STANDARD OPERATING PROCEDURE


Name of the Instrument: Polarimeter (Make: Rajdhani Scientific Instruments)
S.O.P. No.: ADI/SAM-001 Issue No.: 4 Page No.: 1 of 2
Supercedes: Issue No.3, dated
Date of Issue: March 2006 Date of Review: March 2009
November 2003
1.0 Purpose: This SOP describes the procedure for operating Polarimeter. (Make: Rajdhani Scientific
Instruments)

2.0 Responsibility: Analyst

3.0 Precautions:
 Optical elements must be kept clean and in exact alignment. The match point should be close
to the normal zero mark.
 The field must be evenly illuminated and it should be supplemented by a filtering system
capable of isolating the D line from sodium light.
 Specific attention should be paid to temperature control of the solution and of the polarimeter.
 Polarimeter tubes should be filled in such a way so as to avoid air bubbles. Particular care is
necessary for semi-micro and micro tubes.
 For tubes with removable end plates fitted with gaskets and caps, tighten the end plates only
enough to ensure leak proof seal between the end plate and body of the tube.
 Liquids and solutions of solids must be clear.
 Differences between the observed and corrected optical rotation calculated as either specific
optical rotation or optical rotation should not be more than ¼th of the range specified in the
monograph of the substance.

4.0 Procedure:
 Sample solution as per monograph is prepared and filled in Polarimeter tube.
 The tube is inserted in the slot provided ensuring that there is no air bubble in the path of the
light and the analyzer is rotated until two halves of field shows equal intensity of light. Adjust
the reading with fine adjustment knob such that the cross wires exactly coincide with one of
the readings on scale.
Reading (optical rotation) x 100
Specific optical rotation = ----------------------------------------------
L x C
(where, L = length of tube, C = Concentration of the substance in percentage w/v)

Prepared By Approved By Issued By

(Ms. Sabina Golatkar) (Mr. Hemant Patil) (Mr. P. Gopinath)


(Senior Executive- Q. A.) (Sr. Manager - Corporate Q.A) (Gen. Manager-Tech. Services)
CHARAK CONFIDENTIAL

STANDARD OPERATING PROCEDURE


Name of the Instrument: Polarimeter
S.O.P. No: ADI/SAM-001 (Make: Rajdhani Scientific Page No: 2 of 2
Instruments)
5.0 Calibration:
The polarimeter can be calibrated by measuring the optical rotation in a 2 dm table at 25° using
the concentrations of dry sucrose indicated in the table below

Concentration g/ 100 ml Angle of rotation (+) at 25°


10.0 13.33
20.0 26.61
30.0 39.86
40.0 53.06
50.0 66.23

6.0 Frequency of calibration: Monthly

HISTORY OF CHANGES
Issue No. Date Remark
1. March 2000 Original Issue
2. November 2001 Format revised
3. November 2003 Reviewed & issued without any change
4. March 2006 Reviewed & issued without any change

Prepared By Approved By Issued By

(Ms. Sabina Golatkar) (Mr. Hemant Patil) (Mr. P. Gopinath)


(Senior Executive- Q. A.) (Sr. Manager - Corporate Q.A) (Gen. Manager-Tech. Services)