Abstract
Background
The application of image-guided surgery (IGS) to laparoscopic liver resection and ablation is currently limited, but it would assist in intraoperative decision making regarding oncologic margins, ablation probe placement, and ablation tracking.
Methods
Eight spherical surface targets on a liver phantom were imaged with an optically tracked laparoscopic ultrasound (US) probe. Ten US images of each target were registered to computer tomography (CT) images of the phantoms and then mapped to the CT scans. Accuracy of the registration was assessed by comparing the distance between the predicted target location and the position obtained directly from CT.
Results
The average localization error was 5.3 mm. The errors resulted primarily from inaccurate US probe tracking but were otherwise insensitive to the variability that arises from manually identifying targets in US and CT images.
Conclusions
The results obtained for US-to-CT registration in a phantom model suggest that further investigations into its clinical use are warranted and that other IGS technologies could be applied to laparoscopic liver surgery as well.
Similar content being viewed by others
References
MY Abdel-Atty O Farges P Jagot J Belghiti (1999) ArticleTitleLaparoscopy extends the indications for liver resection in patients with cirrhosis Br J Surg 86 1397–1400 Occurrence Handle10.1046/j.1365-2168.1999.01283.x Occurrence Handle1:STN:280:DC%2BD3c%2FltFWrsA%3D%3D Occurrence Handle10583285
P Bao JR Warmath B Poulose RL Galloway SuffixJr AJ Herline (2004) ArticleTitleTracked ultrasound for laparoscopic surgery Proc SPIE 5367 237–246 Occurrence Handlefull_text||10.1117/12.535429
RA Beasley JD Stefansic JL Herring WA Bass AJ Herline WC Chapman BM Dawant et al. (2000) ArticleTitleImplementation and incorporation of liver 3-D surface renderings into interactive, image-guided hepatic surgery Proc SPIE 3976 282–289 Occurrence Handlefull_text||10.1117/12.383050
D Cherqui E Husson R Hammoud B Malassagne F Stephan S Bensaid N Rotman et al. (2000) ArticleTitleLaparoscopic liver resections: a feasibility study in 30 patients Ann Surg 232 753–762
MH Chung TF Wood GJ Tsioulias DM Rose AJ Bilchik (2001) ArticleTitleLaparoscopic radiofrequency ablation of unresectable hepatic malignancies: a phase 2 trial Surg Endosc 15 1020–1026 Occurrence Handle10.1007/s00464-001-9000-2 Occurrence Handle1:STN:280:DC%2BD3MrlvVCmsw%3D%3D Occurrence Handle11443478
JM Fitzpatrick JB West CR Maurer SuffixJr (1998) ArticleTitlePredicting error in rigid-body point-based registration IEEE Trans Med Imag 17 694–702 Occurrence Handle10.1109/42.736021 Occurrence Handle1:STN:280:DyaK1M%2FovFKruw%3D%3D
RL Galloway SuffixJr RJ Maciunas CA Edwards (1992) ArticleTitleInteractive image guided neurosurgery IEEE Tra Biomed Eng 39 1226–1231 Occurrence Handle10.1109/10.184698
JF Gigot D Glineur J Santiago Azagra M Goergen M Camrick M Morino J Etienne et al. (2002) ArticleTitleLaparoscopic liver resection for malignant liver tumors: preliminary results of a multicenter European study Ann Surg 236 90–97
A Herline JD Stefansic J Debelak RL Galloway WC Chapman (2000) ArticleTitleTechnical advances toward interactive image-guided laparoscopic surgery Surg Endosc 14 675–679 Occurrence Handle10.1007/s004640000023 Occurrence Handle1:STN:280:DC%2BD3M7gvFWltg%3D%3D Occurrence Handle10948308
AJ Herline JD Stefansic JP Debelak SL Hartmann CW Pinson RL Galloway WC Chapman (1999) ArticleTitleImages-guided surgery: preliminary feasibility studies of frameless stereotactic liver surgery Arch Surg 134 644–650 Occurrence Handle10.1001/archsurg.134.6.644 Occurrence Handle1:STN:280:DyaK1M3psVaqtw%3D%3D Occurrence Handle10367875
DL Hill P Batchelor (2001) Registration methodology: concepts and algorithms JV Hajnal DL Hill DJ Hawkes (Eds) Medical image registration CRC Press New York 39–70
M Lesurtel D Cherqui A Laurent C Tayar PL Fagniez (2003) ArticleTitleLaparoscopic versus open left lateral hepatic lobectomy: a case-control study J Am Coll Surg 196 236–242 Occurrence Handle10.1016/S1072-7515(02)01622-8 Occurrence Handle12595052
DM Muratore JH Russ BM Dawant RL Galloway SuffixJr (2002) ArticleTitleThree-dimensional image registration of phantom vertebrae for image-guided surgery: a preliminary study Comput Aided Surg 7 342–352 Occurrence Handle10.1002/igs.10055 Occurrence Handle12731097
N Pagoulatos WS Edwards DR Haynor Y Kim (1999) ArticleTitleInteractive 3-D registration of ultrasound and magnetic resonance images based on a magnetic position sensor IEEE Trans Inf Technol Biomed 3 278–288 Occurrence Handle10.1109/4233.809172 Occurrence Handle1:STN:280:DC%2BD3c7osVeqsA%3D%3D Occurrence Handle10719478
GP Penney JM Blackall MS Hamady T Sabharwal A Adam DJ Hawkes (2004) ArticleTitleRegistration of freehand 3D ultrasound and magnetic resonance liver images Med Image Anal 8 81–91 Occurrence Handle10.1016/j.media.2003.07.003 Occurrence Handle1:STN:280:DC%2BD3srntValsA%3D%3D Occurrence Handle14644148
A Siperstein A Garland K Engle S Rogers E Berber A Foroutani A String et al. (2000) ArticleTitleLocal recurrence after laparoscopic radiofrequency thermal ablation of hepatic tumors Ann Surg Oncol 7 106–113 Occurrence Handle10.1007/s10434-000-0106-x Occurrence Handle1:STN:280:DC%2BD3c3it1Chtw%3D%3D Occurrence Handle10761788
JD Stefansic AJ Herline Y Shyr WC Chapman JM Fitzpatrick BM Dawant RL Galloway SuffixJr (2000) ArticleTitleRegistration of physical space to laparoscopic image space for use in minimally invasive hepatic surgery IEEE Trans Med Imaging 19 1012–1023 Occurrence Handle10.1109/42.887616 Occurrence Handle1:STN:280:DC%2BD3M7it1OksA%3D%3D Occurrence Handle11131491
S Umeyama (1991) ArticleTitleLeast-squares estimation of transformation parameters between two point patterns IEEE Trans Pattern Anal Mach Intell 13 376–380 Occurrence Handle10.1109/34.88573
TF Wood DM Rose M Chung DP Allegra LJ Foshag AJ Bilchik (2000) ArticleTitleRadiofrequency ablation of 231 unresectable hepatic tumors: indications, limitations, and complications Ann Surg Oncol 7 593–600 Occurrence Handle1:STN:280:DC%2BD3cvltVGktg%3D%3D Occurrence Handle11005558
Acknowledgments
We thank the Vanderbilt Physician Scientist Development Program, the American Society of Colon and Rectal Surgeons Career Development Award, B&K Medical, members of the Surgical Navigation Apparatus Research Lab (SNARL), and the CT technicians of the Vanderbilt Department of Radiology.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Bao, P., Warmath, J., Galloway, R. et al. Ultrasound-to-computer-tomography registration for image-guided laparoscopic liver surgery. Surg Endosc 19, 424–429 (2005). https://doi.org/10.1007/s00464-004-8902-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-004-8902-1