Biosensors and Bioelectronics 24 (2009) 2643–2648
Contents lists available at ScienceDirect
Biosensors and Bioelectronics
journal homepage: www.elsevier.com/locate/bios
User-friendly, miniature biosensor flow cell for fragile high fundamental
frequency quartz crystal resonators
Brigitte P. Sagmeister a , Ingrid M. Graz a,1 , Reinhard Schwödiauer a,∗ , Hermann Gruber b , Siegfried Bauer a
a
b
Institute of Experimental Physics, Soft Matter Physics Division, Johannes Kepler University, Altenbergerstrasse 69, 4040 Linz, Austria
Institute of Biophysics, Johannes Kepler University, Linz, Austria
a r t i c l e
i n f o
Article history:
Received 30 October 2008
Received in revised form
23 December 2008
Accepted 19 January 2009
Available online 29 January 2009
Keywords:
Quartz crystal microbalance
High frequency fundamental
Flow cell
a b s t r a c t
For the application of high fundamental frequency (HFF) quartz crystal resonators as ultra sensitive acoustic biosensors, a tailor-made quartz crystal microbalance (QCM) flow cell has been fabricated and tested.
The cell permits an equally fast and easy installation and replacement of small and fragile HFF sensors. Usability and simple fabrication are two central features of the HFF-QCM flow cell. Mechanical,
thermal, electrical and chemical requirements are considered. The design of the cell combines these,
partially contradictory, requirements within a simple device. Central design concepts are discussed and
a brief description of the fabrication, with a special focus on the preparation of crucial parts, is provided.
For test measurements, the cell was equipped with a standard 50 MHz HFF resonator which had been
surface-functionalised with a self-assembled monolayer of 1-octadecanethiol. The reliable performance
is demonstrated with two types of experiments: the real time monitoring of phospholipid monolayer
formation and its removal with detergent, as well as step-wise growth of a protein multilayer system by
an alternating immobilisation of streptavidin and biotinylated immunoglobulin G.
© 2009 Elsevier B.V. All rights reserved.
1. Introduction
During the last decade acoustic biosensors which detect massand viscosity-alterations of surface adsorbed biofilms via a corresponding shift in resonance frequency (Steinem and Janshoff, 2007;
Bizet et al., 1999; Ballantine et al., 1997), have found widespread
acceptance as versatile tools for the analysis of biomolecular
interactions and related phenomena. Especially the quartz crystal
microbalance (QCM) with a thickness-shear mode (TSM) resonator
in combination with a flow cell is gaining increasing relevance for
biological and biochemical research (Steinem and Janshoff, 2007).
This trend is well reflected by the steady growth of related publications as recently reported by Cooper and Singleton (2007), who
documented, for the period 2001–2005, more than 1400 articles
referencing “quartz crystal microbalance” or “QCM” in the Web of
Science database.
Various properties of QCM systems may account for their extensive usage. Beside the scientific merits of QCM biosensors (Marx,
2007), the principle of operation is simple and the basic elements of
such a system are easily available at relatively low costs. Hence, for
many research groups home made systems with a reasonable per-
∗ Corresponding author. Fax: +43 732 2468 9273.
E-mail address: reinhard.schwoediauer@jku.at (R. Schwödiauer).
1
Present address: University of Cambridge, Nanoscience Centre, 11 JJ Thomson
Ave, Cambridge, Cambridgeshire CB3 0FF, United Kingdom.
0956-5663/$ – see front matter © 2009 Elsevier B.V. All rights reserved.
doi:10.1016/j.bios.2009.01.023
formance are easy to build and an alternative to more sophisticated,
but also far more expensive commercial machines.
A crucial factor for the performance of QCM systems – home
made or commercial – is the fundamental resonance frequency, f0 ,
of the oscillating piezoelectric sensor. A mass alteration per unit
area, m, of a thin biofilm on the sensor surface is, to a first order
approximation, the main detectable effect of biochemical binding
interactions under study. It can be measured via a corresponding
frequency shift, f , according to the Sauerbrey relation (Sauerbrey,
√
1959)f = −2f02 / · m, with the density = 2648 g/cm3 , and
the shear modulus = 29.47 GPa (AT-cut) of the quartz crystal.
Since the theoretical sensitivity increases with f02 , the use of a sensor
with a high fundamental resonance frequency
is clearly desirable.
//2h, the fabrication
Following the resonance condition f0 =
of TSM resonators with higher values for f0 is basically achieved by
reducing the plate thickness h of the sensor.
For practical purposes however, the minimum thickness is constrained by mechanical stability issues. Most applications utilise
either 5 MHz or 10 MHz resonators, with a corresponding plate
thickness of 0.33 mm or 0.17 mm. Such thicknesses, in combination with a diameter of approximately 14 mm, provide a sufficient
mechanical stability for an easy installation in a flow cell. Occasionally sensors with resonance frequencies between 20 MHz and
30 MHz are used (Okahata et al., 2000, 2007; Sota et al., 2002;
Michalzik et al., 2005; Sota et al., 2002). These sensors are smaller
in diameter (approximately 8 mm) and already rather thin and fragile. The corresponding plate thicknesses between 83 m and 56 m
2644
B.P. Sagmeister et al. / Biosensors and Bioelectronics 24 (2009) 2643–2648
require more careful handling, and also stress-free mounting to flow
cells becomes increasingly difficult (Sota et al., 2002). Thus, a further improvement towards higher sensitivities may not be expected
from the exploitation of even thinner standard resonators, though
they are available up to about 45 MHz.
An alternate approach to higher frequencies is the utilisation of
quartz resonators with an “inverted mesa” structure. These quartz
disks with a diameter and thickness of about 5 mm and 0.1 mm,
respectively, have a membrane of reduced thickness only in their
small central circular area of about 5 mm2 . The surrounding thicker
material provides a better mechanical stability and the small membrane can become as thin as 8.3 m, resulting in a high fundamental
frequency (HFF) up to f0 = 200 MHz.
The first application of a 30 MHz HFF-QCM sensor in a liquid
environment was demonstrated in 1993 by Lin et al. (1993) who
used glucose and electrodeposited copper on the sensor surface
to investigate the acoustic properties. The first application as a
biosensor was presented eight years later. In a comparative study,
Uttenthaler et al. (2001) investigated the detection of M13 phages in
liquids and the acoustic properties of glycerol/water mixtures with
four different HFF quartz crystals, operating at 39 MHz, 56 MHz,
70 MHz and 110 MHz, and with a 19 MHz standard quartz resonator. With respect to the 19 MHz standard the increase of the
relative sensitivity with increasing frequency was even stronger
than theoretically predicted. The 56 MHz sensor showed the best
performance, with a relative improvement of the signal-to-noise
ratio by a factor of 6.5 and an enhanced detection limit for phages
by a factor of 200.
In spite of these encouraging results, it seems that HFF-QCM
systems have still not yet appeared in research laboratories. The reason for this is rather obvious. High fundamental frequency quartz
crystals are too small and, although mechanically more stable than
standard resonators of equal frequency, they are still too fragile,
which makes them difficult to handle and to support. Both, the
small size and the fragile nature of a HFF resonator hardly allow the
application of common design concepts which are frequently used
for standard QCM flow cells. O-rings and electric spring contacts, as
they find wide application in 5 MHz and 10 MHz systems, are not
easy to down-scale. Standardised O-rings are much too hard and
too bulky, electric spring contacts will find very little room to connect to the sensor and they would be tricky to fabricate. In addition,
both elements will create a highly nonuniform stress distribution
through the small quartz crystal and the active membrane unavoidably will be affected. It has been shown by Sota et al. (2002) that
O-ring based holders for a 27 MHz quartz crystal resonator already
cause less stable sensor signals. Therefore, most researchers who
investigate and/or apply high frequency TSM biosensors glue the
sensor irreversibly onto a larger carrier chip, e.g. see (Uttenthaler
et al., 2001; Sota et al., 2002; Michalzik et al., 2005). This solution to
the mounting problem actually works well, but it is very delicate,
time consuming and awkward in practice. All HFF QCM sensors used
in the study of Uttenthaler et al. (2001), for instance, were fabricated
with bond wired crystals glued by a silicone adhesive to a carrier
with open flow channels. In such a configuration the sensor crystal,
once mounted, can hardly be removed from the carrier, a fact that
might easily turn out to be annoying, if not problematic.
In the present work we introduce a revised version of a HFFQCM flow cell which we have tailored to the special requirements
of small and fragile HFF quartz resonators. The design allows for fast
and reliable installation and replacement of a HFF sensor within
a few seconds. This high user friendliness is a key feature of our
system and it is accomplished without the use of complicated
engineering or delicate parts. Both, the essential design guidelines and the manufacturing of crucial elements are discussed. We
demonstrate the effective operation of the system with two kinds
of experiments, both using a self-assembled monolayer (SAM) of
1-octadecanethiol on the gold-covered surface of a 50 MHz HFF
resonator: In a first experiment we have tested the reversible
functionalisation of an installed sensor by phospholipid monolayer formation from an aqueous buffer suspension of phospholipid
vesicles, followed by complete lipid removal by the detergent
octyl- -D-glucopyranoside. In the second experiment, formation
of an alternating protein multilayer composed of streptavidin and
biotinylated immunoglobulin G was monitored.
2. Experimental
2.1. Requirements and design guidelines
The general purpose of a flow cell for QCM applications is primarily to support and connect the sensor mechanically and electrically
in such a way that a flowing liquid can wet the sensitive surface
in a defined region without leaking into the surrounding area.
Whereas these demands are relatively easy to meet with large and
stable quartz crystals, this becomes considerably more challenging for small and fragile HFF quartz resonators. For a user-friendly
HFF-QCM flow cell however, there are even more requirements to
specify: First, (i) the cell has to allow for fast, easy, and reliable
installation and replacement of the HFF quartz sensor. Gluing or
pre-wiring of the sensor shall be avoided by an effective clamping
mechanism. (ii) The clamping mechanism has to act on the fragile
quartz crystal with a gentle and uniform pressure; a shear-stress
across the crystal must not build up. (iii) Under no circumstances,
liquid must leak through the flow cell/sensor interface. Further, (iv)
all materials which get in contact with the liquid have to be biologically inert. (v) All electrical lines and contacts shall be fit for high
frequency signals; wires shall be short and with few discontinuities in impedance. (vi) The complete system shall be fully reusable
with only little need for maintenance. Also important for a good
performance are (vii) the thermal properties of the cell. A small
heat capacity, in combination with a high heat conductivity, allows
for fast and accurate temperature stabilisation via an external active
control unit. Thermal fluctuations on the outer cell-surface shall not
reach the flowing liquid, the sensor, and the probe volume. Thus,
(viii) the probe volume itself shall be located in the centre, and the
probe volume shall be very small, providing for fast sensor response.
In addition to all these requirements, (ix) the fabrication should be
inexpensive and technologically feasible even for a small workshop
with conventional machine tools.
2.2. The flow cell, fabrication and assembling
All requirements mentioned above, and the related conflicting
requirements of the constructions could be met and resolved with
the design given in Fig. 1. For the purpose of a better illustration,
the presented schematic outline is not drawn to scale and some
features of small size are depicted out of proportion. A detailed
exploded assembly drawing together with a complete part list and
additional photographs are presented as supporting online material
in Figs. S1 and S2 respectively.
The complete flow cell system measures 30 mm in diameter and
26 mm in height. The flow cell is made up of two main components:
The lower flow-channel unit, which supports the HFF quartz crystal,
constitutes the major part of the system. The top structure – the flow
cell head – provides the high frequency signal via a planar electrode
and simultaneously generates a uniform pressure upon the quartz
crystal. The accurate positioning between the two parts is ensured
by two precise guide bolts with threads which are tightly fixed to the
lower unit and which stick through, and out of, the flow cell head, so
that the two parts can be screwed together with two curled nuts on
top. The quartz crystal is sandwiched in between these two parts.
B.P. Sagmeister et al. / Biosensors and Bioelectronics 24 (2009) 2643–2648
2645
a 0.5 mm strong spring steel wire exerts a single point force on the
semi-spherical insulation cap. With a space of about 0.15 mm to the
surrounding aluminium, the cap is free to move and thus the ring
electrode can sit on the quartz crystal with a uniform pressure. From
the top electrode, the electric contact is made with a short, flexible,
50 m thick wire to a RG 316/D coaxial cable. The outer conductor
of the cable is connected to the aluminium structure. The closing of
the flow cell system also closes the electric circuit.
A stable and reliable, drift-free sensor operation depends on
two important criteria. The first criterion is a perfectly planar and
smooth surface of the sensor platform and the top planar electrode.
The second criterion is the fabrication and installation of the soft,
slightly adhesive, and elevated PDMS sealant. The first step in fabrication is to cover the polished sensor platform with a thin Scotch
tape. The Scotch tape must accurately match with the rim of the
base electrode, the excess tape is trimmed away with the aid of a
scalpel under a microscope. Afterwards the top of the flow-channel
unit is covered with a thin polyimide sheet and a microscope slide
which are firmly pressed against the sensor platform by two clamps.
The PDMS, prepared with 2.5 vol.% of a curing agent, is injected into
the mold with a syringe through the aligned filler holes. It is essential to observe the filling process under a microscope in order to
see and eliminate air bubbles around the base electrode. Approximately 12 h later the microscope slide can be removed and the
polyimide sheet can be stripped off carefully. The Scotch tape can
then be detached with care from the sensor platform as well, and
any remainders may be washed off with alcohol and a cotton bud.
At this stage, the flow cell system is ready to use. With the aid of
a centreing device, e.g. a thin concentric ring, the HFF quartz crystal
is exactly positioned and pressed against the adhesive PDMS. The
centring device can immediately be removed and the flow cell head
can be installed. Because of the weak adhesion to the PDMS, the
quartz crystal can also be removed or replaced with ease at any time.
A damaged or ineffective sealant can be pulled out, together with
the reusable mold. After cleaning the mold and the base electrode,
a new sealant can be prepared.
Fig. 1. Cross-sectional view from the A–A direction of the flow cell head (a) and the
flow-channel unit with the HFF resonator above the base electrode (b) and the top
view (c) of the flow cell.
The main structure of the flow cell system is made of aluminium
because of its favourable thermal and electrical properties; stainless
steel is used for the electrodes. The lower unit incorporates the flow
cell-core, made of poly(etheretherketon) (PEEK). PEEK is chemically
inert and easy to machine, it thus allowed for conventional drilling
of the long and small flow channels with a diameter of 0.3 mm. The
upper half of the flow cell–core, with the smaller diameter of 4 mm,
is ringed by a slightly conical stainless steel tube which is also in firm
contact with the main aluminium structure. This element forms the
base electrode for the HFF quartz crystal. With an outer diameter
of 4.5 mm, the electrode is 0.5 mm smaller than the diameter of the
quartz crystal. The resulting narrow ring-area of the crystal beyond
the base electrode rests on a ∼50 m elevated, soft and slightly
adhesive polydimetylsiloxane (PDMS) sealant. A detailed description about the fabrication of the PDMS sealant, which is a crucial
element, is given below. When the HFF quartz crystal is in full contact with the sensor platform, constituted by the PEEK surface and
the coplanar base electrode, the actual flow chamber is formed by
the void beneath the resonator membrane with a volume of about
0.164 l. Only little pressure is needed to compress the PDMS; it
adheres to the quartz and seals the interface effectively. The pressure on the quartz is transferred via the contact area of the upper
planar electrode which exactly matches with the area of the HFF
quartz crystal. A plastic insulation cap covers the planar electrode
and separates it from the surrounding aluminium structure. On top,
2.3. Flow injection analysis system
The HFF-QCM flow cell system is part of a flow injection analysis
system (FIA). As schematically depicted in Fig. 2 the FIA system
consists of three subunits, the active thermal control unit, the fluid
transport channel unit and the measurement unit.
The active thermal control unit consists of a 70 W Peltier element which is mounted on an air-cooled radiator. The Peltier
element is powered by a general purpose dual output power supply
via a homemade 15 W bipolar current amplifier which is controlled
by an Eurotherm 2408 PID temperature controller. The temperature
of the flow cell is sensed with a type-K thermoelement which sits
centrally inside the flow cell core 0.5 mm below the sensor platform. At an ambient room temperature of 21 ◦ C a constant flow
cell-temperature between 7 ◦ C and 45 ◦ C can be maintained with a
deviation of less than ±0.05 ◦ C.
The fluid transport channel unit comprises a Smartline 1000
HPLC pump, a Rehodyne 7725 sample injector with a sample
volume of 200 l, a 25 cm long PEEK supply capillary, and the connected flow cell. The central instrument of the measurement unit is
a Hewlett Packard 4291 B impedance analyser which drives the HFF
quartz crystal and measures the resonance frequency every 2.5 s. A
small program on a connected computer reads, stores and displays
the data in real time.
2.4. Materials
Blank HFF quartz crystals with a resonance frequency of f0 ≈
50 MHz were obtained from KVG Quartz Crystal Technology GmbH
2646
B.P. Sagmeister et al. / Biosensors and Bioelectronics 24 (2009) 2643–2648
Fig. 2. Flow injection analysis system.
(http://www.kvg-gmbh.de). The measured crystal dimensions are
5 mm and 0.1 mm for the diameter and thickness of the quartz
disk and 2.5 mm and 33.4 m for the diameter and thickness
of the symmetric membrane. The PDMS Silicone Elastomer, Sylgard 184, was purchased from Dow Chemical (Germany). For
SAM formation on gold, 1-octadecanethiol (ODT) was obtained
from Aldrich. The phospholipids 1,2-dioleoylphosphatidylcholin
(DOPC), 1,2-dioloeoylphosphatidylserin (DOPS) and N-(6-biotinylamidocaproyl) 1,2-dioleoylphosphatidylethanolamin (biotin-capDOPE) were obtained from Avanti Polar Lipids. The detergent
octyl- -D-glucopyranoside (octylglucoside) was bought from
Sigma–Aldrich. Biotinylated immunoglobolin G (IgG) antibodies
with a molecular weight of 150,000 g/mol were prepared as
described (Kamruzzahan et al., 2004). Streptavidin was supplied
by Rockland. Phosphate-buffered saline (PBS) was prepared with
water from a Milli-Q50 system (Millipore Corp., Bedford, WA) and
contained 140 mM NaCl, 2.7 mM KCl, 10 mM Na2 HPO4 and 1.8 mM
KH2 PO4 . The pH of this solution was found to be 7.3, as required
and no further adjustment was necessary.
2.5. Sensor preparation
The blank HFF quartz crystal is first sonicated in ethanol. After
cleaning, the crystal is put in an appropriate metal shadow mask
for electrode evaporation. The shadow mask covers just the outer
0.4 mm of the quartz disk. A 70 nm Au electrode on 3 nm Cr is
evaporated on both sides and all over the HFF quartz. The resulting electrode diameter of 4.2 mm will always find electrical contact
to the base electrode, the remaining Au-free ring at the rim is the
clean contact area for the adhesive sealant. Ideally, the metallised
quartz is transferred immediately after electrode evaporation into a
1 mM ODT solution. After about twelve hours, a monolayer of ODT
has self-assembled on the Au surface and the functionalised HFF
sensor is removed from the ODT solution, extensively rinsed with
ethanol and distilled water, dried with nitrogen gas, and installed
in the flow cell, as described above.
double-sided Au-coated 50 MHz HFF resonator in the flow cell
and compared with measurements of an industrially packaged
51 MHz HFF resonator in a conventional two-point mount package with removed cover. Both resonators were measured at room
temperature (22 ◦ C) under dry condition and under liquid loading
with distilled, purified water. The loading of the packaged 51 MHz
resonator happened with a drop of water from a syringe on its horizontal top surface. In each case the noise of the resonance frequency
and the quality factor
were recorded. The numerical results for the
mean frequency, f0 , the standard deviation, , the maximum frequency
difference, fmax = fmax − fmin , and the mean quality factor,
Q , are listed in Table 1, the corresponding figures of the frequency noise are presented in the supplementary part online with
Figs. S3 and S4 (also characteristic resonance spectra are presented
with Fig. S5, noise figures at different temperatures are discussed
and presented with Fig. S6 and Table S2). Both resonators show
very little difference in their frequency noise. The noise of the resonator in the flow cell is even a bit lower than the noise of the
resonator supported by standard two-point mounting clips. This
result is a good indication that the shear stress in the resonator
induced by the flow cell clamping mechanism does not exceed
the already low shear stress which is produced by conventional
mounting clips.
It has to be mentioned, however, that the frequency noise is
still considerably higher than reported by Uttenthaler et al. (2001).
The reason for this notable difference may be attributed to electronic issues of our circuitry. We have used a broadband impedance
analyser instead of a high quality low bandwidth oscillator circuit.
Even resonators contacted stress-free with silver paste and thin
gold wires did not show lower noise levels than those given in
Table 1. The values mark the low resolution limit of our electronic
measurement system. A specially designed oscillator circuit with
Table 1
Statistical parameters of the signal noise for a HFF resonator in the flow cell and for a
HFF resonator bonded between standard two-point mounting clips in dry condition
and under liquid loading (l.l.).
3. Results and discussion
3.1. Performance test
The quality of the flow cell clamping mechanism is crucial
for low frequency noise. It was tested by measurements of a
f0
Flow cell dry
Open HFF dry
Flow cell l.l.
Open HFF l.l.
(MHz)
50.828
51.823
50.806
51.795
(Hz)
fmax (Hz)
0.67
0.98
14.70
35.25
3.96
6.60
89.27
176.00
9820
28050
1034
964
Q
B.P. Sagmeister et al. / Biosensors and Bioelectronics 24 (2009) 2643–2648
Fig. 3. Reversible functionalisation of the sensor surface with a phospholipid monolayer by alternating injection of lipid vesicles (DOPC/DOPS = 9/1, 1 mg/ml in PBS) and
the detergent octylglucoside (40 mM in water).
a narrow bandwidth is likely to improve the signal-to-noise ratio
significantly.
3.2. Reversible functionalisation of the sensor surface with a
phospholipid monolayer
The assembled HFF sensor was maintained at 22 ◦ C and perfused with a continuous flow of PBS at a rate of 30 l/min. In
this configuration, a base line with a mean resonance frequency
of about 50.8549 MHz and a standard deviation of ±10.2 Hz could
be measured, the corresponding Q-factor, Q = f0 /fFWHM , with the
frequency width, fFWHM , at the 50% value of the conductance, was
about 985.
As can be seen in Fig. 3 the injection of lipid vesicles (DOPC/DOPS
= 9/1, 1 mg/ml in PBS) resulted in a rapid change of the resonance
frequency and after about 7 min a stable level was reached which
was almost 700 Hz lower than before the injection. This process is
well known to reflect the formation of a phospholipid monolayer on
the hydrophobic SAM of octadecanethiol (Lingler et al., 1987). The
phospholipid monolayer could perfectly be removed by injection
of octylglucoside (40 mM), as regularly used to regenerate BIAcore
chips with octadecanethiol SAMs (Gamsjäger et al., 2005).
The different viscosity and density of the detergent causes a
strong response with a frequency shift of almost −2 kHz. After the
octylglucoside injection the sensor surface is again superfused with
PBS and now the resonance frequency returns to the original value
from before the application of phospholipid, as previously observed
for octadecanethiol-coated BIAcore chips (Gamsjäger et al., 2005).
The procedure of lipid monolayer formation and its removal with
detergent can be repeated many times, the sensor response is
always reproducible with a frequency shift of about f ≈ −560 Hz
for the monolayer formation and ≈ −2 kHz for the response during
detergent injection.
The surfactant activity of the detergent is a special challenge
for the PDMS sealant. It is worth mentioning therefore, that the
sealant/sensor interface was always working reliably and no leakage has been observed.
3.3. Multilayer formation
A second HFF quartz crystal was coated with 1-octadecanethiol
and installed in the flow cell. The sensor was first exposed to a
continuous flow of PBS at a rate of 30 l/min for approximately
25 min and the flow cell temperature was adjusted to 23 ◦ C. After
stabilisation of the base line, a mean resonance frequency of about
50.8423 MHz and a standard deviation of ±10.2 Hz could be mea-
2647
Fig. 4. Formation of protein multilayer on top of a phospholipid monolayer. Lipid
vesicles (DOPC/DOPS/biotin-cap-DOPE = 70/10/20, 1 mg/ml in PBS) were injected,
followed by alternating injections of streptavidin (2 M, in PBS) and biotinylated
IgG (2 M, in PBS).
sured, the corresponding Q-factor was about 1007. Then a vesicle
suspension with another lipid composition (DOPC/DOPS/biotincap-DOPE = 70/10/20, w/w/w, 1 mg/ml in PBS) was injected,
resulting in monolayer formation which carried biotin residues
on 20% of the phospholipid head groups. Monolayer formation
was accompanied by a decrease in the resonance frequency of
about −5 kHz and a stable plateau was reached within 5 min. The
biotin-presenting monolayer allowed for biospecific binding of
streptavidin which is a tetramer having two pairs of biotin-binding
sites of which only two are used for docking to the lipid monolayer
(Hahn et al., 2007). The second pair of biotin-binding sites is then
available for binding of antibodies (IgG) carrying about 5–6 biotin
residues per protein molecule (Kamruzzahan et al., 2004), allowing
for protein multilayer formation (Hahn et al., 2007), as exemplified
in Fig. 4. In consecutive steps of about 11 min a 2 M streptavidin
solution and a 2 M biotin-IgG solution were injected. Each immobilisation caused a strong decrease in the resonance frequency. The
mean frequency change is −6000 Hz for binding of a strepatavidin
monolayer and −4400 Hz for each layer of biotin-IgG.
4. Conclusion
The flow cell presented in this study allows for simple and
efficient use of small and fragile HFF quartz crystal resonators as
high-sensitivity QCM biosensors. Any circular HFF resonator with
suitable diameter can be installed and/or replaced within a few seconds. No pre-configuration like gluing, sealing or wiring is required.
This high user friendliness is one of the key features of the system. It is achieved with a multifunctional clamping mechanism
which proved to be robust and reliable throughout all experiments.
Another advantage of the flow cell is the “low-tech” design. No
sophisticated machinery or technology is necessary to fabricate
all parts, including the crucial elements of the clamping mechanism.
In general, the presented flow cell demonstrates a good working
concept and a first approach towards a system that makes the application of HFF quartz crystal resonators not only beneficial in terms
of sensitivity but also more easily accessible, more convenient, and
thus attractive for practical application.
Acknowledgments
Work supported by the Austrian Science Funds within the Translational Research Program, FFG L442-N14.
2648
B.P. Sagmeister et al. / Biosensors and Bioelectronics 24 (2009) 2643–2648
Appendix A. Supplementary Data
Supplementary data associated with this article can be found, in
the online version, at doi:10.1016/j.bios.2009.01.023.
References
Ballantine, D.S., White, R.M., Martin, S.J., Ricco, A.J., Zellers, E.T., Frye, G.C., Wohltjen
Jr., H., 1997. Acoustic Wave Sensors, 1st ed. Academic Press, San Diego.
Bizet, K., Gabrielli, C., Perrot, H., 1999. Analysis 7, 609–616.
Cooper, M.A., Singleton, V.T., 2007. J. Mol. Recognit. 3, 154–184.
Gamsjäger, R., Johs, A., Gries, A., Gruber, H.J., Romanin, C., Prassl, R., Hinterdorfer, P.,
2005. Biochem. J. 389, 665–673.
Hahn, C.D., Leitner, C., Weinbrenner, T., Schlapak, R., Tinazli, A., Tampé, R., Lackner,
B., Steindl, C., Hinterdorfer, P., Gruber, H.J., Hölzl, M., 2007. Bioconj. Chem. 18,
247–253.
Kamruzzahan, A.S.M., Kienberger, F., Stroh, C.M., Berg, J., Huss, R., Ebner, A., Zhu, R.,
Rankl, C., Gruber, H.J., Hinterdorfer, P., 2004. Biol. Chem. 385, 955–960.
Lingler, S., Rubinstein, I., Knoll, W., Offenhausser, A., 1987. Langmuir 26, 7085–
7091.
Lin, Z., Yip, C.M., Joseph, I.S., Ward, M.D., 1993. Anal. Chem. 11, 1546–1551.
Marx, K.A., Steinem, C., Janshoff, A., 2007. Piezoelectric Sensors, 1st ed. Springer,
Berlin, Heidelberg, pp. 371–424.
Michalzik, M., Wendler, J., Rabe, J., Büttgenbach, S., Bilitewski, U., 2005. Sens. Actuat.
B. 105, 508–515.
Ogi, H., Motoshisa, K., Matsumoto, T.,Hatanaka, K., Hirao, M., 2006. Anal. Chem. 78,
19, 383 6903–6909.
Okahata, Y., Niikura, K., Furusawa, H., Hisao, M., 2000. Anal. Sci. 11, 1113–1119.
Okahata, Y., Mori, T., Furusawa, H., Nihira, T., 2007. In: Steinem, C., Janshoff, A. (Eds.),
Piezoelectric Sensors, 1st ed. Springer, Berlin, Heidelberg, pp. 341–369.
Sauerbrey, G., 1959. Z. Phys. 155, 206–222.
Sota, H., Yoshimine, H., Whittier, R.F., Gotoh, M., Shiohara, Y., Hasegawa, Y., Okahata,
Y., 2002. Anal. Chem. 74 15, 3592–3598.
Steinem, C., Janshoff, A. (Eds.), 2007. Piezoelectric Sensors, 1st ed. Springer, Berlin,
Heidelberg.
Uttenthaler, E., Schräml, M., Mandel, J., Drost, S., 2001. Biosens. Bioelectron. 12,
734–735.