Abstract
Free full text
Feasibility of improving vascular imaging in the presence of metallic stents using spectral photon counting CT and K-edge imaging
Abstract
Correct visualization of the vascular lumen is impaired in standard computed tomography (CT) because of blooming artifacts, increase of apparent size, induced by metallic stents and vascular calcifications. Recently, due to the introduction of photon-counting detectors in the X-ray imaging field, a new prototype spectral photon-counting CT (SPCCT) based on a modified clinical CT system has been tested in a feasibility study for improving vascular lumen delineation and visualization of coronary stent architecture. Coronary stents of different metal composition were deployed inside plastic tubes containing hydroxyapatite spheres to simulate vascular calcifications and in the abdominal aorta of one New Zealand White (NZW) rabbit. Imaging was performed with an SPCCT prototype, a dual-energy CT system, and a conventional 64-channel CT system (B64). We found the apparent widths of the stents significantly smaller on SPCCT than on the other two systems in vitro (p<0.01), thus closer to the true size. Consequently, the intra-stent lumen was significantly larger on SPCCT (p<0.01). In conclusion, owing to the increased spatial resolution of SPCCT, improved lumen visualization and delineation of stent metallic mesh is possible compared to dual-energy and conventional CT.
Introduction
Coronary artery stenting has become the most important nonsurgical coronary revascularization procedure1. Coronary angiography using computed tomography (CT) is limited in regards to a correct assessment of in-stent restenosis and its impact on coronary circulation, one of the major complications following the stenting procedure. Indeed, blooming artifacts impair a correct visualization of the vascular lumen. Blooming artifacts result in an apparent increase in the size of strongly attenuating objects, such as the stent’s metallic struts and vascular calcifications, leading to an apparent reduction in size of the vascular lumen. Metal strut induced blooming artifacts are more significant in stents of diameter 3mm or less, reducing intra-stent lumen interpretability to nearly 51% compared to 81% for larger stents2. Furthermore, severe calcifications also induce blooming artifacts and their presence at a stent location may impair the detection of the stent itself.
The recent development of a clinically based small field-of-view (FOV) spectral photon counting CT (SPCCT) prototype achieves higher spatial resolutions than standard CT at similar energies, owing to an important reduction in detector size3. The higher spatial resolution is possible owing to the direct conversion photon counting detectors, which convert single x-rays to electrical pulses. Ultra-high resolution mode resulting in a 0.25×0.25mm pixel size at iso-center was demonstrated recently for a photon-counting CT system4. This increase in spatial resolution is expected to strongly reduce blooming artifacts, potentially improving restenosis detection and interpretation5. Reduction in blooming artifacts should allow morphological assessment of the mesh of the stent which is expected to improve detection of stent deployment problems and stent integrity degradation6. Furthermore, owing to the spectral separation capabilities of photon counting, specific K-edge images can be obtained7–12. When a stent contains a material compatible with k-edge imaging, such as platinum, the element specific K-edge images may allow visualization of only the metal stent, eliminating all other sources of attenuation.
Thus, the objective of the study was to assess the capability of the SPCCT scanner to improve vascular lumen delineation and visualization of stent architecture, in vitro and in a single live rabbit, in comparison with dual-energy and standard CT.
Results
Representative conventional HU images of the stent containing phantom in water acquired on the B64, IQon, and SPCCT are presented in Fig. 1. The higher spatial resolution of SPCCT resulted in a better visualization of the intra-stent lumen. Furthermore, the visual separation of the calcification from the stent was only possible on images acquired with the SPCCT (Fig. 2, arrows indicate the calcification). 3D Volume Rendering views of the platinum coated stent (Promus Premier) performed on conventional HU images allowed clear visualization of the metallic mesh on SPCCT images (Fig. 2). Furthermore, the deformation of the metallic mesh due to the presence of the calcification was also clearly visible on the SPCCT images. However, visualization of the metallic mesh was not possible on B64 and IQon images.
The Pt-specific K-edge images allowed visualization of only the platinum coated stent (Fig. 3, Promus Premier). Signal from all other sources of attenuation such as the calcification or the luminal contrast agent were absent (Fig. 3). The other two types of stent were not visible on the Pt-specific K-edge images, except for the gold markers of Stent 3 (Stainless steel), because these two materials have similar k-edge energies (78.4 for Pt and 80.7 for Au). Furthermore, the deformed shape of the platinum-coated stent due to the presence of the calcification was clearly visible (Fig. 3 center columns).
The apparent width of the metallic struts, measured in the calcification free region, was smaller on SPCCT than both on IQon, and on B64 (Table 1). There was no difference in the apparent width of the Pt-containing stent between the conventional HU and the Pt-specific images. Similarly, the intra-stent lumen, measured in the same region, was larger on SPCCT than on IQon, and on B64. However the true size of 3.4mm remained strongly underestimated (Table 1).
Table 1
Coroflex Blue Neo (Co-Cr) | Promus Premier (Pt-Cr) | Tsunami Gold (Stainless Steel) | ||
---|---|---|---|---|
SPCCT apparent strut width (mm) | Conventional HU | 0.69±0.002 | 0.71±0.002 | 0.69±0.004 |
Pt K-edge images | — | 0.71±0.007 | — | |
IQon apparent strut width (mm) | Conventional HU | 0.98±0.06 | 0.99±0.05 | 1.01 ± 0.16 |
B64 apparent strut width (mm) | Conventional HU | 1.03±0.004 | 1.00±0.003 | 1.04±0.009 |
SPCCT intra-stent lumen (mm) | Conventional HU | 2.44±0.01 | 2.28±0.01 | 2.71±0.04 |
IQon intra-stent lumen (mm) | Conventional HU | 2.04±0.08 | 2.21±0.06 | 1.94±0.21 |
B64 intra-stent lumen (mm) | Conventional HU | 1.96±0.006 | 1.9±0.003 | 2.18±0.04 |
Similarly, in vivo images demonstrated improved visualization of the metallic mesh of stents in the abdominal aorta of the NZW rabbit (Fig. 4). Pt-specific K-edge imaging enabled exclusive visualization of the stent containing Pt only, and removal of other backgrounds and contrast media. Finally, the apparent width of metallic struts was larger in vivo (Table 2) compared to in vitro both on SPCCT and on B64.
Table 2
Coroflex Blue Neo (Co-Cr) | Promus Premier (Pt-Cr) | Tsunami Gold (Stainless Steel) | ||
---|---|---|---|---|
SPCCT apparent strut width (mm) | Conventional HU | 0.79±0.006 | 0.74±0.003 | 0.75±0.004 |
Pt K-edge images | — | 0.70±0.012 | — | |
B64 apparent strut width (mm) | Conventional HU | 1.13±0.009 | 1.11±0.005 | 1.06±0.008 |
SPCCT intra-stent lumen (mm) | Conventional HU | 2.06±0.006 | 1.85±0.004 | 1.81±0.003 |
B64 intra-stent lumen (mm) | Conventional HU | 1.73±0.008 | 1.60±0.004 | 1.50±0.008 |
Discussion
In this study, we present the added value of increased spatial resolution and element specific detection of SPCCT for stented vessel imaging. Firstly, the increased spatial resolution of the SPCCT system compared to dual-energy and standard CT resulted in a significant reduction of stent related blooming artifacts that improved visualization of the intra-stent lumen. Furthermore, the increased spatial resolution allowed the evaluation of the morphology of the stent’s metallic mesh. Secondly, the capability of SPCCT to provide element specific K-edge images allowed visualization of the platinum containing stent and of the gold marker of the stainless steel stent, while all other sources of attenuation were removed from the image. These results demonstrate a high potential of SPCCT in detecting correct stent deployment.
Spatial resolution is a limiting factor for imaging small and strong attenuating objects, such as coronary stents. Neither system was capable to visualize the true size of the metallic struts (~80 microns). However, the measured width of the struts, which was measured approximately 30% more accurately on the SPCCT system than on the other systems, reflects the improved spatial resolution of the SPCCT7,13,14. These results are in agreement with the very recent study of Mannil and colleagues, who found a 28.8% increase in intra stent luminal diameter when comparing a photon counting detector (PCD) to a conventional energy integrating detector (EID)13. To be noted that in our study, different slice thicknesses were used between the SPCCT and the IQon and B64 system. While, slice thickness has an important role in blooming along the Z-axis as seen in the coronal/sagittal images, it has very little impact on the in-plane blooming. The stents were placed along the Z-axis and the apparent strut width was measured in-plane. Consequently, the slice thickness is not expected to have an influence on our finding. Also note that our acquisitions on the two clinical systems (IQon and B64) were performed using Standard Resolution mode with large focal spot (1.1×1.2mm2) while High Resolution with small focal spot (0.6×0.7mm2) was used on SPCCT. The justification for this is that the resolution on the B64 and IQon systems is limited by the Nyquist limit due to the detector pitch. Therefore, Standard Resolution is used in cardiac mode where scan is limited to the equivalent of 180 degrees’ parallel beam recon (high resolution requires higher sampling rate achieved by 360 degrees’ recon utilizing quarter detector offset). Futhermore, High Resolution with small focal spot is usually not used for cardiac CT in routine practice because it doesn’t provide enough X-ray flux for this fast rotation acquisition. In the case of SPCCT, the detectors are much smaller and the limiting factor becomes the focal spot size. Using large focal spot on SPCCT does not allow benefiting from the improved spatial resolution brought by the small pixel size of the photon counting detector, while small focal spot does. As SPCCT is expected to be more dose efficient than energy integrating detector CT, owing to the counting mode versus the integrating mode and smaller detectors with no electronic noise, small focal spot is expected to provide enough X-ray flux for cardiac applications3,15. Our in vitro results for B64 are in agreement with Beohar and colleagues who found that both 16- and 64-MDCT scanners underestimated lumen size by as much as 1.2 mm in stents 3mm in diameter and strut thicknesses comparable to the ones used in the present study16.
Using a third generation scanner in combination with novel iterative reconstruction, visibility of on average 80% of the in-stent lumen was obtained for commonly used coronary stents inflated in plastic tubes to a diameter of 3mm17. These results are very encouraging with respect to our findings as they indicate the strength of iterative reconstruction, currently not available on our SPCCT system, which is expected to further improve stent visualization. This explains the current effort to develop advanced iterative reconstruction methods such as one-step algorithms18. We found an average underestimation of ~1.4mm for B64 and 0.9mm for SPCCT. The added value of the higher spatial resolution of SPCCT is a reduction of blooming artifacts to under a millimeter. The reduction in blooming artifacts allowed morphological assessment of the mesh of the stent even in the presence of calcification. This is of particular importance in diagnosing in-stent restenosis, one of the main complications of coronary artery stenting. In-stent restenosis is due to neo-intimal hyperplasia, proliferation of smooth muscle cells along the inner vessel wall. Strong blooming artifacts impair a correct diagnosis of neointimal hyperplasia. The apparent decrease in the inner lumen diameter may be wrongly interpreted as neointimal hyperplasia. Thus the strong reduction in stent related blooming artifacts provided by SPCCT is expected to improve detection of in-stent restenosis, currently impaired by strong blooming.
In vivo measured apparent strut widths were larger than in vitro, probably due to aortic pulsation-related motion artifacts. Nonetheless, the strut width was smaller with SPCCT compared to B64 also in the live animal. This capability is expected to improve detection of stent deployment problems, such as stent malapposition, and stent integrity degradation, e.g. fractures and “jailed struts” occurring after kissing balloon inflation procedure in bifurcation regions6. Stent fracture occurs in almost 30% of cases, as revealed by autopsy reports19, contrary to the 1–2% of cases revealed by imaging due to the current limited spatial resolution of standard CT.
Importantly, the additional capability of the PCD system presented in our study is k-edge imaging. K-edge images represent specific element maps, such for example the Pt specific K-edge images of the Pt containing stent (Promus Premier) presented here. We did not obtain a statistically significant difference between the apparent strut thicknesses in the conventional HU image and the Pt K-edge image. However, this is to be expected since Platinum is a coating material. The added value of the K-edge is the specificity of the image, i.e. the only thing visible is the Platinum, while other highly attenuating objects such as the vascular calcifications are not present, and by such do not hamper the visualization of the stent. It should be noted that Au markers of the stainless steel stent were also visible in the Pt-specific K-edge images. This is due to the fact that these metals have very close atomic numbers, ZPt=78 and ZAu=79 and thus their respective k-edge energies are very similar 78.4keV (Pt) and 80.7keV (Au). The high specificity of SPCCT for these materials enforces the added value of the SPCCT for coronary stent imaging. Material separation capabilities in elements such as gold was demonstrated previously for another photon-counting system using 4 energy levels20. It is to be noted that this advantage is only applicable to metals of relatively high atomic weights21–25. Stainless steel (Fe) and Co-Cr stents cannot be detected by specific K-edge imaging because their K-edge energies are smaller than 10keV, energies that are filtered from the incident photon spectrum in CT. However, this is not a limitation, since relatively high atomic weight metals, such as Pt, have been added to stent substrates to increase their radio-opacity and thus their visibility on radiological images, and to improve their bio-compatibility.
Several drawbacks could be identified regarding the current capabilities of the SPCCT prototype with respect to cardiac imaging: no ECG-gating is available, the rotation time of 1second is potentially long for dynamic acquisitions, and the z-collimation of 2.5mm limits the current use to small animals. It should be noted however, that these limitations will be overcome in the next generation of the system.
Methods
Spectral photon-counting computed tomography (SPCCT)
The SPCCT prototype system (Philips, Haifa, Israel) is based on a standard clinical CT gantry and X-ray tube but with a direct conversion semiconductor detector technology operated in single photon-counting mode with energy discrimination26. The detector characteristics are presented in Table 3. The scanner has the same geometry as a standard Philips iCT system, with source to detector distance of 1040mm and source to isocenter distance of 570mm (magnification of 1.825). The pixel size is 0.274×0.274 mm2 at isocenter and z-coverage of ~2.5mm (9 pixels). On this pre-clinical prototype, the maximum FOV is 168mm and the rotation time is 1s, enabling acquisition of 2400 projections/rotation. As references for comparison, we used two commercially available systems: a conventional Brilliance 64 CT (B64, Philips, Cleveland, USA) and a dual-energy (DE) CT (IQon, Philips Healthcare, Haifa, Israel). The IQon uses dual-layer detector to segregate low-energy photons from high-energy photons resulting in two raw-data sets that are used in reconstruction algorithms. This concept is different from other DECT systems in that it allows the CT user to process spectral data without need to select a specific dual-energy protocol prior to an examination.
Table 3
Pixel Size | 500µm pitch, flip-chip to sensor |
Observable count rate | >10 Mcps/pixel, Paralyzable |
Input referred noise | <350 |
Leakage compensation | Static 200nA Dynamic 60/600nA 3dB at 10kHz |
Thresholds | 5 |
Energy range | >160keV |
Energy resolution | 0.5keV/LSB |
Frame rate | >10kHz, zero dead-time |
Imaging protocols
Three types of coronary stents of different metal composition: Cobalt-Chromium (Stent1), Platinum (Pt)-Chromium (Stent2), and Stainless steel with gold (Au) markers (Stent3) (see Table 4) were imaged using the SPCCT prototype, the IQon and the B64 systems in vitro, and the SPCCT and the B64 in vivo, all with 120 kVp and 100 mAs tube settings. IQon and SPCCT have same tube filtration while B64 has additional Ti filter resulting with 25–30% lower dose.
Table 4
Stent 1 | Stent 2 | Stent 3 | |
---|---|---|---|
Commercial Name | Coroflex Blue Neo (Braun, Melsungen AG, Germany) | Promus Premier Monorail (Boston Scientific, France) | Tsunami Gold (Terumo Europe NV) |
Composition | Cobalt-Chromium | Platinum (Pt)-Chromium | Stainless steel with gold markers |
Strut thickness (mm) | 0.081 | 0.060 | 0.080 |
Length (mm) | 19 | 24 | 10 |
On the B64 and IQon, the acquisitions were performed in spiral mode using 64×0.625mm collimation, gantry rotation time of 1second, pitch of 0.56, and a large focal spot on the X-ray tube (standard resolution mode). On the SPCCT, the acquisitions were performed in axial step and shoot mode with 9×0.274mm collimation, rotation time of 1second and with small focal spot on the X-ray tube (high Resolution mode) to benefit from the potential spatial resolution improvement allowed by the small detector size.
The reconstructed FOV and voxel size were: 160mm and 0.2×0.2 in plane pixel size and slice thickness of 0.25mm on SPCCT using a “detailed” filter; and 154mm and 0.2×0.2 in plane pixel size, effective slice thickness of 0.67mm and increment of 0.33mm on IQon and B64, using the YA sharp recon kernel. For the SPCCT the photon counting thresholds were adjusted to allow for simultaneous detection of iodinated contrast material and platinum K-edge.
Reconstructions of SPCCT acquired data were performed using maximum-likelihood processing method of counting data to compute several basis material sinograms (water, Iodine, and Platinum)27. The individual sinograms are then reconstructed to obtain conventional HU images, water and Iodine material density maps and Platinum specific K-edge images.
In vitro imaging
The stents were deployed inside individual plastic tubes (internal lumen diameter of 3.5mm). Presence of calcification was simulated by using a mixture of Calcium Phosphate powder (25%), flour (45%), and water (30%). Small spheres (1–2mm in diameter) were obtained from the mixture and allowed to dry for 24hours and were then placed inside the plastic tubes before stent deployment. To assess the impact of the presence of vascular contrast agents on stent imaging, tubes were subsequently filled with water and with a standard iodinated contrast agent (a concentration of ~11mg/ml of elemental iodine (I)). The entire length of the stents was imaged for each type of filling on the three systems.
In vivo imaging
This study was approved by the local Institutional Animal Care and Use Committee (Council Directive No. 2010/63/UE on the protection of animals used for a scientific purpose) under the authorization n° APAFIS#1732-2015091411181645v3 and performed under relevant guidelines and regulations.
Similar stents were sequentially deployed in the abdominal aorta of one New Zealand White rabbit (NZW, male, 3.2kg) while under full anesthesia achieved with an intramuscular injection of medetomidine (0.4mg/kg) and ketamine (20mg/kg). An incision was performed at the level of the left hind paw to reach the iliac artery. The guidewire supporting the stent was then inserted in the artery up to the abdominal aorta at the level of the diaphragm and the first stent was deployed, followed by the other two stents. Stent placement was verified by DSA. During the procedure, the animal’s heart rate and oxygen saturation was monitored. At the end of the procedure the incision was sutured and anti-inflammatory cream was applied locally and the animal was allowed to recover for 2 weeks before the first imaging session.
The animal underwent two imaging sessions, one on SPCCT and one on B64. On SPCCT, cross-sectional views of each stent were acquired before and after manual intravenous injection of a standard iodinated contrast agent (400mg/ml of elemental I, IOMERON, Bracco, Italy). The three stents were imaged individually, each one approximately 5seconds after the administration of 5ml of IOMERON, equivalent to a high clinical dose of 1.6ml/kg, with at least a 5minute interval between the injections, to allow the blood concentration of iodine to decrease. On B64, one injection of 5ml was performed and the entire length of the abdominal aorta was imaged approximately 5seconds after the manual injection.
Image analysis
To evaluate the stent-related blooming artifacts, automatic measurements of the apparent width of the metallic struts and of the intra-stent lumen diameter were performed on images acquired in vitro in the calcification free region and in vivo, using in house developed code as follows: the center of the stent cross-section was identified on 8 consecutive images and intensity profiles were drawn radially center-out using 5° increments, resulting in 576 intensity profiles. The apparent strut width was obtained as the full-width at half maximum (FWHM) value of a Gaussian curve fitted to the interpolated intensity profile using least square fitting. Profiles located at a 180° angle from each other were paired and used to compute the inter-stent lumen diameter following identification of points corresponding to half the maximum intensity on each side of the curve. Then, the inter-stent lumen diameter was defined as the distance between the closest identified points with respect to the center (288 measurements). This analysis was performed on conventional HU images and on Pt-specific K-edge images acquired in water in vitro and before contrast agent administration in vivo.
Data is presented as mean±standard error of the mean.
Conclusion
SPCCT enables improved lumen delineation and visualization of stent metallic mesh owing to increased spectral resolution and increased spatial resolution that resulted in a significant reduction of blooming artifacts compared to dual-energy and conventional CT. Platinum specific K-edge images allow validation of correct deployment of Pt containing stents.
Acknowledgements
This work was supported with funding from the European Union H2020 program (#643694), France Life Imaging (FLI). We thank employees from the CERMEP-Imagerie du Vivant for helping with the experiment.
Author contributions
M.S. and S.S.M. contributed equally to this work. M.S., S.S.P., J.M., D.B.N., P.C.N., L.B. and P.D. conceived and designed the experiments. S.S.M., D.B.N., J.M. and J.B.L. performed the experiments. M.S., S.S.M. and D.B.N. analyzed the data. P.C., E.R., I.B. and M.R. contributed reagents/materials/analysis tools. All authors contributed to the writing of the manuscript.
Competing interests
P.C., I.B., M.R. and E.R. are employees of Philips Healthcare. S.S.M., M.S., P.D., L.B., J.B.L., J.M., D.B. and G.R. declare no potential conflict of interest
Footnotes
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
These authors contributed equally: Monica Sigovan and Salim Si-Mohamed.
References
Articles from Scientific Reports are provided here courtesy of Nature Publishing Group
Full text links
Read article at publisher's site: https://doi.org/10.1038/s41598-019-56427-6
Read article for free, from open access legal sources, via Unpaywall: https://www.nature.com/articles/s41598-019-56427-6.pdf
Citations & impact
Impact metrics
Citations of article over time
Alternative metrics
Smart citations by scite.ai
Explore citation contexts and check if this article has been
supported or disputed.
https://scite.ai/reports/10.1038/s41598-019-56427-6
Article citations
Potential benefits of photon counting detector computed tomography in pediatric imaging.
Br J Radiol, 96(1152):20230189, 25 Oct 2023
Cited by: 6 articles | PMID: 37750939 | PMCID: PMC10646626
Review Free full text in Europe PMC
Clinical Applications of Photon-counting CT: A Review of Pioneer Studies and a Glimpse into the Future.
Radiology, 309(1):e222432, 01 Oct 2023
Cited by: 12 articles | PMID: 37787672 | PMCID: PMC10623209
Review Free full text in Europe PMC
Cardiac imaging with photon counting CT.
Br J Radiol, 96(1152):20230407, 24 Oct 2023
Cited by: 7 articles | PMID: 37750856 | PMCID: PMC10646663
Review Free full text in Europe PMC
Photon-counting computed tomography of coronary and peripheral artery stents: a phantom study.
Sci Rep, 13(1):14806, 08 Sep 2023
Cited by: 2 articles | PMID: 37684412 | PMCID: PMC10491813
Impact of the confluence of cardiac motion and high spatial resolution on performance of ECG-gated imaging with an investigational photon-counting CT system: A phantom study.
Phys Med, 114:102683, 20 Sep 2023
Cited by: 0 articles | PMID: 37738807 | PMCID: PMC10798551
Go to all (24) article citations
Similar Articles
To arrive at the top five similar articles we use a word-weighted algorithm to compare words from the Title and Abstract of each citation.
First In-Human Results of Computed Tomography Angiography for Coronary Stent Assessment With a Spectral Photon Counting Computed Tomography.
Invest Radiol, 57(4):212-221, 01 Apr 2022
Cited by: 40 articles | PMID: 34711766 | PMCID: PMC8903215
Spectral Photon-Counting Computed Tomography for Coronary Stent Imaging: Evaluation of the Potential Clinical Impact for the Delineation of In-Stent Restenosis.
Invest Radiol, 55(2):61-67, 01 Feb 2020
Cited by: 35 articles | PMID: 31524765
Quarter-millimeter spectral coronary stent imaging with photon-counting CT: Initial experience.
J Cardiovasc Comput Tomogr, 12(6):509-515, 15 Oct 2018
Cited by: 30 articles | PMID: 30509378
First Experience With a Whole-Body Spectral Photon-Counting CT Clinical Prototype.
Invest Radiol, 58(7):459-471, 22 Feb 2023
Cited by: 6 articles | PMID: 36822663 | PMCID: PMC10259214
Review Free full text in Europe PMC