See discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/7545647
Characterization of Portland Cement for Use as
a Dental Restorative Material
Article in Dental Materials · June 2006
DOI: 10.1016/j.dental.2005.06.005 · Source: PubMed
CITATIONS
READS
43
111
4 authors, including:
Josette Camilleri
Franco E. Montesin
102 PUBLICATIONS 2,698 CITATIONS
8 PUBLICATIONS 638 CITATIONS
University of Malta
SEE PROFILE
University of Malta
SEE PROFILE
Richard Curtis
St Benedict's, Ealing, UK
51 PUBLICATIONS 781 CITATIONS
SEE PROFILE
Some of the authors of this publication are also working on these related projects:
European Network of Bioadhesion Expertise (ENBA) View project
All content following this page was uploaded by Josette Camilleri on 19 March 2014.
The user has requested enhancement of the downloaded file.
Dental Materials (2006) 22, 569–575
www.intl.elsevierhealth.com/journals/dema
Characterization of Portland cement for use
as a dental restorative material
Josette Camilleria,b,*, Franco E. Montesina, Richard V. Curtisb,
Thomas R. Pitt Fordb
a
Department of Building and Civil Engineering, Faculty of Architecture and Civil Engineering,
University of Malta, Malta
b
Department of Conservative Dentistry and Dental Biomaterials of the Guy’s, King’s and St Thomas’ Dental
Institute, King’s College London, London, UK
Received 27 January 2005; received in revised form 2 June 2005; accepted 16 June 2005
KEYWORDS
Compressive strength;
Dental filling;
Portland cement;
Setting time
Summary Objectives: The aim of this study was to evaluate the suitability of fastsetting cement formulations based on Portland cement as dental core build-up
materials using two different methods of testing compressive strength and evaluation
of setting times.
Methods: Four fast-setting cements based on Portland cement and their four
respective densified with small particle (DSP) mortars were tested for setting
time, constitution of cement by EDAX, and compressive strength using
International and British Standards. Ordinary Portland cement (OPC) was used
as a control.
Results: All the fast-setting cements had a similar elemental composition to OPC
and the setting times were less than 7 min. The compressive strength of OPC was
different between the two methods (P!0.001). All the fast-setting cements
tested showed no difference in compressive strength regardless of the method of
testing at 1 and 7 days (PO0.05), but the cylinders showed a lower compressive
strength at 28 days (P!0.05). The OPC DSP mortar showed poorer compressive
strength than OPC (P!0.01) at all times for cube testing but not for cylinder
testing, where no difference was observed. The fast-setting DSP mortars had a
lower compressive strength at 1 day (P!0.005) with both methods. At later
times, there was no difference between the cements and DSP mortars for the
cubes.
Significance: The pure fast-setting cements set in !7 min and were not susceptible
to changes in the compressive strength testing procedure at 1 and 7 days but at
28 days all the fast-setting cements had a significantly higher strength with the test
using cubes (P!0.05). A reduction in strength was observed at 28 days in cylinder
testing. Most of the cements tested did not show encouraging strengths, however,
* Corresponding author. Address: Department of Building and Civil Engineering, Faculty of Architecture and Civil Engineering,
University of Malta, Malta. Tel.: C356 2340 2894.
E-mail address: joz@global.net.mt (J. Camilleri).
0109-5641/$ - see front matter Q 2005 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
doi:10.1016/j.dental.2005.06.005
570
J. Camilleri et al.
one of the prototype cements tested could be a prospective dental restorative
material.
Q 2005 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
Introduction
With the introduction of mineral trioxide aggregate
(MTA) as a root-end filling material, a new example
of the class described as ‘water-based cement’ has
been introduced in dentistry [1]. As MTA is a dental
water-based cement, its strength testing should
conform to EN ISO 9917-1-2003 Part 1 [2]. MTA has a
similar constitution to ordinary Portland cement
(OPC) and is composed primarily of tricalcium
silicate and dicalcium silicate [3]. The setting
time of MTA is 2 h 45 min [4], which is too long for
the material to be used as a conventional dental
filling. The long setting time initiated the need to
improve the properties of the existing material and
produce one that sets faster and has compressive
strength at least similar to glass ionomer cement.
The improved characteristics of such a prototype
material could make it suitable for a wider variety
of applications in dentistry.
The mechanical properties of Portland cement
can be improved by producing mortars, which are
densified with small particles (DSP mortars). In the
production of DSP mortars, the cement grains are
dispersed at low water/cement ratios with the use
of high levels of dispersing agent and the inter-grain
hydrate gel is densified by the addition of very fine
silica (mean diameter 0.1 mm) [5]. Superplasticizers
are very effective dispersing agents used to reduce
the water content of the mix and retain its
workability. Water can be reduced at an equivalent
flow with an increase in compressive strength [6].
The aim of this study was to evaluate the
suitability of fast-setting cement formulations
based on OPC as core build-up material using two
different methods of testing compressive strength
and evaluation of setting times.
Materials and methods
Four accelerated cements were used for this study.
1. A mixture of 80% OPC (Central Cement, Italy)
and 20% high alumina cement (Ciment Fondu
Lafarge, Chilton, UK): OPC/HAC;
2. Emaco Ultra-rapid (MBT-FEB, Manchester, UK),
which was passed through a 300 mm sieve prior
use to eliminate the larger particles and fibers;
3. Sika Rapid (Sika, Switzerland), which was passed
through a 90 mm sieve prior use to eliminate the
larger particles;
4. Exocem R1 rapid (Ruredil spa, Milan, Italy).
The control material was OPC. The Emaco and
Sika were sieved prior use to eliminate the larger
particles. In addition, all these cements were
modified to become densified with small particles
(DSP) mortars to improve potentially their mechanical properties; DSP mortars were produced by
replacing 10% of the cement with microsilica
(Meyco 610, MBT-FEB, Manchester, UK), and
addition of 1% water reducing admixture (Rheobuild
1000, MBT-FEB, Manchester, UK). The cements
were mixed with water for testing of setting time
and compressive strength in ratios recommended
by the manufacturer for the proprietary brands and
according to the standard consistence test for the
other materials (Table 1).
Chemical constitution
The chemical constitution of the fast-setting
cements was determined using Energy Dispersive
Analysis by X-rays (EDAX) in the scanning electron
microscope (SEM) (Hitachi S3500, Hitachi, Wokingham, UK) and compared with that of OPC. A thin
Table 1 Water to cement percentage and mean
setting times of cements (SD), nZ4.
Cement
type
%Water to
cement
Initial set
minutes
Final set
minutes
OPC
OPC/HAC
Emaco
(300 mm)
Sika rapid
(90 mm)
Exocem
OPC DSP
OPC/HAC
DSP
Emaco DSP
(300 mm)
Sika DSP
(90 mm)
Exocem
DSP
32.5
35
25
262.0 (13.6)
3.6 (0.14)
4.2 (0.3)
479.5 (12.8)
6.4 (0.5)
5.8 (0.5)
32.5
2.5 (0.4)
6.2 (1.9)
32.5
30
33.5
4.3 (1.4)
301.3 (18.4)
4.4 (0.5)
6.9 (1.8)
497.5 (9.6)
6.8 (0.4)
35
4.1 (1.2)
5.8 (0.1)
35
3.3 (0.2)
7.0 (0.4)
27.5
6.0 (0.4)
8.9 (0.3)
Use of Portland cement as a dental restorative material
layer of powder was dispersed over a polymethylmethacrylate slab mounted on an aluminum stub
(Agar Scientific, Stansted, UK). The stubs were
carbon-coated (Emitech K250, Ashford, UK) for
electrical conductivity. The specimens were then
viewed under the SEM and the energy dispersive
analysis by X-ray (EDAX) was carried out to
determine the constituent elements of the
powders.
Setting time
The setting time of the cements was tested using
the Vicat apparatus according to BS EN 196-3
(1995) [7]; 400 g of the cement were mixed with
water on a moistened aluminum tray with a small
trowel. The amount of water required was
calculated with the standard consistency needle
[7] for Portland cement and mixtures of Portland
and high alumina cement, according to manufacturers’ instructions for the fast-setting cements.
The cement was then compacted in a two-part
brass mold and initial and final setting times were
determined. The Vicat assembly was placed in an
incubator at 37 8C. The test was repeated four
times.
Compressive strength
The compressive strength of the cements was
determined according to BS 4550 Part 3 Section
3.4 1978 [8] for testing Portland cement and
according to BS EN ISO 9917-1: 2003 for testing
dental cements [2]. For the first part of the
experiment, 740 g of cement without standardized sand were mixed on a moistened aluminum
tray with a trowel for 2 min at the established
water/cement ratio. After mixing, the cement
was loaded into cubes of side length 70.7 mm
held together in a vice. The metal mold was
previously coated with mold oil (Separol, Sika,
Switzerland). The mix was vibrated for 1 min to
allow all the air bubbles to come to the surface.
The surface of the cube was then made smooth
with the flat end of a trowel. The mixes were
allowed to cure for 24 h at room temperature.
They were then removed from the molds and
stored in an incubator at 37 8C and 100%
humidity. Three cubes were prepared for each
time point for different materials.
For the second part of the test, 10 g of cement
was mixed on a glass slab with water at the
established water/cement ratio. The mix was then
compacted in brass molds of 6 mm in diameter and
16 mm high with 2 mm caps at both ends to produce
571
cylinders of 6 mm in diameter and 12 mm high. A
vibrator was used for 1 min during the filling and
compaction of the cements into the cylinder to
avoid entrapped air. The cylinders were stored in an
incubator at 37 8C and 100% humidity. Six cylinders
were prepared for each time point for different
materials.
Compressive strength testing was performed
after 1 day, 7 and 28 days. The cubes were
compressed using a compression machine (Avery
Denison Impact Laboratory and Engineering
Supplies, Leeds, UK) with a loading rate of
0.6 N mmK2 sK1 until they failed. The cylinders
were compressed using a Universal testing machine
(Instron 1195, High Wycombe, UK) with a cross head
speed of 1 mm/min until they failed. The maximum
load required to fracture the samples was noted.
Compressive strength was calculated using the
formula:
Compressive strength
Z
Applied load at failure ðNewtonsÞ
Original cross sectional area ðmm2 Þ
Effect of specimen size on compressive
strength.
The effect of specimen size on compressive
strength was evaluated by casting cubes of OPC.
Thirty cubes were cast in molds of side length
70.7 mm while another 30 cubes were cast in 50 mm
side length molds. The specimens were left at room
temperature covered with a sheet of plastic for 24 h
after which they were demolded and cured at 100%
humidity at 37 8C for 28 days. The cubes were
crushed using a compression machine (Avery
Denison) with a loading rate of 0.6 N mmK2 sK1
until they failed.
Statistical analysis
The data were analyzed statistically using SPSS. The
distribution was first evaluated to determine the
statistical test that would be performed. The data
was plotted and the distribution curve was analyzed
together with the Kolmogorov–Zmirnov test with
PZ0.05. This signified a normal distribution and
thus parametric tests could be performed accordingly. With normally distributed data, Analysis of
Variance (ANOVA) with PZ0.05 was first performed
to determine any variation between the means.
Then two-tailed independent sample t-test at 95%
572
J. Camilleri et al.
confidence level with PZ0.05 was used to analyze
the data.
Results
Chemical constitution
The three fast-setting cements had a similar
elemental constitution to OPC (Fig. 1). They were
composed of calcium, silicon, aluminum and a small
proportion of sulphur. A small iron peak was visible
in all samples but was not labeled due to the small
proportion; the iron peak was only labeled in Sika
rapid.
Setting time
The setting times of the cements and their
respective DSP mortars are shown in Table 1. All
the cements tested except for OPC had adequate
setting time (!9 min) and thus could be potential
dental restorative materials. There was no statistically significant difference between the setting
time of cements and the DSP mortars with PO0.05
(independent sample t-test at 95% confidence level).
Figure 1
Compressive strength testing
The results of compressive strength tests performed on cubes cast according to BS 4550
Section 3 Part 3.4 1970 are shown in Table 2.
The cubes failed in normal mode (Fig. 3a). The
standard deviation was small for all the cubes
tested. The results of compressive strength
testing according to BS EN ISO 9917: 2003 are
shown in Table 3. The cylinders failed in shear
mode (Fig. 3b). The standard deviation for the
cylinder testing was larger than for the cube
testing. For the cylinder testing, the strength
values did not rise incrementally with material
aging in contrast to the cubes. The data tested
had a normal distribution. The OPC DSP mortar
showed poorer compressive strength than the OPC
(P!0.01) at all times for cube testing but not for
cylinder testing, where no difference was
observed. The fast-setting DSP mortars had
lower compressive strength at 1 day (P!0.005)
when testing both cubes and cylinders. At later
ages, there was no difference between these
cements and their DSP mortars for the cube
testing. The OPC/HAC and Emaco DSP cylinders
showed poor compressive strength values at
EDAX of (a) OPC, (b) Emaco ultra-rapid, (c) Sika rapid and (d) Exocem, respectively.
Use of Portland cement as a dental restorative material
573
Table 2 Mean compressive strength of 70.7!
70.7 mm cement cubes, nZ3 (SD).
Cement type
Compressive strength (N/mm2)
1 day
7 days
28 days
OPC
OPC/HAC
Emaco
(300 mm)
Sika rapid
(90 mm)
Exocem
OPC DSP
OPC/HAC DSP
Emaco DSP
(300 mm)
Sika DSP
(90 mm)
Exocem DSP
48.4 (2.6)
24.1 (1.4)
43.0 (0.8)
68.6 (2.9)
39.2 (3.3)
57.0 (6.0)
83.3 (4.4)
40.5 (2.0)
64.8 (1.1)
25.8 (0.8)
26.4 (2.2)
29.9 (1.3)
14.7
38.9
17.0
34.0
25.1
47.6
37.3
48.1
31.5
58.5
42.0
61.0
(0.5)
(2.5)
(0.7)
(3.4)
(1.0)
(3.9)
(0.9)
(1.1)
(1.9)
(4.3)
(1.1)
(1.7)
16.7 (1.5)
22.4 (2.2)
34.0 (5.6)
12.0 (0.6)
21.3 (2.5)
29.0 (1.5)
7 days (P!0.02) but improved at 28 days (P!
0.05). The Sika and Exocem showed the opposite
with no difference at 7 days but poorer strengths
at 28 days (P!0.001).
Comparison of cylinder to cube compressive
strength showed that that there was a statistically
significant difference between the tests for OPC at
all times (P!0.001). The mixture of OPC and HAC
also showed a statistically significant difference
between the tested cubes and cylinders at 1 day
with PZ0.017. The fast-setting cements were not
affected by the different method of testing at 1 and
7 days, but there was a statistically significant
difference at 28 days (P!0.05), with the cylinders
having lower compressive strength.
Table 3 Mean compressive strength of 6!12 mm
cement cylinders, nZ6 (SD).
Cement type
Compressive strength (N/mm2)
1 day
7 days
28 days
OPC
OPC/HAC
Emaco
(300 mm)
Sika rapid
(90 mm)
Exocem
OPC DSP
OPC/HAC DSP
Emaco DSP
(300 mm)
Sika DSP
(90 mm)
Exocem DSP
21.2 (3.3)
19.2 (0.6)
48.5 (9.6)
21.2 (6.0)
31.7 (7.9)
56.1 (9.5)
26.8 (4.8)
38.8 (8.1)
47.5 (3.0)
22.7 (2.4)
36.5 (14.5)
40.2 (4.3)
17.5 (2.9)
21.5 (5.2)
9.4 (2.1)
29.1 (4.8)
24.7 (7.2)
16.3 (2.2)
18.0 (5.8)
40.4 (11.4)
24.9
20.5
31.1
52.4
9.4 (3.8)
29.0 (2.0)
26.6 (4.1)
17.4 (2.7)
22.9 (5.0)
11.7 (3.8)
(4.2)
(5.5)
(3.2)
(9.0)
Figure 2 Means and standard deviation of compressive
strength on different sized cubes for OPC (nZ30).
Effect of specimen size on compressive
strength
Fig. 2 shows the means and standard deviation
of compressive strength of different cubes. The
data distribution is shown in Table 4. The data were
normally distributed. The larger cubes had a higher
compressive strength (P!0.001, t-test).
Discussion
Although OPC has been developed as a dental
material, only limited information has been published on its physical and chemical properties [3,4].
Since the cement is mixed with water in order to
set, it can be regarded as dental water-based
cement. Thus, its testing could conform to BS EN
ISO 9917 2003 [2]. Normally OPC is used as a binder
in concrete for construction purposes; it is thus not
used on its own. For construction purposes, there
are standards that stipulate the way the material
should be tested. The testing for setting time and
compressive strength of OPC is defined by BS EN
196-3 (1995) [7] and BS 4550 Section 3, Part 3.4
(1978) [8], respectively. Like all water-based
Table 4 One-Sample Kolmogorov–Smirnov Test for
comparison of cubes of different sizes.
Cube
2
Surface area (mm )
N
Mean
Std dev.
Minimum
Maximum
Kolmogorov–Smirnov Z
P (two-tailed)
70.7 mm2
5000
30
65.2
3.9
55.04
72.6
0.424
0.994
50 mm2
2500
29
51.5
5.3
40
64.2
0.637
0.812
574
cements, OPC is very susceptible to additions of
excess water. Thus, the quantity of water added to
the mix is important, as excess water may lead to a
marked decrease in strength [9], while too little
water may result in a dry mix with no water
available for the hydration reaction [10]. The right
amount of water added to the powder can be
calculated using the Vicat and the standard
consistence needle [7]. For the fast-setting
cements, this procedure was difficult to perform
as the working time of the cement was limited;
thus, manufacturer’s instructions were used.
In this experiment, cylinders of 6 mm in diameter
and 12 mm in height were used. This was done to
have a height to diameter ratio of 2. The EN ISO
9917 2003 [2] advises the use of molds producing
cylinders of 4 mm in diameter and 6 mm in height.
The strength of cylinders for height to diameter
ratio 2 is not influenced by the restraining effects of
the loading plates. Values higher than 2 may lead to
buckling of the specimens and lower values require
the use of a correction factor when calculating the
compressive strength. No packing material was
used to minimize friction between the loading
plates and the specimen. The ends were made
plane by the use of capping ends and very fine grit
sandpaper. Lack of planarity can reduce the
effective contact area between the specimen and
the bearing plate [11]. The cubes were cast in steel
molds and were tested on surfaces cast against the
mold rather than the open cast surface.
Cubes were crushed at a loading rate of
0.6 N mmK2 sK1 [8] while cylinders were crushed
at a cross head speed of 0.75 mm minK1 (loading
rate of 50 N minK1) [2]. The observed fracture
mode of cements and mortars was in the form of a
pyramid the base of which coincided with the top
and bottom of the cube and the apex pointing
inwards (Fig. 3a). Cylinders failed in shear mode
(Fig. 3b).
All the cements tested had similar elemental
constitution, which was verified by EDAX. The high
alumina cement added to the Portland cement
reacts with the calcium sulphate thus allowing the
free calcium aluminate to hydrate leading to rapid
setting of the cement. Portland cement was used as
the major constituent as this enabled control over
the setting time [11].
The DSP mortars reduced the compressive
strengths especially at early ages in contrast to
what was reported by Bache [5] who reported
improvement in compressive strengths of cements
by addition of superplasticizers and fine particles.
Some of the cements tested did not show an
incremental increase in measured compressive
strength particularly Emaco ultra-rapid which
J. Camilleri et al.
Figure 3
(a) Failed cube, and (b) failed cylinder.
even exhibited a reduction in compressive strength
between 7 and 28 days. There is little published
research on testing small Portland cement samples.
The measured strength of cement is not unique, as
it is known to be dependent on several factors
among which are specimen size and testing
conditions. The effect of size on measured strength
is well documented [11]. It is also well documented
that as hydration proceeds, micro-cracks tend to
form within the cement. These micro-cracks have a
larger influence when the sample is smaller [12].
Besides size, the samples were not cured under
water as stipulated by standards for normal, larger
cement/concrete samples. The loss in compressive
strength could have been caused by internal
cracking due to shrinkage/thermal cracks which,
given the size of the sample, would affect the
measured strength to a greater degree. The effect
of the method of curing on the measured compressive strength of cements should be investigated
further. The reduction in compressive strength
at later curing times could present a problem if
Use of Portland cement as a dental restorative material
the material is used clinically, as failure of the
restoration could result due to a material weakness.
The data obtained for changes in compressive
strength produced by altering the surface area of
the specimen verified that compressive strength
testing depends on the specific test conditions. In
the present study, Portland cement was used on its
own. The use of sand was omitted as the sand
particles were too large, and the sand was not
added to the potential restorative material.
The material was cured at 37 8C as these are the
conditions usually present in the mouth. The
Portland cement was susceptible to the testing
method unlike fast-setting cements. Smaller
samples were shown to produce lower compressive
strength values when tested. The fast-setting
cements might have less porosities; thus, the
flaws created, when the material is subjected to
compression, would not be as critical as those of
Portland cement.
Conclusions
The pure fast-setting cements set in !7 min were
not susceptible to changes in the compressive
strength testing procedure at 1 and 7 days but at
28 days, all the fast-setting cements had a
significantly higher strength with the test using
cubes (P!0.05).
Acknowledgements
The Commonwealth Scholarship Commission for
funding. Mr Philip Tabone, and Attrans Commercials
Limited for providing the materials. Ms Mahtab
View publication stats
575
Behrouzi for her assistance with procuring the
standards. Mr Pio Saliba and Mr Richard Mallett for
the technical assistance, and Mr Ken Brady for
assistance with the chemical analysis.
References
[1] Torabinejad M, Watson TF, Pitt Ford TR. Sealing ability of a
mineral trioxide aggregate when used as a root end filling
material. J Endod 1993;19:591–5.
[2] British Standard Institution: Dentistry: Water-based
cements. Part 1. Powder/liquid acid-base cements. BS EN
ISO 9917-1; 2003.
[3] Camilleri J, Montesin FE, Brady K, Sweeney R, Curtis RV, Pitt
Ford TR. The constitution of mineral trioxide aggregate.
Dent Mater 2005;21:297–03.
[4] Torabinejad M, Hong CU, McDonald F, Pitt Ford TR. Physical
and chemical properties of a new root-end filling material.
J Endod 1995;21:349–53.
[5] Bache HH. Dense cement ultrafine particle based material.
Proceedings of the 2nd symposium on superplasticizers in
concrete, Ottawa Canada, 1981.
[6] Hattori K. Superplasticizers in concrete. In: proceedngs
from the internatoinal symposium Ottawa, Canada; 1978. p.
49–86.
[7] British Standard Institution: Methods of testing cement.
Determination of setting time and soundness. BS EN 196-3;
1995.
[8] British Standard Institution: Specification for physical
testing of Portland cement: compressive strength. BS 4550
Section 3 Part 3.4; 1978.
[9] Abrams DA. Design of concrete mixtures Bulletin 1.
Structural materials research laboratory. Chicago: Lewis
Institute; 1918.
[10] Klieger P. Effect of mixing and curing temperature on
concrete strength. ACI J Proc 1958;54:1063–81.
[11] Neville AM. Properties of concrete Longman scientific and
technical essex UK. 3rd ed 1981.
[12] Griffith AA. The phenomena of rapture and flow in solids.
Philos Trans R Soc A 1920;221:163–98.