Abstract
Free full text
Modelling secondary lymphatic valves with a flexible vessel wall: how geometry and material properties combine to provide function
Abstract
A three-dimensional finite-element fluid/structure-interaction model of an intravascular lymphatic valve was constructed, and its properties investigated under both favourable and adverse pressure differences, simulating valve opening and valve closure respectively. The shear modulus of the neo-Hookean material of both vascular wall and valve leaflet was varied, as was the degree of valve opening at rest. Also investigated was how the valve characteristics were affected by prior application of pressure inflating the whole valve. The characteristics were parameterised by the volume flow-rate through the valve, the hydraulic resistance to flow, and the maximum sinus radius and inter-leaflet-tip gap on the plane of symmetry bisecting the leaflet, all as functions of the applied pressure difference. Maximum sinus radius on the leaflet-bisection plane increased with increasing pressure applied to either end of the valve segment, but also reflected the non-circular deformation of the sinus cross-section caused by the leaflet, such that it passed through a minimum at small favourable pressure differences. When the wall was stiff, the inter-leaflet gap increased sigmoidally during valve opening; when it was as flexible as the leaflet, the gap increased more linearly. Less pressure difference was required both to open and to close the valve when either the wall or the leaflet material was more flexible. The degree of bias of the valve characteristics to the open position increased with both the inter-leaflet gap in the resting position and valve inflation pressure. The characteristics of the simulated valve were compared with those specified in an existing lumped-parameter model of one or more collecting lymphangions, and used to estimate a revised value for the constant in that model which controls the rate of valve opening/closure with variation in applied pressure difference. The effects of the revised value on the lymph pumping efficacy predicted by the lumped-parameter model were evaluated.
1. Introduction
The lymphatic vascular system contains endothelial primary valves in the wall of initial lymphatics, and intravascular secondary valves (Bohlen et al. 2009) which divide collecting lymphatics into lymphangions. The valves are essential for lymph transport, since there is no equivalent of the heart at the upstream end of the system to propel flow. Instead lymphangions, as their name implies, act as pumping chambers in series, utilising either the intermittent passive squeezing of lymphatic vessels resulting from the relative motion of surrounding tissues or the contractions of the unique muscle type in their walls. Like venous valves, lymphatic valves also break up the hydrostatic column that would otherwise create high transmural pressure in lower leg lymphatics when standing.
Unlike cardiac valves, most of the secondary valves are located in very small vessels; even the largest lymphatic vessel in the body is only some 2.2 mm in diameter (Telinius et al. 2010). Working with rat spinotrapezius muscle Mazzoni et al. (1987) found valves in lymphatic vessels as small as 13 μm wide. Consequently the valves must operate at low Reynolds number and in an almost completely quasi-steady flow field. These viscous-flow conditions are thought to be the reason why lymphatic valves have a complex downstream-extending funnel shape. Unlike cardiac valves, where the leaflets insert into the wall at essentially a single axial location called the valve ring, the leaflet insertions of a secondary lymphatic valve extend over an axial distance greater than one vessel diameter (Zawieja 2009). The shape also differs from that of a typical valve in large veins of the leg, which more resembles two pouches butting up to one another, but has much in common with that of valves in micro-veins (Phillips et al. 2004). For more details of valve configuration, see Moore and Bertram (2018). Whereas venous valves are typically more diaphanous than the vessels they inhabit, the walls of small lymphatic vessels are scarcely more substantial than the valve leaflets themselves; see fig. 1 of Zawieja (2009). Mazzoni et al. (1987) proposed a simple mechanistic model to explain how the valve operated, identifying the essential features as being the valve morphology and the flexibility of the leaflets.
The sole major study of secondary lymphatic valve function thus far is that by Davis et al. (2011). Their measurements of secondary lymphatic valve function showed that the valves are biased to the open position, to an extent depending on transmural pressure, and that they display opening/closing hysteresis. However, the closure measurements, which involved backflow through the valve, did not account for pressure drops due to micropipette resistance, and so are of uncertain accuracy. Thus the extent of the hysteresis remains unknown at this point. Davis has gone on to evolve his valve tests into an effective tool for assessing valve incompetence in genetically altered mice (Lapinski et al. 2017).
A numerical model can provide a concise description of valve properties and fill in gaps in knowledge from experiments. The simplest possible model of a secondary lymphatic valve simply prohibits backward flow and assigns constant resistance for forward flow (Reddy et al. 1975; Venugopal et al. 2007; Kunert et al. 2015). Macdonald (2008) added finite opening and closing time, but this had more to do with stabilizing the numerical scheme than emulating reality more closely. Bertram et al. (2011a) introduced a model in which valve resistance varied sigmoidally with the trans-valvular pressure difference. This utilized four constants to define minimum and maximum valve resistance, the pressure difference offset from zero on which the transition was centred, and the steepness of the transition. Later, this model was elaborated (Bertram et al. 2014a) to include both the dependence of the offset or bias on the valve transmural pressure and the hysteresis which had been observed experimentally (Davis et al. 2011). The performance of the resulting model subsequently threw doubt on the closure measurements, and motivated attenuation of the valve-closure offset (Bertram et al. 2014b). Adapting a pre-existing circulatory model to lymphatic vessels, Contarino & Toro (2018) used the Mynard cardiac valve model, thereby reintroducing parameters for opening and closing time to the lymphatic valve field. However, lacking data for these parameters, they resorted to guessed values1.
All the above valve models were lumped-parameter descriptions of function only; they did not attempt to explain how the valve architecture led to a given behaviour. The task of understanding why the valves behave as they do requires a higher-dimensional numerical model. Macdonald (2008) conducted limited 2D finite-volume simulations of steady flow through idealised rigid lymphatic valve geometries, showing how small valves and creeping flow inhibit fluid recirculation behind the leaflets.
Recently, Li et al. (2019) simulated lymphatic valves in 2D again, using the lattice-Boltzmann method. The valves had leaflet stiffness decreasing from the root to the free edge by a factor of almost 2 million, and were incorporated in a 2D model of a lymphangion with flexible walls. It is difficult to picture leaflet stiffness varying spatially in reality by this amount, and it is likely that extreme property variation was more a response to the artefactual difficulties of obtaining realistic lymphatic valve behaviour from a 2D flexible model. The model lymphangion did not conform to what has been measured (Davis et al. 2011); it ceased to produce forward flow when the adverse pressure difference was only 0.025 cmH2O.
Wilson et al. (2013) modelled the space inside a lymphatic valve in 3D, using data from a stack of confocal images of a rat mesenteric lymphatic. Both steady and time-dependent flows through this space were computed using a finite-volume code and assuming rigid boundaries. Streamlines and wall shear stress (WSS) distributions were computed, and the focus was on determining the concentration distribution of nitric oxide produced in a WSS-dependent rate at the flow domain boundaries, i.e. the vessel wall and valve leaflets.
Wilson et al. (2015) defined an idealised geometry for a 3D lymphatic valve based on measurement of critical dimensions in confocal images from 74 isolated vessels. Three computations were performed successively. In the first, fully developed flow was simulated with the leaflets 2 μm apart on the plane of symmetry cutting them, to determine a pressure field over each leaflet surface and from those data evaluate an average trans-leaflet pressure difference. Then this pressure difference was applied to leaflets with defined thickness and shear modulus, rigidly attached at the rigid vessel boundary, to determine leaflet deflection. Then flow as in step 1 was applied to the rigid valve with this leaflet geometry, to determine valve resistance defined as the trans-leaflet pressure difference divided by the volume flow-rate. Results were obtained for various values of the ratio of maximum sinus diameter to inlet vessel diameter. They found that orifice area was a poor predictor of resistance, and that the combined resistance to forward flow of valve and sinus was less than that of a uniform duct having the diameter of the inlet when the sinus-to-vessel diameter ratio was 1.39 or more.
Wilson et al. (2018) went on to perform two-way coupled fluid-structure interaction computations on the same geometry, still with a rigid vessel boundary. Despite having specified a neo-Hookean elastic material for the leaflet, with no viscoelasticity, they found less deflection of the leaflet for a given trans-valvular pressure during valve opening than during valve closure, for waveforms which varied trans-valvular pressure from zero to peak and back to zero in 1.5 s, with the bulk of each pressure change occurring in 0.2 s. The source of this effect, which is qualitatively different from the static hysteresis observed by Davis et al. (2011), was not specified, but is likely to have been the combined inertia of the leaflet and fluid.
A similar response to a cyclic pressure gradient was found by Ballard et al. (2018), who applied the lattice-Boltzmann method to create a 3D fluid/structure-interaction simulation of a valve consisting of flexible leaflets that were planar when undeformed, intersecting with and fixed to a rigid cylinder representing the vessel wall. They applied steady pressure differences in both forward and backward directions, and varied the leaflet length and stiffness. They found that resistance to forward flow increased with valve length, but that very short valves, of axial length scarcely more than diameter, failed to close fully when flow reversed.
Thus existing models have not yet explained some of the observed behaviour of real lymphatic valves, in particular what structural features set the extent of the bias, and why the bias to the open position is transmural-pressure-dependent (Davis et al. 2011; Bertram et al. 2014b). Accordingly the present paper reports the first 3D computational fluid-dynamic study of lymphatic valves involving fluid/structure interaction with flexibility of both leaflets and vessel wall. By varying the stiffness of the valve wall material and of the leaflet material separately, the contribution of each to the overall valve properties is defined.
2. Methods
2.1. Description of the model
The fluid/structure-interaction models were realized in the finite-element software ADINA v9.4 (ADINA R&D Inc., Watertown, MA, USA), utilizing dimensions based on confocal images of real valves as defined by van Loon (Wilson et al. 2015). Two different geometries were modelled, with the difference being the distance between the leaflets on the plane of symmetry cutting them in the unstressed state. For each model, surface mesh-point grids defining the boundaries of the valve wall, and the fluid space within, excluding the valve leaflet, were constructed in Matlab (see Figure 1) and imported as STL files into ADINA-Structures and ADINA-Fluids, where volume meshes were generated for the wall and the fluid. In the case of the leaflet, point coordinates only were imported into ADINA, and all the surface and volume meshing was performed therein, before marrying the resulting leaflet with the tube wall. Symmetry was assumed about two perpendicular cut-planes intersecting on the axis of the valve; thus only a quarter valve was computed. For each geometry, a rigid-wall model was also defined, in which the wall was merely a fixed no-slip boundary to the fluid space.
Starting from the unstressed or resting configuration with no flow, simulated flow through the valve occurred when the trans-valvular pressure difference was gradually ramped up. To open the valve, the inlet pressure was raised; to close it, the outlet pressure. The coordinate position at each end of the vessel was kept constant (corresponding to securing to pipettes experimentally), but elsewhere the entire structure was free to deform. Over 2s, the pressure was linearly ramped up to 1000 dyn cm−2 in time-steps of 5 ms; given the minute dimensions of the valve, this was sufficiently slow that the flow was quasi-steady, although it was solved with no simplifying assumptions. For runs which involved pressurisation of the whole structure, the inlet and outlet pressures were initially ramped up simultaneously before one of the two pressures was held constant and the other continued to climb. In every case, the initial fluid mesh was used only for one or a few time-steps, then a re-meshing was carried out which concentrated cells in regions of high shear. Re-meshing was repeated regularly, so that the mesh remained well-formed in the face of the deforming valve and well adapted to the changing flow (Figure 2). An initial list of times for re-meshing was prescribed, but additional re-meshings were organized by restarting the run as needed when the rate of geometry change was high. Whereas the initial fluid grid consisted of 84,892 tetrahedral elements in the first fluid model, and 59,462 in the second, the re-meshed grids consisted of in the order of half a million elements. In many instances, particularly when computing the highest-compliance part of the valve opening process, the time-step had to be reduced to achieve convergence or avoid fluid/structure element overlap; time-steps down to 0.5 ms were employed as necessary.
The wall and leaflet were both modelled as Mooney-Rivlin materials, via the neo-Hookean formulation
where W is the strain energy density, G is the shear modulus, l1 is the volume-preserving first invariant of the Cauchy-Green deformation tensor C, κ is the bulk modulus and the reduced invariant J3 is the ratio of the deformed volume to the undeformed volume. The first and second terms on the right-hand side of the equation are the deviatoric and the volumetric strain energy density respectively. The shear modulus was 20 or 40 kPa for the wall, and 10, 20 or 40 kPa for the leaflet. In addition models were run which replaced the wall with a rigid no-slip boundary to which the leaflet insertion adhered, corresponding to the work of Wilson et al. (2018), and an extremely high value of Gwall (2000 kPa) was computed, to check that the flexible-wall model converged to the rigid-wall model in the stiff limit. Where the wall was flexible, both materials were given density 103 kg m−3; the bulk modulus was calculated as
with Poisson ratio vP = 0.49. As noted by Wilson et al. (2018), a shear modulus of 20 kPa is slightly less than that of arterial elastin; this is commensurate with the imposition in these simulations of relatively small deformations, such that the lymphatic solids are confined to the part of their passive stress-strain characteristic where the properties of elastin dominate. A Poisson ratio close to 0.5 assumes that lymphatic vascular wall and valve materials are essentially incompressible as a result of being water-saturated, like the material of blood vessel walls (Carew et al. 1968); for the same reason their density is very close to that of water.
The fluid was taken as Newtonian, with density (103 kg m−3) and bulk modulus (2×109 Pa) similar to water, but with viscosity of 1.5 mPa s for intestinal lymph (Moore and Bertram 2018). All dimensional properties were entered into the simulations in μm-gm-s units, then internally made dimensionless on the basis of L0 = 1 μm, U0 = 104 μm s−1, ρ0 = 10−12 gm μm−3.
The modelled valve occurred in a vessel of inside diameter 100 μm, and the associated sinus had maximum diameter 160 μm and total length 417 μm. In the unstressed configuration, the valve leaflets were separated on the plane of symmetry x = 0 by 20 μm in the first model. This compares with a real valve (Figure 3). In a second model, this distance was halved. A small but important difference between the implementations of the idealised geometry here and by Wilson et al. (2015) is that they assumed the orifice of the valve projected onto a constant-z plane to be elliptical; here, circular arcs define the inboard trailing edges of the two leaflets and thus the initial valve orifice in z-projection (see the left panel of Fig. 3).
Both the valve wall and the leaflet were assumed to have uniform thickness of 5 μm. The maximum pressure difference applied during opening was 1000 dyn/cm2, or approximately 1 cmH2O; during closure, the leaflets reached apposition before this value was attained, except in one instance. Closure was computed up to the last possible apposition configuration before actual leaflet contact; thus contact was not explicitly modelled.
2.2. Numerical considerations
All simulations were computed as simultaneous (rather than iterative) solutions of the fluid and structure component models. In the structure model with 5 μm half-gap, the tube wall and valve leaflet consisted of 9,613 and 3,908 11-node tetrahedral elements respectively. In the equivalent 10 μm model, the wall and leaflet had 10,261 and 10,271 11-node elements respectively.
The fluid space was initially meshed with 84,892 tetrahedral cells in the 10 μm model, and 59,462 tetrahedral cells in the 5 μm model. The greatly increased number of cells in the re-meshed fluid grids was controlled by an adaptive-meshing parameter which nominally set the maximum number of cells. In practice the re-meshed grids typically contained some 50% more cells than this setting. In early runs the parameter was set to 800,000. To check for grid independence, a run with this number set to 400,000 was compared, using the four characterising plot parameters that will be shown below in Figs. 6–9. The curves from the two runs merged everywhere to within the thickness of the (thin) plotted line, i.e. the results with the lower value fully overlaid those from the run with the higher value. With the result thus confirming grid independence, the lower value was used for all subsequent runs, including those giving rise to Figs. 6–9.
Volume flow-rate was calculated from the computed axial velocity on the centre-line x = y = 0 at the extremities z = 150 and 667 μm of the conduit, assuming a fully developed Poiseuille profile. The assumption of fully developed flow was checked both by comparison of the inlet and outlet flow-rates so calculated, and by comparing computed velocities with assumed velocities on the four radial lines x = 0, z = 150 μm, y = 0, z = 150 μm, x = 0, z = 667 μm, and y = 0, z = 667 μm. These comparisons were performed for the maximum favourable pressure drop of 1000 dyn/cm2, giving rise to the maximum flow-rate through the valve. The Reynolds number based on diameter for this flow was 2.8. The departures from the velocities expected in a parabola having the same centre-line velocity reached a maximum discrepancy of 0.73%, but most were very much less. The maximum outlet-minus-inlet difference in computed velocity was 0.77% of the outlet centre-line velocity; on the centre-line it reached 0.75%.
2.3. Lumped-model parameter estimation
In section 3.6, results from this 3D finite-element model are used to estimate a value for a parameter of the lumped-parameter model of a lymphatic vessel which has been developed by our group over several years. Description of the lumped-parameter model in single-lymphangion form is given in the Appendix.
3. Results
3.1. Pressure and velocity fields
The results from a 3D simulation can be interrogated in many different ways; examples of pressure and velocity distributions when the opening pressure difference is 100 dyn cm−2 are shown in Figures 4 and and5.5. The jet that emerges from the leaflet orifice gives rise to recirculation behind the leaflet, but the flow is 3-dimensional, with complex secondary velocities, so fluid is not trapped indefinitely in closed eddies. Velocities in Fig. 4 on the axis of symmetry x = y = 0 reach 6.25 mm/s at the undilated inlet/outlet regions, and a maximum of 11.6 mm/s through the valve aperture, reflecting the imposition of the pressure difference over an axial distance of just 0.517 mm. The velocity vectors in Fig. 4 show a single recirculation eddy in the lee of the valve leaflet; Margaris et al. (2016) observed a similar phenomenon in a rat mesentery lymphatic valve subjected to an axial pressure difference of 1 cmH2O (981 dyn cm−2), of which most would have been dissipated in their perfusion system. The centre-line velocity in their undilated vessel reached 4.5 mm/s where the diameter was about 92 μm; see their fig. 9.
3.2. Variation of valve wall stiffness
With Gleaflet = 20 kPa, the deformation of the leaflet can be characterised by the y-position (initially 5 μm here) of the inner trailing edge at the plane x = 0; see Figure 6 (upper left). Valve closure is represented by this gap approaching zero. Increasing adverse pressure difference (Δp < 0) is needed to close the valve when the wall is stiffer; extrapolating the left end of the curves in the upper left panel the short distance to y = 0 where the leaflets touch, 136 dyn/cm2 sufficed when Gwall = 20 kPa, and 184 dyn/cm2 when Gwall = 40 kPa, but some 560 dyn/cm2 was needed when Gwall = 2000 kPa, while some 660 dyn/cm2 was required when the wall was entirely rigid. These numbers defined a highly nonlinear relation between the shear modulus of the wall and the adverse pressure difference at valve closure which was fitted as
where α = 0.3262, ϐ = 4.908, Gwall is in kPa, Δpclos is the adverse pressure difference at valve closure (in dyn/cm2), and Δpc-rgd is the value of Δpclos for the rigid wall.
The deformation of the leaflet is highly three-dimensional; it does not simply vary its curvature while maintaining self-similar shape. In the opening direction, a maximum inter-leaflet gap of 64.8 μm was achieved with Gwall = 20 kPa. The compliance represented by the slope dy/dp at Δp = 0 varied inversely with Gwall. When the wall was stiff, the compliance varied greatly as the valve opened, going through a maximum at Δp ≈ 240 dyn/cm2 for Gwall = 2000 kPa, and at Δp ≈ 258 dyn/cm2 for a rigid wall. Because of the less nonlinear shape of the curves for lower wall stiffness, the degree of central leaflet opening with a rigid wall even exceeds that for all simulations with flexible valve walls in a small range of opening pressures (281.5 to 388.8 dyn/cm2). The crossing-over of these curves emphasizes that the leaflet opens by a complex three-dimensional deformation that is not wholly parameterised by the single measure of the y-position of the inner trailing edge at the plane x = 0, but also demands consideration of what is happening to the vessel wall.
Leaflet opening is clearly affected by the extent to which the wall flexibility accommodates the rotational torques applied where the leaflet joins the wall. The deformation of the vessel wall at the valve can be characterised by the maximum sinus radius at the plane x = 0; see Fig. 6 (upper right). During valve closure, the sinus is progressively distended by the increasing pressure applied at what is normally the valve exit. During valve opening, the maximum sinus radius in the plane normal to the valve orientation decreases for small values of the increasing pressure applied at the valve entrance, then increases. The decrease is a result of the pressure gradient in the y-direction which is associated with the spreading of the post-valvular jet up into the sinus, where recirculation takes place. The increase at higher values of pressure difference reflects the overall raised distending pressure, and also the reduced pressure drop through the valve leaflets as the valve progressively opens. With Gwall = 2000 kPa, the changes in maximum sinus radius are small, but the radius still exhibited qualitatively the same behaviour, reducing by about 0.08 μm as the forward pressure difference increased from 0 to 245 dyn/cm2 before rising again, but not reaching values greater than its unstressed starting value for pressure differences up to 1000 dyn/cm2.
The very different slope of all the curves of flow-rate Q vs. pressure drop Δp for opening (Δp > 0) and for closing (Δp < 0), as shown in Fig. 6 (lower left), defines the nonlinear property of a one-way valve. During closure, the retrograde flow goes through a shallow maximum before decaying towards zero as the leaflets reach apposition. For the four curves shown in Fig. 6, the maxima of retrograde flow-rate are (in descending order of wall stiffness) −13.2, −11.8, −5.4 and −4.4 μL/hr, occurring at pressure difference −200, −180, −77 and −53 dyn/cm2, when the trailing-edge half-gap at x = 0 is 3.15, 3.02, 2.55 and 2.74 μm, respectively.
The valve resistance Δp/Q (Fig. 6, lower right) is the reciprocal of the slope of the plot of Q(Δp). As the inter-leaflet gap approaches zero (adverse pressure differences), it rises to a maximum, whereas for favourable pressure differences it decreases continuously as the inter-leaflet gap increases. The maximum closed-valve resistance reached was 6.97 × 108 dyn s/cm5, for both Gwall = 2000 kPa and a rigid wall. The minimum open-valve resistance of 2.92 × 106 dyn s/cm5 was reached at Δp = 1000 dyn/cm2 with Gwall = 20 kPa (red curve, obscured by the blue curve).
3.3. Variation of valve leaflet stiffness
Whereas the shear modulus of the vessel wall was varied over a wide range, including up to infinity for the sake of comparison with a previous model, the shear modulus of the leaflet was varied only over a modest range relating to that of arterial elastin (Wilson et al. 2018). Consequently the dramatic qualitative changes in curve shape seen in the upper left panel of Fig. 6 were not seen when the leaflet stiffness was varied. However, quantitatively the valve is more sensitive to variation of the stiffness of the leaflet than to variation of that of the wall; see Figure 7, where Gwall = 20 kPa, and compare the extent of the curve changes between Gwall = 20 kPa and Gwall = 40 kPa in Fig. 6. The maximum valve opening here (trailing edge y-position 35.4 μm when Gleaflet = 10 kPa, purple curve) is greater than any attained in Fig. 6, but the leaflet here starts from a more open position (10 μm half-gap). Across all simulations reported here, the maximum change in trailing-edge position occurred when the half-gap was 5 μm initially and became 32.4 μm at maximum opening (Fig. 6, red curve).
Because the valve leaflets span the lymphatic vessel conduit, their stiffness has a considerable effect on the deformation of the sinus when pressure is applied at valve inlet or outlet to open or close the valve respectively (Fig. 7, upper right). The maximum sinus radius varies through both overall distension of the flexible conduit and variation in the degree of sinus cross-sectional circularity.
Again here there is a retrograde flow-rate maximum during closure (Fig 7, lower left). For leaflet stiffness 10, 20 and 40 kPa, the maximum is −16.4, −23.9 and −33.7 μL/hr, at pressure difference −80, −124 and −178 dyn/cm2, when the trailing-edge half-gap at x = 0 is 4.64, 4.57 and 4.72 μm respectively.
The maximum valve resistance Δp/Q reached here (Fig. 7, lower right) was 9.53 × 107 dyn s/cm5, at Gleaflet = 40 kPa (grey curve), or about 12.5 times the interpolated resistance at Δp = 0. However, the orifice flow is not properly resolved where the inter-leaflet gap approaches zero, so the computed resistance is not reliable in this limit. Note that there is still backward flow at this extremity, a point that will be returned to below. Fig. 7 also shows (lower right, dashed line) the Poiseuille resistance that a straight cylindrical conduit having the diameter of the inlet and outlet of the modeled valve but no leaflets and no sinus would have. Because of the sinus, the additional valve resistance falls to zero at favourable pressure drops, even becoming slightly negative (see inset). The minimum open-valve resistance reached in this study was 2.59 × 106 dyn s/cm5, for the model with 10 μm half-gap and Gleaflet = 10 kPa, Gwall = 20 kPa, at the maximum forward pressure difference of 103 dyn/cm2.
3.4. Variation of valve Inflating pressure
The model with 5 μm half gap and Gleaflet = Gwall = 20 kPa was made the subject of an experiment in which the two boundary pressures were simultaneously ramped up to 500 dyn/cm2 (≈ 0.51 cmH2O) or 1000 dyn/cm2 (≈ 1.02 cmH2O) before any differential pressure was applied. The same procedure as usual was then applied, i.e. in one run the inlet pressure was ramped up a further 1000 dyn/cm2 while outlet pressure remained unchanged, and in the other run the outlet pressure was ramped up while the inlet pressure stayed fixed. In the case of the 1 cmH2O initial inflation, the maximum sinus radius was thereby increased from 80 μm to 81.8 μm, and the inter-leaflet gap on the plane of symmetry also increased, from 10 μm to 16.8 μm; see Figure 8. Because of these enlarged dimensions, the valve admitted slightly more flow-rate for a given pressure difference, in both forward and backward directions2. This in turn translated into a slightly lower resistance at a given pressure difference, which can alternatively be viewed as shifting the resistance characteristic to the left, i.e. to greater adverse pressure differences. Thus, relative to the uninflated valve (Fig. 8, lower right, red curve), the inflated valve (beige and light blue curves) opens at a more negative trans-valvular pressure, i.e. its bias to stay open is increased.
3.5. Variation of resting inter-leaflet gap
Increasing the initial gap between the leaflets has a qualitatively similar effect on the resistance characteristic to increasing the degree of valve inflation by pressure. This experiment required the building of the two models, with the gap between the leaflets on the plane x = 0 varied from 10 to 20 μm. The models with Gleaflet = Gwall = 20 kPa were again made the subject of this comparison. The model with increased gap admits more flow-rate for a given pressure drop in either direction, i.e. has reduced resistance; see Figure 9. The valve resistance characteristic is effectively shifted to the left, i.e. the valve exhibits increased bias to the open position. A maximum closing resistance of 2.05 × 108 dyn s/cm5 was attained for the model with reduced initial gap, but because the resistance in the unstressed position was increased, this only amounted to some 10.8 times the resistance at Δp = 0.
3.6. Comparison with lumped-parameter modeling
Using a lumped-parameter model, Bertram et al. (2017) showed that realistically low rates of leakback flow are achieved with a closed-valve resistance much higher than can be computed here. The maximum resistance of a competent valve is a parameter for which the lower bound is around 1010 dyn s/cm5, established by simulations involving slow leak-back through a closed valve over many cycles of lymphatic contraction; see, e.g., figs. 9 and and1010 of Bertram et al. (2017). On the other hand, the upper limit to closed-valve resistance in the simulations here is set by the topological change that occurs when the inter-leaflet gap on the symmetry plane x = 0 reaches zero. At this point there is still space between the leaflets off the symmetry plane; see Figure 10. As with the crossing-over of curves for the y-position of the trailing edge of the leaflet during valve opening noted in Fig. 6, this is another manifestation of the three-dimensional deformation of the valve leaflet, but here seen during valve closure.
Thus the 3D modelling here does not supersede the values for maximum and minimum valve resistance arrived at previously in the lumped-parameter modelling through comparison with experiments (Bertram et al. 2014b, 2017). Nor does it supersede the lumped-parameter modelling of the valve bias to the open state, since the real extent to which lymphatic valves are anatomically slightly open in the unstressed state is unknown. The comparison of numerical and real orifices shown in Fig. 3 would suggest that even a half-gap of 10 μm may underestimate the true value, but further experimental data are needed before the extent of actual bias can be made more precise. However, one parameter in the lumped-parameter modelling can be compared rather well with the present 3D modelling: the constant so, which controls the steepness of the open/close transition in resistance with trans-valvular pressure difference via the equation
where RV = valve resistance, ΔpV is the trans-valvular pressure difference, RVn = minimum valve resistance, RVx + RVn is the maximum valve resistance, so is the steepness parameter, and Δpo is the opening pressure-difference offset. Recent lumped-parameter modelling (Bertram et al. 2017, 2018) used so = 0.2 cm2/dyn; this value (chosen to limit closing regurgitation and closed-valve leakage) produces very steeply rising transitions. A resistance curve which closely matches that in Fig. 7 for 10 μm half-gap and Gleaflet = Gwall = 20 kPa (cyan curve) is produced with the parameter values RVn = 2.7 × 106 dyn s/cm5, RVx = 1010 − RVn dyn s/cm5, so = 0.01 cm2/dyn, and Δpo = −755 dyn/cm2; see Figure 113.
The result of changing so from 0.2 to 0.01 cm2/dyn was tested in the lumped-parameter model of a single lymphangion (Bertram et al. 2017); see Figure 12. Under control conditions as stipulated in the figure [see parameter values above left panels—explanation of all parameters is given in the Appendix], including pumping against an adverse pressure difference of 2.5 cmH2O, the cycle-average flow-rate was 27.4 μL/hr. At the reduced so value, the cycle-average flow-rate increased to 36.3 μL/hr. The intra-lymphangion pressures varied with time during the contraction over a larger range than in the control run, because the more gradually opening valves provided more resistance to flow when nominally open. By itself, this factor would have reduced the overall ejection volume during the cycle, but its effect was more than compensated by the absence of the regurgitation prior to valve closure that was prominent in the control situation. With the reduced so value, Δpo varied between −0.24 and −0.07 cmH2O for valve V1, with cycle-average −0.13 cmH2O, and between −0.34 and +0.08 cmH2O for V2, with cycle-average −0.10 cmH2O. To match more closely the value of −0.77 cmH2O, the model was again run with so = 0.01 cm2/dyn, but the two curves4 of Δpo(Δptmv) which define the thresholds at which a valve switches from open to closed and vice versa were both shifted by −0.6 cmH2O, i.e. to more negative values of Δpo, for both valves. Under these conditions (right panels), the cycle-average flow-rate was 36.4 μL/hr, i.e. almost unchanged; a small amount of regurgitation is compensated by reduced valve resistance when open. But Δpo varied between −0.90 and −0.67 cmH2O for V1, with cycle-average −0.76 cmH2O, and between −0.94 and −0.58 cmH2O for V2, with cycle-average −0.74 cmH2O5. Other small differences attributable to the changed valve-switching characteristics are noticeable; for instance the isovolumic periods at the beginning and end of systole when both valves are closed are slightly lengthened in the middle panels of Fig. 12, and considerably shortened in the right-hand panels. The leftward shift (in terms of Fig. 11) of the valve characteristic created by the offsets in the right-hand panels of Fig. 12 restores the prompt diastolic filling of the lymphangion that was present in the control run but was hindered under the conditions of the middle panels of Fig. 12. This is seen both in the peak value of flow-rate Q1 through the inlet valve and in terms of the speed with which the intra-lymphangion pressures p1, pm and p2 all reach the inlet reservoir value of 2.5 cmH2O after the inlet valve opens.
4. Discussion
The simulations predict that, during that part of valve closure which takes the valve from its rest configuration to first leaflet apposition, there is a retrograde (leakage) flow-rate which goes through a maximum before decaying towards zero. Because it is slightly open at rest, the valve requires an adverse trans-valvular pressure difference to cause the leaflets to deflect enough to achieve closure. This necessarily invokes retrograde flow, which is eventually gradually suppressed as increasing degrees of closure are achieved. Transient backflow in lymphatics with valves was observed by Dixon et al. (2006) and Blatter et al. (2016). The specific prediction will be difficult to verify experimentally, where it is hard to attain perfectly smooth control of the tiny pressures involved, and closure tends to become an unstable process, occurring suddenly6. The leaflets are then immediately pushed together in a transient which may momentarily bring unsteady dynamics to the situation.
With rigid wall, the model with 5 μm half-gap and Gleaflet = 20 kPa had minimum open-valve resistance of 3.26 × 106 dyn s/cm5. This compares with the value of 2.68 × 106 dyn s/cm5, computed by Wilson et al. (2018) in their rigid conduit, with otherwise only small geometry differences. But there is a difference in definition: Δp as used here is the pressure difference from end to end of the conduit, whereas Wilson et al. (2018) used the pressure difference from end to end of the sinus alone. An equivalent value can be approximated here by subtracting the Poiseuille resistance of the cylindrical inlet and outlet tubes, giving 2.65 × 106 dyn s/cm5.
In agreement with Wilson et al. (2015), the computations here showed that the structure simulated, which included both the valve proper and the sinus, reached an overall resistance in the open state which was as low or lower than that of a uniform cylindrical conduit having the diameter of the inlet and outlet tubes (3.16 × 106 dyn s/cm5). However, Wilson et al. used a rigid tubular geometry, whereas all the structure here is flexible. The comparison here ignores the distension of the structure when one of the boundary pressures is increased to create a pressure difference. An initially cylindrical but flexible tube would also distend; thus overall the resistance of the open valve may not become less than that of the equivalent lymphatic vessel with no valve.
The only measurement of open-valve resistance is the value 0.6 × 106 dyn s/cm5, measured by Bertram et al. (2014b). However, the experimental value was for the added resistance of the valve mechanism itself, over that of an empty equivalent conduit. It does not include any component of viscous resistance along the conduit at all. To make approximate comparison with the computations here, it is necessary to subtract the Poiseuille resistance of a valve-less tube having the dimensions of the sinus plus the inlet and outlet tubes. Using the rigid-wall figure from above, the result is 1.48 × 106 dyn s/cm5. Using instead the minimum open-valve resistance for the flexible-wall model with 10 μm half-gap (section 4.3), the result is 0.81 × 106 dyn s/cm5. That the experimental value is some 25% lower may relate to the numerical valve here, based on the dimensions defined by Wilson et al. (2015), not being able to open as fully as the real valves measured by Bertram et al. (2014b), which were not part of the Wilson dataset.
Fig. 10 showed that, despite the application of significant adverse pressure drops (reaching 425 dyn/cm2 in the case of the simulation with 10pm half-gap and Gleaflet = Gwall = 20 kPa7), the valve orifice never closed entirely, with a narrow gap remaining on each side of the valve centre-line. It is not known whether such configurations ever occur transiently in vivo. It is likely that in real valves any remaining off-centre spaces are progressively squeezed to ever-smaller dimensions by more adverse pressure difference than can be applied here. Many other factors may operate to allow the real valve to achieve full leak-free closure; these may include subtle small-scale spatial material-property changes where the leaflet trailing edge attaches to the vessel wall in the sinus. A lymphatic valve leaflet consists of a monolayer of endothelial cells on either side of an extracellular matrix which was reported by Rahbar et al. (2012) to be composed predominantly of elastin, with collagen bands at the insertion points. Clearly a homogeneous and isotropic hyper-elastic solid describing simple rubber-like materials, as used here, cannot be expected to emulate the fine detail of the in vivo contact conditions when two such leaflets are pressed against each other by a substantial adverse pressure drop across the valve.
The simulations here suggest that the value of one of the parameters in the lumped-parameter model of a secondary lymphatic valve (of comparable dimensions pertaining to rat mesentery collecting lymphatics) should be modified. Specifically so, the parameter which controls the steepness of the open/closed transition in resistance with changing trans-valvular pressure, produces a characteristic which matches that simulated here with 10 μm half-gap and Gleaflet = Gwall = 20 kPa if its value is decreased by a factor of 20 and Δpo is set to an appropriately changed value. Note that what is approximately fitted here is only a small part of the foot of a valve resistance characteristic that in the lumped-parameter model extends up to a maximum closed-valve resistance of 1010 dyn s/cm5; this is why the fitted values of so and Δpo are so interdependent. When the lumped-parameter model is run with the reduced so value (and the valve characteristic offsets set to produce the changed Δpo value on average), the nature of the pumping cycle changes in small but important ways. Slightly increased open-valve resistance at modest transvalvular pressure drops in the open-valve part of the characteristic does not reduce the mean flow-rate in the trials shown here, involving an adverse pressure difference of 2.5 cmH2O across a single lymphangion, because reduced regurgitation before valve closure more than compensates. However, it should be emphasized that this result was achieved with the lumped-parameter valve open/close threshold curves shifted by −0.6 cmH2O relative to those indicated by experiment (Davis et al. 2011; Bertram et al. 2014b). The comparison shows that the 3D simulations are compatible with lumped-parameter valve characteristics that produce a workable lymphangion pump. However, even the detailed 3D valve model is still greatly simplified in terms of material properties, etc., relative to real lymphatic valves in vivo, so this agreement does not guarantee that these shared characteristics are then necessarily closely aligned with the properties of real valves.
To elaborate on some of the simplified aspects, all structures here were assumed neo-Hookean, whereas the passive mechanics of collecting lymphatic vessels is highly non-linear (Rahbar et al. 2012). Partly for that reason, the simulated changes in transmural pressure here (0.5 and 1 cmH2O) were considerably smaller than the range investigated experimentally by Davis et al. (2011). Furthermore, their evidence suggested that valve function can be influenced by vessel tone, here effectively assumed to be zero or a small constant value.
Lymphatic muscle can generate both tone and phasic contractions. In a lymphangion, the active deformation of the vessel wall generates the pressure within the vessel. In the model here the wall passively deforms in response to externally applied inlet and outlet pressures; this is analogous to the experiments performed by Davis et al. (2011) on lymphatic valves under calcium-free conditions to inhibit vessel contraction. However, muscle cells are scarce (Bridenbaugh et al. 2013a; Zawieja et al. 2018) or absent (Sabine and Petrova 2014) in the immediate vicinity of lymphatic valves. The model here was confined to the vicinity of one valve, with total length 517 μm, or just over five non-sinus vessel diameters.
Another factor that can vary considerably is the valve geometry. The idealised geometry used here is closely based on that defined by Wilson et al. (2015) using dimensions measured in 74 rat mesenteric lymphatic vessels. However, their fig. 2 showed that each dimensional parameter varied quite widely, even in the one tissue and species, while Ballard et al. (2018) show images comparing valve morphology across four locations in the rat and one in the sheep.
This paper has shown how the characteristics of a lymphatic valve are affected by both the stiffness of the leaflets and that of the vessel wall into which the leaflets insert. For the first time, they confirm experiments (Davis et al. 2011) which showed that the valve opens and closes at greater adverse pressure difference if distended by transmural pressure. They also show, again for the first time, how the extent of the valve bias to remain open until significant adverse pressure difference is a function of the extent of the gap between the leaflets in the unstressed condition, with all other geometric parameters unchanged. This work extends our knowledge of how the naturally occurring geometry creates a valve structure which functions competently even at the microscopic spatial scale of typical lymphatic collecting vessels.
Table 1
colour | line-type | so | RVn | Δpo |
---|---|---|---|---|
cm2/dyn | dyn s/cm5 | dyn/cm2 | ||
green | solid | 0.2 | 2 × 106 | 0 |
green | dashed | 0.2 | 8 × 106 | 0 |
green | dot-dash | 0.2 | 2 × 106 | −488 |
green | dotted | 0.2 | 8 × 106 | −488 |
blue | solid | 0.2 | 2.7 × 106 | −755 |
red | solid | 0.01 | 2.7 × 106 | −755 |
Acknowledgements
Particularly excellent and essential help and advice were given by Mr. Nicholas Medeiros, a former support engineer at ADINA R&D Inc., as well as assistance from other ADINA support staff. I acknowledge extensive discussions with Associate Professor Charlie Macaskill through the course of this work. I thank Professors Michael J. Davis, James E. Moore Jr. and Raoul van Loon for valuable discussions, and three anonymous reviewers for useful suggestions. I acknowledge funding from NIH grant U01-HL-123420.
Appendix - equations of the model
Mass conservation:
Momentum conservation:
where p1, pm, and p2 = pressure at the upstream end, centre, and downstream end of the lymphangion, pa and pb = upstream and downstream reservoir pressures, p0 = pressure upstream of the inlet valve but downstream of the inlet cannula, and pp = pressure downstream of the outlet valve but upstream of the outlet cannula; Q1 and Q2 = flow-rates into and out of the lymphangion, and D = lymphangion diameter. The resistances of the inlet and outlet valves, RV1 and RV2, are given by
where ΔpV1 = p0 − p1, ΔpV2 = p2 − pp, the open/close threshold Δpo1 depends on either p0 − pe or p1 − pe according to the current valve state, and Δpo2 depends on either p2 − pe or pp − pe (Bertram et al. 2014a). The constants in eqns. A1–4 are the lymphangion length L (0.3 cm), the lymph viscosity μ (1 cP), the resistances of the inlet and outlet cannulae (Ra and Rb respectively), the external pressure pe, and RVn, RVx and so as defined in the main text.
Constitutive relation:
two curves fitted to twitch-contraction data [see Bertram et al. (2017)] describe the maximally active and passive states as
where subscripts act and psv indicate the peak-twitch and pre-twitch states respectively.
Time-course of active contraction:
where the activation waveform M(t) is calculated in two steps; in the first,
where fm = mf, tc is the beginning of a contraction, and 1/f is its duration. A function Mj(t) modifies the preliminary M(t) to its final form, where
Constant m multiplies the rate of onset and decay of the active state; with m = 1, M(t) is a sinusoid.
Pressure-dependent diastole:
the duration of the period of relaxation tr that starts at t = tc + 1/f is
where ftw = 60(−1.39q2 + 12.6q + 0.647), with
where Δptm = pm − pe (in cmH2O).
Footnotes
Publisher's Disclaimer: This Author Accepted Manuscript is a PDF file of an unedited peer-reviewed manuscript that has been accepted for publication but has not been copyedited or corrected. The official version of record that is published in the journal is kept up to date and so may therefore differ from this version.
Conflict of interest
The author declares that he has no conflict of interest.
1Blatter et al. (2017) have observed mouse lymphatic valves to open in less than one unit and close in less than two units of 0.28 s, this being the limit of time resolution of their technique. The 30 frame/s videos from the Davis lab show that opening and closing times vary greatly with factors controlling the rate at which the transvalvular pressure difference changes (personal observation, CDB). Such factors include among others the prevailing adverse pressure difference for the vessel, the transmural pressure, and the vigour of contractions.
2The maxima of retrograde flow-rate were −4.4, −8.8 and −17.3 μL/hr for inflation pressure 0, 0.5 and 1 cm H2O respectively. They occurred at pressure differences −53, −87 and −125 dyn cm−2, when the half-gaps were 2.74, 3.52 and 4.62 μm respectively.
3In the lumped-parameter valve model with transmural-pressure-dependent bias (Bertram et al. 2014b), Δpo is not constant, but varies between approximately 0 and −2210 dyn/cm2.
4See Bertram et al. (2014b) for these curves (fig. 4 of that paper) and for full explanation of how the transmural-pressure-dependent bias and hysteresis are organized. The definition of transmural pressure for a valve (ΔptmV) changes with valve state.
5If this same degree of valve characteristic offset is introduced with so taking its accustomed value of 0.2 cm2/dyn, the result (with all other parameters and boundary conditions unchanged) is that V1 stays open and V2 stays shut all the time, and there is constant slight leakage backflow (0.9 μL/hr) past the closed valve, propelled by the 2.5 cmH2O adverse pressure drop. Contraction produces only an ineffectual to-and-fro sloshing past the open inlet valve.
6Unpublished observations by M.J. Davis suggest that, while there is much variation between individual valves, murine valve closure tends to be a stable, fully reversible process at low transmural pressure, only becoming unstable at higher transmural pressure.
7In a simulation for the 10 μm half-gap model with Gleaflet = 20 kPa and Gwall = 2000 kPa (not shown), closure required an adverse pressure difference of over 1000 dyn/cm2.
References
- Ballard M, Wolf KT, Nepiyushchikh Z, Dixon JB, Alexeev A (2018) Probing the effect of morphology on lymphatic valve dynamic function. Biomechanics and Modeling in Mechanobiology 17 (5):1343–1356. 10.1007/s10237-018-1030-y [Abstract] [CrossRef] [Google Scholar]
- Bertram CD, Macaskill C, Davis MJ, Moore JE jr (2014b) Development of a model of a multi-lymphangion lymphatic vessel incorporating realistic and measured parameter values. Biomechanics and Modeling in Mechanobiology 13 (2):401–416. 10.1007/s10237-013-0505-0 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Bertram CD, Macaskill C, Davis MJ, Moore JE jr. (2017) Valve-related modes of pump failure in collecting lymphatics: numerical and experimental investigation. Biomechanics and Modeling in Mechanobiology 16 (6):1987–2003. 10.1007/s10237-017-0933-3 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Bertram CD, Macaskill C, Davis MJ, Moore JE jr (2018) Contraction of collecting lymphatics: organization of pressure-dependent rate for multiple lymphangions. Biomechanics and Modeling in Mechanobiology 17 (5):1513–1532. 10.1007/s10237-018-1042-7 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Bertram CD, Macaskill C, Moore JE jr (2011a) Simulation of a chain of collapsible contracting lymphangions with progressive valve closure. ASME J Biomech Eng 133 (1):011008-1–011008-10. 10.1115/1.4002799 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Bertram CD, Macaskill C, Moore JE jr (2014a) Incorporating measured valve properties into a numerical model of a lymphatic vessel. Computer Methods in Biomechanics and Biomedical Engineering 17 (14):1519–1534. 10.1080/10255842.2012.753066 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Blatter C, Meijer EFJ, Nam AS, Jones D, Bouma BE, Padera TP, Vakoc BJ (2016) In vivo label-free measurement of lymph flow velocity and volumetric flow rates using Doppler optical coherence tomography. Scientific Reports 6:29035 (1–10). 10.1038/srep29035 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Bohlen HG, Wang W, Gashev A, Gasheva O, Zawieja D (2009) Phasic contractions of rat mesenteric lymphatics increase basal and phasic nitric oxide generation in vivo. American Journal of Physiology - Heart and Circulatory Physiology 297 (4):H1319–H1328. 10.1152/ajpheart.00039.2009 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Bridenbaugh EA, Nizamutdinova IT, Jupiter D, Nagai T, Thangaswamy S, Chatterjee V, Gashev AA (2013a) Lymphatic muscle cells in rat mesenteric lymphatic vessels of various ages. Lymphatic Research and Biology 11 (1):35–42. 10.1089/lrb.2012.0025 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Carew TE, Vaishnav RN, Patel DJ (1968) Compressibility of the arterial wall. Circulation Research 23:61–68 [Abstract] [Google Scholar]
- Contarino C, Toro EF (2018) A one-dimensional mathematical model of collecting lymphatics coupled with an electro-fluid-mechanical contraction model and valve dynamics. Biomechanics and Modeling in Mechanobiology 17 (6):1687–1714. 10.1007/s10237-018-1050-7 [Abstract] [CrossRef] [Google Scholar]
- Davis MJ, Rahbar E, Gashev AA, Zawieja DC, Moore JE jr (2011) Determinants of valve gating in collecting lymphatic vessels from rat mesentery. American Journal of Physiology - Heart and Circulatory Physiology 301 (1):H48–H60. 10.1152/ajpheart.00133.2011 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Dixon JB, Greiner ST, Gashev AA, Cote GL, Moore JE jr., Zawieja DC (2006) Lymph flow, shear stress, and lymphocyte velocity in rat mesenteric prenodal lymphatics. Microcirculation 13 (7):597–610. 10.1080/10739680600893909 [Abstract] [CrossRef] [Google Scholar]
- Kunert C, Baish JW, Liao S, Padera TP, Munn LL (2015) Mechanobiological oscillators control lymph flow. Proceedings of the National Academy of Sciences of the USA 112 (35):10938–10943. 10.1073/pnas.1508330112 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Lapinski PE, Lubeck BA, Chen D, Doosti A, Zawieja SD, Davis MJ, King PD (2017) RASA1 regulates the function of lymphatic vessel valves in mice. Journal of Clinical Investigation 127 (7):2569–2585. 10.1172/JCI89607 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Li H, Mei Y, Maimon N, Padera TP, Baish JW, Munn LL (2019) The effects of valve leaflet mechanics on lymphatic pumping assessed using numerical simulations. Scientific Reports 9:10649 (1–17). 10.1038/s41598-019-46669-9 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Macdonald AJ (2008) The Computational Modelling of Collecting Lymphatic Vessels. Ph.D, University of Exeter, Exeter, UK [Google Scholar]
- Margaris KN, Nepiyushchikh Z, Zawieja DC, Moore J jr., Black RA (2016) Microparticle image velocimetry approach to flow measurements in isolated contracting lymphatic vessels. Journal of Biomedical Optics 21 (2):025002-1–025002-11. 10.1117/1.JBO.21.2.025002 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Mazzoni MC, Skalak TC, Schmid-Schonbein GW (1987) Structure of lymphatic valves in the spinotrapezius muscle of the rat. Blood Vessels 24 (6):304–312 [Abstract] [Google Scholar]
- Moore JE jr., Bertram CD (2018) Lymphatic system flows. Annual Review of Fluid Mechanics 50:459–482. 10.1146/annurev-fluid-122316-045259 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Phillips MN, Jones GT, van Rij AM, Zhang M (2004) Micro-venous valves in the superficial veins of the human lower limb. Clinical Anatomy 17:55–60. 10.1002/ca.10141 [Abstract] [CrossRef] [Google Scholar]
- Rahbar E, Weimer J, Gibbs H, Yeh AT, Bertram CD, Davis MJ, Hill MA, Zawieja DC, Moore JE jr (2012) Passive pressure-diameter relationship and structural composition of rat mesenteric lymphangions. Lymphatic Research and Biology 10 (4):152–163. 10.1089/lrb.2011.0015 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Reddy NP, Krouskop TA, Newell PH jr. (1975) Biomechanics of a lymphatic vessel. Blood Vessels 12 (5):261–278 [correction publ. 14 (2): 128] [Abstract] [Google Scholar]
- Sabine A, Petrova TV (2014) Interplay of mechanotransduction, FOXC2, connexins, and calcineurin signaling in lymphatic valve formation In: Kiefer F, Schulte-Merker S (eds) Developmental Aspects of the Lymphatic Vascular System. Advances in Anatomy, Embryology and Cell Biology, vol 214 Springer-Verlag, Wien, pp 67–80. 10.1007/978-3-7091-1646-3_6 [Abstract] [CrossRef] [Google Scholar]
- Telinius N, Drewsen N, Pilegaard H, Kold-Petersen H, de Leval M, Aalkjaer C, Hjortdal V, Boedtkjer DB (2010) Human thoracic duct in vitro: diameter-tension properties, spontaneous and evoked contractile activity. American Journal of Physiology (Heart and Circulatory Physiology) 299 (3):H811–H818. 10.1152/ajpheart.01089.2009 [Abstract] [CrossRef] [Google Scholar]
- Venugopal AM, Stewart RH, Laine GA, Dongaonkar RM, Quick CM (2007) Lymphangion coordination minimally affects mean flow in lymphatic vessels. American Journal of Physiology - Heart and Circulatory Physiology 293 (2):H1183–H1189. 10.1152/ajpheart.01340.2006 [Abstract] [CrossRef] [Google Scholar]
- Wilson JT, Edgar LT, Prabhakar S, Horner M, van Loon R, Moore JE jr. (2018) A fully coupled fluid-structure interaction model of the secondary lymphatic valve. Computer Methods in Biomechanics and Biomedical Engineering 21 (16):813–823. 10.1080/10255842.2018.1521964 [Abstract] [CrossRef] [Google Scholar]
- Wilson JT, van Loon R, Wang W, Zawieja DC, Moore JE jr (2015) Determining the combined effect of the lymphatic valve leaflets and sinus on resistance to forward flow. J Biomech 48 (13):3584–3590. 10.1016/j.jbiomech.2015.07.045 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Wilson JT, Wang W, Hellerstedt AH, Zawieja DC, Moore JE jr (2013) Confocal image-based computational modeling of nitric oxide transport in a rat mesenteric lymphatic vessel. ASME J Biomech Eng 135 (5):051005-1–051005-8. 10.1115/1.4023986 [erratum publ. vol.135(11), 117001, doi 10.1115/1.4025336] [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Zawieja DC (2009) Contractile physiology of lymphatics. Lymphatic Research and Biology 7 (2):87–96. 10.1089/lrb.2009.0007 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
- Zawieja SD, Castorena-Gonzalez JA, Scallan JP, Davis MJ (2018) Differences in L-type Ca2+ channel activity partially underlie the regional dichotomy in pumping behavior by murine peripheral and visceral lymphatic vessels. American Journal of Physiology - Heart and Circulatory Physiology 314 (5):H991–H1010. 10.1152/ajpheart.00499.2017 [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
Full text links
Read article at publisher's site: https://doi.org/10.1007/s10237-020-01325-4
Read article for free, from open access legal sources, via Unpaywall: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7572455
Citations & impact
Impact metrics
Citations of article over time
Article citations
Connexin-45 is expressed in mouse lymphatic endothelium and required for lymphatic valve function.
JCI Insight, 9(16):e169931, 18 Jul 2024
Cited by: 1 article | PMID: 39074069 | PMCID: PMC11343601
Computational fluid dynamic modeling of the lymphatic system: a review of existing models and future directions.
Biomech Model Mechanobiol, 23(1):3-22, 30 Oct 2023
Cited by: 3 articles | PMID: 37902894 | PMCID: PMC10901951
Review Free full text in Europe PMC
Effect of valve spacing on peristaltic pumping.
Bioinspir Biomim, 18(3), 09 Mar 2023
Cited by: 2 articles | PMID: 36821859 | PMCID: PMC9997067
Multiple aspects of lymphatic dysfunction in an ApoE -/- mouse model of hypercholesterolemia.
Front Physiol, 13:1098408, 06 Jan 2023
Cited by: 6 articles | PMID: 36685213 | PMCID: PMC9852907
An Enhanced 3D Model of Intravascular Lymphatic Valves to Assess Leaflet Apposition and Transvalvular Differences in Wall Distensibility.
Biology (Basel), 12(3):379, 27 Feb 2023
Cited by: 2 articles | PMID: 36979071 | PMCID: PMC10044971
Go to all (7) article citations
Similar Articles
To arrive at the top five similar articles we use a word-weighted algorithm to compare words from the Title and Abstract of each citation.
The effects of valve leaflet mechanics on lymphatic pumping assessed using numerical simulations.
Sci Rep, 9(1):10649, 23 Jul 2019
Cited by: 9 articles | PMID: 31337769 | PMCID: PMC6650476
Simulation of a chain of collapsible contracting lymphangions with progressive valve closure.
J Biomech Eng, 133(1):011008, 01 Jan 2011
Cited by: 39 articles | PMID: 21186898 | PMCID: PMC3356777
An Enhanced 3D Model of Intravascular Lymphatic Valves to Assess Leaflet Apposition and Transvalvular Differences in Wall Distensibility.
Biology (Basel), 12(3):379, 27 Feb 2023
Cited by: 2 articles | PMID: 36979071 | PMCID: PMC10044971
Determining the combined effect of the lymphatic valve leaflets and sinus on resistance to forward flow.
J Biomech, 48(13):3584-3590, 11 Aug 2015
Cited by: 20 articles | PMID: 26315921 | PMCID: PMC4600660
Funding
Funders who supported this work.
NHLBI NIH HHS (1)
Grant ID: U01 HL123420
NIH HHS (1)
Grant ID: U01-HL-123420-01
National Institutes of Health (1)
Grant ID: U01-HL-123420-01