Intensive and Critical Care Nursing (2009) 25, 21—30
available at www.sciencedirect.com
journal homepage: www.elsevier.com/iccn
REVIEW
Endotracheal suctioning of the adult intubated
patient—–What is the evidence?
Carsten M. Pedersen a, Mette Rosendahl-Nielsen b,
Jeanette Hjermind c, Ingrid Egerod d,∗
a
Intensive Care Unit, Frederiksberg Hospital, DK, Denmark
Intensive Care Unit, Rigshospitalet, DK, Denmark
c
Bispebjerg Hospital, DK, Denmark
d
The University Hospitals Centre for Nursing and Care Research, Rigshospitalet, Department 7331,
Blegdamsvej 9, DK-2100 Copenhagen O, Denmark
b
Accepted 28 May 2008
KEYWORDS
Endotracheal
suctioning;
Evidence-based
practice;
Intensive care;
Nursing;
Suction
Summary Intubated patients may be unable to adequately cough up secretions. Endotracheal
suctioning is therefore important in order to reduce the risk of consolidation and atelectasis
that may lead to inadequate ventilation. The suction procedure is associated with complications and risks including bleeding, infection, atelectasis, hypoxemia, cardiovascular instability,
elevated intracranial pressure, and may also cause lesions in the tracheal mucosa. The aim of
this article was to review the available literature regarding endotracheal suctioning of adult
intubated intensive care patients and to provide evidence-based recommendations The major
recommendations are suctioning only when necessary, using a suction catheter occluding less
than half the lumen of the endotracheal tube, using the lowest possible suction pressure, inserting the catheter no further than carina, suctioning no longer than 15 s, performing continuous
rather than intermittent suctioning, avoiding saline lavage, providing hyperoxygenation before
and after the suction procedure, providing hyperinflation combined with hyperoxygenation on
a non-routine basis, always using aseptic technique, and using either closed or open suction
systems.
© 2008 Elsevier Ltd. All rights reserved.
Introduction
Intubated patients may be unable to adequately cough up
secretions. Based on the work by Day et al. (2002), endotra-
Corresponding author. Tel.: +45 35457365; fax: +45 35457399.
E-mail addresses: carstenmp@hotmail.com (C.M. Pedersen), mette.rosendahl@get2net.dk (M. Rosendahl-Nielsen),
jh46@bbh.regionh.dk (J. Hjermind), ie@ucsf.dk (I. Egerod).
∗
cheal suctioning (ET suctioning) is an important activity in
reducing the risk of consolidation and atelectasis that may
lead to inadequate ventilation. ET suctioning is defined as:
‘‘A component of bronchial hygiene therapy and mechanical
ventilation and involves the mechanical aspiration of pulmonary secretions from a patient with an artificial airway
in place’’ (AARC, 1993 p. 500). The procedure is associated
with complications and risks including bleeding, infection,
atelectasis, hypoxemia, cardiovascular instability, elevated
intracranial pressure, and may also cause lesions in the
0964-3397/$ — see front matter © 2008 Elsevier Ltd. All rights reserved.
doi:10.1016/j.iccn.2008.05.004
22
tracheal mucosa (Branson et al., 1993; Thomson et al.,
2000).
Endotracheal suctioning is described by patients as
painful and uncomfortable, and may result in a choking
sensation initiating a violent cough, and also cause an
unpleasant sensation that the lungs are actually being suctioned into the catheter (Gjengedal, 1994; Patak et al.,
2004). Despite the discomfort, patients report that the procedure is necessary and subsequently eases their breathing
(Bergbom-Engberg, 1989; Jablonski, 1994). The discomfort
associated with suctioning does not diminish during a course
of mechanical ventilation; on the contrary, it may worsen
(Bergbom-Engberg, 1989).
The apparent discomfort and the potential complications
of suctioning may intimidate inexperienced nurses. As Day
et al. (2002) stated practice is not always based on current
research recommendations, which may lead to inconsistent
practice among nurses, affecting the patient experience
(Sole et al., 2003; Celik and Elbas, 2000).
Due to the frequency and risk associated with endotracheal suctioning, there is a need to examine clinical practice
critically, and to identify clinical research to guide practice.
The aim of this article was to review the available literature
regarding the endotracheal suctioning of adult intubated
intensive care patients and to provide evidence-based recommendations.
Methods
Design
A search of literature was performed in Medline, Cinahl
and Cochrane Library. The following key words were used:
intensive care, critical care, suction, endotracheal suctioning. The initial strategy was to use broad terms in order
to increase the sensitivity of the search. The search was
then limited to adult patients (≥19 years old) including
Danish, Swedish, Norwegian and English language articles
from 1995 to 2008. References cited in bibliographies of
past reports were included. In order to retrieve the primary
source for some of the widely accepted recommendations,
it was necessary to include articles and studies dating back
to 1962. In order to examine the patient perspective, qualitative research on patient’s experiences during ventilator
treatment was included, but very few data were retrieved.
Studies involving elevated intracranial pressure (ICP) were
excluded, because ICP monitoring requires a specialised protocol. A total of 77 papers were included in the final review,
4 studies describing patient experiences of suctioning, 19
literature reviews, 2 meta-analyses and 52 clinical trials.
Data analysis
Papers were categorised and reviewed according to the following subjects: ET suctioning performed as a routine or
when necessary (prn), appropriate catheter size, suction
pressure, depth of insertion, duration of procedure, intermittent or continuous suction pressure, pre-oxygenation,
hyperinflation (bagging), instillation of normal saline,
hygiene guidelines, and open versus closed endotracheal
suction systems. These themes were chosen based on clini-
C.M. Pedersen et al.
Figure 1
Literature appraisal instrument.
cal practice and are all essential in preventing or minimising
physiological complications associated with endotracheal
suctioning.
The literature was reviewed using an appraisal instrument (Fig. 1). The studies were reviewed critically, to
establish the current best research evidence to recommend
a specific procedure (Sackett et al., 1996). The current best
research evidence was defined as patient centred research
into the efficacy and safety of treatment and care, in which
new evidence invalidates previously accepted regimens and
replaces them with new ones. The evidence included quantitative as well as qualitative studies that were judged by
how well the study question related to the research method,
and the consistency of results, discussion and conclusion
(Rolfe, 1999; Rycroft-Malone et al., 2004). The focus of this
review was on the knowledge base of the recommendations
rather than the level of evidence refering exclusively to the
bio-medical research hierarchy.
Results
Suctioning performed routinely or when necessary
Traditionally, endotracheal suctioning was performed routinely every 1—2 h to ensure the removal of secretions,
and to prevent pneumonia and endotracheal tube (ET-tube)
occlusion (Glass and Grap, 1995). Due to the considerable
risk and adverse effects, it is now recommended, that ET
suctioning should be performed only when necessary (Young,
1984; Branson et al., 1993; Odell et al., 1993; Glass and
Grap, 1995; Day et al., 2002). A prospective randomised
study of 383 patients demonstrated that prn suctioning was
associated with fewer adverse effects. There was no significant variation in ICU mortality, incidence of pulmonary
infections, duration of intubation, or ICU-stay in patients
that were suctioned routinely or only when necessary (Van
de Leur et al., 2003a,b).
A study including 14 patients, showed no significant variation in complications (i.e. decreased saturation, increased
airway pressure, changes in heart rate, heart rhythm and
Endotracheal suctioning of the adult intubated patient
mean arterial pressure (MAP), ET-tube occlusion and infection rate) between routine and prn suctioning (Wood,
1998a). The clinical implications of this study were limited,
due to a small sample size, and similar suction frequency in
the two groups studied.
Suctioning only when necessary requires the nurse to be
able to determine the patient’s need for suctioning. Indications for suctioning are cough, visible or audible secretions,
coarse or absent respiratory sounds, increased airway pressure, desaturation, or increased respiratory work (Wood,
1998b).
A clinical study including 66 patients has shown that
the risk of overlooking residual secretions decreased when
the assessment was done using a stethoscope. Retained
secretions were best identified from the appearance of a
saw-tooth pattern on the flow-volume loop, on the monitor screen of the ventilator (Guglielminotti et al., 2000).
Flow-volume loops are ventilator graphics that provide information on leaks and resistance. The clinical relevance of
the study is uncertain as retained secretions were defined
as >0.5 ml obtained on subsequent suctioning. However, this
study indicates that clinical assessment can be inadequate,
and nurses may not have the necessary skills to carry out the
assessment (Wood, 1998a).
A number of studies show that the internal lumen of
endotracheal tubes decreases significantly after a few days
of intubation, sometimes only after 8 h, due to formation
of biofilm and the adherence of secretions on the surface
(Tenaillon, 1990; Glass et al., 1999; Shah and Kollef, 2004;
Chirag and Kollef, 2004). A minimum frequency of ET suctioning should be considered due to the risk of not detecting
retained secretions and partial tube occlusion, as long as the
patient is unable to adequately clear the secretions.
What is the recommendation on routine versus ‘‘prn
suctioning’’?
Based on clinical experience and a number of studies, it is
recommended that endotracheal suctioning should be performed only when necessary. There is no evidence to support
maximum suction intervals, but it may be advisable to perform suctioning at least every 8 h to reduce the risk of partial
occlusion of the ET-tube and the accumulation of secretions.
Auscultation is recommended as part of clinical assessment.
Appropriate catheter size
There is consensus in the literature that suction catheters
should be as small as possible, yet large enough to facilitate secretion removal. When the catheter size is small, air
may enter the lungs around the catheter during suctioning.
This prevents a sudden drop in functional residual capacity
Figure 2
23
(FRC), and thus reduces the risk of atelectasis (Glass and
Grap, 1995; Wood, 1998b). One study suggests that nurses
may choose larger catheters than recommended and necessary (Day et al., 2001). This could be due to experience,
habit, lack of evidence, or the convention that a larger
catheter may be necessary in the case of thick secretions
(Wood, 1998b).
It is generally recommended that the maximum external
diameter of the suction catheter should be half the size of
the internal diameter of the ET-tube (Demers and Saklad,
1973; Branson et al., 1993; McKelvie, 1998; Odell et al.,
1993; Glass and Grap, 1995; Wood, 1998b; Day et al., 2002).
This recommendation can be traced to a single paper, in
which it is demonstrated by a mathematical model, that a
larger catheter increases the negative pressure applied on
the lungs (Rosen and Hillard, 1962). There is, however, no
evidence to support this exact ratio. It is also stated that
the effectiveness of the suction catheter depends on the
size; the larger the catheter, the greater the effectiveness
of secretion removal.
The widely accepted formula for calculating catheter size
is: Suction catheter size [Fr] = (ET-tube size [mm] − 2) × 2.
Since the lumen of the ET-tube is calculated as:
× r2 × length, a doubling of the diameter of the ET-tube
leads to a quadrupling of the tube lumen. Thus, it would be
more logical to recommend that the suction catheter should
occlude less than half the internal lumen, rather than half
the diameter of the ET-tube. A lung model test supports
this, as it shows that when a suction catheter occludes less
than half the lumen of the ET-tube, the negative pressure in
the lungs is minimised, because the space in the tube that
allows air to pass to the lungs during suctioning correlates
to the tube lumen (Tiffin et al., 1990). According to this, the
following formula for calculating catheter size is suggested:
Suction catheter size [Fr] = (ET-tube size [mm] − 1) × 2 (see
Fig. 2). This formula may be controversial, as it leads to a
slightly larger catheter size than generally recommended,
but it is perhaps more efficient for the removal of thick
secretions.
What is the recommendation on suction catheter size?
Based on clinical experience, calculation of ET-tube lumen
and a lung model test, it is recommended that the suction
catheter should occlude less than half of the internal lumen
of the endotracheal tube.
Level of suction pressure
A number of studies recommend the use of the lowest
possible suction pressure to reduce the risk of atelectasis, hypoxia and damage to the tracheal mucosa (Branson
Calculation of suction catheter lumen size.
24
et al., 1993; Glass and Grap, 1995; Wood, 1998a,b; Day
et al., 2002). At the same time suction pressure must
be sufficient to clear the secretions. The recommendation of maximum suction pressure is exclusively based
on experience; no clinical studies support an exact limit.
A meta-analysis demonstrates that a suction pressure of
80—120 mmHg was used in more than 50% of the studies of
endotracheal suctioning examined (Oh and Seo, 2003). The
negative pressure that is actually applied to the lungs during suctioning cannot be reliably assessed on the manometer
dial of the suctioning equipment; it depends on the suction catheter-ET-tube ratio, the duration of the procedure,
and the volume and viscosity of the secretions (Rosen and
Hillard, 1962). One study has shown that there was no difference in induced tracheobroncheal lesions using negative
pressure of 145 mmHg versus 500 mmHg, when the suction
catheter occludes less than half the internal lumen of the
endotracheal tube. One study on dogs has demonstrated
any application of negative pressure with endotracheal suction produces significant damage to tracheal tissue (Czarnik
et al., 1991). It is concluded that the degree of negative
pressure applied to the lungs depends on catheter size and
duration of suctioning. As high pressure is more effective in
removing secretions, a pressure of 200—300 mmHg may be
applied when using the appropriate catheter size (Tenaillon,
1990).
What is the recommendation on suction pressure?
Based on clinical experience, it is recommended using
the lowest possible suction pressure during endotracheal
suctioning, usually 80—120 mmHg. A negative pressure of
200 mmHg may be applied provided that the appropriate
suction catheter size is used.
Depth of catheter insertion
A number of papers recommend that the suction catheter
should be inserted to the carina and retracted 1—2 cm before
applying suction (Wood, 1998b; Celik and Elbas, 2000; Day
et al., 2002). A single study supports this practice as there is
no difference in duration of intubation, ICU-stay, or mortality using minimally invasive suctioning in which the suction
catheter is inserted to the length of the ET-tube only, or
deep endotracheal suctioning. The minimally invasive procedure is associated with fewer adverse effects, but this is
not exclusively related to the suction depth. It is indicated
that deep suctioning may be necessary in patients with large
amounts of secretions in the lower airways (Van de Leur et
al., 2003a,b). In ETS, secretions are removed only from the
central airways. The movement of secretions depends upon
mucociliary clearance, level of sedation, and the patient’s
ability to cough. Deep endotracheal suctioning may cause a
greater negative pressure applied to the lungs, due to occlusion of more than half the lumen of the bronchial branch
(Rosen and Hillard, 1962). This may lead to bradycardia (Day
et al., 2002).
C.M. Pedersen et al.
which the suction catheter is inserted to the length of the ETtube only. The suction catheter is inserted to the carina and
then retracted 1—2 cm before suctioning is performed, or
the length of the suction catheter is estimated by measuring
an identical endotracheal tube.
Duration of suctioning procedure
The duration of the suctioning procedure affects the severity of adverse effects. It may be difficult to determine
exactly which complications can be attributed to the duration of suctioning. The maximum duration of suctioning
is inadequately documented. One study has shown that
the correct catheter size and duration of less than 15 s
reduced the drop in FRC and PaO2 (partial pressure of oxygen) (Tenaillon, 1990). A clinical trial conducted on dogs
has shown that the reduction of PaO2 increases after 5 s of
suctioning (George, 1983). In a comparative study of complications associated with suctioning, a maximum of 10 s
is recommended for the suctioning itself and 15 s for the
entire procedure (Celik and Elbas, 2000; Wood, 1998b; Day
et al., 2002). Suction should not be applied during catheter
insertion.
What is the recommendation on the duration of the
suction procedure?
Based on clinical experience, it is recommended that the
suctioning procedure should last no longer than 15 s.
Intermittent suctioning technique
The argument in favour of using intermittent suctioning is
that it minimises inward folding (invagination) of the tracheal mucosal membrane into the suction catheter. There
are no clinical studies supporting this argument, or studies
that show a difference in complications between intermittent and continuous suctioning. The literature does not
provide an answer to which should be applied in the individual suctioning procedure (Wood, 1998b; Glass and Grap,
1995). If the suction catheter is inserted no further than
2 cm above the carina, the risk of invagination is assumed
to be minimal. One study has shown that continuous as well
as intermittent application of negative pressure with endotracheal suction produces significant damage to tracheal
tissue on dogs (Czarnik et al., 1991). When the closed suctioning system is used, risk of alveolar collapse increases
with the use of intermittent suctioning (Stenqvist et al.,
2001).
What is the recommendation on intermittent versus
continuous suctioning?
Based on clinical experience and a single study, the recommendation is continuous rather than intermittent suctioning
during the individual suction procedure.
Saline lavage prior to endotracheal suctioning
What is the recommendation on catheter
insertion depth?
Based on clinical experience and a single study, it is recommended using minimally invasive endotracheal suctioning in
Conventional wisdom has it that saline lavage prior to endotracheal suctioning eases catheter insertion, induces cough,
facilitates secretion removal, and ultimately improves the
Endotracheal suctioning of the adult intubated patient
patient’s oxygenation. The procedure is conducted by
injecting 2—5 ml sterile normal saline into the ET-tube prior
to suctioning (Wood, 1998b; Blackwood, 1999; Celik and
Kanan, 2006). A number of studies have advised against routine use of normal saline instillation (Odell et al., 1993;
McKelvie, 1998; Wood, 1998b; Blackwood, 1999; Day et
al., 2002; Celik and Kanan, 2006). Most studies have been
unable to document that normal saline increases aspiration of secretions (Demers and Saklad, 1973; Bostick and
Wendelgass, 1987; Ackerman, 1993), while a few studies
have suggested a positive effect on the amount of secretions
obtained after the instillation of normal saline (Reynolds et
al., 1990; Isea et al., 1993).
Several studies argue against routine instillation of normal saline because saline fails to dilute tenacious secretions
and lubricate the tube (Odell et al., 1993; Wood, 1998b;
McKelvie, 1998; Blackwood, 1999; Day et al., 2002). Studies
of the effect of normal saline instillation on patient’s vital
signs show no significant reduction in minute volume, airway
pressure, pulse and blood pressure (Bostick and Wendelgass,
1987; Gray et al., 1990; Reynolds et al., 1990; Ackerman,
1993; Akgul and Akyolcu, 2002; Celik and Kanan, 2006).
Some studies show that normal saline does not affect oxygen saturation, while others show reduced saturation for
as much as five minutes post suctioning (Ackerman, 1993;
Kinloch, 1999; Akgul and Akyolcu, 2002).
Instillation of normal saline prior to suctioning may dislodge bacteria from the biofilm inside the endotracheal
tube. A study performed on used ET-tubes shows that bacterial release is greater with saline lavage than the insertion
of the suction catheter (Hagler and Traver, 1994). Whether
this increases the infection rate in intubated patients is
uncertain, but a study by Hanley et al. (1978) documents
that only 18.7% of the saline is regained during suctioning.
Patient discomfort during suctioning is considerable, but
the discomfort is not affected by the instillation of normal
saline (Jablonski, 1994; O’Neal et al., 2001). One small study
suggested that the sensation of dyspnoea increased in older
patients after instillation of normal saline (O’Neal et al.,
2001).
What is the recommendation on saline lavage?
Based on clinical studies, routine instillation of normal saline
prior to endotracheal suctioning is not recommended. No
reliable positive effects in terms of secretion removal, saturation, or ventilation have been demonstrated. The only
reliable effect of the procedure is cough stimulation, but
due to the increased risk of infections and patient discomfort, this should not be an indication.
Hyperoxygenation prior to endotracheal suctioning
Hyperoxygenation is performed by increasing the intake of
oxygen immediately prior to suctioning (3—6 ventilations),
and when appropriate, after endotracheal suctioning. Preoxygenation is performed to minimise hypoxaemia and
related complications induced by suctioning (Branson et al.,
1993; Thomson et al., 2000). It is concluded in meta-analysis
that pre-oxygenation with 100% oxygen, reduces the occurrence of suction-induced hypoxaemia by 32%. Combining pre
25
and post-oxygenation, reduced the occurrence of hypoxia
by 49%. Compared with no intervention, the combination
of hyperoxygenation and hyperinflation reduced the occurrence of hypoxemia by 55% (Oh and Seo, 2003).
The most frequently used oxygen setting when performing pre-oxygenation is FiO2 of 1.00. Ventilation with 100%
oxygen for longer periods of time may lead to absorption
atelectasis, and thus reduce pulmonary volume. In a study
of short-term hyperoxygenation, this adverse effect was not
evident (Fernandez et al., 2004). A clinical trial of patients
with chronic obstructive pulmonary disease (COPD) showed
that pre-oxygenation with 20% above the baseline FiO2 , was
sufficient to prevent hypoxemia (Rogge et al., 1989). The
study did not provide adequate documentation for recommending pre-oxygenation with 20% above baseline FiO2 due
to small sample size.
What is the recommendation on hyperoxygenation prior
to suctioning?
Based on clinical controlled trials, pre-oxygenation by the
delivery of 100% oxygen for at least 30 s prior to and after the
suctioning procedure is recommended to prevent decrease
in oxygen saturation.
Hyperinflation (manual/mechanical) prior to
endotracheal suctioning
Hyperinflation is performed by inflating the patient’s lungs,
manually with a ventilation bag (bagging), or via the
mechanical ventilator. A volume of 1.5× baseline tidal
volume is the most common for hyperinflation prior to suctioning. The procedure is assumed to improve the patient’s
oxygenation capacity by recruiting pulmonary volume and
loosening secretions (Day et al., 2002). Hyperinflation is
associated with the risk of barotrauma, cardiovascular
instability and increased intracranial pressure (Day et al.,
2002). The effects of hyperinflation are usually studied
in combination with hyperoxygenation. In a meta-analysis
of interventions to prevent suction-induced hypoxemia, it
was found that the combination of hyperoxygenation and
hyperinflation reduced the occurrence of hypoxemia by 55%
compared to no intervention (Oh and Seo, 2003). As the
effect of short-term hyperoxygenation is significantly positive in adult patients, it is difficult to differentiate the effect
of hyperinflation. In one study the interventions hyperoxygenation, hyperinflation, or a combination of the two were
compared. Hyperinflation alone caused a decrease in oxygen
saturation, while the most positive effect on saturation was
found, when hyperinflation was used in combination with
hyperoxygenation. Some of the patients experienced dyspnoea when hyperinflated (Lookinland and Appel, 1991). In
addition, patients also report increased discomfort during
manual ventilation compared with mechanical ventilation
(Jablonski, 1994).
The use of pulmonary recruitment immediately after
open endotracheal suctioning resulted in earlier return
to baseline values for pulmonary volume and oxygen saturation, in a single study. Pulmonary recruitment was
performed on patients with acute lung injury (ALI) and acute
respiratory distress syndrome (ARDS), by applying an inspiration pressure of 45 cm H2 O to the lungs for 20 s (Dyhr
26
et al., 2003). This procedure is not comparable with traditional hyperinflation, but similar results from recruitment,
were found in other studies (Almgren, 2005; Maggiore et al.,
2003).
Nurses perform manual ventilation with considerable
individual variation. Manual ventilation with insufficient
tidal volume and a peak airway pressure of up to 96 cm H2 O
were performed (Clapham et al., 1995; Robson, 1998). In
a study of 100 nurses performing manual hyperinflation by
a ventilation bag, it was shown that only one-third were
able to deliver a tidal volume greater than the one delivered by the ventilator. The mean volume delivered was 17%
lower during manual ventilation, and some patients were
hypo-ventilated compared with the volume delivered by the
ventilator (Glass et al., 1993). It should be noted that hyperinflation prior to suctioning and pulmonary recruitment are
different procedures.
What is the recommendation on hyperinflation prior to
suctioning?
Based on clinical controlled trials, hyperinflation combined
with hyperoxygenation prior to suctioning is recommended
on a non-routine basis, as it prevents the decrease in oxygen saturation. Hyperinflation used alone results in a larger
decrease in oxygen saturation. Hyperinflation should be
delivered by the ventilator, in order to control tidal volume
and inspiration peak pressure. Routine use of hyperinflation
is not recommended due to the risk of barotrauma from
large volumes, high peak pressures and patient discomfort.
Recruitment manoeuvres following endotracheal suctioning
in patients with ALI and ARDS is recommended by individual
order only.
Hygiene guidelines for open endotracheal
suctioning system
Ventilator-treated patients are at high risk of acquiring nosocomial infections, which increase mortality and morbidity,
ICU-stay, and costs. Intubation and ET suctioning impede the
effect of normal defence mechanisms (Khatib et al., 1999;
Combes et al., 2000). It is highly recommended to maintain aseptic technique, including hand washing and use of
gloves, because ET suctioning is an invasive procedure that
may lead to contamination of the lower airways (Thomson
et al., 2000; Day et al., 2002). When using an open suction system, a new sterile disposable suction catheter for
every insertion in the tube is widely recommended (Day et
al., 2002; Celik and Elbas, 2000). In a single clinical controlled trial, no significant difference in infection rate from
reuse of the suction catheter for 24 h in the same patient
was shown (Scoble et al., 2001). The clinical relevance of
the study can be questioned, as the risk of contamination of
the surroundings is considerable when the suction catheter
is handled and stored between suctioning procedures. After
suctioning, the catheter is discarded and the suction hose is
flushed with tap water from a canister. It is recommended to
change the canister every 8 h (Statens Serum Institut, 2001).
When ET suctioning is performed with an open suction system, there is considerable risk of contaminating the hands
of the staff. To protect reciprocal contamination between
patient and staff, it is widely recommended to use dispos-
C.M. Pedersen et al.
able gloves as standard precaution (Danish Standards, 2001;
DoH, 2007). Aseptic technique is also necessary when using
closed suction systems because condensation from the suction system may splash on the staff (Blackwood and Webb,
1998). According to universal precautions the eyes should
be protected by goggles whenever splashing is a possibility, but no suction guidelines include goggles as standard
procedure.
What is the recommendation on infection control in
relation to suctioning?
Based on increased risk of infection, the use of aseptic technique is recommended. When using open suction system, a
sterile suction catheter must be used, for each insertion in
the endotracheal tube. To protect the staff, gloves should
be used with both open and closed suction systems. Hand
washing should always be carried out before and after the
procedure.
Closed versus open suction systems
The ‘‘closed suction system’’ (CSS) consists of a reusable sterile suction catheter protected by a transparent
sleeve that prevents contact between the catheter and the
environment. CSS is connected to the ET-tube as an integrated part of the mechanical ventilator system and allows
ET suctioning without disconnecting the ventilator, which
minimises the decrease in pulmonary volume caused by disconnection of the positive end-expiratory pressure (PEEP).
Closed suction systems have been regarded as less effective for secretion removal than ‘‘open suction systems’’
(OSS) (Johnson et al., 1994; Blackwood, 1998). A single
study supports this assumption, although the suction system used is only a semi-closed system, where suctioning
is performed via the swivel-adapter (Lascoki et al., 2006).
In another study there was no significant difference in the
amount of secretions removed with CSS compared with
OSS (Witmer et al., 1991). A bench test showed that the
quantity of secretions removed with CSS was significantly
lower than with OSS, when using same catheter size and
suction pressure, as the high inspiratory flow from the
ventilator ‘‘pushes’’ the secretions away from the suction catheter and further down the lungs (Lindgren et al.,
2004).
Some clinical trials conducted on patients with ALI show
that the use of closed systems minimises the loss of pulmonary volume in volume-controlled ventilated patients.
The decrease in pulmonary volume and oxygen saturation
was significantly greater with CSS than OSS. As the loss of
pulmonary volume is critical in patients with ALI, CSS is recommended in patients with impaired oxygenation and risk of
alveolar collapse (Johnson et al., 1994; Cereda et al., 2001;
Maggiore et al., 2003).
Another clinical trial conducted on patients with mild
to moderate lung injury showed the same effect on pulmonary volume when suctioning was performed via the
swivel-adapter (semi-closed system) and when CSS was used
(Fernandez et al., 2004). No difference in suction-induced
saturation decrease was demonstrated between CSS and OSS
in a clinical trial conducted on pigs on pressure control
(Almgren et al., 2004).
Endotracheal suctioning of the adult intubated patient
In two clinical trials CSS was shown to minimise hypertension and arrhythmias induced by endotracheal suctioning.
OSS and CSS resulted in an increase in heart rate, but 30 s
after OSS suctioning patients had significantly lower heart
rate compared to CSS (Cereda et al., 2001; Johnson et al.,
1994).
A test performed on a lung model demonstrates that CSS
used with volume-controlled ventilation resulted in extreme
negative airway pressure if the suction flow exceeded the
flow output of the ventilator. An increase in auto-PEEP was
found, due to resistance to expiration when the suction
catheter was inserted in the endotracheal tube. Based on
this, CSS in connection with volume-controlled and pressureregulated control modes is not recommended (Stenqvist et
al., 2001).
A number of papers discuss the risk of ventilatorassociated pneumonia (VAP) with CSS (Day et al., 2002). Two
trials demonstrated increased bacterial colonization, but an
increased incidence of VAP was not shown (Deppe et al.,
1990; Topeli et al., 2004). Other controlled trials show that
in terms of VAP, there is no significant difference between
the two suction systems (Johnson et al., 1994; Kollef et al.,
1997; Combes et al., 2000; Zeitoun et al., 2003; Lorente et
al., 2005). The findings are supported by an evidence-based
guideline on VAP, and a meta-analysis (Dodek et al., 2004;
Jongerden et al., 2007). In a small clinical controlled trial,
it was demonstrated that the incidence of lung to stomach
contamination was more frequent with OSS than with CSS.
Five cases of VAP were reported with OSS, and none with
CSS (Rabitsch et al., 2004).
One non-randomised crossover study deals with environmental bacterial contamination after ET suctioning. In OSS
bacterial contamination could not be detected for either
OSS or CSS in a radius of 3 m from the ET-tube (Cobley et
al., 1991). It is argued that the use of a CSS reduces the
exposure of the staff to contamination from the patient’s
airways (Cobley et al., 1991; Freytag et al., 2003). Increased
bacterial growth may be the result if CSS are not completely
cleaned between suctions (Blackwood and Webb, 1998). It
is shown in a single clinical trial, that within 24 h of use,
microbiological growth (bacteria/yeast) is detectable in 89%
of CSS, and increased with longer term use (Freytag et al.,
2003). Clinical controlled trials conclude, however, that the
frequency of VAP does not increase, even if the same suction
system is used for 48 h or even longer (Kollef et al., 1997;
Darvas and Hawkins, 2003).
The cost effectiveness of OSS and CSS depends on the
price of the suction catheters and duration of use (Kollef
et al., 1997). Most publications profess that CSS are more
expensive than OSS, but one states that OSS are the more
expensive if the patient is suctioned more than 16 times a
day (Johnson et al., 1994). If CSS is applied for four days
without a change, it is less expensive than OSS (Maggiore,
2006).
The effectiveness of closed suction systems needs to
be investigated further, based on the clinical experience
and the referenced bench test. Clinical trial examining the
effect of CSS in non-sedated patients and patients ventilated
by other ventilation methods than volume-controlled are
needed. The cost effectiveness of CSS depends on, whether
it is safe to use for en extended period. This should be
further investigated.
27
What is the recommendation on closed versus open
suction systems?
Based on clinical controlled trials, closed and open suction
systems are both recommended. There is little evidence to
support that one system is superior to the other in terms
of oxygen saturation, cardiovascular instability, secretion
removal, incidence of VAP, environmental contamination,
and cost.
Discussion
The literature review has demonstrated a paucity of good
clinical trials covering the many aspects of the suctioning
procedure. The widely accepted recommendations appear
to be based on clinical experience and fairly limited documentation. When searching for the original sources for
recommendations on the timing of suctioning, catheter size,
pressure, insertion depth, duration, and oxygenation, most
studies are relatively old, based on small sample sizes, or
based on animal studies. As a consequence, there is room
for improvement in the establishment of exact recommendations for clinical practice. The major recommendations
seen in Table 1 may be used to guide nurses to provide safe
and consistent pulmonary care, while reducing the complications and risks associated with suctioning.
In the question of open versus closed endotracheal suctioning there appears to be more available evidence based
on newer, larger and more substantial clinical controlled trials. More accurate recommendations for clinical practice
are still needed, because most trials were conducted on
heavily sedated and relaxed patients on volume-controlled
ventilation, which is a thing of the past. Contemporary
practice is determined by clinical protocols and lighter sedation together with more sophisticated modes of ventilation
(Dellinger et al., 2008).
A limitation to this study is the lack of strong evidence
on the subject. In keeping with Sackett et al. (1996) we
have attempted to provide the ‘best current evidence’ in
this review. At this time the recommendations rely on the
interpretation of a combination of experience and research.
It is our hope that more reliable external evidence will be
available in the future.
Nurses need to be aware of the current best research
evidence in order to make informed decisions. This literature review is intended as a step in the direction towards
safer practice. The next step is the establishment of an
expert group that will update the guideline every couple of
years. When using a guideline, the nurse should always consider the external evidence in relation to the needs of the
individual patient. The literature review has demonstrated
the need for further investigations ranging from randomised
controlled trials to qualitative studies of comfort and meaning. Finally, it may be stressed that the absence of evidence
is not evidence of absence.
Contribution/authorship
CMP, MRN, JH contributed in the conception and design of
the study; CMP, MRN, JH, IE in drafting the article; and the
final approval was made by IE.
28
Table 1
C.M. Pedersen et al.
Major recommendations for endotracheal suctioning
Research question
Recommendation
What is the recommendation on
routine versus ‘‘prn suctioning’’?
What is the recommendation on
suction catheter size?
What is the recommendation on
suction pressure?
What is the recommendation on
catheter insertion depth?
What is the recommendation on the
duration of the suction procedure?
What is the recommendation on
intermittent versus continuous
suctioning?
What is the recommendation on
saline lavage?
What is the recommendation on
hyperoxygenation prior to suctioning?
It is recommended that endotracheal suctioning should be performed only
when necessary
It is recommended that the suction catheter should occlude less than half
of the lumen of the endotracheal tube
It is recommended using the lowest possible suction pressure during
endotracheal suctioning, usually 80—120 mmHg
It is recommended using minimally invasive endotracheal suctioning in
which the suction catheter is inserted to the length of the ET-tube only
It is recommended that the suctioning procedure should last no longer than
15 s
The recommendation is continuous rather than intermittent suctioning
during the individual suction procedure
What is the recommendation on
hyperinflation prior to suctioning?
What is the recommendation on
infection control in relation to
suctioning?
What is the recommendation on
closed versus open suction systems?
Routine instillation of normal saline prior to endotracheal suctioning is not
recommended
Pre-oxygenation by the delivery of 100% oxygen for at least 30 s prior to
and after the suctioning procedure is recommended to prevent decrease in
oxygen saturation
Hyperinflation combined with hyperoxygenation prior to suctioning is
recommended on a non-routine basis
Based on increased risk of infection, the use of aseptic technique is
recommended
Closed and open suction systems are both recommended
References
Ackerman MH. The effect of saline lavage prior to suctioning. Am J
Crit Care 1993;2(4):326—30.
Akgul S, Akyolcu N. Effects of normal saline on endotracheal suctioning. J Clin Nurs 2002;11(6):826—30.
Almgren B, Wickerts CJ, Heinonen E, Hogman M. Side effects of
endotracheal suction in pressure- and volume-controlled ventilation. Chest 2004;125(3):1077—80.
Almgren B. Endotracheal suction a reopened problem [dissertation].Uppsala: Acta Universitatis Upsaliensis; 2005.
American Association of Respiratory Care. AARC clinical practice
guidelines: endotracheal suctioning of mechanically ventilated adults and children with artificial airways. Respir Care
1993;38(4):500—4.
Bergbom-Engberg I. Patients’ experiences of respirator treatment:
a retrospective study of the influence of medical and nursing
care factors on recall, experience of discomforts, and feelings
of security or insecurity [dissertation]. Göteborg: University of
Göteborg; 1989.
Blackwood B. The practice and perception of intensive care
staff using the closed suctioning system. J Adv Nurs
1998;28(5):1020—9.
Blackwood B, Webb CH. Closed tracheal suctioning systems and
infection control in the intensive care unit. J Hosp Infect
1998;39(4):315—21.
Blackwood B. Normal saline instillation with endotracheal suctioning: primum non nocere (first do no harm). J Adv Nurs
1999;29(4):928—34.
Bostick J, Wendelgass ST. Normal saline instillation as part of the
suctioning procedure: effects on PaO2 and amount of secretions.
Heart Lung 1987;16(5):532—7.
Branson RD, Cambell RS, Chatburn RL, Covington J. AARC clinical
practice guideline: endotracheal suctioning of mechanically ven-
tilated adults and children with artificial airways. Respir Care
1993;38(5):500—4.
Celik SS, Elbas NO. The standard of suction for patients
undergoing endotracheal intubation. Intensive Crit Care Nurs
2000;16(3):191—8.
Celik SA, Kanan N. A current conflict: use of isotonic sodium cloride
solution on endotracheal suctioning in critically ill patients.
Dimens Crit Care Nurs 2006;25(1):11—4.
Cereda M, Villa F, Colombo E, Greco G, Nacoti M, Pesenti A. Closed
system endotracheal suctioning maintains lung volume during
volume-controlled mechanical ventilation. Intensive Care Med
2001;27(4):648—54.
Chirag S, Kollef MH. Endotracheal tube intraluminal loss
among mechanically ventilated patients. Crit Care Med
2004;31(1):120—5.
Clapham L, Harrison J, Raybould T. A multidisciplinary audit of
manual hyperinflation technique (sigh breath) in a neurosurgical intensive care unit. Intensive Crit Care Nurs 1995;11(5):265—
71.
Cobley M, Atkins M, Jones PL. Environmental contamination during tracheal suction: a comparison of disposable conventional
catheters with a multiple-use closed system device. Anaesthesia
1991;46(11):957—61.
Combes P, Fauvage B, Oleyer C. Nosocomial pneumonia in mechanically ventilated patients, a prospective randomised evaluation
of the Stericath closed suctioning system. Intensive Care Med
2000;26(7):878—82.
Czarnik RE, Stone KS, Everhart Jr CC, Preusser BA. Differential
effects of continuous versus intermittent suction on tracheal
tissue. Heart Lung 1991;20(2):144—51.
Danish Standards. Infection control in the health care sector. Part
2: Requirements for hand washing practice for the prevention of
nokocomial infection. DS 2451-2. 1st ed. Charlottenlund: Danish
Standards; 2001.
Endotracheal suctioning of the adult intubated patient
Darvas JA, Hawkins LG. The closed tracheal suction catheter: 24 h
or 48 h change? Aust Crit Care 2003;16(3):86—92.
Day T, Wainwright SP, Wilson-Barnett J. An evaluation of a teaching
intervention to improve the practice of endotracheal suctioning
in intensive care units. J Clin Nurs 2001;10(5):682—96.
Day T, Farnell S, Wilson-Barnett J. Suctioning: a review of
current research recommendations. Intensive Crit Care Nurs
2002;18(2):79—89.
Demers RR, Saklad M. Minimizing the harmful effects of mechanical
aspiration. Heart Lung 1973;2(4):542—5.
Department of Health, UK. Saving Lives, reducing infection, delivering clean and safe care. DoH 2007. http://www.dh.gov.uk.
Deppe SA, Kelly JW, Thoi LL, Chudy JH, Longfield RN, Ducey JP, et al.
Incidence of colonization, nosocomial pneumonia, and mortality
in critically ill patients using a Trach Care closed-suction system
versus an open-suction system: prospective, randomized study.
Crit Care Med 1990;18(12):1389—93.
Dellinger RP, Levy MM, Carlet JM, Bion J, Parker MM, Jaeschke R, et
al. Surviving sepsis campaign: international guidelines for management of severe sepsis and septic shock: 2008. Crit Care Med
2008;36(1):296—327.
Dodek P, Keenan S, Cook D, Heyland D, Jacka M, Hand L, et
al. Evidence-based clinical practice guideline for the prevention of ventilator-associated pneumonia. Ann Intern Med
2004;141(4):305—13.
Dyhr T, Bonde J, Larsson A. Lung recruitment manoeuvres are
effective in regaining lung volume and oxygenation after open
endotracheal suctioning in acute respiratory distress syndrome.
Crit Care 2003;7(1):55—62.
Fernandez MD, Piacentini E, Blanch L, Fernandez R. Changes in lung
volume with three systems of endotracheal suctioning with and
without pre-oxygenation in patients with mild-to-moderate lung
failure. Intensive Care Med 2004;30(12):2210—5.
Freytag CC, Thies FL, Konig W, Welte T. Prolonged application of
closed in-line suction catheters increases microbial colonization
of the lower respiratory tract and bacterial growth on catheter
surface. Infection 2003;31(1):31—7.
George RB. Duration of suctioning: an important variable. Respir
Care 1983;28(4):457—9.
Gjengedal E. Understanding a world of critical illness: a phenomenological study of the experiences of respirator patients and their
caregivers [dissertation]. Bergen: Department of Public Health
and Primary Health Care, Division for Nursing Science, University
of Bergen; 1994.
Glass C, Grap MJ, Corley MC, Wallace D. Nurses’ ability to
achieve hyperinflation and hyperoxygenation with a manual
resuscitation bag during endotracheal suctioning. Heart Lung
1993;22(2):158—65.
Glass C, Grap MJ. Ten tips for safer suctioning. Am J Nurs
1995;95(5):51—3.
Glass C, Grap MJ, Sessler CN. Endotracheal tube narrowing after
closed-system suctioning: prevalence and risk factors. Am J Crit
Care 1999;8(2):93—100.
Gray JE, MacIntyre NR, Kronenberger WG. The effects of bolus
normal-saline instillation in conjunction with endotracheal suctioning. Respir Care 1990;35(8):785—90.
Guglielminotti J, Alzieu M, Maury E, Guidet B, Offenstadt G. Bedside
detection of retained tracheobronchial secretions in patients
receiving mechanical ventilation: is it time for tracheal suctioning? Chest 2000;118(4):1095—9.
Hagler DA, Traver GA. Endotracheal saline and suction catheters:
sources of lower airway contamination. Am J Crit Care
1994;3(6):444—7.
Hanley MV, Ruud T, Butler J. What happens to endotracheal instillations? [abstract]. Am Rev Respir Dis 1978;117(4 II):124.
Isea JO, Poyant D, O’Donnell C, Faling LJ, Karlinsky J,
Celli BR. Controlled trial of a continuous irrigation suction
catheter vs. conventional intermittent suction catheter in
29
clearing bronchial secretions from ventilated patients. Chest
1993;103(4):1227—30.
Jablonski RS. The experience of being mechanically ventilated. Qual
Health Res 1994;4(2):186—207.
Johnson KL, Kearney PA, Johnson SB, Niblett JB, MacMillan
NL, McClain RE. Closed versus open endotracheal suctioning: costs and physiologic consequences. Crit Care Med
1994;22(4):658—66.
Jongerden IP, Rovers MM, Grypdonck MH, Bonten MJ. Open and
closed endotracheal suction systems in mechanically ventilated intensive care patients: a meta-analysis. Crit Care Med
2007;35(1):260—70.
Khatib M, Jamaleddine G, Abdallah A, Ibrahim Y. Hand washing and
use of gloves while managing patients receiving mechanical ventilation in the ICU. Chest 1999;116(1):172—5.
Kinloch D. Instillation of normal saline during endotracheal suctioning: effects on mixed venous oxygen saturation. Am J Crit Care
1999;8(4):231—40.
Kollef MH, Prentice D, Shapiro SD, Fraser VJ, Silver P, Trovillion E,
et al. Mechanical ventilation with or without daily changes of
in-line suction catheters. Am J Respir Crit Care Med 1997;156(2
Pt 1):466—72.
Lascoki S, Qin L, Sartorius A, Fouillat D, Remerand F, Rouby J. Open
and closed-circuit endotracheal suctioning in acute lung injury.
Anesthesiology 2006;104(1):39—47.
Lindgren S, Almgren B, Hogman M, Lethvall S, Houltz E, Lundin
S, et al. Effectiveness and side effects of closed and open
suctioning: an experimental evaluation. Intensive Care Med
2004;30(8):1630—7.
Lookinland S, Appel PL. Hemodynamic and oxygen transport changes
following endotracheal suctioning in trauma patients. Nurs Res
1991;40(3):133—9.
Lorente L, Lecuona M, Martin MM, Garcia C, Mora ML, Sierra A.
Ventilator-associated pneumonia using a closed versus an open
tracheal suction system. Crit Care Med 2005;33(1):115—9.
Maggiore SM, Lellouche F, Pigeot J, Taille S, Deye N, Durrmeyer X,
et al. Prevention of endotracheal suctioning-induced alveolar
derecruitment in acute lung injury. Am J Respir Crit Care Med
2003;167(9):1215—24.
Maggiore SM. Endotracheal suctioning, ventilator-associated pneumonia, and costs: open or closed issue? Intensive Care Med
2006;32(4):244—8.
McKelvie S. Endotracheal suctioning. Nurs Crit Care
1998;3(5):244—8.
O’Neal PV, Grap MJ, Thomson C, Dudley W. Level of dyspnoea experienced in mechanically ventilated adults with and without saline
instillation prior to endotracheal suctioning. Intensive Crit Care
Nurs 2001;17(6):356—63.
Odell A, Allder A, Bayne R, Everett C, Scott S, Still B, et al. Endotracheal suction for adult, non-head-injured, patients. A review
of the literature. Intensive Crit Care Nurs 1993;9(4):274—8.
Oh H, Seo W. A meta-analysis of the effects of various interventions
in preventing endotracheal suction-induced hypoxemia. J Clin
Nurs 2003;12(6):912—24.
Patak L, Gawlinski A, Fung I, Doering L, Berg J. Patients’ reports
of health care practitioner interventions that are related
to communication during mechanical ventilation. Heart Lung
2004;33:308—21.
Rabitsch W, Kostler WJ, Fiebiger W, Dielacher C, Losert H, Sherif
C, et al. Closed suctioning system reduces cross-contamination
between bronchial system and gastric juices. Anesth Analg
2004;99(3):886—92.
Reynolds P, Hoffman LA, Schlichtig R, Davies PA, Zullo TG. Effects of
normal saline instillation on secretion volume, dynamic compliance and oxygen saturation. Am Rev Res Dis 1990;141(4 Suppl.
2):A574.
Robson WP. To bag or not to bag? Manual hyperinflation in intensive
care. Intensive Crit Care Nurs 1998;14(5):239—43.
30
Rogge JA, Bunde L, Baun MM. Effectiveness of oxygen concentrations
of less than 100% before and after endotracheal suctioning in
patients with chronic obstructive pulmonary disease. Heart and
Lung 1989;18(January (1)):64—71.
Rolfe G. Insufficient evidence: the problems of evidence-based nursing. Nurse Educ Today 1999;19(6):433—42.
Rosen M, Hillard EK. The effects of negative pressure during tracheal
suction. Anesth Analg 1962;41:50—7.
Rycroft-Malone J, Seers K, Titchen A, Harvey G, Kitson A, McCormack B. What counts as evidence in evidence-based practice? J
Adv Nurs 2004;47(1):81—90.
Sackett DL, Rosenberg W, Gray JAM, Haynes RB. Evidence based
medicine: what it is and what it isn’t. BMJ 1996;312:71—2.
Scoble MK, Copnell B, Taylor A, Kinney S, Shann F. Effect of reusing
suction catheters on the occurrence of pneumonia in children.
Heart Lung 2001;30(3):225—33.
Shah C, Kollef MH. Endotracheal tube intraluminal volume
loss among mechanically ventilated patients. Crit Care Med
2004;32(1):120—5.
Sole ML, Byers JF, Ludy JE, Zhang Y, Banta CM, Brummel K. A multisite survey of suctioning techniques and airway management
practices. Am J Crit Care 2003;12(3):220—30.
Statens Serum Institut. Råd og anvisninger om infektionsprofylakse
i relation til ventilations-og respirationsudstyr i sundhedssektoren [engelsk oversættelse af titel]. 1. ed. Copenhagen: Central
Department for Hospital Hygiene, SSI; 2001.
Stenqvist O, Lindgren S, Karason S, Sondergaard S, Lundin S. Warning! Suctioning. A lung model evaluation of closed suctioning
systems. Acta Anaesthesiol Scand 2001;45(2):167—72.
Tenaillon A. Tracheobronchial suctions during mechanical ventilation. Update Intensive Care Emerg Med 1990;10:196—200.
C.M. Pedersen et al.
Thomson L, Morton R, Cuthbertson S. Tracheal suctioning of adults
with an artificial airway. Best Practice 2000;4(4):1—6.
Tiffin NH, Keim MR, Frewen TC. The effects of variations in
flow trough an insufflating catheter and endotracheal-tube
and suction-catheter size on test-lung pressures. Respir Care
1990;35(9):889—97.
Topeli A, Harmanci A, Cetinkaya Y, Akdeniz S, Unal S. Comparison of
the effect of closed versus open endotracheal suction systems
on the development of ventilator-associated pneumonia. J Hosp
Infect 2004;58(1):14—9.
Van de Leur JP, Zwaveling JH, Loef BG, Van der Schans CP. Endotracheal suctioning versus minimally invasive airway suctioning in
intubated patients: a prospective randomised controlled trial.
Intensive Care Med 2003a;29(3):426—32.
Van de Leur JP, Zwaveling JH, Loef BG, Van der Schans CP.
Patient recollection of airway suctioning in the ICU: routine
versus a minimally invasive procedure. Intensive Care Med
2003b;29(3):433—6.
Witmer MT, Hess D, Simmons M. An evaluation of the effectiveness of secretion removal with the Ballard closed-circuit suction
catheter. Respir Care 1991;36(8):844—8.
Wood CJ. Can nurses safely assess the need for endotracheal suction in short-term ventilated patients, instead of using routine
techniques? Intensive Crit Care Nurs 1998a;14(4):170—8.
Wood CJ. Endotracheal suctioning: a literature review. Intensive
Crit Care Nurs 1998b;14(3):124—36.
Young CS. Recommended guide lines for suction. Physiotherapy
1984;70(3):106—8.
Zeitoun SS, de Barros AL, Diccini S. A prospective, randomized study
of ventilator-associated pneumonia in patients using a closed vs.
open suction system. J Clin Nurs 2003;12(4):484—9.