CONTROVERSIES IN
CARDIOVASCULAR MEDICINE
Should Transcatheter Aortic Valve
Replacement be Expanded to
Lower-Risk and Younger Patients?
Transcatheter Aortic Valve Replacement Indications Should
Not Be Expanded to Lower-Risk and Younger Patients
Volkmar Falk, MD
S
ince its introduction, transcatheter aortic valve replacement
(TAVR) has evolved rapidly from an experimental to a routine intervention. Randomized trials clearly demonstrated that
TAVR offers a survival benefit in inoperable patients1 and has
early mortality and midterm survival outcomes equal to those of
standard surgical aortic valve replacement (SAVR) in a high-risk
population.2,3 These excellent results have led to guideline recommendations in the United States and Europe promoting the use
of TAVR in populations of inoperable and high-risk patients provided that the decision is based on a heart team discussion.4 With
improvements in valve design, the development of smaller delivery systems, and the introduction of improved mechanisms for
steering, stable positioning, and recapturing, the use of TAVR has
expanded continuously. Nationwide registries5–7 have contributed
to large data sets confirming acceptable results outside clinical
trials. Parallel to the publication of registry data and large center
series, a plethora of case reports have documented the growing
off-label use of TAVR for the treatment of degenerated bioprosthesis (valve-in-valve implantation) and aortic insufficiency and in
lower-risk patient subsets who until now have not been addressed
by randomized trials. As TAVR has become a standardized procedure, many operators have become proficient in performing it, and
patient recruitment has become more challenging, particularly in
high-volume centers. The natural evolution seems to be an expansion of the indication for TAVR to lower-risk patients.
Response by Haussig and Linke on p 2342
The European guidelines on valvular heart disease4 clearly
state that “at the present stage, TAVR should not be performed
in patients at intermediate risk for surgery and trials are required
in this population.” Despite the absence of sufficient data to
support TAVR in low-risk patients, aggressive direct patient
marketing, self-referral bias, and ignorance of the heart team
concept have slowly led to an erosion of these guideline recommendations. In some countries, TAVR is already routinely
applied in lower-risk and younger patient groups8 (Figure 1).
The use of TAVR shows a large variation across nations in
Europe. The number of TAVR implants per million ranged in
2011 from 6.1 in Portugal to 88.7 in Germany. Germany (36.2%)
and Switzerland (34.5%) had the highest and Portugal (3.4%) and
Spain (8.4%) had the lowest penetration rates. The observed correlation between TAVR use and healthcare spending per capita
(r=0.80, P=0.005)10 indicates that reimbursement rather than medical indication is a major driver for the expanded use of TAVR. Not
surprisingly, TAVR-specific reimbursement systems were associated with higher TAVR use than restricted systems (698±232 versus 213±112 implants per 1 million individuals ≥75 years of age;
P=0.002).10 A percutaneous therapy, particularly when delivered
under local anesthesia, will be preferred by patients over a surgical procedure involving cardiopulmonary bypass and some sort
of a thoracotomy or sternotomy, even if the risk-benefit ratio and
long-term outcomes are largely unknown. If patients are advised
to or choose to undergo TAVR outside the current guideline indications, they should be made aware of the following.
The opinions expressed in this article are not necessarily those of the editors or of the American Heart Association.
From the University of Zürich, Zürich, Switzerland.
This article is Part II of a 2-part article. Part I appears on p 2321.
Correspondence to Volkmar Falk, MD, Department of Cardiothoracic and Vascular Surgery, German Heartcenter Berlin, Augustenburger Platz 1,13353
Berlin, Germany. E-mail falk@dhzb.de
(Circulation. 2014;130:2332-2342)
© 2014 American Heart Association, Inc.
Circulation is available at http://circ.ahajournals.org
DOI: 10.1161/CIRCULATIONAHA.114.008145
Downloaded from http://circ.ahajournals.org/
by guest on March 30, 2016
2332
Falk
TAVR in Lower-Risk and Younger Patients
2333
Figure 1. Development of surgical (SAVR) and transcatheter (TAVR) aortic valve replacement in Germany from 2000 to 2012 according to
the annual statistics of the German Association of Cardiothoracic and Vascular Surgery (not all nonsurgical centers are included; annual
TAVR numbers exceed those reported in this database). Initially, TAVR opened a new segment on top of the volume of SAVR. Along with
the exponential use of TAVR beginning in 2009, a decline in annual SAVR volume is observed (A). The curve follows the pattern of a typical innovation S curve (B). Graph in A is derived from Beckmann et al.9
About SAVR
SAVR is a procedure with a history of >50 years of continuous improvements. It is safe and can be performed through
a limited thoracotomy or ministernotomy with excellent outcomes and proven long-term durability. Morbidity and mortality rates are low, and hemodynamic performance is good.
Additional procedures, including root replacement, aortic surgery, left ventricular outflow tract resection, other valve surgery, or revascularization, can be performed at the same time.
In a review of some 108 687 patients who had an isolated SAVR
and were included in the Society of Thoracic Surgeons database
from 1997 to 2006, the overall mortality and risk-adjusted mortality decreased by 24% and 33%, respectively. Mortality was
0.6% in patients <60 years old, <1.3% in patients <70 years old,
<3.5% in patients <80 years old, and <5% in patients <85 years
old.11 These data have been confirmed by numerous national
databases. In a recent meta-analysis including 48 studies of
SAVR in octogenarians, the rates for mortality were 6.7% (in
the period from 2000–2006, 5.8%); stroke, 2.4%; dialysis, 2.6%;
and pacemaker implantation, 4.6%. Survival was 88% at 1 year,
79% at 3 years, 65% at 5 years, and 30% at 10 years of followup.12 In a contemporary study of high-risk patients (mean age,
80.4±3.6 years; mean logistic EuroSCORE, 13%±7%), the hospital mortality after SAVR was 1.3%.12 Quality of life is good
after SAVR, and the procedure is cost-effective.
significantly different between groups (relative risk [RR] 1.00;
95% confidence interval [CI], 0.8–1.23; P=0.97). The subgroup
analysis of randomized, controlled trials also did not show any
significant difference in 30-day all-cause mortality between
the TAVR and SAVR groups (RR, 1.53; 95% CI, 0.99–2.36;
P=0.06). Midterm all-cause mortality was also not significantly
different between the TAVR and SAVR groups (RR, 0.78; 95%
CI, 0.59–1.02; P=0.07) at 85 weeks of mean follow-up.14
Because TAVR does not improve survival even in the
group of patients considered at the highest risk for surgery, it
is unlikely that it will improve survival in younger, low-risk
patients. Some rare but potentially lethal complications such
as coronary occlusion, annular rupture, ventricular perforation, and vascular injury are procedure related. These complications are largely independent of age and the underlying risk
and can therefore be expected to contribute to the procedurerelated mortality in all patient subsets.
There are in fact indications for excess mortality with TAVR
in a low-risk group of patients. Results from the German
Aortic Valve Registry (GARY) show no difference for mortality in high-risk stratified groups with a euroSCORE >20.
In lower-risk groups, however, SAVR shows a significant survival benefit in both groups with a logistic euroSCORE of 10
to 20 (P<0.001) and <10 (P<0.001).15
Stroke Remains an Issue With TAVR
TAVR Does Not Improve Survival
in Operable High-Risk Patients
A meta-analysis including 4873 patients from 17 studies (randomized, clinical trials and adjusted observational comparative studies) showed that TAVR does not reduce early (at 30
days or in hospital) or midterm (at 3 months–3 years) all-cause
mortality compared with SAVR in high-risk patients with aortic
stenosis.13 In another meta-analysis comparing 4659 patients
with severe aortic stenosis who underwent TAVR (n=2267)
and SAVR (n=2392), 30-day all-cause mortality was not
Using diffusion-weighted magnetic resonance imaging, new
ischemic lesions can be detected in up to 90% of patients after
TAVR, a rate that is significantly higher than in patients undergoing SAVR.16 The periprocedural incidence of cerebrovascular events after TAVR therefore exceeds that after any other
cardiac intervention or valve surgery.17
Randomized trials in high-risk populations have consistently shown an increased stroke rate with TAVR compared
with SAVR.3,18,19 A pooled meta-analysis of these randomized, clinical trials also demonstrated a significantly higher
Downloaded from http://circ.ahajournals.org/ by guest on March 30, 2016
2334
Circulation
December 23/30, 2014
incidence of stroke with TAVR compared with SAVR (RR,
0.56; 95% CI, 0.36–0.88; P=0.01) at a mean follow-up of
99 weeks.13 Despite some recent reports that have shown a
reduction in stroke rate associated with TAVR,6 the French
Aortic National CoreValve and Edwards (FRANCE) II and
Australian registries consistently demonstrate a 30-day stroke
rate of 3.4% and 3.9% in the current era.5,20 It is worth mentioning that periprocedural stroke has a dismal prognosis with
an up to >3.5-times increase in 30-day mortality.20,21
Passage of guide wires and large-bore catheters during valvuloplasty and device delivery, positioning, and implantation may
lead to the disruption of atheromatous plaques of valvular and
vascular origin during TAVR.16 Most of the events contributing to the embolic load during TAVR are related to relatively
uncontrolled and remotely induced manipulation of valvular or
vascular structures. Retrograde passage of the aorta and the aortic arch and dilatation and displacement of the native, calcified
aortic valve may cause dislodgement of calcific debris independently of age and the underlying risk. Hemodynamic instability
during or after rapid ventricular pacing can impair cerebral perfusion pressure, which may also contribute to the risk of stroke.
In addition, subacute embolization may be caused by persistent lesions on the displaced (not resected) native valve
leaflets. A “dead” space between the implanted prosthesis and
the native leaflets or between the leaflets and the aortic wall
may be prone to thrombus formation. As in SAVR, new onset
of atrial fibrillation occurs frequently after TAVR and poses an
additional stroke risk in the postprocedural phase.
More recently, a number of embolic protection devices have
been introduced to reduce the risk of stroke during TAVR.
Although these devices may effectively capture some debris,
so far, none of them has proven to effectively reduce the risk
of stroke.
Given the low risk of stroke in all-comers after isolated SAVR
(0.7% in patients <70 years of age, <2.0% in patients 70 to 80
years of age, and <2.5% in the octogenarian population),11 the
question of risk versus benefit must be raised before the indication of TAVR is expanded to younger, low-risk populations.
TAVR Leaks
Paravalvular aortic leakage (PVL) after SAVR is an infrequent
event and occurs in <1% of patients.22 This was also confirmed
in the surgical arm of the Placement of Aortic Transcatheter
Valve Trial (PARTNER), with an incidence of PVL after
SAVR of 0.9% after 2 years.3 On the contrary, PVL is a frequent complication after TAVR. The GARY registry reported
the outcomes of 2763 patients undergoing transfemoral TAVR
in 2011; 55.5% of patients had grade I and 7.0% grade II or
higher aortic insufficiency as a result of PVL.6 Similar numbers have been reported from the PARTNER trial (7.0% at 1
year and 6.9% at 2 years).3 TAVR patients had significantly
more total aortic regurgitation at every postimplantation time
than patients in the SAVR cohort. Mild, moderate, or severe
paravalvular aortic regurgitation was more common in the
TAVR group at every follow-up time (P<0.0001).
In a meta-analysis, the incidence of postoperative moderate
or severe aortic regurgitation, which included both paravalvular and transvalvular regurgitation, was significantly higher
after TAVR than SAVR (7.8% versus 0.6%; RR, 6.82; 95% CI,
3.57–13.04; P<0.00001).23
After TAVR, patients with mild or moderate paravalvular
aortic regurgitation have larger postimplantation left ventricular end diastolic dimension (P=0.009), left ventricular diastolic
volume (P=0.003), left ventricular mass (P=0.001), and left
ventricular mass index (P=0.007), indicating the absence of
reverse remodeling in the presence of PVL. Of all echocardiographic parameters after valve implantation, mild or moderate
paravalvular aortic regurgitation was the strongest predictor of
death (hazard ratio [HR], 2.11; 95% CI, 1.43–3.10; P=0.0002).24
In the Italian (CoreValve) registry, postprocedural paravalvular leak of >2+ was an independent predictor of mortality
beyond 30 days and up to 1year of follow-up (HR, 3.79).7 In
another report, 15% of patients with a CoreValve had moderate
or severe paravalvular AR, which again was an independent predictor of 1-year mortality.25 In the GARY registry, the mortality
of patients with severe aortic insufficiency was 37.5% at 30 days
and 50.0% at 1 year.14 In the FRANCE II registry including 3195
consecutive patients, aortic regurgitation of grade 2 or higher was
observed in 15.8% of patients undergoing TAVR and was found
to be a strong independent predictor of 1-year mortality for both
balloon-expandable (HR,=2.50; P=0.0001) and self-expandable
(HR, 2.11; P=0.0001) valves.25 Patients with a nontransfemoral
approach had a much lower risk of postprocedural aortic regurgitation than those treated through a femoral approach. This benefit was observed both for balloon-expandable (RR reduction,
30%) and for self-expanding (RR reduction, 55%) devices and
was confirmed by multivariable analysis.26
An interesting finding of the FRANCE II registry was that
the use of a self- expanding valve was an independent predictor of postprocedural aortic regurgitation (self-expanding
valve, 21.5%; balloon expandable valve, 13.0%). These data
are in line with findings from other studies.27 The difference
was more striking for the transfemoral approach (21.9% versus 13.9%) than for the nonfemoral approach (10.7% versus
8%). The authors suggest that, with a transfemoral approach,
control of the catheter and implantation depth is more difficult
with an increased risk of PVL.26
In TAVR, the presence of bulky native aortic valve calcifications leading to incomplete prosthesis apposition, annular
eccentricity, inappropriate size selection (undersizing), and
problems with positioning (high or low malpositioning) has
been identified as an independent predictor of PVL in patients
undergoing TAVR25,26,28,29 (Figure 2).
Appropriate sizing with multidetector computed tomography
and template-based planning may help to may prevent inappropriate valve selection.28,30 New-generation devices with supraannular, infra-annular, or intra-annular cuffs for sealing and
delivery systems that allow better control of device deployment,
easy repositioning, or even removal of the device may also
reduce PVL in the future. Eccentric annuli, bulky calcifications,
Downloaded from http://circ.ahajournals.org/ by guest on March 30, 2016
Falk
TAVR in Lower-Risk and Younger Patients
2335
Figure 2. Perivalvular leak can be caused by inappropriate size (undersizing; A), high (B) or low (C) valve malpositioning, or eccentric or
bulky annular or leaflet calcification (D). Data derived from Sinning et al.25
bicuspid anatomy, variations in annular compliance, and deformation of stents (Figure 3), all of which are associated with
PVL, will, however, pose a continuous challenge for TAVR and
are independent of age and the underlying risk.
Moderate or severe paravalvular regurgitation after AVR
cannot be regarded a benign disease and is therefore unacceptable in a younger, low-risk population.
Pacing Is a Risk
Left bundle-branch block (LBBB) frequently occurs after
TAVR. LBBB results in intraventricular asynchrony and may
have a causative role in the development of cardiac remodeling and heart failure.31 The high incidence of atrioventricular
block and LBBB after TAVR, especially after the implantation of self-expanding valves, has led to a remarkable need for
pacemaker implantation after TAVR. The deleterious effect
of new-onset, persistent LBBB on left ventricular function
after TAVR has been shown in a number of studies. Patients
with persistent LBBB showed either a lack of improvement or even a decrease in left ventricular ejection fraction
compared with patients with no new conduction abnormalities.32–34 New onset of LBBB was reported with an incidence
Figure 3. Stent deformation occurs in >90% of valves and may impair long-term valve function. Example of a deformed CoreValve stent
in postoperative computed tomographic and 3-dimensional reconstruction with color coding (orange) of areas exposed to high stress
(lower right). ACS indicates non-coronary sinus; LCS, left coronary sinus; and RCS, right coronary sinus.
Downloaded from http://circ.ahajournals.org/ by guest on March 30, 2016
2336
Circulation
December 23/30, 2014
of 19.2% after TAVR with balloon-expandable devices and
was persistent in 11.8% after discharge. Persistent LBBB was
associated with a higher rate of pacemaker implantation, a
lack of improvement in left ventricular ejection fraction, and
a poorer functional status.35 In another series of 476 patients
undergoing TAVR (including balloon-expandable and selfexpanding valves), 175 (36.8%) developed new LBBB,
85.7% within 24 hours after TAVR, 12.0% before and 2.3%
after hospital discharge. LBBB was persistent in 111 patients
(63.4%). Implantation of the Medtronic CoreValve System
more frequently led to new LBBB than implantation of the
balloon-expandable Edwards SAPIEN valve (53.8% versus
21.7%) with less recovery during follow-up (39.0% versus
9.5%).36 In the Houthuizen et al36 study, which had a much
longer follow-up than previous trials (median follow-up, 915
days [578–1234 days]), persistent LBBB was also associated
with a significant increase in mortality compared with the
absence of LBBB and temporary LBBB combined (HR, 1.49;
95% CI, 1.10–2.03; P=0.01).36 In the GARY registry, the
rate of new pacemaker implantation after SAVR was 4.6%
as opposed to 23.6% after transfemoral TAVR.6 In a metaanalysis including 7 observational and 2 randomized trials,
the rate of pacemaker implantation was significantly higher
after TAVR than SAVR (13.2% versus 3.0%; RR, 3.53; 95%
CI, 1.79–6.97; P=0.0003).23
Chronic right ventricular pacing is known to be associated
with left atrial remodeling,37 reduced atrial function,38 and left
ventricular dyssynchrony,39 resulting in impaired left ventricular function, limited exercise capacity, and progressive
left ventricular remodeling, finally leading to heart failure.39
Permanent right ventricular pacing may cause or worsen heart
failure, especially in patients with impaired left ventricular
function.40–42 Patients with chronic right ventricular pacing are
also at higher risk for development of atrial fibrillation.40
At 6 months after TAVR, patients requiring pacemaker
implantation demonstrated attenuated improvement in left
ventricular ejection fraction (−0.9±8.7% versus 2.3±10.8%,
respectively; P=0.03) and left ventricular stroke volume
(−2±16 versus 4±10 mL/m2, respectively; P=0.015), a trend
toward a smaller reduction in systolic pulmonary artery pressure (−1±12 versus −6±10 mm Hg, respectively; P=0.09), and
deterioration of the right ventricular index of myocardial performance (−3±17% versus 5±26%, respectively; P=0.05).43
Besides the negative hemodynamic effects of right ventricular pacing, other complications are associated with pacemaker implantation. In a population-based study including
46 299 consecutive pacemaker patients with long-term follow-up, the incidence of infection and removal of the device
after first pacemaker implantation was 1.82 in 1000 pacemaker-years and after replacement procedures was 5.32 in
1000 pacemaker-years. Surgical site infection after first pacemaker implantation occurred at an incidence rate of 4.82 per
1000 pacemaker-years after first pacemaker implantation and
12.12 per 1000 pacemaker-years after replacement.44 Leadrelated reinterventions can be expected at a rate of 2.3% after
dual-chamber pacemaker procedures.45 Other complications
such as tricuspid regurgitation and the need for subsequent
device and battery changes have to be factored in.
If the indication for transcatheter aortic valve implantation
is to be expanded to a younger population at less risk, a substantial number of patients will be subjected to new onset of
persistent LBBB, which may impair left ventricular function,
and to decades of pacing with all the associated consequences.
Bioprosthetic Valves in Younger Patients?
Most risk scores (euroSCORE, Society of Thoracic Surgeons)
weigh age heavily. Expanding indications to lower-risk
groups therefore automatically implies expanding indications
to younger patients. This brings the question of valve selection (mechanical versus biological) into focus. Although for
decades surgeons have promoted and expanded the use of biological valves in younger patients, recent evidence suggests an
increased mortality in younger patients receiving bioprostheses.
An analysis of data from 24 410 patients with biological
AVR and 14 789 patients with mechanical SAVR 65 to 80
years of age included in the Society of Thoracic Surgeons
Adult Cardiac Surgery Database from 1991 to 1999 compared the long-term outcomes of patients with bioprosthetic
versus mechanical AVR. In the youngest patient group
(65–69 years of age), patients receiving a bioprosthesis had
a substantially elevated 12-year absolute risk of reoperation
(10.5%) and a 23% increased mortality rate (adjusted HR,
1.23; 95%, CI, 1.16–1.31).46
This finding was confirmed by 2 propensity-matched studies in patients undergoing bioprosthetic or mechanical AVR.
In a study of 206 patients ≤60 of age,47 biological AVR was
associated with reduced midterm survival despite similar
valve-related event rates in both groups.
In another study on 440 patients 50 to 70 years of age,11
multivariate analysis demonstrated that mechanical valves
were protective for late survival (HR, 0.48; 95% CI, 0.35–
0.67; P<0.01).
The freedom from structural valve deterioration of surgically implanted bioprosthetic valves is dependent of the age
at implantation. At >70 years of age, a 15-year freedom from
structural valve deterioration of >90% can be expected. In
patients ≤60 years of age, only 63% of the valves are free
of structural deterioration at 15 years.48 There is little reason
to believe that TAVR will change this pattern. The need for
crimping, balloon dilatation, stent deformation, and overexpansion or underexpansion, which may lead to unfavorable
shear stress distribution of the leaflets, may cause early valve
deterioration after TAVR. Most data on TAVR outcomes have
been obtained in elderly, high-risk, or even inoperable patient
groups with high early (10% at 30 days) and 1-year (20%–
40%) mortalities and limited follow-up. The durability of the
implanted valves, especially in a younger population, is therefore largely unknown.
Bleeding and thrombotic complications in patients under
anticoagulant treatment have been the major drawback of
Downloaded from http://circ.ahajournals.org/ by guest on March 30, 2016
Falk
mechanical valves. A decrease in anticoagulant-related complications of mechanical valves can be expected with the more
widespread use of home international normalized ratio monitoring systems that allow lower levels of anticoagulation (target international normalized ratio, 1.5–2.0) in patients with
mechanical aortic valves.49
In the debate to expand the indications of TAVR to a
younger population, the possible survival benefit of mechanical valves needs to be discussed with the patient.
Valve in Valve in Valve…
A frequent assumption in the TAVR age is that degenerative
valve disease will be routinely treated by a single or even multiple valve-in-valve implantations in the future. This argument,
which as yet is not based on any solid data, is brought forward
to promote the indication for bioprosthetic valve implantation
(SAVR or TAVR) in younger patients.
Several procedural adverse events and complications are
more common after valve-in-valve implantation, including
device malpositioning (15%), coronary obstruction (3.5%),
and elevated postprocedural gradients (28% >20 mm Hg).
Severe patient-prosthesis mismatch can be found in up to 32%
of patients after a valve-in-valve procedure.50,51
In an analysis based on data from the PARTNER trial,
immediate insertion of a second transcatheter prosthetic valve
within the first after TAVR was an independent predictor of
1-year cardiovascular mortality (HR, 1.86; 95% CI, 1.03–3.38;
P=0.041), with a nonsignificant trend toward greater all-cause
mortality (HR, 1.43; 95% CI, 0.88–2.33; P=0.15).52
Very few data exist on the valve-in-valve concept for
treating patients long term after TAVR. Given the results for
valve-in-valve after SAVR and immediately after TAVR, the
enthusiasm for this therapy in the setting of failing TAVR
does not seem justified. Although the valve-in-valve concept
is an interesting option for treating selected patients with
degenerated bioprosthetic valves, its mere technical feasibility does not justify the use of TAVR in a younger population. Even when, and if, multiple valve-in-valve procedures
become a realistic option, such a strategy must be compared
with a single-stage surgical approach in a trial setting before
the indications can be expanded.
Other Complications
In addition to major adverse events, paravalvular leak, and
LBBB, a number of procedure-related complications have
been reported in the context of TAVR. Among others, device
embolization, access complications, annular or ventricular
rupture, and coronary obstruction or occlusion occur.
Procedure-related vascular complications occur frequently
after TAVR. In a meta-analysis including 16 studies, the rate of
major vascular complications was 11.9%.29 In the transfemoral
TAVR group of the GARY registry, the rate of vascular complications (15.9%) was in the same range.6 The incidence of vascular complications is significantly higher in patients who undergo
TAVR compared with those who undergo SAVR (13.8% versus
TAVR in Lower-Risk and Younger Patients
2337
2.0%; RR, 5.65; 95% CI, 3.36–9.50; P<0.00001).23 Vascular
complications are associated with increased mortality.27
Coronary obstruction during TAVR occurs at a rate of 0.6%
to 1% and remains a life-threatening complication (30-day
mortality, 40%).53
Annular (aortic root) rupture is rare, but the combined rate
of annular or left ventricular outflow tract rupture is reported
to be 1.1%.27 Left ventricular outflow tract calcification and
aggressive annular area oversizing are associated with an
increased risk of aortic root rupture during TAVR with balloon-expandable prostheses.54 Aortic root rupture is associated with a 48% to 100% mortality.21,54
Although overly aggressive oversizing can be avoided, the
rate of annular rupture has not changed over the years and
remains a concern, particularly when TAVR is offered to lowrisk patients.
Emergency cardiopulmonary bypass is required in ≈1.1%
to 4.9% of patients undergoing TAVR, usually as a result of
annular or ventricular rupture, aortic dissection, coronary
occlusion, or device embolization.21,55 Emergency cardiopulmonary bypass is associated with a mortality rate of 35%
to 50% (Table).21,55
Money Matters
Although TAVR is cost-effective in inoperable patients, conflicting results exist with regard to its cost-effectiveness compared
with surgery in high-risk operable patients.56 For high-risk operable patients included in the PARTNER trial, the incremental cost-effectiveness ratios were well beyond the generally
accepted maximum threshold.57 The currently available data do
not conclusively demonstrate that TAVR is incrementally costeffective compared with SAVR for high-risk surgical candidates
because of the wide variability of incremental cost-effectiveness
ratio values and probabilities of economic acceptability reported
in different countries and reimbursement systems.58 Six of 9
analyses reported an incremental cost-effectiveness ratio value
that would have been deemed unacceptable with the use of a
US $100 000 per quality-adjusted life year gained threshold, and
7 would have been deemed unacceptable with the use of a US
$50 000 per quality-adjusted life year gained threshold. In the
United States, the total costs of TAVR and SAVR were generally similar because the higher device- and procedure-related
expenses for TAVR were offset by the comparatively greater
costs of hospitalization after SAVR.58 If cost-effectiveness in a
(surgical) high-risk population cannot be demonstrated, it can
hardly be expected in a low -risk group of patients. Low-risk
patients undergoing SVR have a short (if any) stay in the intensive care unit and can usually be discharged shortly after surgery,
effectively decreasing the costs for hospitalization. The costs for
implants (for both SAVR and TAVR) are independent and unrelated to the underlying patient risk.
If TAVR is performed in a low-risk population, the expected
long-term survival implies the need for multiple reinterventions (because of valve degeneration), thereby introducing
another health economy challenge.
Downloaded from http://circ.ahajournals.org/ by guest on March 30, 2016
2338
Table.
Circulation
December 23/30, 2014
Outcomes in High-Risk Operable Patients
SAVR
Mortality at 30 d
TAVR
Impact on
Long-Term
Survival
Discussion
Impact on
Long-Term
LV Function
Same
Same
Yes
Same/less
Same/higher
Yes
No
Same
Same
Yes
Yes
Risk of AI >II
Minimal
High (7%)
Yes
Yes
Persistent LBBB,
need for pacemaker
Minimal
High
(15%–20%)
Likely
Yes
LOS
Same
or longer
Same
or less
No
No
Risk of perioperative
bleeding
Higher
Lower
Likely
No
Durability
Known
Unknown
Likely
Stroke rate
MI
Patient informed consent should include procedure-related risks and their
impact on long-term outcomes and raise awareness for uncertainties such as
durability. AI indicates aortic insufficiency; LBBB, left bundle-branch block; LOS,
length of stay; LV, left ventricular; MI, myocardial infarction; SAVR, surgical
aortic valve replacement; and TAVR, transcatheter aortic valve replacement.
Principal Shortcomings of Outcome Reporting
Shortcomings of outcome reporting are not unique to but also
observed in the TAVR literature. As pointed out in a metaanalysis,22 a substantial number of studies comparing TAVR
and SAVR did not report outcomes based on an intentionto-treat analysis. Some excluded patients with unsuccessful
implantations or periprocedural death, whereas others did
not analyze crossovers from TAVR to SAVR. These reporting problems apply also to some of the national registries.
Because the rate of missing data or crossovers after unsuccessful TAVR is not always known, some uncertainty about
the reported outcomes remains.
The invention of TAVR is an example of disruptive technology or innovation, a term originally coined by Bower
and Christensen.59 The concept describes a new technology (TAVR) that unexpectedly displaces an established one
(SAVR). Sustaining technology (SAVR) relies on incremental
improvements of an existing product (eg, the introduction of a
new anticalcification treatment to improve the long-term performance of a bioprosthesis) and usually does not create new
markets (the market for surgical treatment of aortic stenosis
is not affected by the new anticalcification treatment). On the
contrary, a disruptive technology (or innovation) and the resulting business model can either create new markets (treatment
of inoperable aortic stenosis) or have a disruptive impact on
existing markets (Figures 1 and 4). The Bower and Christensen
definition of disruptive technology reads like a blueprint for
Anderson’s innovation that laid the foundation for TAVR:
Disruptive innovations are technologically straightforward, consisting of off-the-shelf components put
together in a product architecture [TAVR] that is often
simpler than prior approaches [SAVR]. They offer less
of what customers in established markets want and so
can rarely be initially used there. They offer a different
package of attributes [no cardiopulmonary bypass,
percutaneous access] valued only in emerging markets remote from, and unimportant to, the mainstream
[standard SVR in low-risk patients].60
According to the EuroHeart survey and other series, 30%
of patients with severe aortic stenosis were refused treatment
until TAVR was introduced into clinical practice. After opening the market of inoperable patients, the procedure is currently at the stage of “low-end” disruption. The performance
Figure 4. A disruptive technology (innovation) such as transcatheter aortic valve replacement (TAVR) initially targets “low-end” customers who do not need full product performance (very high-risk elderly patients) or target customers who have needs that were previously
unserved by existing technologies. The latter applies to patients deemed inoperable by the existing “sustainable” technology (surgical
aortic valve replacement [SAVR]) and represents a new market disruption. TAVR is currently at the stage of low-end disruption (*), and the
performance of TAVR (the product) starts to exceed the needs of the lower-end customer segments. At this point, technology and riskbenefit assessments are crucial to direct further market penetration. Figure derived from Christensen.60
Downloaded from http://circ.ahajournals.org/ by guest on March 30, 2016
Falk
of TAVR (the product) starts to exceed the needs of the lowerend customer segments. At this point, technology and riskbenefit assessments are absolutely crucial to direct further
market penetration (Figure 4).
TAVR has proven to be a safe and efficient treatment for inoperable patients with aortic stenosis, and there is equipoise with
regard to early and midterm mortality for operable high-risk
patients. Before the indications of this technology are extended
to a lower-risk group and younger patients, the net risk-benefit
ratio needs to be evaluated. As it is the case with most disruptive
technologies, a lack of refinement and performance problems
exist also with TAVR. These problems materialize in a stroke
rate in excess of surgery, a high rate of paravalvular leaks, a
high rate of persistent LBBB after the procedure, vascular complications, and uncertain durability, all of which may impair the
long-term outcomes of this promising technology.
Patient preference, often triggered by direct patient marketing and misleading advertisements by both companies
and healthcare providers, has already generated a demand for
TAVR outside the high-risk population. Off-label use, selfreferral bias, and economic incentives continue to drive TAVR
out of the corridor of current guideline recommendations and
underline the need for patient informed consent and heart team
discussions as a basis for medical decision making. The potential bias that self-referral can introduce into decision making
has long been recognized in the field of cardiology61 and led
to the current recommendations for interdisciplinary heart
teams.4 These teams can provide the appropriate framework for
conducting clinical trials in lower-risk populations. These trials
are necessary and timely and are currently recruiting.
Severe aortic stenosis has a pooled prevalence of 3.4% in
the elderly. Under the current indications, ≈290 000 elderly
patients with severe aortic stenosis are TAVR candidates, and
nearly 27 000 patients become eligible for TAVR annually.62
These patients already qualify for TAVR and should preferably
be served first before the indication is expanded to other patient
populations without sound scientific proof of superiority.
Disclosures
Dr Falk reports receiving research grants from Boston Scientific and
Phillips, receiving honoraria from Edwards and Aesculap, and serving as a consultant for Edwards and Medtronic.
References
1. Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG,
Tuzcu EM, Webb JG, Fontana GP, Makkar RR, Brown DL, Block
PC, Guyton RA, Pichard AD, Bavaria JE, Herrmann HC, Douglas PS,
Petersen JL, Akin JJ, Anderson WN, Wang D, Pocock S; PARTNER Trial
Investigators. Transcatheter aortic-valve implantation for aortic stenosis in
patients who cannot undergo surgery: the PARTNER trial investigators. N
Engl J Med. 2010;363:1597–1607.
2. Smith CR, Leon MB, Mack MJ, Miller DC, Moses JW, Svensson LG,
Tuzcu EM, Webb JG, Fontana GP, Makkar RR, Williams M, Dewey T,
Kapadia S, Babaliaros V, Thourani VH, Corso P, Pichard AD, Bavaria JE,
Herrmann HC, Akin JJ, Anderson WN, Wang D, Pocock SJ; PARTNER
Trial Investigators. Transcatheter versus surgical aortic-valve replacement
in high-risk patients: the PARTNER Trial Investigators. N Engl J Med.
2011;364:2187–2198.
TAVR in Lower-Risk and Younger Patients
2339
3. Kodali SK, Williams MR, Smith CR, Svensson LG, Webb JG, Makkar
RR, Fontana GP, Dewey TM, Thourani VH, Pichard AD, Fischbein M,
Szeto WY, Lim S, Greason KL, Teirstein PS, Malaisrie SC, Douglas PS,
Hahn RT, Whisenant B, Zajarias A, Wang D, Akin JJ, Anderson WN, Leon
MB; PARTNER Trial Investigators. Two-year outcomes after transcatheter or surgical aortic-valve replacement: the PARTNER Trial Investigators.
N Engl J Med. 2012;366:1686–1695.
4. Vahanian A, Alfieri O, Andreotti F, Antunes MJ, Barón-Esquivias G,
Baumgartner H, Borger MA, Carrel TP, De Bonis M, Evangelista A,
Falk V, Lung B, Lancellotti P, Pierard L, Price S, Schäfers HJ, Schuler
G, Stepinska J, Swedberg K, Takkenberg J, Von Oppell UO, Windecker
S, Zamorano JL, Zembala M; ESC Committee for Practice Guidelines
(CPG); Joint Task Force on the Management of Valvular Heart Disease
of the European Society of Cardiology (ESC); European Association for
Cardio-Thoracic Surgery (EACTS). Guidelines on the management of valvular heart disease (version 2012): the Joint Task Force on the Management
of Valvular Heart Disease of the European Society of Cardiology (ESC)
and the European Association for Cardio-Thoracic Surgery (EACTS). Eur
J Cardiothorac Surg. 2012;42:S1–44.
5. Gilard M, Eltchaninoff H, Iung B, Donzeau-Gouge P, Chevreul K,
Fajadet J, Leprince P, Leguerrier A, Lievre M, Prat A, Teiger E, Lefevre
T, Himbert D, Tchetche D, Carrié D, Albat B, Cribier A, Rioufol G, Sudre
A, Blanchard D, Collet F, Dos Santos P, Meneveau N, Tirouvanziam A,
Caussin C, Guyon P, Boschat J, Le Breton H, Collart F, Houel R, Delpine
S, Souteyrand G, Favereau X, Ohlmann P, Doisy V, Grollier G, Gommeaux
A, Claudel JP, Bourlon F, Bertrand B, Van Belle E, Laskar M; FRANCE 2
Investigators. Registry of transcatheter aortic-valve implantation in highrisk patients. N Engl J Med. 2012;366:1705–1715.
6. Hamm CW, Möllmann H, Holzhey D, Beckmann A, Veit C, Figulla HR,
Cremer J, Kuck KH, Lange R, Zahn R, Sack S, Schuler G, Walther T,
Beyersdorf F, Böhm M, Heusch G, Funkat AK, Meinertz T, Neumann T,
Papoutsis K, Schneider S, Welz A, Mohr FW; GARY-Executive Board.
The German Aortic Valve Registry (GARY): in-hospital outcome. Eur
Heart J. 2014;35:1588–1598.
7. De Carlo M, Giannini C, Fiorina C, Bedogni F, Napodano M, Klugmann
S, Tamburino C, Maisano F, Santoro G, Ettori F, Petronio AS. Paravalvular
leak after CoreValve implantation in the Italian Registry: predictors and
impact on clinical outcome. Int J Cardiol. 2013;168:5088–5089.
8. Lange R, Bleiziffer S, Mazzitelli D, Elhmidi Y, Opitz A, Krane M, Deutsch
MA, Ruge H, Brockmann G, Voss B, Schreiber C, Tassani P, Piazza N.
Improvements in trans-catheter aortic valve implantation outcomes in
lower surgical risk patients: a glimpse into the future. J Am Coll Cardiol.
2011;59:280–287.
9. Beckmann A, Funkat AK, Lewandowski J, Frie M, Schiller W, Hekmat K,
Gummert JF, Mohr FW. Cardiac surgery in Germany during 2012: a report
on behalf of the German Society for Thoracic and Cardiovascular Surgery.
Thorac Cardiovasc Surg. 2014;62:5–17.
10. Mylotte D, Osnabrugge RL, Windecker S, Lefèvre T, de Jaegere P, Jeger
R, Wenaweser P, Maisano F, Moat N, Søndergaard L, Bosmans J, Teles
RC, Martucci G, Manoharan G, Garcia E, Van Mieghem NM, Kappetein
AP, Serruys PW, Lange R, Piazza N. Transcatheter aortic valve replacement in Europe: adoption trends and factors influencing device utilization.
J Am Coll Cardiol. 2013;62:210–219.
11. Brown ML, Schaff HV, Lahr BD, Mullany CJ, Sundt TM, Dearani JA,
McGregor CG, Orszulak TA. Aortic valve replacement in patients aged 50
to 70 years: improved outcome with mechanical versus biologic prostheses. J Thorac Cardiovasc Surg. 2008;135:878–884.
12. Vasques F, Lucenteforte E, Paone R, Mugelli A, Biancari F. Outcome of
patients aged ≥80 years undergoing combined aortic valve replacement
and coronary artery bypass grafting: a systematic review and meta-analysis of 40 studies. Am Heart J. 2012;164:410–418.e1.
13. Takagi H, Niwa M, Mizuno Y, Goto SN, Umemoto T; All-Literature
Investigation of Cardiovascular Evidence (ALICE) Group. A meta-analysis of transcatheter aortic valve implantation versus surgical aortic valve
replacement. Ann Thorac Surg. 2013;96:513–519.
14. Panchal HB, Ladia V, Desai S, Shah T, Ramu V. A meta-analysis
of mortality and major adverse cardiovascular and cerebrovascular
events following transcatheter aortic valve implantation versus surgical aortic valve replacement for severe aortic stenosis. Am J Cardiol.
2013;112:850–860.
Downloaded from http://circ.ahajournals.org/ by guest on March 30, 2016
2340
Circulation
December 23/30, 2014
15. Mohr FW, Holzhey D, Möllmann H, Beckmann A, Veit C, Figulla HR,
Cremer J, Kuck KH, Lange R, Zahn R, Sack S, Schuler G, Walther T,
Beyersdorf F, Böhm M, Heusch G, Funkat AK, Meinertz T, Neumann T,
Papoutsis K, Schneider S, Welz A, Hamm CW; GARY Executive Board.
The German Aortic Valve Registry: 1-year results from 13 680 patients
with aortic valve disease. Eur J Cardiothorac Surg. 2014;46:808–816.
16. Astarci P, Glineur D, Kefer J, D’Hoore W, Renkin J, Vanoverschelde JL,
El Khoury G, Grandin C. MRI evaluation of cerebral embolization during
percutaneous aortic valve implantation: comparison of transfemoral and
transapical approaches using Edwards Sapiens valve. Eur J Cardiothorac
Surg. 2011;40:475–479.
17. Fanning JP, Walters DL, Platts DG, Eeles E, Bellapart J, Fraser JF.
Characterization of neurological injury in transcatheter aortic valve
implantation: how clear is the picture? Circulation. 2014;129:504–515.
18. Miller DC, Blackstone EH, Mack MJ, Svensson LG, Kodali SK, Kapadia S,
Rajeswaran J, Anderson WN, Moses JW, Tuzcu EM, Webb JG, Leon MB,
Smith CR; PARTNER Trial Investigators and Patients; PARTNER Stroke
Substudy Writing Group and Executive Committee. Transcatheter (TAVR) versus surgical (AVR) aortic valve replacement: occurrence, hazard, risk factors,
and consequences of neurological events in the PARTNER trial: the PARTNER
trial investigators and patients: the PARTNER Stroke Sub- study Writing Group
and Executive Committee. J Thorac Cardiovasc Surg. 2012;143:832–843.
19. Nielsen HH, Klaaborg KE, Nissen H, Terp K, Mortensen PE, Kjeldsen
BJ, Jakobsen CJ, Andersen HR, Egeblad H, Krusell LR, Thuesen L,
Hjortdal VE. A prospective, randomised trial of transapical transcatheter aortic valve implantation versus surgical aortic valve replacement
in operable elderly patients with aortic stenosis: the STACCATO trial.
EuroIntervention. 2012;8:383–389.
20. Walters DL, Sinhal A, Baron D, Pasupati S, Thambar S, Yong G, Jepson
N, Bhindi R, Bennetts J, Larbalestier R, Clarke A, Brady P, Wolfenden H,
James A, El Gamel A, Jansz P, Chew DP; SOURCE ANZ Investigators.
Initial experience with the balloon expandable Edwards-SAPIEN transcatheter heart valve in Australia and New Zealand: the SOURCE ANZ registry: outcomes at 30 days and one year. Int J Cardiol. 2014;170:406–412.
21. Eggebrecht H, Mehta RH, Kahlert P, Schymik G, Lefevre T, Lange R,
Macaya C, Mandinov L, Wendler O, Thomas M. Emergent cardiac surgery during transcatheter aortic valve implantation (TAVI): insights
from the Edwards SAPIEN Aortic Bioprosthesis European Outcome
(SOURCE) registry [published online ahead of print November 12, 2013].
EuroIntervention. http://www.pcronline.com/eurointervention/ahead_of_
print/201311-05. Accessed December 10, 2014.
22. Weisenberg D, Sagie A, Vaturi M, Monakier D, Sharoni E, Porat E, Shapira
Y. The value of intraoperative transesophageal echocardiography in patients
undergoing aortic valve replacement. J Heart Valve Dis. 2011;20:540–544.
23. Cao C, Ang SC, Indraratna P, Manganas C, Bannon P, Black D, Tian D,
Yan TD. Systematic review and meta-analysis of transcatheter aortic valve
implantation versus surgical aortic valve replacement for severe aortic stenosis. Ann Cardiothorac Surg. 2013;2:10–23.
24. Hahn RT, Pibarot P, Stewart WJ, Weissman NJ, Gopalakrishnan D, Keane
MG, Anwaruddin S, Wang Z, Bilsker M, Lindman BR, Herrmann HC,
Kodali SK, Makkar R, Thourani VH, Svensson LG, Akin JJ, Anderson WN,
Leon MB, Douglas PS. Comparison of transcatheter and surgical aortic
valve replacement in severe aortic stenosis: a longitudinal study of echocardiography parameters in cohort A of the PARTNER trial (placement of
aortic transcatheter valves). J Am Coll Cardiol. 2013;61:2514–2521.
25. Sinning JM, Vasa-Nicotera M, Chin D, Hammerstingl C, Ghanem A,
Bence J, Kovac J, Grube E, Nickenig G, Werner N. Evaluation and management of paravalvular aortic regurgitation after transcatheter aortic
valve replacement. J Am Coll Cardiol. 2013;62:11–20.
26. Van Belle E, Juthier F, Susen S, Vincentelli A, Iung B, Dallongeville J,
Eltchaninoff H, Laskar M, Leprince P, Lievre M, Banfi C, Auffray JL,
Delhaye C, Donzeau-Gouge P, Chevreul K, Fajadet J, Leguerrier A, Prat
A, Gilard M, Teiger E; FRANCE 2 Investigators. Post-procedural aortic
regurgitation in balloon-expandable and self-expandable TAVR procedures: analysis of predictors and impact on long-term mortality: insights
from the FRANCE2 Registry. Circulation. 2014;129:1415–1427.
27. Généreux P, Head SJ, Hahn R, Daneault B, Kodali S, Williams MR,
van Mieghem NM, Alu MC, Serruys PW, Kappetein AP, Leon MB.
Paravalvular leak after transcatheter aortic valve replacement: the new
Achilles’ heel? A comprehensive review of the literature. J Am Coll
Cardiol. 2013;61:1125–1136.
28. Willson AB, Webb JG, Labounty TM, Achenbach S, Moss R, Wheeler M,
Thompson C, Min JK, Gurvitch R, Norgaard BL, Hague CJ, Toggweiler
S, Binder R, Freeman M, Poulter R, Poulsen S, Wood DA, Leipsic J.
3-Dimensional aortic annular assessment by multidetector computed
tomography predicts moderate or severe paravalvular regurgitation after
transcatheter aortic valve replacement: a multicenter retrospective analysis. J Am Coll Cardiol. 2012;59:1287–1294.
29. Yared K, Garcia-Camarero T, Fernandez-Friera L, Llano M, Durst R,
Reddy AA, O’Neill WW, Picard MH. Impact of aortic regurgitation after
transcatheter aortic valve implantation: results from the REVIVAL trial.
JACC Cardiovasc Imaging. 2012;5:469–477.
30. Jacobs S, Gessat M, Walther T, Falk V. Three-dimensional template-based
planning for transapical aortic valve implantation. J Thorac Cardiovasc
Surg. 2011;141:1541–1543.
31. Zannad F, Huvelle E, Dickstein K, van Veldhuisen DJ, Stellbrink C, Køber
L, Cazeau S, Ritter P, Maggioni AP, Ferrari R, Lechat P. Left bundle-branch
block as a risk factor for progression to heart failure. Eur J Heart Fail.
2007;9:7–14.
32. Tzikas A, van Dalen BM, Van Mieghem NM, Gutierrez-Chico JL, Nuis RJ,
Kauer F, Schultz C, Serruys PW, de Jaegere PP, Geleijnse ML. Frequency
of conduction abnormalities after transcatheter aortic valve implantation
with the Medtronic-CoreValve and the effect on left ventricular ejection
fraction. Am J Cardiol. 2011;107:285–289.
33. Hoffmann R, Herpertz R, Lotfipour S, Aktug Ö, Brehmer K, Lehmacher
W, Autschbach R, Marx N, Lotfi S. Impact of a new conduction defect
after transcatheter aortic valve implantation on left ventricular function.
JACC Cardiovasc Interv. 2012;5:1257–1263.
34. Urena M, Mok M, Serra V, Dumont E, Nombela-Franco L, DeLarochellière
R, Doyle D, Igual A, Larose E, Amat-Santos I, Côté M, Cuéllar H,
Pibarot P, de Jaegere P, Philippon F, Garcia del Blanco B, Rodés-Cabau
J. Predictive factors and long-term clinical consequences of persistent left
bundle-branch block following transcatheter aortic valve implantation
with a balloon-expandable valve. J Am Coll Cardiol. 2012;60:1743–1752.
35. Urena M, Webb JG, Cheema A, Serra V, Toggweiler S, Barbanti M,
Cheung A, Ye J, Dumont E, DeLarochellière R, Doyle D, Al Lawati
HA, Peterson M, Chisholm R, Igual A, Ribeiro HB, Nombela-Franco L,
Philippon F, Garcia del Blanco B, Rodés-Cabau J. Impact of new-onset
persistent left bundle-branch block on late clinical outcomes in patients
undergoing transcatheter aortic valve implantation with a balloon-expandable valve. JACC Cardiovasc Interv. 2014;7:128–136.
36. Houthuizen P, van der Boon RM, Urena M, Van Mieghem N, Brueren GB, Poels
TT, Van Garsse LA, Rodés-Cabau J, Prinzen FW, de Jaegere P. Occurrence,
fate and consequences of ventricular conduction abnormalities after transcatheter aortic valve implantation. EuroIntervention. 2014;9:1142–1150.
37. Lin YS, Guo GB, Chen YL, Tsai TH, Pan KL, Liu WH, Chen MC. Atrial
enlargement in symptomatic heart block patients with preserved left ventricular function: possibly related to atrioventricular dyssynchrony. Int J
Cardiol. 2011;148:280–284.
38. Xie JM, Fang F, Zhang Q, Chan JY, Yip GW, Sanderson JE, Lam YY, Yan
BP, Yu CM. Left atrial remodeling and reduced atrial pump function after
chronic right ventricular apical pacing in patients with preserved ejection
fraction. Int J Cardiol. 2012;157:364–369.
39. Wolber T, Haegeli L, Huerlimann D, Brunckhorst C, Lüscher TF, Duru F.
Altered left ventricular contraction pattern during right ventricular pacing:
assessment using real-time three-dimensional echocardiography. Pacing
Clin Electrophysiol. 2011;34:76–81.
40. Sweeney MO, Hellkamp AS, Ellenbogen KA, Greenspon AJ, Freedman
RA, Lee KL, Lamas GA; MOde Selection Trial Investigators. Adverse
effect of ventricular pacing on heart failure and atrial fibrillation among
patients with normal baseline QRS duration in a clinical trial of pacemaker
therapy for sinus node dysfunction. Circulation. 2003;107:2932–2937.
41. Sharma AD, Rizo-Patron C, Hallstrom AP, O’Neill GP, Rothbart S,
Martins JB, Roelke M, Steinberg JS, Greene HL; DAVID Investigators.
Percent right ventricular pacing predicts outcomes in the DAVID trial.
Heart Rhythm. 2005;2:830–834.
42. Steinberg JS, Fischer A, Wang P, Schuger C, Daubert J, McNitt S, Andrews
M, Brown M, Hall WJ, Zareba W, Moss AJ; MADIT II Investigators. The
clinical implications of cumulative right ventricular pacing in the multicenter
automatic defibrillator trial II. J Cardiovasc Electrophysiol. 2005;16:359–365.
43. Biner S, Michowitz Y, Leshem-Rubinow E, Topilsky Y, Ben-Assa E,
Shimiaie J, Banai S, Keren G, Steinvil A, Finkelstein A. Hemodynamic
Downloaded from http://circ.ahajournals.org/ by guest on March 30, 2016
Falk
44.
45.
46.
47.
48.
49.
50.
51.
52.
impact and outcome of permanent pacemaker implantation following
transcatheter aortic valve implantation. Am J Cardiol. 2014;113:132–137.
Johansen JB, Jørgensen OD, Møller M, Arnsbo P, Mortensen PT, Nielsen
JC. Infection after pacemaker implantation: infection rates and risk factors associated with infection in a population-based cohort study of 46299
consecutive patients. Eur Heart J. 2011;32:991–998.
Kirkfeldt RE, Johansen JB, Nohr EA, Jørgensen OD, Nielsen JC.
Complications after cardiac implantable electronic device implantations:
an analysis of a complete, nationwide cohort in Denmark. Eur Heart J.
2014;35:1186–1194.
Brennan JM, Edwards FH, Zhao Y, O’Brien S, Booth ME, Dokholyan RS,
Douglas PS, Peterson ED; DEcIDE AVR (Developing Evidence to Inform
Decisions about Effectiveness–Aortic Valve Replacement) Research
Team. Long-term safety and effectiveness of mechanical versus biological aortic valve prostheses in older patients: results from the Society of
Thoracic Surgeons Adult Cardiac Surgery National Database. Circulation.
2013;127:1647–1655.
Weber A, Noureddine H, Englberger L, Dick F, Gahl B, Aymard T, Czerny
M, Tevaearai H, Stalder M, Carrel TP. Ten-year comparison of pericardial tissue valves versus mechanical prostheses for aortic valve replacement in patients younger than 60 years of age. J Thorac Cardiovasc Surg.
2012;144:1075–1083.
David TE, Armstrong S, Maganti M. Hancock II bioprosthesis for aortic
valve replacement: the gold standard of bioprosthetic valves durability?
Ann Thorac Surg. 2010;90:775–781.
Puskas J, Gerdisch M, Nichols D, Quinn R, Anderson C, Rhenman B,
Fermin L, McGrath M, Kong B, Hughes C, Sethi G, Wait M, Martin T,
Graeve A; PROACT Investigators. Reduced anticoagulation after mechanical aortic valve replacement: interim results from the prospective randomized on-X valve anticoagulation clinical trial randomized Food and
Drug Administration investigational device exemption trial. J Thorac
Cardiovasc Surg. 2014;147:1202–1210.
Dvir D, Webb J, Brecker S, Bleiziffer S, Hildick-Smith D, Colombo A,
Descoutures F, Hengstenberg C, Moat NE, Bekeredjian R, Napodano M,
Testa L, Lefevre T, Guetta V, Nissen H, Hernández JM, Roy D, Teles RC,
Segev A, Dumonteil N, Fiorina C, Gotzmann M, Tchetche D, Abdel-Wahab
M, De Marco F, Baumbach A, Laborde JC, Kornowski R. Transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: results
from the global Valve-in-Valve Registry. Circulation. 2012;126:2335–2344.
Dvir D, Barbanti M, Tan J, Webb JG. Transcatheter aortic valve-in-valve
implantation for patients with degenerative surgical bioprosthetic valves.
Curr Probl Cardiol. 2014;39:7–27.
Makkar RR, Jilaihawi H, Chakravarty T, Fontana GP, Kapadia S,
Babaliaros V, Cheng W, Thourani VH, Bavaria J, Svensson L, Kodali S,
Shiota T, Siegel R, Tuzcu EM, Xu K, Hahn RT, Herrmann HC, Reisman
M, Whisenant B, Lim S, Beohar N, Mack M, Teirstein P, Rihal C, Douglas
PS, Blackstone E, Pichard A, Webb JG, Leon MB. Determinants and outcomes of acute transcatheter valve-in-valve therapy or embolization: a
study of multiple valve implants in the US PARTNER trial (Placement of
TAVR in Lower-Risk and Younger Patients
53.
54.
55.
56.
57.
58.
59.
60.
61.
62.
2341
AoRTic TraNscathetER Valve Trial Edwards SAPIEN Transcatheter Heart
Valve). J Am Coll Cardiol. 2013;62:418–430.
Ribeiro HB, Webb JG, Makkar RR, Cohen MG, Kapadia SR, Kodali S,
Tamburino C, Barbanti M, Chakravarty T, Jilaihawi H, Paradis JM, de
Brito FS Jr, Cánovas SJ, Cheema AN, de Jaegere PP, del Valle R, Chiam
PT, Moreno R, Pradas G, Ruel M, Salgado-Fernández J, Sarmento-Leite
R, Toeg HD, Velianou JL, Zajarias A, Babaliaros V, Cura F, Dager AE,
Manoharan G, Lerakis S, Pichard AD, Radhakrishnan S, Perin MA,
Dumont E, Larose E, Pasian SG, Nombela-Franco L, Urena M, Tuzcu
EM, Leon MB, Amat-Santos IJ, Leipsic J, Rodés-Cabau J. Predictive
factors, management, and clinical outcomes of coronary obstruction following transcatheter aortic valve implantation: insights from a large multicenter registry. J Am Coll Cardiol. 2013;62:1552–1562.
Barbanti M, Yang TH, Rodès Cabau J, Tamburino C, Wood DA, Jilaihawi
H, Blanke P, Makkar RR, Latib A, Colombo A, Tarantini G, Raju R, Binder
RK, Nguyen G, Freeman M, Ribeiro HB, Kapadia S, Min J, Feuchtner G,
Gurtvich R, Alqoofi F, Pelletier M, Ussia GP, Napodano M, de Brito FS
Jr, Kodali S, Norgaard BL, Hansson NC, Pache G, Canovas SJ, Zhang H,
Leon MB, Webb JG, Leipsic J. Anatomic and procedural features associated with aortic root rupture during balloon-expandable transcatheter aortic valve replacement. Circulation. 2013;128:244–253.
Griese DP, Reents W, Kerber S, Diegeler A, Babin-Ebell J. Emergency cardiac surgery during transfemoral and transapical transcatheter aortic valve
implantation: incidence, reasons, management, and outcome of 411 patients
from a single center. Catheter Cardiovasc Interv. 2013;82:E726–E733.
Boothroyd LJ, Spaziano M, Guertin JR, Lambert LJ, Rodés-Cabau J,
Noiseux N, Nguyen M, Dumont É, Carrier M, de Varennes B, Ibrahim
R, Martucci G, Xiao Y, Morin JE, Bogaty P. Transcatheter aortic valve
implantation: recommendations for practice based on a multidisciplinary
review including cost-effectiveness and ethical and organizational issues.
Can J Cardiol. 2013;29:718–726.
Sehatzadeh S, Doble B, Xie F, Blackhouse G, Campbell K, Kaulback K,
Chandra K, Goeree R. Transcatheter aortic valve implantation (TAVI) for
treatment of aortic valve stenosis: an evidence-based analysis (part B). Ont
Health Technol Assess Ser. 2012;12:1–62.
Indraratna P, Ang SC, Gada H, Yan TD, Manganas C, Bannon P, Cao C.
Systematic review of the cost-effectiveness of transcatheter aortic valve
implantation. J Thorac Cardiovasc Surg. 2014;148:509–514.
Bower JL, Christensen CM. Disruptive technologies: catching the wave.
Harvard Business Review, January–February 1995.
Christensen CM. The Innovator’s Dilemma: When New Technologies Cause
Great Firms to Fail. Boston, MA: Harvard Business School Press; 1997.
DeMaria AN. Self-referral in cardiology. J Am Coll Cardiol. 2004;43:
1500–1501.
Osnabrugge RL, Mylotte D, Head SJ, Van Mieghem NM, Nkomo VT,
LeReun CM, Bogers AJ, Piazza N, Kappetein AP. Aortic stenosis in the
elderly: disease prevalence and number of candidates for transcatheter
aortic valve replacement: a meta-analysis and modeling study. J Am Coll
Cardiol. 2013;62:1002–1012.
Downloaded from http://circ.ahajournals.org/ by guest on March 30, 2016
2342
Circulation
December 23/30, 2014
Response to Falk
Stephan Haussig, MD; Axel Linke, MD
Dr Falk rightfully acknowledges the excellent outcomes of conventional aortic valve replacement (AVR), even in an elderly
population. In Placement of Aortic Transcatheter Valve Trial Edwards SAPIEN Transcatheter Heart Valve (PARTNER)
cohort A, transcatheter AVR (TAVR) did not improve survival in operative high-risk patients. Given that the life expectancy
is reduced in this population, especially because of comorbidities that are often life-limiting, it will be hard to show a survival benefit of TAVR compared with surgical AVR (SAVR) in those who are extremely sick. However, patients recovered
much more quickly after TAVR and gained quality of life earlier after the intervention. This achievement is often ignored
when the discussion between TAVR and SAVR comes down to a comparison of survival curves. Recently, in a patient population that was less sick on the basis of the Society of Thoracic Surgeons score compared with PARTNER A, Adams et al
reported that TAVR with a self-expanding transcatheter aortic valve bioprosthesis was associated with a higher survival rate
at 1 year than SAVR. This treatment effect was consistent and independent of age, sex, Society of Thoracic Surgeons score,
and ejection fraction. It appears that the lower invasiveness and higher aortic valve orifice areas after TAVR might outweigh
the lower level of aortic regurgitation and conduction abnormalities after SAVR, despite all present TAVR shortcomings. In
addition, there was no evidence of increased stroke rates at short- or longer-term follow-up with TAVR. These data provide
further evidence that TAVR might be an alternative to SAVR even in a lower-risk population. Nevertheless, results from
randomized trials are required before the indications can be expanded.
Downloaded from http://circ.ahajournals.org/ by guest on March 30, 2016
Transcatheter Aortic Valve Replacement Indications Should Not Be Expanded to
Lower-Risk and Younger Patients
Volkmar Falk
Circulation. 2014;130:2332-2342
doi: 10.1161/CIRCULATIONAHA.114.008145
Circulation is published by the American Heart Association, 7272 Greenville Avenue, Dallas, TX 75231
Copyright © 2014 American Heart Association, Inc. All rights reserved.
Print ISSN: 0009-7322. Online ISSN: 1524-4539
The online version of this article, along with updated information and services, is located on the
World Wide Web at:
http://circ.ahajournals.org/content/130/25/2332
Permissions: Requests for permissions to reproduce figures, tables, or portions of articles originally published
in Circulation can be obtained via RightsLink, a service of the Copyright Clearance Center, not the Editorial
Office. Once the online version of the published article for which permission is being requested is located,
click Request Permissions in the middle column of the Web page under Services. Further information about
this process is available in the Permissions and Rights Question and Answer document.
Reprints: Information about reprints can be found online at:
http://www.lww.com/reprints
Subscriptions: Information about subscribing to Circulation is online at:
http://circ.ahajournals.org//subscriptions/
Downloaded from http://circ.ahajournals.org/ by guest on March 30, 2016