Nothing Special   »   [go: up one dir, main page]

Academia.eduAcademia.edu

Design of compliance chamber and after-load in apparatus for cultured endothelial cells subjected to stresses

2006, Cell Biology International

Cell Biology International 30 (2006) 439e444 www.elsevier.com/locate/cellbi Design of compliance chamber and after-load in apparatus for cultured endothelial cells subjected to stresses Hao Ding a,b, Aike Qiao a, Lixing Shen b, Mingyang Li b, Zhenglong Chen b, Xiaojun Yu c, Yanjun Zeng a,c,* a Biomechanics & Medical Information Institute, Beijing University of Technology, No. 100 PingLeYuan, Beijing 100022, China b The Medical Instrumentation College, University of Shanghai for Science and Technology, Shanghai 200093, China c Medical College of Shantou University, Shantou 515031, China Received 28 November 2005; revised 10 January 2006; accepted 14 February 2006 Abstract In order to create a hemodynamic environment that can simulate the physiological condition of arteries, an in vitro experiment apparatus was designed whose key modules were compliance chamber and after-load. These two modules were developed based on the theories of hemodynamics. Both the normal and shear stress to which endothelial cells are exposed can be controlled with these modules, thus facilitating the research of endothelial cells subjected to stresses. Ó 2006 International Federation for Cell Biology. Published by Elsevier Ltd. All rights reserved. Keywords: Biomechanics; Hemodynamic environment; Experiment; Simulation; Blood flow; Artery 1. Introduction Endothelial cells (ECs), which form a layer of membrane covering heart valves and blood vessels, play a very important role in the physiological and pathological activities of the cardiovascular system. Changes in their structure and function are key events in the occurrence and development of some vascular diseases, such as hypertension and atherosclerosis (Dewey et al., 1981; Franke et al., 1984; Eskin et al., 1984; Helmlinger et al., 1991). ECs in vivo are always exposed to a hemodynamic environment. Many previous research studies have considered that shear stress of blood flow is the main factor influencing ECs (Frame et al., 1998; Passerini et al., 2003; Chen et al., 2003; Greisler et al., 1990; Brooks et al., 2002; Mu and Du, 2004). In fact, besides shear stress, normal stress (i.e. blood pressure) also acts on ECs. Under normal physiological * Corresponding author. Biomechanics & Medical Information Institute, Beijing University of Technology, No. 100 PingLeYuan, Beijing 100022, China. Tel.: þ86 10 67391685. E-mail address: yjzeng@bjut.edu.cn (Y. Zeng). conditions, normal stress, with magnitudes of 15,996/10,664 Pa (120/80 mm Hg) in the systolic and diastolic cardiac phases, respectively, is 5300e8000 times greater than shear stress (2 Pa) (Liu et al., 1997, 2001). Obviously normal stress is an important factor affecting ECs that cannot be ignored. In view of the fact that it is impossible to measure the shear stress of blood flow in arteries both in vivo and in vitro, many in vitro experiments have been conducted to calculate shear stress indirectly (Liu et al., 2001). In order to create a realistic hemodynamic environment (including both shear stress and normal stress) to which the ECs are exposed, we developed an apparatus that can simulate steady and pulsatile blood flow phenomena in arteries. In this apparatus, the compliance chamber and after-load modules are two key modules for the simulation of in vitro experiments. The purpose of this paper is to show the design of these two important modules in the apparatus. 2. Methodology Technically, the compliance chamber and after-load should satisfy the following requirements: first, they are able to simulate normal human aorta 1065-6995/$ - see front matter Ó 2006 International Federation for Cell Biology. Published by Elsevier Ltd. All rights reserved. doi:10.1016/j.cellbi.2006.02.003 440 H. Ding et al. / Cell Biology International 30 (2006) 439e444 systolic and diastolic pressure (systolic pressure: 15,996e23,994 Pa, i.e. 120e 180 mm Hg; diastolic pressure: 7998e15,996 Pa, i.e. 60e120 mm Hg); second, the above-mentioned parameters should be adjustable within a stated range. In our design, the test apparatus has a rectangular flow chamber whose height is much less than its width and where the ECs are planted. For Poiseuille flow in a rectangular chamber, the wall shear stress can be denoted as t¼ 2.1. Principle of the compliance chamber The function of the compliance chamber is to simulate elastic function of the aorta, and change the intermittent blood ejection of the pulsatile pump into a continuous pulsatile flow. The compliance chamber, which is composed of a gas chamber and a liquid chamber to simulate the elastic artery, stores and discharges energy for the test system. Its gas chamber volume can be calculated with (Zhang and Huang, 1991): V0 ¼ P2  DV  1 1  P2 P1 ð1Þ where DV is pump’s output per pulse; P1 and P2 are the systolic pressure and the diastolic pressure, respectively. The total gas volume V0 correspondingly varies with different P1 and P2 . But the following law is always satisfied: the smaller the difference between P1 and P2 , the bigger the V0 . When P1 ¼ P2 , V0 is infinite. 6m Q bd2 ð3Þ where b is the width of the rectangular chamber, and d the height of the rectangular chamber. Obviously, the shear stress is proportional to the flow rate. The flow rate in the pipe can be obtained from the following equation (Diao, 1991; Liu and Li, 1997): Q¼ DP R ð4Þ where DP ¼ P1  P2 is the pressure drop along the pipeline, and R the total peripheral resistance. The flow rate of the heart pump keeps constant if the output per pulse and the heart rate do not vary. Thus, the pressure drop is proportional to the resistance. Increasing R can induce the increase of DP. Blood flow downstream from the after-load is open to the atmosphere directly in the apparatus; therefore, the pressure drop is just the blood pressure in the artery. Thus, the blood pressure in the artery increases with the increase of peripheral resistance and vice versa. 2.2. Principle of the after-load 3. Implementation The function of the after-load is to simulate the peripheral resistance of the micro-vessels and capillary vessels (‘‘lump resistance model’’). Blood pressure is the only normal stress acting on the ECs. The blood pressure level can be changed by adjusting the opening of the after-load. We have found that the throttle, as an effective load, plays a key role in adjusting peripheral resistance of the vascular system. The throttle is a ‘‘lump resistance model’’ whose role is not different from that of capillary vessels when studying large arteries in hemodynamics. Physiologically, the total length of capillary vessels is very long and the resistance distributes along the total length without converging at one point. The distribution of resistance is different depending on the difference in diameter, length and branch of capillary vessels. In this case, the ‘‘lump resistance model’’ has no effect when studying the behavior of capillary vessels. However, when studying the blood flow in large arteries far upstream from capillary vessels, it is not only feasible but also technically convenient to consider the influence of capillary vessels as ‘‘lump resistance’’. As a preliminary study, we assume that the blood flow is a laminar flow in a rigid circular pipe. The shear stress acting on the ECs is just the shear wall stress of arteries. According to Poiseuille’s law, parabolic velocity profile leads to wall shear stress t as follows (Diao, 1991; Liu and Li, 1997): This apparatus consists of a pulsatile pump, a compliance chamber, a rectangular chamber, a reservoir, a thermostatic device, an after-load system and joint pipelines. A diagram of the testing circulation system is shown in Fig. 1. t¼ 32mQ pd3 ð2Þ where Q is the volume flow rate, m the viscosity and d the diameter of blood vessel. 3.1. Compliance chamber The compliance chamber (54 mm long, 72 mm wide, 82 mm high) includes a gas chamber, gas loading valve and purge valve (Fig. 2). The compressed liquid medium from the pulsatile pump is divided into two ways: one enters the pipeline system; the other enters the compliance chamber, which can compress air in the gas chamber and store pressure energy. The pressure level of the pipeline decreases when the pump sucks liquid in. The compressed air in the compliance chamber extrudes pressure liquid into the pipeline. So the compliance chamber functions like a buffer that not only decreases higher pressure by absorbing pressure energy but also increases lower pressure by discharging pressure energy. Thus, the compliance chamber Fig. 1. Semantic of the testing circulation system. H. Ding et al. / Cell Biology International 30 (2006) 439e444 441 Fig. 2. The structure (a) and photograph (b) of compliance chamber. transforms intermittent blood ejection of the heart pump into a continuous pulsatile flow. In practical apparatus, when the compliance chamber is not working, the minimum output pressure of the pump is zero or even negative. But when it is working, the pressure wave presents arterial flow with both the maximum and minimum pressure positive, thereby simulating systolic and diastolic pressure in the blood vessels. Adjusting the ratio of gas to liquid (gas/liquid) changes the amplitude of pulsatile flow. Compressing the gas loading valve and boosting air into the gas chamber, the gas/liquid increases and the amplitude of pulsatile flow decreases. Turning on the purge valve, the gas/liquid reduces and the amplitude of pulsatile flow augments. 3.2. After-load The after-load (43 mm long, 19 mm wide, 50 mm high) is composed of a valve body, a valve needle and an adjustment knob (Fig. 3). The cross-sectional size of the opening is changeable by adjusting the knob and moving the valve needle up and down, which consequently regulates the magnitude of the after-load. Regulating after-load can precisely control normal stress. 3.3. Specifications of the design The rectangular chamber in our test apparatus has an orifice of B24 mm  0.5 mm on which the cover-slip of Fig. 3. The structure (a) and photograph (b) of the after-load. 442 H. Ding et al. / Cell Biology International 30 (2006) 439e444 B24 mm  0.5 mm is located. The ECs are planted on the cover-slip (Fig. 4). The following specifications were applied in our design. Table 1 Relationship between the value of gas/liquid and P2/P1 Value of gas/ liquid Gas volume V0 Liquid volume P2/P1 (1) Systolic pressure is 15,996e23,994 Pa. (2) Diastolic pressure is 7998e15,996 Pa. (3) The pump’s flow rate must be in the range of 20e600 ml/ min. 28:1 4:1 2:1 1:1 1:2 1:4 1:28 241.4 200 167 125 83 50 8.6 8.6 50 83 125 167 200 241.4 0.964 0.957 0.948 0.931 0.896 0.828 0 As mentioned above, the shear stress is proportional to the flow rate and varies within a certain range in the human body. If we want to obtain physiological shear stresses of 2e300 dyn/cm3, then the flow rate of the pump should be 20e 600 ml/min according to the dimension of our designed apparatus. According to Eq. (3), the shear stress is 2 dyn/cm3 when m ¼ 11.68  104 Pa S, d ¼ 0.56 mm, b ¼ 30 mm and Q ¼ 20 ml/min; the shear stress is 300 dyn/cm3 when m ¼ 11.68  104 Pa S, d ¼ 0.56 mm, b ¼ 30 mm and Q ¼ 600 ml/min. (4) Pulsatile frequency should be 40e200 r/min. Pulsatile frequency reflects heart rate, and the normal heart rate of man is 75 beats per min. But in order to simulate the normal heart rate and the abnormal one we have to choose a larger measuring range in this apparatus, therefore the value of pulsatile frequency is 40e200 r/min. (5) In order to make the pulsatile amplitude adjustable within a stated range, the total compliance chamber volume V0 should be between 200 and 300 ml. Here, V0 ¼ gas volume þ liquid volume. The cavity of the compliance chamber is 250 ml, and the maximum flow rate of the designed apparatus is 600 ml/min. Suppose that the heart rate is 70 beats per minute and the heart output per pulse is be DV ¼ 8:6 ml, we can obtain the following data (Table 1) according to Eq. (1). Analyzing the data in Table 1, we can obtain the following conclusion: first, when the value of gas/liquid changed, P1/P2 changed correspondingly in the range of (1e0) that can be regulated according to the experiment requirements. Second, the greater the value of gas/liquid, the smaller the rigidity of compliance chamber, and the smaller the difference of P1 and P2. Fig. 4. The structure of rectangular chamber. Third, the smaller the value of gas/liquid, and thus the greater the rigidity of compliance chamber, the greater the difference of P1 and P2. 4. Results The working procedure of compliance chamber and afterload in the apparatus for cultured endothelial cells subjected to stresses designed in this paper is shown as follows. Not blood but nourishing fluids were employed in the apparatus to feed the ECs on the rectangular chamber. The liquid pumped by pulsatile pump is divided into two branches after the compliance chamber: one is connected to a throttle and then flows back to the reservoir, the other enters the rectangular chamber, and then flows back to the reservoir through the after-load (Fig. 1). This device can simulate the normal systolic and diastolic pressure of 15,996/10,664 Pa (120/80 mm Hg) (Fig. 5). It can adjust the amplitude of pulsatile flow by changing the value of gas/liquid in the compliance chamber (Fig. 6). It can also regulate the absolute value of blood pressure by adjusting the after-load (Fig. 7). Shunt volume can be regulated by changing the opening of the throttle to regulate the shear stress in the rectangular chamber. 5. Discussion Although the apparatus is able to accomplish our intention, it still has some drawbacks as follows: (1) The purpose of this apparatus is to create a hemodynamic environment that is similar to the physiological condition of arteries. In order to observe ECs’ changing states more conveniently, three-dimensional flow field in the arteries is simplified into one-dimensional flow field in the rectangular chamber (Fig. 4), but the simplification creates a difference between the realistic condition and the experimental one. (2) The ECs are planted on an inflexible cover-slip, which is different to the elastic foundation base of arteries. (3) The shear stress is obtained indirectly from the calculation of pressure difference, and there is no direct and simple measurement at the present time. (4) The after-load mentioned by this paper is a ‘‘lump resistance model’’, which simulates the resistance of capillary vessels to blood flow. However, the ‘‘lump resistance model’’ has no effect when studying the behavior of capillary vessels. H. Ding et al. / Cell Biology International 30 (2006) 439e444 443 Fig. 5. Simulating the normal systolic pressure and diastolic pressure (15,996/10,664 Pa or 120/80 mm Hg). 6. Conclusion Since the shear stress acting on vascular endothelium is inaccessible in vivo, we designed an in vitro experiment apparatus in which the compliance chamber and the after-load are two key modules in order to simulate the hemodynamic environment for cultured ECs subjected to both normal stress and shear stress. By adjusting these two modules, we can change the normal stress and the shear stress to which ECs are exposed. This apparatus is supplied as a trial production to Shanghai Cerebrovascular Disease Prevention Research Institute and Life Sciences College of Tongji University. Their experimental findings were inspiring and confirmed the practicability of the designed apparatus. Therefore, with this apparatus, more realistic hemodynamic conditions can be obtained for in vitro experiments using ECs under physiological conditions, thus facilitating the research of ECs subjected to stresses. Fig. 6. Regulating the amplitude of pulsatile flow. 444 H. Ding et al. / Cell Biology International 30 (2006) 439e444 Fig. 7. Regulating the absolute value of blood pressure. Acknowledgments For significant contribution to this work, the authors wish to thank Prof. Shixiong Xu from Fudan University for his support and assistance. References Brooks AR, Lelkes PI, Rubanyi GM. Gene expression profiling of human aortic endothelial cells exposed to disturbed flow and steady laminar flow. Physiol Genomic 2002;9(1):27e41. Chen H, Wu L, Liu X, Chen Y, Wang B. Effects of laminar shear stress on IL-8 mRNA expression in endothelial cells. Biorheology 2003;40(1e3):53e8. Dewey Jr CF, Bussolari SR, Gimbrone Jr MA. The dynamic response of vascular endothelial cells to fluid shear stress. J Biomech Eng 1981;103:177e81. Diao Y. Principles and application of biomechanics. Shanghai: Tongji University Press; 1991. p. 28e31. Eskin SG, Ives CL, Mcintire LV. Response of cultured endothelial cells to steady flow. Microvasc Res 1984;28:87e94. Franke RP, Grafe M, Schnittler H. Induction of human vascular endothelial stress fibers by shear stress. Nature 1984;307:648e9. Greisler HP, Johnson S, Joyce K, Henderson S, Patel NM, Alkhamis T, et al. The effects of shear stress on endothelial cell retention and function on expanded polytetrafluoroethylene. Arch Surg 1990;125(11): 1622e5. Helmlinger G, Geiger RV, Schreck S. Effects of pulsatile flow on cultured vascular endothelial cell morphology. J Biomech Eng 1991;113:123e31. Liu Q, Cai S, Wu Y. A new calculating method for the dynamic wall shear stress of experimental vessel segments. Acta Biophys Sin 1997;13(3): 507e11. Liu Z, He F, Xu G, Chen Y. The determination of artery wall’s shear stress. Sci China Ser A 2001;31(7):651e60. Liu Z, Li X. The principle and method of hemodynamics. Shanghai, China: Fudan University Press; 1997. p. 46e100. Frame MDS, Chapman Gary B, Markino Yoshia, Sarelius Ingrid H. Shear stress gradient over endothelial cells in a curved microchannel system. Biorheology 1998;35(4):245e61. Mu Y, Du G. Advance in the culture methods of endothelial cell in vitro. Foreign Med Sci Sect Pharm 2004;31(6):361e5. Passerini AG, Milsted A, Rittgers SE. Shear stress magnitude and directionality modulate growth factor gene expression in preconditioned vascular endothelial cells. J Vasc Surg 2003;37(1):182e90. Zhang H, Huang Y. Hydraulic transmission. Beijing: Chinese Machine Press; 1991. p. 103e106.