3509
J. Sep. Sci. 2011, 34, 3509–3516
Joseph J. Pesek1
Maria T. Matyska1
Steven M. Fischer2
1
Department of Chemistry, San
Jose State University, San Jose,
CA, USA
2
Agilent Technologies, Santa
Clara, CA, USA
Received July 11, 2011
Revised August 30, 2011
Accepted August 31, 2011
Research Article
Improvement of peak shape in aqueous
normal phase analysis of anionic
metabolites
The problem of poor peak shape for multiply charged negative-ion analytes under aqueous
normal phase (ANP) conditions is investigated. Because less than adequate efficiency and
symmetry can occur with a variety of mobile phases, gradients and additives, and to
varying degrees depending on the instrument, sources other than solute/stationary phase
interactions are more likely the cause. Since it is known that many of these compounds
can interact strongly with metal ions, addition of a chelating agent to the mobile phase
and/or the sample solvent was tested. In particular, ethylenediaminetetraacetic acid
(EDTA) is a compound that forms strong complexes with most di-and tri-valent metal ions
and can be used to verify whether trace amounts of these species are the source of the
problem. In addition, the retention of a number of anionic compounds was measured at
various concentrations of ammonium acetate and formate with EDTA in the mobile phase.
Keywords: Hydrophilic compounds / Metal ions / Silica hydride stationary phase
DOI 10.1002/jssc.201100607
1 Introduction
Over the last several years, silica hydride stationary phases
have demonstrated capabilities that are superior than many
existing materials and possess the potential to provide
solutions for some of the most demanding biological,
clinical and pharmaceutical analyses [1–19]. These columns
are referred to as TYPE-CTM silica because they are
fundamentally different from many current HPLC stationary phases that utilized ordinary silica. The essential
difference between the two materials is that TYPE-CTM
phases have a surface that is populated with Si-H (silica
hydride) groups while Si-OH (silanols) groups dominate
ordinary silica. While this might seem like a trivial
difference, it has profound effects on the fundamental
nature of the material and hence its chromatographic
properties. Silanols are very polar and can often interact
irreversibly with polar compounds, especially bases, while
silica hydride is weakly hydrophobic and results in less
strongly adsorptive properties that are advantageous for
good chromatographic performance.
Several prominent innovations have emerged from
investigations using silica hydride-based stationary phases
and aqueous normal phase (ANP) chromatography. One
important feature is the dual retention capability of all silica
hydride columns fabricated to date. Both hydrophobic and
Correspondence: Professor Joseph J. Pesek, Department of
Chemistry, San Jose State University, San Jose, CA 95192, USA
E-mail: joseph.pesek@sjsu.edu
Fax: 11-408-924-4945
Abbreviations: ANP, aqueous normal phase;
hydrophilic interaction liquid chromatography
& 2011 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
HILIC,
hydrophilic compounds can be retained efficiently when a
silica hydride column is used. The elution order depends on
the type of gradient used (RP gradient from high aqueous to
high organic versus an ANP gradient from high organic to
high aqueous). This retention behavior distinguishes ANP
from hydrophilic interaction liquid chromatography
(HILIC). While retention of polar compounds occurs at high
organic content in the mobile phase ordinary reversed-phase
retention can be easily obtained by using high-aqueous
content mobile phases. This significant dual retention
capability (either simultaneous or sequential) cannot be
obtained using either standard reversed-phase or HILIC
separation materials. Another characteristic is that the
hydrophobic silica hydride surface absorbs less water than
ordinary silica thus creating a more uniform solvent environment around the stationary phase and at the particle
interface. The lack of a substantial water layer on the
hydride surface (a single mobile phase environment) is
likely responsible for the rapid equilibration of the stationary phase after gradient analyses and perhaps the higher
efficiency observed in the ANP compared with HILIC [4].
Another feature of the silica hydride materials that has been
established for a wide range of samples and stationary
phases is the reproducibility of retention times from run to
run. Reproducibilities for a particular analysis are usually in
the range of 0.1–0.5% RSD, even for samples in physiological matrices. Another attractive characteristic of the silica
hydride materials is their long-term durability. The bonding
of an organic moiety to a silica hydride surface via hydrosilation results in a direct silicon–carbon bond.
Additional correspondence: Maria T. Matyska
E-mail: maria.matyska-pesek@sjsu.edu
www.jss-journal.com
3510
J. Sep. Sci. 2011, 34, 3509–3516
J. J. Pesek et al.
One problem that has been identified in our laboratory
for highly negatively charged compounds is poor peak
shape. Similar observations have also been reported for
HILIC columns using a nano-column format [20]. Nucleotides and organic acids with more than one carboxylate
group are the primary analytes affected. Originally, it was
thought that this was just a result of the high negative
charge of the analyte and the resulting poor exchange
kinetics. However, it was shown in the HILIC study that the
real cause was the presence of trace metals, particularly iron,
that is often found in very low concentrations from a variety
of sources within the chromatographic system and the
solvents. Our work has confirmed this conclusion using
several LC/MS and LC/MS/MS instruments. In addition, it
appears that copper may also contribute to the poor
performance of the highly negatively charged ions. The
solution in the HILIC study was to add a trace amount of
EDTA to the mobile phase. Thus, in order to make the ANP
method suitable for every type of negatively charged species
a remedy must be found to remove the trace amounts of
iron and copper in the system. This study explores the use
of EDTA as a possible solution to improve the peak shape of
negatively charged analytes in the ANP mode in a standard
analytical column format.
2 Materials and methods
2.1 Materials
The silica hydride stationary phase used in this study was
the Cogent Diamond Hydride (DH) material in
150 2.1 mm columns (MicroSolv Technology, Eatontown,
NJ, USA). The phase contains a small amount of an
organic moiety (2% carbon as reported by the manufacturer) on a silica hydride surface. The analytes and
mobile phase additives used in this study were purchased
from Sigma-Aldrich (Milwaukee, WI, USA) in the highest
purity available. Mobile phase solvents used were HPLC
grade.
Figure 1. EIC of ATP in negative-ion mode at m/z 505.9885. (A) No addition of EDTA and (B) after injection of 100 mL of a 1-mg/mL solution
of EDTA. Mobile phase: Solvent A, 50:50 MeOH/DI water with 0.05% formic acid. Solvent B, 90:10 acetonitrile/DI water with 10 mM
ammonium acetate adjusted to pH 7. Injection vol. 5 1 mL. Gradient: 0.0–1.0 min at 100% B; 1.0–3.0 min to 90% B; 3.0–6.0 min at 90% B;
6.0–7.0 min to 80% B; 7.0–9.0 min at 80% B; 9.0–10.0 min to 50% B; 10.0–12.0 min at 50% B; 12.0–13.0 min to 30% B; 13.0–15.0 min at 30% B.
Flow rate 5 0.4 mL/min.
& 2011 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.jss-journal.com
J. Sep. Sci. 2011, 34, 3509–3516
2.2 Instrumentation
The HPLC was an Agilent (Little Falls, DE, USA) 1200SL
Series LC system, including degasser, binary pump,
temperature-controlled autosampler and temperaturecontrolled column compartment. The mass spectrometer
system was an Agilent (Santa Clara, CA, USA) Model 6220
MSD TOF with a dual sprayer electrospray source (ESI).
2.3 Methods
Stock solutions of the analytes were made in deionized (DI)
water in the range of 0.2–0.7 mg/mL. Sample solutions were
Liquid Chromatography
3511
made by diluting the stock 1:100 in 50:50 acetonitrile/water
containing the mobile phase additive used in the analysis.
The flow rate was 0.4 mL/min. The column temperature
was 201C.
3 Results and discussion
Negatively charged analytes like ATP with more than a
single ionizable group often appear as irregular peaks when
eluted in the ANP mode. An example of a typical
chromatogram for ATP is shown in Fig. 1A. At first this
was believed to be caused by poor exchange kinetics between
the stationary phase and the mobile phase solvent. A
Figure 2. EICs for nucleotides utilizing addition of 10 mM EDTA to the B solvent. Gradient: 0.0–1.0 min at 100% B; 1.0–15.0 min to 0% B;
15.0–17.0 min at 0% B. Other conditions same as Fig. 1. Peak identification: 1, AMP; 2, ADP; 3, ATP; 4, NAD.
Figure 3. EICs for nucleotides utilizing addition of 10 mM EDTA to the A solvent. All conditions and peak identification are the same as Fig. 2.
& 2011 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.jss-journal.com
3512
J. J. Pesek et al.
number of approaches were attempted to try to remedy this
situation including varying the pH of the mobile phase,
utilizing different gradients and increasing the column
temperature. None of these proved satisfactory in achieving
both high symmetry and efficiency. In addition, the degree
of asymmetry varied considerably between LC systems even
under identical experimental conditions. Because low
concentrations of metal ions, particularly iron, were
suspected as the cause of poor peak shape for anionic
metabolites in nano-LC [5], it was assumed that a similar
problem might be present in typical analytical HPLC
systems. In order to test this assumption, EDTA was
introduced as part of either the mobile phase or the sample
solvent.
There are three possible means of utilizing EDTA in the
LC system in order to remove potential trace metal ion
contaminants: (i) injecting a significant volume of a high-
J. Sep. Sci. 2011, 34, 3509–3516
concentration solution of EDTA in order to remove the trace
metals in a single treatment; (ii) including EDTA at low
concentration as part of the mobile phase; and (iii) preparing the sample with EDTA as part of the solvent. Each of
these approaches was tested in order to determine which
one would produce good peak shape and efficiency as well
as have the least effect on sensitivity when using MS for
detection.
The first approach was tested by injecting 100 mL of a
1-mg/mL solution of EDTA into the column and then
equilibrating the column for 15 min using the starting
composition of the mobile phase for the gradient being used
in the subsequent injection. The initial injection resulted in
peaks of very low intensity. The chromatogram in Fig. 1B
shows the result of the fifth injection after the EDTA
treatment of the same ATP sample solution and using the
same gradient as in Fig. 1A with the DH column. A very
Figure 4. EICs for nucleotides and related compounds utilizing addition of 5 mM EDTA to Solvents A and B. Mobile Phase: Solvent A, 50:50
MeOH/DI water with 0.025% formic acid. Solvent B, 90:10 acetonitrile/DI water with 5 mM ammonium acetate adjusted to pH 7. Injection
vol. 5 1 mL. Gradient same as Fig. 1. Flow rate 5 0.4 mL/min. Peak identification: 1–4 same as Fig. 2; 5, NADP; 6, UDP-galactose; 7, UDPglucose; 8, GTP; 9, galactose-1-phosphate.
& 2011 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.jss-journal.com
J. Sep. Sci. 2011, 34, 3509–3516
Liquid Chromatography
3513
Figure 5. EIC at m/z 115.0037. (A) No EDTA in the mobile phase and (B) after injection of 10 mL of 3.4 mM EDTA. All conditions same as
Fig. 1. Peak identification: 1, maleic acid and 2, fumaric acid.
significant improvement is seen in the peak shape so that
both retention time identification and quantitation are
possible. The reproducibility of this data including peak
intensity is quite good for approximately ten subsequent
injections after which noticeable peak tailing is evident
indicating that the presence of trace metal ions such as iron
and copper has returned to levels that have measureable
effects. Peak shape identical to that shown in Fig. 1B can be
obtained by another injection of the high concentration
EDTA solution. Similar behavior is observed for a number
of other di- and tri-phosphate containing compounds. These
results indicate that the source of peak tailing is external to
the column and as reported in the nano-LC study is probably
due to trace metal ions from either the instrument or the
solvents used for preparing the mobile phase. However,
from a practical aspect, the use of repeated injections of
EDTA and equilibration of the column on a frequent basis
does not seem desirable.
The next approach tested was to place EDTA in the
sample solution. At concentrations of 5–10 mM there was a
& 2011 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
noticeable improvement in peak shape (AS5) but not
sufficient for most analytical determinations. At 10–20 mM
there was significant improvement in peak shape (AS3)
over that shown in Fig. 1A, but there was a noticeable
decrease (450%) in the sample signal since it was determined that EDTA has retention comparable to ATP and
several other phosphate containing species on the DH
column. The retention of EDTA on the DH column was
verified by monitoring the extracted ion chromatogram at
m/z 291. Thus it was concluded that this was not a viable
approach for improving peak shape for these anionic species.
The third means of adding EDTA to the system is to
place a low concentration in the mobile phase. There are
several possibilities for accomplishing this. The first case
involves adding EDTA to only Solvent B (the organic
component). The results of this test are shown in Fig. 2 for
several nucleotides with optimization of the gradient under
these conditions. As can be seen, there are reasonably good
peak shapes for some compounds but others were still not
suitable for developing a good analytical method. Next, the
www.jss-journal.com
3514
J. J. Pesek et al.
J. Sep. Sci. 2011, 34, 3509–3516
Figure 6. EIC at 521.9834. (A) Mobile
phase same as Fig. 1 except with
10 mM ammonium formate and (B)
mobile phase same as Fig. 1. Injection
vol. 1 mL and flow rate 5 0.4 mL/min.
Gradient:
0.0 min
100%
B;
0.0–10.0 min to 20% B. Peak identification: 1, UDP-galactose and 2, UDPglucose.
EDTA was added only to the aqueous component of the
mobile phase (Solvent A). The results for this approach are
shown in Fig. 3 for a representative number of analytes. As
can be seen, the peak shape has improved in comparison to
the results obtained for EDTA in the B solvent. In all cases,
the chromatographic peak shapes are satisfactory for most
analytical methods. The final approach investigated was to
add EDTA to both the A and B solvents. The results of this
mobile phase composition are shown in Fig. 4. Under these
conditions each of the species tested gives both good efficiency as well as peak symmetry. In addition, UDP-glucose
and UDP-galactose are almost completely resolved. The
concentrations of all the mobile phase constituents are
relatively low: 5 mM for EDTA, 0.025% for formic acid and
5 mM for ammonium formate. Under these conditions,
signal suppression in the MS is minimized which is beneficial for method development requiring low detection
limits. The use of 10 mM EDTA caused measurable signal
depression for many compounds. At 20 mM EDTA signal
intensity for most compounds was significantly decreased
(450%). The ionization technique used is gas-assisted
& 2011 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
electrospray. This technique is capable of producing charged
aerosol at much higher flow rates than pure electrospray,
but is subject to the effect of non-volatile material in the
solvent. Ammonia acetate and formate buffers are essentially non-volatile in the time frame of the electrospray
ionization and desorption process. The result is sensitivity
decreases with increasing buffer concentration in a nonlinear fashion. EDTA’s effect on signal response arises from
a different mechanism than the buffers. It is an intermediately strong acid resulting in charge competition with
the analytes to be ionized by the electrospray process. The
continuous addition of 5 mM of EDTA to the mass spectrometer ion source will not damage the instrument. The
EDTA will deposit on the surfaces of the ion source but
cause no damage. Occasional cleaning of the ion source will
be required but the ‘‘volatile’’ buffers ammonium formate
or ammonium acetate accumulate faster than the EDTA.
For some compounds, the use of EDTA in the mobile
phase is not necessary. Good examples are the isobaric
diprotic acids, maleic and fumaric. The results of one
gradient run in the ANP mode are shown in Fig. 5. Figure
www.jss-journal.com
J. Sep. Sci. 2011, 34, 3509–3516
Liquid Chromatography
3515
Figure 7. EICs for NAD and NADP.
Gradient and other conditions same
as Fig. 6. Peak identification: 1A, NAD
in 5 mM ammonium acetate; 1B, NAD
in 10 mM ammonium acetate; 2A,
NADP in 5 mM ammonium acetate;
2B, NADP in 10 mM ammonium
acetate.
Figure 8. EICs for ATP and GTP.
Gradient and other conditions same
as Fig. 6. Peak identification: 1A, ATP
in 5 mM ammonium acetate; 2A, GTP
in 5 mM ammonium acetate; 1B, ATP
in 10 mM ammonium acetate; 2B,
GTP in 10 mM ammonium acetate.
5A is the EIC at m/z 115 for the two compounds using a
mobile phase with 10 mM ammonium acetate as the additive. The column is then injected with 10 mL of a 3.4-mM
solution of EDTA. Figure 5B shows the chromatogram that
is obtained after the EDTA treatment. For all practical
purposes, it is essentially the same as the one shown in Fig.
5A. Similar results are obtained for the mobile phase used
in Fig. 4. Using a variety of gradients, it is possible to obtain
good peak shape for both of these acids in ammonium
acetate or formate buffers. The compounds that are affected
by trace metals are those capable of coordination to multi& 2011 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
valent cations. Phosphate-containing compounds can coordinate to multivalent cations and the level of impact is
dependent on the number of phosphates in the molecule;
34241. The di-acids are not affected because they do not
contain phosphate and do not have a geometry that supports
cation coordination. The result is that the di-acids do not
coordinate with multivalent ions while ATP strongly coordinates with them.
In addition to the amount of EDTA, factors such as type
of additive as well as its concentration were also evaluated
with respect to retention and signal intensity. Figure 6
www.jss-journal.com
3516
J. Sep. Sci. 2011, 34, 3509–3516
J. J. Pesek et al.
shows the effect of the type of additive on retention, resolution and signal intensity. In this case, the evaluation is
done for UDP-glucose and UDP-galactose, two compounds
that are difficult to resolve due to the minor structural
differences between them. There are some differences
observed in the results between the additives. There is
slightly longer retention for ammonium acetate but slightly
better resolution for ammonium formate. The signal levels
are comparable since the longer retention in ammonium
acetate results in somewhat broader peaks. Another example is shown in Fig. 7 for the separation of NAD and NADP.
Retention is longer for both compounds in the 10 mM
ammonium acetate mobile phase than in the 5 mM buffer.
There is some decrease in signal intensity for NAD when a
10-mM solution of the additive is used in comparison with
the 5 mM buffer. Figure 8 shows another comparison of
experimental conditions. In this case using ammonium
acetate as the buffer it can be seen that pH effects retention
similarly to concentration. In this example both ATP and
GTP elute at essentially the same retention time with the
gradient used. The higher concentration mobile phase
(10 mM ammonium acetate) also is adjusted to a higher pH
than the lower concentration eluent (5 mM). Longer retention is observed in the higher concentration of ammonium
acetate at elevated pH. In addition, there is a small signal
increase at the higher ammonium acetate concentration for
both compounds.
4 Concluding remarks
Peak shape (efficiency and symmetry) can be significantly
improved by the addition of EDTA at the low micromolar
level to the mobile phase. Continuous treatment is more
effective than a single injection of a high-concentration
EDTA solution or addition of the chelating agent to the
sample solvent. It is assumed that the cause of the poor
performance is the presence of trace amounts of metal ions
in the system due to the effect of EDTA and the variability of
peak shape from instrument to instrument. The use of
EDTA is in most instances a more practical method of
removing trace metals rather than vigorous leaching of the
instrument with an agent such as a strong acid. The main
drawback to the use of EDTA is the suppression of the
analyte signal because under the mobile phase conditions
used it also has more than a signal negative charge. Therefore,
investigations are continuing into other reagents that might
be used to complex the trace metal ions in the system without
having a significant impact on analyte sensitivity.
The authors acknowledge the support of the National
Science Foundation (Grant 0724218) and Agilent Technologies,
& 2011 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
Santa Clara, CA for donation of the equipment used in this
study.
The authors have declared no conflict of interest.
5 References
[1] Brown, L., Ciccone, B., Pesek, J. J., Matyska, M. T.,
American Lab. 2003, 35, 23–29.
[2] Pesek, J. J., Matyska, M. T., LCGC 2006, 24, 296–303.
[3] Pesek, J. J., Matyska, M. T., J. Sep. Sci. 2009, 32,
3999–4011.
[4] Pesek, J. J., Matyska, M. T., Loo, J. A., Fischer, S. M.,
Sana, T. R.., J. Sep. Sci. 2009, 32, 2200–2208.
[5] Pesek, J. J., Matyska, M. T., Fischer, S. M., Sana, T. R.,
J. Chromatogr. A 2008, 1204, 48–55.
[6] Pesek, J. J., Matyska, M. T., Hearn, M. T. W., Boysen, R.
I., J. Chromatogr. A 2009, 1216, 1140–1146.
[7] Pesek, J. J., Matyska, M. T. in: Grushka, E., Grinberg, N.
(Eds.), Advances in Chromatography, vol. 48, CRC
Press, Boca Raton, FL 2010, pp. 255–288.
[8] Pesek, J. J., Matyska, M. T., Duley, J., Zamzami, M.,
Fischer, S. M., J. Sep. Sci. 2010, 33, 930–938.
[9] Young, J. E., Matyska, M. T., Pesek, J. J., J. Chromatogr.
A 2011, 1218, 2121–2126.
[10] Pesek, J. J., Matyska, M. T., in: Wang, P. G., He, H.
(Eds.), Hydrophilic Interaction Chromatography (HILIC)
and Advanced Applications, CRC Press, Boca Raton, FL
2011, pp. 1–26.
[11] Watson, D. G., Parasitology 2010, 137, 1409–1423.
[12] Dejaegher, B., Vander Heyden, Y., J. Sep. Sci. 2010, 33,
698–715.
[13] Dell’mour, M., Jaitz, L., Oburger, E., Puschenreiter, M.,
Koellensperger, G., Hann, S., J. Sep. Sci. 2010, 33,
911–922.
[14] Cheng, C., Chen, C.-S., Hsieh, P.-H., J. Chromatogr. A
2010, 1217, 2104–2110.
[15] Callahan, D. L., De Souza, D., Bacic, A., Roessner, U.,
J. Sep. Sci. 2009, 32, 2273–2280.
[16] Butterfield, T. R., Fischer, S. M., Rhee, K. Y., J. Sep. Sci.
2009, 32, 2262–2265.
[17] de Carvalho, L. P. S., Zhao, H., Dickinson, C. E., Arango,
N. M., Lima, C. D., Fischer, S. M., Ouerfelli, O., Rhee, K.
Y., Chem. Biol. 2010, 17, 323–332.
[18] Marrero, J., Rhee, K. Y., Schnappinger, D., Pethe, K.,
Ehrt, S., Proc. Natl. Acad. Sci. USA 2010, 107,
9819–9824.
[19] De Carvalho, L. P. S., Fischer, S. M., Marrero, J., Nathan,
C., Ehrt, S., Rhee, K. Y., Chem. Biol. 2010, 17, 1122–1131.
[20] Myint, K. T., Uehara, T., Aoshima, K., Oda, Y., Anal.
Chem. 2009, 81, 7766–7772.
www.jss-journal.com