Clinical guidelines and position paper
Clinically oriented device programming in bradycardia
patients: part 2 (atrioventricular blocks and neurally
mediated syncope). Proposals from AIAC (Italian
Association of Arrhythmology and Cardiac Pacing)
Pietro Palmisanoa,, Matteo Ziacchib,, Mauro Biffib, Renato P. Riccic,
Maurizio Landolinad, Massimo Zoni-Berissoe, Eraldo Occhettaf,
Giampiero Magliag, Gianluca Bottoh, Luigi Padelettii, Giuseppe Borianij,
on behalf of the Italian Association of Arrhythmology and Cardiac Pacing (AIAC)
The purpose of this two-part consensus document is to
provide specific suggestions (based on an extensive
literature review) on appropriate pacemaker setting in
relation to patients’ clinical features. In part 2, criteria for
pacemaker choice and programming in atrioventricular
blocks and neurally mediate syncope are proposed. The
atrioventricular blocks can be paroxysmal or persistent,
isolated or associated with sinus node disease. Neurally
mediated syncope can be related to carotid sinus syndrome
or cardioinhibitory vasovagal syncope. In sinus rhythm, with
persistent atrioventricular block, we considered appropriate
the activation of mode-switch algorithms, and algorithms
for auto-adaptive management of the ventricular pacing
output. If the atrioventricular block is paroxysmal, in addition
to algorithms mentioned above, algorithms to maximize
intrinsic atrioventricular conduction should be activated.
When sinus node disease is associated with atrioventricular
block, the activation of rate-responsive function in patients
with chronotropic incompetence is appropriate. In
permanent atrial fibrillation with atrioventricular block,
algorithms for auto-adaptive management of the ventricular
pacing output should be activated. If the atrioventricular
block is persistent, the activation of rate-responsive
Introduction
Symptomatic atrioventricular block is the most common
indication for permanent pacemaker implantation. In
Italy, 58.3% of pacemakers implanted in 2015 were
implanted for atrioventricular blocks.1,2 Trials have
shown that in patients receiving a pacemaker for atrioventricular block, the correct choice of pacing modality
and the appropriate device setting can maximize the
benefit of pacing therapy.
Although neurally mediated syncope is a very common
disorder, and the most common cause of syncope, it is a
rare indication for pacemaker implantation.3 Specifically,
permanent pacing may have a role in preventing syncopal
Pietro Palmisano and Matteo Ziacchi contributed equally to the writing of this
article.
1558-2027 ß 2018 Italian Federation of Cardiology. All rights reserved.
function is appropriate. In carotid sinus syndrome, adequate
rate hysteresis should be programmed. In vasovagal
syncope, specialized sensing and pacing algorithms
designed for reflex syncope prevention should be activated.
J Cardiovasc Med 2018, 19:000–000
Keywords: atrial fibrillation, atrioventricular block, bifascicular block,
bradyarrhythmias, carotid sinus syndrome, device setting, pacemaker,
pacing mode, permanent pacing, vasovagal syncope
a
Cardiology Unit, ‘Card. G. Panico’ Hospital, Tricase, bInstitute of Cardiology,
University of Bologna, S. Orsola-Malpighi University Hospital, Bologna, cSan
Filippo Neri Hospital, Rome, dOspedale Maggiore, Crema, CR, eDepartment of
Cardiology, ASL 3, Padre A Micone Hospital, Genoa, fDivision of Cardiology,
University of Eastern Piedmont, Maggiore della Carità Hospital, Novara,
g
Cardiology–Coronary Care Unit, Pugliese-Ciaccio Hospital, Catanzaro, hS.
Anna Hospital, Como, iOspedale Careggi, Florence and jCardiology Division,
Department of Diagnostics, Clinical and Public Health Medicine University of
Modena and Reggio Emilia Policlinico di Modena, Modena, Italy
Correspondence to Professor Giuseppe Boriani, MD, PhD, FESC, Cardiology
Division, Department of Diagnostics, Clinical and Public Health Medicine,
University of Modena and Reggio Emilia, Policlinico di Modena, Via del Pozzo, 71,
41124 Modena, Italy
Tel: +39 059 4225836; fax: +39 059 4224498;
e-mail: giuseppe.boriani@unimore.it
Received 26 October 2017 Revised 5 January 2018
Accepted 11 January 2018
recurrences in selected patients with cardioinhibitory
form.4,5 In this setting, the correct use of specific algorithms can increase the effectiveness of pacing.
The purpose of this two-part consensus document is to
provide a state-of-the-art review of the appropriate use of
pacemaker algorithms in relation to patients’ clinical
features, and to report suggestions of a consensus writing
group, convened by the Italian Association of Arrhythmology and Cardiac Pacing (AIAC), for appropriate pacemaker setting in different clinical scenarios. In this
second part, we propose criteria for a correct pacemaker
choice and programming in atrioventricular blocks and
neurally mediated syncope. In order to identify the
device characteristics and settings required, we divided
these rhythm disorders needing pacing into six clinical
categories, as follows: sinus rhythm with persistent
DOI:10.2459/JCM.0000000000000629
© 2018 Italian Federation of Cardiology. All rights reserved.
2 Journal of Cardiovascular Medicine 2018, Vol 00 No 00
Table 1
Appropriate type of device according to clinical scenarios
Clinical scenarios
Type of device
Comments
References
SND with or without
documentend atrial
tachyarrhythmias/AFa
DDDR
11,40,42,43
SR with persistent AV block
DDD
SR with paroxysmal AV block
DDD
SND with paroxysmal AV
block
Permanent AF with AV block
DDDR
VVIR
Carotid sinus syndrome
DDD
Cardioinhibitory vasovagal
syncope
DDD
Compared with VVIR, DDDR reduces the risk of AF, HF, and stroke; no
significant reduction in mortality has been reported. Compared with AAIR,
DDDR reduces the risk of AF and of PM re-operation; no difference in allcause mortality has been reported
Compared with VVI, DDD prevents PM syndrome with positive impact on
exercise capacity and quality of life; no significant reduction in mortality, HF,
AF, and stroke has been reported. Compared with VDD, DDD ensures a
better atrial sensing performance in the long term, and the possibility of an
atrial pacing in the event of SND development. CRT should be considered if
clinical symptoms of HF and a severely reduced LVEF are present
DDD pacing ensures the maintenance of a physiological AV synchronism in
case of progression of the block to a persistent form
Compared with VVIR, DDDR reduces the risk of AF, HF, stroke, and prevents
PM syndrome; no significant reduction in mortality has been reported
In patients requiring a high percentage of ventricular pacing (including patients
underwent AV node ablation), CRT should be considered if clinical
symptoms of HF and a severely reduced LVEF are present2
Compared with VVI, DDD ensures less hemodynamic deterioration
induced by CSM, and seems to be more effective in the prevention
of syncopal relapses
VVI and VDD pacing are not able to abort the syncope induced by tilt test
10 –14,17,20–23
23
1,44– 46
23
63 –65
78,79
AF, atrial fibrillation; AV, atrioventricular; CRT, cardiac resynchronization therapy; CSM, carotid sinus massage; HF, heart failure; LVEF, left ventricular ejection fraction; PM,
pacemaker; SND, sinus node disease; SR, sinus rhythm. a See part 16 for details.
atrioventricular block; sinus rhythm with paroxysmal
atrioventricular block; sinus node disease (SND) with
paroxysmal atrioventricular block; permanent atrial fibrillation with atrioventricular block; carotid sinus syndrome
(CSS); cardioinhibitory vasovagal syncope (VVS). In the
specific sections, we have summarized and discussed all
the available evidence in support of the proposed pacing
modality and device settings. The appropriate type of
device and the minimal appropriate device settings
required, according to clinical features, are summarized
in Tables 1 and 2, respectively. In Table 3 are summarized the appropriate device settings in the different
clinical scenarios according to goals of programming
detailed in part 1.6
Sinus rhythm with persistent atrioventricular
block
Although no randomized controlled trials evaluating the
effectiveness of pacing in patients with persistent atrioventricular block have been performed, several observational studies performed at the beginning of the
pacemaker era suggest that pacing prevents the recurrence of syncope and improves survival.7,8
Type of device
Compared with single-chamber ventricular pacing, dualchamber pacing maintains physiological atrioventricular
synchronism in patients with persistent atrioventricular
block in sinus rhythm.9
Large randomized parallel trials, however, have been
unable to show the superiority of dual-chamber pacing
over ventricular pacing with regard to the hard clinical
endpoints of mortality and morbidity. In the Pacemaker
Selection in the Elderly study, a single-blind, randomized,
10
controlled comparison of ventricular pacing and dualchamber pacing, 407 patients (50% of whom with atrioventricular block) aged 65 years or older were enrolled.
During a 30-month follow-up, no differences in either
quality of life or prespecified clinical outcomes (including death, stroke, heart failure, and atrial fibrillation)
emerged between the two pacing modes. In the Canadian Trial of Physiologic Pacing (CTOPP) trial, patients
who had undergone pacemaker implantation for symptomatic bradycardia (52% of which atrioventricular
blocks) were randomly assigned to VVI or DDD pacing.
After a mean follow-up of 3 years, physiologic pacing
provided little benefit over ventricular pacing in the
prevention of stroke or death due to cardiovascular
causes.11 The United Kingdom Pacing and Cardiovascular Events (UK-PACE) study12 enrolled 2021 elderly
patients (70 years or older) with a pacing indication due
to atrioventricular block. During a median follow-up 4.6
years, dual-chamber cardiac pacing did not reduce allcause mortality, atrial fibrillation, heart failure, or a
composite of stroke, transient ischemic attack, or other
thromboembolism. A meta-analysis of five randomized,
controlled, parallel-design trials comparing dual-chamber pacing with ventricular-based pacing in patients with
bradycardia showed no significant reduction in mortality,
heart failure, atrial fibrillation, or stroke in the subgroup
of patients with atrioventricular block.13
Several crossover trials have consistently demonstrated
that dual-chamber pacing prevents pacemaker syndrome,
which occurs in more than a quarter of patients with
atrioventricular block treated with single-chamber ventricular pacing. This effect impacts positively on exercise
capacity and quality of life.10,14 This evidence justifies
the preference for dual-chamber pacing.
© 2018 Italian Federation of Cardiology. All rights reserved.
Table 2
Minimal appropriate device settings required according to clinical scenarios
SND with atrial
tachyarrhythmias/
AFa
SR with
persistent
AV block
SR with
paroxysmal
AV block
SND with
paroxysmal
AV block
Permanent
AF with
paroxysmal
AV block
Pacing mode
Lower rate limit (b.p.m.)
Rate-responsive function
(yes/no)
Maximum sensor rate
(if rate-responsive function
is activated) (b.p.m.)
Maximum atrial tracking rate
(b.p.m.)
DDD (R)
50–70
Yes in patients CI
DDD (R)
50–70
Yes in patients CI
DDD
<50
No
DDD
<50
No
DDD (R)
50–70
Yes in patients CI
VVI
50
No
VVIR
70–80
Yes
DDD
50
No
DDD
50
No
According to patient’s
clinical characteristics
–
–
110–120
–
–
130
–
–
Yes
No
According to
patient’s clinical
characteristics
Yes
–
–
According to
patient’s clinical
characteristics
Yes
According to
patient’s clinical
characteristics
Yes
Rate hysteresis (yes/no)
Mode-switch algorithms
(yes/no)
Algorithms for preventing
and termitating atrial
tachyarrhythmias (yes/no)
Specialized sensing and
pacing algorithms for VVS
prevention (yes/no)
Algorithms for auto-adaptive
ventricular output (yes/no)
Algorithms for auto-adaptive
atrial output (yes/no)
No
Yes
No
Yes
No
Yes
No
Yes
According to
patient’s clinical
characteristics
According to
patient’s clinical
characteristics
Yes (with the
exception of
patients with a
prolonged PQ
interval at
baseline)
No
Yes
–
Algorithms to promote
intrinsic AV conduction
(yes/no)
According to
patient’s clinical
characteristics
According to
patient’s clinical
characteristics
Yes
Yes
–
No
–
Yes
Yes
No
Yes
No
Yes
No
No
No
–
–
No
No
No
No
No
No
No
–
–
No
Yes
No
No
Yes
No
No
No
Yes
No
No
Yes
Yes
No
No
Yes
–
–
No
No
Settings
According to patient’s
clinical characteristics
AF, atrial fibrillation; AV, atrioventricular; CI, chronotropic incompetence; SND, sinus node disease; SR, sinus rhythm; VVS, vasovagal syncope.
a
See part 16 for details.
Carotid sinus
syndrome
Cardioinhibitory
vasovagal
syncope
Device programming in bradycardia patients Palmisano et al. 3
© 2018 Italian Federation of Cardiology. All rights reserved.
SND
without atrial
tachyarrhythmias/
AFa
Permanent
AF with slow
mean ventricular
response
(including AV
junction ablation)
4 Journal of Cardiovascular Medicine 2018, Vol 00 No 00
Table 3
Appropriate device settings in different clinical scenarios according to goals of programming detailed in part 1
Appropriate device settings in different clinical scenarios
SNDa
SR with persistent
AV block
SR with
paroxysmal
AV block
SND with
paroxysmal
AV block
To preserve or
restore the heart
rate response
Activation of RR
function in
patients with CI
Setting a high
maximum atrial
tracking rate
Setting a high
maximum atrial
tracking rate
To maintain
physiological
conduction
Activation of
algorithms to
promote intrinsic
AV conduction
–
Activation of
algorithms to
promote intrinsic
AV conduction
Activation of RR
function in
patients with CI,
and setting a
high maximum
atrial tracking
rate
Activation of
algorithms to
promote intrinsic
AV conduction
To maximize device
longevity
Activation of
algorithms for
auto-adaptive
atrial output
Activation of
algorithms for
auto-adaptive
ventricular
output
Activation of
algorithms for
auto-adaptive
ventricular
output
To detect atrial
arrhythmia
Activation of modeswitch
algorithms
Activation of
mode-switch
algorithms
Activation of
mode-switch
algorithms
Goals of
programminga
Activation of
algorithms for
auto-adaptive
atrial and
ventricular
output
Activation of
mode-switch
algorithms
Permanent AF
with AV block
Neurally mediate
syncope (including
CSS and VVS)
Activation of RR
function in
patients with CI
Setting a high
maximum atrial
tracking rate
Activation of rate
hysteresis
Activation of
algorithms to
promote intrinsic
AV conduction,
and of rate
hysteresis
Activation of
algorithms for
auto-adaptive
atrial and
ventricular
output
Activation of modeswitch
algorithms
Activation of
algorithms for
auto-adaptive
ventricular
output
–
AF, atrial fibrillation; AV, atrioventricular; CI, chronotropic incompetence; RR, rate-responsive; SND, sinus node disease; SR, sinus rhythm; VVS, vasovagal syncope. a See
part 16 for details.
Single-lead VDD pacing is a possible alternative to dualchamber pacing in patients with atrioventricular block
and normal sinus node function. VDD pacing combines
the physiological benefits of synchronous atrioventricular
pacing with the convenience of a single-lead system
(lower cost, shorter operation and fluoroscopic times,
lower complication rate).15–19 However, the main limitation of single-lead VDD pacing is its poor atrial sensing
performance in the long term. Indeed, atrial under-sensing requiring a switch to the VVIR pacing mode has been
reported in 6.9–27.6% of patients.17,20–22 Another limitation is the potential development of SND, which is
observed in 1–2% of patients with atrioventricular block
and normal sinus node function at the time of pacemaker
implantation.17
On the basis of this evidence, we consider the dualchamber (DDD) pacemaker to be the optimal choice
for patients with complete atrioventricular block. In the
subgroup of patients with clinical symptoms of heart
failure and a severely reduced left ventricular ejection
fraction (LVEF), the implantation of a biventricular
pacemaker should be considered, in order to avoid the
deleterious effect of chronic right ventricular pacing.23–26
Goals of programming
In patients with isolated atrioventricular block, setting a
low lower rate level (LRL) (<50 b.p.m.) is appropriate, in
order to avoid forced atrial stimulation with a nonphysiological atrioventricular interval due to interatrial delay for
several hours daily. This has been reported to be associated with atrial fibrillation development during follow-
up.27 In order to preserve a physiological heart rate
fluctuation, it is appropriate to set a high maximum atrial
tracking rate (130 b.p.m.), especially in the youngest
patients and in those with an active lifestyle.
Activation of the mode-switch algorithm is also appropriate in this clinical category of patients.23 The report of
time in mode switch is a very sensitive measure of atrial
fibrillation, and should be used to assess the evolution of
atrial fibrillation during follow-up, with a view to administering anticoagulant therapy when necessary.
Patients with complete atrioventricular block require a
high percentage of ventricular pacing and often have a slow
and unstable intrinsic rhythm, or are completely pacemaker-dependent. In this setting, the algorithms for automatic management of the ventricular pacing output
improve patient safety by automatically tailoring the pacing output to threshold fluctuations, while maximizing
device longevity, with consequent reduction in replacement-related complications.28–30 There is much evidence
showing that, beyond preventing loss of capture in the
event of an unexpected pacing threshold increase,29 these
algorithms enable the service life of the pacemaker to be
matched with patients’ life expectancy, thereby reducing
the replacement rate.30 On the basis of this evidence,
activation of algorithms for the automatic management
of ventricular pacing is appropriate.
Sinus rhythm with paroxysmal atrioventricular
block
The intermittent/paroxysmal atrioventricular block that
occurs in patients with underlying heart disease and/or
© 2018 Italian Federation of Cardiology. All rights reserved.
Device programming in bradycardia patients Palmisano et al. 5
bundle branch block is usually regarded as a manifestation of intrinsic disease of the atrioventricular conduction system (Stokes-Adams). When paroxysmal
atrioventricular block occurs, the cardiac rhythm may
become dependent on subsidiary (often unreliable) pacemaker sites. The delay in the emergence of these subsidiary pacemakers and their relatively slow rates (typically
25–40 b.p.m.) can cause inadequate cerebral perfusion,
resulting in syncope and presyncope.23,31 In patients with a
history of syncope and presyncope and documented intermittent/paroxysmal third or second-degree atrioventricular block due to intrinsic disease of the atrioventricular
conduction, permanent pacing is effective in preventing
syncope recurrence and may improve survival.23
Type of device
In some patients with paroxysmal atrioventricular block,
the episodes of block are sporadic; consequently, pacing
may be required only for short periods of time. In these
cases, single-chamber ventricular pacing would be
enough to prevent asystoles and related symptoms. However, when the paroxysmal atrioventricular block is secondary to intrinsic disease of the atrioventricular
conduction system, there is a risk of progression of the
conduction disturbance over time to a persistent form.
For this reason, even paroxysmal forms of atrioventricular
back the optimal pacing mode is dual-chamber, because
this pacing mode ensures the maintenance of physiological
atrioventricular synchronism in the case of progression of
the block to a persistent form.23
Goals of programming
In patients with paroxysmal atrioventricular block without SND, setting a low LRL (<50 b.p.m.) is appropriate
in order to avoid forced atrial stimulation with a prolonged atrioventricular interval for several hours daily,
and hence reduce unnecessary right ventricular pacing.
In patients with transient atrioventricular block, preserving
spontaneous atrioventricular conduction by activating algorithms to minimize unnecessary right ventricular pacing is
appropriate, in order to prevent the deleterious effects of
excessive ventricular pacing (increased risk of atrial fibrillation or heart failure).32,33 In this setting, the promotion of
spontaneous atrioventricular conduction may also slow or
prevent the progression of the paroxysmal/intermittent
atrioventricular block to complete atrioventricular block,
and consequently to pacemaker dependency.33–35
Patients with transient atrioventricular blocks often
require a very low percentage of ventricular pacing.
Therefore, during routine follow-up, the percentage of
ventricular pacing should be systematically assessed in
order to minimize it as much as possible.23 Moreover, the
conduction disturbance should be reassessed at each
follow-up examination because, if progression of the
atrioventricular block to a persistent form is found, it
may be advisable to deactivate algorithms that search for
spontaneous atrioventricular conduction. Indeed, in
patients with persistent atrioventricular block, these algorithms can, in rare cases, cause inadvertent bradycardia36
and lead to potentially lethal bradycardia-related tachyarrhythmias.37–39 On the contrary, the use of these algorithms allows a continuous atrioventricular conduction
assessment and management in patients with intermittent and permanent atrioventricular block.40
The evaluAtioN of the SafeR mode in patients With a dualchambER pacemaker indication (ANSWER) study41
enrolled 650 consecutive patients with a pacemaker indication and randomized them to programming either with
an algorithm to minimize unnecessary right ventricular
pacing (the SafeR algorithm, LivaNova, Clamart, France)
or conventional DDD pacing. About 40% of patients with
either intermittent or permanent atrioventricular block
randomized to SafeR mode had a ventricular pacing less
than 50% over a 3-year time, indicating that, in a sizeable
proportion of patients with atrioventricular conduction
disturbance, the atrioventricular conduction shows significant circadian and monthly variation. Atrioventricular
block is therefore more dynamic than previously thought
and the use of these algorithms allows adjustment of the
atrioventricular conduction in response to continuous
changes in patients’ conduction disturbance.41
Sinus node disease with paroxysmal
atrioventricular block
In SND patients, a 0.6–1.9% per year risk of developing
an atrioventricular block has been reported.42 No trial has
specifically addressed the role of cardiac pacing in
patients with binodal dysfunction (SND þ atrioventricular block). However, a subgroup of patients with concurrent atrioventricular block has been included in many
trials of patients with SND. For example, a first-degree
atrioventricular block was present in 8.1% of the patients
enrolled in the THEOPACE (Effects of Permanent
Pacemaker and Oral Theophylline in Sick Sinus Syndrome) trial,43 and a concurrent atrioventricular block of
any degree was present in 8.3% of the patients enrolled in
the CTOPP trial,11 and in 21% of patients enrolled in the
Mode Selection Trial (MOST).44
Type of device
Considering the benefits of atrial-based pacing in patients
with SND (see part 1),6,11,13,44 –46 and the need for backup ventricular pacing in the event of an atrioventricular
block, dual-chamber pacing mode (DDDR) is appropriate in patients with SND and concurrent paroxysmal
atrioventricular block.
Goals of programming
Setting a LRL between 50 and 70 b.p.m. (according to
the patients’ clinical features) is appropriate, in order to
avoid inappropriate sinus bradycardia and to restore a
physiological atrial rate.
© 2018 Italian Federation of Cardiology. All rights reserved.
6 Journal of Cardiovascular Medicine 2018, Vol 00 No 00
As discussed in part 1,6 activation of the rate-responsive
function is advisable in patients with documented
chronotropic incompetence, in whom this feature can
lead to an increase in exercise capacity and an improvement in quality of life.47–49
In patients with transient atrioventricular blocks,
unnecessary right ventricular pacing should be systematically avoided by activating an algorithm to maximize
intrinsic atrioventricular conduction.32,50 Considering the
deleterious effects (both hemodynamic and clinical) of
prolonged atrioventricular conduction,51,52 programming
an excessively long atrioventricular interval (>220 ms),42
and/or the activation of features to minimize ventricular
pacing is debatable in patients with a prolonged baseline
PQ interval.42,50
As reported in the setting of patients with isolated SND or
atrioventricular block (see subparts ‘Type of device’ and
‘Sinus node disease with paroxysmal atrioventricular
block’ under the section heading ‘Sinus rhythm with
paroxysmal atrioventricular block’), activation of the
mode-switch algorithm is also appropriate in patients with
both sinus node and atrioventricular node dysfunction.
In patients with SND and paroxysmal atrioventricular
block, the need for atrial pacing prevail over the need for
ventricular pacing. In addition, in patients with a low
intrinsic sinus rate, a high percentage of atrial pacing is
required. In this setting, the activation of auto-adaptive
management of the atrial pacing output is appropriate, in
order to ensure constant atrial capture and to maximize
battery longevity.53,54
After pacemaker implantation, it is recommended that
the patient’s conduction disturbances be reassessed at
each follow-up examination, in order to detect possible
progression of transient atrioventricular block to a persistent form, and to update the device settings to the new
clinical needs of the patient.
Permanent atrial fibrillation with
atrioventricular block
Patients with permanent atrial fibrillation needing pacing
fall into two different clinical categories: those with a
normal mean ventricular rate and sporadic prolonged
asystoles due to paroxysmal atrioventricular blocks; those
with a slow mean ventricular rate due to a persistent
atrioventricular block. In the former, the most common
clinical manifestations are syncope and presyncope; in
the latter, heart failure and reduced exercise capacity due
to chronotropic incompetence.
Type of device
In permanent atrial fibrillation, or when atrial fibrillation
is the predominant rhythm or the ‘chosen’ rhythm for the
patient, atrioventricular synchrony cannot be restored
and single-chamber ventricular pacing (VVIR) is the
appropriate choice.
Goals of programming
Atrial fibrillation with paroxysmal atrioventricular block
Setting a low LRL (from 40 to 50 b.p.m. according to the
patients’ clinical features) is appropriate, in order to
prevent the detrimental effects of a high percentage
of right ventricular pacing and to maximize battery
longevity.
Adequate rate hysteresis should be programmed in order
to allow the spontaneous rhythm to emerge and to restrict
pacing to the short periods of time in which the
asystoles occur.
Atrial fibrillation with slow mean ventricular response
In comparison with fixed-rate pacing, rate-responsive
pacing is associated with better exercise performance,
decrease of symptoms, and improved quality of life.55,56
The LRL rate should be programmed high (70 b.p.m.)
than for sinus rhythm patients, in an attempt to compensate for the loss of active atrial filling, and the
maximum sensor rate should be programmed restrictively (110 –120 b.p.m.), in order to avoid ‘overpacing’,
that is, pacing with a heart rate faster than necessary,
which can elicit symptoms, especially in patients with
coronary artery disease.23
An ‘ablate and pace’ strategy (ablation of the atrioventricular junction followed by permanent pacemaker
implantation) is effective in controlling symptoms,
improving quality of life and functional capacity, and
reducing morbidity in patients with symptomatic
permanent atrial fibruillation and uncontrolled, drugrefractory high ventricular rate.57–59 Patients treated with
this strategy come within the setting of atrial fibrillation
with slow mean ventricular response. Indeed, after atrioventricular junction ablation, these patients become
totally pacemaker-dependent and lose the ability to
increase their heart rate appropriately during physical
activity (a condition of ‘iatrogenic’ chronotropic incompetence). This condition can be corrected by the use of
rate-responsive pacing. In RESPONSIBLE – a randomized, single-blind study60 enrolling 60 patients with
permanent uncontrolled atrial fibrillation and left
ventricular dysfunction, treated with atrioventricular
junction ablation and biventricular pacing – rate-responsive
pacing induced an acute significant increase in functional
capacity in comparison with fixed-rate pacing.
Patients with a spontaneous slow ventricular response,
and those treated with atrioventricular junction ablation,
require a high percentage of ventricular pacing. In order
to ensure constant ventricular capture, and to maximize
battery longevity, the activation of algorithms for the
automatic management of ventricular pacing output
is recommended.
In patients with clinical symptoms of heart failure and a
severely reduced LVEF, cardiac resynchronization therapy should be considered.23,25
© 2018 Italian Federation of Cardiology. All rights reserved.
Device programming in bradycardia patients Palmisano et al. 7
Carotid sinus syndrome
Carotid sinus syndrome is part of reflex syncope (neurally
mediated syncope). In its rare spontaneous form, it is a
syncope triggered by mechanical manipulation of the
carotid sinuses. In the more common form, no mechanical
trigger is found and it is diagnosed when carotid sinus
massage induces an asystole of above 3 s (cardioinhibitory
form), or a fall in systolic blood pressure of greater than
50 mmHg (vasodepressor form), or both (mixed form),
with reproduction of the spontaneous syncope.3,23 A few
small controlled trials and retrospective observational
studies have shown the benefits of pacing in reducing
the burden of syncope and of related morbidities in
patients affected by CSS.61 On the basis of this evidence,
permanent pacing is indicated in patients with dominant
cardioinhibitory CSS and recurrent unpredictable
syncope.3,23 CSS is a rare indication for pacemaker
implantation. Indeed, in pacemaker studies, the rate of
patients undergoing implantation for CSS is 2.5–3%.1,27,62
Type of device
In a small acute intrapatient study63 enrolling 11 patients
with carotid hypersensitivity VVI pacing caused a marked
deterioration in comparison with the DVI pacing mode;
VVI pacing induced a greater fall in systolic blood pressure (59 vs. 37 mmHg; P < 0.001) and a higher rate of
symptom persistence (91 vs. 27%; P ¼ 0.008). In a 2month randomized crossover study on DVI vs. VVI
pacing, performed in 23 patients affected by mixed
CSS,64 syncope occurred in 0 vs. 13% (P ¼ 0.25) and
presyncope in 48 vs. 74% (P ¼ 0.04), respectively; DVI
was the mode preferred by 64% of the patients, whereas
the remaining 36% did not express any preference
(P < 0.001). In the Westminster study of 202 patients,65
syncope recurred in 9% of DDD-paced patients, whereas
in VVI-paced patients, the rate of recurrence was twice as
high (18%). More recently, in a 6-month randomized,
double-blind crossover study enrolling 21 patients with
CSS,66 three pacing modalities were compared: VVI vs.
DDDR vs. DDDR, with a specialized sensing and pacing
algorithm for the prevention of neurally mediated
syncope (see below for details). The results of the study
were unable to confirm the initial study hypothesis of a
superiority of one pacing modality over another, both in
preventing syncopal and presyncopal recurrences, and in
improving quality of life.
Despite the lack of large randomized clinical trials,
review of the literature supports the hypothesis that
the VVI pacing mode hampers the efficacy of pacing
therapy in patients with CSS. Therefore, the optimal
pacing mode is dual-chamber.
Goals of programming
Carotid sinus syndrome is a condition characterized by
intermittent bradycardia, in which pacing may be
required only for rare and short periods of time. In this
situation, device programming should be aimed at preventing reflex bradycardia and pauses, while at the same
time avoiding the detrimental effects of a high percentage of pacing.23
To avoid forced atrial stimulation for several hours daily,
setting a low LRL (50 b.p.m., according to the patients’
clinical features) is appropriate.
Adequate rate hysteresis should be programmed in order
to allow spontaneous sinus rate to emerge and to restrict
pacing to the short period of time in which reflex bradycardia occurs.23 Manual adaptation of the atrioventricular
interval (>250 ms) or programming atrioventricular hysteresis to prevent unnecessary right ventricular pacing is
recommended, in order to minimize the percentage of
ventricular pacing.23
Some specialized sensing and pacing algorithms for the
faster detection of an insidious drop in heart rate and
short-lasting intervention pacing at a high rate (see subsection ‘Goals of programming’ under the section ‘Cardioinhibitory vasovagal syncope’ for details) have been
developed by manufacturers, in order to improve the
effect of pacing in reflex syncope. Initial clinical experiences67 and the results of a small noncontrolled study68
suggest that these algorithms are effective in reducing the
incidence of syncopal recurrences and symptoms in the
setting of CSS. To date, however, no comparison with
conventional dual-chamber pacing has ever been made.
Therefore, the evidence is not sufficient to recommend
routine activation of these algorithms in patients
with CSS.
Cardioinhibitory vasovagal syncope
Vasovagal syncope is an abnormal cardiovascular reflex
characterized by an inappropriate reduction in heart rate
and systemic hypotension caused by arteriolar vasodilatation. In patients with severe recurrent VVS that is
refractory to conventional and/or pharmacological treatment, and a prevalent cardioinhibitory response to the
head-up tilt test (HUTT), permanent pacing prevents
important bradycardia and/or systole leading to clinical
benefits and reducing episodes of syncope.3,4,69,70 However, several studies have shown that VVS may not be
completely prevented by conventional cardiac pacing.69–
72
These suboptimal results are probably justified by the
inability of conventional electrical cardiac stimulation to
counteract the vasodepressor component of the vasovagal
reflex that is present in almost all subjects during syncopal episodes and that usually precedes cardioinhibition
and bradycardia.72 Another possible cause of the poor
efficacy of pacing in VVS patients selected on the basis of
their HUTT response might be the weak correlation
existing between the mechanism of spontaneous
neuro-mediated syncope and HUTT response.73,74 In
this regard, the Third International Study on Syncope
of Uncertain Etiology (ISSUE-3) trial75 suggested that
© 2018 Italian Federation of Cardiology. All rights reserved.
8 Journal of Cardiovascular Medicine 2018, Vol 00 No 00
patients with VVS should be selected for permanent
pacing on the basis of the documentation of spontaneous
asystole through prolonged ECG monitoring by implantable loop recorder (ILR). Initially, 511 patients received
an ILR for recurrent SVV; 89 of these had documentation
of syncope with moiré than 3 s asystole or more than 6 s
asystole without syncope and met criteria for pacemaker
implantation; 77 of the 89 patients were randomly
assigned to dual-chamber pacing with rate-drop response
or to sensing only. After 2 years of follow-up, patients
with pacemaker on showed a significantly reduced risk of
syncope recurrence. In accordance with these results, the
‘ISSUE-3 approach’ to selecting patients with suspected
SVV to receive a pacemaker was implemented by current
European guidelines for cardiac pacing.23
Type of device
During the cardioinhibitory vasovagal reflex, the efferent
vagal activity leads more frequently to a transient depression of the sinus node function, resulting in sinus bradycardia and sinus pauses; transient atrioventricular blocks
are less frequently observed and are almost always associated to sinus node depression.76,77 These are the pathophysiological bases for preferring dual-chamber pacing to
ventricular pacing. In a small acute study, VVI pacing was
not able to abort the syncope or to prevent the symptoms
of vagal reaction induced by HUTT. During the vagal
reflex, the onset of ventricular pacing with retrograde
atrioventricular conduction led to hemodynamic deterioration and the rapid reproduction of syncope.78 In a
crossover study, the effectiveness of VDD pacing with a
short atrioventricular interval in preventing tilt-induced
syncope was evaluated in a series of 11 patients affected
by cardioinhibitory VVS.79 VDD pacing did not prevent
the initiation or progression of tilt-induced VVS. In
comparison with unpaced tilt tests, VDD pacing
appeared to accelerate the onset of syncope. Although
weak, this evidence is sufficient to suggest that dualchamber pacing should be considered the optimal pacing
mode in the setting of VVS.
Goals of programming
As already discussed with regard to CSS (see subsection
‘Goals of programming’ under the section heading
‘Carotid sinus syndrome’) in patients with cardioinhibitory VVS, pacing should be restricted to the short period
of time in which reflex bradycardia and/or pause occur. In
this view, setting a low LRL (50 b.p.m., according to the
patients’ clinical features), and the activation of algorithms to maximize intrinsic atrioventricular conduction
are appropriate, in order to minimize unnecessary atrial
and ventricular pacing.23
Some specialized sensing and pacing algorithms designed
to provide better prevention of VVS have been developed
by various manufacturers. Two examples of these algorithms are the Rate Drop Response (RDR) (Medtronic
Inc., Minneapolis, Minnesota, USA) and the Sudden
Brady Response (SBR) (Boston Scientific Corp., Place
Natick, Massachusetts, USA). The operation of these
algorithms is similar. The device monitors the heart rate
in order to detect any significant rate drop. When a rapid
rate drop is detected, the device institutes rapid DDD
pacing for a programmable time interval. While several
studies have shown that, in patients with severe VVS,
these algorithms are more effective than conservative
treatments in reducing syncopal episodes,68–69,75 no comparison with conventional dual-chamber pacing has ever
been made. In a small randomized study, DDD pacing
with the RDR function was more effective than DDI
pacing with rate hysteresis in cardioinhibitory VVS.80
Closed Loop Stimulation (CLS) (Biotronik GmbH &
Co., Germany) is an algorithm for pacing rate modulation
which responds to myocardial contraction dynamics by
measuring variations in right ventricular intracardiac
impedance. In the first stage of VVS, CLS detects the
increased myocardial contractility and activates high-rate
DDD pacing; this may anticipate the loss of sympathetic
tone and counterbalance the increase in vagal tone, thus
preventing arterial hypotension, bradycardia, and possibly syncope.81,82 According to this rationale, some studies have used CLS pacing in patients with VVS. In the
Inotropy controlled pacing in vasovagal syncope
(INVASY) – a randomized, single-blind study – CLS
pacing proved more effective than DDI pacing in preventing the recurrence of VVS in patients with a cardioinhibitory response to HUTT during a mean follow-up
of 19 months.83 In this study, no patient treated with CLS
pacing experienced syncopal recurrence. The effectiveness of CLS pacing in preventing syncopal recurrence is
maintained even in the long term, as shown by the results
of a single-center long-term follow-up study,84 in which
the burden of syncope episodes was compared before and
after CLS implantation in 35 patients followed up for
a mean of 61 months. The results of a more recent,
randomized, single-blind, crossover study85 have conclusively demonstrated the superiority of CLS to DDD
pacing (without hysteresis or dedicated algorithms for
VVS prevention) in the prevention of syncopal relapses in
50 patients with cardioinhibitory VVS. Over a period of 36
months, fewer syncopal and presyncopal episodes were
registered during CLS pacing than during DDD pacing
(syncopal episodes: 2 vs. 15; P ¼ 0.007; presyncopal
episodes: 5 vs. 30; P ¼ 0.004). CLS also seems to be more
effective than the conventional algorithms for VVS
prevention. A single-center, retrospective study86 evaluated 41 patients with recurrent, refractory VVS: 25 were
treated with a dual-chamber CLS pacemaker and 16 with
a dual-chamber pacemaker with conventional algorithms
for VVS prevention (RDR or SBR). Over a median of 4.4
years, patients with a dual-chamber CLS pacemaker
showed a significantly lower risk of syncopal relapses
than the others.
© 2018 Italian Federation of Cardiology. All rights reserved.
Device programming in bradycardia patients Palmisano et al. 9
In summary, although no large randomized trials have
compared these algorithms with conventional pacing,
some evidence suggests that they can increase the effectiveness of pacing in the prevention of syncopal recurrences. For these reasons, we consider their activation
appropriate in patients with cardioinhibitory VVS.
3
4
5
6
Conclusions
Manufacturers have developed several algorithms
designed for optimization of pacing therapy in different
clinical scenarios. The appropriate use of these algorithms based on suggestions proposed in this document
(derived from an extensive literature review) allows the
maximum benefit from pacing therapy to be achieved.27
In patients with persistent atrioventricular blocks setting a
low LRL and a high maximum atrial tracking rate is
appropriate in order to preserve a physiological heart rate
fluctuation. Activation of the mode-switch algorithm is also
appropriate to detect atrial tachyarrhythmias and to adjust
automatically the pacing mode. The activation of algorithms
for the automatic management of ventricular pacing output
is recommended to improve patient safety, and maximize
device longevity. If the atrioventricular block is intermittent, the activation of algorithms to minimize unnecessary
right ventricular pacing is appropriate, in order to prevent
the deleterious effects of excessive ventricular pacing.
7
8
9
10
11
12
13
In the setting of neurally mediate syndromes, the activation of specialized sensing and pacing algorithms for
prevention of VVS can increase the effectiveness of
pacing in the prevention of syncopal recurrences.
14
Acknowledgements
15
Conflicts of interest
P.P. reports personal fees from Speaker Honoraria from
Metronic, and personal fees from Proctorship agreement
with Boston Scientific for training healthcare professionals on Subcutaneous Implantable Defibrillator (SICD), outside the submitted work. R.P.R. received minor
consultancy fees by Medtronic and Biotronik, outside the
submitted work. M.L. has a speakers’ bureau appointment with St Jude Medical, Medtronic and Boston Scientific, and an advisory board relationship with St Jude
Medical and Medtronic, outside the submitted work.
L.P. has received grants and personal fees from Livanova;
personal fees from Biotronik, Medtronic, St Jude Medical, Livanova, Boston Scientific, outside the submitted
work. G.B. reported speaker’s fees of small amount from
Biotronik, Boston, and Medtronic, outside the submitted
work. There are no conflicts of interest.
References
1
2
16
17
18
19
20
21
22
23
Proclemer A, Zecchin M, D’Onofrio A, et al. The Pacemaker and
Implantable Cardioverter-Defibrillator Registry of the Italian Association of
Arrhythmology and Cardiac Pacing: annual report 2015. G Ital Cardiol
(Rome) 2017; 18:67–79.
Giammaria M, Cerrato E, Imazio M, Curnis A. Active implanted cardiac
devices and magnetic resonance: results of a survey among cardiologists
of Piedmont. J Cardiovasc Med (Hagerstown) 2017; 18:185–195.
24
Moya A, Sutton R, Ammirati F, et al. Guidelines for the diagnosis and
management of syncope (version 2009). Eur Heart J 2009; 30:
2631–2671.
Alboni P, Dinelli M, Gianfranchi L, Pacchioni F. Current treatment of
recurrent vasovagal syncope: between evidence-based therapy and
common sense. J Cardiovasc Med (Hagerstown) 2007; 8:835–839.
Christodoulides T, Ioannides M, Yiangou K, Nicolaides E. Syncope due to
prolonged asystole: an indication for pacing? J Cardiovasc Med
(Hagerstown) 2012; 13:465–467.
Ziacchi M, Palmisano P, Biffi M, et al. Clinically-oriented device
programming in bradycardia patients: part 1 (sinus node disease).
Proposals from AIAC (Italian Association of Arrhythmias and Cardiac
Pacing). J Cardiovasc Med (in press).
Edhag O, Swahn A. Prognosis of patients with complete heart block or
arrhythmic syncope who were not treated with artificial pacemakers. A longterm follow-up study of 101 patients. Acta Med Scand 1976; 200:
457–463.
Johansson BW. Complete heart block. A clinical, hemodynamic and
pharmacological study in patients with and without an artificial pacemaker.
Acta Med Scand Suppl 1966; 451:1–127.
Deniz HB, Caro JJ, Ward A, Moller J, Malik F. Economic and health
consequences of managing bradycardia with dual-chamber compared to
single-chamber ventricular pacemakers in Italy. J Cardiovasc Med
(Hagerstown) 2008; 9:43–50.
Lamas GA, Orav EJ, Stambler BS, et al. Quality of life and clinical outcomes
in elderly patients treated with ventricular pacing as compared with dualchamber pacing. Pacemaker Selection in the Elderly Investigators. N Engl J
Med 1998; 338:1097–1104.
Connolly SJ, Kerr CR, Gent M, et al. Effects of physiologic pacing versus
ventricular pacing on the risk of stroke and death due to cardiovascular
causes. Canadian Trial of Physiologic Pacing Investigators. N Engl J Med
2000; 342:1385–1391.
Toff WD, Camm AJ, Skehan JD, United Kingdom Pacing and
Cardiovascular Events Trial Investigators. Single-chamber versus dualchamber pacing for high-grade atrioventricular block. N Engl J Med 2005;
353:145–155.
Healey JS. Cardiovascular outcomes with atrial-based pacing compared
with ventricular pacing: meta-analysis of randomized trials, using individual
patient data. Circulation 2006; 114:11–17.
Castelnuovo E, Stein K, Pitt M, Garside R, Payne E. The effectiveness and
costeffectiveness of dual-chamber pacemakers compared with singlechamber pacemakers for bradycardia due to atrioventricular block or sick
sinus syndrome: systematic review and economic evaluation. Health
Technol Assess 2005; 9:; iii, xi–xiii,1–246.
Migliore F, Curnis A, Bertaglia E. Axillary vein technique for pacemaker and
implantable defibrillator leads implantation: a safe and alternative
approach? J Cardiovasc Med (Hagerstown) 2016; 17:309–313.
Busacca P, Gheller G, Pupita M, et al. Long-term follow-up of patients
paced in VDD mode for advanced atrioventricular block: a pilot study. J
Cardiovasc Med (Hagerstown) 2008; 9:39–42.
Wiegand UK, Bode F, Schneider R, et al. Atrial sensing and AV synchrony
in single lead VDD pacemakers: a prospective comparison to DDD
devices with bipolar atrial leads. J Cardiovasc Electrophysiol 1999;
10:513–520.
Wiegand UK, Bode F, Bonnemeier H, Eberhard F, Schlei M, Peters W.
Long-term complication rates in ventricular, single lead VDD, and dual
chamber pacing. Pacing Clin Electrophysiol 2003; 26:1961–1969.
Eberhardt F, Bode F, Bonnemeier H, et al. Long term complications in
single and dual chamber pacing are influenced by surgical experience and
patient morbidity. Heart 2005; 91:500–506.
Wiegand UK, Potratz J, Bode F, et al. Age dependency of sensing
performance and AV synchrony in single lead VDD pacing. Pacing Clin
Electrophysiol 2000; 23:863–869.
Huang CC, Tuan TC, Fong MC, Lee WS, Kong CW. Predictors of
inappropriate atrial sensing in long-term VDD-pacing systems. Europace
2010; 12:1251–1255.
Marchandise S, Scavée C, le Polain de Waroux JB, et al. Long-term followup of DDD and VDD pacing: a prospective nonrandomized single-centre
comparison of patients with symptomatic atrioventricular block. Europace
2012; 14:496–501.
Brignole M, Auricchio A, Baron-Esquivias G, et al. 2013 ESC Guidelines on
cardiac pacing and cardiac resynchronization therapy: the Task Force on
cardiac pacing and resynchronization therapy of the European Society of
Cardiology (ESC). Developed in collaboration with the European Heart
Rhythm Association (EHRA). Europace 2013; 15:1070–1118.
Ziacchi M, Diemberger I, Biffi M, et al. Predictors of nonsimultaneous
interventricular delay at cardiac resynchronization therapy optimization. J
Cardiovasc Med (Hagerstown) 2016; 17:299–305.
© 2018 Italian Federation of Cardiology. All rights reserved.
10 Journal of Cardiovascular Medicine 2018, Vol 00 No 00
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
Boriani G, Ziacchi M, Diemberger I, Biffi M, Martignani C, Leyva F. BLOCK
HF: how far does it extend indications for cardiac resynchronization
therapy? J Cardiovasc Med (Hagerstown) 2016; 17:306–308.
Gierula J, Cubbon RM, Jamil HA, et al. Patients with long-term permanent
pacemakers have a high prevalence of left ventricular dysfunction. J
Cardiovasc Med (Hagerstown) 2015; 16:743–750.
Biffi M, Melissano D, Rossi P, et al. The OPTI-MIND study: a prospective,
observational study of pacemaker patients according to pacing modality
and primary indications. Europace 2014; 16:689–697.
Biffi M, Bertini M, Saporito D, et al. Actual pacemaker longevity: the benefit
of stimulation by automatic capture verification. Pacing Clin Electrophysiol
2010; 33:873–881.
Biffi M, Bertini M, Mazzotti A, et al. Long-term RV threshold behavior by
automated measurements: safety is the standpoint of pacemaker longevity!.
Pacing Clin Electrophysiol 2011; 34:89–95.
Boriani G, Rusconi L, Biffi M, et al. Role of ventricular Autocapture function
in increasing longevity of DDDR pacemakers: a prospective study.
Europace 2006; 8:216–220.
Lee S, Wellens HJ, Josephson ME. Paroxysmal atrioventricular block. Heart
Rhythm 2009; 6:1229–1234.
Wilkoff BL, Cook JR, Epstein AE, et al. Dual-chamber pacing or ventricular
backup pacing in patients with an implantable defibrillator: the Dual
Chamber and VVI Implantable Defibrillator (DAVID) Trial. J Am Med Assoc
2002; 288:3115–3123.
Rosenqvist M, Edhag O. Pacemaker dependence in transient high-grade
atrioventricular block. Pacing Clin Electrophysiol 1984; 7:63–70.
Nagatomo T, Abe H, Kikuchi K, Nakashima Y. New onset of pacemaker
dependency after permanent pacemaker implantation. Pacing Clin
Electrophysiol 2004; 27:475–479.
Korantzopoulos P, Letsas KP, Grekas G, Goudevenos JA. Pacemaker
dependency after implantation of electrophysiological devices. Europace
2009; 11:1151–1155.
Pascale P, Pruvot E, Graf D. Pacemaker syndrome during managed
ventricular pacing mode: what is the mechanism? J Cardiovasc
Electrophysiol 2009; 20:574–576.
Sweeney MO, Ruetz LL, Belk P, Mullen TJ, Johnson JW, Sheldon T.
Bradycardia pacing-induced short-long-short sequences at the onset of
ventricular tachyarrhythmias: a possible mechanism of proarrhythmia? J Am
Coll Cardiol 2007; 50:614–622.
van Mechelen R, Schoonderwoerd R. Risk of managed ventricular pacing in
a patient with heart block. Heart Rhythm 2006; 3:1384–1385.
Vavasis C, Slotwiner DJ, Goldner BG, Cheung JW. Frequent recurrent
polymorphic ventricular tachycardia during sleep due to managed
ventricular pacing. Pacing Clin Electrophysiol 2010; 33:641–644.
Auricchio A, Ellenbogen KA. Reducing ventricular pacing frequency in
patients with atrioventricular block: is it time to change the current pacing
paradigm? Circ Arrhythm Electrophysiol 2016; 9:.
Stockburger M, Boveda S, Moreno J, et al. Long-term clinical effects of
ventricular pacing reduction with a changeover mode to minimize ventricular
pacing in a general pacemaker population. Eur Heart J 2015; 36:151–157.
Nielsen JC, Thomsen PE, Højberg S, et al. A comparison of single-lead
atrial pacing with dual-chamber pacing in sick sinus syndrome. Eur Heart J
2011; 32:686–696.
Alboni P, Menozzi C, Brignole M, et al. Effects of permanent pacemaker and
oral theophylline in sick sinus syndrome the THEOPACE study: a
randomized controlled trial. Circulation 1997; 96:260–266.
Lamas GA, Lee KL, Sweeney MO, et al. Mode selection trial in sinus-node
dysfunction. Ventricular pacing or dual-chamber pacing for sinus-node
dysfunction. N Engl J Med 2002; 346:1854–1862.
Andersen HR, Thuesen L, Bagger JP, Vesterlund T, Thomsen PE.
Prospective randomised trial of atrial versus ventricular pacing in sick-sinus
syndrome. Lancet 1994; 344:1523–1528.
Link MS, Hellkamp AS, Estes NA 3rd, et al. High incidence of pacemaker
syndrome in patients with sinus node dysfunction treated with ventricularbased pacing in the Mode Selection Trial (MOST). J Am Coll Cardiol 2004;
43:2066–2071.
Sulke N, Chambers J, Dritsas A, Sowton E. A randomized double-blind
crossover comparison of four rate-responsive pacing modes. J Am Coll
Cardiol 1991; 17:696–706.
Capucci A, Boriani G, Specchia S, Marinelli M, Santarelli A, Magnani B.
Evaluation by cardiopulmonary exercise test of DDDR versus DDD pacing.
Pacing Clin Electrophysiol 1992; 15:1908–1913.
Padeletti L, Pieragnoli P, Di Biase L, et al. Is a dual-sensor pacemaker
appropriate in patients with sino-atrial disease? Results from the
DUSISLOG study. Pacing Clin Electrophysiol 2006; 29:34–40.
Sweeney MO, Bank AJ, Nsah E, et al. Minimizing ventricular pacing to
reduce atrial fibrillation in sinus-node disease. N Engl J Med 2007;
357:1000–1008.
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
Cheng S, Keyes MJ, Larson MG, et al. Long-term outcomes in individuals
with prolonged PR interval or first-degree atrioventricular block. J Am Med
Assoc 2009; 301:2571–2577.
Nielsen JC, Thomsen PE, Hojberg S, et al. Atrial fibrillation in patients with
sick sinus syndrome: the association with PQ-interval and percentage of
ventricular pacing. Europace 2012; 14:682–689.
Biffi M, Bertini M, Saporito D, et al. Automatic management of atrial and
ventricular stimulation in a contemporary unselected population of
pacemaker recipients: the ESSENTIAL Registry. Europace 2016;
18:1551–1560.
Sperzel J, Milasinovic G, Smith TW, et al. Automatic measurement of atrial
pacing thresholds in dual-chamber pacemakers: clinical experience with
atrial capture management. Heart Rhythm 2005; 2:1203–1210.
Lau CP, Rushby J, Leigh-Jones M, et al. Symptomatology and quality of life
in patients with rate-responsive pacemakers: a double-blind, randomized,
crossover study. Clin Cardiol 1989; 12:505–512.
Oto MA, Muderrisoglu H, Ozin MB, et al. Quality of life in patients with rate
responsive pacemakers: a randomized, cross-over study. Pacing Clin
Electrophysiol 1991; 14:800–806.
Brignole M, Menozzi C, Gianfranchi L, et al. Assessment of atrioventricular
junction ablation and VVIR pacemaker versus pharmacological treatment in
patients with heart failure and chronic atrial fibrillation: a randomized,
controlled study. Circulation 1998; 98:953–960.
Lim KT, Davis MJ, Powell A, et al. Ablate and pace strategy for atrial
fibrillation: long-term outcome of AIRCRAFT trial. Europace 2007; 9:
498–505.
Orlov MV, Gardin JM, Slawsky M, et al. Biventricular pacing improves
cardiac function and prevents further left atrial remodeling in patients with
symptomatic atrial fibrillation after atrioventricular node ablation. Am Heart J
2010; 159:264–270.
Palmisano P, Aspromonte V, Ammendola E, et al. Effect of fixed-rate vs.
rate-RESPONSIve pacing on exercise capacity in patients with permanent,
refractory atrial fibrillation and left ventricular dysfunction treated with
atrioventricular junction aBLation and bivEntricular pacing
(RESPONSIBLE): a prospective, multicentre, randomized, single-blind
study. Europace 2017; 19:414–420.
Brignole M, Menozzi C. The natural history of carotid sinus syncope and the
effect of cardiac pacing. Europace 2011; 13:462–464.
Ziacchi M, Palmisano P, Ammendola E, et al. Clinically guided pacemaker
choice and setting: pacemaker expert programming study. Europace 2017;
19:1500–1507.
Madigan NP, Flaker GC, Curtis JJ, Reid J, Mueller KJ, Murphy TJ. Carotid
sinus hypersensitivity: beneficial effects of dual-chamber pacing. Am J
Cardiol 1984; 53:1034–1040.
Brignole M, Sartore B, Barra M, Menozzi C, Lolli G. Is DDD superior to VVI
pacing in mixed carotid sinus syndrome? An acute and medium-term study.
Pacing Clin Electrophysiol 1988; 11:1902–1910.
Sutton R. Pacing in patients with carotid sinus and vasovagal syndromes.
Pacing Clin Electrophysiol 1989; 12:1260–1263.
McLeod CJ, Trusty JM, Jenkins SM, et al. Method of pacing does not affect
the recurrence of syncope in carotid sinus syndrome. Pacing Clin
Electrophysiol 2012; 35:827–833.
Johansen JB, Bexton RS, Simonsen EH, Markowitz T, Erickson MK.
Clinical experience of a new rate drop response algorithm in the
treatment of vasovagal and carotid sinus syncope. Europace 2000;
2:245–250.
Benditt DG, Sutton R, Gammage MD, et al. Clinical experience with
Thera DR rate-drop response pacing algorithm in carotid sinus
syndrome and vasovagal syncope. The International Rate-Drop
Investigators Group. Pacing Clin Electrophysiol 1997; 20 (3 Pt
2):832–839.
Connolly SJ, Sheldon R, Roberts RS, Gent M, The North American
Vasovagal Pacemaker Study (VPS). A randomized trial of permanent
cardiac pacing for the prevention of vasovagal syncope. J Am Coll Cardiol
1999; 33:16–20.
Sutton R, Brignole M, Menozzi C, et al. Dual- chamber pacing in the
treatment of neurally mediated tilt-positive cardioinhibitory syncope:
pacemaker versus no therapy: a multicenter randomized study. The
Vasovagal Syncope International Study (VASIS) Investigators. Circulation
2000; 102:294–299.
Connolly SJ, Sheldon R, Thorpe KE, et al. Pacemaker therapy for prevention
of syncope in patients with recurrent severe vasovagal syncope: Second
Vasovagal Pacemaker Study (VPS II): a randomized trial. J Am Med Assoc
2003; 289:2224–2229.
Raviele A, Giada F, Menozzi C, et al. A randomized, double-blind, placebocontrolled study of permanent cardiac pacing for the treatment of recurrent
tilt-induced vasovagal syncope: the vasovagal syncope and pacing trial
(SYNPACE). Eur Heart J 2004; 25:1741–1748.
© 2018 Italian Federation of Cardiology. All rights reserved.
Device programming in bradycardia patients Palmisano et al. 11
73
74
75
76
77
78
79
80
Brignole M, Sutton R, Menozzi C, et al. Lack of correlation between the
responses to tilt testing and adenosine triphosphate test and the
mechanism of spontaneous neurally mediated syncope. Eur Heart J 2006;
27:2232–2239.
Brignole M, Sutton R. Pacing for neurally mediated syncope: is placebo
powerless? Europace 2007; 9:31–33.
Brignole M, Menozzi C, Moya A, et al. Pacemaker therapy in patients with
neurally mediated syncope and documented asystole: Third International
Study on Syncope of Uncertain Etiology (ISSUE-3): a randomized trial.
Circulation 2012; 125:2566–2571.
Ellenbogen KA, Morillo CA, Wood MA, Gilligan DM, Eckberg DL, Smith ML.
Neural monitoring of vasovagal syncope. Pacing Clin Electrophysiol 1997;
20:788–794.
Fitzpatrick A, Theodorakis G, Vardas P, Sutton R. Methodology of head-up
tilt testing in patients with unexplained syncope. J Am Coll Cardiol 1991;
17:125–130.
Fitzpatrick AP, Travill CM, Vardas PE, et al. Recurrent symptoms after
ventricular pacing in unexplained syncope. Pacing Clin Electrophysiol
1990; 13:619–624.
Petersen ME, Price D, Williams T, et al. Short AV interval VDD pacing does not
prevent tilt induced vasovagal syncope in patients with cardioinhibitory
vasovagal syndrome. Pacing Clin Electrophysiol 1994; 17 (5 Pt 1):882–891.
Ammirati F, Colivicchi F, Toscano S, et al. DDD pacing with rate drop
response function versus DDI with rate hysteresis pacing for
cardioinhibitory vasovagal syncope. Pacing Clin Electrophysiol 1998;
21:2178–2181.
81
82
83
84
85
86
Occhetta E, Bortnik M, Vassanelli C. The DDDR closed loop stimulation for
the prevention of vasovagal syncope: results from the INVASY prospective
feasibility registry. Europace 2003; 5:153–162.
Palmisano P, Dell’Era G, Russo V, et al. Effects of closed-loop stimulation
vs. DDD pacing on haemodynamic variations and occurrence of syncope
induced by head-up tilt test in older patients with refrac\tory cardioinhibitory
vasovagal syncope: the Tilt test-Induced REsponse in Closed-loop
Stimulation multicentre, prospective, single blind, randomized study.
Europace 2017; doi: 10.1093/europace/eux015 [Epub ahead of print].
Occhetta E, Bortnik M, Audogliob R, Vassanellia C. Closed loop stimulation
in prevention of vasovagal syncope: inotropy controlled pacing in vasovagal
syncope (INVASY): a multicentre randomized, single blind, controlled
study. Europace 2004; 6:538–547.
Bortnik M, Occhetta E, Dell’Era G, et al. Long-term follow-up of DDDR
closed-loop cardiac pacing for the prevention of recurrent vasovagal
syncope. J Cardiovasc Med (Hagerstown) 2012; 13:242–245.
Russo V, Rago A, Papa AA, et al. The effect of dual-chamber closed-loop
stimulation on syncope recurrence in healthy patients with tilt-induced
vasovagal cardioinhibitory syncope: a prospective, randomised, singleblind, crossover study. Heart 2013; 99:1609–1613.
Palmisano P, Zaccaria M, Luzzi G, Nacci F, Anaclerio M, Favale S. Closedloop cardiac pacing vs. conventional dual-chamber pacing with specialized
sensing and pacing algorithms for syncope prevention in patients with
refractory vasovagal syncope: results of a long-term follow-up. Europace
2012; 14:1038–1043.
© 2018 Italian Federation of Cardiology. All rights reserved.