HARDWARE AND
INSTRUMENTATION Full Paper
Magnetic Resonance in Medicine 69:583–593 (2013)
Gradient System Characterization by Impulse Response
Measurements with a Dynamic Field Camera
Signe J. Vannesjo,1 Maximilan Haeberlin,1 Lars Kasper,1,2 Matteo Pavan,1
Bertram J. Wilm,1 Christoph Barmet,1 and Klaas P. Pruessmann1*
This work demonstrates a fast, sensitive method of characterizing the dynamic performance of MR gradient systems. The accuracy of gradient time-courses is often compromised by field
imperfections of various causes, including eddy currents and
mechanical oscillations. Characterizing these perturbations is
instrumental for corrections by pre-emphasis or post hoc signal
processing. Herein, a gradient chain is treated as a linear timeinvariant system, whose impulse response function is determined by measuring field responses to known gradient inputs.
Triangular inputs are used to probe the system and response
measurements are performed with a dynamic field camera consisting of NMR probes. In experiments on a whole-body MR
system, it is shown that the proposed method yields impulse
response functions of high temporal and spectral resolution.
Besides basic properties such as bandwidth and delay, it also
captures subtle features such as mechanically induced field
oscillations. For validation, measured response functions were
used to predict gradient field evolutions, which was achieved
with an error below 0.2%. The field camera used records
responses of various spatial orders simultaneously, rendering
the method suitable also for studying cross-responses and
dynamic shim systems. It thus holds promise for a range of
applications, including pre-emphasis optimization, quality
assurance, and image reconstruction. Magn Reson Med
C 2012 Wiley Periodicals, Inc.
69:583–593, 2013. V
Key words: gradient impulse response; GIRF; linear timeinvariant systems; mechanical oscillations; gradient calibration;
dynamic shimming; magnetic field monitoring
Time-varying magnetic field gradients are essential for signal preparation and encoding in magnetic resonance
imaging and spectroscopy. Most MR methods rely on
highly accurate gradient time-courses for correct signal
encoding and suffer from artifacts when significant deviations from the prescribed time-courses occur. In practice,
effective gradient waveforms usually do differ somewhat
from the ideal shapes defined in the underlying pulse
sequence. These deviations are largely due to a variety of
1
Institute for Biomedical Engineering, University and ETH Zurich, Zurich,
Switzerland.
2
Laboratory for Social and Neural Systems Research, University of Zurich,
Zurich, Switzerland.
Grant sponsor: Philips Healthcare.
*Correspondence to: Klaas P. Pruessmann, Ph.D., Institute for Biomedical
Engineering, University and ETH Zurich, Gloriastrasse 35, CH-8092 Zurich,
Switzerland. E-mail: pruessmann@biomed.ee.ethz.ch
Received 23 November 2011; revised 3 February 2012; accepted 25
February 2012.
DOI 10.1002/mrm.24263
Published online 12 April 2012 in Wiley Online Library (wileyonlinelibrary.
com).
C 2012 Wiley Periodicals, Inc.
V
hardware imperfections including bandwidth limitations
of gradient amplifiers, eddy currents induced in gradient
coils and in other conducting structures of the scanner
(1,2), field fluctuations caused by mechanical vibrations
after gradient switching (3,4), and thermal variation in
hardware components (5). Slight perturbations can also
stem from physiologically induced fields that originate in
the subject under examination (6,7) or from magnetic
sources and currents external to the MR system.
Besides further hardware optimization, the most common ways of addressing dynamic field imperfections are
precompensation of gradient waveforms (8–10) and postcorrection of acquired data (11). Both of these options
are most feasible for mechanisms of perturbation that are
reproducible and can be accurately modeled. Physiological and external field contributions as well as thermal
effects are challenging to address in this way. However,
eddy currents, which are usually the dominant causes of
field imperfections, as well as mechanical effects have
been successfully modeled and predicted.
Eddy currents arise due to the self- and cross-inductance of gradient coils as well as the cross-inductance
between gradient coils and other conductive structures of
the scanner, such as shim coils or parts of the cryostat. To
model eddy currents, the conductors involved are usually
approximated as LR circuits, i.e., as combinations of an
inductance and a resistance, in which induced currents
decay exponentially (1,2,12). Using linear systems theory,
the net step response of a gradient chain can then be
modeled as a superposition of an ideal response, i.e., a
Heaviside function, and a set of exponentially decaying
field terms representing eddy currents of different sources. Response models of this kind are traditionally used
to predict eddy current effects and to determine pre-emphasis settings for a given system (8,9,12,13).
Field oscillations due to mechanical vibrations are a
well-known disturbance in MR spectroscopy, where they
can cause modulation sidebands particularly of unsuppressed water peaks. Similar to eddy current effects, the
underlying mechanics and conversion from vibrations to
magnetic fields have been viewed as linear mechanisms
that contribute to a net linear impulse response (14). Using
this approach, mechanically mediated field oscillations
have been modeled as exponentially damped sinusoidal
responses, which have been used to simulate sideband
effects (4,15). Linear systems theory has also been used to
characterize the acoustic response of a gradient system,
relying on sound pressure measurements in the magnet
bore (16–18).
583
584
So far, the models of field responses due to eddy currents and mechanical effects have relied on simplifying
abstractions. For both mechanisms, it has been assumed
that only a limited number of distinct instances occur,
which obey an elementary pattern such that they can be
represented by small sets of exponentials and exponential
sinusoids, respectively. In general, this picture will hold
only approximately. Due to the electrical and mechanical
complexity of gradient systems and MR systems as a
whole, both eddy-current and mechanical behavior may
require expansion into very many components of different
time constants. Eddy currents may be supported by structures that cannot be fully modeled as mere LR circuits.
Moreover, coupling may occur both among structures that
can carry eddy currents and among mechanical resonances, giving rise to yet more complex behavior.
To better capture the variety of possible behavior, it is
attractive to remove model assumptions on individual
eddy-current processes or mechanical resonances and
rather characterize a gradient system in terms of its complete impulse response function (19,20). According to
linear systems theory, this approach should permit
jointly representing all response mechanisms that are
linear and time-invariant (LTI). A net gradient impulse
response function (GIRF) should hence incorporate all
influences on the gradient waveform between the console and the magnet bore. This would include amplifier
and coil characteristics as well as eddy currents and
vibration-induced fields, without the need to consider
individual underlying mechanisms. Knowledge of the
comprehensive GIRF could form the basis of advanced
pre-emphasis and serve for quality assurance purposes.
It could also yield more accurate estimates of effective kspace trajectories for image reconstruction and of other
encoding parameters such as b-values in diffusion imaging or gradient moments in velocity mapping.
The key challenge toward this goal is determining
GIRFs accurately, with sufficient bandwidth and frequency resolution, and within reasonable measurement
times. Probing the GIRF must generally involve observing
a system’s response to given gradient input waveforms.
This could be achieved with known NMR-based techniques for measuring gradients and k-space trajectories (21–
26). Partly, these methods rely on selective excitation of
larger phantoms or imaging objects and signal detection
with large receiver coils (22–24). Such detection methods
offer only limited sensitivity, while the extent of the imaging object makes it prone to dephasing by potential inplane gradient responses. Methods that rely on a dedicated small sample are also limited in sensitivity unless
the sample is mounted in a tight-fitting detector with a
high filling factor (21,25,26). All of these methods have
the disadvantage that each measurement needs to be
repeated several times either to achieve temporal resolution or to distinguish different spatial field components.
An efficient alternative is to record the field evolution
with a dynamic field camera (27), i.e., with an array of
miniature NMR probes that are operated simultaneously
and positioned such as to distinguish different spatial
components of interest (28,29). This approach was
recently proposed for monitoring magnetic field evolutions concurrently with image acquisition (30). However,
Vannesjo et al.
it is equally suited for mere field observations as
required for GIRF measurements. Relying on tightly
wound detectors around small sample droplets, it combines high sensitivity with robustness against dephasing.
The temporal resolution of the camera measurement is
limited only by the acquisition bandwidth of the spectrometer used and the simultaneous recording at different positions immediately yields differentiation of a gradient’s self- and cross-responses.
In this work, it is proposed to determine comprehensive gradient impulse response functions using field
observations with a dynamic field camera. Starting from
LTI systems theory, a strategy is derived for obtaining a
full GIRF from suitable combinations of input functions.
The method is demonstrated by GIRF measurements on
a 3 T whole-body human MRI system and validated by
comparing measured field evolutions with GIRF-based
predictions.
THEORY AND METHODS
Probing an LTI System
For a linear time-invariant system, the relation between
the input to the system, i(t), and its output, o(t), is determined by the impulse response, h(t), as described by
(31):
oðtÞ ¼
Z1
FT
iðtÞ hðt tÞdt $ OðvÞ ¼ IðvÞ HðvÞ;
½1
1
where t and v denote time and angular frequency, and
O(v), I(v), and H(v) represent the Fourier transforms
(FT) of o(t), i(t), and h(t), respectively. Accordingly, the
system response to any given input can be predicted
based on the impulse response function. Knowing the
impulse response function thus amounts to having full
information about the behavior of the system, within its
range of linearity, and the task of characterizing a linear
system amounts to determining its impulse response.
To do so by measurement, the output of the system
must be observed for known input, requiring an appropriate measurement technique. The impulse response
function can then be obtained by deconvolution of the
measured output by the input, which can be efficiently
performed by division in the frequency domain. Let
^
OðvÞ
¼ OðvÞ þ hðvÞ denote the observed output, including the measurement noise g(x) of standard deviation
r(x). Then, the resulting estimate of HðvÞ is
^
^ ðvÞ ¼ OðvÞ ¼ OðvÞ þ hðvÞ ¼ H ðvÞ þ hðvÞ
H
I ðvÞ
I ðvÞ
I ðvÞ
½2
with the signal-to-noise (SNR) ratio
SNRH^ ðvÞ ¼
jHðvÞj jIðvÞj
:
sðvÞ
½3
I(v) should therefore be large and ideally contain equal
amplitudes of all frequencies in the bandwidth of interest. A perfect Dirac delta function would evenly cover
an infinite bandwidth but is not a feasible input for a
Gradient System Characterization With a Dynamic Field Camera
585
real-world gradient system. In modern MR systems, the
gradient input is typically defined digitally with a finite
dwell time and is subject to limitations in terms of the
maximum amplitude and slope of the gradient waveform. Under these constraints, a feasible approximation
of a Dirac delta function is a short triangle pulse. A triangular function can be described as the convolution of
two identical boxcar functions:
iðtÞ ¼ p ðboxT boxT ÞðtÞ ¼
p ðT jtjÞ; jtj T
0;
jtj > T
½4
where boxT has length T and amplitude 1, and p is the
slope of the triangle (Fig. 1). In the frequency domain, a triangular input thus corresponds to a squared sinc function,
IðvÞ / p
sin2 ðvT=2Þ
;
v2
½5
which has several relevant properties with respect to the
choice of its parameters. First, as can be seen from Eq. 5,
the envelope of the sinc-squared is determined by the
slope, p, of the triangle. To optimize the sensitivity of
the measurement, the slope should therefore generally be
maximized. Second, the amplitude of the main lobe of
I(v) scales with the total moment of the pulse:
Zþ1
Ið0Þ /
iðtÞdt ¼ p T 2 ;
½6
1
and third, the sinc-squared input exhibits zeros with a
spacing that scales inversely with the length of the triangle:
vjI¼0 ¼
2pn
; n ¼ 6 1; 2; 3::::
T
½7
These zeros amount to blind spots in the measurement. Therefore, in the choice of the triangle length, a
trade-off needs to be made between the bandwidth covered by the main lobe, which is larger for a shorter pulse,
and the sensitivity at low frequencies, which increases
strongly with the pulse duration.
This dilemma can be circumvented by using multiple
triangular pulses of different length, giving complementary response information. When using multiple input
functions, the data obtained in the different measure^
ments need to be combined in calculating HðvÞ.
A leastsquares estimate at each frequency is obtained by
^
HðvÞ
¼
j Ij ðvÞ
P
P
j
^ j ðvÞ
O
jIj ðvÞj
2
¼ HðvÞ þ
j Ij ðvÞ
P
P
j
hj ðvÞ
jIj ðvÞj2
;
½8
where I*(v) denotes the complex conjugate of I(v), and
the index j counts the different input functions. Assuming equal noise levels and no noise correlation between
the different measurements, the variance of the noise
term in Eq. 8 sums up to
s2 ðvÞ
s2H^ ðvÞ ¼ P
2;
j jIj ðvÞj
½9
FIG. 1. The set of triangular input functions used for GIRF measurements, shown in the time-domain (a), and corresponding sincsquared functions in the frequency-domain (b). The frequency-domain plot includes the envelope of the side-lobes falling off with p/v2,
and the root-sum-of-squares of all inputs, indicating the combined
sensitivity of the set of input functions. A zoom of the side-lobes is
plotted in (c), showing how the different input functions complement
each other in terms of frequency content. [Color figure can be viewed
in the online issue, which is available at wileyonlinelibrary.com.]
yielding the net SNR
jHðvÞj
SNRH^ ðvÞ ¼
qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
P
2
j jIj ðvÞj
sðvÞ
:
½10
As seen from Eq. 10, the SNR of the combined measurement is governed by the spectral root-sum-of-squares profile of the set of input functions and by the basic noise
586
Vannesjo et al.
level of the individual measurements. Based on this relationship, the specific choice of input waveforms can be
made according to the desired sensitivity profile. To boost
net sensitivity, certain or all waveforms may be performed
multiple times, yielding an additional SNR gain by
implicit averaging through the summation in Eq. 8.
As described so far, the proposed approach is geared to
determining the self-response of any actuated field component, which may be a gradient field, a shim field, or
else. However, it can readily be extended to incorporate
crossresponses of undesired spatial field patterns as well.
To this end, the magnetic field response in the magnet
bore is expanded into a suitable set of spatial basis functions, which each constitute a separate output channel.
Conversely, each control input of the system forms a separate input channel. For a gradient system, three input
channels are considered, one each for the x, y, and z gradients. On this basis, individual impulse response functions are introduced for each pair of input channel l and
output channel m, yielding the complete description
om ðtÞ ¼
¼
1
X Z
l
X
FT
il ðtÞ hl;m ðt tÞdt $ Om ðvÞ
1
I ðvÞ
l l
Hl;m ðvÞ:
½11
Given the capability to record the different outputs
simultaneously, the multiple-channel impulse response
can be determined simply by repeating the response
measurements for each input channel. The resulting
^ ðlÞ ðvÞ yield
observations O
m;j
^ l;m ðvÞ ¼
H
j Il;j ðvÞ
P
P
j
^ ðlÞ ðvÞ
O
m;j
jIl;j ðvÞj2
½12
with SNR as derived in Eq. 10.
Dynamic Field Camera
Measurements of magnetic field responses were performed
with a third-order dynamic field camera (27), using the field
monitoring methodology described in (29). In short, NMR
probes (28,32) are used to sense the magnetic field dynamics at a set of different positions with high temporal resolution. The actual observable of each probe is the phase of its
NMR signal, which is proportional to the integral of the
magnetic field magnitude at the probe position. Based on
knowledge of the probe positions, the spatiotemporal field
evolution is then expressed in terms of the previously chosen spatial basis functions. The resulting time-varying coefficients, k^m ðtÞ, indicate the observed phase accumulated
due to fields of the respective spatial structure. The timederivative of the phase coefficients gives corresponding
field coefficients, which constitute the observed multiplechannel field output according to Eq. 11:
^m ðtÞ ¼
o
^m ðtÞ
1 dk
;
g dt
½13
where g is the gyromagnetic ratio of the NMR-active nucleus in the probes. To fully determine the chosen spatial
Table 1
Real-Valued Spherical Harmonics Used as Spatial Basis Functions
Nr
Spherical harmonic
0:
1:
2:
3:
4:
5:
6:
7:
8:
9:
10:
11:
12:
13:
14:
15:
1
x
y
z
xy
zy
2z2 (x2 þ y2)
xz
x2 y2
3yx2 y3
xyz
5yz2 y(x2 þ y2 þ z2)
2z3 3z(x2 þ y2)
5xz2 x(x2 þ y2 þ z2)
z(x2 y2)
x3 3xy2
Order
0
1
2
3
model, the number of field probes must be equal to or larger
than that of basis functions and the probes must be placed
appropriately to permit a well-conditioned fit. The temporal
resolution of the measurement is limited only by the acquisition bandwidth of the spectrometer used and the length of
individual field observations is limited by the signal lifetime of the NMR probes.
The dynamic field camera consisted of 16 unshielded
transmit-receive NMR probes based on water samples of
0.8 mm in diameter, permitting field observations up to
k values of 11,000 rad/m (28,29). The water was doped
with 3.3 g/L of CuSO4 to reduce T1 for fast re-excitation,
while keeping T2 long enough to maintain a signal lifetime of about 100 ms. The probes were evenly distributed on the surface of a sphere of 20 cm in diameter.
Their positions were determined by measuring NMR frequency shifts under static gradients of 2.5 mT/m in the
x, y, and z directions, respectively.
GIRF Measurements
In this work, real-valued spherical harmonics up to full
third order were used to represent phase and field
responses (Table 1). In total, this expansion comprises
16 basis functions, matching the number of available
field probes. The single zeroth-order function captures
the uniform component of phase and field, whereas the
three first-order harmonics represent linear gradients in
the x, y, and z directions. For better readability, the first^1 ; k^2 ; k^3 ) and field (^
^2 ; o
^3 ) responses will
o1 ; o
order phase (k
be equivalently referred to by the common notations
kx ; ky ; kz and Gx ; Gy ; Gz .
Measurements were performed on a 3 T Philips Achieva
whole-body system with a maximum gradient strength of
40 mT/m and a slew rate limit of 200 mT/(mms). Gradient
responses were acquired with and without built-in eddycurrent compensation and cross-term correction. The
dwell time for discrete definition of the gradient waveforms
was 6.4 ms. To avoid significant discretization errors, the
shortest triangular gradient pulse used had a time-to-peak
of 50 ms. Such a pulse has the first spectral zero at 20 kHz,
Gradient System Characterization With a Dynamic Field Camera
587
FIG. 2. Magnitude (a, b, c) and phase (d, e, f) of measured GIRFs for all three gradient directions, with and without built-in eddy current
compensation (ECC). The low-pass characteristic of the gradient system is apparent in (a). Details of the responses in (b) show how the
eddy current compensation serves to broaden the response plateau and align the responses of the three different gradient channels. In
a further zoom of the magnitude (c) and phase (f) plots, several mechanical resonances are apparent at low frequencies. Flat phase at
low frequencies reflects near-zero net delay.
where the system response is still non-negligible. Therefore, a combination of different pulse lengths was necessary. Overall, 12 triangular waveforms were used with a
slope of p ¼ 180 mT/(m ms) and time-to-peak T ranging
between 50 and 160 ms at increments of 10 ms (Fig. 1). The
slope was chosen large for high sensitivity while keeping it
well within the system’s specified range. The moment of
the longest pulse was around 1200 rad/m and thus well
within the observable range of the NMR field probes.
In each of the response measurements, signal acquisition
from the field probes was started 5 ms before the nominal
start of the gradient pulse and had a duration of 68 ms,
yielding a frequency resolution of about 14.7 Hz. The primary sampling bandwidth was 398 kHz and thus amply
above the expected response bandwidth. The repetition
time of the acquisitions was 2 s to allow for field settling
between the individual measurements. Each of the 12 triangular inputs was performed 50 times per gradient channel.
This results in a measurement time of 20 min for one gradient channel and a total of 1 h for all three gradient channels. From each individual set of probe signals, field outputs were calculated according to Eq. 13. All field outputs
were then combined to yield the GIRF using Eq. 12. The
whole procedure was performed separately for the x-, y-,
and z-gradient channels. To verify the measured GIRFs,
they were used to predict field responses to select gradient
inputs according to Eq. 11. The predictions were then compared with direct measurements of the responses to the
same gradient inputs. The gradient waveforms used for validation were a set of triangular pulses of 0.05 to 3.2 ms
time-to-peak and 1 to 31 mT/m amplitude, and an echo-planar imaging (EPI) readout sequence (45 echoes, resolution
¼ 2.5 mm, acquisition time ¼ 62.5 ms). To actually probe
the LTI assumption and accuracy of the GIRFs, the lengths
of the triangular validation pulses were different from and
partly much larger than those used for the GIRF measurement. The EPI was included to confirm that the linear
model holds also for a full gradient sequence including
both triangular and trapezoidal pulse shapes.
RESULTS
GIRF Measurements
Magnitude and phase plots of the measured GIRFs in the
frequency domain are shown in Fig. 2. By design, the
gradient chains exhibit overall low-pass characteristics.
The built-in eddy-current compensation broadens the
response plateaus at low frequencies and aligns the
responses of the three gradient subsystems. With compensation, the bandwidth at 3 dB is around 4.6 kHz,
with the response tapering off toward zero at around 22–
25 kHz. As can be seen in the plot, the noise in the GIRF
measurement increases considerably toward higher frequencies, where the power of the input waveforms is
low. The phase response of the x and y gradients around
DC (zero frequency) is essentially flat up to 400 Hz,
with eddy current compensation. The corresponding zgradient phase response exhibits a minor slope at low
frequencies indicating a residual group delay of 0.5 ms
close to DC. The virtual absence of group delay in all
three channels reflects appropriate delay calibration,
which consists in constant shifts of the nominal time of
the gradient chains relative to that of the acquisition
subsystem.
588
FIG. 3. Time-domain representations of the GIRFs in all three gradient directions, with and without eddy current compensation
(ECC). Vertical shifts (D) serve for better comparison of the GIRFs
of the different gradient channels. [Color figure can be viewed in
the online issue, which is available at wileyonlinelibrary.com.]
At low frequencies, the GIRFs show channel-specific patterns of distinct peaks with amplitudes up to 1% of the
full response. The x and y gradients exhibit several such
peaks between 600 and 1800 Hz, while the GIRF of the z
gradient shows mainly one rather narrow peak at 1300 Hz
(Fig. 2). Based on their frequency range, amplitudes, and
widths, these peaks are ascribed to mechanical resonances
of the gradient coils. This interpretation is supported by a
preceding acoustic observation of the single resonance at
1300 Hz (27). Acoustic responses at similar frequencies
have also previously been noted in the literature (4,17).
The prominent GIRF peaks at 25 kHz reflect the transistor switching frequency of the gradient amplifiers.
The switching events cause slight modulation peaks at
the switching frequency. However, these modulation
peaks are not related to the actual input at 25 kHz and
are consequently misinterpreted in the GIRF calculation.
As the input at 25 kHz is extremely small, division
according to Eq. 8 causes the small switching artifact to
appear with large amplitude in the GIRF.
Figure 3 shows plots of the three impulse responses,
with and without eddy current compensation, in the time
domain, filtered with a low-pass filter (raised cosine,
60 kHz full-width half-maximum (FWHM), 12 kHz transition band) to reduce high-frequency noise. The response
kernels rise slightly more quickly than they fall and exhibit somewhat delayed second lobes, which are more
pronounced for the y and z gradients. For the x and y gradients, the peak amplitude of the pre-emphasized
responses is somewhat higher than without the eddy current compensation, whereas for the z gradient, the opposite is observed. From this representation, the width of
the time-domain kernels was estimated at 40 ms with
peaks at around 20 ms. The presence of a gradient
response at t < 0 implies that a gradient field is observed
before the nominal start of the gradient. This occurs as a
result of the MR system’s aforementioned delay correction, which advances the nominal gradient time to play
out gradient waveforms slightly sooner.
Selected cross-responses are shown in Fig. 4, scaled to
percent of the input. To compare field distributions of dif-
Vannesjo et al.
ferent spatial order, the basis functions were normalized
to unit maximum field within a sphere of 20 cm diameter,
centered at gradient isocenter (33). For instance, a linear
field of 10 mT/m amounts to 1 mT in the normalized representation and is thus comparable with a zeroth-order
field of 1 mT as they both produce the same maximum
field shift within the defined sphere. The cross-responses
were generally on the order of 0.1% or less for all three
gradient chains, except at specific resonances and in the
B0 term where it reached 1% for x-gradient input. In Fig.
4, several distinct cross-peaks can be seen at previously
observed mechanical resonance frequencies of the respective gradient channel. In the x response to z-gradient
input, there is a peak at the 1300 Hz resonance of the z
gradient while the B0 and the higher-order 2z2 (x2 þ y2)
(^
o6 ) terms both contain a resonance of the x gradient at
o11 ) term shows
1780 Hz, and the 5yz2 – y(x2 þ y2 þ z2) (^
a resonance of the y gradient at 1700 Hz. The y response
to x-gradient input shows peaks at 1060 Hz and 1750 Hz,
both of which are less pronounced as resonances of the x
gradient, the former however being a strong resonance in
the self-response of the y gradient. Peaks at 1060 Hz
appear in several higher-order field terms as well (data
not shown), however only as responses to x-gradient
inputs and not to the y gradient. A speculative explanation for this appearance of a resonance more strongly
related to the y gradient than the driving x gradient may
lie in the proximity of the two gradients and the shared
structure embedding them both. Switching the x gradient
may induce vibrations that propagate through the surrounding structure and decay more slowly in the y-gradient coil than in the x-gradient coil. The details, however,
of how different vibrational modes are maintained, and
how they affect the field inside the scanner remain to be
investigated. The small distinct peaks seen at DC in several of the cross-terms largely stem from slow system
drifts unrelated to gradient inputs as well as residual susceptibility broadening of the field probe samples (28),
which mimics slight field variation at very low
frequencies.
GIRF-Based Predictions
Figure 5 shows predictions of triangular gradient pulses
with a time-to-peak T of 0.1, 0.4, and 1.05 ms, and peak
amplitudes of 1 (Fig. 5a) and 31 mT/m (Fig. 5c), compared with nominal and directly measured (filtered to 60
kHz FWHM) pulses. Error plots of nominal vs. measured,
and predicted vs. measured gradients are plotted for each
pulse (Fig. 5b,d). For shorter pulses, the actual gradient
deviates more from the nominal waveform than for longer
ones, which is expected due to the weaker response at
higher frequencies. The directly measured pulses follow
the predictions closely with a maximal prediction error of
about 0.2% of the pulse amplitude. Due to different
dynamic ranges, the measurement noise appears more
prominently in the upper plot. Increased noise toward the
end of the 31 mT/m, 1.05 ms pulse, as seen in the error
plot (Fig. 5d), was caused by nearly complete dephasing
of the field probes in the validation measurement. The
size of the NMR field probes thus limits the capability of
validating predictions of large gradient pulses by single
Gradient System Characterization With a Dynamic Field Camera
589
FIG. 4. Measured cross-responses of the x, y, z, B0, 2z2 - (x2 þ y2) and 5yz2 – y(x2 þ y2 þ z2) field terms due to operating the x- (blue),
y- (green), and z-gradients (red), expressed in percent of the input. Several distinct peaks coincide with mechanical resonances previously identified in the self-responses (Fig. 2). Note the different scaling in the B0 plot.
direct measurements. It does not, however, hinder GIRFbased predictions of large gradients per se.
To study subtle long-term features of the gradient
response, Fig. 6a shows a detail of the first 15 ms after a
gradient pulse (30 mT/m amplitude, T ¼ 0.25 ms). In the
prediction, a distinct oscillatory pattern is seen long after
the pulse, reflecting the subtle resonance peaks in the
underlying GIRF. The direct measurement, however, was
too noisy to confirm the oscillations on this small scale.
To see them more clearly, Fig. 6b shows the phase accumulated under the same gradient pulse. The underlying
integration suppresses high-frequency noise and thus
reveals the oscillations also in the direct measurement. It
is now apparent that the predicted and the measured
patterns again agree closely.
The accumulated phase, expressed as the k-space position, is also the key parameter underlying MR image
encoding and reconstruction. In Fig. 7, the measured
GIRFs are used to predict a whole EPI readout sequence
in terms of its effective gradient waveforms (left) and kspace trajectory (right). The directly measured trajectory
again shows good accordance with the prediction, following it closely throughout, including the turns where
it deviates significantly from the nominal trajectory.
DISCUSSION
The impulse response function of a linear time-invariant
system is a well known and much used concept in engineering. Herein, we have investigated the feasibility of
determining the impulse response functions of the gradient chains of an MR system. The investigation concerned
each gradient chain as a whole between the two endpoints that are relevant for the user, i.e., between gradient waveform definition at the console and the actual
magnetic fields seen by an object placed in the scanner
bore. Using a third-order dynamic field camera based on
NMR probes and a set of input functions that are easily
implemented in standard scanner software, the impulse
response measurement has proven a straightforward and
fast procedure, yielding highly accurate results.
When defined as above, GIRFs constitute a composite
response reflecting all linear distortions that a gradient
waveform undergoes between the software interface of
the scanner and the resulting magnetic fields. This
includes software signal processing, eddy-current compensation, gradient amplifier characteristics, potential
cable effects, coil characteristics, coil coupling, eddy currents, and mechanical responses of the gradient system.
590
Vannesjo et al.
stage of a gradient chain acts on the waveform, the input
and output of that stage would need to be measured.
Measuring the combined response, as done here, is a fast
and comprehensive way of characterizing how a gradient
system performs as one entity.
Applications
Knowledge of the gradient impulse response can be beneficial in a variety of ways. Most immediately, it forms a
comprehensive basis of tuning the gradient response by
pre-emphasis. Existing methods of eddy current compensation commonly rely on a limited number of exponentially
decaying correction terms found by time-domain fitting.
With the GIRFs, corresponding fits can be based on a complete picture of the native response in the frequency domain (34), including correction for mechanically induced
field oscillations. Knowledge of the GIRF also permits preemphasis by a general filter equal to the frequency-domain
ratio of the desired system response and the GIRF. In this
fashion, LTI theory can be deployed to shape the net system response comprehensively within the technical limitations of the available gradient hardware.
More generally, fast GIRF measurement may be useful
for the calibration and quality assurance of gradient systems and as a diagnostic tool in cases of suboptimal gradient performance. As shown in this work, known GIRFs
can also be used to calculate the actual gradient output
that is obtained with any given pulse sequence. Given
sufficient system stability, image reconstruction may thus
be based on actual k-space trajectories and higher-order
encoding (33) without the need for sequence-by-sequence
field monitoring. Calculated field evolutions may also
yield more accurate accounts of specialized signal preparation such as, e.g., in phase-contrast methods (35). In
some situations, retrospective correction for field imperfections is not or hardly feasible. This is true, for instance,
for spatially selective excitation based on k-space trajectories played out during radiofrequency transmission. In
these cases, GIRF-based predictions could serve as a basis
for designing trajectories such that they can be realized
perfectly with a suitable precompensated demand.
Model Limitations
FIG. 5. GIRF-based predictions of gradient blips of 1 mT/m peak
amplitude, T ¼ 0.1, 0.4, and 1.05 ms (a) and 31 mT/m peak amplitude,
T ¼ 0.4 and 1.05 ms (c), compared with nominal and directly measured gradient waveforms. For each pulse, the differences between
nominal and measured (black) as well as predicted and measured
(blue) traces are plotted in (b, d). The prediction error amounts to maximally 0.2% of the gradient amplitude. [Color figure can be viewed in
the online issue, which is available at wileyonlinelibrary.com.]
Several features of the measured GIRFs have been linked
to underlying phenomena. To fully understand how each
The LTI approach is only valid to the extent that the system considered is actually LTI. Although typically
minor, there are a number of influences on gradient
responses that are nonlinear or vary over time. Power
amplifiers generally exhibit nonlinear behavior to varying extent and procedures to linearize the output are a
standard part of gradient amplifier engineering. Gradient
responses also depend somewhat on the thermal state of
the system. High-gradient-duty-cycle scans, such as for
fMRI or diffusion imaging, heat up the gradient coils and
supporting structures. Ensuing changes in electrical and
mechanical properties will cause the gradient response
to change slightly over time. Likewise, any other structures that support eddy currents may slightly heat up
and thus change their eddy current characteristics. In as
far as the heating is caused by driving the gradients
themselves, it could be seen as an extension to the
Gradient System Characterization With a Dynamic Field Camera
591
FIG. 6. Nominal, measured, and predicted waveforms of a triangular gradient pulse (30 mT/m peak amplitude, T ¼ 0.25 ms) in field (top)
and accumulated-phase (bottom) representation. The zoomed insets on the right show details of the system response over about 15 ms after the pulse. As seen there, the GIRF predicts subtle oscillations after the pulse, which cannot be distinguished from noise in the direct field
measurement. They are largely confirmed, however, by plots of the corresponding phase coefficient, ky , in which high-frequency noise is
attenuated by integration over time. [Color figure can be viewed in the online issue, which is available at wileyonlinelibrary.com.]
response of the system, which will however generally be
nonlinear. There may also be external factors that influence the net field independently of how the system is
driven and are thus inherently nonlinear. Any perturbation from the environment, such as fields produced by
other electrical equipment, nearby traffic or power lines
fall into this class. Nonetheless, as long as these model
violations are suitably small, the LTI approach permits
useful approximations of the system behavior. In this
work, the validity of measured GIRFs and thus of the
underlying LTI picture has been demonstrated by predictions of effective gradient waveforms. For a variety of
gradient inputs, a high degree of prediction accuracy
was achieved with a maximal error on the order of 0.2%
of the pulse amplitude. Moreover, preliminary data indicate good stability of the system response over years.
mechanical displacements in the system translate into
field changes. Such studies are beyond the scope of this
work. However, the observed agreements in vibration
peaks between self-, cross-, and higher-order responses
indicate that GIRF measurements will be instrumental in
elucidating these mechanisms. Cross-responses via eddy
currents are pronounced with unshielded gradient or
shim coils, which couple strongly to the cryostat. In this
situation, measurements of the cross-responses could
form the basis of cross pre-emphasis and higher-order
image reconstruction (33). In this work, a field camera
with 16 NMR probes permitted GIRF measurements of full
third order in terms of spherical harmonics. Further discrimination of field responses up to arbitrary spatial order
could be readily achieved provided a suitable number and
distribution of field probes.
Mechanical Oscillations and Cross-terms
Sensitivity and Resolution
Among the features that the measured GIRFs capture and
predict well are field oscillations due to mechanical vibrations. Judging by the achieved prediction accuracy, such
oscillations seem to behave largely linearly. Understanding these effects in more detail will require models of how
The sensitivity of the GIRF measurement has been
shown to depend on the set of input functions used and
on the noise level of individual field measurements. Triangular pulses offer the highest spectral sensitivity close
to DC. Toward higher frequencies, their net sensitivity
592
Vannesjo et al.
FIG. 7. Nominal, measured, and
predicted gradient time-courses
during an EPI readout sequence
(left). Zoomed details show one
of the phase-encoding blips and
the settling of a frequencyencoding gradient (bottom left).
The resulting full k-space trajectories are shown on the right,
including a detail of one of the kspace corners including the end
of the prephaser gradient. [Color
figure can be viewed in the
online issue, which is available
at wileyonlinelibrary.com.]
drops approximately as 1=v2 as seen from Eq. 5. The
transition from phase data to field data further emphasizes high-frequency noise since taking the temporal
derivative corresponds to multiplication by x in the
frequency domain. Jointly, these effects caused the
GIRF sensitivity to drop as 1=v3 overall.
A straightforward way of increasing the overall SNR is
to acquire averages of the field response measurements.
To specifically increase sensitivity at higher frequencies,
sinuisodally modulated pulses could be used in principle,
shifting the spectral main lobe away from DC. To keep
with slew rate limitations, however, the slope of a pulse
to be modulated with a sinusoid would need to be kept
lower than the system maximum, thereby reducing the
total power of the pulse. For this reason, the sensitivity of
a single measurement at a high frequency will necessarily
be lower than what can be achieved around DC with the
corresponding unmodulated pulse. The optimal choice of
input pulses ultimately depends on the desired sensitivity
profile for the system to be measured. Equations 8 and 12
describe how to obtain a least-squares estimate of the
GIRFs based on the sensitivity profile of each input pulse,
and hold for any combination of modulated and unmodulated pulses alike. Combining simple triangular pulses as
done here offers the advantage of straightforward implementation on standard MR equipment.
In this work, the gradient response measurements covered 68 ms, yielding a frequency resolution of 14.7 Hz.
Particularly for unshielded gradient coils, it is known
that significant eddy currents may persist for several
hundreds of milliseconds (9), thus exceeding the range
possible to capture in one single measurement with the
field camera used in this work. To overcome this, multiple response measurements with interleaved timing can
be acquired (34), which however comes at the expense
of increasing the total time required for the measurements. With hardware able to perform continuous field
monitoring, as described by (36), long-living responses
could be captured straightforwardly.
Measurement Time
The time required for measuring GIRFs with the proposed
approach depends on the total number of input pulses
and the time allowed for gradient settling between pulses.
The aim of the measurements presented here was to elucidate even subtle behavior of the gradient responses, to
which end a large number of averages was acquired. For
three gradient channels, the total measurement time
amounted to 1 h. For a coarser measurement, even one
observation of each pulse would suffice, reducing the
measurement time to about 1 min. Initial experience suggests that even such coarser GIRFs yield useful predictions of gradients and trajectories. This is largely due to
the fact that common input gradient waveforms have
most power at relatively low frequencies, where the SNR
of the measured GIRF is favorable.
CONCLUSION
GIRF measurements with a third-order dynamic field
camera enable fast and easy determination of gradient
system characteristics. Among others, GIRFs measured in
Gradient System Characterization With a Dynamic Field Camera
593
this way readily reveal subtle field oscillations due to
mechanical resonances. These have been found to be
largely linear and time-independent in experiments
without significant gradient heating. GIRFs also permit
highly accurate predictions of net system behavior. Such
predictions will be useful for image reconstruction and
gradient waveform optimization. Notably, suitable sets of
field probes straightforwardly yield GIRFs up to high
spatial order along with the common gradient responses.
This capability renders the proposed approach promising
also for the characterization of dynamic shim systems.
The GIRF concept assumes linearity and time-independence of the system to be characterized but is otherwise
free of mechanistic assumptions. For more advanced
models incorporating, e.g., thermal or nonlinear electromechanical effects, response measurements with a
dynamic field camera will still be instrumental.
14. Barnett A. Comments on ‘‘Gradient-induced acoustic and magnetic
field fluctuations in a 4 T whole-body MR imager.’’ Magn Reson Med
2001;46:207.
15. Nixon TW, McIntyre S, Rothman DL, de Graaf RA. Compensation of
gradient-induced magnetic field perturbations. J Magn Reson 2008;
192:209–217.
16. Hedeen RA, Edelstein WA. Characterization and prediction of gradient acoustic noise in MR imagers. Magn Reson Med 1997;37:7–10.
17. Li W, Mechefske CK, Gazdzinski C, Rutt BK. Acoustic noise analysis
and prediction in a 4-T MRI scanner. Concept Magn Res 2004;21B:
19–25.
18. Sierra CVR, Versluis MJ, Hoogduin JM, Duifhuis H. Acoustic fMRI
noise: linear time-invariant system model. IEEE Trans Biomed Eng
2008;55:2115–2123.
19. Addy NO, Wu HH, Nishimura DG. Simple Method for MR Gradient
System Characterization. In Proceedings of the 17th Annual Meeting
of ISMRM, Honolulu, Hawaii, USA, 2009. p. 3068.
20. Vannesj€
o SJ, H€
aberlin M, Kasper L, Barmet C, Pruessmann KP. A
Method for Characterizing the Magnetic Field Response of a Gradient
System. In Proceedings of the 18th Annual Meeting of ISMRM,
Stockholm, Sweden, 2010. p. 1536.
21. Wysong RE, Lowe IJ. A simple method of measuring gradient
induced eddy currents to set compensation networks. Magn Reson
Med 1993;29:119–121.
22. Terpstra M, Andersen PM, Gruetter R. Localized eddy current compensation using quantitative field mapping. J Magn Reson 1998;131:
139–143.
23. Duyn JH, Yang YH, Frank JA, van der Veen JW. Simple correction
method for k-space trajectory deviations in MRI. J Magn Reson 1998;
132:150–153.
24. Alley MT, Glover GH, Pelc NJ. Gradient characterization using a
Fourier-transform technique. Magn Reson Med 1998;39:581–587.
25. Goodyear DJ, Shea M, Beyea SD, Shah NJ, Balcom BJ. Single point
measurements of magnetic field gradient waveform. J Magn Reson
2003;163:1–7.
26. Balcom BJ, Bogdan M, Armstrong RL. Single-point imaging of gradient rise, stabilization, and decay. J Magn Reson Ser A 1996;118:
122–125.
27. Barmet C, Wilm BJ, Pavan M, Pruessmann KP. A Third-Order Field
Camera with Microsecond Resolution for MR System Diagnostics. In
Proceedings of the 17th Annual Meeting of ISMRM, Honolulu,
Hawaii, USA, 2009. p. 781.
28. De Zanche N, Barmet C, Nordmeyer-Massner JA, Pruessmann KP.
NMR probes for measuring magnetic fields and field dynamics in
MR systems. Magn Reson Med 2008;60:176–186.
29. Barmet C, De Zanche N, Pruessmann KP. Spatiotemporal magnetic
field monitoring for MR. Magn Reson Med 2008;60:187–197.
30. Barmet C, Wilm BJ, Pavan M, Katsikatsos G, Keupp J, Mens G,
Pruessmann KP. Concurrent Higher-Order Field Monitoring for Routine Head MRI: An Integrated Heteronuclear Setup. In Proceedings of
the 18th Annual Meeting of ISMRM, Stockholm, Sweden, 2010.
p. 216.
31. Lathi BP. Signal processing and linear systems, Indian Edition. New
York: Oxford University Press; 2008.
32. Barmet C, De Zanche N, Wilm BJ, Pruessmann KP. A transmit/
receive system for magnetic field monitoring of in vivo MRI. Magn
Reson Med 2009;62:269–276.
33. Wilm BJ, Barmet C, Pavan M, Pruessmann KP. Higher order reconstruction for MRI in the presence of spatiotemporal field perturbations. Magn Reson Med 2011;65:1690–1701.
34. Vannesj€
o J, Fillmer A, Barmet C, Boesiger P, Henning A, Pruessmann
KP. Fast Characterization of Higher-Order Shim Dynamics by
Impulse Response Measurements with a Dynamic Field Camera. In
Proceedings of the 19th Annual Meeting of ISMRM, Montreal, Canada, 2011. p. 719.
35. Giese D, Haeberlin M, Barmet C, Pruessmann KP, Schaeffter T,
Kozerke S. Analysis and correction of background velocity offsets in
phase-contrast flow measurements using magnetic field monitoring.
Magn Reson Med 2012;67:1294–1302.
36. Dietrich BE, Barmet C, Brunner D, Pruessmann KP. An Autonomous
System for Continuous Field Monitoring with Interleaved Probe Sets.
In Proceedings of the 19th Annual Meeting of ISMRM, Montreal,
Canada, 2011. p. 1842.
ACKNOWLEDGMENTS
The authors thank Geran Peeren at Philips Healthcare, as
well as Frederik Testud at the University of Freiburg, for
reading and providing valuable comments on this work.
Technical support from Philips Healthcare is gratefully
acknowledged.
REFERENCES
1. Boesch C, Gruetter R, Martin E. Temporal and spatial analysis of
fields generated by eddy currents in superconducting magnets: optimization of corrections and quantitative characterization of magnet/
gradient systems. Magn Reson Med 1991;20:268–284.
2. Liu Q, Hughes DG, Allen PS. Quantitative characterization of the
eddy current fields in a 40-cm bore superconducting magnet. Magn
Reson Med 1994;31:73–76.
3. Wu YH, Chronik BA, Bowen C, Mechefske CK, Rutt BK. Gradientinduced acoustic and magnetic field fluctuations in a 4 T wholebody MR imager. Magn Reson Med 2000;44:532–536.
4. Clayton DB, Elliott MA, Leigh JS, Lenkinski RE. 1H Spectroscopy
without solvent suppression: characterization of signal modulations
at short echo times. J Magn Reson 2001;153:203–209.
5. Foerster BU, Tomasi D, Caparelli EC. Magnetic field shift due to mechanical vibration in functional magnetic resonance imaging. Magn
Reson Med 2005;54:1261–1267.
6. Van de Moortele P, Pfeuffer J, Glover GH, Ugurbil K, Hu X. Respiration-induced B0 fluctuations and their spatial distribution in the
human brain at 7 Tesla. Magn Reson Med 2002;47:888–895.
7. Versluis MJ, Peeters JM, van Rooden S, van der Grond J, van Buchem
MA, Webb AG, van Osch MJP. Origin and reduction of motion and
f0 artifacts in high resolution T2*-weighted magnetic resonance
imaging: application in Alzheimer’s disease patients. NeuroImage
2010;51:1082–1088.
8. Jehenson P, Westphal M, Schuff N. Analytical method for the compensation of eddy-current effects induced by pulsed magnetic field
gradients in NMR systems. J Magn Reson 1990;90:264–278.
9. van Vaals JJ, Bergman AH. Optimization of eddy-current compensation. J Magn Reson 1990;90:52–70.
10. Wysong RE, Madio DP, Lowe IJ. A novel eddy current compensation
scheme for pulsed gradient systems. Magn Reson Med 1994;31:572–575.
11. Jezzard P, Barnett AS, Pierpaoli C. Characterization of and correction
for eddy current artifacts in echo planar diffusion imaging. Magn
Reson Med 1998;39:801–812.
12. Zur Y, Stokar S. An algorithm for eddy currents symmetrization and
compensation. Magn Reson Med 1996;35:252–260.
13. Eccles CD, Crozier S, Westphal M, Doddrell DM. Temporal spherical-harmonic expansion and compensation of eddy-current fields
produced by gradient pulses. J Magn Reson Ser A 1993;103:
135–141.