the glass lock button, loosen the conveyor belt,remove travel switch bracket,
remove weld screws for securing front panel, remove the front panel.

c) Remove ducts and wiring harness.

42 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

d) Remove press bar.

e) Remove inner front panel

www.biolabscientific.com 43
Biological Safety Cabinet Maintenance Manual

f) Cut off air duct sealant with a blade.

g) Remove travel switch.

h) Remove the fan installation plate

44 www.biolabscientific.com
Biological Safety Cabinet Maintenance Manual

i) Remove upper filter bracket bolt

a) Remove upper filter bracket and upper filter.

k) After repla

the filter, reinstall the disassembled components to the device

l) Replace the fan, take out the fan from safety cabinet, remove uniform net and fan,

replace same type fan

www.biolabscientific.com 45
Biological Safety Cabinet Maintenance Manual

46 www.biolabscientific.com
47
Biological Safety Cabinet Maintenance Manual

www.biolabscientific.com
1 2 3 4 5 6 7 8
D T&H SENSOR D

D O O R O P EN D ET EC TI ON

F O O T SW I T C H U P

F O O T S W IT CH D OW N
BLACK(Φ0.5)

W IN DO W H

W IN DO W L
BLACK(Φ1)
U V BA L L A S T

BLACK(Φ0.5)
WHITE(Φ0.5)
UV LAMP WHITE(Φ0.5)
BLUE(Φ1)

R E D ( Φ 0 .5 )
R E D ( Φ 0 .5 )

W H I T E ( Φ 0 .5 )

B L A C K ( Φ 0 .5 )

B L U E ( Φ 0 .5 )
1 3
2

W in do w
H
L

GND

V CC
H T UV Ctrl

G ND
BLUE(Φ0.5)
RED(Φ0.5)
up

F oo t
YELLOW(Φ0.5) down
C BLACK(Φ0.5) C
LAMP YELLOW(Φ0.5)
220 V/50 Hz

A NJ IA N
1 MCU EIGHT CORE CABLE
LCD

L IG H T BA L LA S T
2
RED(Φ0.5) GREEN(Φ1)
3
BCBS-201 / BCBS-202 wiring diagram

RED(Φ0.5)
4
TRANS-IN
5 BLUE(Φ1)
04 Wiring Diagram

BLACK(Φ0.5) 6 L
EX-FAN 12V/0 .5A
BLACK(Φ0.5) 7 N
8
BLUE(Φ0.5) L
LAMP
N TRANS-OUT SPEAK
BLUE(Φ0.5)
L

UV
N
SIXTEEN CORE CABLE
SOCKET L

L IG HT S OC KE T L
N
RED(Φ1) FUSE3 RED(Φ1) SPEAKER
L
5A N
BLUE(Φ1)
YELLOW GREEN (Φ1.0) BROWN(Φ1.0)

RE D( Φ 1)

D OW N

D OO R MO TO R
BLUE(Φ1.0) DOWN-FAN N UP

N
L N
B SOCKET B
YELLOW GREEN (Φ1.0)
BLUE(Φ1) BLACK(Φ1) BROWN(Φ1)
YELLOW(Φ1) BLUE(Φ1)
BROWN(Φ1.0) FUSE2 DOOR MOTOR
L

W A V E FIL T ER
10A DOWN FAN
POWER POWER LOCK
YELLOW GREEN (Φ1.0)
YELLOW GREEN (Φ1.0)
BLUE(Φ1.0)
N
A A
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8
D T&H SENSOR D

D OO R OPEN D ET EC TI ON

F OO T SWIT CH U P

F OO T SW IT CH D OW N
BLACK( Φ 0.5)

W IN DO W H

W IN DO W L
BLACK( Φ1)
U V BA LL A S T

BLACK( Φ0.5)
WHITE( Φ 0.5)
UV LAMP WHITE( Φ 0.5)
BLUE( Φ1)

R E D( Φ 0 .5 )
R E D ( Φ 0.5 )

W H I T E ( Φ 0 .5 )

B L A C K ( Φ 0 .5 )

B L U E ( Φ 0 .5 )
1 3
2

W in do w
H
Biological Safety Cabinet Maintenance Manual

GND

V CC
H T UV Ctrl

G ND
BLUE( Φ0.5)
RED( Φ 0.5)
up

Foo t
YELLOW(Φ 0.5) down
C BLACK( Φ 0.5) C
LAMP YELLOW( Φ 0.5)
220 V/ 50 Hz

A NJ IA N
1 MCU EIGHT CORE CABLE
LCD
LIGHT BALLAS T

2
RED( Φ 0.5) GREEN( Φ1)
3
RED( Φ 0.5)
4
TRANS-IN
5 BLUE( Φ 1)
BLACK( Φ0.5) 6 L
EX-FAN 12V/0 .5A
BLACK( Φ0.5) 7 N
8
BLUE( Φ 0.5) L
LAMP
N TRANS-OUT SPEAK
BLUE( Φ 0.5)
L

UV
N
SIXTEEN CORE CABLE
SOCKET L

L IG HT
N
RED( Φ 1) FUSE3 RED( Φ 1) SPEAKER
L

S OC KE T
5A N
BLUE( Φ1)
YELLOW GREEN ( Φ 1.0) BROWN( Φ 1.0)
R E D( Φ 1 )

D OW N

D OO R MO TO R
BLUE( Φ1.0) DOWN-FAN N

www.biolabscientific.com
UP

N
L N
B SOCKET B
YELLOW GREEN (Φ 1.0)
BCBS-203 wiring diagram

FUSE4
10A BLUE( Φ 1) BLACK( Φ 1) BROWN( Φ 1)
YELLOW( Φ 1) BLUE( Φ1)
BROWN( Φ 1.0) FUSE2 DOOR MOTOR
L

W A V E FIL T ER
10A DOWN FAN
POWER POWER LOCK
YELLOW GREEN (Φ 1.0)
YELLOW GREEN ( Φ 1.0)
BLUE( Φ1.0)
N
A A

48
1 2 3 4 5 6 7 8
49
Biological Safety Cabinet Maintenance Manual

1 2 3 4 5 6 7 8

www.biolabscientific.com
T&H SENSOR

D OOR OP EN D ET ECTI ON
D D

FOO T SWIT CH D OW N
BLACK( Φ0.5)

F OOT SW ITCH U P
U V BA LL AS T

BLACK( Φ1)

WIN DO W L
BLACK( Φ0.5)

W INDO W H
WHITE( Φ0.5)
UV LAMP WHITE( Φ0.5)
BLUE(Φ 1)

WH I T E( Φ 0 .5 )

BL U E ( Φ 0.5 )
BL A C K( Φ 0 .5 )
RE D ( Φ 0 .5 )
RE D ( Φ 0 .5 )
BLACK( Φ0.5) 1 3
U V BA LL A S T

BLACK( Φ1)
BLACK( Φ0.5) 2
WHITE( Φ0.5)
BLUE( Φ1)

W in do w
UV LAMP WHITE( Φ0.5) H
L

GND

V CC
H T UV Ctrl

G ND
BLACK( Φ0.5) BLUE( Φ0.5)
L IGH T B A L LA S T

GREEN( Φ1) RED(Φ 0.5)


BLACK( Φ0.5) up

Foot
down
BLUE(Φ 0.5) BLACK( Φ0.5)
BLUE( Φ1)
BLUE( Φ0.5) 220V/ 50 Hz

ANJ IA N
MCU EIGHT CORE CABLE LCD
C C
YELLOW( Φ0.5) TRANS-IN
1
L IGH T BA L LA S T

YELLOW( Φ0.5) L
EX-FAN 12V/0 .5A
N
RED(Φ 0.5)
L
RED(Φ 0.5) 18B2-EF
N TRANS-OUT SPEAK
L

UV
N SIXTEEN CORE CABLE
SOCKET L

LIG HT
N
RED(Φ 1) FUSE3 RED(Φ 1) SPEAKER
L

SOC KE T
5A N
BLUE(Φ 1)
BROWN( Φ1.5)
红 ( Φ 1)

YELLOW GREEN (Φ 1.0)

DOW N

DOO R MOTO R
BLUE(Φ 1.5) DOWN-FAN N UP

N
L N
SOCKET
B B
YELLOW GREEN (Φ 1.0)
FUSE4
BCBS-204 wiring diagram

10A BLUE(Φ 1) BLACK( Φ1) BROWN( Φ1)


YELLOW(Φ 1) BLUE(Φ 1)
FUSE2 DOOR MOTOR
BROWN( Φ1.5)
L

W A V E FIL T ER
10A BLOWER
POWER LOCK
POWER
YELLOW GREEN ( Φ1.0)
YELLOW GREEN (Φ 1.5)
N BLUE( Φ1.5)
A A
1 2 3 4 5 6 7 8
Biological Safety Cabinet Maintenance Manual

BIOLAB SCIENTIFIC LTD.


3660 Midland Avenue, Suite 300, Toronto, Ontario M1V 0B8 Canada
Email: contact@biolabscientific.com Tel: +1 707 533 1445
Website: www.biolabscientific.com

50 www.biolabscientific.com