Programat CS: Operating Instructions
Programat CS: Operating Instructions
Programat CS: Operating Instructions
Operating Instructions f
Valid as o n 3.0
o f t w a r e Versio
S
2
Table of Contents
Views of the Furnace, List of Parts 4
1. Introduction / Signs and Symbols 8
1.1 Preface
1.2 Introduction
1.3 Notes regarding the Operating Instructions
1.4 Notes on the different voltage versions
2. Safety First 9
2.1 Indications
2.2 Health and safety instructions
3. Product Description 12
3.1 Components
3.2 Hazardous areas and safety equipment
3.3 Functional description
3.4 Accessories
3.5 Indication/Contraindication
6. Practical Use 23
6.1 Switching on/off
6.2 Firing using a standard program
6.3 Firing using an individual program
6.4 Other options and special features of the furnace
6.5 Programming
8. What if ... 27
8.1 Error messages
8.2 Additional error messages
8.3 Technical malfunctions
8.4 Repair
9. Product Specifications 30
9.1 Delivery form
9.2 Technical data
9.3 Acceptable operating conditions
9.4 Acceptable transportation and storage conditions
10. Appendix 31
10.1 Program table
10.2 Program structure
3
List of Parts
Front view:
1 Sealing surface
2 Furnace head sealing ring
3 Insulation
4 Thermocouple
5 Firing plate
6 Display
7 Frame plate
8 QTK heating muffle
9 Furnace housing
10 Keypad
11 On/Off switch
12 Heating element fuse
13 Vacuum pump fuse
14 USB-Device interface
15 Fuse holder
16 Power cord
17 Power socket
18 Vacuum pump socket
19 Rating plate
20 Thermocouple cable
21 Vacuum hose connection
23 Rubber feet
24 Firing plate holder
25 Furnace head housing
26 Thermocouple plug
27 Plug fuse
28 Heater plug
29 Heater plug socket
30 Thermocouple plug socket
32 Leaf spring
33 Air vents (base)
34 Cooling tray
35 Screw for cooling tray
36 Hood
37 Knurled screw for hood
38 Air vents furnace head
39 Air vents rear panel
40 Warnings
41 Furnace head mounting mark
42 Furnace base mounting mark
43 Furnace head mounting
44 Quartz-glass tube
46 Vacuum hose
47 Silicone rest
48 Connecting rod
49 Plug-in console
4
25
40
47
35
10 34
23
33
4 8
2 3
24
7
5
38
25
36
37
39
11 12 13 19
21
46
17
16
15
18
19
12 13
32 43
49
48
11 20 26 30 29 27 28
6
3
49
48
4
41
42
44
Control unit:
71 ESC key
72 ENTER key
78
73 START key
74 Start LED 87
75 STOP key
76 + key 81 88
77 – key
78 Settings key
79 Cursor key right
80 Cursor key left
75
81 Information key
82 Program 1
83 Program 2
80 79 77 76 71 73
84 Program 3 72
85 Program 4 74
86 Next program
87 Open furnace head
88 Close furnace head
82 83 84 85 86
7
1. Introduction / Signs and Symbols
The Operating
1.4 Notes on the different
Instructions must
be read voltage versions
The furnace is available with
different voltage versions.
– 100 V / 50–60 Hz
– 110–120 V / 50–60 Hz
– 200–240 V / 50–60 Hz
8
2. Safety First
This chapter is especially important for personnel who work with the 2.1.1
Programat CS or who have to carry out maintenance or repair work.
This chapter must be read and the corresponding instructions fol- Contraindication
lowed.
Do not carry the furnace head by
the cooling tray.
2.1 Indications
The Programat CS must only be used to fire dental ceramic materials
and it should be used for this purpose only. Other uses than the
ones stipulated, e.g. cooking of food, firing of other materials, etc.
are contraindicated. The manufacturer does not assume any liability
for damage resulting from misuse. The user is solely responsible for
any risk resulting from failure to observe these Instructions.
2.1.2
Further instructions to assure proper use of the furnace:
Contraindication
– The instructions, regulations, and notes in these Operating
Instructions must be observed. Do not carry the furnace head by
– The instructions, regulations, and notes in the material’s the cables, since the cables and
Instructions for Use must be observed. connections may be damaged.
– The furnace must be operated under the indicated environmental
and operating conditions (Chapter 9).
– The Programat CS must be properly maintained.
2.1.3
Contraindication
2.1.4
Contraindication
2.1.5
Contraindication
9
2.1.6 2.1.11
Risks of burn hazard
Contraindication
Never place objects in the firing
Never use the furnace without a
chamber by hand, since there is a
firing table. burn hazard. Always use the tongs
(accessories) supplied for this pur-
pose. Never touch the hot surface
of the furnace head, as there is a
burn hazard. Please also refer to
the warnings on the furnace.
2.1.7
Contraindication 2.1.12
Risk of crushing and burn hazard
Firing trays must not be placed in
the area surrounding the firing
Never reach under the furnace
table, since this will obstruct the
head with the hand or other parts
closing of the furnace head.
of the body during operation,
since there is a risk of crushing
and a burn hazard.
2.1.8
Contraindication
2.1.13
Foreign objects must not be Risks and dangers
placed on the furnace head or the
air vents. Make sure that no liq-
uids or other foreign objects enter This product contains ceramic
the air vents, since this may result fibres and may release fibre dust.
in an electrical shock. Do not use compressed air, or
blow, on the furnace thus
distributing the dust in the
environment and observe the
additional notes on page 11.
2.1.9
Contraindication
2.1.14
Risks and dangers
The furnace must not be operated if the quartz tube in the firing
2.1.10 chamber is damaged. There is a risk of electric shock upon contact
with the heating wire.
Contraindication
10
2.2 Health and Safety Instructions Any disruption of the protective conductor either inside
or outside the furnace or any loosening of the protective
This furnace has been designed according to EN 61010-1 and has conductor connection may lead to danger for the user in
been shipped from the manufacturer in excellent condition as far as case of malfunction. Deliberate interruptions are not
safety regulations are concerned. To maintain this condition and to tolerated.
assure risk-free operation, the user must observe the notes and
warnings contained in these Operating Instructions. Materials developing harmful gases must not be fired.
– The user must especially become familiar with the warnings and
the operating conditions to prevent injury to personnel or damage
to materials. The manufacturer is not responsible for damage Warnings regarding the dismounting of the heating muffle
resulting from misuse or failure to observe the Operating Instruc- This product contains ceramic fibres and may release fibre
tions. Warranty claims cannot be accepted in such cases. dust. Fibre dust has proved to be carcinogenic in animal
experiments. The corresponding EU Safety Data Sheet must
– Before switching on the furnace, make sure that the voltage be observed.
indicated on the rating plate complies with your local power
supply. The heat insulation of the firing chamber in the Programat CS
consists of ceramic fibres. After prolonged use of ceramic fibres at
– The power socket must be equipped with a residual current circuit
temperatures of over 900 °C (1652 °F), silicogenic substances
breaker.
(Cristobalite) may be produced. In certain cases, e.g. upon changing
– The furnace must be plugged into a socket with protected of the heating muffle, the possible resulting dust exposure may
contacts. cause irritation of the skin, eyes, and respiratory organs. Therefore,
proceed as follows when changing the heating muffle:
– Place furnace on a fire-proof table (observe local regulations, e.g.
distance to combustible substances or objects, etc.) – Make sure the corresponding staff wears long-sleeved clothing, as
well as headgear, goggles, and gloves.
– Always keep the air vents at the rear and the side of the furnace
– Place suction equipment at the source of the dust or, if not
free from obstruction.
possible, provide the staff with FFP3 facemasks or similar items.
– Do not touch any parts that become hot during the operation of – Once the procedure has been completed, any dust possibly
the furnace. There is a burn hazard! adhering to exposed skin must first be rinsed off with cold water.
Only after that should soap and warm water be used.
– Clean furnace only with a dry or slightly moist cloth. Do not use
– The corresponding work clothes should be washed separately.
any solvents! Disconnect power before cleaning.
– The furnace must be cool before it is packed for transportation Warning
purposes. The insulation on this product contains refractory ceramic fibres (RCF)
which pose a possible cancer hazard, if agitated and inhaled. May be
– Use original packaging for transportation purposes.
irritating to the skin, eyes or respiratory tract if insulation is cracked
– Before calibration, maintenance, repair, or exchange of parts, the or corrupted.
power must be disconnected if the furnace is to be opened.
California Proposition 65
– If calibration, maintenance, or repair has to be carried out with Warning: ”This product contains Refractory
the power connected and the furnace open, only qualified Ceramic Fibres, a substance known to the State of California to
personnel, who are familiar with the risks and dangers, may cause cancer.”
perform these procedures.
Disposal:
– After maintenance, the required safety tests (high voltage
The furnaces must not be disposed of in the normal
resistance, protective conductor, etc.) have to be carried out.
domestic waste. Please correctly dispose of old
– Ensure that only fuses of the indicated type and rated current are furnaces according to the corresponding EU council
used. directive. Information on the correct disposal may
also be found on your local Ivoclar Vivadent home-
– If it is assumed that safe operation is no longer possible, the page.
power must be disconnected to avoid accidental operation.
Safe operation is no longer possible if
– the furnace is visibly damaged
– the furnace does not work
– the furnace has been stored under
unfavourable conditions over an extended period of time
– Use only original spare parts.
– The temperature range for faultless operation is +5 °C to +40 °C
(+41 °F to +104 °F).
– If the furnace has been stored at very low temperatures or high
atmospheric humidity the head has to be opened and the unit
dried or left to adjust to room temperature for approx.
1 hour (do not connect the power yet).
– The furnace has been tested for use at altitudes of up to 2000 m
(6562 ft) above sea level.
– The furnace may only be used indoors.
11
3. Product Description
3.4 Accessories
(not part of the delivery form)
– Automatic Temperature Checking Set 2 (ATK 2)
– Programat Accessories Set (large and small firing trays, firing
tongs, Temperature Checking Set)
– Vacuum pump
3.5 Indication/Contraindication
Indication
– Glaze and crystallization firings (chairside area)
Contraindication
– The Programat CS is not suitable as ceramic furnace for dental
laboratories
12
4. Installation and Initial Start-Up
4.1 Unpacking and checking the contents 4.2 Selecting the location
The packaging provides the following advantages: Place the furnace on a flat table using the rubber feet. Make sure
– Reusable packaging that the furnace is not placed in the immediate vicinity of heaters or
– Closing mechanism with integrated transporta-tion grips other sources of heat. Make sure that air may properly circulate
– Ideal protection by Styrofoam inserts between the wall and the furnace.
– Easy handling / optimum unpacking
– The packaging may be used in several ways (modules) Also ensure that there is enough space between the furnace and the
user, as the furnace releases heat during the opening of the furnace
Remove the furnace components from their head.
packaging and place it on a suitable table. Please observe the
instructions on the outer packaging. The furnace should neither be placed nor operated in areas
where there is an explosion hazard.
There are no special transportation grips on the furnace. Support the
bottom of the furnace to carry it.
4.3 Assembly
Make sure the voltage indicated on the rating plate (19) complies
with the local power supply. If this is not the case, the furnace must
not be connected.
35
35
47
47
13
Place the cooling tray (34) on the frame plate (7).
Make sure that the cooling tray (34) is correctly
positioned on the frame plate (7).
34
35 47
Step 2: 5
Placing the firing plate (5)
Place the firing plate (5) on the firing plate holder
(24).
24
14
Step 3:
Mounting the furnace head
The complete furnace head is best mounted with
the rear panel of the furnace pointing towards the
user. Lift the furnace head with both hands (see
picture) and carefully position it on the furnace
head mounting.
48
41
42
Step 4:
Connections
Connect the cables of the furnace head with the
furnace base. Proceed as follows:
– Insert the thermocouple plug (26) (make sure
that the polarity of the plug is correct)
– Insert the heater plug (28)
28
26 27
30
15
Secure the heater plug (28) with the plug fuse (27)
by turning it until the heater plug (28) has been
secured.
27
28
Step 5:
Mounting the hood (36)
Once all cables are properly connected to the
furnace base, the hood (36) can be mounted.
Subsequently, secure the hood with the knurled
screw (37).
Step 6:
Establishing additional connections
Power connection
Please make sure that the voltage indicated on the
rating plate complies with the local power supply.
Connect the power cord (16) with the power 21
socket (17) of the furnace. 46
12 13
16
4.4 Removing the furnace head
Before the hood (36) is removed, the furnace has
to be switched off and the power cord (16)
disconnected from the power socket (17).
93
4
2
1 Status bar
2 Firing hours
3 SW version
4 Current supply voltage
17
5. Operation and configuration
82 83 84 85 86
18
5.4 Program structure 5.5 Adjustable parameters and possible value ranges
The furnace offers three types of programs:
Symbol Parameter Value range Value range
a. Standard programs for Ivoclar Vivadent m
aterials
P Program number P 1–20
b. Free programs
c. Test programs B Stand-by temperature 100–700 °C 212–1292 °F
S Closing time (min : sec) 00:18–30:00
a) Standard programs for Ivoclar Vivadent t➚ (*) Temperature increase rate 30–140 °C/min 54–252 °F/min
materials (see chapter 10.1) T Holding temperature 100–1200 °C 212–2192 °F
– IPS e.max CAD
H Holding time (min : sec) 00.01–60:00
– IPS Empress CAD
V1 Vacuum on 0 or 1–1200 °C 0 or 34–2192 °F
b) Free programs V2 Vacuum off 0 or 1–1200 °C 0 or 34–2192 °F
All free programs are available as equivalent t➚ (*) Temperature increase rate
and thus fully-fledged programs. All the t2➚ Second stage 30–140 °C/min 54–252 °F/min
parameters can be individually set for each T Holding temperature
program.
Second stage 100–1200 °C 212–2192 °F
H Holding time
When the furnace is delivered ex
works, the standard programs already Second stage (min : sec) 00.01–60:00
contain the recommended material V1 (V1 2) Vacuum on
parameter settings. Second stage 0 or 1–1200 °C 0 or 34–2192 °F
However, the parameters can be changed and V2 (V2 2) Vacuum off
overwritten at any time, if required, if the Second stage 0 or 1–1200 °C 0 or 34–2192 °F
programs are to be used for other purposes. L Long-term cooling 0 or 50–1200 °C 0 or 122–2192 °F
Therefore, the 50 programs are also available as tL Cooling temperature rate 0 or 1–50 0 or 32–90
free programs.
(*) 100 V Version: 140°C/min (252°F/min)
The programs are designed in such a way that
they can be either used as conventional, one-stage
programs or as two-stage programs, if required.
The mode can be changed via the symbol (one- or
two-stage program) by using the + or – key. Automatic plausibility check
The furnace is equipped with an automatic plausibility check function. The parameters
c) Test programs (e.g. T 960 but L 1000) are checked upon each program start. In case of contradictory
Various test programs are available. Please refer parameter combinations, the program stops automatically and the respective error
to chapter 5.5 Settings / configuration and number is indicated.
information.
Examples of firings
– Typical glaze firing
19
5.6 Settings / test programs Settings Display Short description
By pressing the „Settings“ key (78), the Settings will be displayed
(indication of the last selected Settings). Renaming Enables the renaming
of the currently
selected program. *
Enables the renaming
of the materials.
The cursor keys (79, 80) are used to toggle between the possible Time The time can be
settings. This screen can be exited with the ESC key (71) or a entered by using the
Program key (82, 83, …). – / + keys
20
Settings Display Short description 5.5.2 Information
By pressing the ‚Information‘ key (81), you can access the informa-
“Ivoclar Vivadent Only after entry of tion display (the latest information selected is being displayed). You
optimized the STD code. can browse through the various information using the cursor keys
temperature control Enables the deactiva- (79, 80). You can exit this display by hitting ESC (71) or one of the
function“ tion of the “Ivoclar program keys (82, 83, …).
Vivadent optimized
temperature control
function“
Settings Display
Selection of factory With this setting, all Serial number of the
settings values and para- Serial number furnace
meters can be reset
to the factory
settings.
Attention:
All individual pro-
grams which have Software version
been created and
saved will be deleted
with this function.
Furnace head
firing hours
Important information
The user code (6725) is required for some settings.
Operating hours
furnace
Operating hours
vacuum pump
Latest start of
calibration program
Calibration value
Calibration value 660°C and 963°C
Short description
21
5.7 Symbols in the display 5.8 Explanation of beeper tunes
“One-stage Indicates that a conventional, Beeper lasting for approx. Self-test has been completed
program“ one-stage program is used 2 seconds with unchangeable
“Self-test signal” to indicate
the completion of the self-test.
“Two-stage Indicates that a specific,
program“ two-stage program is used. Beeper lasting for approx. The firing process has been
The bold line indicates the 5 seconds with the beeper completed and the furnace
values for the first stage signal set by the user. head is open *.
22
6. Practical Use
The operating procedure for the Programat CS will be explained with 6.3 Firing using an individual program
the help of two examples: one standard and one individual
program. Step 1:
Select a free program.
Step 2:
6.1 Switching on/off Select the parameter list via the
setting ‚Programming‘ (7/20) and
Put ON/OFF switch (11) on position “I”. The furnace conducts an
change the program parameters
automatic self-test, which will be indicated in the beginning.
using the ‚–/+‘ keys. After that,
Subsequently, a status bar shows that the self-test is being
change back to the Stand-by screen
conducted. Make sure that the furnace is not manipulated during
by pressing ESC or the Program
this time.
key.
6.1.1 Stand-by mode Step 3:
After successful completion of the self-test, the Stand-by mode is Now, open the furnace head using
shown in the display. the ‚Open furnace head‘ key (87)
and position the firing tray with the
object in the furnace. Press the
START key (73) and the program is
started. The program can only be
started if the furnace head is open.
The sequence can be observed on
the firing curve screen.
6.4.4 Renaming
e)
The keyboard can be selected via the setting ‚Renaming‘ (8/20,
The following information is always displayed: 9/20), provided that the currently selected program is not write-
a) Status of program protected. The desired letters can be selected using the cursor keys
b) Remaining time (circulating). The letter is selected with the ENTER key. Individual
c) Current temperature letters can be deleted using the ‚delete‘ key (arrow symbol). The
d) Program number changes are saved using the ‚save‘ key (disk symbol) or the ESC key.
e) Program name With this action, the keyboard display is also exited.
f) Status bar
This key enables the shift to lower case letters, numbers/
symbols and back to upper case
6.2 Firing using a standard program letters.
Step 1:
Select with the program keys the desired program (1 to 50)
according to the firing table.
Step 2:
Open the furnace head with the “Open furnace head” key (87) and
place the object on the firing tray in the furnace.
Step 3:
Press the START key (73). The program can only be started with the
furnace head open. The process is indicated in the firing curve
display. 23
6.5 Programming
One-stage program *
The parameter list can be selected via the setting ‚Programming‘
(7/19). In this list, all the program parameters are shown. The desired
parameter can be selected using the cursor keys. The value can be
changed using the ‚-/+‘ keys. This screen can be exited with the ESC
or the Program keys.
Two-stage program
If a two-stage program is used, the parameters are displayed for the
first and the second stage in a separate screen. Switching between
the two screens is carried out using the symbol ‚Change program
stage indication‘.
* The programs are protected with a code. If there will be necessary changes, you will be
informed about the corresponding code.
24
7. Maintenance, Cleaning, and Diagnosis
This chapter describes the user maintenance and cleaning procedures This furnace has been developed for typical use in dental
for the Programat CS. All the other tasks must be performed by laboratories. If the product is used in a production
qualified service personnel at a certified Ivoclar Vivadent Service enterprise, for industrial applications, and for continuous
Center. use, premature ageing of the expendable parts has to be
expected.
The expendable parts are as follows:
7.1 Monitoring and maintenance – Heating muffle
The time for these maintenance procedures depends on the – Insulation material
frequency of use and the working habits of the users. For that
Expendable parts are not covered by the warranty. Please also
reason, the recommended times are only approximations.
observe the shorter service and maintenance intervals.
7.2 Cleaning
The furnace may only be cleaned when it is cool, since
there is a burn hazard. Do not use any cleaning solutions.
25
7.3 Test programs 7.6 Temperature calibration
Press the the key ‘Settings’ and select the desired test program by 1. Select the calibration program.
using the Cursor-keys.
2. Remove the firing plate from
Vacuum pump test program: the furnace using the furnace
With this program, the vacuum performance of the furnace vacuum tongues and place it on the
system can be automatically tested. For that purpose, the achieved cooling tray.
(minimum) pressure in mbar is measured and indicated. If the
pressure value is below 80 mbar, the vacuum performance of the
system is adequate.
Heating muffle test program
With this heater test, the quality of the heating muffle can be auto- 3. Carefully grip the upper part
matically tested (duration approximately 7 minutes). The heating of the ATK 2 using the furnace
muffle test should only be conducted with the firing chamber empty, tongs (Caution: Fracture risk of
since any additional item in the firing chamber (e.g. firing tray) the ceramic) and insert it into
influences the result. Please run the heating muffle test program the holes designated for this
immediately after switching on the furnace and before starting the purpose until it snaps into
actual firing procedures. If the furnace is too hot, an incorrect place. The orientation of the
heating muffle quality will be indicated. If the quality of the heating calibration sample (left or right)
elements drops below 50%, changing the heating element is is not important.
strongly recommended. 4. If necessary, use the furnace
Keypad test program tongs to apply slight pressure
With every keystroke a short beep sounds. The keypad test is ended to the center of the calibration
by pressing the ESC key. base until the calibration
sample clicks into place.
Cleaning program Observe the corresponding
The heating muffle is ‚cleaned‘ using the cleaning program (duration markings.
approximately 17 minutes).
5. Start the calibration program
7.5 Dehumidification program 7. Replace the firing plate using the furnace tongs.
The condensation of water in the insulation of the firing chamber 8. Close the furnace head and select a firing program.
and the vacuum pump will result in a lower vacuum and thus
impaired firing results. For that reason, the furnace head should be 9. The ATK 2 can only be used once. Use a new calibration set for
kept closed when the furnace is switched off, in order to prevent the the next calibration procedure.
absorption of humidity. Start the dehumidification program if
required (humidity in the insulation).
26
8. What if …
3 Entry V2x <= V1x 9 Enter a logical value for the vacuum-on temperature Vx1 or the vacuum-off temperature Vx2
4 Entry V2x > Tx + 1°C 10 Change either the vacuum values or the holding time T
5 Entry Incorrect values for V1x, V2x 11 Enter a logical value for V1x, V2x
6 System Current temperature after 13 Excess temperature! Program aborted, furnace head opens to allow the furnace to cool down.
Start > Tx + 50 °C *, **
8 Entry T2 < T1 16 Enter a lower value for T1 or a higher value for T2.
9 System Power failure > 10 s during 17 A firing program in progress was interrupted for more than 10 s. The program cannot be continued!
a firing program in progress
10 Entry T1 > V12 18 Enter a lower value for T1 or a higher value for V12
11 Entry vV set, but V2 is 19 Pre-vacuum activated! V2 must be higher than B.
missing or invalid
12 System Error in the heating system 20 no Check the heater fuse. If the fuse is O.K., contact your local Ivoclar Vivadent Service Center.
**
13 System Heating muffle very old 23 The heating muffle is very old. It is recommended to replace it. After the error message has been
acknowledged, a firing program may still be started.
14 System Heating muffle defective 24 The condition of the muffle is so poor that is has to be replaced immediately.
16 Entry T is > B + 200 °C at the 26 Firing chamber too hot to start a firing program.
start of a firing program
17 System Furnace head cannot 27 The furnace head cannot be moved to the final position. It might be blocked by an external mechanical
be initialized **, *** source! If this is not the case, please contact your local Ivoclar Vivadent Service Center!
18 System The furnace head does not 28 The furnace head does not open/close correctly. The furnace head was manually moved or is obstructed.
reach the target position ** The furnace head must only be moved using the keys intended for this purpose!
21 System Necessary vacuum (xxxmbar) 33 The vacuum cannot be established. Check the seal of the firing chamber, vacuum hose, vacuum pump,
is not reached within 1 min. pump fuse.
33 Entry HV > H (H2) 110 Enter a lower value for HV or a higher value for H (H2)
35 Entry „Share of the holding time 120 Activate the vacuum during the holding time Tx or deactivate HV.
with vacuum“ is activated,
but Vx2 does not
correspond to Tx or Tx+1
38 System Brief power failure during a 702 A firing program in progress was interrupted by a brief power failure. The program is continued!
firing program in progress
46 System The vacuum does not 802 No vacuum increase could be measured. Check the following points: Is the firing chamber tight (no
increase (self-test) contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?
Is the fuse F1 o.k.?
75 System ATK2 calibration: 1302 Error during calibration. Sample may not be correctly inserted. Try again with a new sample and make sure
Pre-heating to 963 °C ** the sample makes ampl contact.
79 Note Calibration reminder 1310 Some time has passed since the last calibration procedure. Calibrate the furnace soon.
80 Note Dehumidification reminder 1312 Some time has passed since the last dehumidification procedure. Dehumidificat the furnace soon.
25, 29
32
43, 44, 45, 46, 47, 48
54, 56
103, 107
700, 701, 703, 704, 705, 706, 707
800
1010, 1011, 1012, 1013, 1014, 1015, 1016
1024, 1025, 1026, 1028
1202, 1203, 1204, 1205, 1206, 1207
1300, 1301, 1303, 1304, 1305
1400, 1401, 1402
1500
28
8.3 Technical malfunctions
These malfunctions may occur without an error message being displayed.
* If there are any questions, please contact the Ivoclar Vivadent After Sales Service.
8.4 Repair
Repairs may only be carried out by a certified Ivoclar
Vivadent Service Center. Please refer to the addresses on
the last page of these Operating Instructions.
If repairs during the warranty period are not carried out by a certified
Ivoclar Vivadent Service Center, the warranty will expire immediately.
Please also refer to the corresponding warranty regulations.
29
9. Product Specifications
Recommended accessories
– Programat Accessories Set
9.4 Acceptable transportation and storage conditions
– Automatic Temperature Checking Set 2 (ATK 2) Acceptable temperature range -20 to +65 °C (-4 °F to +149 °F)
– Vacuum Pump VP3 easy Acceptable humidity range Max. 80 % relative humidity
Acceptable ambient pressure 500 mbar to 1060 mbar
110–120 V / 50–60 Hz
200–240 V / 50–60 Hz
Overvoltage category II
Contamination level 2
Tolerated voltage fluctuations +/- 10%
Max. power consumption 11 A at 100 V
12 A at 110–120 V
8.5 A at 200–240 V
Acceptable data for vacuum pump of other manufacturers
Max. output: 250 W / max. leakage current 0.75 mA
Final vacuum: < 50 mbar
Use only tested pumps
Electrical fuses: 100 V / 110–120 V:
250 V / T 15 A (heating circuit)
250 V / T 5 A (vacuum pump)
200–240 V:
250 V / T 8 A ( heating circuit)
250 V / T 3.15 A ( vacuum pump)
Dimensions of electrical fuses 110–120 V:
Diameter 6.3 x 32 mm
200–240 V:
Diameter 5 x 20 mm
Dimensions of the closed furnace
Depth: 430 mm / width: 305 mm / 410 mm (with Cooling tray)
Height: 320 mm
Usable size of the firing chamber Diameter 80 mm
Height 48 mm
Max. firing temperature 1200 °C (2192 °F)
Weight Furnace base: 12.0 kg
Furnace head: 4.5 kg
Safety information
The CS complies with the following guidelines:
– IEC 1010-1/EN 61010, Part 1
– UL and cUL standards
Radio protection / electromagnetic compatibility EMC tested
30
10. Appendix
Important Information
You also will find the program tables in the Internet:
www.ivoclarvivadent.com/downloadcenter
You can download the program tables as PDF files from this side.
Please note, that the version of the program tables should
correspond to the Software version which is used on your furnace.
Program Description
31
Ivoclar Vivadent – worldwide
Ivoclar Vivadent AG Ivoclar Vivadent SAS Ivoclar Vivadent S.A. de C.V. Ivoclar Vivadent S.L.U.
Bendererstrasse 2 B.P. 118 Av. Insurgentes Sur No. 863, c/ Emilio Muñoz Nº 15
FL-9494 Schaan F-74410 Saint-Jorioz Piso 14, Col. Napoles Entrada c/ Albarracin
Liechtenstein France 03810 México, D.F. E-28037 Madrid
Tel. +423 235 35 35 Tel. +33 450 88 64 00 México Spain
Fax +423 235 33 60 Fax +33 450 68 91 52 Tel. +52 (55) 50 62 10 00 Tel. + 34 91 375 78 20
www.ivoclarvivadent.com www.ivoclarvivadent.fr Fax +52 (55) 50 62 10 29 Fax + 34 91 375 78 38
www.ivoclarvivadent.com.mx www.ivoclarvivadent.es
Ivoclar Vivadent Pty. Ltd. Ivoclar Vivadent GmbH
1 – 5 Overseas Drive Dr. Adolf-Schneider-Str. 2 Ivoclar Vivadent Ltd. Ivoclar Vivadent AB
P.O. Box 367 D-73479 Ellwangen, Jagst 12 Omega St, Rosedale Dalvägen 14
Noble Park, Vic. 3174 Germany PO Box 303011 North Harbour S-169 56 Solna
Australia Tel. +49 (0) 79 61 / 8 89-0 Auckland 0751 Sweden
Tel. +61 3 979 595 99 Fax +49 (0) 79 61 / 63 26 New Zealand Tel. +46 (0) 8 514 93 930
Fax +61 3 979 596 45 www.ivoclarvivadent.de Tel. +64 9 914 99 99 Fax +46 (0) 8 514 93 940
www.ivoclarvivadent.com.au Fax +64 9 914 99 90 www.ivoclarvivadent.se
Ivoclar Vivadent Marketing (India) www.ivoclarvivadent.co.nz
Ivoclar Vivadent Ltda. Pvt. Ltd. Ivoclar Vivadent Liaison Office
Alameda Caiapós, 723 503/504 Raheja Plaza Ivoclar Vivadent Polska Sp. z o.o. : Tesvikiye Mahallesi
Centro Empresarial Tamboré 15 B Shah Industrial Estate Al. Jana Pawla II 78 Sakayik Sokak
CEP 06460-110 Barueri – SP Veera Desai Road, Andheri (West) 00-175 Warszawa Nisantas’ Plaza No:38/2
Brazil Mumbai, 400 053 Poland Kat:5 Daire:24
Tel. +55 11 2424 7400 India Tel. +48 22 635 54 96 34021 Sisli – Istanbul
Fax +55 11 3466 0840 Tel. +91 (22) 2673 0302 Fax +48 22 635 54 69 Turkey
www.ivoclarvivadent.com.br Fax +91 (22) 2673 0301 www.ivoclarvivadent.pl Tel. +90 212 343 08 02
www.ivoclarvivadent.in Fax +90 212 343 08 42
Ivoclar Vivadent Inc. Ivoclar Vivadent Marketing Ltd. www.ivoclarvivadent.com
1-6600 Dixie Road Ivoclar Vivadent s.r.l. Prospekt Andropova 18 korp. 6/
Mississauga, Ontario Via Isonzo 67/69 office 10-06 Ivoclar Vivadent Limited
L5T 2Y2 40033 Casalecchio di Reno (BO) 115432 Moscow Ground Floor Compass Building
Canada Italy Russia Feldspar Close
Tel. +1 905 670 8499 Tel. +39 051 611 35 55 Tel. +7 499 418-03-00 Warrens Business Park
Fax +1 905 670 3102 Fax +39 051 611 35 65 Fax +7 499 418-03-10 Enderby
www.ivoclarvivadent.us www.ivoclarvivadent.it www.ivoclarvivadent.ru Leicester LE19 4SE
United Kingdom
Ivoclar Vivadent Shanghai) Ivoclar Vivadent K.K. Ivoclar Vivadent Marketing Ltd. Tel. +44 116 284 78 80
Trading Co., Ltd. 1-28-24-4F Hongo Qlaya Main St. Fax +44 116 284 78 81
2/F Building 1, 881 Wuding Road, Bunkyo-ku Siricon Building No.14, 2nd Floor www.ivoclarvivadent.co.uk
Jing An District Tokyo 113-0033 Office No. 204
200040 Shanghai Japan P.O. Box 300146 Ivoclar Vivadent, Inc.
China Tel. +81 3 6903 3535 Riyadh 11372 175 Pineview Drive
Tel. +86 21 6032 1657 Fax +81 3 5844 3657 Saudi Arabia Amherst, N.Y. 14228
Fax +86 21 6176 0968 www.ivoclarvivadent.jp Tel. +966 1 293 83 45 USA
www.ivoclarvivadent.com Fax +966 1 293 83 44 Tel. +1 800 533 6825
Ivoclar Vivadent Ltd. www.ivoclarvivadent.com Fax +1 716 691 2285
Ivoclar Vivadent Marketing Ltd. 12F W-Tower, 1303-37 www.ivoclarvivadent.us
Calle 134 No. 7-B-83, Of. 520 Seocho-dong, Seocho-gu, Ivoclar Vivadent Pte. Ltd.
Bogotá Seoul 137-855 171 Chin Swee Road
Colombia Republic of Korea #02-01 San Centre
Tel. +57 1 627 33 99 Tel. +82 (2) 536 0714 Singapore 169877
Fax +57 1 633 16 63 Fax +82 (2) 596 0155 Tel. +65 6535 6775
www.ivoclarvivadent.co www.ivoclarvivadent.co.kr Fax +65 6535 4991
www.ivoclarvivadent.com
Version: 4
Date information prepared: 08/2012
Valid as of Software V3.0
This apparatus has been developed solely for use in dentistry. Start-up and
operation should be carried out strictly according to the Operating Instructions.
Liability cannot be accepted for damages resulting from misuse or failure to
observe the Instructions. The user is solely responsible for testing the apparatus
for its suitability for any purpose not explicitly stated in the Instructions.
Descriptions and data constitute no warranty of attributes.
Printed in Austria
© Ivoclar Vivadent AG, Schaan/Liechtenstein
604261/03.09.2011/e