Diagnostic accuracy of sonication fluid cultures from prosthetic components in periprosthetic joint infection: an updated diagnostic meta-analysis

Abstract

Background

Periprosthetic joint infection (PJI) is the most serious complication following total joint arthroplasty (TJA) and has a significant impact on patients and the national healthcare system. To date, the diagnosis of PJI is still confronted with dilemmas. The present study investigated the validity of sonication fluid culture (SFC) for removing implants in the diagnosis of PJI after joint replacement.

Methods

From database establishment to December 2020, relevant literature was retrieved from the PubMed, Web of Science, Embase and Cochrane Library databases. Two reviewers independently performed quality assessment and data extraction to calculate the pooled sensitivity, specificity, positive likelihood ratio (PLR), negative likelihood ratio (NLR), area under the curve (AUC) and diagnostic odds ratio (DOR) to evaluate the diagnostic value of overall SFC for PJI.

Results

A total of 38 eligible studies including 6302 patients were selected in this study. The pooled sensitivity, specificity, PLR, NLR, and DOR of SFC for PJI diagnosis were 0.77 (95% confidence interval [CI], 0.76–0.79), 0.96 (95% CI, 0.95–0.96), 18.68 (95% CI, 11.92–29.28), 0.24 (95% CI, 0.21–0.29), and 85.65 (95% CI, 56.46–129.94), respectively, while the AUC was 0.92.

Conclusion

This meta-analysis showed that SFC was of great value in PJI diagnosis, and the evidence of SFC on PJI was more favorable but not yet strong. Therefore, improvement of the diagnostic accuracy of SFC is still necessary, and the diagnosis of PJI continues to warrant a multiplex approach before and during a revision procedure.

Introduction

Periprosthetic joint infection (PJI) after primary total joint arthroplasty (TJA) is regarded as a devastating complication [1], with an infection rate ranging from 0.88% to 2.18% [2, 3]. The prevalence of PJI in TJA is increasing as the number of joint replacements grows, leading to significant influences on the patient and the national healthcare system owing to multiple surgeries, prolonged hospital stays, patient suffering, and the high cost of treatments [4, 5]. Currently, the diagnosis of PJI depends on serum inflammatory markers, arthrocentesis-based studies, and intraoperative tissue cultures. However, traditional diagnostic techniques are limited in their capacities to detect low-grade infections caused by bacteria, which exist on the surface of implants in the form of biofilms rather than plankton [6]. Therefore, accurately identifying shielded microbes on the prosthesis surface and their respective sensitivities is a major challenge in the management of periprosthetic joint infections (PJIs).

Synovial fluid and intraoperative periprosthetic tissue cultures are considered to be the standard method for the diagnosis of PJI as defined by the Infectious Diseases Society of America (IDSA), the Musculoskeletal Infection Society (MSIS) and the International Consensus Meeting (ICM) [7,8,9]. However, the high rate of false negatives and low sensitivity of microbial cultures have attracted the attention of scholars. Studies have reported an unacceptably high false-negative rate for these procedures, ranging from 17 to 53% [10,11,12]. Trampuz et al. reported a low sensitivity of 61% (95% CI: 49–72) for microbial cultures [13]. In addition, microbial detection of polymicrobial infections by culture is as low as 13% to 17% [14]. Therefore, it is urgent for joint surgeons to find an efficient, inexpensive and convenient diagnostic method for PJI. In recent years, the role of sonication fluid culture (SFC) of the removed implant as a diagnostic tool of PJI continues to evolve. An inherent advantage of sonication is that it can destroy bacteria in biofilms and increase the number of culturable bacterial cells [15]. Several groups have demonstrated that SFC has higher sensitivity than periprosthetic tissue culture in the diagnosis of PJI, especially when antibiotics have been used shortly before explantation [13, 16,17,18,19]. In contrast, Van Diek et al. [20] reported that the sensitivity of ultrasound liquid analysis was lower than that of tissue culture, with 0.47 (95% CI: 0.35–0.59) and 0.68 (95% CI: 0.56–0.78), respectively. Currently, some studies have shown that the sensitivities (range 0.47 to 0.93) and specificities (range 0.67 to 1.0) of SFC were inconsistent in evaluating the diagnostic value of PJI [14, 15, 20,21,22,23,24,25,26,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41,42,43,44,45,46,47,48,49,50,51,52]. To our knowledge, there is no consensus as to the most appropriate tests for excluding PJI after primary TJA.

Therefore, we conducted a systematic review and meta-analysis to synthesize the available evidence on the accuracy of SFC in the diagnosis of PJI and provide further evidence for its clinical application.

Methods

Our study strictly followed the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines [53]. This protocol was determined by all the authors. We performed a literature search, screened the studies identified, the data statistics, the results combination, and the manuscript, and evaluated the studies that related to the application of SFC in PJI diagnosis.

Search strategy

We (Guanrong Peng and Qiang Liu) systematically searched online databases such as PubMed, Cochrane Library, Embase, and Web of Science (from the time of database inception to December 2020) under the guidance of the Cochrane review method. The medical subject headings (MeSH) and keywords were used as follows: “periprosthetic joint infection” or “prosthesis-related infection” or “periprosthetic infection” represents the disease, “ultrasonics” or “sonications” or “sonication” or “ultrasonic” stands for diagnostic method, “sensitive” or “sensitivity and specificity” or “predictive and value” or “predictive value of tests” or “accuracy” represents the research type. Throughout the retrieval process, only studies in English were included. To obtain valuable articles for this study, we also manually searched the reference lists of eligible studies and review articles after database screening. The full database search strategies for our study can be found in “Appendix”.

Eligibility criteria

We included all studies that reported the accuracy of SFC in the diagnosis of PJI after TJA. Two authors (Guanrong Peng and Qiang Liu) independently scanned the titles, abstracts and full texts sequentially, and eligible studies were included in this systematic review. When there was a disagreement between the two authors regarding inclusion, consensus was reached by consultation with another author (Professor Yirong Zeng). Inclusion criteria that eligible studies had to meet were as follows: (1) focusing on the diagnosis of SFC in PJI; (2) cutoff or range definitions of the tests; (3) reference standard, such as “Musculoskeletal Infection Society (MSIS)”, “International Consensus Meeting (ICM)”, “Infectious Diseases Society of America (IDSA)” or “European Bone and Joint Infection Society (EBJIS)”, and other culture or clinical diagnosis criteria; and (4) providing data (including true positive, false negative, false positive, and true negative) for completion of 2-by-2 tables. Studies lacking sensitivity and specificity values or having duplicate data were excluded (Fig. 1).

Fig. 1
figure 1

Flow diagram for study selection

Data extraction and quality assessment

Relevant information was extracted by the two reviewers (Guanrong Peng and Qiang Liu) independently from all selected studies with a standardized data collection form. The following information was abstracted: (1) study baseline characteristics (such as the author name, publication year, country, study design, time of follow-up, number ratio of non-PJI vs. PJI, sex distribution, average age, and joint type, among others); (2) study intervention characteristics (the cutoff value of SFC, diagnostic criteria, etc.); and (3) outcome indicators, including sensitivity and specificity. Then, the true-positive (TP), false-positive (FP), true-negative (TN) and false-negative (FN) data used to construct the 2-by-2 tables were further calculated. Finally, we obtained the positive likelihood ratio (PLR), negative likelihood ratio (NLR), area under the curve (AUC) and diagnostic odds ratio (DOR) of the SFC subgroup analysis for PJI diagnosis.

The risk of bias and concerns regarding applicability of each included study were independently assessed by utilizing the Quality Assessment of Diagnostic Accuracy Studies-2 (QUADAS-2) tool in Revman (version 5.3) software [54]. QUADAS-2 consists of four key domains: patient selection, index test, reference standard, and flow and timing. All domains of the QUADAS-2 were assessed for risk of bias, and the first three areas were also evaluated for clinical practicability. The risk of bias level was determined as “low”, “high” or “uncertain” according to the answers (“yes”, “no” or “uncertain”) to the relevant landmark questions contained in each domain. When the two authors disagreed, the third author made the final decision. The results of the QUADAS-2 evaluation of the studies included in the meta-analysis are shown in Fig. 2.

Fig. 2
figure 2

Quality assessment of included studies based on QUADAS-2 tool criteria

Statistical analysis

Meta-DiSc 1.4 software and Stata 14.0 were used for data analysis and image production. We calculated the values of TP, FP, FN, and TN according to the sensitivity and specificity of each eligible study to construct 2-by-2 tables. The bivariate random effect model was used to calculate the pooled sensitivity, specificity, PLR, NLR, AUC, DOR, and 95% confidence intervals (CI) for each study by using Meta-DiSc software.

Forest plots obtained from the random effect model were used to summarize the results and test the heterogeneity. The heterogeneity among studies was expressed as the inconsistency index (I2) statistic, with values ranging from 0 to 100%. High heterogeneity was defined as I2 values of > 75%; 50% ≤ I2 ≤ 75% indicated moderate heterogeneity and I2 < 50% indicated low heterogeneity. If large heterogeneity was caused by the threshold effect, the Spearman correlation coefficient of the logarithms of sensitivity and 1-specificity was performed to evaluate the threshold effect. The threshold effect was considered to have a significant difference at p < 0.05. When there was high heterogeneity, we conducted subgroup analysis to find the potential sources of heterogeneity. The type of study design, the sample type, the publication year, the threshold used in the study, the number of joints, the reference standard, and the country or region were considered to be the variables that may have caused heterogeneity. Then, the heterogeneity caused by the nonthreshold effect was detected by calculating the heterogeneity of the Chi-squared value (Cochran Q) and the inconsistency index (I2) of the DOR of each test. When I2 > 50% or p0.1, significant heterogeneity was considered among the studies, and we used a random effects model to pool the related effects and perform a subgroup analysis; otherwise, the fixed-effects model was used (I250% or p > 0.1). In addition, we drew a summarized receiver operating characteristic curve (SROC) and AUC. The higher values of the AUC indicate more accurate test results of SFC.

Furthermore, sensitivity analysis was performed for each eligible study using the “Midas” command of Stata software. Deeks' funnel plot was used to estimate publication bias, while the Fagan plot clearly displayed the change in diagnostic value of SFC for PJI. We performed the Z-test to compare sensitivity, specificity and value of the AUC among the different subgroups when p < 0.05 was considered statistically significant.

Results

Study selection

After systematically searching the aforementioned four databases, we initially obtained a total of 385 articles, with 199 from Web of Science, 128 from PubMed, 56 from Embase, and 2 from the Cochrane Library. A total of 139 duplicate studies were removed from these 385 studies, leaving 246 for title and abstract review. A total of 82 studies were eligible for full-text article review after reviewing the titles and abstracts. After reviewing the full text of each study in detail, another 44 studies were excluded. Eventually, a total of 38 studies were included in our systematic review and meta-analysis (Fig. 1).

Quality assessment

Two authors (Jinlun Chen and Xiaobo Sun) used the QUADAS-2 tool to assess the quality of all included studies. As shown in Fig. 2, two studies showed “high risk” for patient selection, while other studies showed “low risk” or “unclear” for patient selection, reference standard, index test, and flow and timing bias. For the index test and the reference standard bias, 14 (36.8%) and 16 studies (42.1%) showed unclear risk, respectively. The reason for this assessment was to interpret the results of the index test without knowing the results of the reference standard. Most of the studies in this meta-analysis were evaluated as “low risk” in the applicability section, and there was no “high risk”.

Study characteristics

These 38 studies [13,14,15,16,17, 20,21,22,23,24,25,26,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41,42,43,44,45,46,47,48,49,50,51,52] included 19 prospective studies, 17 retrospective studies and 2 not described in detail. Among the 38 studies, 8 reported MSIS criteria, 5 reported ICM criteria, 5 reported IDSA criteria, and 20 reported other reference standards. In terms of the diagnostic threshold, 10 studies used 50 colony-forming units (CFU)/ml, 6 used 20 CFU/plate, and the remaining 22 used other values. Twenty-five of the studies were from European countries, 9 were from the USA, and the other 4 were from Asian countries. All studies included in this meta-analysis were selected because they cultured the sonication fluids obtained from the prostheses removed by surgery after ultrasonic treatment. A total of 6302 patients were enrolled in our study, among whom 2290 (36.3%) were confirmed to have PJI. The detailed characteristics of all the studies are shown in Table 1. The data extraction results of each study are summarized in Table 2.

Table 1 Characteristics of the studies in meta-analysis for the diagnosis of PJI applying sonication
Table 2 Data extracted for the construction of 2 × 2 table

Heterogeneity analysis

The data extracted in this study were imported into Meta DiSc software for threshold effects analysis, and the Spearman correlation coefficient between logarithms of sensitivity and logarithms of (1-specificity) was 0.249 (p = 0.132 > 0.05), indicating that there was no threshold effect in this study. Meanwhile, by drawing the asymmetric SROC curve, there was no “shoulder arm shape”, which further showed that heterogeneity might be independent of the threshold effect in our study (Fig. 4b). Furthermore, the Cochran-Q test of DOR showed that Cochran-Q was 107.71 (p < 0.001), indicating heterogeneity caused by a nonthreshold effect.

Threshold and diagnostic accuracy of SFC for PJI

For the cutoff value, 10 and 6 studies used common thresholds of 50 CFU/ml and 20 CFU/plate, respectively, while the remaining 22 studies used different thresholds.

Significant heterogeneity was found in sensitivity (I2 = 74.7%, p < 0.01), specificity (I2 = 87.7%, p < 0.01), PLR (I2 = 87.3%, p < 0.01), NLR (I2 = 76.5%, p < 0.01) and DOR (I2 = 65.6%, p < 0.01) (Fig. 3, Fig. 4a, Fig. 5); thus, the random-effects model was performed. As shown in Table 3, the pooled sensitivity and specificity of SFC for diagnosing PJI were 0.77 (95% CI, 0.76–0.79) and 0.96 (95% CI, 0.95–0.96), respectively. The pooled DOR was 85.65 (95% CI, 56.46–129.94) (Fig. 4a). The AUC of SFC for PJI was 0.92 (95% CI, 0.89–0.94) (Fig. 4b).

Fig. 3
figure 3

Forest plots with sensitivity and specificity of sonication fluid cultures for PJI

Fig. 4
figure 4

Pooled diagnostic odds ratio (a) and SROC curve of included studies (b)

Fig. 5
figure 5

Forest plots of pooled likelihood ratio (positive likelihood ratio and negative likelihood ratio)

Table 3 Subgroup analysis of sonication fluid cultures for PJI diagnosis

Evaluation of the clinical utility

As shown in Table 3 and Fig. 6a, the pooled PLR and NLR of SFC for PJI diagnosis were 18.68 (95% CI, 11.92–29.28) and 0.24 (95% CI, 0.21–0.29), respectively. According to previous studies, the incidence of PJI is approximately 20% in revision arthroplasty. Therefore, a 0.2 pretest probability was used to calculate the posttest probability by the likelihood ratio and pretest probability [55]. The posttest probability of PJI was 5%, indicating negative SFC results (Fig. 6b).

Fig. 6
figure 6

Pooled likelihood ratio scatter diagrams (a) and Fagan’s nomogram of sonication fluid cultures for diagnosis of PJI (b)

Subgroup analysis

The Cochran-Q test of DOR showed that the heterogeneity was caused by a nonthreshold effect, and the heterogeneity of this study was large. Therefore, we performed the following subgroup analysis to explore the source of heterogeneity according to the study design, the sample type, the publication year, the threshold used in the study, the number of joints, reference standard, antibiotic use before revision, sample processing time, solution for prosthesis, centrifugation, vortexing, culture period, and geographical location. The subgroup analysis showing the pooled sensitivity, specificity, PLR, NLR, DOR, and AUC of each subgroup are presented in Table 3. The pooled sensitivity, specificity, PLR, NLR, DOR, and AUC estimates for detecting PJI by using SFC at 50 CFU/ml were 0.76 (95% CI, 0.72–0.79), 0.96 (95% CI, 0.95–0.97), 21.44 (95% CI, 9.90–46.42), 0.26 (95% CI, 0.17–0.40), 91.26 (95% CI, 42.43–196.31), and 0.94 (95% CI, 0.92–0.96), respectively. In the subgroup of the reference standard, the pooled sensitivity, specificity, and AUC were 0.81 (95% CI, 0.78–0.84), 0.94 (95% CI, 0.93–0.96), and 0.94 (95% CI, 0.92–0.96), respectively, for MSIS or ICM, while they were 0.73 (95% CI, 0.69–0.77), 0.99 (95% CI, 0.98–1.00), and 0.80 (95% CI, 0.77–0.83), respectively, for IDSA or EBJIS.

Sensitivity analysis and publication biases

As shown in Fig. 7, there were four original studies with relatively strong sensitivity, and other studies did not cause sensitivity in the pooled results. Overall, the results of our study should be relatively sound. The funnel plot based on Deeks’ test of the pooled DOR was asymmetric, indicating a possible publication bias (p = 0.01 < 0.05) (Fig. 8).

Fig. 7
figure 7

Sensitivity analysis: a Goodness of fit, b bivariate normality, c influence analysis, d outlier detection

Fig. 8
figure 8

Funnel plot for publication bias assessment of included studies

Discussion

As the diagnosis of PJI remains a challenge and PJI brings a huge economic burden to society, an increasing number of orthopaedists pay attention to this devastating complication after TJA. Accurate diagnosis is the key to treating PJI because it has a major influence on the direction of treatment (antibiotic use and surgical rehabilitation) and the course of intervention. Therefore, an increasing number of preoperative and intraoperative tests have been performed to diagnose PJI [7,8,9, 13, 27, 28, 32,33,34, 40]. Historically, intraoperative tissue cultures have been used as the gold standard. It is known that bacteria can exist in two main forms, planktonic and sessile. Banousi et al. [24] suggested that the biofilms formed by bacteria at the surface of implants are not only a major factor for chronic PJI but also one of the main causes for the lack of positive cultures in periprosthetic soft tissue samples obtained intraoperatively. Moreover, bacteria in the form of biofilms are resistant to antibiotics and difficult to detect by conventional tissue culture. Therefore, previous studies reported that intraoperative tissue cultures lacked adequate sensitivity (ranging from 0.51 to 0.90) and specificity (ranging from 0.67 to 1.00) [15, 20, 30, 34, 38, 56]. Moreover, Renz et al. [38] found that the sensitivity of intraoperative tissue cultures for the diagnosis of PJI was only 51.0%. Parvizi et al. [56] suggested that there was no significant difference in clinical features between chronic deep PJI and aseptic joint loosening. Fortunately, with the development of medicine, various detection methods have been found for the diagnosis of PJI, such as laboratory tests (e.g., white blood cell count, erythrocyte sedimentation rate, C-reactive protein, interleukin-6, alpha-defensin, procalcitonin, D-dimer, and fibrinogen), synovial fluid characteristics, histopathological studies, microbiological studies (e.g., conventional synovial fluid and tissue cultures, and SFC), and radiological studies. To the best of our knowledge, recovering qualified bacteria from samples is a prerequisite for improving diagnostic accuracy. As Janz et al. [47] showed, the true diagnostic ability of cultures depends on the accurate recovery of bacteria from samples. Fortunately, Tunney et al. [57] suggested that sonication can dislodge adherent bacteria from explanted prosthetic surfaces by ultrasound. In addition, a study by Trampuz et al. [13] confirmed that cultures of microorganisms from removed orthopaedic implants are more sensitive than tissue cultures. To date, many studies have reported that SFC improves the accuracy of PJI diagnosis [13, 15, 16, 18, 30, 35, 45, 48, 52]. Meanwhile, we found that SFC had a high sensitivity of 0.77 (95% CI, 0.76–0.79) and a very high specificity of 0.96 (95% CI, 0.95–0.96) for the diagnosis of PJI.

As an important diagnostic method for PJI, SFC is still widely used in clinical practice. Since this technique is simple, it can be employed in most microbiology laboratories. Trampuz et al.[13] proposed that another advantage of SFC is the improved detection rate of polymicrobial PJI; meanwhile, they also popularized the application of sonication to detect PJI and described a significant improvement in the sensitivity of SFCs (78.5%) when compared with periprosthetic tissue cultures (PTCs) (60.8%). Since then, SFCs have shown superior sensitivity as opposed to tissue cultures in the diagnosis of PJI, which has been reported in many previous studies [13, 15,16,17,18, 24, 30,31,32, 35, 44,45,46,47, 51, 52]. For example, Banousi et al.[24] reported on 234 patients with revision arthroplasty (91 with PJI), and SFCs were more sensitive than PTCs (91% vs. 43%), while specificities were similar (100% vs. 100%). In another study of 200 patients and 276 samples, the culture sensitivities of sonicate fluid, periprosthetic tissue, synovial fluid and a combination of periprosthetic tissue and/or synovial fluid were 69.5, 52.8, 54.8 and 60.2%, respectively; the specificities were 97.6, 90.3, 93.0 and 89.9%, respectively [35]. In a study of 503 patients who underwent revision total hip arthroplasty and total knee arthroplasty, Adam et al. [15] demonstrated that the sensitivity of SFC (0.97; 95% CI, 0.89–0.99) was greater than that of synovial fluid culture (0.57; 95% CI, 0.44–0.69) and PTC (0.70; 95% CI, 0.58–0.80), although the specificity of SFC was not significantly different from that of synovial culture or tissue culture (0.90, 95% CI 0.72–0.97 vs. 1.0, 95% CI 0.86–1.0; 0.90, 95% CI 0.72–0.97 vs. 0.97, 95% CI 0.81–1.0, respectively). Sonication of retrieved implants in particular seems to have additional value in the detection of low-virulent biofilm-producing microorganisms that cause chronic PJIs. Trampuz et al. [13] reported that the sensitivities of PTCs and SFCs were 60.8% and 78.5%, respectively, and found that SFCs detected 14 cases but not PTCs. One recent study claimed that SFC detected positive results in 8 of 87 patients with PJI, while PTC was negative, and Hoekstra et al. emphasized that SFC should be used to rule out infection in patients with suspected early aseptic loosening of the prosthesis and negative preoperative synovial fluid culture [23]. Furthermore, previous studies have found a strong trend toward greater sensitivity of SFC against PTC in chronic/delayed PJI [29, 31, 35, 58]. In addition, Prieto et al. [35] further revealed that SFC (91.3%) was more sensitive for all infection types of implants than PTC (60.0%), synovial fluid (63.2%) and a combination of PTC and/or synovial fluid (66.7%); when conventional cultures were combined with SFC, the sensitivity improved significantly in total (from 60.2 to 77.1%) and delayed PJI (from 45.1 to 71.7%). It is interesting to note that SFC significantly improves pathogen detection, especially for patients who were administered antimicrobial therapy [13, 26, 32], and the sensitivity of SFCs was significantly superior to that of PTCs in patients who received antibacterial treatment within 14 days before operation (75.0% vs. 45.0%, p < 0.001), as reported by Trampuz et al. [13].

Various factors, including antibiotic use and specimen contamination, may affect the accuracy of SFC diagnosis [45, 47, 49, 52]. Meanwhile, in our subgroup analysis, we found that the diagnostic ability of SFC varied with the study design, cutoff value (CFU value), reference standard, sample type (such as hip, knee, shoulder and elbow), number of joints, antibiotic use before revision, sample processing time, solution used for prosthesis, centrifugation, vortexing, culture period, geographic location and year of publication (Table 3). It is well known that traditional methods such as PTCs also have the same disadvantages for PJI in patients with these conditions.

Since no consensus has been reached on the use of a single threshold to date, different cutoff values of SFC (0.1–50 CFU/mL) were applied in recent studies to diagnose PJI [13,14,15,16,17, 20,21,22,23,24,25,26,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41,42,43,44,45,46,47,48,49,50,51,52]. For example, due to different infection criteria, the culture protocols and reference standards in the study of Piper et al. and Grosso et al. differed [16, 37]. Using a cutoff of 20 CFU/mL, Piper et al. [16] reported an increased SFC sensitivity compared with PTC (66.7% vs. 54.5%), and specificities were not significantly different (98% vs. 95.1%), whereas a higher sensitivity (96%) and lower specificity (75%) for standard intraoperative cultures as well as a lower sensitivity for SFCs (56%) were reported by Grosso et al. [37]. Our results are similar to those of Piper et al., and we found a higher sensitivity (78%), and specificity (99%) for SFCs in this meta-analysis at the 20 CFU/plate cutoff. In our study, the superior sensitivity and specificity of SFC for detecting hip and knee prosthesis infections were 0.78 (95% CI, 0.76–0.81), 0.97 (95% CI, 0.96–0.97), respectively. In fact, previous studies have concluded that SFCs as a successful diagnostic tool may be most beneficial for detecting organisms in this population of patients [13, 58, 59]. Because the PJI diagnostic threshold of SFC was different for various joints, the sensitivity and specificity of diagnosis were also different. Therefore, the appropriate cutoff value of SFC in the diagnosis of PJI still needs to be studied.

In addition, the findings of comparable studies confirmed that previous administration of antibiotics had no effect on the sensitivity of SFC [60, 61]. However, our subgroup analysis showed a trend toward higher sensitivity and specificity of SFC in the antibiotic groups as opposed to when antibiotics were not administered (79% vs. 74%, 97% vs. 94%, respectively). In this meta-analysis, it must be noted that the antibiotic group was characterized by a relatively large sample size (n = 3799). Furthermore, the criteria used for the definition of PJI differed between studies, including MSIS, ICM, IDSA, EBJIS, and so on. In our study, higher sensitivity was shown when using MSIS or ICM as a reference standard (0.81; 95% CI, 0.78–0.84). To the best of our knowledge, the duration of the incubation period remains controversial. In our study, two- to 14-day incubations were reported [13,14,15,16,17, 20,21,22,23,24,25,26,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41,42,43,44,45,46,47,48,49,50,51,52]. For PJI caused by low-virulence microorganisms, however, most studies suggested prolonging the anaerobic culture period for up to 14 days [14, 15, 17, 21, 22, 24, 25, 29, 32,33,34, 36, 38, 42,43,44,45,46,47, 49,50,51]. By comparing the time to culture tissue and sonication fluid in PJI and aseptic failure, Butler-Wu SM et al. [62] found that anaerobic organisms were detected in SFC up to 13 days, whereas aerobes required only 7 days of incubation. In this meta-analysis, 10 studies detected aerobic and anaerobic bacteria in SFC within 7 days, 15 studies detected aerobic bacteria within 7 days and anaerobic bacteria within 7–14 days, and aerobic and anaerobic bacteria in SFC were detected in 10 studies within 7–14 days. In addition, the subgroup analysis showed that patients with the second condition (aerobic (≤ 7 days) and anaerobic (7–14 days) culture were performed in SFC) had the highest sensitivity of 0.81 (95% CI, 0.79–0.84). Therefore, our results supported that 7-day aerobic culture and 7–14-day anaerobic culture may improve the sensitivity of PJI diagnosis. Of note, previous studies [13, 17, 58, 59, 63] have reported that SFC has gained acceptance in detecting hip and knee prosthetic infections. Nevertheless, Carlos et al. [21] did not support the routine use of SFC to detect infection in shoulder implants. Consistent with previous studies, in our study, SFC had the best sensitivity and specificity for detecting hip and knee prosthesis infection, with 0.79 (95% CI, 0.76–0.81) and 0.97 (95% CI, 0.96–0.97), respectively. Additionally, in this meta-analysis, 34 studies recorded the implants that were covered with solution. It should be mentioned that Ringer's solution was the most commonly used solution (23/38, 60.5%), with a relatively high sensitivity and specificity of 0.79 (95% CI, 0.76–0.81), and 0.95 (95% CI, 0.96–0.97), respectively. Portillo et al. [45] reported that the vortexing-sonication procedure demonstrated higher biofilm removal efficiency than vortexing alone. In this study, compared to SFC without the use of centrifugation or vortexing, SFC with the use of centrifugation or vortexing had a higher sensitivity (centrifugation, 0.78 [CI, 0.76–0.80] versus 0.74 [CI, 0.70–0.78] and vortexing, 0.78 [CI, 0.76–0.80] versus 0.73 [CI, 0.67–0.79], respectively) and specificity (centrifugation, 0.96 [CI, 0.95–0.97] versus 0.93 [CI, 0.91–0.95] and vortexing, 0.96 [CI, 0.95–0.97] versus 0.87 [CI, 0.82–0.91], respectively).

Although an excellent meta-analysis on SFC in the diagnosis of PJI has been published by Zhai et al. [64], we noted that their four articles were inconsistent with other studies and should not be included in the meta-analysis. One of their studies involved 16S rRNA gene PCR analysis for the diagnosis of PJI [65]. One study involved multiplex PCR of sonication fluid for the diagnosis of PJI [66]. The subjects of the other two studies were not only considering joint prostheses but also other orthopaedic implants, such as fixation devices and spinal devices[58, 67]. Therefore, the reliability of their conclusions may be limited. Compared with Zhai et al. [64], our meta-analysis included more up-to-date studies and patients (6302 patients in 38 studies) after rigorous screening and literature quality evaluations. In addition, our pooled AUC of SFC for the diagnosis of PJI was 0.92 (95% CI, 0.89–0.94), which was higher than the results of Zhai et al. (AUC = 0.89). Furthermore, the overall pooled sensitivity and specificity were calculated to be 0.77 and 0.96, respectively. However, heterogeneity still exists in our study. For example, in the 4 studies from 2020, inconsistent results are shown in "Forest plots with sensitivity and specificity of sonication fluid cultures for PJI" (Fig. 3) and "Pooled Diagnostic Odds Ratio (a) and SROC curve of included studies (b)" (Fig. 4). LR and DOR have been generally used to demonstrate the validity of diagnostic indicators [68]. In fact, a previous guideline clearly defined PLR > 2, NLR < 0.5, or DOR > 4 as a viable predictor, while PLR > 5, NLR < 0.2, or DOR > 10 was considered a good predictor [69]. The pooled PLR, NLR and DOR of this meta-analysis were 18.68, 0.24 and 85.65, respectively. From the results of our study, evidence of SFC on PJI was more favorable but not yet strong. Therefore, there is still much work to be done in the future, and more time and evidence are needed to prove that SFC is a reliable detection method and of great value in PJI diagnosis.

Admittedly, this meta-analysis has certain limitations. First, there is still no established gold standard for the detection of PJI. Individual studies used different reference standards, which may affect the diagnostic accuracy of a test method. Second, there was significant heterogeneity among the included studies because they were completed in different institutions and used different test methods or sample sources. Third, only 23 studies recorded the threshold, and the optimal diagnostic threshold of SFC could not be calculated due to incomplete original data. Finally, nearly half of the included studies were retrospective studies, which may reduce the strength of our study conclusions.

Conclusion

In summary, our findings were somewhat in line with a previous meta-analysis published by Zhai et al. Based on the results of the current study, we conclude that the evidence of SFC on PJI is more favorable but not yet strong because the results from recent publications (until 2020 in our study) are still inconsistent. Therefore, improvement of the diagnostic accuracy of SFC is still necessary, and the diagnosis of PJI continues to warrant a multiplex approach before and during a revision procedure. Additionally, another meta-analysis on this topic should be conducted after years.

Availability of data and materials

The datasets used and/or analyzed during this study are not publicly available due to feasibility, but are available from the corresponding authors on reasonable request.

Abbreviations

PJI:

Periprosthetic joint infection

TJA:

Total joint arthroplasty

IDSA:

Infectious Diseases Society of America

MSIS:

Musculoskeletal Infection Society

ICM:

International Consensus Meeting

SFC:

Sonication fluid culture

PTC:

Periprosthetic tissue cultures

TP:

True positive

FP:

False positive

FN:

True negative

TN:

False negative

PLR:

Positive likelihood ratio

NLR:

Negative likelihood ratio

DOR:

Diagnostic odds ratio

AUC:

Area under the curve

QUADAS-2:

Quality Assessment of Diagnostic Accuracy Studies-2

CFU:

Colony-forming unit

References

  1. Boyle KK, Kapadia M, Landy DC, Henry MW, Miller AO, Westrich GH. Utilization of debridement, antibiotics, and implant retention for infection after total joint arthroplasty over a decade in the United States. J Arthroplasty. 2020;35(8):2210–6.

    Article  PubMed  Google Scholar 

  2. Tande AJ, Patel R. Prosthetic joint infection. Clin Microbiol Rev. 2014;27(2):302–45.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Marang-van de Mheen PJ, Bragan Turner E, Liew S, Mutalima N, Tran T, Rasmussen S, Nelissen R, Gordon A. Variation in Prosthetic Joint Infection and treatment strategies during 45 years of follow-up after primary joint arthroplasty using administrative data of 41397 patients across Australian, European and United States hospitals. BMC Musculoskelet Disord. 2017;18(1):207.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Zmistowski B, Karam JA, Durinka JB, Casper DS, Parvizi J. Periprosthetic joint infection increases the risk of one-year mortality. J Bone Joint Surg Am. 2013;95(24):2177–84.

    Article  PubMed  Google Scholar 

  5. Kurtz SM, Lau E, Watson H, Schmier JK, Parvizi J. Economic burden of periprosthetic joint infection in the United States. J Arthroplasty. 2012;27(8 Suppl):61-65.e61.

    Article  PubMed  Google Scholar 

  6. Costerton JW, Stewart PS, Greenberg EP. Bacterial biofilms: a common cause of persistent infections. Science. 1999;284(5418):1318–22.

    Article  CAS  PubMed  Google Scholar 

  7. Osmon DR, Berbari EF, Berendt AR, Lew D, Zimmerli W, Steckelberg JM, Rao N, Hanssen A, Wilson WR. Diagnosis and management of prosthetic joint infection: clinical practice guidelines by the Infectious Diseases Society of America. Clin Infect Dis. 2013;56(1):e1–25.

    Article  PubMed  Google Scholar 

  8. Parvizi J, Zmistowski B, Berbari EF, Bauer TW, Springer BD, Della Valle CJ, Garvin KL, Mont MA, Wongworawat MD, Zalavras CG. New definition for periprosthetic joint infection: from the Workgroup of the Musculoskeletal Infection Society. Clin Orthop Relat Res. 2011;469(11):2992–4.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Parvizi J, Gehrke T. Definition of periprosthetic joint infection. J Arthroplasty. 2014;29(7):1331.

    Article  PubMed  Google Scholar 

  10. Atkins BL, Athanasou N, Deeks JJ, Crook DW, Simpson H, Peto TE, McLardy-Smith P, Berendt AR. Prospective evaluation of criteria for microbiological diagnosis of prosthetic-joint infection at revision arthroplasty. The OSIRIS Collaborative Study Group. J Clin Microbiol. 1998;36(10):2932–9.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  11. Spangehl MJ, Masri BA, O’Connell JX, Duncan CP. Prospective analysis of preoperative and intraoperative investigations for the diagnosis of infection at the sites of two hundred and two revision total hip arthroplasties. J Bone Joint Surg Am. 1999;81(5):672–83.

    Article  CAS  PubMed  Google Scholar 

  12. Baré J, MacDonald SJ, Bourne RB. Preoperative evaluations in revision total knee arthroplasty. Clin Orthop Relat Res. 2006;446:40–4.

    Article  PubMed  Google Scholar 

  13. Trampuz A, Piper KE, Jacobson MJ, Hanssen AD, Unni KK, Osmon DR, Mandrekar JN, Cockerill FR, Steckelberg JM, Greenleaf JF, et al. Sonication of removed hip and knee prostheses for diagnosis of infection. N Engl J Med. 2007;357(7):654–63.

    Article  CAS  PubMed  Google Scholar 

  14. Portillo ME, Salvadó M, Sorli L, Alier A, Martínez S, Trampuz A, Gómez J, Puig L, Horcajada JP. Multiplex PCR of sonication fluid accurately differentiates between prosthetic joint infection and aseptic failure. J Infect. 2012;65(6):541–8.

    Article  PubMed  Google Scholar 

  15. Rothenberg AC, Wilson AE, Hayes JP, O’Malley MJ, Klatt BA. Sonication of arthroplasty implants improves accuracy of periprosthetic joint infection cultures. Clin Orthop Relat Res. 2017;475(7):1827–36.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Piper KE, Jacobson MJ, Cofield RH, Sperling JW, Sanchez-Sotelo J, Osmon DR, McDowell A, Patrick S, Steckelberg JM, Mandrekar JN, et al. Microbiologic diagnosis of prosthetic shoulder infection by use of implant sonication. J Clin Microbiol. 2009;47(6):1878–84.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Portillo ME, Salvadó M, Alier A, Martínez S, Sorli L, Horcajada JP, Puig L. Advantages of sonication fluid culture for the diagnosis of prosthetic joint infection. J Infect. 2014;69(1):35–41.

    Article  PubMed  Google Scholar 

  18. Portillo ME, Salvadó M, Trampuz A, Siverio A, Alier A, Sorli L, Martínez S, Pérez-Prieto D, Horcajada JP, Puig-Verdie L. Improved diagnosis of orthopedic implant-associated infection by inoculation of sonication fluid into blood culture bottles. J Clin Microbiol. 2015;53(5):1622–7.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Bjarnsholt T, Ciofu O, Molin S, Givskov M, Høiby N. Applying insights from biofilm biology to drug development—Can a new approach be developed? Nat Rev Drug Discov. 2013;12(10):791–808.

    Article  CAS  PubMed  Google Scholar 

  20. Van Diek FM, Albers CGM, Van Hooff ML, Meis JF, Goosen JHM. Low sensitivity of implant sonication when screening for infection in revision surgery. Acta Orthop. 2017;88(3):294–9.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Torrens C, Fraile A, Santana F, Puig L, Alier A. Sonication in shoulder surgery: Is it necessary? Int Orthop. 2020;44(9):1755–9.

    Article  PubMed  Google Scholar 

  22. Randau TM, Molitor E, Fröschen FS, Hörauf A, Kohlhof H, Scheidt S, Gravius S, Hischebeth GT. The performance of a dithiothreitol-based diagnostic system in diagnosing periprosthetic joint infection compared to sonication fluid cultures and tissue biopsies. Zeitschrift fur Orthopadie und Unfallchirurgie. 2021;159(4):447–53.

    Article  PubMed  Google Scholar 

  23. Hoekstra M, Veltman ES, Nurmohamed R, van Dijk B, Rentenaar RJ, Vogely HC, van der Wal BCH. Sonication leads to clinically relevant changes in treatment of periprosthetic hip or knee joint infection. J Bone Joint Infect. 2020;5(3):128–32.

    Article  Google Scholar 

  24. Banousi A, Evangelopoulos DS, Stylianakis A, Fandridis E, Chatziioannou S, Sipsas NV, Pneumaticos SG. A comparative study of heterogeneous antibiotic resistance of microbial populations in conventional periprosthetic tissue cultures and sonication fluid cultures of orthopaedics explanted prostheses. Eur J Orthop Surg Traumatol. 2020;30(7):1307–18.

    Article  PubMed  Google Scholar 

  25. Akgün D, Maziak N, Plachel F, Siegert P, Minkus M, Thiele K, Moroder P. The role of implant sonication in the diagnosis of periprosthetic shoulder infection. J Shoulder Elbow Surg. 2020;29(6):e222–8.

    Article  PubMed  Google Scholar 

  26. Ueda N, Oe K, Nakamura T, Tsuta K, Iida H, Saito T. Sonication of extracted implants improves microbial detection in patients with orthopedic implant-associated infections. J Arthroplasty. 2019;34(6):1189–96.

    Article  PubMed  Google Scholar 

  27. <New Methods in the Diagnosis of Prosthetic Joint Infection.pdf>.

  28. Di Benedetto P, Dalla Vecchia G, Dante F, Gisonni R, Cainero V, Causero A. Leukocyte esterase strip test as a reliable intraoperative PJIs biomarker. Our experience. Acta Bio-medica Atenei Parmensis. 2019;90(12):43–7.

    PubMed  PubMed Central  Google Scholar 

  29. Bellova P, Knop-Hammad V, Königshausen M, Mempel E, Frieler S, Gessmann J, Schildhauer TA, Baecker H. Sonication of retrieved implants improves sensitivity in the diagnosis of periprosthetic joint infection. BMC Musculoskelet Disord. 2019;20(1):623.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Yan Q, Karau MJ, Greenwood-Quaintance KE, Mandrekar JN, Osmon DR, Abdel MP, Patel R. Comparison of diagnostic accuracy of periprosthetic tissue culture in blood culture bottles to that of prosthesis sonication fluid culture for diagnosis of prosthetic joint infection (PJI) by use of Bayesian Latent Class Modeling and IDSA PJI Criteria for Classification. J Clin Microbiol. 2018;56(6):e00319.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  31. Tani S, Lepetsos P, Stylianakis A, Vlamis J, Birbas K, Kaklamanos I. Superiority of the sonication method against conventional periprosthetic tissue cultures for diagnosis of prosthetic joint infections. Eur J Orthop Surg Traumat. 2018;28(1):51–7.

    Article  Google Scholar 

  32. Stylianakis A, Schinas G, Thomaidis PC, Papaparaskevas J, Ziogas DC, Gamaletsou MN, Daikos GL, Pneumaticos S, Sipsas NV. Combination of conventional culture, vial culture, and broad-range PCR of sonication fluid for the diagnosis of prosthetic joint infection. Diagn Microbiol Infect Dis. 2018;92(1):13–8.

    Article  CAS  PubMed  Google Scholar 

  33. Sebastian S, Malhotra R, Sreenivas V, Kapil A, Chaudhry R, Dhawan B. Sonication of orthopaedic implants: A valuable technique for diagnosis of prosthetic joint infections. J Microbiol Methods. 2018;146:51–4.

    Article  PubMed  Google Scholar 

  34. Sambri A, Cadossi M, Giannini S, Pignatti G, Marcacci M, Neri MP, Maso A, Storni E, Gamberini S, Naldi S, et al. Is treatment with dithiothreitol more effective than sonication for the diagnosis of prosthetic joint infection? Clin Orthop Relat Res. 2018;476(1):137–45.

    Article  PubMed  Google Scholar 

  35. Prieto-Borja L, Auñón Á, Blanco A, Fernández-Roblas R, Gadea I, García-Cañete J, Parrón R, Esteban J. Evaluation of the use of sonication of retrieved implants for the diagnosis of prosthetic joint infection in a routine setting. Eur J Clin Microbiol Infect Dis. 2018;37(4):715–22.

    Article  PubMed  Google Scholar 

  36. Huang Z, Wu Q, Fang X, Li W, Zhang C, Zeng H, Wang Q, Lin J, Zhang W. Comparison of culture and broad-range polymerase chain reaction methods for diagnosing periprosthetic joint infection: analysis of joint fluid, periprosthetic tissue, and sonicated fluid. Int Orthop. 2018;42(9):2035–40.

    Article  PubMed  Google Scholar 

  37. Grosso MJ, Frangiamore SJ, Yakubek G, Bauer TW, Iannotti JP, Ricchetti ET. Performance of implant sonication culture for the diagnosis of periprosthetic shoulder infection. J Shoulder Elbow Surg. 2018;27(2):211–6.

    Article  PubMed  Google Scholar 

  38. Renz N, Feihl S, Cabric S, Trampuz A. Performance of automated multiplex PCR using sonication fluid for diagnosis of periprosthetic joint infection: a prospective cohort. Infection. 2017;45(6):877–84.

    Article  CAS  PubMed  Google Scholar 

  39. Prieto-Borja L, Rodriguez-Sevilla G, Auñon A, Pérez-Jorge C, Sandoval E, Garcia-Cañete J, Gadea I, Fernandez-Roblas R, Blanco A, Esteban J. Evaluation of a commercial multiplex PCR (Unyvero i60(®)) designed for the diagnosis of bone and joint infections using prosthetic-joint sonication. Enferm Infecc Microbiol Clin. 2017;35(4):236–42.

    Article  PubMed  Google Scholar 

  40. Fernández-Sampedro M, Fariñas-Alvarez C, Garces-Zarzalejo C, Alonso-Aguirre MA, Salas-Venero C, Martínez-Martínez L, Fariñas MC. Accuracy of different diagnostic tests for early, delayed and late prosthetic joint infection. BMC Infect Dis. 2017;17(1):592.

    Article  PubMed  PubMed Central  Google Scholar 

  41. Fernandez-Sampedro M, Salas-Venero C, Fariñas-Álvarez C, Sumillera M, Pérez-Carro L, Fakkas-Fernandez M, Gómez-Román J, Martínez-Martínez L, Fariñas MC. 26Postoperative diagnosis and outcome in patients with revision arthroplasty for aseptic loosening. BMC Infect Dis. 2015;15:232.

    Article  PubMed  PubMed Central  Google Scholar 

  42. Ryu SY, Greenwood-Quaintance KE, Hanssen AD, Mandrekar JN, Patel R. Low sensitivity of periprosthetic tissue PCR for prosthetic knee infection diagnosis. Diagn Microbiol Infect Dis. 2014;79(4):448–53.

    Article  PubMed  Google Scholar 

  43. Dapunt U, Lehner B, Burckhardt I, Zimmermann S, Hänsch GM, Ewerbeck V. Evaluation of implant sonication as a diagnostic tool in implant-associated infections. J Appl Biomater Funct Mater. 2014;12(3):135–40.

    PubMed  Google Scholar 

  44. Bogut A, Niedźwiadek J, Kozioł-Montewka M, Strzelec-Nowak D, Blacha J, Mazurkiewicz T, Macias J, Marczyński W. Sonication as a diagnostic approach used to investigate the infectious etiology of prosthetic hip joint loosening. Pol J Microbiol. 2014;63(3):299–306.

    Article  PubMed  Google Scholar 

  45. Portillo ME, Salvadó M, Trampuz A, Plasencia V, Rodriguez-Villasante M, Sorli L, Puig L, Horcajada JP. Sonication versus vortexing of implants for diagnosis of prosthetic joint infection. J Clin Microbiol. 2013;51(2):591–4.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  46. Janz V, Wassilew GI, Hasart O, Tohtz S, Perka C. Improvement in the detection rate of PJI in total hip arthroplasty through multiple sonicate fluid cultures. J Orthop Res. 2013;31(12):2021–4.

    Article  CAS  PubMed  Google Scholar 

  47. Janz V, Wassilew GI, Hasart O, Matziolis G, Tohtz S, Perka C. Evaluation of sonicate fluid cultures in comparison to histological analysis of the periprosthetic membrane for the detection of periprosthetic joint infection. Int Orthop. 2013;37(5):931–6.

    Article  PubMed  PubMed Central  Google Scholar 

  48. Drago L, Signori V, De Vecchi E, Vassena C, Palazzi E, Cappelletti L, Romanò D, Romanò CL. Use of dithiothreitol to improve the diagnosis of prosthetic joint infections. J Orthop Res. 2013;31(11):1694–9.

    Article  CAS  PubMed  Google Scholar 

  49. Cazanave C, Greenwood-Quaintance KE, Hanssen AD, Karau MJ, Schmidt SM, Gomez Urena EO, Mandrekar JN, Osmon DR, Lough LE, Pritt BS, et al. Rapid molecular microbiologic diagnosis of prosthetic joint infection. J Clin Microbiol. 2013;51(7):2280–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  50. Gomez E, Cazanave C, Cunningham SA, Greenwood-Quaintance KE, Steckelberg JM, Uhl JR, Hanssen AD, Karau MJ, Schmidt SM, Osmon DR, et al. Prosthetic joint infection diagnosis using broad-range PCR of biofilms dislodged from knee and hip arthroplasty surfaces using sonication. J Clin Microbiol. 2012;50(11):3501–8.

    Article  PubMed  PubMed Central  Google Scholar 

  51. Vergidis P, Greenwood-Quaintance KE, Sanchez-Sotelo J, Morrey BF, Steinmann SP, Karau MJ, Osmon DR, Mandrekar JN, Steckelberg JM, Patel R. Implant sonication for the diagnosis of prosthetic elbow infection. J Shoulder Elbow Surg. 2011;20(8):1275–81.

    Article  PubMed  PubMed Central  Google Scholar 

  52. Trampuz A, Piper KE, Hanssen AD, Osmon DR, Cockerill FR, Steckelberg JM, Patel R. Sonication of explanted prosthetic components in bags for diagnosis of prosthetic joint infection is associated with risk of contamination. J Clin Microbiol. 2006;44(2):628–31.

    Article  PubMed  PubMed Central  Google Scholar 

  53. Moher D, Liberati A, Tetzlaff J, Altman DG. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. J Clin Epidemiol. 2009;62(10):1006–12.

    Article  PubMed  Google Scholar 

  54. Whiting PF, Rutjes AW, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, Leeflang MM, Sterne JA, Bossuyt PM. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011;155(8):529–36.

    Article  PubMed  Google Scholar 

  55. Xie K, Dai K, Qu X, Yan M. Serum and synovial fluid interleukin-6 for the diagnosis of periprosthetic joint infection. Sci Rep. 2017;7(1):1496.

    Article  PubMed  PubMed Central  Google Scholar 

  56. Parvizi J, Adeli B, Zmistowski B, Restrepo C, Greenwald AS. Management of periprosthetic joint infection: the current knowledge: AAOS exhibit selection. J Bone Joint Surg Am. 2012;94(14):e104.

    Article  PubMed  Google Scholar 

  57. Tunney MM, Patrick S, Gorman SP, Nixon JR, Anderson N, Davis RI, Hanna D, Ramage G. Improved detection of infection in hip replacements: a currently underestimated problem. J Bone Joint Surg Br Vol. 1998;80(4):568–72.

    Article  CAS  Google Scholar 

  58. Puig-Verdié L, Alentorn-Geli E, González-Cuevas A, Sorlí L, Salvadó M, Alier A, Pelfort X, Portillo ME, Horcajada JP. Implant sonication increases the diagnostic accuracy of infection in patients with delayed, but not early, orthopaedic implant failure. Bone Joint J. 2013;95(2):244–9.

    Article  PubMed  Google Scholar 

  59. Janz V, Wassilew GI, Kribus M, Trampuz A, Perka C. Improved identification of polymicrobial infection in total knee arthroplasty through sonicate fluid cultures. Arch Orthop Trauma Surg. 2015;135(10):1453–7.

    Article  CAS  PubMed  Google Scholar 

  60. Sampedro MF, Huddleston PM, Piper KE, Karau MJ, Dekutoski MB, Yaszemski MJ, Currier BL, Mandrekar JN, Osmon DR, McDowell A, et al. A biofilm approach to detect bacteria on removed spinal implants. Spine. 2010;35(12):1218–24.

    Article  PubMed  Google Scholar 

  61. Scorzolini L, Lichtner M, Iannetta M, Mengoni F, Russo G, Panni AS, Vasso M, Bove M, Villani C, Mastroianni CM, et al. Sonication technique improves microbiological diagnosis in patients treated with antibiotics before surgery for prosthetic joint infections. New Microbiol. 2014;37(3):321–8.

    PubMed  Google Scholar 

  62. Butler-Wu SM, Burns EM, Pottinger PS, Magaret AS, Rakeman JL, Matsen FA 3rd, Cookson BT. Optimization of periprosthetic culture for diagnosis of Propionibacterium acnes prosthetic joint infection. J Clin Microbiol. 2011;49(7):2490–5.

    Article  PubMed  PubMed Central  Google Scholar 

  63. Holinka J, Bauer L, Hirschl AM, Graninger W, Windhager R, Presterl E. Sonication cultures of explanted components as an add-on test to routinely conducted microbiological diagnostics improve pathogen detection. J Orthop Res. 2011;29(4):617–22.

    Article  PubMed  Google Scholar 

  64. Zhai Z, Li H, Qin A, Liu G, Liu X, Wu C, Li H, Zhu Z, Qu X, Dai K. Meta-analysis of sonication fluid samples from prosthetic components for diagnosis of infection after total joint arthroplasty. J Clin Microbiol. 2014;52(5):1730–6.

    Article  PubMed  PubMed Central  Google Scholar 

  65. Bjerkan G, Witsø E, Nor A, Viset T, Løseth K, Lydersen S, Persen L, Bergh K. A comprehensive microbiological evaluation of fifty-four patients undergoing revision surgery due to prosthetic joint loosening. J Med Microbiol. 2012;61(Pt 4):572–81.

    Article  CAS  PubMed  Google Scholar 

  66. Achermann Y, Vogt M, Leunig M, Wüst J, Trampuz A. Improved diagnosis of periprosthetic joint infection by multiplex PCR of sonication fluid from removed implants. J Clin Microbiol. 2010;48(4):1208–14.

    Article  PubMed  PubMed Central  Google Scholar 

  67. Esteban J, Alonso-Rodriguez N, del-Prado G, Ortiz-Pérez A, Molina-Manso D, Cordero-Ampuero J, Sandoval E, Fernández-Roblas R, Gómez-Barrena E. PCR-hybridization after sonication improves diagnosis of implant-related infection. Acta Orthop. 2012;83(3):299–304.

  68. Glas AS, Lijmer JG, Prins MH, Bonsel GJ, Bossuyt PM. The diagnostic odds ratio: a single indicator of test performance. J Clin Epidemiol. 2003;56(11):1129–35.

    Article  PubMed  Google Scholar 

  69. Jaeschke R, Guyatt GH, Sackett DL. Users’ guides to the medical literature. III. How to use an article about a diagnostic test. B. What are the results and will they help me in caring for my patients ? The Evidence-Based Medicine Working Group. JAMA. 1994;271(9):703–7.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

One of the first authors (Guanrong Peng) would like to thank his wife Yujie Zeng for her help in the grammar expression. In addition, we would like to thank Professor Yirong Zeng and his team for their expertise and commitment.

Funding

The article received financial support from the High-Level Hospital Construction Project of the First Affiliated Hospital of Guangzhou University of Chinese Medicine (Grant no: 211010010722).

Author information

Authors and Affiliations

Authors

Contributions

GP, ZZ and YZ did conceptualization. GP, QL, ZG, and ML have done literature review and search. GP, QL, JC and WF performed data extraction and quality assessment. XS, XZ, and JL contributed to software and formal analysis. JZ, JL, and YZ validated the study. GP, QL, and ML prepared and edited the manuscript. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Zhangrong Zhong or Yirong Zeng.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Appendix: Search strategy

Appendix: Search strategy


Pubmed


#1"Prosthesis-Related Infections"[Mesh]


#2


(((((((((Prosthesis Related Infections[Title/Abstract]) OR (Prosthesis Related Infection[Title/Abstract])) OR (Infection, Prosthesis Related[Title/Abstract])) OR (Related Infection, Prosthesis[Title/Abstract])) OR (Related Infections, Prosthesis[Title/Abstract])) OR (Prosthesis-Related Infection[Title/Abstract])) OR (Infections, Prosthesis-Related[Title/Abstract])) OR (Periprosthetic Joint Infection[Title/Abstract])) OR (PJI[Title/Abstract])) OR (periprosthetic infection[Title/Abstract])


#3 #1 OR #2


#4: "Ultrasonics"[Mesh]


#5


((Ultrasonic[Title/Abstract]) OR (Sonication[Title/Abstract])) OR (Sonications[Title/Abstract])


#6 #4 OR #5


#7


"sensitiv*"[Title/Abstract] OR "sensitivity and specificity"[MeSH Terms] OR ("predictive"[Title/Abstract] AND "value*"[Title/Abstract]) OR ("predictive value of tests"[MeSH Terms] OR ("predictive"[All Fields] AND "value"[All Fields] AND "tests"[All Fields]) OR "predictive value of tests"[All Fields]) OR "accuracy*"[Title/Abstract]


#8 #3 AND #6 AND #7


Embase


#1 'prosthesis related' AND ('infections'/exp OR infections)


#2


'prosthesis related infections':ab,ti OR 'prosthesis related infection':ab,ti OR 'infection, prosthesis related':ab,ti OR 'related infection, prosthesis':ab,ti OR 'related infections, prosthesis':ab,ti OR 'prosthesis-related infection':ab,ti OR 'infections and infestations':ab,ti OR 'periprosthetic joint infection':ab,ti OR 'pji':ab,ti OR 'periprosthetic infection':ab,ti


#3: #1 OR #2


#4 'ultrasound'/exp


#5 'ultrasonic':ab,ti OR 'sonication':ab,ti OR 'sonications':ab,ti


#6: #4 OR #5


#7 'sensitiv':ab,ti OR 'sensitivity and specificity':ab,ti OR 'value':ab,ti OR 'predictive':ab,ti OR 'predictive value of tests':ab,ti OR 'accuracy':ab,ti


#8 #3 AND #6 AND #7


Cochrane


#1


MeSH descriptor: [Prosthesis-Related Infections] explode all trees


#2


(Prosthesis Related Infections):ti,ab,kw OR (Prosthesis Related Infection):ti,ab,kw OR (Infection, Prosthesis Related):ti,ab,kw OR (Related Infection, Prosthesis):ti,ab,kw OR (Related Infections, Prosthesis):ti,ab,kw OR (Prosthesis-Related Infection):ti,ab,kw OR (Infections, Prosthesis-Related):ti,ab,kw OR (Periprosthetic Joint Infection):ti,ab,kw OR (PJI):ti,ab,kw OR (periprosthetic infection):ti,ab,kw


#3: #1 OR #2


#4


MeSH descriptor: [Ultrasonics] explode all trees


#5


(Ultrasonic):ti,ab,kw OR (Sonication):ti,ab,kw OR (Sonications):ti,ab,kw


#6: #4 OR #5


#7 (sensitiv):ti,ab,kw OR (sensitivity and specificity):ti,ab,kw OR (value):ti,ab,kw OR (predictive):ti,ab,kw OR (predictive value of tests):ti,ab,kw OR (accuracy):ti,ab,kw


#8 #3 AND #6 AND #7


Web of Science


#1


TS = (Prosthesis-Related Infections or Prosthesis Related Infections or Prosthesis Related Infection or Infection, Prosthesis Related or Related Infection, Prosthesis or Related Infections, Prosthesis or Prosthesis-Related Infection or Infections, Prosthesis-Related or Periprosthetic Joint Infection or PJI or periprosthetic infection)


#2


TS = (Ultrasonics or Ultrasonic or Sonication or Sonications)


#3


TS = (sensitiv or sensitivity and specificity or value or predictive or predictive value of tests or accuracy)


#4 #1 AND #2 AND #3

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Peng, G., Liu, Q., Guan, Z. et al. Diagnostic accuracy of sonication fluid cultures from prosthetic components in periprosthetic joint infection: an updated diagnostic meta-analysis. J Orthop Surg Res 18, 175 (2023). https://doi.org/10.1186/s13018-023-03662-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-023-03662-3

Keywords