Abstract
Objective
This study aimed to compare the signal-to-noise ratios (SNRs) and apparent diffusion coefficients (ADCs) obtained using two fat suppression techniques in breast diffusion-weighted imaging (DWI) of a phantom.
Materials and methods
The breast phantom comprised agar gels with four different concentrations of granulated sugar (samples 1, 2, 3, and 4). DWI with short tau inversion recovery (STIR-DWI) and that with spectral attenuated inversion recovery (SPAIR-DWI) were performed using 3.0-T magnetic resonance imaging, and the obtained SNRs and ADCs were compared. ADCs were also compared between the right and left breast phantoms.
Results
For samples 3 and 4, SNRs obtained using STIR-DWI were lower than those obtained using SPAIR-DWI. For samples 2, 3, and 4, overall ADCs obtained using STIR-DWI were significantly higher than those obtained using SPAIR-DWI (p < 0.001 for all), although no significant difference was observed for sample 1 (p = 0.62). STIR-DWI shows a positive bias and wide limits of agreement in Bland–Altman plot. The coefficients of variance of overall ADCs were good in STIR-DWI and SPAIR-DWI. For all samples, STIR-DWI demonstrated slightly larger percentage differences in ADCs between the right and left phantoms than SPAIR-DWI.
Conclusion
SNRs and ADCs obtained using STIR-DWI are influenced by the T 1 value; a shorter T 1 value decreases SNRs, overestimates ADCs, and induces the measurement error in ADCs. STIR-DWI showed a larger difference in ADCs between the right and left phantoms than SPAIR-DWI.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.References
Yabuuchi H, Matsuo Y, Okafuji T, Kamitani T, Soeda H, Setoguchi T, Sakai S, Hatakenaka M, Kubo M, Sadanaga N, Yamamoto H (2009) Enhanced mass on contrast-enhanced breast MR imaging: lesion characterization using combination of dynamic contrast-enhanced and diffusion-weighted MR images. J Magn Reson Imaging 28:1157–1165
Zhang L, Tang M, Min Z, Lu J, Lei X, Zhang X (2016) Accuracy of combined dynamic contrast-enhanced magnetic resonance imaging and diffusion-weighted imaging for breast cancer detection: a meta-analysis. Acta Radiol 57:651–660
Lo GG, Ai V, Chan JK, Li KW, Cheung PS, Wong TT, Ma M, Lee R, Chien D (2009) Diffusion-weighted magnetic resonance imaging of breast lesions: first experiences at 3 T. J Comput Assist Tomogr 33:63–69
Guo Y, Cai YQ, Cai ZL, Gao YG, An NY, Ma L, Mahankali S, Gao JH (2002) Differentiation of clinically benign and malignant breast lesions using diffusion-weighted imaging. J Magn Reson Imaging 16:172–178
Nogueira L, Brandao S, Matos E, Nunes RG, Ferreira HA, Loureiro J, Ramos I (2014) Diffusion-weighted breast imaging at 3 T: preliminary experience. Clin Radiol 69:378–384
Richard R, Thomassin I, Chapellier M, Scemama A, de Cremoux P, Varna M, Giacchetti S, Espié M, de Kerviler E, de Bazelaire C (2013) Diffusion-weighted MRI in pretreatment prediction of response to neoadjuvant chemotherapy in patients with breast cancer. Eur Radiol 23:2420–2431
Subhawong TK, Jacobs MA, Fayad LM (2014) Diffusion-weighted MR imaging for characterizing musculoskeletal lesions. Radiographics 34:1163–1177
Delfaut EM, Beltran J, Johnson G, Rousseau J, Marchandise X, Cotten A (1999) Fat suppression in MR imaging: techniques and pitfalls. Radiographics 19:373–382
Lin C, Rogers CD, Majidi S (2015) Fat suppression techniques in breast magnetic resonance imaging: a critical comparison and state of the art. Rep Med Imaging 2015:37–49
Mürtz P, Tsesarskiy M, Kowal A, Träber F, Gieseke J, Willinek WA, Leutner CC, Schmiedel A, Schild HH (2014) Diffusion-weighted magnetic resonance imaging of breast lesions: the influence of different fat-suppression techniques on quantitative measurements and their reproducibility. Eur Radiol 24:2540–2551
Stadlbauer A, Bernt R, Gruber S, Bogner W, Pinker K, van der Riet W, Haller J, Salomonowitz E (2009) Diffusion-weighted MR imaging with background body signal suppression (DWIBS) for the diagnosis of malignant and benign breast lesions. Eur Radiol 19:2349–2356
Ouyang Z, Ouyang Y, Zhu M, Lu Y, Zhang Z, Shi J, Li X, Ren G (2014) Diffusion-weighted imaging with fat suppression using short-tau inversion recovery: clinical utility for diagnosis of breast lesions. Clin Radiol 69:e337–e344
Nogueira L, Brandão S, Nunes RG, Ferreira HA, Loureiro J, Ramos I (2015) Breast DWI at 3 T: influence of the fat-suppression technique on image quality and diagnostic performance. Clin Radiol 70:286–294
Brandão S, Nogueira L, Matos E, Nunes RG, Ferreira HA, Loureiro J, Ramos I (2015) Fat suppression techniques (STIR vs. SPAIR) on diffusion-weighted imaging of breast lesions at 3.0 T: preliminary experience. Radiol Med 120:705–713
Norris DG (2001) Implications of bulk motion for diffusion-weighted imaging experiments: effects, mechanisms, and solutions. J Magn Reson Imaging 13:486–495
Yoshida T, Urikura A, Shirata K, Nakaya Y, Terashima S, Hosokawa Y (2016) Image quality assessment of single-shot turbo spin echo diffusion-weighted imaging with parallel imaging technique: a phantom study. Br J Radiol 89:20160512
Celik A (2016) Effect of imaging parameters on the accuracy of apparent diffusion coefficient and optimization strategies. Diagn Interv Radiol 22:101–107
Bogner W, Gruber S, Pinker K, Grabner G, Stadlbauer A, Weber M, Moser E (2009) Diffusion-weighted MR for differentiation of breast lesions at 3.0 T: how does selection of diffusion protocols affect diagnosis? Radiology 253:341–351
Malyarenko D, Galbán CJ, Londy FJ, Meyer CR, Johnson TD, Rehemtulla A, Ross BD, Chenevert TL (2013) Multi-system repeatability and reproducibility of apparent diffusion coefficient measurement using an ice-water phantom. J Magn Reson Imaging 37:1238–1246
Hancu I, Govenkar A, Lenkinski RE, Lee SK (2013) On shimming approaches in 3T breast MRI. Magn Reson Med 69:862–867
Price RR, Axel L, Morgan T, Newman R, Perman W, Schneiders N, Selikson M, Wood M, Thomas SR (1990) Quality assurance methods and phantoms for magnetic resonance imaging: report of AAPM nuclear magnetic resonance Task Group No. 1. Med Phys 17:287–295
Bland JM, Altman D (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 327:307–310
Thomassin-Naggara I, De Bazelaire C, Chopier J, Bazot M, Marsault C, Trop I (2013) Diffusion-weighted MR imaging of the breast: advantages and pitfalls. Eur J Radiol 82:435–443
Pereira FP, Martins G, de Oliveira RD (2011) Diffusion magnetic resonance imaging of the breast. Magn Reson Imaging Clin N Am 19:95–110
Moschetta M, Telegrafo M, Rella L, Ianora AA, Angelelli G (2014) Effect of gadolinium injection on diffusion-weighted imaging with background body signal suppression (DWIBS) imaging of breast lesions. Magn Reson Imaging 32:1242–1246
Edden RA, Smith SA, Barker PB (2010) Longitudinal and multi-echo transverse relaxation times of normal breast tissue at 3 Tesla. J Magn Reson Imaging 32:982–987
Giannotti E, Waugh S, Priba L, Davis Z, Crowe E, Vinnicombe S (2015) Assessment and quantification of sources of variability in breast apparent diffusion coefficient (ADC) measurements at diffusion weighted imaging. Eur J Radiol 84:1729–1736
Zhao J, Guan H, Li M, Gu H, Qin J, Wu X (2016) Significance of the ADC ratio in the differential diagnosis of breast lesions. Acta Radiol 57:422–429
Funding
There is no funding.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have conflict of interest.
Ethical standards
This article does not contain any studies with human participants or animals performed by any of the authors.
Rights and permissions
About this article
Cite this article
Yoshida, T., Urikura, A., Shirata, K. et al. Short tau inversion recovery in breast diffusion-weighted imaging: signal-to-noise ratio and apparent diffusion coefficients using a breast phantom in comparison with spectral attenuated inversion recovery. Radiol med 123, 296–304 (2018). https://doi.org/10.1007/s11547-017-0840-9
Received:
Accepted:
Published:
Issue Date: