Abstract
The present study is intended to investigate and compare the hemodynamics in two different sizes of hemodialysis arteriovenous grafts for upper arm hemodialysis vascular access: 8-mm tapered to 6-mm at the arterial side and straight 6 mm. A computational simulation approach is presented for this study, which is validated against the available experimental and numerical pressure measurements in the literature. The imposed boundary conditions at the arterial inlet and venous outlet boundaries of the models are physiological velocity and pressure waveforms, respectively. Blood flow fields and distribution patterns of the hemodynamic indices including wall shear stress (WSS) as one of the major hemodynamic parameters of the cardiovascular system and spatial wall shear stress gradient (SWSSG) as an indicator of disturbed flow patterns and hence susceptible sites of lesion developments are analyzed and compared between the two grafts. The tapered 6- to 8-mm graft seemingly is associated with less disturbed flow patterns within the venous anastomosis (VA) and the vein downstream while benefiting from higher blood flow rates within. Also, it shows a definitive advantage in terms of WSS and SWSSG distribution patterns around the VA and throughout the vein downstream with significantly lower values, which reduce the risk of thrombosis formation and stenotic lesion developments. The only disadvantage encountered in using 6- to 8-mm tapered graft is higher values of hemodynamic parameters at the arterial junction attributable to its significantly higher mean blood flow rate within. The results clearly indicate that the tapered 6- to 8-mm graft entirely outperforms straight 6-mm graft hemodynamically as an upper arm hemodialysis vascular access graft and confirms clinical data in the literature, which suggests advantageous use of tapered 6- to 8-mm grafts in the creation of upper arm brachioaxillary hemodialysis vascular access grafts in selected groups of patients with expectably higher patency rates and lower complications.
Similar content being viewed by others
References
Akoh JA (2009) Prosthetic arteriovenous grafts for hemodialysis. J Vasc Access 10:137–147
Allon M (2007) Current management of vascular access. Clin J Am Soc Nephrol 2:786–800
Buchanan JR, Kleinstreuer C, Hyun S, Truskey GA (2003) Hemodynamics simulation and identification of susceptible sites of atherosclerotic lesion formation in a model abdominal aorta. J Biomech 36:1185–1196
DePaola N, Gimbrone MA Jr, Davies PF, Dewey CF Jr (1992) Vascular endothelium responds to fluid shear stress gradients. Arterioscler Thromb 12(11):1254–1257
Dewitz TS, McIntire LV, Martin RR, Sybers HD (1979) Enzyme release and morphological changes in leukocytes induced by mechanical trauma. Blood Cells 5(3):499–512
El Zahab Z, Divo E, Kassab A (2010) Minimisation of the wall shear stress gradients in bypass grafts anastomoses using meshless CFD and genetic algorithms optimization. Compu Methods Biomech Biomed Engin 13(1):35–47
Ene-Iordache B, Mosconi L, Remuzzi G, Remuzzi A (2001) Computational fluid dynamics of a vascular access case for hemodialysis. J Biomech Eng 123(3):284–292
Ene-Iordache B, Remuzzi A (2012) Disturbed flow in radial-cephalic arteriovenous fistulae for haemodialysis: low and oscillating shear stress locates the sites of stenosis. Nephrol Dial Transplant 27(1):358–368
Ethier CR, Prakash S, Steinman DA et al (2000) Steady flow separation in a 45 degree junction. J Fluid Mech 411:1–38
Fan Y, Xu Z, Jiang W et al (2008) An S-type bypass can improve the hemodynamics in the bypassed arteries and suppress intimal hyperplasia along the host artery floor. J Biomech 41:2498–2505
Fillinger MF, Reinitz ER, Schwartz RA, Resetarits DE, Paskanik AM, Bruch D, Bredenberg CE (1990) Graft geometry and venous intimal-medial hyperplasia in arteriovenous loop grafts. J Vasc Surg 11:556–566
Friedman MH, Bargeron CB, Duncan DD, Hutchins GM, Mark FF (1992) Effects of arterial compliance and non-Newtonian rheology on correlations between intimal thickness and wall shear. J Biomech Eng 114(3):317–320
Fry DL (1968) Acute vascular endothelial changes associated with increased blood velocity Gradients. Circ Res 22:165–197
Garcia-Pajares R, Polo JR, Flores A, Gonzalez-Tabares E, Solis JV (2003) Upper arm polytetrafluoroethylene grafts for dialysis access: analysis of two different graft sizes: 6 mm and 6–8 mm. Vasc Endovasc Surg 37:335–343
Haruguchi H, Teraoka S (2003) Intimal hyperplasia and hemodynamic factors in arterial bypass and arteriovenous grafts: a review. J Artif Organs 6:227–235
He X, Ku DN (1996) Pulsatile flow in the human left coronary artery bifurcation: average conditions. J Biomech 118(1):74–82
Himburg HA, Grzybowski DM, Hazel AL et al (2004) Spatial comparison between wall shear stress measures and porcine arterial endothelial permeability. Am J Physiol Heart Circ Physiol 286(5):H1916–H1922
Hofstra L, Bergmans DC, Hoeks AP, Kitslaar PJ, Leunissen KM, Tordoir JH (1994) Mismatch in elastic properties around anastomoses of interposition grafts for hemodialysis access. J Am Soc Nephrol 5(5):1243–1250
Hofstra L, Bergmans DC, Leunissen KM et al (1995) Anastomotic intimal hyperplasia in prosthetic arteriovenous fistulas for hemodialysis is associated with initial high flow velocity and not with mismatch in elastic properties. J Am Soc Nephrol 6(6):1625–1633
Hofstra L, Bergmans DC, Leunissen KM et al (1996) Prosthetic arteriovenous fistulas and venous anastomotic stenosis: influence of a high flow velocity on the development of intimal hyperplasia. Blood Purif 14(5):345–349
Holme PA, Orvim U, Hamers MJ, Solum NO, Brosstad FR, Barstad RM, Sakariassen KS (1997) Shear-induced platelet activation and platelet microparticle formation at blood flow conditions as in arteries with a severe stenosis. Arterioscler Thromb Vasc Biol 17(4):646–653
Huber TS, Carter JW, Carter RL, Seeger JM (2003) Patency of autogenous and polytetrafluoroethylene upper extremity arteriovenous hemodialysis accesses: a systematic review. J Vasc Surg 38:1005–1011
Keynton RS, Evancho MM, Sims RL, Rodway NV, Gobin A, Rittgers SE (2001) Intimal hyperplasia and wall shear in arterial bypass graft distal anastomoses: an in vivo model study. J Biomech Eng 123(5):464–473
Keynton RS, Rittgers SE, Shu MC (1991) The effect of angle and flow rate upon hemodynamics in distal vascular graft anastomoses: an in vitro model study. J Biomech Eng 113(4):458–463
Kharboutly Z, Deplano V, Bertrand E, Legallais C (2010) Numerical and experimental study of blood flow through a patient-specific arteriovenous fistula used for hemodialysis. Med Eng Phys 32(2):111–118
Kharboutly Z, Fenech M, Treutenaere JM, Claude I, Legallais C (2007) Investigations into the relationship between hemodynamics and vascular alterations in an established arteriovenous fistula. Med Eng Phys 29(9):999–1007
Kleinstreuer C, Hyun S, Buchanan JR, Longest PW, Archie JP, Truskey GA (2001) Hemodynamic parameters and early intimal thickening in branching blood vessels. Crit Rev Biomed Eng 29(1):1–64
Kleinstreuer C, Lei M, Archie JP Jr (1996) Flow input waveform effects on the temporal and spatial wall shear stress gradients in a femoral graft-artery connector. J Biomech 118(4):506–510
Knox RC, Berman SS, Hughes JD, Gentile AT, Mills JL (2002) Distal revascularization-interval ligation: a durable and effective treatment for ischemic steal syndrome after hemodialysis access. J Vasc Surg 36:250–256
Konner K et al (2003) The arteriovenous fistula. J Am Soc Nephrol 14:1669–1680
Krueger U, Zanow J, Scholz H (2002) Computational fluid dynamics and vascular access. Artif Organs 26(7):571–575
Ku DN, Giddens DP, Zarins CK, Glagov S (1985) Pulsatile flow and atherosclerosis in the human carotid bifurcation. Positive correlation between plaque location and low oscillating shear stress. Arteriosclerosis 5(3):293–302
Leask RL, Butany J, Johnston KW, Ethier CR, Ojha M (2005) Human saphenous vein coronary artery bypass graft morphology, geometry and hemodynamics. Ann Biomed Eng 33(3):301–309
Lee SW, Antiga L, Steinman DA (2009) Correlations among indicators of disturbed flow at the normal carotid bifurcation. J Biomech Eng 131(6):061013. doi:10.1115/1.3127252
Lee SW, Smith DS, Loth F, Fischer PF, Bassiouny HS (2007) Importance of flow division on transition to turbulence within an arteriovenous graft. J Biomech 40(5):981–992
Lei M, Archie JP, Kleinstreuer C (1997) Computational design of a bypass graft that minimizes wall shear stress gradients in the region of the distal anastomosis. J Vasc Surg 25(4):637–646
Lei M, Kleinstreuer C, Archie JP (1996) Geometric design improvements for femoral graft-artery junctions mitigating restenosis. J Biomech 29(12):1605–1614
Leon C, Asif A (2007) Arteriovenous access and hand pain: the distal hypoperfusion ischemic syndrome. Clin J Am Soc Nephrol 2(1):175–183
Li L, Terry CM, Shiu YT, Cheung AK (2008) Neointimal hyperplasia associated with synthetic hemodialysis Grafts. Kidney Int 74(10):1247–1261
Longest PW, Kleinstreuer C (2000) Computational haemodynamics analysis and comparison study of arterio-venous grafts. J Med Eng Technol 24(3):102–110
Loth F, Fischer PF, Arsalan N, Bertram CD, Lee SE, Royston TJ, Shaalan WE, Bassiouny HS (2003) Transitional flow at the venous anastomosis of an arteriovenous graft: potential activation of the ERK1/2 mechanotransduction pathway. J Biomech 125:49–61
McVeigh GE, Bratteli CW, Morgan DJ et al (1999) Age-related abnormalities in arterial compliance identified by pressure pulse contour analysis: aging and arterial compliance. Hypertension 33(6):1392–1398
Meichelboeck W (2011) End stage renal disease (ESRD) epidemiology—Where are we going? J Vasc Access 12(2):137–181
Meyerson SL, Skelly CL, Curi MA, Shakur UM, Vosicky JE, Glagov S, Schwartz LB (2001) The effects of extremely low shear stress on cellular proliferation and neointimal thickening in the failing bypass graft. J Vasc Surg 34(1):90–97
Nagel T, Resnick N, Deway CF Jr et al (1999) Vascular endothelial cells respond to spatial gradients in fluid shear stress by enhanced activation of transcription factors. Arterioscler Thromb Vasc Biol 19:1825–1834
NKF, KDOQI (2006) Clinical practice guidelines and clinical practice recommendations for vascular access; 2006 updates (suppl 1). Am J Kidney Dis 48:S227–S364
O’Callaghan S, Walsh M, McGloughlin T (2006) Numerical modelling of Newtonian and non-Newtonian representation of blood in a distal end-to-side vascular bypass graft anastomosis. Med Eng Phys 28:70–74
Ojha M (1993) Spatial and temporal variations of wall shear stress within an end-to-side arterial anastomosis model. J Biomech 26(12):1377–1388
Ojha M (1994) Wall shear stress temporal gradient and anastomotic intimal hyperplasia. Circ Res 74:1227–1231
Polo JR, Ligero JM, Diaz-Cartelle J, Garcia-Pajares R, Cervera T, Reparaz L (2004) Randomized comparison of 6-mm straight grafts versus 6- to 8-mm tapered grafts for brachial-axillary dialysis access. J Vasc Surg 40:319–324
Rouleau L, Farcas M, Tardif JC, Mongrain R, Leask RL (2010) Endothelial Cell Morphologic Response to Asymmetric Stenosis Hemodynamics: effects of Spatial Wall Shear Stress Gradients. J Biomech Eng 132(8):081013. doi:10.1115/1.4001891
Roy-Chaudhury P, Kelly BS, Melhem M et al (2005) Vascular access in hemodialysis: issues, management, and emerging concepts. Cardiol Clin 23:249–273
Salam TA, Lumsden AB, Suggs WD, Ku DN (1996) Low shear stress promotes intimal hyperplasia thickening. J Vasc Invest 2:12–22
Shaik E, Hoffmann KA, Dietiker JF (2008) Numerical simulations of pulsatile non-Newtonian flow in an end-to-side anastomosis model. Simul Model Pract Theory 16:1123–1135
Sivanesan S, How TV, Black RA, Bakran A (1999) Flow patterns in the radiocephalic arteriovenous fistula: an in vitro study. J Biomech 32(9):915–925
Slack SM, Jennings LK, Turitto VT (1994) Platelet size distribution measurements as indicators of shear stress-induced platelet aggregation. Ann Biomed Eng 22(6):653–659
Steinman DA, Vinh B, Ethier CR et al (1993) A numerical simulation of flow in a two-dimensional end-to-side anastomosis model. J Biomech Eng 115(1):112–118
Sutera SP, Mehrjardi MH (1975) Deformation and fragmentation of human red blood cells in turbulent shear flow. Biophys J 15(1):1–10
Tardy Y, Resnick N, Nagel T et al (1997) Shear stress gradients remodel endothelial monolayers in vitro via a cell proliferation-migration-loss cycle. Arterioscler Thromb Vasc Biol 17(11):3102–3106
Van Canneyt K, De Santis G, Eloot S, Segers P, Verdonck P (2011) Swirlgraft versus conventional straight graft as vascular access: a full CFD-analysis. Conference proceedings CMBE11 383–386
Van Canneyt K, Morbiducci U, Eloot S, De Santis G, Segers P, Verdonck P (2013) A computational exploration of helical arterio-venous graft designs. J Biomech 46(2):345–353
Van Canneyt K, Pourchez T, Eloot S et al (2010) Hemodynamic impact of anastomosis size and angle in side-to-end arteriovenous fistulae: a computer analysis. J Vasc Access 11(1):52–58
Van Tricht I, De Wachter D, Tordoir J, Verdonck P (2004) Hemodynamics in a compliant hydraulic in vitro model of straight versus tapered PTFE arteriovenous graft. J Surg Res 116:297–304
Van Tricht I, De Wachter D, Tordoir J, Verdonck P (2006) Comparison of the hemodynamics in 6 mm and 4–7 mm hemodialysis grafts by means of CFD. J Biomech 39:226–236
Van Tricht I, De Wachter D, Tordoir J, Verdonck P (2005) Hemodynamics and complications encountered with arteriovenous fistulas and grafts as vascular access for hemodialysis: a review. Ann Biomed Eng 33(9):1142–1157
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Sarmast, M., Niroomand-Oscuii, H., Ghalichi, F. et al. Evaluation of the hemodynamics in straight 6-mm and tapered 6- to 8-mm grafts as upper arm hemodialysis vascular access. Med Biol Eng Comput 52, 797–811 (2014). https://doi.org/10.1007/s11517-014-1181-7
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11517-014-1181-7