ARTICLE IN PRESS
Atmospheric Environment 37 (2003) 4781–4793
Field evaluation of a personal cascade impactor
sampler (PCIS)
Manisha Singh, Chandan Misra, Constantinos Sioutas*
Department of Civil and Environmental Engineering, University of Southern California, 3620 South Vermont Avenue,
Los Angeles, CA 90089, USA
Received 16 May 2003; accepted 9 August 2003
Abstract
This paper presents the field evaluation of a personal cascade impactor sampler (PCIS). PCIS is a miniaturized
cascade impactor, consisting of four impaction stages, followed by an after-filter. Particles are separated in the
following aerodynamic particle diameter ranges: o0.25, 0.25–0.5, 0.5–1.0, 1.0–2.5 and 2.5–10 mm. The PCIS operates at
a flow rate of 9 liters per minute (l/min) using a very high efficiency, battery-operated light pump at a pressure drop of
11 in H2O (2.7 kPa). For field evaluation, the PCIS was collocated with other samplers including the micro-orifice
uniform deposit impactor (MOUDI), scanning mobility particle sizer (SMPS) and aerodynamic particle sizer (APS) in
Los Angeles and Claremont, CA. PCIS and MOUDI agree very well for coarse particulate matter (PM) (PM2.510)
mass concentrations. The fine PM (PM2.5) mass as measured by PCIS is in excellent agreement with SMPS–APS
measurement (B1.02 times) and slightly higher (B1.1 times) than the MOUDI measurements. Time-integrated (size
fractionated) PM2.5 mass, inorganic ions (nitrate and sulfate), elemental carbon (EC) and organic carbon (OC)
concentrations obtained with PCIS and MOUDI were found to be in very good agreement with few differences in the
o0.25 mm size fraction, especially for OC and nitrate measurements. Near-continuous and size fractionated PM2.5
nitrate and total carbon measurements by PCIS and MOUDI using the ADI and Sunset labs monitors are in close
agreement for all size fractions, indicating that any differences between MOUDI and PCIS measurements for timeintegrated data might be due to artifacts associated with long-term sampling and not to differences in individual cutpoints. The performance of the PCIS was also evaluated in wind tunnel tests at wind speeds up to 8 km/h. These tests
showed that particle sampling efficiency and separation characteristics of the PCIS are unaffected by the wind speeds
for particles up to 10 mm in aerodynamic diameter.
r 2003 Elsevier Ltd. All rights reserved.
Keywords: Personal sampler; Particulate matter (PM); Battery operated personal pump; MOUDI; SMPS; Continuous nitrate and
carbon monitors
1. Introduction
Ambient particulate matter (PM) is an air pollutant
that has become a major public health concern. A great
number of clinical and epidemiological studies have
indicated cause and effect associations between respira*Corresponding author. Tel.: +1-213-740-6134; fax: +1213-744-1426.
E-mail address: sioutas@usc.edu (C. Sioutas).
tory effects and exposures to ambient PM (Dockery
et al., 1993a; Gordian et al., 1996; Schwartz and
Dockery, 1992). Most of the health effects have been
associated with the ambient mass concentration of
particles smaller than 10 mm (PM10) or, more recently,
2.5 mm in aerodynamic diameter (PM2.5). Results from
other studies suggest that particle components, such as
+
sulfate (SO2
4 ) and aerosol strong acidity (H ), also may
be associated with increased mortality and other adverse
health impacts (Ayres et al., 1989; Bates and Sizto, 1989;
1352-2310/$ - see front matter r 2003 Elsevier Ltd. All rights reserved.
doi:10.1016/j.atmosenv.2003.08.013
ARTICLE IN PRESS
4782
M. Singh et al. / Atmospheric Environment 37 (2003) 4781–4793
Bates et al., 1990; Dockery et al., 1993b; Raizenne et al.,
1989, 1993; Thurston et al., 1992, 1993). More recent
studies in Southern California (Li et al., 2003) demonstrated increased biological potency of the spatially
inhomogeneous urban ultrafine particles, which was
related to their high content of redox cycling organic
chemicals and their ability to damage mitochondria.
In light of these findings, the sufficiency of the
National Ambient Air Quality Standard (NAAQS) for
PM, in terms of both its regulated level and species has
become a matter of concern. In order to address these
concerns, knowledge of relationship among the various
particle measures and their personal exposure profiles
becomes essential. Both indoor and outdoor concentrations have been found to be poor estimators of personal
exposures to PM10 and its components, as neither
indoor nor outdoor concentrations suffice to account
for the observed interpersonal variability in their
exposures. Daytime personal PM10 exposures were
found to be approximately 50% higher than corresponding indoor and outdoor levels (Thomas et al., 1993),
+
while personal SO2
exposures were found to
4 and H
be higher than indoor, but lower than outdoor
concentrations (Suh et al., 1994).
Understanding of individual exposures to PM can be
significantly improved by the use of personal monitors,
as these samplers incorporate the effects of factors like
indoor pollutant sources and human time-activity
patterns. Coupled with recent advances in the area of
molecular source tracers (Schauer et al., 2002), these
samplers can enhance the quality of studies linking
personal exposures to specific outdoor or indoor sources
and help determine degrees to which they influence
personal exposure.
A personal sampler for PM that allows separation of
airborne particles in several size ranges has been
developed and evaluated in the laboratory (Misra et al.,
2002). The personal cascade impactor sampler (PCIS) is
a miniaturized cascade impactor, consisting of four
impaction stages, followed by an after-filter. Particles
are separated in the following aerodynamic particle
diameter ranges: o0.25, 0.25–0.5, 0.5–1.0, 1.0–2.5 and
2.5–10 mm. It operates at a high flow rate (9 l/min) by
personal sampling standards that make chemical analyses of the size-fractionated particles possible within a
period of 24-h or less. The only other personal cascade
impactors reported in the aerosol literature were
developed by Rubow et al. (1987) and Demokritou
et al. (2002). The Rubow et al. impactor operates at a
flow of 2 l/min and its smallest cutpoint size is 0.5 mm.
Demokritou et al. (2002) have developed a personal
cascade impactor that also operates at a flow rate of 5 l/
min with smallest cutpoint size of 0.5 mm. The development of the PCIS therefore constitutes a major
improvement over the prior state-of-the-art in the field
of personal monitoring because of its much higher flow
rate and its ability to classify particles as small as
0.25 mm in aerodynamic diameter.
In this paper, we present results from a field
evaluation of the PCIS conducted in two locations of
Southern California, at Claremont and Los Angeles,
respectively. The field evaluation was performed via a
comparison to collocated samplers including microorifice uniform deposit impactor (MOUDI, Model 110,
MSP Corp, Minneapolis, MN), scanning mobility
particle sizer (SMPS, TSI Model 3936) and aerodynamic
particle sizer (APS, TSI Model 3320). Correlations
between the instruments will serve to validate the
sampler’s field performance. The performance of PCIS
is also evaluated in the wind tunnel.
2. Methods
2.1. Description of the PCIS
Design and a comprehensive laboratory evaluation of
the PCIS are described in detail by Misra et al. (2002).
Thus, only a brief description is presented in this paper.
The PCIS is a miniaturized cascade impactor, consisting
of four impaction stages and an after-filter that allows
the separation and collection of airborne particles in five
size ranges (Figs. 1a–d). The sampling flow rate is 9 l/
min and the measured total pressure drop across the
sampler is 11 in of H2O (2.7 kPa). The PCIS is used in
combination with the Leland Legacy Sample Pump
(SKC Inc., Cat. No. 100–3000). The total weight of the
sampler is approximately 150 g, thus easy to be used by
subjects such as children or elderly. The pump weighs
about 450 g (including the battery) and is placed inside a
small pouch with snap latch and foam inserts to protect
the pump during transport. Particles in the size range of
0.25–10 mm are accelerated in rectangular-shaped nozzles and collected on commercially available 25 mm
substrates made of quartz (Pallflex Corp., Putnam, CT),
aluminum or Zefluor (0.5 mm pore, Gelman Science,
Ann Arbor, MI). Particles smaller than 0.25 mm are
collected on a 37 mm after-filter made of either quartz or
Teflon (PTEF, 2 mm pore, Gelman Science, Ann Arbor,
MI). For real-life usage, the PCIS is designed to clip
onto a subject’s collar in the breathing zone and the
pump clips on the subject’s belt.
2.2. Field experiments
The performance of the PCIS was evaluated in a field
study, which was conducted from March to November
2002. Outdoor sampling was initiated at Claremont,
CA, the location at the time of the Southern California
Supersite mobile laboratory facility. This site is located
approximately 45 km east of downtown Los Angeles.
Sampling was conducted during 24-h periods once a
ARTICLE IN PRESS
M. Singh et al. / Atmospheric Environment 37 (2003) 4781–4793
(a)
(b)
(c)
(d)
4783
Fig. 1. Pictures of the PCIS.
week. The PCIS was collocated with the micro-orifice
uniform deposit impactor (MOUDIt, MSP Corp.
Minneapolis, MN) and the scanning mobility particle
sizer (SMPSt, TSI Model 3936)—aerodynamic particle
sizer (APSt, TSI Model 3320) tandem inside the Particle
Instrumentation Unit, a mobile laboratory trailer that
was developed through funds provided by the US
Environmental Protection Agency and is being currently
used in large-scale field studies that are part of the
Southern California Particle Center and Supersite
activities. As of July 2002, sampling was continued
outdoors in the environment of University of Southern
California, the next sampling site of the Southern
California Supersite. The school is located about 3 km
south of downtown Los Angeles and represents a typical
urban area in which the aerosol is primarily emitted by
vehicles.
Size-segregated PM2.5 mass, inorganic ions (nitrate
and sulfate), elemental carbon (EC) and organic carbon
(OC) concentrations measured by PCIS and collocated
MOUDI were compared. MOUDI operates at a higher
flow rate of 30 l/min compared to PCIS. Concentrations
obtained from the MOUDI were grouped in the
following size ranges: o0.18, 0.18–0.32, 0.32–0.5, 0.5–
1.0, 1.0–2.5 and 2.5–10 mm, whilst for PCIS, the ranges
were: o0.25, 0.25–0.5, 0.5–1.0, 1.0–2.5 and 2.5–10 mm.
The MOUDI does not have a 0.25 mm cutpoint stage. In
order to therefore make the MOUDI cutoff size ranges
comparable to those of the PCIS, 50% of the mass
measured in the 0.18–0.32 mm size range was added to
that in the 0–0.18 mm range and 50% to that in the 0.32–
0.50 mm. It should be noted that this conversion was
only an effort to bring the MOUDI and PCIS aerosol
size ranges closer to each other; it assumes that the
aerosol size distribution in that range is log-normal, an
assumption that may or may not be valid, and is thus
likely to introduce some errors in the comparison
experiments between MOUDI and PCIS. For mass
and inorganic ions concentration measurements, the
MOUDI substrates were 47 mm Teflon filters, followed
by a 37-mm Teflon after-filter, whereas the PCIS
substrates were 25 mm Zefluor substrates followed by
a 37 mm Teflon after-filter as the last stage. For EC and
OC analysis, 25 and 35 mm prebaked aluminum foil
substrates were used for the PCIS and MOUDI stages,
respectively, followed by 37 mm prebaked quartz filters.
For mass concentration measurements, the substrates
were first weighed before and after each field tests using
a Mettler 5 Micro-balance (MT 5, Mettler-Toledo Inc.,
Highstown, NJ), under controlled relative humidity (e.g.
40–45%) and temperature (e.g., 22–24 C) conditions. At
the end of each experiment, filters were stored in the
ARTICLE IN PRESS
4784
M. Singh et al. / Atmospheric Environment 37 (2003) 4781–4793
control humidity and temperature room for 24 h prior to
weighing in order to ensure removal of particle-bound
water. Particle-bound sulfate and nitrate concentrations
were determined by means of ion chromatography (IC)
similar to that employed by Harrison and Peak (1996).
To determine the EC and OC content of PM, the
substrates were analyzed by means of a thermo-analysis
technique (Fung, 1990).
The semi-continuous SMPS–APS particle mass concentrations were determined by integrating the cumulative number count for the respective size fractions, and
converting to mass by assuming particles to be perfect
spheres and an average particle density of 1.6 g/cm3.
Prior to this conversion, for the SMPS data each particle
size interval was converted from mobility equivalent
diameter to aerodynamic diameter using the following
equation (Peters et al., 1993):
sffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
pffiffiffiffiffi
Cme rp
Ca da ¼
dme ;
ð1Þ
wr0
where da is aerodynamic diameter, dme is the mobility
equivalent diameter, Ca is the slip correction factor for
the aerodynamic diameter, Cme is the slip correction
factor for the mobility equivalent diameter, w is the
dynamic shape factor, rp is the density of the particle
(1.6 g/cm3), and r0 is the unit density (1 g/cm3). When
performing this conversion, the mobility equivalent
diameter was assumed to be equal to the equivalent
volume diameter, while particles were assumed to be
perfect spheres (dynamic shape factor, w ¼ 1). The
hourly SMPS-APS PM concentrations were integrated
into 24 h intervals and compared to those measured
concurrently by the PCIS.
In addition to the IC analysis, size segregated PM2.5
nitrate concentrations from PCIS and MOUDI were
measured using a continuous nitrate monitor from
aerosol dynamics Inc. (ADI) described in more detail
by Stolzenburg et al. (2002). Near continuous (10-min)
data were obtained by connecting the ADI monitor
(sampling at a flow of 1 l/min) to ports placed downstream of each MOUDI and PCIS stages. The concentrations downstream of a MOUDI stage were measured
immediately after its corresponding PCIS stage for
10 min. For any given particle range, this procedure
was repeated 2–3 times in order to ensure that there are
no significant changes in the concentrations of the
ambient PM2.5 aerosol during the course of these
experiments.
Size segregated total carbon concentrations in PM2.5
by MOUDI and PCIS were also measured using a
continuous carbon monitor (Sunset Laboratory SemiContinuous OC-EC Carbon Aerosol Analyzer, Sunset
Labs, Forest Grove, OR) operating at a flow of 9 l/min.
Given that the carbon monitor flow rate is equal to that
of the PCIS, the entire flow after a given PCIS stage was
drawn into the carbon analyzer, whereas MOUDI
concentrations in each stage were measured by diverting
9 of the 30 l/min downstream of each stage, as described
above for nitrate measurements. The experimental
procedure was identical to that described in the previous
paragraph regarding nitrate measurements, with the
exception that direct-reading measurements in each
PCIS and MOUDI stage were alternately taken every
30 min (as opposed to 10 min in the case of nitrate tests).
In both series of experiments, the PCIS was used with
Zefluor impaction substrates and the MOUDII with
Teflon substrates. A total of 8 field experiments were
conducted for nitrate and 10 for carbon measurements.
2.3. Wind tunnel tests
The ability of the PCIS to sample isokinetically
particles at moderate wind speeds was evaluated in the
wind tunnel facility of School of Public Health, UCLA.
The wind tunnel is described in detail in Hinds and Kuo
(1995) and by Kennedy et al. (2001). The tunnel has a
1.6 1.6 m2 cross-section and was operated at two wind
speeds (3 and 8 km/h). A plywood baffle was placed
about 0.5 m upstream of the aerosol generation system
to promote mixing. A vibrating orifice aerosol generator
(VOAG) (Model 3450, TSI Inc., St. Paul, MN) was used
to generate monodisperse particles. The droplet size
primarily depends on the orifice size for a given solution
feed rate and the frequency. For these experiments, a
20 mm orifice was used for generating Typical VOAG
operating parameters were 0.150 ml/min of feed rate at
65–70 kHz. Uranine tagged oleic acid was used as a nonvolatile solute for generating particles with acetone as
the solvent. Approximately, 2 g of uranine dye was
dissolved in 50 ml of methanol to prepare the tracer
solution and was left overnight to dissolve the dye in the
solution with concomitant settling of the undissolved
uranine dye. A 20–40% of this solution was then added
to the oleic acid–acetone solution to generate monodisperse aerosols in the size of 2–10 mm. The monodispersity of the generated aerosols was confirmed by
observing the generated particles under a microscope,
which also corroborated the size of the particles.
The vibrating aerosol orifice was itself mounted on a
shaft, which moved both up and down and sideways to
promote uniform injection. Three isokinetic samplers
were placed around the PCIS. The PCIS was placed such
that it was equidistant from the three-isokinetic samplers. Two of these samplers were lateral to the PCIS
while the third one was above the PCIS. The positioning
of the isokinetic samplers corresponded to uniformity in
concentration around the PCIS. An earlier work by
Hinds and Kuo (1995) describes the positioning of
isokinetic samplers in detail.
Considering that isokinetic sampling is an issue of
concern mainly for super-micrometer particles, tests
ARTICLE IN PRESS
M. Singh et al. / Atmospheric Environment 37 (2003) 4781–4793
4785
borate in 1000 ml of water (solution B). Diluting 50 ml of
solution A and 59 ml of solution B to 200 ml using
distilled water yielded the buffer solution.
The Millipore filters were then extracted in glass vials
using the buffer solution. Most of the extractions were
done using 5–10 ml of buffer solution. Standard uranine
dye solutions of 0.005, 0.01, 0.025, 0.05, 0.075 and
0.1 ppm were used to plot the calibration curve.
For each wind speed, comparison between the mass
concentrations obtained by means of the three isokinetic
samplers and the PCIS 2.5 mm stage was performed. For
each particle size, the averaged value of the mass
concentration for the three isokinetic samplers was
used. Finally, the particle penetration through the PCIS
was plotted against the particle diameter for each wind
speed.
were only conducted for the particle range of 2–10 mm.
The PCIS was thus used in a modified configuration,
whereby the all impaction stages were moved except of
the 2.5 mm cut off stage, which was followed by the PCIS
after-filter. The sampling characteristics of the 2.5 mm
cut PCIS stage were determined by comparing the
coarse (2.5–10 mm) particle mass concentration obtained
by the PCIS 2.5 mm stage to that measured by the
isokinetic samplers. The isokinetic samplers and the
PCIS 2.5 mm stage were positioned at the same distance
from the sample injection point (same axial plane). The
isokinetic samplers were constructed from 2.5 cm in-line
stainless-steel filter holders (P/N 1209, Gelman Sciences
Inc., Ann Arbor, MI) fitted with 8.5 mm ID brass probes
that extended 32 mm from the face of the filter holder
and sampled at a flow rate of 10 l/min for wind speeds of
3 and 8 km/h. Millipore membrane filters (SMWP
02500, Millipore, Bedford, MA) were used to collect
uranine tagged oleic acid particles in the isokinetic
samplers, whereas a 25 mm Zefluor filter was used as an
impaction substrate for the 2.5 mm stage of the PCIS.
Each of the experiments was characterized by particle
size and wind speed and lasted for about 10–15 min,
which was sufficient to obtain detectable mass on the
filters.
Detection of the deposited uranine tagged oleic acid
particles on the Millipore filters was performed using a
fluorescence detector (Model FD-500, Programmable
Fluorescence Detector, GTI, Concord, MA). Prior to
their fluorescence detection, the extraction of the
uranine was done using a buffer solution. The buffer
solution was prepared by dissolving 12.4 g of boric acid
in 1000 ml of water (solution A) and 19.05 g of sodium
3. Results and discussion
3.1. Coarse particle (PM102.5) mass concentrations
Coarse particle concentrations measured by PCIS
were compared to concurrent measurements by MOUDI. Inter-comparisons between PCIS and MOUDI
indicate an overall excellent agreement with an average
PM102.5 PCIS to MOUDI ratio of 1.007 (70.07). The
coarse particle mass concentrations obtained with PCIS
and those obtained with MOUDI are plotted in Fig. 2
along with the linear regression lines and the regression
coefficients. As evident from the figure, the PCIS
concentrations are also highly correlated with MOUDI
(R2 ¼ 0:95).
25
y = 0.96x + 0.41
R2 = 0.95
MOUDI (µg/m3)
20
15
10
5
0
0
5
10
15
20
25
PCIS (µg/m3)
Fig. 2. Coarse PM mass concentrations obtained with PCIS and MOUDI.
ARTICLE IN PRESS
M. Singh et al. / Atmospheric Environment 37 (2003) 4781–4793
4786
3.2. Fine particle (PM2.5) mass concentrations
between these monitors, with an average PCIS to
SMPS–APS PM2.5 ratio of 1.02 (70.34). However,
MOUDI PM2.5 concentrations are slightly lower than
those measured by the PCIS, with the average PCIS to
MOUDI PM2.5 concentration ratio being 1.11 (70.05).
The fine particle mass concentrations obtained with
PCIS and those obtained with MOUDI and SMPS-APS
are plotted in Figs. 3a and b, respectively, along with the
The 24-h averaged fine particle (PM2.5) mass concentrations measured by the PCIS were compared with
those obtained with MOUDI and the semi-continuous
SMPS–APS samplers. A summary of these comparisons
is given in Table 1. Inter-comparisons between PCIS and
SMPS–APS indicate an overall excellent agreement
Table 1
Summary of comparison between the PCIS, SMPS–APS and MOUDI for PM2.5 species concentrations obtained from time-integrated
measurements
Instruments Compared
Species
Mean of the ratios7S.D.
Mean difference (mg/m3)7S.D.
(mean relative difference)
p-value
PCIS:
PCIS:
PCIS:
PCIS:
PCIS:
PCIS:
Mass
Mass
Sulfate
Nitrate
EC
OC
1.0270.34
1.1170.05
1.1170.10
1.2270.21
1.1970.37
1.94a70.52
0.3078.43
2.1570.93
0.1970.22
0.6470.63
0.1870.22
5.272.35
0.96
0.82
0.86
0.93
0.39
o0.01
a
SMPS–APS
MOUDI
MOUDI
MOUDI
MOUDI
MOUDI
(1.19%)
(9.28%)
(8.04%)
(6.53%)
(20.28%)
(46.94%)
Statistically significant difference in values at p ¼ 0:05 level.
50
y = 0.98x - 2.01
R2 = 0.90
MOUDI(µg/m3)
40
30
20
10
0
0
5
10
15
20
25
30
35
40
45
50
PCIS (µg/m3)
(a)
SMPS-APS (µg/m3)
70
y = 0.78x + 4.91
R2 = 0.77
60
50
40
30
20
10
0
0
(b)
10
20
30
40
50
60
70
3)
PCIS (µg/m
Fig. 3. (a) PM2.5 mass concentrations obtained with PCIS and MOUDI and (b) PM2.5 mass concentrations obtained with PCIS and
SMPS–APS.
ARTICLE IN PRESS
M. Singh et al. / Atmospheric Environment 37 (2003) 4781–4793
4787
10
Concentrations (µg/m3)
PCIS
MOUDI
8
6
4
2
0
2.5-1.0
1.0-0.5
0.5-0.25
< 0.25
Aerodynamic Diameter (µm)
(a)
10
Concentration (µg/m3)
PCIS
SMPS-APS
8
6
4
2
0
(b)
2.5-1.0
1.0-0.5
0.5-0.25
< 0.25
Aerodynamic Diameter (µm)
Fig. 4. (a) Comparison of size-segregated mass concentrations obtained with PCIS and MOUDI and (b) comparison of size-segregated
mass concentrations obtained with PCIS and SMPS–APS.
linear regression lines and the regression coefficients. As
evident from these figures, the PCIS concentrations are
highly correlated with both MOUDI and SMPS–APS
PM2.5 data, with the R2 for PCIS vs. MOUDI and PCIS
vs. SMPS–APS being 0.90 and 0.77, respectively.
3.3. Size-fractionated PM2.5 mass concentrations
Fig. 4a depicts size-fractionated average PM2.5 mass
concentrations obtained with PCIS and MOUDI (error
bars represent the standard deviation of field measurements) in the respective size fractions, based on a total of
13 experiments. Good overall agreement between the
PCIS and MOUDI was obtained for particles in the
aerodynamic diameter ranges of 1–2.5, 0.5–1.0 and 0.25–
0.5 mm, with the average PCIS to MOUDI concentration ratios being 1.21 (70.15), 0.84 (70.12) and 1.09
(70.18), respectively. However, for the o0.25 mm range,
MOUDI slightly underestimates the mass concentrations; the average PCIS to MOUDI ratio is 1.28
(70.12). This difference may be attributed to the high
flow rate of MOUDI (30 l/min) and consequently a
relatively high pressure drop across its 0.18 mm stage as
well as the after-filter, which would enhance the
volatilization of labile species, resulting in the underestimation of overall mass of particles o0.25 mm in size.
Paired t-tests between PCIS and MOUDI concentrations for these four size ranges indicates that these
concentrations are not statistically significantly different
at the p ¼ 0:05 level (including the last stage).
Fig. 4b shows a comparison of the size fractionated
average PM2.5 mass concentrations obtained with the
PCIS and SMPS–APS samplers, along with standard
deviations, in the respective size fractions. A total of 14
24-h sampling experiments were conducted for these
comparisons. Very good overall agreement was obtained
between the PCIS and SMPS–APS concentrations for
the particle sizes 2.5–1.0, 0.5–1.0 and 0.25–0.5 mm with
average PCIS to SMPS–APS mass ratios of 0.96
(70.14), 1.03 (70.15) and 1.16 (70.21). For particles
ARTICLE IN PRESS
M. Singh et al. / Atmospheric Environment 37 (2003) 4781–4793
4788
Concentations (µg/m3)
1.50
PCIS
MOUDI
1.00
0.50
0.00
2.5-1.0
(a)
1.0-0.5
0.5-0.25
Aerodynamic Diameter (µm)
<0.25
Concentations (µg/m3)
5
PCIS
MOUDI
4
3
2
*
1
0
2.5-1.0
(b)
1.0-0.5
0.5-0.25
<0.25
Aerodynamic Diameter (µm)
Fig. 5. (a) Comparison of size-segregated sulfate concentrations obtained with PCIS and MOUDI and (b) comparison of sizesegregated nitrate concentrations obtained with PCIS and MOUDI. The asterisk denotes statistically different concentrations at the
p ¼ 0:05 level.
o0.25 mm in size, the PCIS overestimates the mass
concentrations as compared to SMPS by a factor of 1.31
(70.24). Although not statistically significant at the p ¼
0:05 level, this difference may be due to the different size
classification principles employed by the PCIS and
SMPS, with the former measuring aerodynamic and
the latter measuring mobility particle diameters. As
indicated by previous studies (McMurry et al., 2002)
particles in urban areas, originating from vehicular
emissions, contain a high fraction of fractal-like
agglomerates. These particles are relatively hollow and
would be classified in the ultrafine PM mode aerodynamically. However, due to their large surface area,
the SMPS would place a substantial portion of these
particles in the accumulation mode due to their
increased mobility diameter. Similar observations between the mass concentrations measured by impactors
and those measured by the SMPS–APS tandem were
made in a recent study by Shen et al. (2002). As in the
case of the PCIS–MOUDI comparisons, paired t-tests
between PCIS and SMPS–APS concentrations for these
four size ranges indicates that these concentrations are
not statistically significantly different at the p ¼ 0:05
level.
3.4. PM2.5 chemical species concentrations
Fig. 5a shows size-segregated PM2.5 average sulfate
concentrations measured by MOUDI and PCIS in
respective size fractions. PM2.5 sulfate concentrations
measured by PCIS are in close agreement with
concurrent MOUDI measurements (within 10% or less,
Table 1). Paired t-tests between the PCIS and MOUDI
concentrations for the 1–2.5, 0.5–1.0, 0.25–0.50 and
o 0.25 mm size ranges indicate that these concentrations
are not statistically significantly different (p ¼ 0:82; p ¼
0:99; p ¼ 0:90 and p ¼ 0:27; respectively).
Overall PM2.5 nitrate concentrations measured
by PCIS are also in close agreement with MOUDI
(Table 1). Fig. 5b depicts the size fractionated PM2.5
nitrate measurements by MOUDI and PCIS along with
standard deviations in respective size fractions. Paired
t-tests between the PCIS and MOUDI concentrations
for size ranges 1.0–2.5, 0.5–1.0, 0.25–0.5 mm indicate that
ARTICLE IN PRESS
M. Singh et al. / Atmospheric Environment 37 (2003) 4781–4793
4789
1
EC Conntrations (µg/m3)
PCIS
MOUDI
0.8
0.6
0.4
0.2
0
2.5-1.0
(a)
1.0-0.5
0.5-0.25
Aerodynamic Diameter (µm)
<0.25
10
PCIS
MOUDI
*
Concentations (µg/m3)
8
6
4
2
0
2.5-1.0
(b)
1.0-0.5
0.5-0.25
Aerodynamic Diameter (µm)
<0.25
Fig. 6. (a). Comparison of size-segregated EC concentrations obtained with PCIS and MOUDI and (b) comparison of OC
concentrations obtained with PCIS and MOUDI. The asterisk denotes statistically different concentrations at the p ¼ 0:05 level.
4
PCIS
MOUDI
Concentration (µg/m3)
3
2
1
0
2.5-1.0
1.0-0.5
0.5-0.25
Aerodynamic Diameter (µm)
<0.25
Fig. 7. Comparison of nitrate concentrations obtained with MOUDI and PCIS using continuous monitor (the MOUDI cutpoint of
the last stage is 0.18 mm).
ARTICLE IN PRESS
M. Singh et al. / Atmospheric Environment 37 (2003) 4781–4793
4790
6
PCIS
MOUDI
Concentrations (µg/m3)
5
4
3
2
1
0
2.5-1.0
1.0-0.5
0.5-0.25
<0.25
Aerodynamic Diameter (µm)
Fig. 8. Comparison of total carbon concentrations obtained with PCIS and MOUDI using continuous monitor (the MOUDI cutpoint
of the last stage is 0.18 mm).
Wind Speeds
1
Collection Efficiency
0.8
3 km/h
8 km/h
0.6
0.4
0.2
0
0
2
4
6
8
10
Particle Diameter (µm)
Fig. 9. Collection efficiency of the 2.5 mm PCIS stage vs. particle diameter at two different wind speeds.
these concentrations are not statistically significantly
different (p ¼ 0:90; p ¼ 0:93; p ¼ 0:78; respectively). The
difference in the PCIS–MOUDI concentrations appears
to be significant (p ¼ 0:04) in the smaller particle size
range of o0.25 mm, with the PCIS measuring about
2.7(70.13) times higher concentrations than the
MOUDI. Lower MOUDI concentrations in the
o0.25 mm range may be due to evaporative losses from
its after-filter stage, as discussed previously. The low
nitrate content of the urban aerosol in that size range
does not affect the overall good PM2.5 agreement
between the concentrations of PCIS and MOUDI.
ARTICLE IN PRESS
M. Singh et al. / Atmospheric Environment 37 (2003) 4781–4793
Figs. 6a and b show a comparison of the average size
fractionated EC and OC concentrations, respectively,
obtained with the PCIS and MOUDI along with
standard errors in the respective size fractions. For
these measurements, PCIS and MOUDI agree well (i.e.,
within 15%) for particle size ranges from 2.5 to 0.25 mm.
However the MOUDI underestimates the OC and EC
contents for particles o0.25 mm, with the average PCIS
to MOUDI ratio being 2.43(70.78) and 1.32(70.37),
respectively, for this size range. While EC concentrations of the PCIS and MOUDI for the o0.25 mm range
are not statistically significantly different (p ¼ 0:12),
those based on OC are (po0:01). The lower EC
concentrations may be due to the slight difference in
the classification of the aerosol in the 0–0.25 mm range
by PCIS and MOUDI. As previously discussed, the
algorithm used to bring the cutpoints 0.0.18 and 0.25 mm
of MOUDI and PCIS closer assumes a log-normal
distribution of aerosol in these size ranges which may
not be always true. Considering that a substantial mass
fraction of EC may be found in the 0–0.25 mm range
(Hughes et al., 1998), small differences in the size ranges
in which PM are classified by the two samplers may
result in appreciable differences in their concentrations.
This is further supported by the slightly higher MOUDI
EC concentrations in the 0.25–0.50 mm range and by the
overall good agreement in the total PM2.5 EC concentrations obtained between these samplers, as shown in
Table 1.
The significantly lower MOUDI OC concentrations
observed in the last stage and after-filter could be
attributed to volatilization of organic particles collected
under low pressure in that stage. Evaporative losses of
semi-volatile compounds from impactor stages are a
function of pressure drop across these stages (Zhang and
McMurry, 1987). The pressure drop across all MOUDI
stages is considerably higher than the corresponding
PCIS stages (the pressure drop across the 0.18 mm and
after-filter MOUDI stages are 0.06 and 0.125 atmospheres, respectively, compared to 0.009 and 0.01 atmospheres for the last two PCIS stages). This difference in
pressure drop may cause some volatilization of labile
organic species from particles collected on the MOUDI
filter. An additional explanation for the discrepancy
between PCIS and MOUDI for the o0.25 mm range may
be increased adsorption of gaseous organic compounds
on the quartz after filter of the PCIS. Organic vapor
adsorption in the preceding PCIS and MOUDI stages
can be ruled out, given that particles are collected on
aluminum substrates, but it is quite likely to occur on the
quartz after-filters of either sampler. The much lower
phase velocity and pressure drop of the PCIS after-filter
compared to MOUDI would favor this process in the
PCIS. This hypothesis is further corroborated by total
carbon measurements using continuous monitors, which
are described in subsequent paragraphs.
4791
Figs. 7 and 8 show the comparisons for nitrate and
total carbon measurements, respectively, using the semicontinuous ADI and Sunset Labs monitors. As evident
from the figures, PCIS and MOUDI agree remarkably
well (within715%) for these measurements for all size
ranges, including the 0–0.25 mm range. This finding is of
particular note, because it suggests that the differences
in labile nitrate and OC concentrations between PCIS
and MOUDI for that size range observed in the timeintegrated experiments are due to the sample collection
process, and not as much to differences in the cutpoints
between the two samplers. Sampling artifacts such as
adsorption to- and desorption from impaction and filter
substrates are phenomena that are associated with
prolonged sampling due to changes in parameter such
as temperature, humidity and vapor phase concentrations that occur typically in time scales of several hours.
The near-continuous carbon and nitrate data indicate
that the particle concentrations of PCIS and MOUDI
penetrating the 0.25 mm stage are virtually identical when
measured by a continuous monitor, the readings of which
would not be affected as much by these sampling
artifacts. However, when sampling these labile species
in longer time periods, the increased pressure drop across
the MOUDI after-filter on which these particle-bound
species are collected will enhance evaporation of volatile
compounds, as evident from the lower nitrate and OC
concentrations measured by that sampler compared to
the PCIS. The possibility of increased adsorption of
organic vapors on the PCIS after-filter also cannot be
ruled out, which would explain the higher OC values
obtained in the time integrated experiments.
3.5. Wind tunnel tests
The results of the wind tunnel test are summarized in
Fig. 9, which reveals that particle penetration characteristics of the PCIS 2.5 mm stage are unaffected by the
wind speeds. Particle collection efficiency increases
sharply as particle diameter becomes larger than about
2 mm to about 50% at 2.5 mm (the design cutpoint of that
stage) and exceed the 90% value at about 3.5 mm. The
collection efficiency curves for all the wind speeds tested,
e.g., 3 and 8 km/h, show a very close agreement to that
determined in control laboratory experiments for that
PCIS stage, described by Misra et al. (2002). This finding
is particularly important because it demonstrates that
the PCIS can be used for personal sampling in indoor or
occupational environments under non-quiescent air
conditions.
4. Summary and conclusions
The performance of the PCIS, a miniaturized cascade
impactor consisting of four impaction stages and an
ARTICLE IN PRESS
4792
M. Singh et al. / Atmospheric Environment 37 (2003) 4781–4793
after-filter, was evaluated in a field study. PM2.5
mass concentrations measured by PCIS and SMPS–
APS were compared and found to be in good agreement for all size fractions. The results also show
excellent agreement between PCIS and MOUDI for
coarse PM (PM2.510) mass. The fine PM (PM2.5)
mass as measured by PCIS is slightly higher than
MOUDI measurement, mostly due to an underestimation of particles o0.25 mm by the MOUDI.
PM2.5 sulfate measurements by MOUDI and PCIS
were found to be in good agreement with some
small differences in individual size bins. A comparison
of PCIS–MOUDI nitrate concentrations shows significant difference in the smaller particle size range
o0.25 mm, with the PCIS measuring about 2.7(70.13)
times higher concentrations than the MOUDI. Lower
MOUDI concentrations in the o0.25 mm range are
attributed to possible evaporative losses from the
MOUDI after-filter. EC and OC measurements follow
a trend similar to mass concentration measurements.
MOUDI and PCIS agree well for particles in the range
2.5–0.25 mm, however MOUDI underestimates the
carbonaceous content for particles o0.25 mm. The
relatively small discrepancy between PCIS and MOUDI
for EC measurements is attributed to a difference in
classification of particles in 0–0.25 mm range by PCIS
and MOUDI. However, the significant difference in OC
concentrations could be attributed to either volatilization of OC particles collected under low pressure in the
last stages and/or to increased adsorption of gaseous
organic compounds on the quartz after-filter of the
PCIS.
In an attempt to eliminate sampling artifacts associated with time-integrated measurements, nitrate and
total carbon measurements by PCIS and MOUDI using
continuous monitors were compared. PCIS and MOUDI agreed well (within715%) for these measurements
for all size ranges, including the o0.25 mm range. We
thus conclude that differences for nitrate and OC
concentrations between PCIS and MOUDI observed
in the time-integrated experiments are due to the sample
collection process and not to differences in the cutpoints
between the samplers.
The sampling efficiency of PCIS was also evaluated in
wind tunnel tests, which confirmed that sampling and
collection of super-micrometer particles by the PCIS is
unaffected at wind speeds as high as 8 km/h.
Acknowledgements
This work was supported in parts by the Mickey
Leland National Urban Air Toxics Center through
Grant #53-4507-7821 and by the Southern California
Particle Center and Supersite (Grants #53-4507-7721
and 53-4507-0482) to USC. The research described in
this article has not been subjected to the Agency’s
required peer and policy review and therefore does not
necessarily reflect the views of the Agency and no official
endorsement should be inferred.
References
Ayres, J., Fleming, D., Williams, M., McInnes, G.,
1989. Measurement of respiratory morbidity in general
practice in the United Kingdom during the acid transport
event of 1985. Environmental Health Perspectives 79,
83–88.
Bates, D.V., Sizto, R., 1989. The Ontario air pollution studies:
identification of the causative agent. Environmental Health
Perspectives 79 (1), 69–72.
Bates, D.V., Baker-Anderson, M., Sizto, R., 1990. Asthma
attack periodicity: a study of hospital emergency visits in
Vancouver. Environmental Research 51 (1), 51–70.
Demokritou, P., Gupta, T., Ferguson, S., Koutrakis, P., 2002.
Development and laboratory performance evaluation of a
personal cascade impactor. Journal of the Air and Waste
Management Association 52 (10), 1230–1237.
Dockery, D.W., Pope, C.A., Xu, X., Spengler, J.D., Ware, J.H.,
Ferris, B.G., Speizer, F.E., 1993a. An association between
air-pollution and mortality in 6 United-States cities. New
England Journal of Medicine 329 (24), 1753–1759.
Dockery, D.W., Damokosh, A.I., Neas, L.M., Raizenne, M.,
Spengler, J.D., Koutrakis, P., Ware, J.H., Speizer, F.E.,
1993b. Health effects of acid aerosols on North-American
children: respiratory symptoms and illness. American
Review of Respiratory Diseases 147 (4), A633.
Fung, K., 1990. Particulate carbon speciation by MnO2
oxidation. Aerosol Science and Technology 12, 122–127.
Gordian, M.E., Ozkaynak, H., Xue, J.P., Morris, S.S.,
Spengler, J.D., 1996. Particulate air pollution and respiratory disease in anchorage, Alaska. Environmental Health
Perspectives 104 (3), 290–297.
Harrison, R.M., Peak, J.D., 1996. Atmospheric aerosol major
ion composition and cloud condensation nuclei over the
Northeast Atlantic. Journal of Geophysical Research 101
(D2), 4425–4434.
Hinds, W.C., Kuo, T.L., 1995. A low-velocity wind tunnel to
evaluate inhalability and sampler performance for large dust
particles. Applied Occupational and Environmental Hygiene 32, 549–556.
Hughes, L.S., Cass, G.R., Gone, J., Ames, M., Olmez, I., 1998.
Physical and chemical characterization of atmospheric
ultrafine particles in the Los Angeles area. Environmental
Science and Technology 32, 1153–1161.
Kennedy, N.J., Tatyan, K., Hinds, W.C., 2001. Comparison of
a simplified and full-size mannequin for the evaluation of
inhalable sampler performance. Aerosol Science and Technology 35, 564–568.
Li, N., Sioutas, C., Cho, A., Schmitz, D.A., Misra, C., Sempf,
J., Oberley, T., Froines, J.R., Nel, A., 2003. Ultrafine
particulate pollutants induce oxidative stress and mitochondrial damage. Environmental Health Perspectives 111 (4),
455–460.
McMurry, P.H., Wang, X., Ehara, K., 2002. The relationship
between mass and mobility for atmospheric particles; a new
ARTICLE IN PRESS
M. Singh et al. / Atmospheric Environment 37 (2003) 4781–4793
technology for measuring particle density. Aerosol Science
and Technology 36 (2), 227–238.
Misra, C., Singh, M., Shen, S., Hall, P.M., Sioutas, C., 2002.
Development and evaluation of a personal cascade impactor
sampler (PCIS). Journal of Aerosol Science 33, 1027–1047.
Peters, T.M., Chein, H.M., Lundgren, D.A., 1993. Comparison
and combination of aerosol size distributions measured with
a low impactor, differential mobility particle sizer, electrical
aerosol analyzer, and aerodynamic particle sizer. Aerosol
Science and Technology 19, 396–405.
Raizenne, M., Burnett, R.T., Stern, G., Franklin, C.A.,
Spengler, J.D., 1989. Acute lung function responses to
ambient acid aerosol exposures in children. Environmental
Health Perspectives 79, 179–185.
Raizenne, M., Neas, L., Damokosh, A., Dockery, D.W.,
Spengler, J.D., Koutrakis, P., Ware, J., Speizer, F., 1993.
Health-effects of acid aerosols on North-American children:
pulmonary function. American Review of Respiratory
Diseases 147 (4), A635.
Rubow, K.L., Marple, V.A., Olin, J., McCauley, M.A., 1987. A
personal cascade impactor: Design, evaluation and calibration. American Industrial Hygiene Association Journal 48
(6), 532–538.
Schauer, J.J., Fraser, M.P., Cass, G.R., Simoneit, B.R.T., 2002.
Source reconciliation of atmospheric gas-phase and
particle-phase pollutants during a severe photochemical
smog episode. Environmental Science and Technology 36,
3806–3814.
Schwartz, J., Dockery, D.W., 1992. Particulate air pollution
and daily mortality in Steubenville, Ohio. American Journal
of Epidemiology 135 (1), 12–19.
4793
Shen, S., Zhu, Y., Jaques, P.A., Sioutas, C., 2002. Evaluation of
the SMPS–APS system as a Continuous Monitor for PM2.5
and PM10. Atmospheric Environment 36, 3939–3950.
Stolzenburg, M.R., Dutcher, D., Kirby, B., Kreisberg, N.,
Hering, S.V., 2002. Automated measurement of the size
distribution of airborne particulate nitrate. Aerosol Science
and Technology 37 (7), 537–546.
Suh, H.H., Koutrakis, P., Spengler, J.D., 1994. Interactions
between aerosol acidity and ammonia in indoor environments. Journal of Exposure Analysis and Environmental
Epidemiology 4, 1.
Thomas, K.W., Pellizzari, E.D., Clayton, C.A., Whitaker,
D.A., Shores, R.C., Spengler, J., Ozkaynak, H., Frohlich,
S.E., Wallace, L.A., 1993. Particle total exposure assessment
methodology (PTEAM) 1990 study: method performance
for personal, indoor, and outdoor monitoring. Journal of
Exposure Analysis and Environmental Epidemiology 3 (2),
203–1226.
Thurston, G.D., Ito, K., Kinney, P.L., Lippmann, M., 1992. A
multiyear study of air-pollution and respiratory hospital
admissions in 3 New-York-state metropolitan areas: results
for 1988 and 1989 summers. Journal of Exposure Analysis
and Environmental Epidemiology 2 (4), 429–450.
Thurston, G., Lippmann, M., Bartoszek, M., Fine, J., 1993.
Air-pollution associations with asthma exacerbations, peak
flow changes, and respiratory symptoms in Children at a
Summer Asthma Camp. American Review of Respiratory
Disease 147 (4), A633.
Zhang, X., McMurry, P.H., 1987. Theoretical analysis of
evaporative losses from impactor and filter deposits.
Atmospheric Environment 21, 1779–1789.